Calibrated Viscosity Cups Designed, Manufactured & Sold by The Paul N. Gardner Company, Inc.
316 N.E. FIRST ST • POMPANO BEACH, FL 33060P.O. BOX 10688 • POMPANO BEACH, FL 33061-6688PHONE 1-800-762-2478 • (954) 946-9454 • FAX (954) 946-9309WEB SITE: www.gardco.com • E Mail: [email protected]
Mini Ford Dip Viscosity Cup
Instruction Manual
PU-G213 Copywright April 6, 1998 Rev:4 01/07
Parlin Viscosity CupsStandard Ford Viscosity CupsI.S.O. Viscosity CupsDin Viscosity CupFisher Standard CupsStandard Ford Dip Viscosity CupsMini Ford Dip Viscosity CupsMini I.S.O. Dip Viscosity CupsMini Din Dip Viscosity CupFisher Dip Viscosity CupEZ™/ZAHN (ASTM) Dip Viscosity CupsS-90/ZAHN Signature Dip Viscosity Cups
COMPANY INCORPORATEDGARDNERPAUL N.
®
S-90/ZAHN Signature Dip Viscosity Cups
VISCOSITY
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
GARDCO MINI FORD DIP VISCOSITY CUPSFOR VISCOSITY MEASUREMENTS OF VERY THIN PAINTS, INKS, SOLVENTS, LACQUERS, ADHESIVES, and COATING MATERIALS OF ALL KINDS
FEATURES■ CAPACITY IS 50% OF STANDARD FORD VISCOSITY CUP.■ ±4% PRODUCTION TOLERANCE■ EASILY REmOVABLE "SNAP IN, SNAP OUT" ORIFICE CAN SAVE USER THE COST OF FIVE CUPS — BUY ONE CUP AND ANY OF SIX "SNAP-IN" ORIFICES.■ SIX ORIFICES NOw AVAILABLE No. 0, 1, 2, 3, 4 and 5■ COmPATIBLE TO ASTm SPECIFICATION D 1200■ FOR USE IN FIELD OR LABORATORY■ 2 TO 1400 CENTISTOKE RANGE■ THE GARDCO SERIES OF SIX mINI FORD DIP CUPS ARE PRODUCED wITH EASILY REmOVABLE ORIFICES OF THE GARDCO "SNAP-IN, SNAP-OUT" TYPE. CUP AND ORIFICE PRODUCTION TOLERANCE PERmITS REPLACE- mENT OR INTERCHANGE OF ORIFICES wITHOUT LOSS OF ACCEPTABLE CUP ACCURACY.■ SECONDS TO CENTISTOKES CONVERSION TABLE FURNISHED wITH EVERY CUP■ DIRECTLY RELATED TO STANDARD FORD CUP SERIES DESIGNED, STANDARDIZED AND PRODUCED BY THE PAUL N. GARDNER COmPANY, INC.■ CALIBRATION TRACEABLE TO THE U.S. NATIONAL INSTITUTE OF STANDARDS & TECHNOLOGY.■ CUP QUALIFIES UNDER MIL STD 45662A & CONFORMS TO ISO 9000 WHEN ORDERED W/ CALIBRATION CERTIFICATION.■ CUP No. 5 COmPATIBLE TO REQUIREmENTS OF FEDERAL SPECIFICATION mmm-A-121 FOR ADHESIVES.■ STANDARD OILS, TRACEABLE TO THE U.S. NATIONAL INSTITUTE OF STANDARDS & TECHNOLOGY ARE NOw AVAILABLE FOR CONFIRmING THE CALIBRATION OF THE CUPS.
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY IS EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT. This information is furnished upon the condition that the person receiving it shall make his own tests to determine the suitability thereof for his particular purpose.
CAUTION: Silicone fl uids should not be used to calibrate viscosity cups. These materials change the interface between the cup surface and the test material and therefore change the cup calibration. The following is taken from ASTM D 445. Viscometers used for silicone fl uids should be reserved for the exclusive use of such fl uids. Solvent washings from these viscometers should not be used for cleaning other viscometers.
A mathematical formula developed for each Mini Ford Dip Cup in the series, relating time in seconds to centistoke viscosity defi nes cup fl ow characteristics with respect to standard oils traceable to U.S. National Institute of Standards and Technology. The resulting formulas have been solved for each tenth second within the recommended cup range. A table with this information as well as the cup formula is furnished free with each cup.
1
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
VISCOSITYCUP DIMENSIONS
Time in seconds obtained with the dip series is just one half that obtained with the standard series. Simply multiply the seconds obtained by two and the results are directly converted to centistokes by the same theory presented for the standard series. This relationship is obtained by producing the dip series with an internal vertical cross section equal to just one half of the standard series and adjusting the cup height to obtain desired results. The orifice series used in the dip cup is identical to the series used in the standard cup. The design of the bottom of the cup, as can be seen from the drawing, causes all material draining from the outside of the cup to flow away from the orifice and not to influence the rate of flow from the orifice. Minimum time for the dip series is one half of that stated earlier for the standard series, or ten seconds, and cup sensitivity is one-half of that shown earlier for the Gardco Standard Ford Cup. There are six cups in the series covering seconds and centistoke range as follows:
FIG 1
FORD DIP CONVERSION CHART CONVERSION CHART AVERAGE CUP NO. SECONDS RANGE CENTISTOKES VISCOSITY CENTISTOKES RANGE PER SECOND 0 28 to 60 2 to 14 0.4 1 30 to 60 12 to 40 1.0 2 18 to 50 25 to 120 2.8 3 12 to 60 29 to 273 5.0 4 10 to 60 53 to 441 7.6 5 10 to 60 215 to 1413 25.0
CUP EFFLUX TIME GRAPHS & FORMULAS
GARDCO MINI FORD DIP CUP #0Centistokes, V= 0.266T - 157 ÷ TWhere T is in tenths of Seconds orSeconds, T= (V + V2 + 167) ÷ 0.532 Where V is in Centistokes
#0 MINI FORD DIP CUPS
0 5 10 15 20 25 30 35 40 45 50 55 60
�
VISCOSITY
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
CUP EFFLUX TIME GRAPHS & FORMULAS (Continued)
#1 MINI FORD DIP CUPS
#3 MINI FORD DIP CUPS
GARDCO MINI FORD DIP CUP #1Centistokes, V= 0.744T - 300 ÷ TWhere T is in tenths of Seconds orSeconds, T= (V + V2 + 893) ÷ 1.488 Where V is in Centistokes
#2 MINI FORD DIP CUPSGARDCO MINI FORD DIP CUP #2Centistokes, V= 2.48T - 385 ÷ TWhere T is in tenths of Seconds or
Seconds, T= (V + V2 + 3819) ÷ 4.96Where V is in Centistokes
GARDCO MINI FORD DIP CUP #3Centistokes, V=4.62T - 275 ÷ TWhere T is in tenths of Seconds or
Seconds, T= (V + V2 + 5082) ÷ 9.24 Where V is in Centistokes
#4 MINI FORD DIP CUPSGARDCO MINI FORD DIP CUP #4Centistokes, V= 7.40T - 200 ÷ TWhere T is in tenths of Seconds or
Seconds, T= (V + V2 + 5920) ÷ 14.80 Where V is in Centistokes
0 5 10 15 20 25 30 35 40 45 50 55 60
0 5 10 15 20 25 30 35 40 45 50 55 60
0 5 10 15 20 25 30 35 40 45 50 55 60
0 5 10 15 20 25 30 35 40 45 50 55 60
�
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
VISCOSITYCUP EFFLUX TIME GRAPHS & FORMULAS (continued)
GARDCO MINI FORD DIP CUP #5Centistokes, V= 23.60T - 204 ÷ TWhere T is in tenths of Seconds or
Seconds, T= (V + V2 + 19258) ÷ 47.20 Where V is in Centistokes
#5 MINI FORD DIP CUPS
0 5 10 15 20 25 30 35 40 45 50 55 60
COPYRIGHTED CONVERSION TABLE
An 8-1/2 x 11 copyrighted table is furnished with each cup. It is printed in reduced form at left. On the reverse side are instructions for cup use and conversion formulas between efflux time in seconds and centistokes.
USE THE TABLE AS FOLLOWS:Assume an efflux time of 51.6 seconds. Read down the left column to the 51.0 line and then to the right on this line to the 0.6 column. The value at the intersection is 32.6 centistokes. The table may be read in reverse to find efflux time in seconds from a known centistoke value.
ORIFICE INTERCHANGE
Orifice interchange is not applicable to certified cups. Production tolerance on the standard cup, the dip cups and the orifice whether it be #0, #1, #2, #3, #4 or #5, is sufficiently close to permit interchange of orifices in either cup or between cups. For convenience and for best pos-
Figure 2 at right is a cross section of the orifice used in both the Gardco Standard and the Gardco Ford Dip Cups. Note that the upper interior portion following the conversion from the slope of the cup interior is spherical in shape rather than conical. It has been found that this shape provides a “cleaner” break of the stream as the cup empties. In addition, the length of the orifice is better defined in accordance with theory. The “0” orifice does not have this feature as it is longer and has less need of this shape.In the new design the exterior of the brass orifice is a truncated cone, sized to match a conical section in the body of the cup. This design insures close fit with no gaps between the body and the orifice either at the top or bottom. This design also eliminates the hard to clean areas that exist in most other designs.
FIG. 2
�
VISCOSITY
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
FORD DIP CUP ORIFICE DETAIL
*
IDENTIFICATIONTOOL
Long Orifice
DoubleShoulder
SingleShoulder
ID 0.10"Under
SingleShoulder
*Punch Mark
SingleShoulder
ID 0.150"OverNo
Shoulder
IDENTIFICATION TOOL, ORIFICE NUMBER, CROSS SECTION AND END VIEW
VISCOSITY STANDARDS RECOMMENDED FOR CALIBRATION OF MINI FORD DIP VISCOSITY CUPS
Catalog No. Calibration For Use PRICE Liquid With per pint Description Cup No.
VI-3801 G-6 0 VI-3803 G-10 1 VI-3810 G-35 2 VI-3815 G-60 3 VI-3815 G-60 4VI-3821 G-200 5
Adjust the temperature of the material to be measured if necessary. Lower the cup into the material so that the top rim is submerged. Place a thermometer into the cup as it is immersed and determine the temperature of the confined sample. Remove thermometer. Quickly raise the cup from the material, high enough to easily observe the efflux stream and at the same time start the timer. Closely observe the efflux stream and stop the timer at the first break in the stream, one to two inches below the orifice.Record the cup used ( ex. Gardco Mini Dip #4), the measured temperature and the efflux time. If the efflux time is immediately converted to Gardco Standard Cup time by multiplying by two, it should be so noted.Promptly clean the cup with close attention to the orifice.
USE OF GARDCO MINI DIP SERIES CUPSThe GARDCO MINI DIP CUPS should have the same care as the Gardco Standard Series as they are produced to the same close tolerance. The use of the dip cup, however, is somewhat different.
