manual gse 460 y 465 sp

40
1 GSE 460 SERIES PROGRAMMABLE WEIGH INDICATORS Revision 1.0 April 2001 Part Number 39-10-37661

Upload: yunnuen-aguilar

Post on 12-Mar-2015

1.230 views

Category:

Documents


73 download

TRANSCRIPT

Page 1: Manual GSE 460 y 465 SP

1

GSE 460 SERIESPROGRAMMABLE WEIGH

INDICATORS

Revision 1.0April 2001

Part Number 39-10-37661

Paul
Logo standard sin Texto
Page 2: Manual GSE 460 y 465 SP

2

GSE Scale SystemsInformation in this manual is subject to change without notice due tocorrection or enhancement. The information described in the guide is solelythe property of GSE, Inc. No part of this manual may be reproduced ortransmitted in any form or by any means – electronic or mechanical, includingphotocopying and recording – nor be sold monetarily without the expresswritten permission of GSE. This prohibition applies to the software, used forcomputer simulation of GSE hardware and its firmware. Setup files that havenot been developed by GSE are neither the property nor the responsibility ofGSE. It is highly recommended that all setup files for the 460 Series indicatorsbe backed on disk in ASCII text format.

Copyright 2001 GSE, Inc.All rights reserved

This user’s guide is intended to be a reference tool for explaining the basicfunctions. Questions concerning the programming, setup, or servicing of thisproduct should be addressed to your GSE distributor.

If you have any questions or need assistance after first reading this guide,please contact a GSE customer service representative at the followingnumber.

Your GSE distributor is:

Page 3: Manual GSE 460 y 465 SP

3

GSE Locations Worldwide

GSE Scale Systems22705 Heslip DriveNovi, MI 48375U.S.A.Phone: (800) 755-7875www.gse-inc.com

GSE Canada, Inc.617 East Lake RoadAirdrie, Alberta Canada T4B 2B8Phone:(403) 948-9921Fax: (403) 948-1449

GSE Advanced IndustrialTechnologies GmbHHausinger Strasse 4Langenfeld, Germany 40764Phone: (49) 21 739 16399-0Fax: (49) 21 739 16399-9

Page 4: Manual GSE 460 y 465 SP

4

Table of ContentsCHAPTER 1 INTRODUCTION....................................................................................7

DISPLAY...........................................................................................................................7MODEL 460 KEYPAD.......................................................................................................8MODEL 465 KEYPAD.......................................................................................................9ALPHA-NUMERIC CHARACTER ENTRY..........................................................................10AVAILABLE OPTIONS ....................................................................................................11

CHAPTER 2 INSTALLATION INSTRUCTIONS ..................................................13

WARNING!..................................................................................................................13TABLE-TOP USE............................................................................................................13PERMANENT MOUNTING...............................................................................................14PANEL MOUNT USE .......................................................................................................15ENVIRONMENTAL SUITABILITY....................................................................................16REMOTE DISPLAY..........................................................................................................16

CHAPTER 3 CALIBRATION MODE.......................................................................17

CALIBRATION MODE.....................................................................................................17QUICK CALIBRATION ....................................................................................................17“NEW ZERO?” CALIBRATION .......................................................................................17"LAST ZERO?" CALIBRATION .......................................................................................18"TEMP ZERO?" CALIBRATION.......................................................................................18"ONLY ZERO?" CALIBRATION ......................................................................................19"CAL RESET”.................................................................................................................19"KNOWN LCOUT”.........................................................................................................20

CHAPTER 4 COUNTING MODE..............................................................................22

PARTS COUNTING..........................................................................................................22COUNTING (FEATURES) ................................................................................................22PARTS COUNTING METHODS ........................................................................................23

Method 1 counting a specific number of parts .......................................................23Counting out a specific number of parts with piece weight enhancement. ..........23Counting the total number of parts on hand with a known container weight......24

PIECE WEIGHT DATABASE MODE (MODEL 465).........................................................25Store APW – Store a specific APW to the database...............................................25Get APW – Retrieve a specific APW from the database........................................25Print APW’s – Prints all stored records in row/column format. ..........................25Clear One APW – Clears current record from the database................................26Clear All APW’s – Clears all records from the database. ....................................26

Page 5: Manual GSE 460 y 465 SP

5

CHAPTER 5 TRUCK WEIGHING (MODEL 465 ONLY) ....................................27

TRUCK IN.......................................................................................................................27TRUCK OUT ...................................................................................................................27PRINT ID’S?...................................................................................................................27CLEAR ONE ID ..............................................................................................................27CLEAR ALL...................................................................................................................28STORE PERMANENT TRUCK IN.....................................................................................28

CHAPTER 6 CHECK WEIGHING AND BATCHING..........................................29

CHECK WEIGH ABSOLUTE AND PERCENTAGE OPERATION.........................................29Set the Hi and Lo weights........................................................................................29Use Check Weigh .....................................................................................................30

BATCHING OPERATION .................................................................................................30Setting setpoints, preact, and dribbles....................................................................30Batching....................................................................................................................30

