T150
T170
T200
manual-version
MCU5-v1.07
revision 15.04.2014
order number:
[cos14310m5-v1.07cme-en]
sales & service
COSMED S.r.l.
Via dei Piani di Monte Savello, 37
IT 00040 Pavona di Albano - Rome
Italy
phone +39 06 9 31 54 92
fax +39 06 9 31 45 80
www.cosmed.it
development & production
h/p/cosmos sports & medical gmbh
Am Sportplatz 8
DE 83365 Nussdorf-Traunstein
Germany
phone +49 86 69 86 42 0
fax +49 86 69 86 42 49
www.h-p-cosmos.com
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 2 of 16
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 3 of 16
Dear customer, We would like to express our gratitude for putting your trust in us, in deciding for this top of the range running machine. Since 1988 h/p/cosmos® has been developing and manufacturing running machines, systems and accessories for applications in sports, rehabilitation, medicine, diagnostics and science. When it comes to technology, ergonomy, design and safety, we have set extremely high standards for ourselves.
Because the running machine is a motor-driven device, you must pay special attention to the mentioned safety regulations. If proper notice is taken of the safety regulations, the operation of our running machines is nearly without risk. Neglect of the safety regulations could result in dangerous situations and accidents with serious injury or death. Therefore please read this operating and service manual and the danger precautions in full before taking the device into operation. Some simple maintenance and monitoring (no repair work!), as described, can easily be done or even have to be done by yourself. All kinds of installation and repair work and most maintenance work are to be performed only by trained and authorized technicians who have been certified by h/p/cosmos. The following symbols will indicate which work can be done by the customer and which work has to be done by authorized technicians only:
The customer/user should perform this maintenance and monitoring work. Some safety checks or monitoring (for example harnesses and ropes, running belt condition and position, etc.) have to be performed on daily basis. It is not expedient to contract certified technicians for such maintenance work. However, where it is practical, all maintenance and monitoring work marked with this symbol can also be performed by certified technicians also.
All installation, maintenance, repair and monitoring work marked with this symbol must only be performed by trained and authorized technicians who have been certified by h/p/cosmos. Customers/users must not perform these kinds of tasks and work.
We recommend calling our competent service team or entering into a maintenance contract for a routine service at an interval of 6 or 12 months for standard machines and standard applications. A form for registration of your institution and device is included in the delivery. In order to be able to supply you with the latest technical information and service, it is important for you to fill out the form. Therefore please fill out the form for registration immediately and send it back via fax, email or mail. This and the manufacturer’s operating and service manual as a firm part of the delivery have to be accessible to the user at any time. They have been written with great care. Should you, however, still find any details which do not correspond with your device, please notify us so that we can correct any mistakes as soon as possible. Subject to alterations without prior notice. Errors and omissions excepted. We wish you a lot of fun and success while exercising and working with your h/p/cosmos running machine.
h Franz Harrer President & CEO h/p/cosmos sports & medical gmbh
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 4 of 16
1 Model overview
n
medical running machines
c sports & fitness running machines
T150 D med
T150 DE med T150 D T150 DE
T170 D med
T170 DE med
T170 DE sportmed
T170 D T170 DE
T200 S
T200 M n.a.
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 5 of 16
11 Technical data
11.1 Treadmill ergometer, 150/50 (T150)
T150 DE med T1
50 D
T150
DE
T150
D m
ed
T150
DE
med
UserTerminal (display & keyboard)
6 LCD displays / 6 keys / display of mode and unit with LED.
Power supply AC (standard, read nameplate), dedicated line required.