These standard oils prepared expressly by the Cannon Instrument Company for the Paul N. Gardner Company are produced in accordance with ISO 9002:1987; EN 29002:1987 BS 5750 Part2:1987; ANSI/ASQC Q92:1987.
�
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
VISCOSITYPRICES
PRICE OF EACH CUP INCLUDES VISCOSITY CUP CONVERSION TABLE, INSTRUCTION BOOKLET AND wOOD
DOwEL FOR USE IN ORIFICE REmOVAL AND INSERTION.
Cat. No. Description PriceVI-3320 Gardco #0 Mini Ford Dip Viscosity CupVI-3321 Gardco #1 Mini Ford Dip Viscosity CupVI-3322 Gardco #2 Mini Ford Dip Viscosity CupVI-3323 Gardco #3 Mini Ford Dip Viscosity CupVI-3324 Gardco #4 Mini Ford Dip Viscosity CupVI-3325 Gardco #5 Mini Ford Dip Viscosity CupVI-3329 MIL STD 45662 Calibration Certifi cation Single Cup- Orifi ce Combination (Conforms to ISO 9000)TH-0482 Thermometer, blue spirit fi lled, 20°- 30°C TM-14-1/2 Stopwatch, 7-jewel, 1/10th SecondTM-705 Stopwatch, Electronic, Digital, #705, 1/100th Sec.TM-705/C Certifi ed Stopwatch, Electronic, Digital, 1/100th Sec. VI-FGS Standard Ford Cup Conversion Grid Set, for Cups #0-5VI-FMDGS Mini Ford Dip Cup Conversion Grid Set, for Cups #0-5
Replacement Orifi ces These orifi ces will fi t both our Standard Series and Dip Series Ford Cups.
VI-3310 Replacement Orifi ce #0 (Snap-in Type)VI-3311 Replacement Orifi ce #1 (Snap-in Type)VI-3312 Replacement Orifi ce #2 (Snap-in Type)VI-3313 Replacement Orifi ce #3 (Snap-in Type)VI-3314 Replacement Orifi ce #4 (Snap-in Type)VI-3315 Replacement Orifi ce #5 (Snap-in Type)
Orifi ce Identifi cation ToolVI-3341 Orifi ce Identifi cation Tool
Gardco Viscosity Cup Equivalent ChartVI-VCC GARDCO VISCOSITY CUP EQUIVALENT CHART
ASTM Mercury In Glass ThermometersTH-02232 Celsius Thermometer, (19-27°C)TH-02233 Fahrenheit Thermometer, (66-80°F)TH-02272 Celsius Thermometer, (23.6-26.4°C)TH-02273 Fahrenheit Thermometer, (74.5-79.5°F)
6
�
VISCOSITY
1-800-762-2478 • (954) 946-9454 • FAX (954) 946-9309316 NE 1st STREET , POMPANO BCH, FL 33060
G SERIES VISCOSITY STANDARDS
Paul N. Gardner Company, Inc.
These standard oils prepared expressly by the Cannon Instrument Company for the Paul N. Gardner Company are produced in accor-dance with: ISO 9002: 1994 EN ISO 9002: 1994 BS EN ISO 9002: 1994 ANSI/ASQC Q9002: 1994 These standards provide the greatest possible accuracy in viscosity measurements with Gard-co produced and calibrated viscosity cups. A list of “G” Series viscosity standards is available for periodic checking of these measuring instru-ments. See list on following page.
IMPORTANT LIMITATIONSUse of standard oils available from the Paul N. Gardner Company are NOT recom-mended for use with cups produced by other suppliers unless the conversion formula between centistoke viscosity and efflux time as well as cup production tolerance for the particular cup in question has been furnished.
USE OF STANDARD OILSRecommended use of a standard oil to check the calibration of a Gardco produced viscosity cup is as follows: Conversion formulas are either provided with the cup to be checked or are available in the Product Leaflet for the cup.
FOR USE WITHGARDCO VISCOSITY CUPS
PACKAGING - LABELINGAll viscosity standards are packaged in sealed, dark glass bottles. Net content of each bottle is 0.47 Liter (1 Pint). A label on each bottle gives the Standard Name, Lot No., Termination Date, Viscosity in Centistokes at 25.00°C (77.00°F)* and a statement that the standard “is prepared expressly for Paul N. Gardner Co., Inc., P.O. Box 10688, Pompano Beach, FL 33061-6688” by the Cannon Instrument Company and that data on the standard is traceable to the National Institute of Standards and Technology. A Material Safety Data Sheet (M.S.D.S.) accompanies each bottle of oil.
VISCOSITY
�
*The
se a
re n
omin
al v
alue
s. F
or th
e ac
tual
cen
tisto
ke v
alue
of t
he s
tand
ard,
refe
r to
the
valu
e on
the
bottl
e la
bel.
*
*NO
TE:
The
8mm
ISO
Cup
is n
ot a
vaila
ble
in th
e D
ip v
ersi
on.
G
-6
8.9
- -
0 -
- 0
- -
-
G-1
0 19
1
- 1
- -
1 3
- -
G
-20
35
- 1
- 1
- -
- -
-
G-3
5 68
-
-
2 2
2 2
4 -
1
G-6
0
1
20
2 2
3, 4
3
-
3, 4
-
- -
G
-100
270
- 3
- -
- -
6 4
2
G-2
00
45
0
3,
4
- 5
4 -
5 -
- -
G
-350
880
5
4, 5
-
-
-
-
*
* 8
-
-
G-3
50
102
0 -
- -
- -
- -
- 3
G-J
3000
10
,800
-
- -
- -
- -
- 4
Stan
dard
Cen
tisto
ke*
EZ™
S90
/
Min
i
Gar
dco/
F
ishe
r
Std
. For
d
Gar
dco/
ISO
Gar
dco/
Din
Par
lin
O
il V
isco
sity
C
up
Zahn
Fo
rd D
ip F
ishe
r Dip
Sta
ndar
d &
Std
. For
d &
ISO
Dip
&
Din
Dip
Cu
p
No
Cup
No
Cup
No
Cu
p N
o
Cup
No
Dip
Cup
No
Cup
mm
C
up m
m
No
G S
ERIE
S V
ISC
OSI
TY S
TAN
DA
RDS
The
info
rmat
ion
cont
aine
d he
rein
, or
sup
plie
d by
us
or o
n ou
r be
half
in a
ny o
ther
man
ner
is b
ased
on
data
obt
aine
d by
our
ow
n re
sear
ch a
nd is
co
nsid
ered
acc
urat
e. H
owev
er, N
O W
ARRA
NTY
IS E
XPRE
SSED
OR
IMPL
IED
REG
ARD
ING
TH
E AC
CURA
CY O
F TH
ESE
DAT
A, T
HE
RESU
LTS
TO B
E O
B -TA
INED
FRO
M T
HE
USE
THER
EOF,
OR
THAT
AN
Y SU
CH U
SE W
ILL
NO
T IN
FRIN
GE
ANY
PATE
NT.
Thi
s in
form
atio
n is
furn
ishe
d up
on th
e co
nditi
on th
at
the
pers
on re
ceiv
ing
it sh
all m
ake
his
own
test
s to
det
erm
ine
the
suita
bilit
y th
ereo
f for
his
par
ticul
ar p
urpo
se.
�
VISCOSITYEFFLUX TImE TO THE NEAREST 0.1 SECOND CAN BE READ FROm THE CONVERSION CHART FURNISHED wITH EACH GARDCO PRODUCED CUP, OR A FORmULA CAN BE USED AS FOLLOwS:
make a minimum of three determinations with the cup in question on the applicable standard, adhering to the method of use prescribed in the product leaflet or in the applicable ASTm method, and calculate the average efflux time.
1.
2.
3.
4.
NOTE: This determined efflux time is for the standard at 25.00°C (77.00°F) only*.
Compare the experimentally determined efflux time with the design efflux time. If experi mental results were at 25.0°C or 77.0°F*, the comparison should be within the following limits:
NOTE: Pay particular attention to temperature at time of measurement. Use a thermometer that can be read to the nearest 0.1°C or 0.2°F. Experimental results should NOT be compared against determined design efflux time unless tempera-ture at time of measurement was within ± 0.2° of 25.0°C or ± 0.4° of 77.0°F*. The standard oils change in viscosity between 3% and 8%, depending on the viscosity of the standard, for each degree change in temperature from 25°C*.
Select the formula relating viscosity to time of efflux in seconds when viscosity in centistokes is known. Substitute the standard label value for centistokes for “V” in the equation and solve the equation. The resulting value is design efflux (drain) time in seconds.
MANUAL TEMPERATURE ADJUSTMENT
For the technician who does not have automatic temperature control equipment the following procedure may be useful with dip type cups: If the oil to be used is below temperature, place the bottle with the cap slightly loosened no closer than about one foot from an incandescent desk lamp. This will slowly raise the temperature of the oil. When within about one degree of measuring temperature, transfer the required amount of the oil to a glass container of sufficient height and diameter to accom-modate the viscosity cup. (Glass is recommended due to its low rate of temperature conductance.) Place the viscosity cup to be used in the oil of the glass container so that it comes to the same temperature as the oil and place a thermometer, preferably glass, in the oil which can also be used as a stirring rod. If necessary to elevate the temperature slightly, simply hold the glass snugly in the hand while stirring the contents gently and observing the change in temperature. If necessary to lower the temperature, dip the bottom of the glass container for a few seconds in water that is 10 to 20 degrees lower than measuring temperature and continue gentle stirring while the temperature slowly lowers. With a little practice it is possible to make measurements very close to the target temperature.
Gardco EZˇ™Zahn (ASTM)Dip Cups 3%Gardco S90/Zahn Sig. Dip Cups 5%Gardco Mini Ford Dip Cups 4%Gardco/Fisher Dip Cups 2%Gardco/Fisher Standard Cup 2% Gardco Standard Ford Cups 2%Gardco Standard Ford Dip #3 3%Gardco Standard Ford Dip #4 2%Gardco Standard Ford Dip #5 3%
Gardco/ISO Cups 2%Gardco/ISO 8mm Cup 3%Gardco/ISO Dip Cup 3%Gardco/DIN Cup 3%Gardco/DIN Dip Cup 3%Gardco Parlin #1 Cup 3%Gardco Parlin #2 Cup 3%Gardco Parlin #3 Cup 3%Gardco Parlin #4 Cup 5%
* With the exception of the Parlin Cups whose Viscosity in Centistokes is measured at 23°C.
