ltv 1200 ems training 12-18-07

101
Mechanical Mechanical Ventilation Ventilation With the LTV With the LTV ® ® 1200 1200 Ventilator: Ventilator: Applications for Applications for EMS EMS

Upload: wiadams

Post on 07-May-2015

1.995 views

Category:

Business


8 download

TRANSCRIPT

Page 1: Ltv 1200   ems training 12-18-07

Mechanical Mechanical VentilationVentilationWith the LTVWith the LTV® ®

12001200Ventilator:Ventilator:

Applications for Applications for EMSEMS

Page 2: Ltv 1200   ems training 12-18-07

LTVLTV ®®’s Versatile Applications’s Versatile Applications

• Adult & Pediatric (> 5kg)

• Invasive (ETT, King LT, or Combitube) or Non-Invasive (Mask) Ventilation

• Emergency and Non-emergency Transport

• Long-Term Care

• Rehab

• Homecare

Page 3: Ltv 1200   ems training 12-18-07

LTVLTV ®® 1200 Ventilators 1200 Ventilators

Small & Light Weight

• 3” x 10” x 12”

• Weighs less than 15 pounds

• Built-in turbine; no compressor or 50 psi air source needed

• Multiple power options

Page 4: Ltv 1200   ems training 12-18-07

Indications for UseIndications for Use

The majority of the time the LTV 1200 will be used during the transport of a patient already supported by a ventilator

Emergency patients for whom the ventilator may be used: – Cardiac and/or Respiratory Arrest (apnea)– Respiratory Failure (inadequate rate and/or volume)– Impending Respiratory Failure (multiple causes)– Patients already on a vent with unassociated problems

Page 5: Ltv 1200   ems training 12-18-07

Mechanical VentilationMechanical Ventilation

The BasicsThe Basics

Page 6: Ltv 1200   ems training 12-18-07

Ventilation vs. OxygenationVentilation vs. Oxygenation

Ventilation is the movement of gas into and out of the lungs and does not necessarily provide oxygenation. (e.g. a patient with no heart beat or circulatory support is not being oxygenated by simply blowing oxygen into their lungs.)

Oxygenation is the delivery of oxygen to the blood and tissue level of the patient and does not necessarily require ventilation. (e.g. a patient on heart-lung bypass is being oxygenated without any ventilation of the lungs)

Page 7: Ltv 1200   ems training 12-18-07

Of Course …Of Course …

Humans usually provide oxygen to their system by ventilating themselves (breathing) with air which includes approximately 21% oxygen.

A functioning circulatory system will send enough blood flow past the alveoli in the lungs to adequately absorb oxygen and excrete CO2.

Page 8: Ltv 1200   ems training 12-18-07

and . . .and . . .

A decrease in blood oxygenation (hypoxemia) or increase in blood carbon dioxide levels (hypercarbia) will increase the body’s need for ventilation.

CO2 and/or O2 Respiratory Drive

Need for More Ventilation

Page 9: Ltv 1200   ems training 12-18-07

Mechanical VentilationMechanical Ventilation

Definition Mechanical ventilation is the use of a

mechanical device to inflate and deflate the lungs.

Purpose Mechanical ventilation provides the force

needed to deliver air to the lungs in a patient whose own ventilatory abilities are diminished or lost.

Page 10: Ltv 1200   ems training 12-18-07

Mechanical VentilatorsMechanical Ventilators

Range of devices:

Bag-Valve-MaskEmergency Vents

Transport VentsHomecare Vents

Critical Care Vents

Page 11: Ltv 1200   ems training 12-18-07

Pressure vs. Volume VentilationPressure vs. Volume Ventilation

Ventilating Pressure is controlled and the resulting tidal volume is based on the physical size of airways and lungs and the patient’s lung compliance (stiffness).

or Tidal Volume is controlled and the

resulting pressure is based on the physical size of airways and lungs and the patient’s lung compliance (stiffness).