U: 100 V ~ / 110 V ~ / 200 V ~ / 208 V ~ / 230 V ~ including N and PE
f: 50/60 Hz Fuse (standard, read nameplate) C 16 A at 230 V ~Power input (long time) 1500 VA
Power input (momentary) 3400 VA
(peak performances for milliseconds; max. 1s/min during heel strike) Drive motor capacity 3300 VA Elevation motor capacity n.a. 390 VA n.a. 390 VA
Safety standards
c
IEC EN 60335-1(VDE 0700) IEC EN 60601-1-2,
IEC EN 60601-1-4, VDE 0701, EN 957-1, EN 957-6, Machinery Directive 2006/42/EG
n
MDD, directive 93/42/EEC + 2007/47/EC, IEC EN 60601-1 (VDE 0750), VDE 0751, IEC EN 60601-1-2, IEC EN 60601-1-4,
EN 957-1, EN 957-6, Machinery Directive 2006/42/EC Leakage current 0.6 mA 0.2 mA Isolation transformer --- 2000 VA
Degree of protection: class I M / --- / IP 20 class I M / type B R / IP 20
Overvoltage category Life-Earth < 300 V
transient over-voltage limit 1500 V Category II: Life-Earth < 300 V
transient over-voltage limit 2500 V Mode of operation continuous operation with intermittent loading (according to IEC 60601-1) Field of application / accuracy sports and fitness: S,I,A (EN957) sports and medical: S,I,A (EN957)
Classification according to MDD --- active therapeutic device and
active diagnostic device, risk class II b Speed Definition
0.0 ... 22.0 km/h0.1 km/h
Speed accuracy* +/- 5 %Option special speed 0.0 ... 10.0 km/h or ... 30.0 km/hAngle of inclination** Definition
n.a. 0.0 ... 25.0 %**
0.1 % n.a.
0.0 ... 25.0 %**0.1 %
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 6 of 16
T150 DE med T1
50 D
T150
DE
T150
D m
ed
T150
DE
med
Dimension of running surface length x width
1500 x 500 mm
Dimension frame length x width x height
2100 x 800 x 1370 mm(UserTerminal + 150 mm width)
Track access height from floor 180 mmWeight of device 200 kg 220 kg
Weight load limits to floors (traffic load to DIN 1055 part 3)
8.2 kN/m² 8.3 kN/m²
In case of any further questions related to load limits onto floors, please ask our service department.
Max. permissible user weight 200 kg (deviation on request specially made)
Environmental conditions: Transport and storage
temperature: –30...+50 °C,humidity: 0...95 % - without condensation
barometric pressure: 700...1060 hPa
Environmental conditions: Operation
temperature: +10...+40 °C (deviation on request), humidity: 30...70% - without condensation (up to 95% on request),
barometric pressure: 700...1060 hPa Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
Pollution degree Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.Central lubrication system manualNumber of stored programmes 6 training profiles / 10 test profiles / 8 user definable profiles
RS232 Digital interface / serial port (USB-converter optional)
COM1: standardCOM2: optional
COM3: optional (standard for service) COM4: optional
Heart rate measurement POLAR, 1 channel receiver, ECG preciselyPC software (not for medical applications)
h/p/cosmos para control® includedoptions: h/p/cosmos para graphics®, h/p/cosmos para analysis® and h/p/cosmos para motion®
Reverse belt rotation optionalSafety arch with chest belt system***
optional
Handrails options: short / long / adjustable handrails, crossbar
Arm support optional
Additional keyboard optional
* Overload and weak power supply (under voltage) could lead to higher discrepancies concerning the speed display or dripping of the fuse.
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
*** For maximum stress tests and with oversize running machines (track sizes) a safety facility (e.g. a safety arch with chest belt system) is necessary and highly recommended. Data of weight and measurements can deviate.
Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & OE.
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 7 of 16
11.2 Treadmill ergometer, 170/65 (T170)
T170 DE med T1
70 D
T170
DE
T170
D m
ed
T170
DE
med
T170
DE
spor
tmed
UserTerminal (display & keyboard)
6 LCD displays / 6 keys / display of mode and unit with LED.
Power supply AC (standard, read nameplate), dedicated line required.
