  • file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.04.2014 printed 15.04.2014 page: 1 of 16

    running machines operat ing and service manual

    Following chapters of the provided h/p/cosmos operating and service manual have to be replaced by the content of this

    attachement; for these chapters please consider the content of this attachement only.

    Foreword, chapter 1 model overview, chapter 11 technical data, chapter 12 options and accessories, chapter 17 contact.

    treadmill ergometer

    T150

    T170

    T200

    manual-version

    MCU5-v1.07

    revision 15.04.2014

    order number:

    [cos14310m5-v1.07cme-en]

    sales & service

    COSMED S.r.l.

    Via dei Piani di Monte Savello, 37

    IT 00040 Pavona di Albano - Rome

    Italy

    phone +39 06 9 31 54 92

    fax +39 06 9 31 45 80

    info@cosmed.it

    www.cosmed.it

    development & production

    h/p/cosmos sports & medical gmbh

    Am Sportplatz 8

    DE 83365 Nussdorf-Traunstein

    Germany

    phone +49 86 69 86 42 0

    fax +49 86 69 86 42 49

    service@h-p-cosmos.com

    www.h-p-cosmos.com

  • file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.04.2014 printed 15.04.2014 page: 2 of 16

  • file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.04.2014 printed 15.04.2014 page: 3 of 16

    Dear customer, We would like to express our gratitude for putting your trust in us, in deciding for this top of the range running machine. Since 1988 h/p/cosmos has been developing and manufacturing running machines, systems and accessories for applications in sports, rehabilitation, medicine, diagnostics and science. When it comes to technology, ergonomy, design and safety, we have set extremely high standards for ourselves.

    Because the running machine is a motor-driven device, you must pay special attention to the mentioned safety regulations. If proper notice is taken of the safety regulations, the operation of our running machines is nearly without risk. Neglect of the safety regulations could result in dangerous situations and accidents with serious injury or death. Therefore please read this operating and service manual and the danger precautions in full before taking the device into operation. Some simple maintenance and monitoring (no repair work!), as described, can easily be done or even have to be done by yourself. All kinds of installation and repair work and most maintenance work are to be performed only by trained and authorized technicians who have been certified by h/p/cosmos. The following symbols will indicate which work can be done by the customer and which work has to be done by authorized technicians only:

    The customer/user should perform this maintenance and monitoring work. Some safety checks or monitoring (for example harnesses and ropes, running belt condition and position, etc.) have to be performed on daily basis. It is not expedient to contract certified technicians for such maintenance work. However, where it is practical, all maintenance and monitoring work marked with this symbol can also be performed by certified technicians also.

    All installation, maintenance, repair and monitoring work marked with this symbol must only be performed by trained and authorized technicians who have been certified by h/p/cosmos. Customers/users must not perform these kinds of tasks and work.

    We recommend calling our competent service team or entering into a maintenance contract for a routine service at an interval of 6 or 12 months for standard machines and standard applications. A form for registration of your institution and device is included in the delivery. In order to be able to supply you with the latest technical information and service, it is important for you to fill out the form. Therefore please fill out the form for registration immediately and send it back via fax, email or mail. This and the manufacturers operating and service manual as a firm part of the delivery have to be accessible to the user at any time. They have been written with great care. Should you, however, still find any details which do not correspond with your device, please notify us so that we can correct any mistakes as soon as possible. Subject to alterations without prior notice. Errors and omissions excepted. We wish you a lot of fun and success while exercising and working with your h/p/cosmos running machine.

    h Franz Harrer President & CEO h/p/cosmos sports & medical gmbh

  • file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.04.2014 printed 15.04.2014 page: 4 of 16

    1 Model overview

    n

    medical running machines

    c sports & fitness running machines

    T150 D med T150 DE med

    T150 D T150 DE

    T170 D med T170 DE med T170 DE sportmed

    T170 D T170 DE

    T200 S T200 M

    n.a.