CHAPTER 7 ACCUMULATION OPERATIONS...................................................31

PERFORMING ACCUMULATIONS ...................................................................................31Set Accumulation......................................................................................................31Get Accumulation.....................................................................................................31View Number of Accumulations ..............................................................................31Preset Accumulation ................................................................................................31Clear Accumulations................................................................................................32

(MODEL 460) ACCUMULATION ....................................................................................33Set Accumulation......................................................................................................33Get Accumulation.....................................................................................................33View Number of Accumulations ..............................................................................33Clear Accumulations................................................................................................33Preset Accumulation ................................................................................................33

CHAPTER 8 PRINTING OPERATIONS..................................................................34

CUSTOM TRANSMIT (MODEL 465 ONLY) ....................................................................34Print (Model 460).....................................................................................................34Print (Model 465).....................................................................................................34

CHAPTER 9 TIME AND DATE .................................................................................35

NOTES ............................................................................................................................35ENTERING TIME AND DATE (MODEL 460)...................................................................35ENTERING TIME AND DATE (MODEL 465)...................................................................35VIEWING THE TIME AND DATE.....................................................................................36

Page 6: Manual GSE 460 y 465 SP

6

Model 465.................................................................................................................36Model 460.................................................................................................................36

CHAPTER 10 TROUBLESHOOTING......................................................................37

ERROR MESSAGES (OVERVIEW )...................................................................................37OPERATIONAL MODE ERROR MESSAGES.....................................................................37HARDWARE PROBLEM ERROR MESSAGES ...................................................................38CALIBRATION ERROR MESSAGES.................................................................................38MISCELLANEOUS MESSAGES........................................................................................39COMMUNICATIONS ERROR MESSAGES ........................................................................39SERVICE.........................................................................................................................40GLOSSARY OF TERMS....................................................................................................40

Page 7: Manual GSE 460 y 465 SP

7

Chapter 1 Introduction

Display

The display for the 460 Series indicators has two sections: a large numericarea on the left, and a smaller two-line-by-10-character dot matrix are on theright side. The six large digits display numeric data. The dot matrix area hasseveral functions. The first two characters of the upper line show the weighingunits of the displayed data, and the last three characters indicate a CENTERZERO ( -->O<-- ) condition, when applicable. The lower line specifies the typeof data, such as Gross, Net, Tare, etc. The dot matrix area also displaysmessages during operation and setup.

Page 8: Manual GSE 460 y 465 SP

8

Model 460 Keypad

The simplified keypad enables operator interface with the Model 460. Keyfunctions are described below. Some of these keys have additional functionsdepending on the specific mode being used.

[ZERO] Press to reset the current weight reading to zero.

[UNITS] Press to select the available displayed units, such as lb, kg,oz, g, etc.

[SELECT] Press to select the programmed operating mode such as NETWeight, TARE Weight, GROSS Weight, Time-Date, etc.

[PRINT] Press to send data to a printer, computer or other device.

[TARE] Press to establish a Tare Weight. This key also doubles as an[ENTER] key.

[UNITS] + These keys are used when making numeric and[PRINT] character entries.

Page 9: Manual GSE 460 y 465 SP

9

Model 465 Keypad

The Model 465 keypad has ten function keys and includes a numeric portionfor the direct entry of numbers. Several of the function keys perform multipletasks. The function depends on which mode of operation the indicator is in.

[ZERO] Press to reset the current weight reading to zero.

[UNITS] Press to select the available displayed units, such as lb, kg,oz, g, etc. This key doubles as the UP arrow key whenscrolling through the ASCII character set.

[SELECT] Press to select the programmed operating mode such as NETWeight, TARE Weight, GROSS Weight, Time-Date.

[SCALE Toggle between enabled scale inputs.SELECT]

[ID] Press to enter, store and use the ID parameters. IDs are usedfor the Truck IN/OUT mode and also standard ID operation.

[TARGET] Initiates macro number 4.

[TARE] Press to establish a Tare Weight. This key doubles as theRIGHT arrow key during data entry, enabling you to move tothe next character position.

Page 10: Manual GSE 460 y 465 SP

10

[PRINT] Press to enter and store ID information. Press to send data toa printer, computer, or other device. This key is also usedwhen making character entries to create Custom Transmits,etc.

[ENTER] Press to accept or enter information to the indicator. Alsodoubles as a [YES] key for conditional operations.

[F1] Initiates macro number 1.

Alpha-Numeric character entry

To enter alphanumeric characters you need an alphanumeric keyboard thatconnects via a serial port to your indicator. Please contact your distributor fora GSE Alpha Numeric Keyboard (Part Number 24574B-401A0).

If your indicator is configured to use alphanumeric characters!���� Function like the arrow keys found on a computer keyboard.

[UNITS � ] can be used as a backspace key[TARE �] advances the cursor to the next character position[TARGET �] scrolls backward through the list of characters[F1 �] scrolls forward through the list of uppercase and lowercase alphabets,

ASCII punctuation, and numeric characters (see figure below for theorder characters are listed)

W

/ hR

~

(space)

}

{zy

(

$ #'& %

"!

|

)

+*

,

.-

] \`

a

_ ^[ZY

eU

dcb

X

V

gf

TS

=

xwvut5

210

43

876

:9

<;

q p o

ms

rn

kO

ji

QP

lNML

C> @ BAH

KJI

ED GF?