Page 12: Ltv 1200   ems training 12-18-07

“ “Modes” of Ventilation on the LTV 1200Modes” of Ventilation on the LTV 1200

AC – Assist/Control SIMV - Synchronized Intermittent Mandatory

Ventilation PSV - Pressure Support Ventilation (PSV) CPAP – Continuous Positive Airway Pressure NPPV – Non-Invasive Positive Pressure

Ventilation (a.k.a. Bi-Level)

Page 13: Ltv 1200   ems training 12-18-07

Assist / Control Ventilation Assist / Control Ventilation

A minimum number of mandatory breaths per minute is set, the breaths are given at either a set tidal volume or a set pressure. The patient may breath faster than the set rate, but the breaths will be measured and may be assisted in order to reach the set tidal volume or set pressure.

– Example: The ventilator is set to a rate of 8. The patient will get at least 8 “controlled” breaths per minute, but the ventilator can respond to patient effort and “assist” with additional breaths, all at the set tidal volume or pressure.

Page 14: Ltv 1200   ems training 12-18-07

Assist / Control VentilationAssist / Control Ventilation If patient DOES NOT have any spontaneous respiration's, then the patient will receive the set number of breaths at either the set

pressure or the set tidal volume each minute.

If the patient DOES try to initiate a spontaneous breath, the patient will receive the set tidal volume or pressure.

Time

Pressure

Machine-initiated and/or patient-initiated breaths are all delivered at the set parameters (volume or pressure)

Controlled breath Assisted breath

Patient effort

Page 15: Ltv 1200   ems training 12-18-07

Basic Assist / Control SettingsBasic Assist / Control Settings

Breath Rate Tidal Volume or Pressure Control level Inspiratory Time (or Peak Inspiratory Flow rate) Oxygen % PEEP Sensitivity

Page 16: Ltv 1200   ems training 12-18-07

PSV - Pressure Support Ventilation PSV - Pressure Support Ventilation

The patient’s spontaneous breathing effort is supported to a set positive pressure from the ventilator.

There are no mandatory breaths from the ventilator. Patient effort determines respiratory rate, inspiratory time, peak flow, and tidal volume

Page 17: Ltv 1200   ems training 12-18-07

PSV - Pressure Support VentilationPSV - Pressure Support Ventilation

Goals– Overcome the work associated with moving

gas through the artificial airway and circuit– Improve patient/ventilator synchrony– Augment spontaneous tidal volume

10cm

Time

Pressure

Patient effort

Pressure Support Setting

Page 18: Ltv 1200   ems training 12-18-07

Basic PSV SettingsBasic PSV Settings

****THE PATIENT MUST HAVE ADEQUATE SPONTANEOUS RESPIRATORY EFFORT

PSV - Range 5-25 cmH2O (typical)

PEEP FiO2

Page 19: Ltv 1200   ems training 12-18-07

Assist / Control vs. Pressure SupportAssist / Control vs. Pressure Support

Assist / Control

Set Tidal volume or set Pressure

Breaths may or may not be spontaneous

Tidal Volume, inspiratory time, flow rate, and minimal set breathing rate are determined by clinician

Pressure Support

Set Pressure assist level

All breaths are spontaneous

Patient determines respiratory rate, inspiratory time, peak flow, and tidal volume

Page 20: Ltv 1200   ems training 12-18-07

Spontaneous breath

SIMV – SIMV – Synchronized Intermittent Mandatory VentilationSynchronized Intermittent Mandatory Ventilation

• This ventilation mode provides a mixture of mandatory (controlled) and spontaneous breath types.

• The LTV will give and/or “sync” with enough breaths to achieve the set breathing rate; the patient may breath in between these mandatory breaths, but the “in between” breaths will not be assisted

Time

Pressure

Synchronized machine breath

Patient effort

Machine breath

Page 21: Ltv 1200   ems training 12-18-07

Basic SIMV SettingsBasic SIMV Settings

Breath Rate Tidal Volume or Pressure Control level Inspiratory Time Oxygen % PEEP Sensitivity

Time

Pressure

Synchronized machine breath

Patient effort

Page 22: Ltv 1200   ems training 12-18-07

PEEP and CPAPPEEP and CPAP

Definition:– PEEP = Positive End Expiratory Pressure– CPAP = Continuous Positive Airway Pressure

