libero cs - elpro · applicaion area cold chain monitoring – applicaions with stability data or...
TRANSCRIPT
LIBERO CS
-Multi-Level PDFLogger
-Noequipment-Nosoftware
USB
Interface
LIBEROPDFLogger
CS
Anyconfigurationononedevice
Autoexpirydatemanagement
SecureStart/Stopbutton
Shipmentinformation
Temperature graphwithalarmlimits
Multi-levelalarm
statistics
LIBEROCSPDFReport
OKandALARMindicator
MKTandduration
TECHNICAL SPECIFICATIONS LIBERO CS
LIBERO CS
Multi-Level PDF Logger with Cold Chain DatabaseLIBERO CS is the PDF Logger that simplifies the shipment process for products with known stability data. Supporting up
to eight alarm levels, MKT and duration as alarm criteria, the LIBERO CS allows the user to create temperature profiles
for individual products. No need to keep different loggers in stock, simply apply the correct configuration profile and add
shipment information at the time of use and send off the logger. No equipment and no software is required. Release
products based on the OK or ALARM status on the display, download the PDF report and forward it to the sender or to
liberoMANAGER, ELPRO’s cold chain database, for further analysis.
we prove it.
- Simple and safe, no equipment, no software- Quick implementation, also by 3rd parties- Compatible with liberoMANAGER cold chain database - 100 % calibrated sensors and electronics- Fully compliant with industry guidelines
Type PDFLoggerwithmultiplealarmlevels
Application area ColdChainMonitoring–applicationswithstabilitydataorotherallowedtime/temperatureexcursions
Recording options and mode Singleuse:start/stopmode
Operating range -30 °C..+70 °C
Measurement accuracy TheLIBEROCSisthemostaccuratedataloggeramongstitspeersandfeaturesaguaranteedaccuracyof ±0.4°Cinthecorerangeof0.0°Cto25.0°C.Thetypicalaccuracy(95%ofallLIBEROs)isevenbetterwith
±0.3°Cinthecorerange.
Accuracyoverfulltemperaturerange: ±0.4°Cfrom0.0°Cto+25.0°C
Resolution 0.1 °
Measurement interval 1to60minutes,userprogrammable
Memory capacity 24'200measurementvalues(equals168daysruntimewith10mininterval)
Expiry date & battery life -Dataloggercanbestartedanytimeduringshelflife -Starteddataloggerrunsuntilmemoryisfullorloggerisstopped(max.expirydate+30days) -Autoexpirydatemanagement:Afterexpiry,displaychangesstatusandloggercannotbestartedanymore
-Usebelow0°Candabove+40°Cforprolongedtimescanlimitbatterylife
Battery type Buttoncellcontainedinequipment,UN3091,exemptfromDGRdeclaration
Certificate Ilac/NIST/ISO17025traceableproductioncalibrationcertificate
Programmable alarms -8temperaturealarmlevelswithsingleorcumulativedelays -AlarmonMKT -2temperaturethresholdswithalarmdelay -Alarmonduration/runtime
Start-up delay 0minutesto24h
Display MultifunctionLCD,size:23.5×23.5mm,withlargeOKandALARMindicator
Display functionality -Status:Conf,Start,Run,Stop -Currenttemperaturemeasurement -OK,ALARMorblank -Configurationprofileidentification
Report Built-inPDFfilegeneratorautomaticallyestablishesanevaluationreportwithembeddeddataupon
connectiontoanUSBport.ComplieswiththeISOstandard19005-1DocumentManagementforthe long-termpreservationofelectronicdocuments(PDF/A)andFDA21CFRPart11. -Customizablereporttitleandfilename
-Textareaforadditionalinformation(e.g.shipmentinformation,instructionsforrecipient,etc.) -Statistics(min/max,average,alarm)anddetailedloggerinformation(ID,configuration,etc.) -Chartvisualizingthetemperaturecurveandalarmlimits
Traceability IDnumber(traceabletocomponentlevel)
Case | dimension | weight ABSplasticmaterial|96×37×11mm(3.8×1.5×0.5inch)|38g(1.34oz)
Certifications CE|EN12830|EMC|RTCADO-160|RoHS|WHO-PQS|SafeTransportofChemicalGoods
Data logger configuration liberoCONFIGandSmartStartsoftwaretocreate,storeandmanageindividualsettingsinasettingsprofile. and analysis software liberoMANAGERdatabaseascloudservicetostoreshipmentreportsandresults,analyzeshipments, managereleasestatusandtriggercustomerspecificworkflowsincoldchainprocesses. elproVIEWERsoftwaretoaccessandexportembeddeddataofPDFreport,fordataanalysis andcomprehensivereportfeatures(forusersnotusingliberoMANAGER).
±0.5°Cfrom-10.0°Cto-0.1°Caswellas+25.1°Cto+50.0°C
Technical Specifications LIBERO CS Version 2019
TS_LIBER
O_C
S_EN
_
V9_1
9
±1.0°Cfrom-30.0°Cto-10.1°Caswellas+50.1°Cto+70.0°C