Presentationon
“Method Development and validation for the
Quantitative Estimation of ONDANSETRON
HYDROCHLORIDE by HPTLC in Bulk Drug & Tablet Dosage Form”
By: LOVE KUMAR
M.Sc Biotechnology
![Page 2: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/2.jpg)
What is Drug?
• Drug is substance meant for cure, mitigation
prevention or treatment of disease.
• A drug is a substance which may have
medicinal, intoxicating, performance
enhancing or other effects when taken or put
into a human body.
![Page 3: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/3.jpg)
ONDANSETRON HYDROCHLORIDE
• Ondansetron hydrochloride is a serotonin 5-HT-3 receptor antagonist used mainly as an antiemetic to treat nausea and vomiting
![Page 4: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/4.jpg)
Mechanism of action
• It effects on both peripheral and central nerves system by
reducing the activity of the vagus nerve which deactivate the
vomiting centre in the medulla oblongota and also block
serotonin receptors in the chemoreceptor trigger zone
• It also has little effect on vomiting caused by motion sickness.
![Page 5: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/5.jpg)
ondansetron hydrochloride
• 1. Nature : Amorphous
• 2. Colour : White
• 3. Odour : Odourless
• 4. Taste : Bitter
• 5. Melting point : 1200 C
• 6. Solubility : freely soluble in methanol
![Page 6: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/6.jpg)
Method Development
• Existing method may not be suitable.
• Unreliable
• Expensive
• Time consuming
• Not Easily automated
• New instrumentation technique may provide improved methods
![Page 7: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/7.jpg)
CHROMATOGRAPHY
• The term ‘Chromatography’ covers those processes aimed at the separation of the various species of a mixture on the basis of their distribution characteristics between a stationary and a mobile phase.
![Page 8: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/8.jpg)
Different modes of chromatography are as follows:
• Normal Phase Chromatography
• Reversed Phase Chromatography
• Reversed Phase – Ion pair Chromatography
• Ion Chromatography
• Ion-Exchange Chromatography
• Affinity Chromatography
• Size Exclusion Chromatography
![Page 9: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/9.jpg)
Instrumentation
Pump
Injector
Silica PlateDetector
Mobile Phases
Gradient Controller
•
![Page 10: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/10.jpg)
Materials and MethodReagents and Chemicals:
• Sample and working standards of ondansetron hydrochloride
were obtained from Sun Pharma Pvt Ltd, HPLC grade
methanol was purchased from E.Merck and ammonium acetate
of analytical reagent grade was purchased from Sisco Research
Laboratories.
Instrumentation:
• Uv –vis spectrophotometer is of backman DU640B
• Pump - water 600controller pump
• Auto sampler – water 717 plus
• Uv detector – water 486
• Software – millinium 32 .
![Page 11: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/11.jpg)
Chromatographic Conditions
• The elution of ondansateron hydrochloride was obtained by running
HPTLC in isocratic mode using methanol and tiriethylamine –
glacial acetic acid in a ratio of 9.5:0.5 v/v, flowrate was maintained
at 1.0 ml per minute with run time of 10 minutes.
• The Rf for ondansateron hydrochloride was obtained at 7.7 and
detection was performed at 309 nm.
• Mobile phase was previously filtered through what mann filter paper
no 41.
![Page 12: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/12.jpg)
• Preparation of Standard Stock solution
• Stock solution (1000 ppm) of ondansateron hydrochloride was
prepared by dissolving 100 mg of Eslicarbazepine Acetate
standard in methanol in a 100 ml volumetric flask, the volume
was made up to the mark with methanol.
• This stock solution was further diluted to obtain desired
concentrations.
• Preparation of Sample solution:
• Weighed 20 tablets of ondansateron hydrochloride (labelled
claim 400 mg of ondansateron hydrochloride ) and average
weight was calculated.
• The tablets were crushed to get homogenous powder and a
quantity equivalent to 100 mg was weighed in a volumetric flask.
• The powder was then allowed to dissolve in methanol by
sonication. Make up the volume to the mark with methanol and
filter the solution through 0.2 μm filters.