PEEP: a technique of assisting breathing by increasing the air pressure in the lungs and air passages near the end

of expiration so that an increased amount of air remains in the lungs following expiration

CPAP: a technique of assisting breathing by maintaining the air pressure in the lungs and air passages constant and above atmospheric pressure throughout the breathing cycle

Page 23: Ltv 1200   ems training 12-18-07

PEEP and CPAPPEEP and CPAP

Function:** Functionally, CPAP and PEEP do the same thing:

Splint open airways and alveoli - Increases functional residual capacity (FRC)

Improves oxygenation Redistributes lung water from alveoli to

perivascular space (very beneficial in CHF and Pulmonary Edema patients)

Page 24: Ltv 1200   ems training 12-18-07

CPAPCPAP

CPAP is actually not a mode of “ventilation” as CPAP does not move gas into and out of the patient. CPAP requires a spontaneous breathing patient.

Paw

SpontaneousSpontaneous Spontaneous Spontaneous

Page 25: Ltv 1200   ems training 12-18-07

CPAPCPAP

CPAP machines used for obstructive sleep apnea are typically used to “splint” upper airway structures open, but have the same effect as CPAP on the ventilator.

Page 26: Ltv 1200   ems training 12-18-07

Sensitivity:Sensitivity:Understanding how the vent cyclesUnderstanding how the vent cycles

Sensitivity determines when the ventilator will recognize a patient’s inspiratory effort.

The LTV 1200 offers flow sensitivity.

Page 27: Ltv 1200   ems training 12-18-07

Flow SensitivityFlow Sensitivity

Flow Sensors

Page 28: Ltv 1200   ems training 12-18-07

SensitivitySensitivity

Ventilator delivers a low level of constant flow (10 lpm) into the patient circuit. This is called the “bias” flow.

Delivered flowReturned flow

No patient effort

Page 29: Ltv 1200   ems training 12-18-07

Sensitivity Sensitivity

As the patient begins to inhale, some of this constant flow is diverted to the patient.

Change in flow in the vent circuit will cause the vent to “cycle” and deliver gas to patient.

Delivered flowLess flow returned

Page 30: Ltv 1200   ems training 12-18-07

SensitivitySensitivity

Sensitivity is usually set 2-3 liters per minute in the hospital.

May require slightly higher setting during transport in field to avoid “auto-cycling” of the ventilator.

Delivered flowLess flow returned

Setting too low can cause auto-cycling of the ventilator (usually due to leak in system)

Setting too high can “lock out” patient from being able to “trigger” any spontaneous breaths

Page 31: Ltv 1200   ems training 12-18-07

Ventilator AlarmsVentilator Alarms

High airway pressure Low airway pressure Low minute volume Apnea

Page 32: Ltv 1200   ems training 12-18-07

Ventilator Alarms- Ventilator Alarms- High pressure limitHigh pressure limit

– Setting High Pressure Limit Alarm:

• Usually set within 10 cmH2O above patient’s average Peak Pressure.

• *** When activated, ventilator will terminate breath and the patient does not receive full tidal volume

Page 33: Ltv 1200   ems training 12-18-07

Ventilator Alarms- Ventilator Alarms- High pressure limitHigh pressure limit

Causes of high pressure alarm violation:– Resistance to gas flow:

• kinks in tubing or monitoring lines• patient coughing• secretions• bronchospasm• gagging, “fighting the ventilator”

– Decrease in lung compliance:• atelectasis• pneumothorax• pulmonary edema

Page 34: Ltv 1200   ems training 12-18-07

Ventilator Alarms- Ventilator Alarms- Low Pressure AlarmLow Pressure Alarm

– Setting Low Pressure Alarm:• 5-8 cmh2o less than ventilating pressure

Page 35: Ltv 1200   ems training 12-18-07

Ventilator Alarms- Ventilator Alarms- Low Pressure AlarmLow Pressure Alarm

Causes of Low Pressure Alarms:– Cuff Leak– Vent Circuit

• Check tubing for holes or kinking• Check monitoring lines for tight fit or kinking (Leur

connections can become loose)

