Technical Features
Hydrogen G200, G400, G600 Laboratory Hydrogen Gas Generator
www.protononsite.com | T 203.949.8697 | F 203.949.8016 | Proton OnSite 10 Technology Drive Wallingford, CT 06492 | [email protected]
TECHNICAL FEATURES G200 G400 G600
FLOW RATE* 200 cc/min 400 cc/min 600 cc/min
TECHNOLOGIES EMPLOYED Proton Exchange Membrane (PEM) Solid Electrolyte
Maintenance-Free and Caustic-Free
OUTPUT PRESSURE 3 to 8 barg / 43 to 116 psig
PURITY 99.9995% (H2O < 5 ppmv, O2 < 1 ppmv)
DI WATER SUPPLY Deionized, ASTM Type II, > 1 Meg Ohm-cm (< 1 microSiemen/cm)
< 0.5 ppm, Total Dissolved Solids
DI WATER FEED PRESSURE
SPEC FOR AUTO-FILL
> 0.7 barg (10 psig)
WATER CONSUMPTION RATE 12 cc/hr 0.29 L/day
24 cc/hr 0.58 L/day
36 cc/hr 0.86 L/day
POWER CONSUMPTION RATE < 110 W < 180 W < 250 W
ELECTRICAL SUPPLY 120 - 230 VAC, Auto Ranging, 50/60 Hz
INGRESS PROTECTION
RATING
IP20
OPERATING TEMPERATURE 40°F to 104°F / 5°C to 40°C
ENVIRONMENTAL MOISTURE 80% RH, non-condensing
MAX ALTITUDE 2000 m / 6560 ft above sea level
UNIT DIMENSIONS 14.5”(H) x 20.5” (D) x 9” (W)
36.8 cm (H) x 52 cm (D) x 22.9 cm (W)
UNIT WEIGHT 67 lbs / 30 kg
ADDITIONAL FEATURES
Additional Desiccant Cartridge
Maintenance Diary
Removable Water Feed Tank
Report Alarms
Touch Screen Display
PD-0600-0100 Rev C
© 2014 Proton OnSite. All Rights Reserved.
Proton, Proton OnSite, Proton Energy Systems and the Proton symbol are trademarks of
Proton Energy Systems, Inc. d/b/a Proton OnSite.
* Measured at standard conditions: 21°C, 1 Atmosphere
Technical Specifications
www.protononsite.com | T 203.949.8697 | F 203.949.8016 | Proton OnSite 10 Technology Drive Wallingford, CT 06492 | [email protected]
Hydrogen G4800 Laboratory Hydrogen Gas Generator
TECHNICAL FEATURES G4800
PURE HYDROGEN OUTPUT RATE* 4800 cc / min
ELECTROLYTE SYSTEM Proton Exchange Membrane (PEM) solid electrolyte, maintenance and caustic-free
HYDROGEN PURIFICATION TECHNOLOGY PSA, Molecular sieve
HYDROGEN INVENTORY AT FULL CAPACITY < 2 grams
DELIVERY PRESSURE PSIG (BAR) 0-200 psig / 0-13.8 barg w/optional regulator
PURITY (ASSAY, %) 99.9999+%
IMPURITIES BY SPECIES <1 ppm water, <1 ppm total N2 & O2, <1 ppb hydrocarbons/halocarbons
AUTO-FILL WATER REPLENISHMENT Standard
WATER QUALITY REQUIRED (MINIMUM) Deionized, ASTM Type II Deionized Water (<1 micro Siemen/cm, >1 MegOhm-cm)
ELECTRICAL SUPPLY REQUIRED 205 to 240 VAC, single phase, 50 or 60 Hz
ELECTRICAL SERVICE RATING 4KVA
CORDSET Hard wired connection
OPERATING ENVIRONMENT Indoor, non-classified, 40°F to 95°F (5°C to 35°C), < 90% RH, ventilated
DUTY CYCLE 100% (24/7/365)
DIMENSIONS (W x D x H)
(Product / Est. Shipping)
31” x 38” x 42” (79x 97x 107 cm) / 38” x 45” x 52” (97 x 114 x 132 cm)
WEIGHT (Product / Est. Shipping) 475 lbs (215 kg) / 625 lbs (284 kg)
PRODUCT CERTIFICATIONS cTUVus (UL and CSA equivalent), CE (PED, ATEX, LVD, PED, EMC) NYFD Approval
CONTROLS AND AUTOMATION LED high visibility display; Adjustable pressure output; Remote alarm terminal/output;
Switchable psi/kPa pressure reading; Ethernet and RS232
DIAGNOSTICS AND WARNINGS Water quality, water level, comprehensive self-diagnostics, leak sensing
MAINTENANCE Annual filter changes
OPTIONS Deionized water supply system, Auto-crossover manifold for backup gas supply, Pressure
reducing regulator
WARRANTY 1 year
*Output measurement at 21°C and 1 atmosphere
Consult Proton Onsite Applications and Department for proper installation guidelines. Specifications subject to change.