U: 100 V ~ / 110 V ~ / 200 V ~ / 208 V ~ / 230 V ~ including N and PE
f: 50/60 Hz Fuse (standard, read nameplate) C 16 A at 230 V ~Power input (long time) 1600 VA
Power input (momentary) 3500 VA
(peak performances for milliseconds; max. 1s/min during heel strike) Drive motor capacity 3300 VAElevation motor capacity n.a. 500 VA n.a. 500 VA
Safety standards
c
IEC EN 60335-1(VDE 0700) IEC EN 60601-1-2,
IEC EN 60601-1-4, VDE 0701, EN 957-1, EN 957-6, Machinery Directive
2006/42/EG
n
MDD, directive 93/42/EEC + 2007/47/EC, IEC EN 60601-1 (VDE 0750), VDE 0751, IEC EN 60601-1-2, IEC EN 60601-1-4,
EN 957-1, EN 957-6, Machinery Directive 2006/42/EC
Leakage current 0.6 mA 0.2 mA Isolation transformer --- 2000 VA
Degree of protection: class I M / --- / IP 20 class I M / type B R / IP 20
Overvoltage category Life-Earth < 300 V
transient over-voltage limit 1500 V Category II: Life-Earth < 300 V
transient over-voltage limit 2500 V Mode of operation continuous operation with intermittent loading (according to IEC 60601-1) Field of application / accuracy sports and fitness: S,I,A (EN957) sports and medical: S,I,A (EN957)
Classification according to MDD --- active therapeutic device and
active diagnostic device, risk class II b Speed Definition
0.0 ... 30.0 km/h0.1 km/h
0.0 ... 40.0 km/h0.1 km/h
Speed accuracy* +/- 5 %
Option special speed 0.0 ... 10.0 km/h, 0.0 ... 40.0 km/h, 0.0... 45.0 km/h 0.0 ... 10.0 km/h0.0... 45.0 km/h
Angle of inclination** Definition
n.a. 0.0 ... 25.0 %
0.1 % n.a.
0.0 ... 25.0 % 0.1 %
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 8 of 16
T170 DE med T1
70 D
T170
DE
T170
D m
ed
T170
DE
med
T170
DE
spor
tmed
Dimension of running surface length x width
1700 x 650 mm
Dimension frame length x width x height
2300 x 1050 x 1410 mm(UserTerminal + 150 mm width)
Track access height from floor 230 mmWeight of device 310 kg 315 kg 330 kg 335 kg Weight load limits to floors (traffic load to DIN 1055 part 3)
6.2 kN/m²In case of any further questions related to load limits onto floors, please ask our service department.
Max. permissible user weight 200 kg (deviation on request specially made)
Environmental conditions: Transport and storage
temperature: –30...+50 °C,humidity: 0...95 % - without condensation
barometric pressure: 700...1060 hPa
Environmental conditions: Operation
temperature: +10...+40 °C (deviation on request), humidity: 30...70% - without condensation (up to 95% on request),
barometric pressure: 700...1060 hPa Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
Pollution degree Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.Central lubrication system manualNumber of stored programmes 6 training profiles / 10 test profiles / 8 user definable profiles
RS232 Digital interface / serial port (USB-converter optional)
COM1: standardCOM2: optional
COM3: optional (standard for service) COM4: optional
Heart rate measurement POLAR, 1 channel receiver, ECG precisely
PC software (not for medical applications)
h/p/cosmos para control® included options: h/p/cosmos para graphics® and h/p/cosmos para analysis®
para control®,para graphics®
options: para analysis®, para motion®
Reverse belt rotation optional standardSafety arch with chest belt system***
optional
Handrails options: short, long, crossbar (standard)Arm support optionalAdditional keyboard optional
* Overload and weak power supply (under voltage) could lead to higher discrepancies concerning the speed display or dripping of the fuse.
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
*** For maximum stress tests and with oversize running machines (track sizes) a safety facility (e.g. a safety arch with chest belt system) is necessary and highly recommended. Data of weight and measurements can deviate.
Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & OE.
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 9 of 16
11.4 Treadmill ergometer, 200/75 up to 300/125 (T200, T250, T300)
T200
T200 S T200 M
UserTerminal (display & keyboard)
External control unit with TouchScreen
Power supply AC (standard, read nameplate), dedicated line required.