• Check connection at “Y” connector

– If using a Ballard suction device, check that all connections are secure (cap for saline port)

– Vent not meeting patient’s inspiratory need (A/C)

Page 36: Ltv 1200   ems training 12-18-07

Ventilator Alarms- Ventilator Alarms- Low minute volumeLow minute volume

Definition: Minute volume = total volume of breaths over 1 minute time

e.g. 10 breaths per minute x 600 ml per breath = minute volume of 6000 ml 6.0 L minute volume

– Setting Low Minute Volume:• Set 3 Liters under patient’s minute volume, with a

minimal setting of 5L/m.• Ensures adequate alveolar ventilation is maintained.

Page 37: Ltv 1200   ems training 12-18-07

Ventilator Alarms- Ventilator Alarms- Low minute volumeLow minute volume

– Causes of Low Minute Volume alarms:

• Neuro changes (A/C or PSV)

• Sedation issues (A/C or PSV)

• Patient fatigue (PSV)

• Decrease in compliance (PSV)

• High pressure alarm active and ventilator dumps delivered tidal volume (A/C or PSV)

Page 38: Ltv 1200   ems training 12-18-07

Ventilator AlarmsVentilator Alarms

Apnea Parameters

– Activated when no exhalation is detected for a selected time period (e.g. 20 seconds)

– Tidal volume and pressure control level should be set appropriately for patient, as these will be used for apnea ventilation.

Page 39: Ltv 1200   ems training 12-18-07

Key Points Key Points BEFOREBEFORE Transporting Transporting

– See how the patient is interacting with their current vent

• If in A/C– Breathing Rate– Minute Volume– Peak Pressures – Peak Flow on the hospital vent?– Sensitivity

Page 40: Ltv 1200   ems training 12-18-07

Key Points Key Points BEFOREBEFORE Transporting Transporting

– See how the patient is interacting with their current vent

• If in PSV– Breathing Rate– Spontaneous tidal volumes – Minute Volume– Sensitivity

Page 41: Ltv 1200   ems training 12-18-07

Key Points Key Points BEFOREBEFORE Transporting Transporting

– Talk to the patients therapist and nurse:• Secretions

– (If already in place, keep the in-line suction device attached to the patient when you go)

• Weaning schedule or ventilator goals for this patient• Any “Quirky” respiratory patterns

– example: pt will breath 50 times per minute when he/she gets anxious.

– See what relieves the “quirkiness” (changing modes, settings, favorite medication, reassurance, etc)

Page 42: Ltv 1200   ems training 12-18-07

Key Points Key Points While While TransportingTransporting

– Set alarms appropriately• if set appropriately, alarms can alert you to subtle

changes before they become large problems.

– Monitor patients vent parameters• Minute Volume (A/C or PSV)• Peak Pressure (A/C)• Tidal Volumes (PSV)

– Anticipate what changes you would make if patients vent needs change?

Page 43: Ltv 1200   ems training 12-18-07

LTV 1200 Ventilator SetupLTV 1200 Ventilator Setup

Overview

Page 44: Ltv 1200   ems training 12-18-07

LTV 1200 Ventilator Setup - OverviewLTV 1200 Ventilator Setup - Overview

Making the connections

Turning the unit ON

Adjusting the settings

Extended Menus

Monitoring the patient

Turning OFF and processing the unit

Page 45: Ltv 1200   ems training 12-18-07

THE LTV 1200 VENTILATORTHE LTV 1200 VENTILATOR

Page 46: Ltv 1200   ems training 12-18-07

Front Panel OverviewFront Panel Overview

Page 47: Ltv 1200   ems training 12-18-07

Left Side Panel OverviewLeft Side Panel Overview

Cooling Fan

Power and Communication

Connections

High/Low Pressure O2 connection

Turbine Intake Filter

Page 48: Ltv 1200   ems training 12-18-07

Power SourcesPower Sources

External AC Adapter 120VAC/12VDC External Lithium Ion Battery 12V (3 hours) Internal Battery (1 hour)

Page 49: Ltv 1200   ems training 12-18-07

Lithium Ion BatteryLithium Ion Battery

Lithium Ion Battery in pocket of Backpack carrying case

Page 50: Ltv 1200   ems training 12-18-07

Oxygen SourceOxygen Source

The LTV 1200 can be used with either a 50-PSI oxygen source or with Low Pressure oxygen.