VISCOSITY
10
1-800-762-2478 • (954) 946-9454 • FAX (954) 946-9309
Paul N. Gardner Company, Inc.316 NE 1st STREET , POMPANO BCH, FL 33060
Approx.Cat. No. Description CentistokesVI-3801 G-6 Viscosity Standard 9VI-3803 G-10 Viscosity Standard 19 VI-3805 G-20 Viscosity Standard 35 VI-3810 G-35 Viscosity Standard 68VI-3815 G-60 Viscosity Standard 120VI-3820 G-100 Viscosity Standard 230VI-3821 G-200 Viscosity Standard 450VI-3825 G-350 Viscosity Standard 880VI-3826 G-350 Viscosity Standard 1020VI-3842 G-J3000 Viscosity Standard 10800
CAUTION: Silicone fluids should not be used to calibrate viscosity cups. These materials change the interface between the cup surface and the test material and therefore change the cup calibration. The following is taken from ASTM D 445; Viscometers used for silicone fluids should be reserved for the exclusive use of such fluids. Solvent washings from these viscometers should not be used for cleaning other viscometers.
A Material Safety Data Sheet is furnished with each bottle of "G" Series Standard Oils.
CARE OF STANDARD OILSThe viscosity standards available from Gardco are precision materials. They are accurate to within 0.25% of the label value at specified temperature. This is a much closer tolerance than viscosity measuring devices normally available to users of viscosity cups. Therefore, there is no easy means of checking the standard to insure that it remains at the value listed on the label except through very careful use of the standard to insure its integrity. Included in this care should be making sure that any container used to receive the standard, as well as the cup to be checked, the thermometer and any other stirring device, be absolutely free of contaminants. Immediately following use, the standard should be returned to its labeled container and capped. If there is valid reason to question the original value of the standard, it should be replaced.The below listed standards are used by the producer of Gardco viscosity cups not only to calibrate the cups but also to certify such calibration under ANSI/NCSL Z540-1 or MIL-STD-45662A, as applicable, & in conformance with ISO 9000, which is available as an extra cost service. One of the requirements under such certification is the replacement of the standard on or before the termination date. Experience has shown, however, that in the absence of contamination, the standards do not materially deteriorate over an extended period of time.
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY IS EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT. This information is furnished upon the condition that the person receiving it shall make his own tests to determine the suit-ability thereof for his particular purpose.
VISCOSITY
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
a. One GARDCO Dip-Type Viscosity Cup.b. The Conversion Grid sheet (for cup chosen) to convert seconds effl ux time (to nearest tenth of a second) to centistoke viscosity. Every Gardco Cup is furnished with such a data sheet.c. One pint of certifi ed calibration oil (appropriate to cup chosen, see chart below)
EXAMPLE OF MATERIALS NEEDEDFOR CALIBRATION OF A DIP-TYPE VISCOSITY CUP
d. To insure highest accuracy of calibration, we recommend the use of ASTM mercury fi lled glass type thermometers having an accuracy of 0.1°C (0.2°F). Use any of the following: 1. Gardco Cat. No. VI-3130 glass mercury fi lled thermometer, 4-1/2" length, 20-30°C. range. 2. Gardco Cat. No. TH-17C ASTM mercury fi lled glass thermometer, 10.8" length, 19-27°C. range. 3. Gardco Cat. No. TH-17F ASTM mercury fi lled glass thermometer, 10.8" length, 66-80°F. range.e. Container, 250 ml capacity.f. Gardco Cat. No. TM-705 digital electronic stopwatch, 0.01 second.g. Reference literature: 1. Gardco Viscosity cup leafl et (for cup chosen) 2. Viscosity Cups-Common Questions and Their Answers leafl et 3. Gardco Viscosity Cup Equivalent Wall Chart (VI-VCC) or Slide Chart (VI-9939) 4. Gardco General Catalog
PROCEDURE FOR CALIBRATION FOR DIP-TYPE VISCOSITY CUPS
THIS INFORMATION PERTAINS ONLY TO VISCOSITY CUPS MADE BY THE PAUL N. GARDNER COMPANY, INC. (GARDCO)
G6 8.9 - - 0 - 0 - - G10 19 1 - 1 - 1 3 - G20 34 - 1 - 1 - - - G35 64 - 2 2 2 4 - G60 120 2 2 3, 4 3 3, 4 - - G100 230 - 3 - - - 6 4 G200 450 3, 4 - 5 4 5 - - G350 880 5 4, 5 - - - * * 8 -
Standard Centistoke* EZ™ S90/ Mini Gardco/ Std. Ford Gardco/ISO Gardco/Din Oil Viscosity Cup Zahn Ford Dip Fisher Dip Dip Dip Dip No Cup No Cup No Cup No Cup No Cup mm Cup mm
*These are nominal values. For the actual centistoke value of the standard, refer to the value on the bottle label. * *NOTE: The 8mm ISO Cup is not available in the Dip version.
DIP-TYPE VISCOSITY CUPS
11
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
VISCOSITY
THIS INFORMATION PERTAINS ONLY TO VISCOSITY CUPS MADE BY THE PAUL N. GARDNER COMPANY, INC. (GARDCO)
PROCEDURE1. Examine the cup for obvious obstruction or damage to the orifi ce area and general cleanliness of cup.2. Select the recommended Certifi ed Viscosity Standard for the cup to be calibrated. make sure the expiration date for the standard has not passed.3. Pour a suffi cient amount of oil into a 500 ml container to totally submerge the cup.4. Using a calibrated ASTm mercury-fi lled thermometer bring oil to target temperature, 25.0°C ± .5°C. Proper temperature is very important. By wrapping a hand around the beaker and slowly stirring oil, the temperature can be slightly raised.5. Place cup in oil and allow 1 to 5 minutes for cup and oil to reach thermal equilibrium.6. Raise the cup vertically to a distance of no more than 6” from the surface of the oil. Start timing with a stop watch with accuracy of one tenth second or better as the top edge of the cup breaks the surface of the oil.7. Leave the thermometer in the beaker to insure the temperature remains constant throughout the run.8. Stop the timer when the fi rst defi nite break in the effl ux stream occurs near the bottom of the cup. Be sure to keep air bubbles at a minimum; they will sometimes cause an apparent break pre- maturely.9. Record time and temperature and repeat. Keep temperature drift between runs at ±.2°C. Disregard any runs that are more than .5 seconds apart.10. Adjust times for temperature deviation and average three valid runs.11. Divide the design time by the average and determine the correction factor.12. Determine if the cup is in or out of tolerance according to manufacturer’s specifi cation.
** THIS PROCEDURE mEETS THE REQUIREmENTS OF ASTM D4212 **
1�
VISCOSITY
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
EXAMPLE OF VISCOSITY CUP CALIBRATION CHECK SHEETDATE 7/31/98CUP NAmE GARDCO S-90 Zahn Signature CUP NO. 2 CUP RING TAG SERIAL NO. 44263STATED TOLERANCE ± 3% CERTIFIED OIL DATACALIBRATION OIL NO. G60 LOT NO. 90102LABEL DUE DATE 5/ 30/98DATE OIL FIRST USED ** 7/31/98CENTISTOKE VISCOSITY ON LABEL 118.4 CS CALIBRATION DATA *RUN NO. 1 32.85 SECS @ 25.0°C RUN NO.3 33.55 SECS @ 25.0°CRUN NO. 2 33.09 SECS @ 25.0°C AVERAGE 33.16 SECS @ 25.0°C
DESIGN EFFLUX TImE (FROm GARDCO CONVERSION GRID CHART) 33.72 SECS @ 25.0°C
IN TOLERANCE ( √ ) OUT OF TOLERANCE ( ) CORRECTION FACTOR 1.017CALIBRATED BY A.B. Smith CHECKED BY L.A. Dawn
(Only For Cups Made By GARDCO)
**
When many different determinations are to be made on similar products in the same viscosity range, it may be helpful to produce a graph for converting measured temperature and viscosity cup effl ux time in seconds to seconds at a specifi ed temperature, normally 25.0° C (77.0°F). Refer to "Temperature" in Viscosity Cup leafl et for a suggested method of preparation for such a graph.
*
When fi nished with standard oil examine it for contamination of any kind. A fi ne wire mesh fi lter may be used if dirt specks are noticed. If oil is accidently contaminated with water settled in bottom of tumbler, pour off clean portion of oil and discard contaminated water portion. Pour oil back into bottle and record the date of fi rst use on the label. Stored in normal ambient temperature the oil is good for a year after date of fi rst use, provided it does not become contaminated.
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY IS EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT. This information is furnished upon the condition that the person receiving it shall make his own tests to determine the suitability thereof for his particular purpose.
CLEANINGAfter use of cup, allow to drain and then clean with suitable solvent and dry cup with soft lint-free cloth. Be sure that orifi ce is clean and dry. Do not use metal object to clean the orifi ce. Use care not to score or scratch orifi ce. A length of nylon fi shing line is suitable in many cases to clean the orifi ce. Clean thermometer with suitable solvent and dry with lint-free cloth before storage. Clean container in suitable solvent and mark for future “viscosity cup calibration use only”.
A point of caution: Even the standard oils change in viscosity in the range of 3% - 8% per degree Celsius at the 25° normal measuring range. Temperature of products being measured should be adjusted to within 0.1° of specifi ed temperature if accurate results are to be obtained.
CAUTION
1�
Paul N. Gardner Company, Inc.316 N.E. First Street • Pompano Beach, Florida 33060
Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
GARDCO MINI FORD DIP VISCOSITY CUPSCUP NUMBER 0
CONVERSION FORMULAS AND TABLE©
GARDCO MINI FORD DIP viscosity cups are designed to provide just one half the efflux time of the GARDCO STANDARD FORD DIP viscosity cups which comply with requirements of ASTM D-1200. Both cups use the identical orifice. Cup dimensions are carefully controlled and cup calibration conditions comply with MIL STD 45662A. Standard viscous oil traceable to the National Institute of Standards and Technology are used in calibration procedures to insure specified efflux time tolerance.
Use this formula derived by the Paul N. Gardner Company research to find viscosity (V) in centistokes when cup efflux time in seconds (T) is known:
V = 0.266T - 157 ÷ T
Use this formula to find cup efflux seconds (T) when viscosity (V) in centistokes is known:
T = ( V + √ V2 + 167 ) ÷ 0.532
Results from the above formulas, solved for each tenth of a second within the cup range, are shown on the reverse side of this page. To find centistoke viscosity for a given cup efflux time in seconds, read down the column on the left to find the nearest second. Then, read to the right to the nearest tenth of a second column to find centistoke value. The chart may be read in reverse to find efflux time seconds when viscosity is known.
The GARDCO MINI FORD DIP series of six viscosity cups are produced, calibrated and sold only by the Paul N. Gardner Company and licensed agents.
©Copyright 1987, Paul N. Gardner Company.
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT.
GARDCO MINI FORD DIP VISCOSITY CUP NUMBER 0EFFLUX TIME - CENTISTOKES CONVERSION TABLE ©
316 N.E. First Street • Pompano Beach, Florida 33060Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
Paul N. Gardner Company, Inc.
