MEDTRONICGASTROENTEROLOGY

Bile Reflux and Bilitec System

Page 2: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

What is DG- or DGE- Reflux?

Retrograde movement of duodenal content such as bile, pancreatic and enteric juices to the stomach and esophagus:

Duodeno-Gastric Reflux (DGR)

Duodeno-Gastro-Esophageal Reflux (DGER)Stomach

Sphincter of Oddi

Gallblader

Pancreas

Duodenum

Bile reflux

Acid reflux

Esophagus

LES

Page 3: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Indications

Patients with symptoms of gastroesophageal reflux (GER) Evaluation in patients with complications of GER such as:

Barrett’s esophagus Strictures Ulceration of the esophagus

Patients that have failed medical therapy for acid reflux Patients with poor response to medical treatment of reflux

esophagitis Postgastrectomized patients with reflux symptoms Pre- and postoperative evaluation of anti-reflux surgery

Page 4: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

How to Measure DG- and DGE- Reflux

The Bilitec System has been recognized as being the best method for detecting duodenogastric/bile reflux

4 studies has validated that:

Bilirubin is a suitable marker to measure the presence of bile acids – i.e. demonstrate that there is a correlation between Bilitec readout (presence of bilirubin) and presence of biliary salts

In-vivo spectrophotometric reading of Bilitec is reliable - i.e. compare Bilitec reading with spectrophotometric measurements carried out on samples drawn from the stomach and/or esophagus; this could be defined as optical validation

Page 5: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Why is Bile Reflux Important?

The main reason to measure bile-reflux is based on the indications that in GERD, duodenal juice is an important component in the refluxed gastric juice that causes damage to the esophageal mucosa.

Page 6: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Composition of Refluxed Juice in GERD*

*From DeMeester et al: Biology of Gastroesophageal Reflux Disease, 1999

Page 7: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Is measurement of DGR/DGER relevant?

Bile Reflux

Increased Risk for Mucosal Damage

The clinical benefit of the Bilitec system is to identify patients at risk for: Mucosal injuryBarret’s esophagusHigh grade dysplasia.

Select patients for further examination of esophageal damage (e.g. biopsy) and for whom anti-reflux surgery should be considered

Page 8: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Examples of Bile reflux studies

“Patients with erosive esophagitis and Barrett’s metaplasia have increased esophageal exposure to duodenal juice compared to normal subjects”, Kauer et al, 1995

“The combined reflux of gastric and duodenal juices causes severe esophageal mucosal damage. The vast majority of duodenal reflux occurs at a pH range of 4 to 7, at which bile acids, the major component of duodenal juice, are capable of damaging the esophageal mucos”, Kauer et al, 1995

“Duodenal juice adds a noxious component to the refluxed gastric juice and potentiates the injurious effects of gastric juice on the esophageal mucosa” Fein et al, 1997

“The prolonged simultaneous attack of bile and acid may play a key role in the development of Barrett's metaplasia”, Menges et al, 2001

...

Page 9: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Detection of Bile/Bilirubin

Bilitec uses two wavelengths for detection of bilirubin: 470 nm and 565 nm

The absorbance is calculated from Beer’s law based on light intensities transmitted in absence and presence of absorbing substance (bilirubin)

The absorbance is proportional to the concentration of bilirubin however, Bilirubin is only a marker and the knowledge of its concentration is not important from a clinical point of view

Bilitec measures the exposure time of bile-containing reflux to the gastric-/esophageal mucosa

nm400 500 600 700 800

A

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

Absorption spectrum of a gastric sample containing bile

Page 10: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Normal values

Normal values proposed by P.Bechi for gastric studies:Pct. Time of bilirubin exposure>0.14 in proximal stomach*

Stomach

Total: 12.9 %Supine: 24.2 %Upright: 4.0 %*Bilitec probe placed 10 cm below LES lower boarder

Normal values from Kauer, DeMeester et al for esophageal studies:Pct. Time of esophageal* bilirubinexposure >0.2

Esophagus

Total: 2.9 %Supine: 0.4 %Upright: 4.0 %*Bilitec probe placed 5 cm above LES upper boarder

Page 11: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Selection of Absorbance Threshold

A default threshold of 0.14 in absorbance values is considered in all the validations. This is the threshold introduced by P. Bechi: “it is the lowest value where the absorbance can be considered as an actual measurement of bilirubin and not as noise due to mucus in the stomach”

There are different opinions on the threshold level, and in other studies higher absorbance values are used as threshold (e.g. 0.2 and 2.5).

From presentation by T.DeMeester, DDW, 2001

Page 12: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Correction of absorbance values rel. pH

Is Bile reflux underestimated in an acidic environment?

This is a controversial subject discussed in a few papers: J.Richter: 30% underestimation of Bilitec reading for pH<3.5 P.Bechi: 18% underestimation of Bilitec reading for pH<3.5 J.Janssens: such underestimation has only minor effect on the

results of Bile reflux testing and thereby, have no clinical impact.

Only few centers consider this in their analysis of Bile reflux

Page 13: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Bilitec System Components

Fiber Optic Catheter

Bilitec 2000 Carrying case Cable for serial interface Optical calibration chambers Manuals

Fiber Optic Bile Catheter Single-use catheter, length 175 cm, diam. 3.0 mm

Bile Upload and analysis option for P98 pH

Note: Available in US when 510(k) cleared

Page 14: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

The Bilitec Procedure

Set-up Bilitec 2000 device Set average time and 1 or 2 channels Calibrate Bilitec system in water

Intubate patient with single-use fiber optic probe w/wo pH recording

Start the measurement Standard Diet: abs. < 0.05 at 470 nm Patient Keeps Diary: symptoms & events

Extubate patient, upload and analyze data

Page 15: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Features and Benefits

Features BenefitsUp to 2 channels can be recordersimultaneously over 24 hours

Perform simultaneous gastric- and esophagealbile studies.

Compatible with both Polygram 98 pH TestingApplication, PW EsopHogram Reflux AnalysisModule and DOS EsopHogram

Current users can analyze data using theirexisting Medtronic software.

Single-Use optic probe and Calibration is easilydone in water before each study

Limited resources and no special componentsneeded for preparation of the system

Absorbance threshold can be defined by user Ability to set reflux threshold to match everystudy design

Correction algorithm of bile values can beenabled in protocol settings

Ability to adjust for underestimation of bile refluxin acid environment (if required)

Record and analyses data for 24 hours Evaluate the circadian pattern of bile reflux, i.e.total, upright, supine periods and correlatesymptoms to reflux episodes.

Page 16: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Polygram Bile Analysis Software

Technical Review of the Software

Page 17: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Protocol Settings

0-4 channels pH and1-2 channels Bile

Adjustment of Bile values

Page 18: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Protocol Settings

Analysis of Bile Reflux:

Page 19: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Bile Analysis

Page 20: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Upload Procedure for Combined Studies

Upload pH data first

…awaiting data from the Bilitec2000 device connect Bilitec to the serial port go to data transfer mode and press “Enter”

NB: Uses Digitrapper time for Combined Studies- the Bilitec 2000 has no Clock

Page 21: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Capture View

Page 22: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Review view

Page 23: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Combined pH and Bile Report

Page 24: MEDTRONIC GASTROENTEROLOGY Bile Reflux and Bilitec System

MEDTRONICGASTROENTEROLOGY

Pitfalls

Variety of substances in food may result in false-positive readings.- Therefore, it is important to use a modified diet.

If long periods of bile reflux are recorded, this may be simply a “contaminated” detector or other artifact.

Presence of blood may distort the readings obtained from the device

Underestimation of bile reflux in an acid environment (pH < 3.5)