Page 51: Ltv 1200   ems training 12-18-07

Oxygen Source Oxygen Source – 50 PSI– 50 PSI

A 50 PSI Oxygen source allows you to use the internal oxygen-air blender to set a specific O2% from 21-100%.

The 50 PSI source can be from a regulated oxygen cylinder or off the ambulance or hospital wall source.

An input O2 of less than 35 PSI will cause an alarm.

Page 52: Ltv 1200   ems training 12-18-07

Oxygen Source Oxygen Source – Low Pressure– Low Pressure

Home ventilator patients may utilize low pressure oxygen (from a flow meter or oxygen concentrator) bled into the unit using a nipple adapter on the oxygen fitting on the unit.

The “Low Pressure O2 Source” button must be activated and the blender will no longer active.

An O2 Input pressure of more than 35 PSI will cause an alarm.

Page 53: Ltv 1200   ems training 12-18-07

Right Side Panel OverviewRight Side Panel Overview

Page 54: Ltv 1200   ems training 12-18-07
Page 55: Ltv 1200   ems training 12-18-07

Front Panel OverviewFront Panel Overview Display of Monitored Data

Ventilation ControlsAlarm Settings Set Value Knob

LED Pressure Bar

Silence + Reset

Other indicators

Page 56: Ltv 1200   ems training 12-18-07

Key LTV Ventilator SettingsKey LTV Ventilator Settings

Rate: 0 to 80 bpm Tidal Volumes: 50 to 2000 ml Press. Control: 1 to 99 cmH2O I-time: 0.3 to 9.9 sec Press. Support: 0 to 60 cmH2O O2%: 21 to 100% PEEP: 0 to 20 cmH2O

Page 57: Ltv 1200   ems training 12-18-07

High Pressure

Low Pressure

Low Minute Volume

Apnea Low/High Oxygen

Pressure Disconnect/Sense Power Low, Power Lost Hardware Fault Battery Low, Battery

Empty Vent Inoperative

LTV® Alarms

Operator Set Preset - Automatic

Page 58: Ltv 1200   ems training 12-18-07

Monitored ParametersMonitored Parameters

Display Monitored Data Description

PIPPeak Inspiratory

Pressure

Greatest pressure measured during the inspiratory phase

Updated at end of inspiration

MAP Mean Airway Pressure The average airway pressure for the last

60 seconds. Updated every 10 seconds

PEEPPositive End Expiratory

Pressure

The pressure in the patient circuit at the end of exhalation.

Updated at the end of exhalation

f Total Breath Rate

Breaths / minute based on the last 8 breaths. Includes all breath types.

Recalculated and updated at the end of each exhalation or 20 seconds

Page 59: Ltv 1200   ems training 12-18-07

Monitored ParametersMonitored Parameters

Display Monitored Data Description

Vte Exhaled Tidal Volume

Displays the Exhaled Tidal Volume as measured at the patient wye

Updated at the end of exhalation

VE Exhaled Minute Volume

Displays the exhaled tidal volume for the last 60 seconds, calculated from the last 8 breaths.

Updated every 10 seconds

I:EInspiratory / Expiratory

Ratio

Displays the ratio between measured inspiratory and expiratory time

Also displays inverse I:E ratios

VcalcCalculated Peak Flow(Volume Breaths only)

The calculated peak flow based on tidal volume and inspiratory time settings

Displayed when selected or whenever these two controls are selected

Page 60: Ltv 1200   ems training 12-18-07

Front Panel – On/OffFront Panel – On/Off

Press to turn ONTo turn OFF, press On/Standby for 3 seconds, then press Silence/Reset

Page 61: Ltv 1200   ems training 12-18-07

LTV 1200 PresetsLTV 1200 Presets

Presets are loaded in the LTV to facilitate the quick initiation of mechanical ventilation when operators with limited knowledge of the equipment must apply it.

