microstream capnography exercises: physio-control lifepak 12 and lifepak 15 monitor/defibrillators

2
MICROSTREAM CAPNOGRAPHY EXERCISES PHYSIO-CONTROL LIFEPAK ®* 12 AND LIFEPAK ®* 15 MONITOR/DEFIBRILLATORS The purpose of this exercise is to gain a better understanding of the practical application of the FilterLine sampling line, the etCO 2 value and the waveform and documentation with the Physio-Control LifePak 12 or LifePak 15 monitor/defribrillators. The exercises will provide you with hands on experience with Microstream capnography and the responsiveness of the technology with changing patient conditions. NOTE: The following exercises should be performed while wearing Smart CapnoLine Adult non-intubated sampling line and a pulse oximetry sensor. Please perform exercises for hypoventilation, hyperventilation and breath-hold while sitting in a chair. BASELINE ASSESSMENT 1. Turn on the ZOLL monitor 2. Plug in the Smart CapnoLine Plus non-intubated Oral-nasal sampling line. 3. Press the CO 2 quick access key ( ) to initiate CO 2 monitoring. 4. Apply the Smart Capnoline Plus sampling line to your face. a. Monitor your breathing for 60 seconds or until a respiratory rate appears on the Physio-Control LifePak 12 or LifePak 15 monitor/defibrillator. 5. Put the pulse oximeter sensor on your index finger. b. What is your CO 2 value?_____mmHg c. What is your RR? __________ d. What is your SpO 2 %? _______ % These CO 2 and SpO 2 values are considered your baseline and are important to capture as the starting point with each patient you are monitoring. IMPORTANT NOTE: If the Physio-Control LifePak 12 or LifePak 15 monitor/ defibrillator does not have a capnography waveform on Channel 2 or 3, use the selector knob to choose the capnography waveform. 6. Observe the capnography waveform on the screen and press print to collect a CO 2 waveform strip. HYPOVENTILATION SIMULATION 1. Decrease respiratory rate by trying to increase your baseline CO 2 level by 5 mmHg. 2. What is your CO 2 value? ______ mmHg 3. Observe waveform and capture the event by pressing print. Smart CapnoLine sampling line with O 2

Upload: ed-praetorian

Post on 29-Jan-2016

199 views

Category:

Documents


1 download

DESCRIPTION

The purpose of this exercise isto gain a better understandingof the practical applicationof the FilterLine™ samplingline, the etCO2 value and thewaveform and documentationwith the Physio-ControlLifePak 12 or LifePak 15monitor/defribrillators. Theexercises will provide you withhands on experience withMicrostream™ capnographyand the responsiveness ofthe technology with changingpatient conditions.

TRANSCRIPT

Page 1: Microstream Capnography Exercises: Physio-Control Lifepak 12 And Lifepak 15 Monitor/Defibrillators

MICROSTREAM™ CAPNOGRAPHY EXERCISESPHYSIO-CONTROL LIFEPAK®* 12 AND LIFEPAK®* 15 MONITOR/DEFIBRILLATORS

The purpose of this exercise is to gain a better understanding of the practical application of the FilterLine™ sampling line, the etCO

2 value and the

waveform and documentation with the Physio-Control LifePak 12 or LifePak 15 monitor/defribrillators. The exercises will provide you with hands on experience with Microstream™ capnography and the responsiveness of the technology with changing patient conditions.

NOTE: The following exercises should be performed while wearing Smart CapnoLine™ Adult non-intubated sampling line and a pulse oximetry sensor. Please perform exercises for hypoventilation, hyperventilation and breath-hold while sitting in a chair.

BASELINE ASSESSMENT1. Turn on the ZOLL monitor

2. Plug in the Smart CapnoLine™ Plus non-intubated Oral-nasal sampling line.

3. Press the CO2 quick access key ( ) to initiate CO2 monitoring.

4. Apply the Smart Capnoline Plus sampling line to your face.

a. Monitor your breathing for 60 seconds or until a respiratory rate appears on the Physio-Control LifePak 12 or LifePak 15 monitor/defibrillator.

5.Putthepulseoximetersensoronyourindexfinger.

b. What is your CO2 value?_____ mmHg

c. What is your RR? __________

d. What is your SpO2%? _______ %

These CO2 and SpO2 values are considered your baseline and are important to capture as the starting point with each patient you are monitoring.

IMPORTANT NOTE: If the Physio-Control LifePak 12 or LifePak 15 monitor/defibrillatordoesnothaveacapnographywaveformonChannel2or3,usetheselector knob to choose the capnography waveform.

6. Observe the capnography waveform on the screen and press print to collect a CO2 waveform strip.

HYPOVENTILATION SIMULATION 1. Decrease respiratory rate by trying to increase your baseline CO2 level by 5 mmHg.

2. What is your CO2 value? ______ mmHg

3. Observe waveform and capture the event by pressing print.

Smart CapnoLine™ sampling line with O2

Page 2: Microstream Capnography Exercises: Physio-Control Lifepak 12 And Lifepak 15 Monitor/Defibrillators

HYPERVENTILATION SIMULATION1. Increase your respiratory rate by breathing rapidly and trying to decrease your

baseline CO2 level by 5 mmHg.

2. What is your CO2 value? _________ mmHg

3. Observe the CO2 waveform and capture the event by pressing print.

APNEA SIMULATION1. With the SpO2probeonfinger,holdbreathtotriggerCO2 apnea alarm (30 seconds).

2. Observe waveform, CO2 value and SpO2 value. Capture the event by pressing print.

3. Observation: __________________________________________________________

SAMPLING LINE OCCLUSION SIMULATION1. Kink FilterLine™ sampling line until Physio-Control LifePak 12 or LifePak 15 monitor/

defibrillatorindicatesthattheCO2 module is performing a purge. To attempt to clear an occlusion, these monitors purge whenever an occlusion in the line or airway adaptor is detected, and will display PURGING while this function is performed. If the CO2modulecannotcleartheocclusionwithin30seconds,themonitor/defibrillatorwill issue an equipment alert and display the message: FILTERLINE OCCLUSION.

2. Observe message, then remove the kink from FilterLine™ sampling line.

3. Do you observe a CO2 waveform and etCO2 value? If not, disconnect the FilterLine™ sampling line and reconnect to reactivate the CO2 pump.

4. If reconnecting fails to produce a CO2 waveform and etCO2 value, replace the FilterLine™ sampling line.

5. Observation: __________________________________________________________

COVIDIEN MEDICAL AFFAIRS

6135 Gunbarrel Avenue Boulder, CO 80301

800-635-5267

covidien.com/PACE

© 2015 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together aretrademarks of Medtronic. ®* and ™* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company. 15-PM-0180