200 V 3N~ / 208 V 3N~ / 400 V 3N~with N and PE
(peak performances for milliseconds; max. 1s/min during heel strike) Fuse (standard, read nameplate) 3 x C 32 A at 400 V 3N~Power input (long time) 7500 VAPower input (momentary) 15200 VADrive motor capacity 11000 VAElevation motor capacity 370 VA
Safety standards
n
MDD, directive 93/42/EEC + 2007/47/EC, IEC EN 60601-1 (VDE 0750), IEC EN 60601-1-2, IEC EN 60601-1-4
VDE 0751, EN 957-1, EN 957-6, Machinery Directive 2006/42/EC Leakage current 0.25 mAIsolation transformer 3 x 3000 VA
Degree of protection: class I M / type B R / IP 20
Overvoltage category Category III: Life-Earth < 300 Volt = transient over-voltage limit 4000 Volt Mode of operation continuous operation with intermittent loading (according to IEC 60601-1)
Field of application / accuracy sports and medical / S,I,A (EN957)
[r] = suitable for bicycle and wheelchair application (from running surface 3000 x 1250 mm standard) Classification according to MDD active therapeutic device and active diagnostic device, risk class II b Speed Definition
0.0 ... 40.0 km/h / 0.1 km/h0.0 ... 11.11 m/s / 0.1 m/s
Speed accuracy* +/- 5 % (class A based on EN 957)
Option special speed 0.0 ... 30.0 or ... 50.0 or ... 60.0 or ... 80.0 km/h
on request higher speed versions than 40 km/h only for sports models Angle of inclination** Definition
-35.0 % ... +35.0 %0.1 %
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 10 of 16
T200 T200
S
T200
M
Dimension of running surface length x width
2000 x 750 2000 x 1000
Dimension frame length x width x height
2400 x 1150 x 1380 2400 x 1400 x 1380
Track access height from floor 480 mmWeight of device 850 kg 900 kg
Weight load limits to floors (traffic load to DIN 1055 part 3)
9.0 kN/m² 7.5 kN/m²
In case of any further questions related to load limits onto floors, please ask our service department.
Max. permissible user weight 280 kg
(deviation on request specially made)
Environmental conditions: Transport and storage
temperature: –30...+50 °C,humidity: 0...95 % - without condensation
barometric pressure: 700...1060 hPa
Environmental conditions: Operation
temperature: +10...+40 °C (deviation on request), humidity: 30...70% - without condensation (up to 95% on request),
barometric pressure: 700...1060 hPa Maximum operating altitude: approx. 10,000 feet (3000m), without pressurization
Pollution degree Degree 2: Normally only nonconductive pollution. Temporary conductivity caused by condensation is to be expected.Central lubrication system manual / running machines with [r] have automatic central lubrication system Number of stored programmes 6 training profiles / 10 test profiles / 8 user defined profiles RS232 Digital interface / serial port (USB-converter optional)
COM1: standard; COM2: standard; COM3: n/a (standard for service); COM4: n/a
Heart rate measurement POLAR W.I.N.D. with WearLink
coded transmission transmission range approx. 10 m
PC software (not for medical applications)
h/p/cosmos para control® and h/p/cosmos para graphics® included option: h/p/cosmos para analysis®, h/p/cosmos para motion®
Reverse belt rotation standardSafety arch with chest belt system***
obligatory option (or alternative fall stop preventive system)
Handrails optional
Wheelchair stabilizer optional
Additional keyboard optional
* Overload and weak power supply (under voltage) could lead to higher discrepancies concerning the speed display or dripping of the fuse.
** The time of operation of the elevation motor (TO = 15 %) converts to approx. 5 full elevation cycles. Overload can result in temporary tripping of the fuse, but can be used again after short cool down period. Treadmills of the h/p/cosmos saturn series built approx. before January 2007 have a maximum elevation of 25 %.
*** For maximum stress tests and with oversize running machines (track sizes) a safety facility (e.g. a safety arch with chest belt system) is necessary and highly recommended. Data of weight and measurements can deviate.
Ask for further details and optional equipment or visit www.h-p-cosmos.com. Subject to technical alterations without prior notice. E & OE.
Chapter 11.3 does not apply, chapter 11.5 up to 11.8 see h/p/cosmos operating and service manual.
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 11 of 16
12 Options and accessories
The following chapter describes the accessories and options available at COSMED.
Only connect and/or use accessories, software and host equipment if it is approved as compatible by all manufacturers. Non-compliance with this clause can lead to serious injury or death. Furthermore, manufacturer warranty and liability expires. Pay special attention to all safety regulations and operation instructions in the manufacturer’s operating and service manual.