Page 11: Manual GSE 460 y 465 SP

11

When the desired character is displayed press the [TARE �] to advance thecursor to the next position where another “A” will be displayed.

Available Options

Hardware options available from GSE provide additional functionality forstandard 460 Series indicators, helping you take advantage of their totalcapabilities. These options can be installed in the field. It is not necessary toreturn the indicator to GSE for installation of these options. Contact your GSEdistributor for part number, price and availability of the options below. Thefollowing options are available for the 460 Series indicators:

Multi-Scale Option:Adds the capability of having two scales.

4-Position I/O Board460 controllers supports the eight following 4-position I/O boards:• Low Voltage AC 2 input 2 output

• High Voltage AC 2 input 2 output• AC 4 output• Low Voltage AC 4 input

• High Voltage AC 4 input• DC 2 input 2 Output• DC 4 output• DC 4 input

Analog Output Option:The capability exists for two analog options to be added.

20 mA Current Loop:Replaces port 1 RS232 communications, TX-active or passive.

2” LCD Remote Display:Allows data to be remotely displayed.

Page 12: Manual GSE 460 y 465 SP

12

RS 485 Option: (Network Module)Addressable networking capability.

Time & Date Option:Adds the capability of battery backed time of day and date.

Alphanumeric Serial KeyboardKeyboard that connects via a serial port allowing easy entry of alphanumeric characters

Dura-ShieldA clear weatherable film lexan that adheres over the keypad anddisplay. Recommended when high-pressure washdown, petroleumdistillates or harsh cleaning agents are present.

Splash GuardsA clear flexible vinyl cover for use in harsh environments. It will keepthe indicator well protected and looking like new.

Page 13: Manual GSE 460 y 465 SP

13

Chapter 2 Installation Instructions

NOTE: in order to connect the load cells and peripherals, the indicator mustbe opened. However, this may void the warranty. Your GSE distributor isavailable to make the proper internal connections.

IMPORTANT! The 460 Series indicators do not include an on/off switch and,therefore, must be installed near a power outlet socket that is easilyaccessible and in keeping with UL/CSA approval requirements.

WARNING!

Make sure the Model 460 or Model 465 indicator is completely disconnectedfrom the power source before working on it. Any operation that involvesopening the indicator enclosure should be performed only by qualified servicepersonnel and only after the power is disconnected! Hazardous voltage isaccessible within the enclosure. Before you begin the installation of your 460Series indicator, please review the guidelines and step-by-step installationprocedures provided in this chapter.

INFORMATION IMPORTANT!Prendre note que les contrôleurs de serie 350 ne sont pas munisd'interrupteurs "Marche / Arrêt". Par conséquent, il devront être installés prèsd'une source d'alimentation secteur accessible pour demeurer sous lesexigences des normes de sécurité UL/CSA.

Table-Top Use

Each of the standard 460 Series indicators comes with a swivel bracket thatpermits you to tilt the face of the instrument to any desired angle. The bottomof the bracket has non-slip rubber pads. These pads prevent the indicatorfrom scratching and sliding across a table surface when you press the keys.

Page 14: Manual GSE 460 y 465 SP

14

Permanent Mounting

Four mounting holes on the swivel bracket enable the indicator to be securedto a fixed surface. The holes are 0.28-in diameter (0.7 mm) to accommodate¼” or M6 metric fasteners.

Figure 1: 460 dimensions

Figure 2: 465 dimensions

Page 15: Manual GSE 460 y 465 SP

15

Panel mount Use

The Model 460 indicator is available in a panel mount version. The panelcutout dimensions are 6.00-6.25” height by 9.00-9.25” width. Once it isinstalled, the panel unit has full washdown capability. The panel mountversion functions just like the tabletop version. Differences are in theenclosure packaging and the positioning of the main board and display.

Figure 3: 460 Panel mount

The Model 465 indicator is available in a panel mount version. The panelcutout dimensions are 8.00-8.25” height by 9.00-9.25” width. Once it isinstalled, the panel unit has full washdown capability. The panel mountversion functions just like the tabletop version. Differences are in theenclosure packaging and the positioning of the main board and display.

Figure 4: 465 Panel mount

Page 16: Manual GSE 460 y 465 SP

16

Environmental Suitability

Each of the standard 460 Series indicators comes in a sealed stainless steelenclosure that can be used in a washdown environment. If the indicator isused in such a setting, be sure the AC power socket outlet is properlyprotected!

The keypad is made of silicon rubber. Specifications are available from GSEfor the reactions of this material to various solvents. The keypad can becleaned with a soft damp non-abrasive cloth and mild detergent. The displaywindow may scratch due to aggressive cleaning, care must be taken to avoidsuch damage.

The display window is made from a polycarbonate material. The gasket for thedisplay is a closed-cell polyethylene foam. The gasket is resistant to water,detergent, and alcohol, but it may be adversely affected by aliphatic andaromatic hydrocarbons.

Remote Display

When setup as a remote display, the 460/465 duplicates the display of theMaster Scale. The 460/465 key-presses are directed to the master scale..To Disable the Remote's keys, disconnect the wire from the Remotes TX onComm 1 or Contact your GSE Distributor to reprogram the 460/465.To partially disable the keyboard, contact your GSE Scale Distributor.