2� 1.�6 1.�1 1.�5 2.00 2.04 2.0� 2.14 2.1� 2.23 2.2�2� 2.32 2.36 2.41 2.45 2.50 2.54 2.5� 2.63 2.6� 2.�230 2.�6 2.�1 2.�5 2.�� 2.�4 2.�� 3.03 3.0� 3.11 3.15
31 3.20 3.24 3.2� 3.33 3.3� 3.41 3.45 3.50 3.54 3.5�32 3.62 3.66 3.�0 3.�5 3.�� 3.�3 3.�� 3.�1 3.�5 3.��33 4.04 4.0� 4.12 4.16 4.20 4.24 4.2� 4.32 4.36 4.4034 4.44 4.4� 4.52 4.56 4.60 4.64 4.6� 4.�2 4.�6 4.�035 4.�4 4.�� 4.�2 4.�6 5.00 5.03 5.0� 5.11 5.15 5.1�
36 5.23 5.2� 5.31 5.34 5.3� 5.42 5.46 5.50 5.54 5.5�3� 5.61 5.65 5.6� 5.�3 5.�6 5.�0 5.�4 5.�� 5.�1 5.�53� 5.�� 6.03 6.06 6.10 6.14 6.1� 6.21 6.25 6.2� 6.323� 6.36 6.40 6.43 6.4� 6.51 6.54 6.5� 6.62 6.65 6.6�40 6.�3 6.�6 6.�0 6.�4 6.�� 6.�1 6.�4 6.�� �.02 �.05
41 �.0� �.12 �.16 �.20 �.23 �.2� �.30 �.34 �.3� �.4142 �.45 �.4� �.52 �.55 �.5� �.62 �.66 �.6� �.�3 �.�643 �.�0 �.�3 �.�� �.�0 �.�4 �.�� �.01 �.04 �.0� �.1144 �.15 �.1� �.22 �.25 �.2� �.32 �.35 �.3� �.42 �.4645 �.4� �.53 �.56 �.60 �.63 �.66 �.�0 �.�3 �.�� �.�0
46 �.�3 �.�� �.�0 �.�4 �.�� �.00 �.04 �.0� �.10 �.144� �.1� �.21 �.24 �.2� �.31 �.34 �.3� �.41 �.44 �.4�4� �.51 �.54 �.5� �.61 �.64 �.6� �.�1 �.�4 �.�� �.�14� �.�4 �.�� �.�1 �.�4 �.�� 10.01 10.04 10.0� 10.10 10.1450 10.1� 10.20 10.24 10.2� 10.30 10.33 10.3� 10.40 10.43 10.46
51 10.50 10.53 10.56 10.60 10.63 10.66 10.6� 10.�3 10.�6 10.��52 10.�2 10.�5 10.�� 10.�2 10.�5 10.�� 11.02 11.05 11.0� 11.1153 11.15 11.1� 11.21 11.24 11.2� 11.31 11.34 11.3� 11.40 11.4354 11.4� 11.50 11.53 11.56 11.5� 11.63 11.66 11.6� 11.�2 11.�555 11.�� 11.�2 11.�5 11.�� 11.�1 11.�4 11.�� 12.01 12.04 12.0�
56 12.10 12.13 12.16 12.20 12.23 12.26 12.2� 12.32 12.35 12.3�5� 12.42 12.45 12.4� 12.51 12.54 12.5� 12.60 12.64 12.6� 12.�05� 12.�3 12.�6 12.�� 12.�2 12.�5 12.�� 12.�2 12.�5 12.�� 13.015� 13.04 13.0� 13.10 13.13 13.1� 13.20 13.23 13.26 13.2� 13.3260 13.35 13.3� 13.41 13.44 13.4� 13.51 13.54 13.5� 13.60 13.63
SECONDS 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.� 0.� 0.�
10/�5 (Accurate for True Liquids Only)
Example: 55.2 Seconds = 11.85 Centistokes. ©Copyright 1��� Paul N. Gardner Co.
< -----------------------------------------VISCOSITY IN CENTISTOKES ------------------------------------------ >
Paul N. Gardner Company, Inc.316 N.E. First Street • Pompano Beach, Florida 33060
Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT.
GARDCO MINI FORD DIP viscosity cups are designed to provide just one half the efflux time of the GARDCO STANDARD FORD DIP viscosity cups which comply with requirements of ASTM D-1200. Both cups use the identical orifice. Cup dimensions are carefully controlled and cup calibration conditions comply with MIL STD 45662A. Standard viscous oil traceable to the National Institute of Standards and Technology are used in calibration procedures to insure specified efflux time tolerance.
Use this formula derived by the Paul N. Gardner Company research to find viscosity (V) in centistokes when cup efflux time in seconds (T) is known:
V= 0.744T - 300 ÷ T
GARDCO MINI FORD DIP VISCOSITY CUPSCUP NUMBER 1
CONVERSION FORMULAS AND TABLE©
Use this formula to find cup efflux seconds (T) when viscosity (V) in centistokes is known:
T= ( V + √ V2 + 893 ) ÷ 1.488
Results from the above formulas, solved for each tenth of a second within the cup range, are shown on the reverse side of this page. To find centistoke viscosity for a given cup efflux time in seconds, read down the column on the left to find the nearest second. Then, read to the right to the nearest tenth of a second column to find centistoke value. The chart may be read in reverse to find efflux time seconds when viscosity is known.
The GARDCO MINI FORD DIP series of six viscosity cups are produced, calibrated and sold only by the Paul N. Gardner Company and licensed agents.
©Copyright 1987, Paul N. Gardner Company
GARDCO MINI FORD DIP VISCOSITY CUP NUMBER 1EFFLUX TIME - CENTISTOKES CONVERSION TABLE ©
10/�5 (Accurate for True Liquids Only)
< -----------------------------------------VISCOSITY IN CENTISTOKES ------------------------------------------ >
SECONDS 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.� 0.� 0.�
30 12.3 12.4 12.5 12.6 12.� 12.� 13.0 13.1 13.2 13.3
31 13.4 13.5 13.6 13.� 13.� 13.� 14.0 14.1 14.2 14.332 14.4 14.5 14.6 14.� 14.� 14.� 15.1 15.2 15.3 15.433 15.5 15.6 15.� 15.� 15.� 16.0 16.1 16.2 16.3 16.434 16.5 16.6 16.� 16.� 16.� 1�.0 1�.1 1�.2 1�.3 1�.4 35 1�.5 1�.6 1�.� 1�.� 1�.� 1�.0 1�.1 1�.2 1�.3 1�.4
36 1�.5 1�.5 1�.6 1�.� 1�.� 1�.� 1�.0 1�.1 1�.2 1�.33� 1�.4 1�.5 1�.6 1�.� 1�.� 1�.� 20.0 20.1 20.2 20.33� 20.4 20.5 20.6 20.6 20.� 20.� 20.� 21.0 21.1 21.23� 21.3 21.4 21.5 21.6 21.� 21.� 21.� 22.0 22.1 22.240 22.3 22.4 22.4 22.5 22.6 22.� 22.� 22.� 23.0 23.1
41 23.2 23.3 23.4 23.4 23.6 23.6 23.� 23.� 23.� 24.042 24.1 24.2 24.3 24.4 24.5 24.6 24.� 24.� 24.� 24.�43 25.0 25.1 25.2 25.3 25.4 25.5 25.6 25.6 25.� 25.�44 25.� 26.0 26.1 26.2 26.3 26.4 26.5 26.5 26.6 26.�45 26.� 26.� 2�.0 2�.1 2�.2 2�.3 2�.3 2�.4 2�.5 2�.6
46 2�.� 2�.� 2�.� 2�.0 2�.1 2�.1 2�.2 2�.3 2�.4 2�.5
4� 2�.6 2�.� 2�.� 2�.� 2�.� 2�.0 2�.1 2�.2 2�.3
Paul N. Gardner Company, Inc.316 N.E. First Street • Pompano Beach, Florida 33060
Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
GARDCO MINI FORD DIP VISCOSITY CUPSCUP NUMBER 2
CONVERSION FORMULAS AND TABLE©
GARDCO MINI FORD DIP viscosity cups are designed to provide just one half the efflux time of the GARDCO STANDARD FORD DIP viscosity cups which comply with requirements of ASTM D-1200. Both cups use the identical orifice. Cup dimensions are carefully controlled and cup calibration conditions comply with MIL STD 45662A. Standard viscous oil traceable to the National Institute of Standards and Technology are used in calibration procedures to insure specified drain time tolerance.
Use this formula derived by the Paul N. Gardner Company research to find viscosity (V) in centistokes when cup drain time in seconds (T) is known:
V = 2.48T - 385 ÷ T
Use this formula to find cup drain seconds (T) when viscosity (V) in centistokes is known:
T = ( V + √ V2 + 3819 ) ÷ 4.96 Results from the above formulas, solved for each tenth second within the cup range, are shown on the reverse side of this page. To find centistoke viscosity for a given cup drain time in seconds, read down the column on the left to find the nearest second. Then, read to the right to the nearest tenth second column to find centistoke value. The chart may be read in reverse to find drain time seconds when viscosity is known.
The GARDCO MINI FORD DIP series of six viscosity cups are produced, calibrated and sold only by the Paul N. Gardner Company and licensed agents.
©Copyright 1987, Paul N. Gardner Company.
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT.
