Manual of AWA6228 I). Calibration

Hangzhou Aihua Instruments Co., Ltd

Page 2: Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd

Functions of keys

Start / Reset Key

Escape Key

Background light Key

Run / Pause Key

Setup Key

Left / Right KeysEnter Key

Delete Key

Up / Down Keys

Page 3: Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd

Boot up meter…

Press the start/reset key for at least 2 seconds to boot up the meter. And enter to the main menu.

Main menu

Page 4: Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd

Calibration…

Press left and right keys to move the cursor to ‘Cali’ and press enter to go to interface of Calibration.

Calibration Window

MIC SerialNo.

Page 5: Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd

MIC SerialNo.

Calibration…

Put on the calibrator and turn it on.Notes: keep the meter and calibrator connected tightly and solidly.

Calibration Window

Press left and right to move the curser to ‘CALI’, and press enter to start calibrating.

Page 6: Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd

MIC Serial

Calibration…

In processing of calibration, the upright corner will shows the # 1-9. (Which means the calibration takes 9 seconds totally.)

MIC SerialNo.

When the digit becomes 9, means the calibration is completed。

MIC SerialNo.No.

MIC Serial

Notes: the LC value should be around 93.8 ±0.2dB. Otherwise, redo the calibration.

After finishing calibration, move cursor to ‘USE ’, press enter to save the setup.

Page 7: Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd

Over 3dB difference …

1) The screen inside the microphone is broken. In this case, the instrument should be sent back to the manufactory to get repaired.

2) Check whether or not the calibrator work normally.

MIC SerialNo.

When the Sensiti and Lpx have over 3dB difference, the new calibration setting won’t be saved. Then please follow the steps to resolve the problem.

Page 8: Manual of AWA6228 I). Calibration Hangzhou Aihua Instruments Co., Ltd

3). We recommend to replace the microphone when you found the Lpx is smaller than sensiti too much. Though, when you are in emergency case, you can try manually adjust the sensitive of microphone. First of all, record down the Lpx value.

Over 3dB difference (Cont.) …

MODI

Press Reset to reset the meter first, then come back to the calibration interface again. Move the cursor to ‘MODI’, press enter. Then use left/right keys to move cursor to ‘Sensiti’ and use up/down keys to adjust the sensitivity to equal the Lpx value that you have recorded down earlier..

MIC SerialNo.