Page 17: Manual GSE 460 y 465 SP

17

Chapter 3 Calibration Mode

Calibration Mode

The 460 Series indicators can be calibrated several ways. The methodsdescribed below assume that the necessary parameters (full-scale value orgraduation size, for example) are selected before the actual calibration isperformed, these parameters normally are set up by a GSE distributor.Information on parameter setup for calibration of the 460 Series indicators canbe found in the 460 Series Technical Reference Manual.

Quick Calibration

Enter the Quick Cal mode from either the Weigh or Counting modes using thefollowing keypad instructions. Entering Quick cal mode (With a 460).

1. Press [ZERO] + [SELECT] simultaneously.2. Press [ZERO] [PRINT] [UNITS] [ENTER].

(With a 465)1. Press/enter [100] [SELECT] [54321] [ID] [ENTER].

“New Zero?” Calibration

Used for a new calibration (establishing the first/new calibration).

1. Please remove any extraneous load.2. Press [SELECT] key until “New Zero” is displayed.3. Press [ENTER] to select “New Zero”.

Note: the display should now read Keyin CalWt.

4. Place the calibration weight on the platform.5. Key in the calibration weight value.6. Press [ENTER]

Note: the display should now read “CAL OK?”

Paul
Highlight
Paul
Highlight
Paul
Highlight
Paul
Highlight
Paul
Highlight
Paul
Highlight
Paul
Highlight
Paul
Highlight
Paul
Highlight
Paul
Highlight
Page 18: Manual GSE 460 y 465 SP

18

7. Press [ENTER] to accept calibration.8. Press [ENTER] to save modifications.9. Press [ENTER] to exit calibration.

"Last Zero?" Calibration

Allows span re-calibration without removing the applied test weight. The lastzero established by pressing [ZERO] from the weigh mode will be used as thezero reference.

1. Press [SELECT] key until “Last Zero?” is displayed.2. Press [ENTER] to select “Last Zero?”.

Note: Indicator briefly displays the units for keying in the calibration weights

3. Key in the applied weight.4. Press [ENTER].

Note: the display should now read “CAL OK?”

5. Press [ENTER] to accept calibration.6. Press [ENTER] to save modifications.7. Press [ENTER] to exit calibration.

"Temp Zero?" Calibration

To Re-Cal (without establishing a new zero) used to perform a calibrationwithout removing the currently applied load.

1 Press [SELECT] key until “Temp Zero?” is displayed1. Press [ENTER] to select “Temp Zero?”.

Note: Indicator briefly displays the units for keying in the calibration weights

2. Add the calibration weight.3. Key in the applied weight.

Paul
Highlight
Paul
Highlight
Page 19: Manual GSE 460 y 465 SP

19

4. Press [ENTER]

Note: the display should now read “CAL OK?”

5. Press [ENTER] to accept calibration.6. Press [ENTER] to save modifications.7. Press [ENTER] to exit calibration.

"Only Zero?" Calibration

To re-establish the indicator’s calibration zero without affecting the establishedgain. Used in weighing applications where the re-zero parameter (P118) is setvery low in order to prevent inadvertent re-zeroing.

1. Press [SELECT] key until “Only Zero?” is displayed.2. Press [ENTER] to select “Only Zero?”.

Note: the display zero’s out and then should read “CAL OK?”

3. Press [ENTER] to accept calibration.4. Press [ENTER] to save modifications.5. Press [ENTER] to exit calibration.

"Cal Reset”

The Cal Reset adjusts the zero and gain factors of the amplifier on the A/D.

1. Press [SELECT] key until “Cal Reset” is displayed.2. Press [ENTER] to select “Cal Reset”.3. Perform a “New Zero?” calibration (Procedure A above).

Note: After performing a Cal Reset, a re-calibration must be performed beforeexiting the calibration or setup modes. The reset will not be saved unless a re-calibration is performed and changes are saved.

Page 20: Manual GSE 460 y 465 SP

20

"Known LCOut”

Known Load Cell is used to calibrate without test weights. The exact full-scalemV/V rating must be known for each load cell. All load cells must be of thesame full-scale capacity.

1. Press [SELECT] until “Known LCOut” is displayed.2. Press [ENTER] to select “Known LCOut”.

Note: The number of load cells specified during the last calibration will also bedisplayed. A value of zero (0) indicates that this calibration method has not yetbeen performed.

3. Enter the number of load cells (8 maximum).

Key in the number of loads cell then press [ENTER]- or -

Press [ENTER] to accept the displayed value.

4. Enter in the load cell’s mV/V value

Key in the load cells mV/V and then press [ENTER]- or -

Press [ENTER] to accept displayed value.

5. The indicator will now automatically repeat steps 3 and 4 for each loadcell.

6. Enter the load cell full scale “LC FS”

Key in the load cells full scale and then press [ENTER]- or -

Press [ENTER] to accept displayed value.7. Establish the current input signal as the zero reference.

1) Press [ENTER] to accept displayed value- or -

2) Press [SELECT] to display Zero=0mVv?3) Press [ENTER] to use a 0mV/V output as the zero reference.