Paul N. Gardner Company, Inc.316 N.E. First Street • Pompano Beach, Florida 33060
Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
GARDCO MINI FORD DIP VISCOSITY CUP NUMBER 2EFFLUX TIME - CENTISTOKES CONVERSION TABLE ©
Paul N. Gardner Company, Inc.316 N.E. First Street • Pompano Beach, Florida 33060
Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
< -----------------------------------------VISCOSITY IN CENTISTOKES ------------------------------------------ >
SECONDS 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.� 0.� 0.�
10/�5 (Accurate for True Liquids Only)
1� 23.3 23.6 24.0 24.3 24.� 25.1 25.4 25.� 26.1 26.51� 26.� 2�.2 2�.6 2�.� 2�.3 2�.6 2�.0 2�.3 2�.� 30.020 30.4 30.� 31.0 31.4 31.� 32.1 32.4 32.� 33.1 33.4
21 33.� 34.1 34.4 34.� 35.1 35.4 35.� 36.1 36.4 36.�22 3�.1 3�.4 3�.� 3�.0 3�.4 3�.� 3�.0 3�.3 3�.� 40.023 40.3 40.6 40.� 41.3 41.6 41.� 42.2 42.5 42.� 43.224 43.5 43.� 44.1 44.4 44.� 45.0 45.4 45.� 46.0 46.325 46.6 46.� 4�.2 4�.5 4�.� 4�.1 4�.4 4�.� 4�.1 4�.4
26 4�.� 50.0 50.3 50.6 50.� 51.2 51.5 51.� 52.1 52.42� 52.� 53.0 53.3 53.6 53.� 54.2 54.5 54.� 55.1 55.42� 55.� 56.0 56.3 56.6 56.� 5�.2 5�.5 5�.� 5�.1 5�.42� 5�.6 5�.� 5�.2 5�.5 5�.� 60.1 60.4 60.� 61.0 61.330 61.6 61.� 62.1 62.4 62.� 63.0 63.3 63.6 63.� 64.2
31 64.5 64.� 65.0 65.3 65.6 65.� 66.2 66.5 66.� 6�.032 6�.3 6�.6 6�.� 6�.2 6�.5 6�.� 6�.0 6�.3 6�.6 6�.�33 �0.2 �0.5 �0.� �1.0 �1.3 �1.6 �1.� �2.2 �2.4 �2.�34 �3.0 �3.3 �3.6 �3.� �4.1 �4.4 �4.� �5.0 �5.2 �5.535 �5.� �6.1 �6.4 �6.6 �6.� ��.2 ��.5 ��.� ��.0 ��.3
36 ��.6 ��.� ��.1 ��.4 ��.� �0.0 �0.2 �0.5 �0.� �1.13� �1.4 �1.6 �1.� �2.2 �2.5 �2.� �3.0 �3.3 �3.6 �3.�3� �4.1 �4.4 �4.� �4.� �5.2 �5.5 �5.� �6.0 �6.3 �6.63� �6.� ��.1 ��.4 ��.� ��.� ��.2 ��.5 ��.� ��.0 ��.340 ��.6 ��.� �0.1 �0.4 �0.� �0.� �1.2 �1.5 �1.� �2.0
41 �2.3 �2.6 �2.� �3.1 �3.4 �3.6 �3.� �4.2 �4.5 �4.�42 �5.0 �5.3 �5.5 �5.� �6.1 �6.3 �6.6 �6.� ��.1 ��.443 ��.� ��.0 ��.2 ��.5 ��.� ��.0 ��.3 ��.6 ��.� 100.144 100.4 100.6 100.� 101.2 101.4 101.� 102.0 102.2 102.5 102.�45 103.0 103.3 103.6 103.� 104.1 104.4 104.6 104.� 105.2 105.4
46 105.� 106.0 106.2 106.5 106.� 10�.0 10�.3 10�.6 10�.� 10�.14� 10�.4 10�.6 10�.� 10�.2 10�.4 10�.� 110.0 110.2 110.5 110.�4� 111.0 111.3 111.5 111.� 112.1 112.3 112.6 112.� 113.1 113.44� 113.� 113.� 114.2 114.5 114.� 115.0 115.2 115.5 115.� 116.050 116.3 116.6 116.� 11�.1 11�.4 11�.6 11�.� 11�.1 11�.4 11�.�
51 11�.� 11�.2 11�.5 11�.� 120.0 120.2 120.5 120.� 121.0 121.3
Example: 35.2 Seconds = 76.4 Centistokes. ©Copyright 1��� Paul N. Gardner Co.
Results from the above formulas, solved for each tenth of a second within the cup range, are shown on the reverse side of this page. To find centistoke viscosity for a given cup efflux time in seconds, read down the column on the left to find the nearest second. Then, read to the right to the nearest tenth of a second column to find centistoke value. The chart may be read in reverse to find efflux time seconds when viscosity is known.
The GARDCO MINI FORD DIP series of six viscosity cups are produced, calibrated and sold only by the Paul N. Gardner Company and licensed agents.
©Copyright 1987, Paul N. Gardner Company
GARDCO MINI FORD DIP viscosity cups are designed to provide just one half the efflux time of the GARDCO STANDARD FORD DIP viscosity cups which comply with requirements of ASTM D-1200. Both cups use the identical orifice. Cup dimensions are carefully controlled and cup calibration conditions comply with MIL STD 45662A. Standard viscous oil traceable to the National Institute of Standards and Technology are used in calibration procedures to insure specified efflux time tolerance.
Use this formula derived by the Paul N. Gardner Company research to find viscosity (V) in centistokes when cup efflux time in seconds (T) is known:
V = 4.62T - 275 ÷ T
Use this formula to find cup efflux seconds (T) when viscosity (V) in centistokes is known:
T = ( V + √ V2 + 5082 ) ÷ 9.24
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT.
Paul N. Gardner Company, Inc.316 N.E. First Street • Pompano Beach, Florida 33060
Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
GARDCO MINI FORD DIP VISCOSITY CUPSCUP NUMBER 3
CONVERSION FORMULAS AND TABLE©
GARDCO MINI FORD DIP VISCOSITY CUP NUMBER 3EFFLUX TIME - CENTISTOKES CONVERSION TABLE ©
< -----------------------------------------VISCOSITY IN CENTISTOKES ------------------------------------------ >
SECONDS 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.� 0.� 0.�
10/�5 (Accurate for True Liquids Only)
12 32.5 33.2 33.� 34.5 35.1 35.� 36.4 3�.0 3�.� 3�.313 3�.� 3�.5 40.2 40.� 41.4 42.0 42.6 43.2 43.� 44.414 45.0 45.6 46.2 46.� 4�.4 4�.0 4�.6 4�.2 4�.� 50.415 51.0 51.6 52.1 52.� 53.3 53.� 54.4 55.0 55.6 56.2
16 56.� 5�.3 5�.� 5�.4 5�.0 5�.6 60.1 60.� 61.2 61.�1� 62.4 62.� 63.5 64.0 64.6 65.1 65.� 66.2 66.� 6�.31� 6�.� 6�.4 6�.0 6�.5 �0.1 �0.6 �1.1 �1.� �2.2 �2.�1� �3.3 �3.� �4.4 �4.� �5.5 �6.0 �6.5 ��.1 ��.6 ��.120 ��.� ��.2 ��.� �0.2 �0.� �1.3 �1.� �2.3 �2.� �3.4
21 �3.� �4.4 �5.0 �5.5 �6.0 �6.5 ��.1 ��.6 ��.1 ��.622 ��.1 ��.� �0.2 �0.� �1.2 �1.� �2.2 �2.� �3.3 �3.�23 �4.3 �4.� �5.3 �5.� �6.4 �6.� ��.4 ��.� ��.4 ��.�24 ��.4 ��.� 100.4 100.� 101.5 102.0 102.5 103.0 103.5 104.025 104.5 105.0 105.5 106.0 106.5 10�.0 10�.5 10�.0 10�.5 10�.0
26 10�.5 110.0 110.5 111.0 111.6 112.1 112.6 113.1 113.6 114.12� 114.6 115.1 115.6 116.1 116.6 11�.1 11�.5 11�.0 11�.5 11�.02� 11�.5 120.0 120.5 121.0 121.5 122.0 122.5 123.0 123.5 124.02� 124.5 125.0 125.5 126.0 126.5 12�.0 12�.5 12�.0 12�.4 12�.�30 12�.4 12�.� 130.4 130.� 131.4 131.� 132.4 132.� 133.4 133.�
31 134.3 134.� 135.3 135.� 136.3 136.� 13�.3 13�.� 13�.3 13�.�32 13�.2 13�.� 140.2 140.� 141.2 141.� 142.2 142.� 143.2 143.633 144.1 144.6 145.1 145.6 146.1 146.6 14�.0 14�.5 14�.0 14�.534 14�.0 14�.5 150.0 150.4 150.� 151.4 151.� 152.4 152.� 153.435 153.� 154.3 154.� 155.3 155.� 156.3 156.� 15�.2 15�.� 15�.2
36 15�.� 15�.2 15�.6 160.1 160.6 161.1 161.6 162.1 162.5 163.03� 163.5 164.0 164.5 165.0 165.4 165.� 166.4 166.� 16�.4 16�.�3� 16�.3 16�.� 16�.3 16�.� 1�0.2 1�0.� 1�1.2 1�1.� 1�2.2 1�2.63� 1�3.1 1�3.6 1�4.1 1�4.6 1�5.0 1�5.5 1�6.0 1�6.5 1��.0 1��.440 1��.� 1��.4 1��.� 1��.4 1��.� 1�0.3 1�0.� 1�1.3 1�1.� 1�2.2
41 1�2.� 1�3.2 1�3.� 1�4.1 1�4.6 1�5.1 1�5.6 1�6.1 1�6.5 1��.042 1��.5 1��.0 1��.4 1��.� 1��.4 1��.� 1�0.4 1�0.� 1�1.3 1�1.�43 1�2.3 1�2.� 1�3.2 1�3.� 1�4.2 1�4.6 1�5.1 1�5.6 1�6.1 1�6.644 1��.0 1��.5 1��.0 1��.5 1��.� 1��.4 1��.� 200.4 200.� 201.345 201.� 202.3 202.� 203.2 203.� 204.2 204.6 205.1 205.6 206.1
46 206.5 20�.0 20�.5 20�.0 20�.4 20�.� 20�.4 20�.� 210.3 210.�4� 211.3 211.� 212.2 212.� 213.2 213.� 214.1 214.6 215.1 215.64� 216.0 216.5 21�.0 21�.5 21�.� 21�.4 21�.� 21�.3 21�.� 220.34� 220.� 221.2 221.� 222.2 222.� 223.1 223.6 224.1 224.6 225.050 225.5 226.0 226.4 226.� 22�.4 22�.� 22�.3 22�.� 22�.3 22�.�
51 230.2 230.� 231.2 231.6 232.1 232.6 233.1 233.5 234.0 234.552 235.0 235.4 235.� 236.4 236.� 23�.3 23�.� 23�.3 23�.� 23�.253 23�.� 240.1 240.6 241.1 241.6 242.0 242.5 243.0 243.4 243.�54 244.4 244.� 245.3 245.� 246.3 246.� 24�.2 24�.� 24�.2 24�.655 24�.1 24�.6 250.0 250.5 251.0 251.5 251.� 252.4 252.� 253.3
56 253.� 254.3 254.� 255.2 255.� 256.2 256.6 25�.1 25�.6 25�.05� 25�.5 25�.0 25�.5 25�.� 260.4 260.� 261.3 261.� 262.3 262.�5� 263.2 263.� 264.2 264.6 265.1 265.6 266.0 266.5 26�.0 26�.45� 26�.� 26�.4 26�.� 26�.3 26�.� 2�0.3 2�0.� 2�1.2 2�1.� 2�2.160 2�2.6 2�3.1 2�3.6 2�4.0 2�4.5 2�5.0 2�5.4 2�5.� 2�6.4 2�6.�
Example: 30.3 Seconds = 130.9 Centistokes. ©Copyright 1��� Paul N. Gardner Co.
GARDCO MINI FORD DIP viscosity cups are designed to provide just one half the efflux time of the GARDCO STANDARD FORD DIP viscosity cups which comply with requirements of ASTM D-1200. Both cups use the identical orifice. Cup dimensions are carefully controlled and cup calibration conditions comply with MIL STD 45662A. Standard viscous oil traceable to the National Institute of Standards and Technology are used in calibration procedures to insure specified efflux time tolerance.
Use this formula derived by the Paul N. Gardner Company research to find viscosity (V) in centistokes when cup efflux time in seconds (T) is known:
V = 7.40T - 200 ÷ T
Use this formula to find cup efflux seconds (T) when viscosity (V) in centistokes is known:
T = ( V + √ V2 + 5920 ) ÷ 14.80
GARDCO MINI FORD DIP VISCOSITY CUPSCUP NUMBER 4
CONVERSION FORMULAS AND TABLE©
Results from the above formulas, solved for each tenth of a second within the cup range, are shown on the reverse side of this page. To find centistoke viscosity for a given cup efflux time in seconds, read down the column on the left to find the nearest second. Then, read to the right to the nearest tenth of a second column to find centistoke value. The chart may be read in reverse to find efflux time seconds when viscosity is known.
