microcont ii - iljin rad -display in cps, cps net, bq, bq/cm 2-automatic background substraction-10...
TRANSCRIPT
www.rados.com
RTM430 for measuringcontamination
- easy to use- display in cps, cps net, Bq, Bq/cm2
- automatic background substraction- 10 profiles, all parameters can only be modified via
normal office PC and software by using an externalIR-interface at the COM-port
- memory for max.500 measurement results, directexport into Excel on PC possible
ααααα/β β β β β or β β β β β/γγγγγ measurement high sensitivity by largedetection area large multifunctional display
MicroCont II
- connection to external detector by adapterplate (option)
- display of selected nuclide and calibrationfactor
- large-area LCD, can be illuminated, asmultifunction display
- measurement and system parameters can bepre-set via PC and special software
The advantages at a glance
alarm set points:can be freely preselected over the whole range,accoustic alarm to be aknowledged, optical alarm byalarm segment on display
system requirement: Win95/98/2000 ME/NT/XP,analogue measurement display as bar chart andacoustic single pulse indicator (switchable)membrane keyboard with large keys having a cleardetent action
accurate determination of activity through long-term
measurements:
in timer mode with a selectable measurement periodof up to 103s
in counter mode using pulse selection, automatictermination of measurement. Storage of data possible
As standards, specifications anddesign change from time to time,please ask for confirmation of the
information given in this publication.
Lamanon - FranceTurku - Finland
Hamburg - GermanySmyrna (GA) - USA
Other countries
128301D Tel +33 (0)4 90 59 59 59Tel +358 2 4684 600Tel +49 40 85193-0Tel +001 (770) 432 2744Tel +33 (0)4 90 59 60 41
Representative address:
Mechanical characteristics
housing: compact housing in impact-resistant ABS, splash-proof (IP 63), surface can bedecontaminated, coloured yellow, ergonomically shaped, handle with integrated on/offbutton
weight:MicroCont II with gasflow detector approx. 2100 g (4.62 lb)MicroCont II with xenon detector approx. 1950 g (4.29 lb)
The job defines the detector
The MicroCont II can be operated with a sealed Xenondetector or with a large-area flow through detector
RXE Xenon Large-Area Detectorhermetically sealed proportional detector for β and γcontamination measurementsno gaz supply necessaryhigh efficiency factor fir the nuclides used in nuclearmedicinelarge detection area (233 cm2 - 36.11 in2)fast localisation of contamination due to the graphicalTitanium foil, 5 mg/cm2 . Can be used at temperaturesdown to -10°C (14°F)
RGZ large-Area Gas Flow Detector
integrated gas suplly (75 cm3 - 4.57 in3)
simple gas flushing via press button
refilling from disposable bottles of liquid butane-propane
RBP 170 plastic scintillator (option)
Electrical characteristics
power supply: 3 standard baby cells with alkaline batteries, 200 hours operationalperiod (battery check in LCD with check key) or external power supply 3 - 5 V/ 230V,50 Hzwarning functions: one warning threshold for each measuring mode (cps, cps net,Bq and Bq/cm2) adjustable over the whole measurement range, optical alarm in LCD,acoustic alarm can be acknowledged, display of selected alarm thresholds with checkkey
Environmental characteristics
service temperature range:with RXE detector: -10 ... +50 °C (+14°F to 122°F)with RGZ detector: 0 ... +50 °C (32°F to 122°F) at a max. of 90% real humiditywithout condensation
storage temperature:-40 ... +50 °C (-40 to 122°F) max 75% humidity as yearly average, withoutcondensation
Detector efficiency factor in %
Nuclide RGZ 190
RXE 260
Nuclide RGZ 190
RXE 260
Nuclide RBP 170
MO-99 /Tc-99
C-14 15 2,5 3,8 4 C-14 6
P-32 40 34 In-111 7,5 10,5 CI-36 30
S-35 15 5,2 I-123 5 7 Co-60 16
K-40 38 32 I-125 5 Sr-90 65,5
Co-57 6 8,5 I-131 27 23 Sr/Y-90 33
Co-60 28 13 Cs-137 38 24 Tc-99 m 1
Sr/Y-90 34 34 Am-241 22 I-131 18,5
TI-204 35 31 Cs-137 27
Am-241 14