OEM options order number article description
COSMED handrail long 150/50 cos100114
COSMED handrail long 170/65
handrails upgrade (during production) order number higher prices than the upgrade prices are applicable for retrofitting
of options and accessories ***
Handrail crossbar 50/36 arched with grip cover cos16571 crossbar with fixation clamps for handrail D=60 mm, for models with
running surface 150/50 cm
Handrail crossbar 65/36 arched with grip cover cos15133-02 crossbar with fixation clamps for handrail D=60 mm, for models with
running surfaces 170&190/65 cm
Handrail long/short/rolled sep.1/3 200 cos14190 2/3 of the rear length of the handrails are detachable for running surface
200 cm long
Bottle holder for handrail (D=60 mm) cos11020 the bottle holder is suitable for all standard drinking bottles (1x included
with every treadmill)
additional keyboard & accessories order number
h/p/cosmos additional keyboard cos100680 6 keys with wire connection; works parallel to user terminal keys - without
mounts
Holder for optional keyboard on handrail (D=60
mm)
cos14135 holder for the additional keyboard on the handrail (D=60 mm)
Rear panel terminal for additional options cos10109 for running machines with user terminal (free with safety arch, additional
keyboard, extra safety stop)
reverse belt rotation order number
Reverse belt rotation (downhill) 150/50 cos00098100045 for models with running surface 150/50 cm, incl. safety-stop waist belt &
magnet switch
Reverse belt rotation (downhill) 170/65 cos10181-01 for models with running surface 170/65 cm, incl. safety-stop waist belt &
magnet switch
sepcial speed * order number higher prices than the upgrade prices are applicable for retrofitting
of options and accessories ***
Special speed 0 - 10 km/h 150/50 cos10000 for rehabilitation and therapy for models with running surface 150/50 cm
Special speed 0 - 30 km/h 150/50 cos00096110008 for athletes for models with running surface 150/50 cm
Variable elevation speed 200+/75+ cos100923 Elevation/gradient can be changed in various speed; for models with
running surface 200/75…300/125 cm. Manufactured after January 2011.
safety systems order number higher prices than the upgrade prices are applicable for retrofitting
of options and accessories ***
Safety arch 50 with harness & chest belt / fall-stop
prevention
cos10079 fall protection for all applications (mandatory for high risk applications); for
running surface 50 cm wide
Safety arch 65 with harness & chest belt / fall-stop
prevention
cos10170 fall protection for all applications (mandatory for high risk applications); for
running surface 65 cm wide
Safety arch 75 with harness & chest belt / fall-stop
prevention
cos10171 fall protection for all applications (mandatory for high risk applications); for
running surface 75 cm wide
Safety arch 100 with harness & chest belt / fall-stop
prevention
cos10172 fall protection for all applications (mandatory for high risk applications); for
running surface 100 cm wide
Safety arch upgrade h=220 cm cos144 ** upgrade for safety arch for users up to 2.20 m
Chest belt XL for safety arch cos14903-03-XL chest belt for safety arch, size XL (chest measurement: 125-155 cm)
Chest belt L for safety arch cos14903-03-L chest belt for safety arch, size L (chest measurement: 105-135 cm)
Chest belt M for safety arch cos14903-03-M chest belt for safety arch, size M (chest measurement: 85-115 cm)
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 12 of 16
Chest belt S for safety arch cos14903-03-S chest belt for safety arch, size S (chest measurement: 65-95 cm)
Chest belt XS for safety arch cos14903-03-XS chest belt for safety arch, size XS (chest measurement: 55-75 cm)
Chest belt XXS for safety arch cos14903-03-XXS chest belt for safety arch, size XXS (chest measurement: 45-65 cm)
Rear panel terminal for additional options cos10109 for running machines with user terminal (free with safety arch, additional
keyboard, extra safety stop)
Magnet with ripcord incl. plastic clip (spare part) cos101699 safety stop lanyard pull-cord with magnet (only spare part "magnet, cord
and clip", not the integrated contact switch)
Emergency stop-button with magnet holder + 5 m
spiral cable
cos15933 and Velcro-Strip for safety reasons. Jack at this treadmill is prerequisite!
Quote treadmill serial no. for refitting.
Emergency stop-button with magnet holder + 10 m
spiral cable
cos100548 and Velcro-Strip for safety reasons. Jack at this treadmill is prerequisite!