The GARDCO MINI FORD DIP series of six viscosity cups are produced, calibrated and sold only by the Paul N. Gardner Company and licensed agents.
©Copyright 1987, Paul N. Gardner Company
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT.
316 N.E. First Street • Pompano Beach, Florida 33060Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
Paul N. Gardner Company, Inc.
GARDCO MINI FORD DIP VISCOSITY CUP NUMBER 4EFFLUX TIME - CENTISTOKES CONVERSION TABLE ©
10 54 55 56 5� 5� 5� 60 60 61 6211 63 64 65 66 6� 6� 6� 6� �0 �112 �2 �3 �4 �5 �6 �� �� �� �� �013 �1 �2 �3 �3 �4 �5 �6 �� �� ��14 �� �0 �1 �2 �3 �4 �4 �5 �6 ��15 �� �� �� 100 101 102 103 103 104 105
16 106 10� 10� 10� 10� 110 111 112 112 1131� 114 115 116 116 11� 11� 11� 120 120 1211� 122 123 124 124 125 126 12� 12� 12� 12�1� 130 131 132 132 133 134 135 136 136 13�20 13� 13� 140 140 141 142 143 144 144 145
21 146 14� 14� 14� 14� 150 151 151 152 15322 154 154 155 156 15� 15� 15� 15� 160 16123 162 162 163 164 165 165 166 16� 16� 16�24 16� 1�0 1�1 1�2 1�2 1�3 1�4 1�5 1�5 1�625 1�� 1�� 1�� 1�� 1�0 1�1 1�2 1�2 1�3 1�4
26 1�5 1�5 1�6 1�� 1�� 1�� 1�� 1�0 1�1 1�22� 1�2 1�3 1�4 1�5 1�5 1�6 1�� 1�� 1�� 1��2� 200 201 202 202 203 204 205 205 206 20�2� 20� 20� 20� 210 211 212 212 213 214 21530 215 216 21� 21� 21� 21� 220 221 221 222
31 223 224 224 225 226 22� 22� 22� 22� 23032 231 231 232 233 234 234 235 236 23� 23�33 23� 23� 240 240 241 242 243 243 244 24534 246 246 24� 24� 24� 250 250 251 252 25335 253 254 255 256 256 25� 25� 25� 25� 260
36 261 262 262 263 264 265 265 266 26� 26�3� 26� 26� 2�0 2�1 2�1 2�2 2�3 2�4 2�4 2�53� 2�6 2�� 2�� 2�� 2�� 2�0 2�0 2�1 2�2 2�33� 2�3 2�4 2�5 2�6 2�6 2�� 2�� 2�� 2�� 2�040 2�1 2�2 2�3 2�3 2�4 2�5 2�6 2�6 2�� 2��
41 2�� 2�� 300 301 302 302 303 304 305 30542 306 30� 30� 30� 30� 310 311 311 312 31343 314 314 315 316 31� 31� 31� 31� 320 32044 321 322 323 323 324 325 326 326 32� 32�45 32� 32� 330 331 332 332 333 334 335 335
46 336 33� 33� 33� 33� 340 341 341 342 3434� 344 344 345 346 34� 34� 34� 34� 350 3504� 351 352 353 353 354 355 356 356 35� 35�4� 35� 35� 360 361 362 362 363 364 365 36550 366 36� 36� 36� 36� 3�0 3�0 3�1 3�2 3�3
51 3�3 3�4 3�5 3�6 3�6 3�� 3�� 3�� 3�� 3�052 3�1 3�2 3�2 3�3 3�4 3�5 3�5 3�6 3�� 3��53 3�� 3�� 3�0 3�1 3�1 3�2 3�3 3�4 3�4 3�554 3�6 3�� 3�� 3�� 3�� 400 400 401 402 40355 403 404 405 406 406 40� 40� 40� 40� 410
56 411 412 412 413 414 415 415 416 41� 41�5� 41� 41� 420 421 421 422 423 424 424 4255� 426 426 42� 42� 42� 42� 430 431 432 4325� 433 434 435 435 436 43� 43� 43� 43� 44060 441 441 442 443 444 444 445 446 44� 44�
< -----------------------------------------VISCOSITY IN CENTISTOKES ------------------------------------------ >
SECONDS 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.� 0.� 0.�
10/�5 (Accurate for True Liquids Only)
Example: 21.8 Seconds = 152 Centistokes. ©Copyright 1��� Paul N. Gardner Co.
GARDCO MINI FORD DIP VISCOSITY CUPSCUP NUMBER 5
CONVERSION FORMULAS AND TABLE©
GARDCO MINI FORD DIP viscosity cups are designed to provide just one half the efflux time of the GARDCO STANDARD FORD DIP viscosity cups which comply with requirements of ASTM D-1200. Both cups use the identical orifice. Cup dimensions are carefully controlled and cup calibration conditions comply with MIL STD 45662A. Standard viscous oil traceable to the National Institute of Standards and Technology are used in calibration procedures to insure specified efflux time tolerance.
Use this formula derived by the Paul N. Gardner Company research to find viscosity (V) in centistokes when cup efflux time in seconds (T) is known:
V = 23.60T - 204 ÷ T
Use this formula to find cup efflux seconds (T) when viscosity (V) in centistokes is known:
T = ( V + √ V2 + 19528 ) ÷ 47.20
Results from the above formulas, solved for each tenth of a second within the cup range, are shown on the reverse side of this page. To find centistoke viscosity for a given cup efflux time in seconds, read down the column on the left to find the nearest second. Then, read to the right to the nearest tenth of a second column to find centistoke value. The chart may be read in reverse to find efflux time seconds when viscosity is known.
The GARDCO MINI FORD DIP series of six viscosity cups are produced, calibrated and sold only by the Paul N. Gardner Company and licensed agents.
©Copyright 1987, Paul N. Gardner Company
The information contained herein, or supplied by us or on our behalf in any other manner is based on data obtained by our own research and is considered accurate. However, NO WARRANTY EXPRESSED OR IMPLIED REGARDING THE ACCURACY OF THESE DATA, THE RESULTS TO BE OBTAINED FROM THE USE THEREOF, OR THAT ANY SUCH USE WILL NOT INFRINGE ANY PATENT.
Paul N. Gardner Company, Inc.316 N.E. First Street • Pompano Beach, Florida 33060
Phone 1-800-762-2478 • (954) 946-9454 • Fax (954) 946-9309
GARDCO MINI FORD DIP VISCOSITY CUP NUMBER 5EFFLUX TIME - CENTISTOKES CONVERSION TABLE ©
SECONDS 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.� 0.� 0.�
< -----------------------------------------VISCOSITY IN CENTISTOKES ------------------------------------------ >
10 216 21� 221 223 226 22� 231 233 236 23�11 241 244 246 24� 251 254 256 25� 261 26412 266 26� 2�1 2�4 2�6 2�� 2�1 2�4 2�6 2��13 2�1 2�4 2�6 2�� 301 303 306 30� 311 31314 316 31� 321 323 326 32� 331 333 335 33�15 340 343 345 34� 350 353 355 35� 360 362
16 365 36� 3�0 3�2 3�5 3�� 3�� 3�2 3�4 3��1� 3�� 3�2 3�4 3�6 3�� 401 404 406 40� 4111� 413 416 41� 421 423 426 42� 430 433 4351� 43� 440 442 445 44� 450 452 455 45� 45�20 462 464 46� 46� 4�1 4�4 4�6 4�� 4�1 4�3
21 4�6 4�� 4�1 4�3 4�6 4�� 500 503 505 50�22 510 512 515 51� 520 522 524 52� 52� 53223 534 536 53� 541 544 546 54� 551 553 55624 55� 560 563 565 56� 5�0 5�2 5�5 5�� 5��25 5�2 5�4 5�� 5�� 5�1 5�4 5�6 5�� 601 603
26 606 60� 611 613 615 61� 620 622 625 62�2� 630 632 634 63� 63� 642 644 646 64� 6512� 654 656 65� 661 663 665 66� 6�0 6�3 6�52� 6�� 6�0 6�2 6�5 6�� 6�� 6�2 6�4 6�6 6��30 �01 �04 �06 �0� �11 �13 �15 �1� �20 �23
31 �25 �2� �30 �32 �35 �3� �3� �42 �44 �4632 �4� �51 �54 �56 �5� �61 �63 �65 �6� ��033 ��3 ��5 ��� ��0 ��2 ��5 ��� ��� ��2 ��434 ��6 ��� �01 �04 �06 �0� �11 �13 �15 �1�35 �20 �23 �25 �2� �30 �32 �34 �3� �3� �42
36 �44 �46 �4� �51 �53 �56 �5� �61 �63 �653� �6� ��0 ��2 ��5 ��� ��0 ��2 ��4 ��� ���3� ��1 ��4 ��6 ��� �01 �03 �06 �0� �10 �133� �15 �1� �20 �22 �25 �2� �2� �32 �34 �3�40 �3� �41 �44 �46 �4� �51 �53 �56 �5� �60
41 �63 �65 �6� ��0 ��2 ��4 ��� ��� ��2 ��442 ��6 ��� ��1 ��3 ��6 ��� 1001 1003 1005 100�43 1010 1012 1015 101� 1020 1022 1024 102� 102� 103144 1034 1036 103� 1041 1043 1046 104� 1050 1053 105545 105� 1060 1062 1065 106� 106� 10�2 10�4 10�6 10��
46 10�1 10�4 10�6 10�� 10�1 10�3 10�5 10�� 1100 11024� 1105 110� 1110 1112 1114 111� 111� 1121 1124 11264� 112� 1131 1133 1136 113� 1140 1143 1145 114� 11504� 1152 1155 115� 115� 1162 1164 1166 116� 11�1 11�450 11�6 11�� 11�1 11�3 11�5 11�� 11�0 11�2 11�5 11��
51 1200 1202 1204 120� 120� 1211 1214 1216 121� 122152 1223 1226 122� 1230 1233 1235 123� 1240 1242 124553 124� 124� 1252 1254 1256 125� 1261 1264 1266 126�54 12�1 12�3 12�5 12�� 12�0 12�2 12�5 12�� 12�0 12�255 12�4 12�� 12�� 1301 1304 1306 130� 1311 1313 1316
56 131� 1320 1323 1325 132� 1330 1332 1335 133� 133�5� 1342 1344 1346 134� 1351 1353 1356 135� 1361 13635� 1365 136� 13�0 13�2 13�5 13�� 13�� 13�2 13�4 13��5� 13�� 13�1 13�4 13�6 13�� 1401 1403 1406 140� 141060 1413 1415 141� 1420 1422 1424 142� 142� 1432 1434
Example: 45.9 Seconds = 1079 Centistokes. ©Copyright 1��� Paul N. Gardner Co.