Page 62: Ltv 1200   ems training 12-18-07

LTV 1200 PresetsLTV 1200 Presets

SAME

NEW

Page 63: Ltv 1200   ems training 12-18-07

PresetsPresets

INFANT

PEDIATRIC

ADULT

Page 64: Ltv 1200   ems training 12-18-07

PresetsPresets

Page 65: Ltv 1200   ems training 12-18-07

PresetsPresets

The preset values are simply recommended starting points and should be safe levels for most patients. Once mechanical ventilation is initiated, adjustments and changes should be made to meet the needs of the patient.

The full range of ventilator settings is available to the operator, regardless of the preset used.

Page 66: Ltv 1200   ems training 12-18-07

Basic Operations – Vent SettingsBasic Operations – Vent Settings

When there are variable settings or options:– Press the button by the parameter or setting to

be changed– Turn the Set Value knob clockwise or

counterclockwise to the desired setting– Press the parameter button again to confirm

setting

Page 67: Ltv 1200   ems training 12-18-07

Rules to keeping it simpleRules to keeping it simple::

Rule # 1

You either set the ventilator to deliver a volume or you set the ventilator to deliver

a pressure.

• If you set volume, you monitor pressure closely.• If you set pressure, you monitor volume closely.

Page 68: Ltv 1200   ems training 12-18-07

Basic Operation - Mode SelectionBasic Operation - Mode Selection

Modes are selected on the bottom row of the green ventilator setting box.

Selections are made by pressing the Mode “Select” button. One push and the next mode selection flashes. A second push confirms the mode and makes the change.

The NPPV mode (Non-invasive Positive Pressure Ventilation) is the Bi-Level setting for the LTV. – A non-vented mask is necessary when applying NPPV with

the LTV.

Page 69: Ltv 1200   ems training 12-18-07
Page 70: Ltv 1200   ems training 12-18-07

Rules to keeping it simpleRules to keeping it simple::

Rule # 2

Monitor minute volume closely

If minute volume changes, understand what caused it to change.

Anticipate what adjustments you will make if patient’s minute volume changes. (dependant on what mode patient is in)

Page 71: Ltv 1200   ems training 12-18-07

Monitor Display (scrolling)Monitor Display (scrolling)

Page 72: Ltv 1200   ems training 12-18-07

Rules to keeping it simpleRules to keeping it simple::

Rule # 3

Alarms:if set and used appropriately, they will alert

you of changes in the patient’s ventilation before they become life threatening.

Page 73: Ltv 1200   ems training 12-18-07

Alarm SettingsAlarm Settings

Page 74: Ltv 1200   ems training 12-18-07

AlarmsAlarms

Audible and visual alarm when parameter is violated.

If situation is corrected, audible alarm will silence, but visual will stay lit until Silence/Reset button is hit.

The monitor display will show the active alarm violation until reset.

Page 75: Ltv 1200   ems training 12-18-07

Basic Ventilator SettingsBasic Ventilator Settings

Breath Rate, Tidal Volume, Pressure Control, Insp. Time, Pressure Support, O2 %, Sensitivity

Set by selecting the parameter button, rotating the set value knob, and pushing parameter button again or waiting 5 seconds

Page 76: Ltv 1200   ems training 12-18-07

Basic Ventilator SettingsBasic Ventilator Settings

Page 77: Ltv 1200   ems training 12-18-07

Monitoring the PatientMonitoring the Patient

Measured and calculated values scroll through the monitor display

Press the ‘Select’ button to find to a desired parameter

Double click the ‘Select’ button to resume the automatic scroll

Airway pressure is dynamically displayed on the light bar above monitoring display

Page 78: Ltv 1200   ems training 12-18-07

Monitor Display (scrolling)Monitor Display (scrolling)

Page 79: Ltv 1200   ems training 12-18-07

Other Settings on Front of LTVOther Settings on Front of LTV

Insp/Exp Hold– These are maneuvers that are used to assess

the lung compliance of the patient and determine if there is any air-trapping happening during ventilation.

– You will not be utilizing these maneuvers. Low O2 Source

– This button must be selected when a low pressure O2 source is used.