• The filtrate was diluted to obtain desired concentrations.
![Page 13: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/13.jpg)
Method Development:
Selection of wavelength
• Appropriate dilutions of ondansateron hydrochloride was
prepared from the stock solution,which were scanned over a
range of 200-400 nm and the UV spectrum was .
• The maximum absorbance was found at 309 nm.
![Page 14: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/14.jpg)
Method Validation
• Linearity• Sensitivity• Precision• Accuracy• Limit of Detection (LOD)• Limit of Quantitation (LOQ)• Ruggedness• Robustness• Stability• System Suitability
![Page 15: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/15.jpg)
Precision
• Precision of an analytical procedure referred to degree of
scatterness between a series of observations obtained from
multiple sampling of same homogenous sample in given
conditions.
• The term intraday precision (repeatability) refers to the use of
analytical procedure within same laboratory conditions over a
short period of time by same analyst and same instrument
where as inter day precision (intermediate precision) refers to
the use of analytical procedure in same laboratory conditions
on different days by different analysts.
• Six injections of ondansateron hydrochloride 50 ppm were
applied twice after a short interval of time on the same day, the
procedure was repeated on two consecutive days by different
analysts and % RSD was reported.
![Page 16: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/16.jpg)
Accuracy
• Accuracy is the closeness of the test results to that of the true
value which can be determined in terms of percent recovery.
• The recovery study was performed at three levels mainly
80%, 100% and 120% by adding known amount of pure drug
in to the 50ppm sample solution.
• The results of recovery studies was tabulated and found
satisfactory.
![Page 17: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/17.jpg)
Linearity
• Different concentrations ranging from 10-90 ppmwere injected in duplicate, out of which fiveconcentrations (10, 30, 50, 70, 90, ppm) wereselected and each concentration was injected induplicate.
• The mean peak areas were recorded and thecalibration curve
• was established by plotting concentration on x-axisand peak area on y-axis.
• The results show that the Beer’s law is obeyed inthe concentration range of 10-90 ppm.
![Page 18: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/18.jpg)
Range
• Range is the minimum and maximum concentration of the
sample at which the analytical procedure gives reproducible
results.
• Range can be determined by linearity, accuracy and precision
studies.
• The method was found acceptable across wide range of
concentration (10-90 ppm).
![Page 19: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/19.jpg)
Specificity
• The retention time of the sample solution of ondansateron
hydrochloride tablet was found to be 6.0 minutes, which is
similar to that of the standard solution of Eslicarbazepine
Acetate.
• This indicates that there is no drug-excipient interference and
the drug is properly resolved by this particular method
![Page 20: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/20.jpg)
Robustness
• Robustness determines the reproducibility of the test result
with small and deliberate variations in the method parameters.
• The experiment was carried out by slightly changing the ratio
of methanol (50±2 ml) in the mobile phase, column oven
temperature (25.0 ± 0.2oC)and flow rate (1.0±0.1 ml), The
effectiveness of the deliberate variations was observed on
retention time of peak.
• The statistical data gives no significant variations in the above
parameters indicating that the method is robust.
![Page 21: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/21.jpg)
Analysis of Pharmaceutical API & Dosage form
• The proposed method was successfully applied for the
estimation of ondansateron hydrochloride in bulk drug and
tablet dosage form.
• The assay results were compiled and chromatogram found
satisfactory and show there is no interference of the tablet
matrix with the drug.
• Low % RSD shows that this methodcan be easily applied for
the estimation of in Eslicarbazepine Acetate in bulk drug and
tablet dosage form.
![Page 22: Love ppt](https://reader033.vdocuments.mx/reader033/viewer/2022052602/55a2082b1a28abe8788b4809/html5/thumbnails/22.jpg)
Conclusion
A high performance liquid chromatography method for the quantitative estimation of in bulk drug and tablet dosage form has been developed as per the requirement of present era.
Statistical result and low % RSD values indicate that the method is precise, accurate, robust, specific and can be used as a wide range of concentration.