10/�5 (Accurate for True Liquids Only)
GARDCO STANDARD FORD DIP VISCOSITY CUPSGARDCO STANDARD FORD VISCOSITY CUPSGARDCO MINI FORD DIP VISCOSITY CUPS
Furnished with each GARDCO Standard, Standard Dip, or Mini Ford Dip Viscosity cup is a conver-sion grid that lists Centistoke viscosity values for true liquids for each tenth of a second within the cup recommended use range. The applicable mathematical formulas are also included with the grid.
CONVERSION BETWEEN EFFLUX TIME AND CENTISTOKES
In order to determine the exact relationship of an unknown cup to a cup in the GARDCO Standard Series, the cup in question should be checked with GARDCO standard oils or oils of known centistoke value from other sources. Determine the efflux time in seconds for the listed centistoke value of the standard oil using the Conversion Grid furnished with each GARDCO cup, or from the mathemati-cal formula included with the grid, or from the GARDCO product leaflet. Run the standard oil in the unknown cup and determine actual efflux time. If the efflux time obtained is materially different than that determined from the grid or mathematical formula, either a correction factor can be determined for future use or, if the difference is substantial, the cup should be replaced.
Results for the GARDCO Standard, Standard Dip, and Mini Ford Dip Cup Series are compatible with results from Ford cups by other major manufacturers who comply with specifications published by the American Society for Testing Materials, when working within imposed efflux time limitations. In-creased range and other advantageous features of the GARDCO viscosity cups are listed in the product leaflets for these cups.
COMPATIBILITY WITH FORD CUPS PRODUCED BY OTHERS
Do not use in place of the dowel rod a metal punch or other similar hard object as damage to the orifice is likely to occur.If the cup/orifice combination has been CERTIFIED, replacing the original orifice with another orifice voids the certification.Use a piece of nylon fishing line to clean orifice side wall.
WARNING
To insert the orifice, place the cup on a solid surface right side up, place the orifice into the cup, insert an end of the dowel rod into the spherical section of the orifice and tap the other end of the dowel rod in a manner similar to orifice removal.
ORIFICE CHANGETo remove the orifice, simply place the cup upside down on a solid surface (let the handle of the dip cup extend below the edge of the surface), place an end of the wood dowel rod, furnished with the cup, squarely against the orifice and tap the other end of the dowel rod smartl;y with a small hammer or similar heavy object, such as a paper weight.
Paul N. Gardner Company, Inc.316 N.E. 1st Street, POmPANO BCH, FL 33060
1-800-762-2478 • (954) 946-9454 • FAX (954) 946-9309
Solution of Cup Formula ***
The following examples are in relation to the GARDCO Standard #4 Ford Cup and the method is applicable to all cups in the GARDCO Standard Ford Cup Series.
Solve for Viscosity (V) when Time (T) in seconds is known:
Viscosity = 3.70T - 400 / T
GARDCO STANDARD FORD DIP VISCOSITY CUPViscosity Measurement By
GARDCO STANDARD FORD DIP VISCOSITY CUPSGARDCO STANDARD FORD VISCOSITY CUPSGARDCO MINI FORD DIP VISCOSITY CUPS
Assume that Time in seconds (T) is 45.5. Substitute this value in the above equation and solve by the following steps:
Viscosity = 3.70 X 45.5 - 400 ÷ 45.5 Viscosity = 168.4 - 8.8 Viscosity = 159.6
Time = 159.5 ÷ 7.40 + √ 159.52 + 5920 ÷ 7.40 Time = 159.5 ÷ 7.40 + √ 25440 + 5920 ÷ 7.40 Time = 159.5 ÷ 7.40 + √ 31360 ÷ 7.40 Time = 159.5 ÷ 7.40 + 177.1 ÷ 7.40 Time = 21.6 + 23.9 Time = 45.5 *** Reference page 4 of Ford Cup leaflet dated R-4/93
Assume that the Viscosity value (V) is 15�.5. Substitute this value in the above equa-tion and solve by the following steps:
Solve for Time in seconds (T) when the Viscosity in centistoke (V) value is known:
Time = V / 7.40 + √ V2 + 5920 / 7.40
CALIBRATION OIL STATEMENTEFFLUX VISCOSITY CUPS
VIBL-CC-1/98
Paul N. Gardner Company, Inc.316 N.E. FIRST STREET • POMPANO BEACH, FL 33060
PHONE (�54) �46-�454 • �00-�62-24�� • FAX (�54) �46-�30�
The National Institute of Standards and Technology traceable standard oils used in the calibration of all certified efflux type viscosity cups made and sold by The Paul N. Gardner Company, Inc. (GARDCO) are standard oils prepared expressly for the Paul N. Gardner Com-pany, Inc. (GARDCO) by the Cannon Instrument Company and are produced in accordance with ISO 9002: 1994, EN ISO 9002: 1994; BS EN ISO 9002: 1994; ANSI/ASQC Q9002: 1994.
The following viscosity cups are calibrated with oils produced in accordance with ISO:
GARDCO Standard Ford Cups ..............Cups #0, #1, #2, #3, #4 and #5GARDCO/ISO Cups .......................3mm, 4mm, 6mm, and 8mm CupsGARDCO/DIN Cup .................................................................4mm CupGARDCO/Parlin Cups........................................#1, #2, #3 and #4 CupsGARDCO/Fisher Standard Cups ...................... #1, #2, #3 and #4 CupsGARDCO Standard Ford Dip Cups.........................Cups #3, #4 and #5GARDCO Mini Ford Dip Cups ...............Cups #0, #1, #2, #3, #4 and #5GARDCO/ISO Dip Cups .......................... 3mm, 4mm, and 6mm CupsGARDCO/DIN Dip Cup ...........................................................4mm CupGARDCO/Fisher Dip Cups ................................ Cups #1, #2, #3 and #4GARDCO EZ™ Zahn (ASTM) Dip Cups .......Cups #1, #2, #3, #4 and #5GARDCO S90/Zahn Signature Dip Cups ....Cups #1, #2, #3, #4 and #5
GARDCO MINI FORD DIP VISCOSITY CUPS
EFFLUX
TI
ME
SECONDS
EFFLUX TIME VERSUS TEMPERATURE
Gardco produced viscosity cups are calibrated with standard "G" Series oils. Centistoke viscosity of these oils is traceable to the National Institute of Standards and Technology. These oils are available from the Paul N. Gardner Company.
Shown in the above graph is the viscosity cup number and the standard "G" oil used for its calibration. Normally, cup calibration is at 25°C, shown on the graph by bold lines intersecting with the curve in the circle.
Viscosity of most liquids, including the standard oils, are dependent on temperature. Ef-flux time in seconds for the indicated cup-oil combination from twenty (20) to twenty-seven (2�) degrees Celsius is shown in the above graph. The cup may be checked with the indicated "G" oil with reasonable accuracy within these limits. For best accuracy, the viscosity of the standard oil in Centistokes, as printed on the standard oil bottle label, at a temperature of 25° Celsius should be used. Conversion from viscosity to efflux time in seconds is by the formula or table furnished with the cup.
This information, furnished as an additional customer service, is included with each viscosity cup sold by the Paul N. Gardner Company or by licensed distributors.
20 21 22 23 24 25 26 2�
54
52
50
4�
46
44
MINI FORD DIP CUP #0G-60 STANDARD OIL
25 DEGREES
46.37 SECONDS
8.95 CENTISTOKES
TEMPERATURE SCALE CONVERSIONBETWEEN
CELSIUS AND FAHRENHEITDEGREES
CELSIUS FAHRENHEIT 20.0 6�.0 20.1 6�.2 20.2 6�.4 20.3 6�.5
20.4 6�.� 20.5 6�.� 20.6 6�.1 20.� 6�.3
20.� 6�.4 20.� 6�.6 21.0 6�.� 21.1 �0.0
21.2 �0.2 21.3 �0.3 21.4 �0.5 21.5 �0.� 21.6 �0.� 21.� �1.1 21.� �1.2 21.� �1.4 22.0 �1.6 22.1 �1.� 22.2 �2.0 22.3 �2.1
22.4 �2.3 22.5 �2.5 22.6 �2.� 22.� �2.� 22.� �3.0 22.� �3.2 23.0 �3.4 23.1 �3.6
23.2 �3.� 23.3 �3.� 23.4 �4.1 23.5 �4.3
DEGREESCELSIUS FAHRENHEIT 23.6 �4.5 23.� �4.� 23.� �4.� 23.� �5.0
24.0 �5.2 24.1 �5.4 24.2 �5.6 24.3 �5.�
24.4 �5.� 24.5 �6.1 24.6 �6.3 24.� �6.5
24.� �6.6 24.� �6.� 25.0 ��.0 25.1 ��.2
25.2 ��.4 25.3 ��.5 25.4 ��.� 25.5 ��.�
25.6 ��.1 25.� ��.3 25.� ��.4 25.� ��.6
26.0 ��.� 26.1 ��.0 26.2 ��.2 26.3 ��.3
26.4 ��.5 26.5 ��.� 26.6 ��.� 26.� �0.1
26.� �0.2 26.� �0.4 2�.0 �0.6
F° = (C° x 1.8) + 32C° = (F° - 32°) ÷ 1.8
42
40
3�
36
34
32
40
3�
36
34
32
30
2�
20 21 22 23 24 25 26 2�
20 21 22 23 24 25 26 2�
EFFLUX
TI
ME
SECONDS
EFFLUX
TI
ME
SECONDS
MINI FORD DIP CUP #1G-10 STANDARD OIL
MINI FORD DIP CUP #2G-35 STANDARD OIL
35.09 SECONDS
17.60 CENTISTOKES
25 DEGREES
31.22 SECONDS
65.09 CENTISTOKES
25 DEGREES
TEMPERATURE IN DEGREES CELSIUS
TEMPERATURE IN DEGREES CELSIUS
MINI FORD DIP CUP #3G-60 STANDARD OIL
MINI FORD DIP CUP #4G-60 STANDARD OIL
MINI FORD DIP CUP #5G-200 STANDARD OIL
24
22
20
1�
16
14
25
23
21
1�
1�
15
2�
2�
25
23
21
1�
1�
20 21 22 23 24 25 26 2�
20 21 22 23 24 25 26 2�
20 21 22 23 24 25 26 2�
EFFLUX
TI
ME
SECONDS
EFFLUX
TI
ME
SECONDS
EFFLUX
TI
ME
SECONDS
17.49 SECONDS
65.09 CENTISTOKES
25 DEGREES
17.57 SECONDS
118.6 CENTISTOKES
25 DEGREES
19.87 SECONDS
458.6 CENTISTOKES
25 DEGREES
GARDNER BUILDING, 316 N.E. FIRST ST., POMPANO BEACH, FL 33060PO BOX 10688, POMPANO BEACH, FLORIDA 33061-6688
1-800-762-2478 • PHONE: (954) 946-9454 • FAX (954) 946-9309WEB SITE: www.gardco.com • E MAIL: [email protected]
VI-Revised 4/4/06
A Service To Industry
GARDCO VISCOSITY CUP CERTIFICATIONand
GARDCO INSTRUMENT CERTIFICATION
QUALIFIES UNDER ANSI/NCSL Z540-1 OR MIL STD 45662A AS APPLICABLE
CALIBRATION OILS TRACEABLE TO N.I.S.T.