– You will be utilizing a high pressure source, so this option should not be On, or lit.

Page 80: Ltv 1200   ems training 12-18-07

Manual Breath– This can be used to temporarily increase

the ventilation for a patient – Sometimes used after a stressful situation

to help the patient “catch up” with their ventilation demands

– Manual breaths also send burst of air through sensing lines to clear them of fluid/secretions

Other Settings on Front of LTVOther Settings on Front of LTV

Page 81: Ltv 1200   ems training 12-18-07

Other Settings on Front of LTVOther Settings on Front of LTV

Control Lock– Pressing this button will lock the controls on

the unit, so they may not be accidentally (or intentionally) changed.

– The indicator is lit when the controls are locked out.

– Simply press the button again to turn off the lock, thus allowing changes.

Page 82: Ltv 1200   ems training 12-18-07

Extended MenusExtended Menus

Accessed by pressing and holding the ‘Select’ button

Access to extended Alarm and Ventilator setup and operations including:

Page 83: Ltv 1200   ems training 12-18-07

Extended MenusExtended MenusAlarm OpAlarm VolApnea IntHP Delay

LPP AlarmHigh Rate

High PEEPPt. Assist

Exit

Vent OpRise TimeFlow TermTime Term

PC Flow TermNPPV ModeLeak CompCtrl UnlockLanguageVer XXX

Usage XXXCom Setting

Set DateSet Time

Date FormatPIP LEDO2 Flush

O2 Cyl DurExit

XDCR ZeroAP XXP

FDw xxPFDn xxP

Event Trace256 Event Codes

455 Events

Set ValueKnob

RT XDCR DataLeak

Page 84: Ltv 1200   ems training 12-18-07

Extended MenusExtended Menus

The extended menu settings can be preset to standard and acceptable levels for most applications, then accessed only when necessary, by properly trained personnel.

Page 85: Ltv 1200   ems training 12-18-07

Extended Menus Extended Menus – O2 Cylinder Duration– O2 Cylinder Duration

Particularly useful for transport teamsAccessed in:

Extended Features Vent Op

O2 Cylinder Duration Cyclinder Type?Cylinder Pressure?Calculate >>>

Page 86: Ltv 1200   ems training 12-18-07

O2 Cylinder DurationO2 Cylinder Duration

O2 DUR 04:55

O2 CYL DUR 04:55

Page 87: Ltv 1200   ems training 12-18-07

IMPORTANT !!!IMPORTANT !!!

NEVER silence an alarm without knowing the cause of the alarm and attempting to correct.

ALWAYS reset the alarm after you’ve taken corrective measures or completed suctioning.

Page 88: Ltv 1200   ems training 12-18-07

IMPORTANT !!!IMPORTANT !!!

Treat the patient, not the machine

Troubleshoot, starting with the patient:– Look at your patient – distressed, moving,

coughing, seizing, disconnected from vent, ???– Look at the vent alarm – which one is activated?– Look at the physiologic monitor – how is patient

responding? If patient is not being ventilated effectively,

solve the problem quickly or ventilate manually (BVM)

Page 89: Ltv 1200   ems training 12-18-07

Suctioning / Clearing SecretionsSuctioning / Clearing Secretions

Utilize the suction catheter on the patient’s existing circuit, or have catheter available

May need to pre-oxygenate and/or post-oxygenate some patients using O2% button

Use of the ‘Manual Breath’ button will deliver a breath at set volume or pressure, and will also send burst of air through sensing lines to clear any fluid/secretions blocking ports

Page 90: Ltv 1200   ems training 12-18-07

Use of a “closed” suction systemUse of a “closed” suction system

ATTACH SALINE BULLET, THEN UNLOCK THUMB VALVE WHILE DEPRESSING THUMB VALVE, SET WALL SUCTION (120-140 mm Hg) HOLD CONNECTOR WITH ONE HAND AND INSERT TIP OF

CATHETER INTO THE ENDOTRACHEAL TUBE LAVAGE (DEPENDING ON PROTOCOL) PASS CATHETER DOWN THE ENDOTRACHEAL TUBE

(measured or until resistance)

