latin america training. ross electrodes training
TRANSCRIPT
Latin America Training
ROSS ElectrodesTraining
3
ROSS Electrodes
• Premium pH electrode available
• Specially formulated reference system provides quickest response at temperature changes
• Offer less drift and greater stability than other pH electrodes
• Double junction reference system
• Can be used with biotech & TRIS samples
• Ability to use different fill solutions based on sample composition
• Environmentally friendly
• Also available as Triode
4
ROSS Electrode Specifications
• Drift: Less than 0.002 pH units/day, minimizes restandardization
• Accuracy of measuring a pH 6.86 buffer after standardization at 25°C: Accurate within 0.03 pH for buffer at any temperature between 0-100°C using automatic temperature compensation.
• Speed of response in 6.86 buffer going from 25° C to 75°C: Response stable to 0.01 pH within 30 seconds.
• Speed of response between 6.86 and 4.01 buffers at 25°C: Response stable to 0.002 pH within 15 seconds.
• Precision: 0.01 pH
• Triode temperature accuracy: Accurate to +/- 1o C within 2 minutes
5
ROSS Temperature Response
• ROSS reference system uses iodide/iodine which is water soluble. ROSS reference system is independent of temperature changes.
• Platinum in reference system prevents junction problems caused in conventional electrodes by temperature changes
6
ROSS pH Electrode Families
• Standard ROSS electrodes• Wide variety to choose from
• Meets all ROSS specifications
• 12 month warranty
• Available with BNC connectors, screw caps and US standard connectors
• ROSS Sure-Flow• Unique free flowing junction prevents clogging
• 3 models to choose from
• Available mostly as BNC connector, but also EDIN
• ROSS Ultra• Improved reference system
• Less drift than standard ROSS models
• Available with BNC only
• 24 month warranty
7
ROSS pH Electrode Families
• ROSS Ultra Triodes• Newest Line of ROSS• Only ROSS available as a low-maintenance electrode• 4 models as epoxy refillable• 4 models as epoxy gel• 1 model to soon be released as glass refillable• Different connectors for thermistor to fit different meters
• PerpHecT ROSS• For use with LogR meters• Utililize glass pH membrane to measure temperature• Temperature accuracy of 0.5oC• Variety of models available• Available only with BNC connector
• ROSS Micro• 1 model with BNC connector• For small sample size• Also can be used with LogR meters
8
Reference junctions Used in ROSS Electrodes
• Ceramic – Used in most glass refillable electrodes
• Glass fiber – Used in most refillable epoxy electrodes
• Glass Capillary – New junction! Used in new ROSS gel triode
9
Reference junctions Used in ROSS Electrodes
• Sure-Flow – Glass to glass or glass to epoxy junction
• Glass sleeve – Specialty electrode, used in 8162SC
10
Electrode Connectors –
• BNC Connector• Waterproof BNC
• US Standard Connector
• Screw Cap Electrode
• Pin Tip Connector
(for reference half cell)
11
ROSS Half Cell Electrodes – Standard & Ultra
800300
Sure Flow Reference Electrode
8101BNWP
pH Half Cell
800500U
ROSS Ultra Reference Electrode
12
Standard ROSS Line – Glass Electrodes
8102BN
Standard size
120 X 12mm
8103BN
Semi-Micro
6mm tip
8104BN
Rugged bulb
8162SC
Fixed glass sleeve titration
joint 14/15
8163BNWP
Spear tip
For piercing, cheese,
fruits
13
Standard ROSS Line – Epoxy Electrodes
8115BN
Semi-Micro
6mm tip
8135BN
Flat surface
For agar gel plates, cheese & meats
815600
Standard size for more rugged use
14
ROSS Sure-Flow Electrodes
8165BNWP
Epoxy Body
Standard size
120 X 12 mm
Has replaceable bulb guard
8172BNWP
Glass Body
Standard size
Has replaceable bulb guard
8175BNWP
Epoxy body
Semi-micro
8mm tip
Has replaceable bulb guard
15
PerpHecT ROSS Line
8202BN
Glass body
Standard size
120 X 12mm
8203BN
Glass Body
Semi-Micro
6mm tip
8220BNWP
Glass Body
Micro tip
3mm diameter
8235BN
Epoxy body
Flat surface
For moist surfaces
8272BN
Glass Body
Sure-Flow
Has replaceable bulb guard
8256
Epoxy body
Standard size
16
ROSS Ultra Electrodes
8102BNUWP
Glass body
Standard size
120 X 12mm
8103BNUWP
Glass Body
Semi-Micro
6mm tip
8104BNUWP
Glass Body
Rugged bulb
8115BNUWP
Epoxy body
Semi-micro
8mm tip
8156BNUWP
Epoxy Body
Standard Size
8135BNUWP
Epoxy Body
Flat Surface
Standard Size
17
ROSS Ultra Epoxy Body Refillable Triodes
• Four models to accommodate different meters• Faster temperature response that Ag/AgCl triodes• Only ROSS electrodes not to have a coil• 24 month warranty• Performs extremely well in clean water samples• Able to stand up to TRIS buffers with no effect on reference junction
• 8157BNUMD – For Star meters, 1 meter cable• 8157UWMMD – For Star meters, 3 meter cable• 8157BNU – For A+ meters, 1 meter cable• 8257BNU – For PerpHecT meters, 1 meter cable
18
Orion Refillable ROSS Ultra Triode Strengths
• Placement and small-diameter design of the Orion ROSS Ultra Triode thermistor makes the temperature response fast and accurate.