- or -4) Press [SELECT] to display Keyin CurWt.

Page 21: Manual GSE 460 y 465 SP

21

5) Enter the known gross weight already applied to the scale6) Press [ENTER].

- or -7) Press [CLR] to bypass the zeroing option.

8. Press [ENTER] to accept “CAL OK”.9. Press [ENTER] to save modifications.10. Press [ENTER] to exit calibration.

NOTE: If the calibration weight is less than 5% of capacity or if there is a large changein the calibration, the indicator will prompt, ReCal Req'd. Press [ENTER] and you willbe returned to the beginning of the calibration selections, or press [CLR] to obtain the“CAL OK?” Prompt as described above and override the re-calibration requirement.GSE recommends that you perform a re-calibration in order to avoid any seriousinaccuracy.

Page 22: Manual GSE 460 y 465 SP

22

Chapter 4 Counting Mode

Parts Counting

The Counting mode enables the 460 Series indicators to count the number ofparts on a weigh platform. In order to do this, a sampling operation must firstbe performed to establish the average weight of each of the pieces to becounted. However, if the average piece weight value is already known, youcan key it in directly, thus avoid the sampling operation.

Counting (Features)

Your distributor will set the following modes/features to your requirements.They might slightly change the procedures listed below.

• Auto Sample: The indicator will automatically assume the correct numberof pieces has been added to the scale when the scale stabilizes.

• Auto Enhance: Works in conjunction with % Accuracy to allow theindicator to automatically enhance the average piece weight whenever thenecessary criteria has been met

• Sample Size: The number of pieces used to calculate the average pieceweight

• % Accuracy: The required piece weight accuracy for sample acceptance.• Accuracy Displayed: shows the accuracy of the current average piece

weight

Page 23: Manual GSE 460 y 465 SP

23

Parts Counting Methods

Method 1 counting a specific number of parts

1. Press [SELECT] until you are in “Qty” mode2. Place an empty container on the platform (optional)3. Press [ENTER] to tare the indicator.

Note: the display will prompt you to add your sample count “Add 10” (default = 10)

4. Add the specified number of parts to the scale

1.) Add the specified number of sample parts to the scale.2.) Press [ENTER].

- or -

1) Add any number of parts to the scale.2) Key in the number of parts added (Note: if you have specified a

“% Accuracy” you may be prompted to add more pieces at thispoint).

3) Press [ENTER].

5. Place the required number of parts on the scale.

Counting out a specific number of parts with piece weightenhancement.

1. Press [SELECT] until you are in “Qty” mode.2. Place an empty container on the platform (optional).3. Press [ENTER] to tare the indicator.

Note: the display will prompt you to add your sample count “Add 10” (default = 10)

4. Add the specified number of sample parts to the scale.

1) Add the specified number of parts to the scale.2) Press [ENTER].

- or -

Page 24: Manual GSE 460 y 465 SP

24

1) Add any number of parts to the scale.2) Key in the number of parts added (Note: you may be prompted to

add more pieces at this point to reach your “%Accuracy”).3) Press [ENTER].

Note: when the scale stabilizes the indicator will calculate the piece weight of thesample parts and display the maximum number of parts that can be added for a pieceweight enhancement to occur. Then the minimum achieved accuracy will be displayed.If a greater accuracy is desired, add more parts but not more than the maximumenhanceable quantity. As soon as motion ceases, the indicator will recalculate thepiece weight and display the new maximum number of pieces that can be added whilestill accurately enhancing the piece weight.

5. Add parts to the scale, allowing the scale to stabilize for further pieceweight enhancements as necessary.

Counting the total number of parts on hand with a knowncontainer weight.

1. Press [SELECT] until you are in “Qty” mode.2. Place a full container of parts on the platform.3. Press [ENTER] to tare the indicator.

Note: the display will prompt you to add your sample count “Add 10” (default =10)

4. Remove the specified number of sample parts from the scale.

1) Remove the specified number of parts from the scale.2) Press [ENTER].

- or -

1) Remove any number of parts from the scale.2) Key in the number of parts removed.3) Press [ENTER].

(Note: you may be prompted to remove more pieces at this point to reach your“%Accuracy”)

5. Add the removed parts back onto the scale.6. Key-in the tare weight of the container

Page 25: Manual GSE 460 y 465 SP

25

7. Press [ENTER].8. The displayed quantity is the total number of parts in the container.

Piece Weight Database Mode (Model 465)

Store APW – Store a specific APW to the database

1. Press [F1].2. Key-in the ID#.3. Press [ENTER]. Prompt = “store new#?”4. Press [ENTER] to store the ID. Or [CLR] to abort5. Add sample parts (default value = 10) to the scale.6. Press [ENTER] to store APW.

[SELECT] [TARE] or [CLR] will abort the process.

Get APW – Retrieve a specific APW from the database

1. Press [F1].2. Key-in the ID#. If the ID# exists, “Found” will be displayed briefly and the

ID# will be displayed.3. Press [ENTER] to accept entry.4. The pieceweight register is updated with the associated value.

Note: If the ID# does not exist you will be prompted to add default number ofpieces to store a new APW.

[SELECT] [TARE] or [CLR] will abort the process.

Print APW’s – Prints all stored records in row/column format.