Page 91: Ltv 1200   ems training 12-18-07

Use of a “closed” suction systemUse of a “closed” suction system

DEPRESS THUMB VALVE ,WAIT FOR 1-2 SECONDS BEFORE SLOWLY PULLING THE CATHETER BACK (CONTINUOUS SUCTION)

WITHDRAW CATHETER UNTIL BLACK STRIPE IS VISIBLE IN SHEATH (*see below)

WHILE CONTINUING TO DEPRESS THE THUMB VALVE, FLUSH THE INSIDE OF THE CATHETER WITH 15 ML OF SALINE ** then release thumb valve**

DISCONNECT SALINE, LOCK THUMB VALVE CHANGE CATHETER EVERY 24 HOURS.

Page 92: Ltv 1200   ems training 12-18-07

Providing Oxygen “Flush”Providing Oxygen “Flush”

Pressing and holding the O2% button for 3 seconds will set the vent to deliver 100% oxygen (oxygen flush) for 2 minutes

This can be used to pre-oxygenate or post-oxygenate the patient during suctioning or a stressful event

Page 93: Ltv 1200   ems training 12-18-07

Oxygen % (100% flush)Oxygen % (100% flush)

Page 94: Ltv 1200   ems training 12-18-07

Oxygen % (100% flush)Oxygen % (100% flush)

Page 95: Ltv 1200   ems training 12-18-07

Sequence of LTV SetupSequence of LTV Setup

1. Connect breathing circuit to the LTV ventilator.2. Make sure ventilator is connected to adequate

power source – battery, UPS, or AC-DC supply (internal battery should only be used for short transport or during switch to alternate power supply).

3. Connect oxygen source to ventilator (if ventilating at greater than 21% O2).

4. Turn unit ON – UNIT SHOULD NOT BE CONNECTED TO THE PATIENT AT THIS TIME.

5. Select the patient type (Adult, Pediatric, Infant) using the Presets in the LTV.

6. Make any necessary adjustments in the ventilator settings.

Page 96: Ltv 1200   ems training 12-18-07

Sequence of LTV SetupSequence of LTV Setup

7. Set proper alarm limits, as appropriate for patient.8. Check the low pressure, high pressure, and disconnect alarms before applying to the patient. 9. Connect LTV breathing circuit to the patient and closely monitor the patient.

*Utilize HME (heat and moisture exchanger) and closed suction catheter, if on patient circuit at facility.10. Keep flow sensing lines up to avoid water and

secretions in these lines.11. Monitor your patient. Make appropriate adjustments,

and CALL FOR HELP, if you’re uncomfortable with what you see. Use the BVM if necessary.

Page 97: Ltv 1200   ems training 12-18-07

Key Points Key Points BEFOREBEFORE Transporting Transporting

See how the patient is interacting with their current vent.• If in A/C

– Breath Rate– Minute Volume– Peak Pressures – Sensitivity

• If in PSV– Breath Rate– Spontaneous tidal volumes – Minute Volume– Sensitivity

Page 98: Ltv 1200   ems training 12-18-07

Key Points Key Points BEFOREBEFORE Transporting Transporting

Talk to the patients therapist and nurse:• Secretions

– (keep the in-line suction device attached to the patient when you go)

• Weaning or goals• Any “Quirky” respiratory patterns

– example: pt will breath 50 times per minute when he/she gets anxious.

– See what relieves the “quirkiness” (changing modes, settings, favorite medication, reassurance, etc)

Page 99: Ltv 1200   ems training 12-18-07

Key Points Key Points While While TransportingTransporting

Set alarms appropriately• if set appropriately, alarms can alert you to subtle

changes before they become large problems. Monitor patient’s monitored vent parameters

• Breath Rate• Minute Volume (A/C or PSV)• Peak Pressure (A/C)• Tidal Volumes (PSV)

Anticipate what changes you would make if patient’s vent needs change?

If you need help, ask for it.

Page 100: Ltv 1200   ems training 12-18-07

TroubleshootingTroubleshooting

Page 101: Ltv 1200   ems training 12-18-07

Scenarios – Practical Training Scenarios – Practical Training