  • file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.04.2014 printed 15.04.2014 page: 5 of 16

    11 Technical data

    11.1 Treadmill ergometer, 150/50 (T150)

    T150 DE med T

    150 D

    T150

    DE

    T150

    D m

    ed

    T150

    DE

    med

    UserTerminal (display & keyboard) 6 LCD displays / 6 keys / display of mode and unit with LED.

    Power supply AC (standard, read nameplate), dedicated line required.

    U: 100 V ~ / 110 V ~ / 200 V ~ / 208 V ~ / 230 V ~ including N and PE

    f: 50/60 Hz Fuse (standard, read nameplate) C 16 A at 230 V ~Power input (long time) 1500 VA

    Power input (momentary) 3400 VA(peak performances for milliseconds; max. 1s/min during heel strike) Drive motor capacity 3300 VA Elevation motor capacity n.a. 390 VA n.a. 390 VA

    Safety standards

    c

    IEC EN 60335-1(VDE 0700) IEC EN 60601-1-2,

    IEC EN 60601-1-4, VDE 0701, EN 957-1, EN 957-6, Machinery Directive 2006/42/EG

    n

    MDD, directive 93/42/EEC + 2007/47/EC, IEC EN 60601-1 (VDE 0750), VDE 0751, IEC EN 60601-1-2, IEC EN 60601-1-4,

    EN 957-1, EN 957-6, Machinery Directive 2006/42/EC Leakage current 0.6 mA 0.2 mA Isolation transformer --- 2000 VA Degree of protection: class I M / --- / IP 20 class I M / type B R / IP 20

    Overvoltage category Life-Earth < 300 Vtransient over-voltage limit 1500 V Category II: Life-Earth < 300 V

    transient over-voltage limit 2500 V Mode of operation continuous operation with intermittent loading (according to IEC 60601-1) Field of application / accuracy sports and fitness: S,I,A (EN957) sports and medical: S,I,A (EN957)

    Classification according to MDD --- active therapeutic device and active diagnostic device, risk class II b Speed Definition

    0.0 ... 22.0 km/h0.1 km/h

    Speed accuracy* +/- 5 %Option special speed 0.0 ... 10.0 km/h or ... 30.0 km/hAngle of inclination** Definition n.a.

    0.0 ... 25.0 %**0.1 % n.a.

    0.0 ... 25.0 %**0.1 %

  • file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc 2014 h/p/cosmos sports & medical gmbh email@h-p-cosmos.com created 15.04.2014 printed 15.04.2014 page: 6 of 16

    T150 DE med T

    150 D

    T150

    DE

    T150

    D m

    ed

    T150

    DE

    med

    Dimension of running surface length x width 1500 x 500 mm

    Dimension frame length x width x height

    2100 x 800 x 1370 mm(UserTerminal + 150 mm width)

    Track access height from floor 180 mmWeight of device 200 kg 220 kg

    Weight load limits to floors (traffic load to DIN 1055 part 3)

    8.2 kN/m 8.3 kN/m

    In case of any further questions related to load limits onto floors, please ask our service department. Max. permissible user weight 200 kg (deviation on request specially made)

    Environmental conditions: Transport and storage

    temperature: 30...+50 C,humidity: 0...95 % - without condensation

    barometric pressure: 700...1060 hPa

    Environmental conditions: Operation

    temperature: +10...+40 C (deviation on request), humidity: 30...70% - without condensation (up to 95% on request),

    barometric pressure: 700...1060 hPa Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization

    Pollution degree Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.Central lubrication system manualNumber of stored programmes 6 training profiles / 10 test profiles / 8 user definable profiles

    RS232 Digital interface / serial port (USB-converter optional)

    COM1: standardCOM2: optional

    COM3: optional (standard for service) COM4: optional

    Heart rate measurement POLAR, 1 channel receiver, ECG preciselyPC software (not for medical applications)

    h/p/cosmos para control includedoptions: h/p/cosmos para graphics, h/p/cosmos para analysis and h/p/cosmos para motion

    Reverse belt rotation optionalSafety arch with chest belt system*** optional

    Handrails options: short / long / adjustable handrails, crossbar

    Arm support

Recommended