Quote treadmill serial no. for refitting.
special colour order number
Special colour treadmill frame 150/50 in RAL cos00098110004 all special (RAL) colours (incl. pure white, signal blue, sulphur yellow); free
of charge for orders of 5 or more
Special colour treadmill frame xx/65 in RAL cos100138 all special (RAL) colours (incl. pure white, signal blue, sulphur yellow); free
of charge for orders of 5 or more
special voltages order number
Special voltage for medical devices from 150/50 to
170/65
cos101214 voltage-M U:100V AC1~/N/PE 25A f50/60Hz (not recommended due to
performance limitations)
Special voltage for sports devices from 150/50 to
170/65
cos101215 voltage-S U:100V AC1~/N/PE 25A f50/60Hz (not recommended due to
performance limitations)
Special voltage for sports devices from 150/50 to
170/65
cos10470 voltage-S U:110V AC1~/N/PE 25A f50/60Hz (not recommended due to
performance limitations)
Special voltage for sports devices from 150/50 to
170/65
cos10468 voltage-S U:200V AC1~/N/PE 20A f50/60Hz (consider performance
limitations)
Special voltage for sports devices from 150/50 to
170/65
cos16172 voltage-S U:208V AC1~/N/PE 20A f50/60Hz (consider performance
limitations)
Special voltage for medical devices from 150/50 to
170/65
cos10469 voltage-M U:110V AC1~/N/PE 25A f50/60Hz (not recommended due to
performance limitations)
Special voltage for medical devices from 150/50 to
170/65
cos12850 voltage-M U:200V AC1~/N/PE 20A f50/60Hz (consider performance
limitations)
Special voltage for medical device 170/65 3p cos16153-01 voltage-M U:200V AC3~/PE 32A f50/60Hz
Special voltage for medical devices from 150/50 to
170/65
cos15990 voltage-M U:208V AC1~/N/PE 20A f50/60Hz (consider performance
limitations)
Special voltage for medical device 170/65 3p cos12068 voltage-M U:208V AC3~/N/PE 32A f50/60Hz
Transformer 3p Dyn5 200V/208/220/400/480 cos100803-01 transformer for special voltage of 3x200V/76A, 3x208V/72A, 3x220V/67A,
3x400V/36A, 3x480V/30A
f: 50/60 Hz, incl. housing in IP21, protectors and jacks of 1p/3 p
wheel chair ramps order number
Wheelchair ramp L: 1.25 m for treadmill 150/50 cos16186-01 wheelchair ramp (L: 125 cm x W: 80 cm) for models with running surface
150/50 cm
Access aid for wheelchair ramp cos16186-01 cos101614 additional hatch (1 pair = left/right), which almost eliminates the step
between wheelchair ramp and running deck
Wheelchair ramp L: 1.25 m for treadmill
170&190/65
cos101057va01 wheelchair ramp (L: 125 cm x W: 105 cm) for models with running surface
170&190/65 cm
user terminals order number
UserTerminal Touchpanel external remote control cos10002-02 connected via RS232 and coscom® interface to the treadmill (available on
request)
UserTerminal Touchpanel retro fit kit for handrail
mounting
cos14569-02 connected via RS232 and coscom® interface to the treadmill (available on
request), for handrail (D=60 mm)
Display & Keyboard 6 keys retro fit kit for
UserTerminal MCU5
cos15312va02 retrofit-kit Display & Keyboard for UserTerminal, extra charge for travel
costs and installation; (available on request)
Folding shelf for external control desk,
UserTerminal oversize treadmill
cos100781 for models with running surface 200+/75+ cm, manufactured after January
2011
Notebookholder (mounted on ext. control desk
oversize treadmill)
cos13320-01 for models with running surface 200+/75+ cm, manufactured after January
2011
interfaces, cables & printers order number
Heart rate receiver set POLAR® WIND incl.