1. Press [TARGET].2. Press [CLR/NO] to advance to print mode.3. Press [ENTER] to print stored APW’s in row/column format.4. Indicator returns to the previous display mode.

Page 26: Manual GSE 460 y 465 SP

26

Clear One APW – Clears current record from the database.

1. Press [TARGET].2. Press [ENTER] at the prompt to confirm deletion.

Note: this will only remove the current APW from the database.

Clear All APW’s – Clears all records from the database.

1. Press [TARGET]2. Press [CLR/NO] until you are in Clear ALL?” mode.3. Press [ENTER] to delete all ID# from the database.4. Press [ENTER] at the prompt to confirm deletion.5. Indicator returns to the previous display mode.

Page 27: Manual GSE 460 y 465 SP

27

Chapter 5 Truck Weighing (Model 465 Only)

Truck In

Weigh the vehicle in.

1. With the truck on the scale.2. Press [F1] prompt “KeyIn ID#”.3. Press [ENTER] for Automatic ID# or key-in ID# and press [ENTER].4. Indicator returns to previous display mode.

Note: look at ticket to find the trucks ID# if Automatic ID# is used.

Truck Out

Weigh the vehicle out.

1. With the truck on the scale.2. Press [F1] prompt = “KeyIn ID#”.3. Key-in truck ID#.4. Press [ENTER] prompt flashes “weigh Out”.

Print ID’s?

Prints all ID’s in a row column with header format.

1. Press [TARGET] prompt = “Print ID’s?”.2. Press [ENTER] prompt = “Clear One?”.3. Press [CLR/ NO] prompt = “Clear ALL?”.4. Press [CLR/NO] indicator returns to previous display mode.

Clear One ID

Clears 1 record from the database.

1. Press [TARGET] prompt = “Print ID’s?”.

Page 28: Manual GSE 460 y 465 SP

28

2. Press [CLR/NO] prompt = “Clear One?”.3. Press [ENTER] prompt = “ID to Delet”.4. Key –in ID# to be deleted.5. Press [ENTER].

Clear ALL

Clears all records from the database.

1. Press [TARGET] prompt = “Print ID’s?”.2. Press [CLR/NO] prompt = “Clear One?”.3. Press [CLR/NO] prompt = “Clear ALL?”.4. Press [ENTER] prompt = “Clear Sure”.5. Press [ENTER] indicator returns to previous display mode.

Store Permanent Truck IN

Make a permanent record in the database for a truck tare weight.

1. Press [F1] prompt = “KeyIn ID#”.2. Press [TARE] prompt = “Press tare or KeyIn”.3. Key-In tare weight of truck or have truck on the scale.4. Press [TARE] prompt = “KeyIn ID#”.5. Key-in ID#.6. Press [ENTER].

Page 29: Manual GSE 460 y 465 SP

29

Chapter 6 Check Weighing and Batching

Check Weigh Absolute and Percentage Operation

The standard 460 Series check-weigh operation is designed to check theweight of an item against a given standard or target weight. In the checkweigh absolute mode, values are entered as target limits. In the check weighpercent mode, the values are entered based on percentage.

Set the Hi and Lo weights (M460)

1. Press [SELECT] repeatedly until prompt = “mName SETUP”.2. Press [TARE] prompt =”10.00 Hi” (by default)3. Use the [PRINT] key to enter a number ( [UNITS] key moves placeholder

over 1 spot)4. Press [TARE] prompt = “number entered in step above”.5. Repeat steps 2 – 4 to enter in low valuePress [TARE] indicator returns to previous display mode.

Set the Hi and Lo weights (M465)

6. Press [TARGET] prompt = “10.00 Hi”. (by default)7. Key-in hi weight.8. Press [ENTER] prompt = “5.00 Lo”. (by default)9. Key-in Lo weight.10. Press [ENTER] indicator returns to previous display mode.

Page 30: Manual GSE 460 y 465 SP

30

Use Check Weigh

1. Press [F1] to start checkweighing2. Add your weight to the scale.3. Press [TARGET] key to stop checkweighing (M465)3a . Press [SELECT] until prompt = “mName SETUP”. Then press [TARE] tostop check weighing (M460)

Note: if your weight is greater than your hi weight then the prompt = “OVER” isdisplayed. If your weight is less than your Lo weight the prompt = “UNDER” isdisplayed. If your weight is between the Hi and Lo then prompt = “GOOD” is displayed.

Batching Operation (465 Only)

Setting setpoints, preact, and dribbles

1. Press [TARGET] prompt = “current setpoint value + SP 1”.2. Key-in Setpoint value.3. Press [ENTER] prompt = “PAct.”.4. Key-in Preact value (if enabled) zero (0) for no preact.5. Press [ENTER] prompt = “Drib”.6. Key-in Dribble value.7. Repeat steps 2 – 6 for each setpoint your systems has.8. Systems returns to previous display mode.

Batching

1. Press [ZERO] to zero the indicator2. Press [F1] Prompt = “Done” when cycle completes3. Empty the scale

Page 31: Manual GSE 460 y 465 SP

31

Chapter 7 Accumulation Operations

The indicator contains two memory registers into which weighing data may beaccumulated. The way in which Accumulations operate depends on selectionsmade during the Setup Mode. Accumulation instructions assumes yoursystem has the default setup, please contact your distributor for further detailsor changes.