PAUL N. GARDNER COMPANY, INC.
GARDCO-produced viscosity cups are calibrated with standard "G" Series oils. Centistoke viscosity of these oils is traceable to the National Institute of Standards and Technology. These standard oils, prepared expressly by the Cannon Instrument Company for the Paul N. Gardner Company, are produced in accordance with ISO 9002: 1994, EN ISO 9002: 1994; BS EN ISO 9002: 1994; ANSI/ASQC Q9002: 1994.
®
GARDCO VISCOSITY CUP CERTIFICATIONQUALIFIES UNDER ANSI/NCSL Z540-1 OR MIL STD 45662A AS APPLICABLE STANDARD "G" CALIBRATING OILS
ARE TRACEABLE TO NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
PAUL N. GARDNER COMPANY, INC.GARDNER BUILDING, 316 N.E. FIRST ST., POMPANO BEACH, FL 33060
POST OFFICE BOX 10688, POMPANO BEACH, FL 33061-6688PHONE: (954)-946-9454 • 1-800-762-2478 • FAX (954) 946-9309
GARDCO VISCOSITY CUP
NAME SIZE
STANDARD "G" OIL NUMBER
G-6G-10G-35G-60G-60G-200
G-10G-35G-100G-350
G-100
G-60G-60G200
0.133 313 0.372 600 1.24 770 2.31 550 3.70 400 11.80 408
CUPFORMULA
CONSTANTSK C
0.443 200 1.37 200 6.90 570 21.78 306
4.57 452
2.31 550 3.70 400 11.80 408
222222
CUPTOLER-ANCE± %
2223
3
323
GARDCO STANDARD FORD CUP 0GARDCO STANDARD FORD CUP 1GARDCO STANDARD FORD CUP 2GARDCO STANDARD FORD CUP 3GARDCO STANDARD FORD CUP 4GARDCO STANDARD FORD CUP 5
GARDCO/ISO CUP 3mmGARDCO/ISO CUP 4mmGARDCO/ISO CUP 6mmGARDCO/ISO CUP 8mm
GARDCO/DIN CUP 4mm
G-6G-10G-35G-60G-60G-200
G-10G-35G-100
G-100G-20G-35G-60G-200
G-10G-60G-200G-200G-350
G-35G-60G-100G-350G-350
0.266 157 0.744 300 2.48 385 4.62 275 7.40 200
23.60 204 0.886 100 2.74 100 13.80 285 9.14 226 0.85 175 2.32 190 5.39 185 18.90 210
0.875 993 2.80 747 10.09 587 13.26 673 23.56 744
1.59 1070 4.18 760 10.23 575 15.13 545 27.27 540
444444
333
2222
3
33333
55555
GARDCO MINI FORD DIP CUP 0GARDCO MINI FORD DIP CUP 1GARDCO MINI FORD DIP CUP 2GARDCO MINI FORD DIP CUP 3GARDCO MINI FORD DIP CUP 4GARDCO MINI FORD DIP CUP 5
GARDCO/ISO DIP CUP 3mmGARDCO/ISO DIP CUP 4mmGARDCO/ISO DIP CUP 6mm
GARDCO/DIN DIP CUP 4mm
GARDCO/FISHER DIP CUP 1GARDCO/FISHER DIP CUP 2GARDCO/FISHER DIP CUP 3GARDCO/FISHER DIP CUP 4 GARDCO EZ™ ZAHN (ASTM) DIP CUP 1GARDCO EZ™ ZAHN (ASTM) DIP CUP 2GARDCO EZ™ ZAHN (ASTM) DIP CUP 3GARDCO EZ™ ZAHN (ASTM) DIP CUP 4GARDCO EZ™ ZAHN (ASTM) DIP CUP 5
GARDCO S90/ZAHN SIGNATURE DIP CUP 1GARDCO S90/ZAHN SIGNATURE DIP CUP 2GARDCO S90/ZAHN SIGNATURE DIP CUP 3GARDCO S90/ZAHN SIGNATURE DIP CUP 4GARDCO S90/ZAHN SIGNATURE DIP CUP 5
GARDCO STANDARD FORD DIP CUP 3GARDCO STANDARD FORD DIP CUP 4GARDCO STANDARD FORD DIP CUP 5
GARDCO /PARLIN CUP 1GARDCO /PARLIN CUP 2GARDCO /PARLIN CUP 3GARDCO /PARLIN CUP 4
G-35G-100G-350G-J3000
1.55 800 4.82 100 20.75 500139.00 750
3335
G-20G-35G-60G-200
0.85 175 2.32 190 5.39 185 18.90 210
2222
GARDCO/FISHER STANDARD CUP 1GARDCO/FISHER STANDARD CUP 2GARDCO/FISHER STANDARD CUP 3GARDCO/FISHER STANDARD CUP 4
2
CONVERSION BETWEEN CENTISTOKES “V” AND CUP DRAIN SECONDS "T":
V = KT – C ÷ T or T = ( V + √ V + 4KC ) ÷ 2KWHERE “K” AND “C” ARE THE ABOVE LISTED CONSTANTS
SAM
PLE
REV 11/98
GARDCO INSTRUMENT ITEM ITEM MINIMUMITEM AND RANGE TOLERANCE CHECK GRADUATIONS POINTS
WEIGHT PER GALLON CUPS U.S. STANDARD CUP 0.5% 83.205gr NA CUP WITH TARE WEIGHT 0.2G Cup Wt NA BRITISH STANDARD CUP 0.5% 99.925gr NA CUP WITH TARE WEIGHT 0.2G Cup Wt NA U.S. MINI CUP 1.2% 8.321gr NA CUP WITH TARE WEIGHT 0.1G Cup Wt NA
TARELITE WEIGHT PER GALLON CUPS U.S. STANDARD CUP 2% 83.205gr NA CUP WITH TARE WEIGHT 0.2G Cup Wt NA BRITISH STANDARD (100cc) CUP 2% 99.925gr NA CUP WITH TARE WEIGHT 0.2G Cup Wt NA BRITISH Standard (50cc) CUP 2% 49.963gr NA CUP WITH TARE WEIGHT 0.1G Cup Wt NA
WET FILM THICKNESS GAGE Mils Mils Mils MODEL C, 0 - 2 MILS 0.2 .5, 1, 1.5 0.1 MODEL C, 0 - 4 MILS 0.2 1, 2, 3 0.2 MODEL C, 2 - 12 MILS 0.4 4, 7, 10 0.5 MODEL C, 10 - 30 MILS 0.5 15, 20, 25 1.0 MODEL C, 20 - 60 MILS 0.5 30, 40, 50 2.0
WET FILM THICKNESS GAGE MILS MICRONS MILS MICRONS MILS MICRONS MODEL IC 0 - 1 MILS / 0 - 25 MICRONS 0.1 2.5 .25,.5,.75 5,13,20 .05 1.0 MODEL IC 0 - 2 MILS / 0 - 50 MICRONS 0.2 5.0 .5,1,7.5 10,26,40 0.1 2.0 MODEL IC 0 - 4 MILS / 0 - 100 MICRONS 0.2 5.0 1,2,3 20,52,80 0.2 4.0 MODEL IC 2 - 12 MILS / 50 - 300 MICRONS 0.4 10 4,7,10 100,180,250 0.5 10 MODEL IC 10 - 30 MILS / 250 - 750 MICRONS 0.5 13 15,20,25 380,500,640 1.0 20 MODEL IC 20 - 60 MILS / 500 - 1500 MICRONS 0.5 13 30,40,50 760,1000,1280 2.0 40
WET FILM THICKNESS GAGE MILS MILS MILS MODEL CC 2-6 MILS 0.2 3,4,5 0.2 MODEL CC 4-8 MILS 0.2 5,6,7 0.2
GUARDED RING TENSIONED THICKNESS STANDARD MILS MILS GRTTS. 1.0 MIL NOMINAL 0.1 1 NA GRTTS, 2.0 MIL NOMINAL 0.1 2 NA GRTTS, 5.0 MIL NOMINAL 0.1 5 NA UNIVERSAL APPLICATOR, ADJUSTABLE BLADE TYPE MILS MILS MILS 2" BLADE, 0 - 50 MILS 1.0 0,25,50 1.0 4" BLADE, 0 - 50 MILS 1.0 0,25,50 1.0 6" BLADE, 0 - 50 MILS 1.0 0,25,50 1.0 8" BLADE, 0 - 50 MILS 1.0 0,25,50 1.0 10" BLADE, 0 - 50 MILS 1.0 0,25,50 1.0 12" BLADE, 0 - 50 MILS 1.0 0,25,50 1.0 2" BLADE, 0 - 10 MILS 0.5 0,5,10 0.2 4" BLADE, 0 - 10 MILS 0.5 0,5,10 0.2 6" BLADE, 0 - 10 MILS 0.5 0,5,10 0.2 8" BLADE, 0 - 10 MILS 0.5 0,5,10 0.2 10" BLADE, 0 - 10 MILS 0.5 0,5,10 0.2 12" BLADE, 0 - 10 MILS 0.5 0,5,10 0.2
PERMEABILITY CUP MM MM STANDARD DIAMETER .05 56.4 NA STANDARD HEIGHT .05 10 NA
P-A-T BLADES MM MM 1.0MM - 6 TEETH .05 1.0 NA 1.0MM - 11 TEETH .05 1.0 NA 1.5MM - 11 TEETH .05 1.5 NA 2.0MM - 6 TEETH .05 2.0 NA 3.0MM - 6 TEETH .05 3.0 NA 5.0MM - 5 TEETH .05 5.0 NA INCH INCH 3/32 - 4 TEETH .002 3/32 NA
INSTRUMENT CERTIFICATIONQUALIFIES UNDER ANCI/NCSL Z540-1 OR MIL STD 45662A AS APPLICABLE AND CONFORMS TO ISO 9000
WHEN ORDERED WITH CALIBRATION CERTIFICATION
NOTES: (1) Prices quoted are FOB, Pompano Beach, Florida. Prices are subject to change without notice (2) Customer Service can provide certification prices for instruments not listed.
®
Paul N. Gardner Co., Inc. • www.gardco.com • (954) 946-9454 • 1-800-762-2478
PAUL N. GARDNER COMPANY, INC.GARDNER BUILDING, 316 N.E. FIRST ST., POMPANO BEACH, FL 33060
POST OFFICE BOX 10688, POMPANO BEACH, FL 33061-6688PHONE: (954)-946-9454 • 1-800-762-2478 • FAX (954) 946-9309