Waerlink transmitter
cos100106 digital heart rate transmission through POLAR® WIND WearLink, coded,
with transmission range approx. 10 m
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 13 of 16
POLAR® HeartRate Sensor H7 Wearlink® Bluetooth
Smart
cos101787 wireless transmitter including chest belt size M-XXL, for smartphones,
iPhone (from iPhone 4S, iOS 5.1)
2nd interface RS232 com2 coscom® 9,600 bps cos00098010025 additional RS232 port for ECG, EEG, PC, printer or UserTerminal
3rd interface RS232 com3 coscom® 115,200 bps cos16487 additional RS232 port for ECG, EEG, PC, or UserTerminal with increased
data speed
Connection cable RS232 5 m cos00097010034 RS232, 5 m (Sub-D 9-pin male/female), USB-Adapter optionally available
(cos12769-01)
Connection cable RS232 10 m cos00097010035 RS232, 10 m (Sub-D 9-pin male/female), USB-Adapter optionally
available (cos12769-01)
Connection cable RS232 15 m cos11889 RS232, 15 m (Sub-D 9-pin male/female), USB-Adapter optionally
available (cos12769-01)
Potential equalization cable 5 m (with 2 POAG-
plugs)
cos10223 potential equalization cable for medical devices
RS232 - Centronix converter for printer cos10056 converter for connecting PCL printer directly to running machine
USB to RS232 converter cos12769-01 converter from USB to serial port RS232 (Sub-D 9-pin male)
Laser printer BW + interface converter cos10073 laser printer BW + interface converter and cable
Laser printer COL cos15511-03 laser printer 4-COL
h/p/cosmos satellite print bw cos14954 laser printer B/W, RS232 converter and cable, h/p/cosmos printer table
h/p/cosmos satellite PC med cos14970 medical PC, 19" LCD Monitor, COL Laser printer, h/p/cosmos PC table
with 4 wheels
DELL Laptop Computer cos13476-01 technical details and specification on request
lactate measurement, supplies & software order number
Software h/p/cosmos para control® cos10071-v4.1.0-txt software for controlling running machines
Software h/p/cosmos para graphics® cos10156-V2.6.14 software for controlling, recording and displaying running machine
parameters
floor protection mats order number
Floor protection mat 150/50 cos14005 for all h/p/cosmos running machines with running surface 150/50 cm
Floor protection mat 170/65 cos14042 for all h/p/cosmos running machines with running surface 170/65 cm
Floor protection, steel base plate (for elevation
wheels 150/50)
cos101393 floor plate for weight distribution, for models 150/50 cm
Floor protection, steel base plate (for elevation
wheels 170/65)
cos101394 floor plate for weight distribution, for models 170/65 cm
* We recommend a dedicated line 3 phase power connection (400 Volt AC3~/N/PE 50/60 Hz 16 to 32A fuse) and 3-phase device for high speed, fast
acceleration, special applications and for heavier subjects due to higher performance. For all single phase powered treadmills the natural performance limitations of single phase voltage supply apply due to the law of physics. For professional performance diagnostics, athletic training and high performance applications we strongly recommend running machines with 3-phase voltage power supply.
** Article in several sizes, variants or versions available (detailed addition of the article no.) Please quote size in your order.
*** In case of retrofitting of options and accessories, prices will be higher than indicated in the price list. Please specify device, model and serial number for
a detailed quotation.
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 14 of 16
17 Contact
For additional orders and technical enquiries please have the model type, the serial
number and the date of installation of your running machine ready. In case of any
further questions about delivery dates, service or maintenance, orders for
consumables etc., please contact the corresponding phone, fax or email for
qualified help.
Sales & service
COSMED S.r.l.
Via dei Piani di Monte Savello, 37
IT 00040 Pavona di Albano - Rome
Italy
phone +39 06 9 31 54 92
fax +39 06 9 31 54 92
email [email protected]
Headquarters of manufacturer
h/p/cosmos sports & medical gmbh
Am Sportplatz 8
DE 83365 Nussdorf-Traunstein
Germany
phone +49 18 05 16 76 67
fax +49 18 05 16 76 69
www.h-p-cosmos.com
Building 1 (top picture)
h/p/cosmos development & production
Am Sportplatz 8
DE 83365 Nussdorf-Traunstein
Building 2 (picture below)
h/p/cosmos sales & service
Feldschneiderweg 5
DE 83365 Nussdorf-Traunstein
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 15 of 16
file: n:\article\cos14310m5-v1_07cme-en\20140318_cos14310m5-v1_07cme-en_manual_cosmed_running_machine_print.doc © 2014 h/p/cosmos sports & medical gmbh [email protected] created 15.04.2014 printed 15.04.2014 page: 16 of 16