Performing Accumulations

(Model 465) Accumulation

Set Accumulation

1. Place weight on the scale2. Press [.] prompt = “.”3. Press [ENTER] prompt flashes = “Add Done”

Note: you must first clear the weight from the scale prior to setting the nextaccumulation

Get Accumulation

1. key-in the number three (3) prompt = “3”2. Press [SELECT] display then shows GrTOT3. Press [SELECT] to return to gross mode if so desired

Note: 6 + [SELECT] returns NtTOT

View Number of Accumulations

1. Key-in the number nine (9) prompt = “9”2. Press [SELECT] prompt = “# Accum”3. Press [SELECT] to return to gross mode

Preset Accumulation

1. Key-in the number three (3) prompt = “3”2. Press [SELECT] display then shows GrTOT

Page 32: Manual GSE 460 y 465 SP

32

3. Key-in desired preset weight i.e. 1000 lbs.4. Press [ENTER] prompt = “New ? Total” , “ENTER =New!”5. Press [ENTER] prompt = “GrTOT”

Clear Accumulations

1. Key-in the number three (3) prompt = “3”2. Press [SELECT] display then shows GrTOT3. Press [CLR/NO] prompt = “ENTER =CLR!”

Page 33: Manual GSE 460 y 465 SP

33

(Model 460) Accumulation

Set Accumulation

1. Press [SELECT] to enter the accumulation mode prompt = “GrTOT”2. Place weight on the scale3. Press [PRINT] prompt = “.”4. Press [ENTER] prompt flashes “Add Done”

Get Accumulation

1. Press [SELECT] to enter the accumulation mode prompt = “GrTOT”2. Press [PRINT] until the prompt equals the number three (3) prompt = “3”3. Press [SELECT] display then shows GrTOT4. Press [SELECT] to return to gross mode if so desired

View Number of Accumulations

1. Press [SELECT] to enter the accumulation mode prompt = “GrTOT”2. Press [PRINT] until the prompt equals the number nine (9) prompt = “9”3. Press [SELECT] prompt = “# Accum”4. Press [SELECT] to return to gross mode

Clear Accumulations

1. Press [SELECT] to enter the accumulation mode prompt = “GrTOT”2. Press [CLEAR] + [ENTER] simultaneously prompt = “ENTER =CLR”3. Press [ENTER]

Preset Accumulation

1. Press [SELECT] to enter the accumulation mode prompt = “GrTOT”2. Press [PRINT] until the prompt equals the number three (3) prompt = “3”3. Press [SELECT] display then shows GrTOT4. Key-in desired preset weight i.e. 1000 lbs.5. Press [ENTER] prompt = “New ? Total” , “ENTER =New!”6. Press [ENTER] prompt = “GrTOT”

Page 34: Manual GSE 460 y 465 SP

34

Chapter 8 Printing Operations

Pressing the [PRINT] key will transmit the information in the custom transmit.The custom transmit refers to a table that can be customized to yourapplication. The 460 series indicators are configured with a default factorytransmission, shown below. If this is not the desired transmission formatcontact your GSE distributor.

9876.54 lbs Gross Wt.9864.20 lbs Net Wt. 12.34 lbs Tare Wt.

Custom Transmit (Model 465 Only)

Four independent custom transmit tables permit the RS232 output to beconfigured for an format required by a printer, computer, remote display orany serial device. Two (2) RS232 serial ports are available to send the customtransmit tables. This permits communication with a variety of serial devices forsystems integration.

Print (Model 460)

1. Press [PRINT] – the custom transmit will send the signal via either serialport

Print (Model 465)

1. Press [PRINT] – The custom transmit will send the signal via one of the(2) serial ports. (the unit may be set up to prompt a user to choose whichcustom transmit to send after pressing the [PRINT] button.

Note: The continuous print will transmit the custom-programmed information eachtime the display is updated. If the receiving device (printer, display or computer) cannotreceive the data being sent for any reason, the message "Tx On Hold" will appear for afew seconds. Press [CLR] to abort the transmission. If this situation happens whileusing the continuous print feature, the transmission is suspended, but can be resumedby pressing the [PRINT] key.

Page 35: Manual GSE 460 y 465 SP

35

Chapter 9 Time and Date

Notes

1. Both models have a battery backed time/date option available, pleasecontact your GSE distributor for more details

2. If the battery backed time/date option is not installed, the time and datewill need to be reset upon power-up. If the clock is not reset at power-up,the time will be the elapsed time since power-up and the date will be01/01/70.