• The ROSS reference system makes the electrode pH response very accurate, fast, and reproducible at elevated temperatures and after use in elevated temperatures.
• Junction testing shows the linear flow glass fiber reference junction performs well: has low stir potential and accurate response upon changing the ionic strength of the sample. It does not clog when soaked in Tris Buffer solution.
• The epoxy body is durable and protects the glass bulb from damage• The refillable Orion ROSS Ultra Triode is the star performer in clean water
samples (tap and DI tested) – fast, stable, accurate. • Even when cycled with a strong acid, strong base, and high ionic strength pH buffer
6.86, response is quick, stable, and accurate.
• It also performs well in Tris Buffer at 37C and in creamy salad dressing at 35C.
19
ROSS Ultra Epoxy Body Low-Maintenance Triodes
• Four models to accommodate different meters• Faster temperature response that Ag/AgCl manuals• Only ROSS electrodes not to have a coil• Only low-maintenance ROSS electrodes• Superior performance to competitors in low ionic strength solutions• More accurate and stable than competitors electrodes at
temperature extremes• 18 month warranty
• 8107BNUMD – For Star meters, 1 meter cable• 8107UWMMD – For Star meters, 3 meter cable• 8107BNU – For A+ meters, 1 meter cable• 8207BNU – For PerpHecT meters, 1 meter cable
20
Competitive Comparison of Gel ROSS Ultra™ Triode in Low Ionic Strength Standard
Low Ionic Strength Standard has expected pH value of 4.005 at 0oC to 30oC
The Gel ROSS Ultra Triode has superior performance in the low ionic strength solutions; it demonstrated better accuracy and response time than the other electrodes
3.70
3.75
3.80
3.85
3.90
3.95
4.00
4.05
4.10
4.15
0 200 400 600 800 1000 1200
Response Time, sec.
pH
Re
ad
ing
Aft
er
Ca
libra
tio
n
ROSS Triode
Brand A
Brand B
Brand C
Expected
21
pH Response in pH 4 Buffer at Different Temperatures
Delta from expected pH in pH 4.01 buffer at ~ 10C
-0.200
-0.100
0.000
0.100
0.200
0 200 400 600
Time, sec.
Del
ta p
H
ROSS Triode 1
ROSS Triode 2
Brand B
Brand C
Delta from expected pH in pH 4.01 buffer at ~ 55C
-0.200
-0.100
0.000
0.100
0.200
0 200 400 600
Time, sec.
Del
ta p
H
ROSS Triode 1
ROSS Triode 2
Brand B
Brand C
Delta from expected pH in pH 4.01 buffer at ~ 23C after 10C and 55C
-0.200
-0.100
0.000
0.100
0.200
0 100 200 300 400 500
Time, sec.
Del
ta p
H
ROSS Triode 1
ROSS Triode 2
Brand B
Brand C
Two ROSS Gel Triodes read close to each other and close to the expected values at about 10oC, 55oC, and again at 23oC.
Brand B is good except at cold temp., when readings deviated from the expected value (about 0.05 pH).
Brand C is less accurate when reading buffers at different temperatures; it has large deviations at cold temp (about 0.2 pH) and after return to RT (>0.1pH).
22
Temperature Response at ~55C
Temperature response at ~55C in pH 4
40.0
44.0
48.0
52.0
56.0
60.0
0 100 200 300 400 500 600 700
Time, sec.
Tem
pera
ture
deg
rees
C
ROSS #1
ROSS #2
Brand B
Brand C
Temp. response ar ~ 57 in pH 10 buffer
40.0
44.0
48.0
52.0
56.0
60.0
0 100 200 300 400
Time, sec.
Tem
p. d
egre
es C
ROSS #1
ROSS #2
Brand B
Brand C
ROSS Triodes are faster at elevated temperature: they reached stable temp. readings in 30 seconds.It takes about 2 min. for Brand B and more than 5 min. for Brand C electrode to reach stable temperature readings at ~ 55C.
23
ROSS Ultra Glass Triode
• Just released in July 2010• 1 model for Star meters, 8302BNUMD• First Orion glass triode• Temperature response faster than most glass triodes on market• New ceramic junction
24
ROSS Electrodes
Thank you to Gayle Gleichauf and Nadia Vilissova for graphs and competitive testing data of ROSS refillable and low maintenance Triode
Thank you!