Mobile Radiography Course

RAD 13041304

Instructor: TAGELDIN ABUGROON

Page 2: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Mobile Image intensifier (C arm)

Page 3: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Mobile Image intensifier Units (C arm Unit)Mobile image intensifier are designed primarily for use in

any case needs screening image or immediate X-ray

Visualization especially in :

(A) The operation room with:

1.Orthopedic patients (for reduction & fixation of the fractured

femoral neck)

2. Urinary cases (e.g. remove calculi).

3. Biliary system cases (e.g. removal gall stones).

(B) The intensive care departments for :

1. Introduction of cardiac catheters.

2. Implantation of cardiac pace-maker.

Page 4: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Central venous Access Catheter

Page 5: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 6: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 7: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Peripherally Inserted Central Catheters (PICC Lines)Peripherally Inserted Central Catheters (PICC Lines)

Page 8: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Placement of the Catheter

Page 9: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Hickman Catheter

Page 10: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 11: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Advantages: 1.Immediate localization & manipulation of fractures, calculi &

catheters under direct vision.

2. Fewer radiographs are needed to provide

the required information.( Low dose to operator & patient).

3. Reduced operation time can be achieved, which

result in;

(A) Less Anesthesia.

(B) More efficient use of theatre operating time.

Page 12: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Disadvantages: Very close control must be kept on the length of

fluoroscopy which should be kept as minimum as you can to

avoid accumulating dose.

The method used to limit the exposure time include:

1.The use of fluoroscopic time display & alarm.

2. The used of pulsed fluoroscopy.

Page 13: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

1

2

3

4

5

6

7

8

Page 14: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 15: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

components of mobile image intensifier (C-arm) 1.X-ray tube with it’s collimator.

2 .Image intensifier & TV camera .

3 .C-arm showing rotation possible .

4 .C- arm support allowing rotation .

5 .Vertical column showing movement .

6 .Transverse movement .

7 .Control unit & HT generator .

8 .Base extending forward to give stability .

9 .Castors for easy movement .

Page 16: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

components of mobile image intensifier (C-arm)

1. The tube head: (X-ray tube & collimator)

- Is mounted on the end of the of the “C” shaped arm.

The X-ray tube has a dual focus ,commonly:

i. 0.6mm for fluoroscopy.

ii. 1.2mm for radiography.

- It has two collimators:

1. One to restrict the beam size to the intensifier

2. Additional one to collimate X-ray field size.

Page 17: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

2. Image intensifier & TV camera: They are mounted on the opposite end of the “C” arm. The image intensifier may be 5 inch input phosphor.

3. The “C” shaped supported arm:

Is mounted on a vertical column to provide:

a. Considerable vertical movement.

b. Rotation about the fixing to allow .

i. a vertical beam.

Or ii. Horizontal beam.

Or iii. Angles between.

Page 18: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 19: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 20: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 21: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

4. The vertical column:Is mounted on the mobile base.The mobile base carries:1. The control unit (fluoroscopy factors are normallyautomatically control) The high tension generator ( behind thevertical column).

2. H.T generator making a counterbalance for “C” arm.All movements of the column must have secure mechanicallockers.The wheels on the base must be large to allow easymovement.

Page 22: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 23: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

5. TV monitor trolley:

A separate trolley carrying the TV monitor.

6. Power supply: 13Amp. 240 volt main.

For fluoroscopy a maximum of 3mA is provided.

For radiography 30-40 mA with a kV range of 50-90 .

Methods of recording the image:1. By video-tape to record the moving image:

2. CD: Compact disc or DVD.

3. By radiography , either by the cassette holder or by sorting imaging during fluoroscopy and transfer it to the X-ray films

Page 24: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 25: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 26: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 27: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Care must be taken in setting up this equipment to ensure that the x-ray tube is as far as possible from the patient to minimize the patient dose especially patient skin.

If the focus-skin distance is short the dose received by the patient will be high & the image quality will be poor which occur because of the high object-phosphor distance & enlarged image.

Page 28: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Exposure levels-CR Vertical

PA Projection Exposure levels-CR 30 degree from vertical

Page 29: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Exposure patterns & levels-CR horizontal

Exposure levels AP projection )Tube on Top (not recommended

Page 30: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Work with C arm Equipment in the Operating Room Undertaking radiography in conditions other than

the controller ones of the x-ray department has psychological effect on the radiographer especially in the operating theatre.

This work has an emergency nature, & the time factor is important. The aim usually in surgical procedures is to be as quick as is consistent with efficient, the prolonging the length of time during the patient under anesthesia will increase the patient risk, so the radiographer should be calm & fast.

Page 31: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 32: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 33: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 34: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Risks in Operating Theatre

Risk of Importing Infection

Risk of Explosion

Radiation Risk

Page 35: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

1. Risk of importing infection (Asepsis)

How we can reduced this risk? A. cleaning the machine surface before and after usage by

sterilized or antiseptic solution in alcohol base.

B. placing a plastic cover on the places of the unit that may

touch the sterilization area.

2.Risk of explosion This risk may occurred by using ignitable anesthetic

gases during exposure which sometimes produces

sparks. These sparks could be produced from the units wheels or

from the electrical parts of the control panel.

Page 36: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 37: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 38: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 39: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 40: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

How the risk of explosion can be eliminated?

(A). using non- ignitable anesthesia gases.

(B). special conducting wheels, which manufactured by

conducting material to pass small currents from static charges

directly to the theatre ground (earthing).

(C). special spark-proof X-ray units and electrical plugs are

used.

Radiation Risk This risk involves anyone in operating theatre. The radiation

protection is the responsibility of the radiographer before starting any exposure.

Page 41: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

How the Radiation Risk can be minimized in OR?

(1).Wearing lead apron , thyroid collar, glasses , gloves and

TLD, the Radiological technologist and the staff must wearing

this accessories during fluoroscopy and during taking

radiographs.

(2) Using the inverse square law with the staff: How?

i. By standing at maximum distance .

ii. By standing outside from the path of radiation source.

(3) Collimation the radiation beams on the required area of the

patient.

(4) Minimize the screening time as possible as you can.

(5)Giving clear instructions to staff before any

exposure is made.

Page 42: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 43: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Inverse square law

I1 / I2 = (d2)2 / (d1)2

Page 44: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 45: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Lead Apron & Thyroid Shield

Page 46: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Lead Gloves & Goggles

Page 47: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Mobile Barrier Mobile Curtain Face Shield

Page 48: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Collimator

Page 49: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

Special Concerns: In many modern theatre suites it is normally that you find a

special room for the x-ray equipment & for processing facilities.

The first thing that the radiographers must be concerned with, is their own personal preparation entering an aseptic controlled area. The radiographer should change clothes & wearing theatre uniform & wear (shoe cover, hair cover, mask & washing hand).

Before unit is removed from it’s store & after it is placed in the theatre, it should be tested.

This procedure should be done before surgical

procedure.

Page 50: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON
Page 51: Mobile Radiography Course 1304 RAD 1304 Instructor: TAGELDIN ABUGROON

The processing equipment should be switched on & tested. Adequate levels of replenisher solutions & supply of cassettes & films should be checked.

All needed accessory equipment should be available

(e.g cassette holder).

Always the examination request should be checked before starting the preparation.