3. To enter PM times please add 12 hours to the actual timeI.e. 9:12 PM is entered as 21.12

4. International time date format is available and should be setup by yourGSE distributor

Entering Time and Date (Model 460)

1. Press [SELECT] until the time/date is displayed2. Press [ENTER] prompt = “date”3. Using a combination of the [PRINT] and [UNITS] keys enter in the date

a. Format mm.dd.yy (decimal separated)4. Press [ENTER] prompt = “Date”5. Press [SELECT] prompt = “time”6. Using a combination of the [PRINT] and [UNITS] keys enter in the time

a. Format hh.mm.ss or hh.mm (decimal separated)7. Press [ENTER] new time date is displayedNote: if you want to only adjust the time press [SELECT] at step 2

Entering Time and Date (Model 465)

1. Key-in the number eleven (11)2. Press [SELECT] date time is displayed3. Key-in the date

a. Format mm.dd.yy (decimal separated)4. Press [ENTER] prompt = “Date”5. Press [SELECT] prompt = “time”6. Key-in the time

a. Format hh.mm.ss or hh.mm (decimal seperated)

Page 36: Manual GSE 460 y 465 SP

36

7. Press [ENTER] prompt = “time”8. Press [SELECT] time date is displayed9. Press [SELECT] to accept the time date enteredNote: if you want to only adjust the time press [SELECT] at step 2

Viewing the Time and Date

Model 465

1. Key-in eleven (11)2. Press [SELECT]

Model 460

1. Press [SELECT] until you are in the time date mode

Page 37: Manual GSE 460 y 465 SP

37

Chapter 10 Troubleshooting

Error Messages (Overview)

The following is a summary of some of the error messages within theIndicator.

Operational Mode Error Messages

02 UnderLoad! Input signal less than negative full scale. Possibly due toexcessive loading, reduce the load. Otherwise, check theload cell connections.

03 Over-Load! Input signal (weigh applied to scale) is greater thanpositive full scale (scale capacity). Use same check as forunderload.

04 # > Dsply Number to be displayed will not fit within 6 digits. This willnot normally occur for the Gross, Net or Tare Weights butmay result while displaying the accumulated totals if theamount exceeds 999,999. Either clear the totals or settlefor only being able to transmit the totals.

05 Zero> Max.! An attempt was made to zero out more weight thanallowed. Use the [TARE] key for subtracting off containerweights or if large dead-load is always to be present,apply this dead-load during the No Load? prompt duringcalibration to permanently eliminate the offset.

06 Tare>F.S.! Tare entry was greater than full scale. Most likely theentered tare value was incorrect.

Page 38: Manual GSE 460 y 465 SP

38

Hardware Problem Error Messages

If any of the following errors occur please contact your GSE distributor

21 EEROMerror Error reading data from the EEPROM. (Call distributor)

22 EEROMerror Error writing data to the EEPROM. (Call distributor)

24 EEROMFull! The setup being attempted requires more EEPROM thanis currently installed. (Call distributor)

25 DefltSetup Upon power-up the Indicator has not found the propercodes. Therefore all parameters have been reset tofactory default values.

27 RE-BOOT! The Indicator cannot use the EEPROM for data storage,so it is attempting to power-up again to cure the problem.

28 NoRAMAVAIL The current setup requires more RAM than is currentlyinstalled. Either contact your dealer or the factory.

Calibration Error Messages

30 F.S.>MAX! The entered calibration weight, together with the currentlyapplied signal, indicates that the full-scale signal will begreater than the allowed maximum of the Indicator. Verifythat correct entries have been made for the capacity andthe calibration weight.

31 F.S.<.1mVv The entered calibration weight, together with the currentlyapplied signal, indicates that the full scale signal will beless than the allowed minimum of the Indicator.

32 ADD MORE! The applied weight during calibration was less than 0.1%of capacity. More weight than this is required to calibrate.

33 ReCALReq'd The just completed calibration is insufficient to guaranteeaccurate results due to either the cal weight being lessthan 5% of capacity or this was the first calibration of this

Page 39: Manual GSE 460 y 465 SP

39

platform to this Indicator and therefore the coarse gainwas adjusted by the Indicator. Press [ENTER] to acceptthis calibration or [CLR] to re-calibrate.

Miscellaneous Messages- - EntryError Primary causes are entering a value preceding a key

(such as [ZERO]) which is not allowed, entering alphadata for a numeric selection, or entering a fractional valuefor an entry which only accepts whole numbers..

Communications Error Messages

par'y error This indicates that the parity of a received character didnot match the parity specified . This could also result ifthe baud rate or the number of data bits is incorrect.

ovrun error This indicates an over-run error where an additionalcharacter was received while the receive buffer of the 460Series indicator was full, and thus the extra receivedcharacter will be lost.

frm'g error This indicates that the stop bit of a received character didnot occur when it was expected. This could be the resultof an incorrect baud rate, incorrect number of data bits, orincorrect parity setting.

port error This indicates that the 460 Series indicator did not checkits receive data register in time, thus missing a character.If this error should occur, please notify your GSEdistributor or the factory.

tx on hold This will occur if a data transmission is held up for twoseconds of more due to a de-asserted handshake. Referto the description of parameter P209 for moreinformation.

Page 40: Manual GSE 460 y 465 SP

40

tx abort This occurs if the [CLR] key is pressed when the tx onhold error message is shown.

tx con'd This will appear briefly when the handshake isre-asserted after the tx on hold message occurs.

Service

There are no user-serviceable items in GSE 460 series indicators! Servicemust be performed by qualified service technicians only! Attempts to servicethis instrument by unqualified personnel may void the warranty!

Glossary of Terms

For a glossary of industry terms please see our websitehttp://www.gse-inc.com