PD-0600-0060 Rev. E
© 2014 Proton OnSite. All Rights Reserved.
Proton, Proton OnSite, Proton Energy Systems and the Proton symbol are trademarks of
Proton Energy Systems, Inc. d/b/a Proton OnSite.
Technical Specifications
HOGEN®
S Series Hydrogen Generation Systems
www.protononsite.com | T 203.949.8697 | F 203.949.8016 | Proton OnSite 10 Technology Drive Wallingford, CT 06492 | [email protected]
S10 S20 S40
DESCRIPTION
On-site hydrogen generator in an integrated, automated, site-ready enclosure. Load Following operation automatically adjusts output to match demand.
ELECTROLYTE
Proton Exchange Membrane (PEM) - caustic-free
HYDROGEN PRODUCTION
Net Production Rate: Nm3/hr @ 0°C, 1 bar SCF/hr @ 70°F, 1 atm SLPM @ 70°F, 1 atm kg per 24 hours
0.265 Nm3/hr 0.53 Nm3/hr 1.05Nm3/hr 10 SCF/hr 20 SCF/hr 40 SCF/hr 4.7 SLPM 9.4 SLPM 18.8 SLPM 0.57 kg/24hr 1.14 kg/24hr 2.27 kg/24hr
Delivery Pressure - Nominal 13.8 barg (200 PSIG)
Power Consumed per Volume of H2 Gas Produced
6.7 kWh / Nm3 17.6 kWh / 100 ft3
Purity (Concentration of Impurities) 99.9995% (Water Vapor < 5 PPM, -65°C (-85°F) Dewpoint, N2 < 2 PPM, O2 < 1 PPM, All Other Undetectable)
Turndown Range 0 to 100% net product delivery
Upgradeability N/A
DI WATER REQUIREMENT
Rate at Max Consumption Rate 0.235 L/hr (0.065 gal/hr) 0.47 L/hr (0.13 gal/hr) 0.94 L/hr (0.25 gal/hr)
Temperature 5°C to 35°C (41°F to 95°F)
Pressure 1.5 to 4 barg (21.8 to 58.0 PSIG)
Input Water Quality ASTM Type II Deionized Water required, < 1 micro Siemen/cm (>1 megOhm-cm) ASTM Type I Deionized Water preferred, < 0.1 micro Siemen/cm (> 10 megOhm-cm)
PD-0600-0061 Rev. C © 2011 Proton Energy Systems, Inc. All Rights Reserved. Proton, Proton OnSite, Proton Energy Systems, and HOGEN are trademarks of Proton Energy Systems, Inc. d/b/a Proton OnSite.
S10 S20 S40
HEAT LOAD AND COOLANT REQUIREMENT
Cooling Air-Cooled; Ambient Air, 5°C to 40°C (41°F to 104°F)
Max. Heat Load from System 1.1 kW / 3,754 BTU/hr 2.2 kW / 7,507 BTU/hr 4.3 kW / 14,673 BTU/hr
ELECTRICAL SPECIFICATIONS
Recommended Breaker Rating 4 kVA 8 kVA 12 kVA
Electrical Specification 205 to 240 VAC, single phase, 50 or 60 Hz
INTERFACE CONNECTIONS *Consult Installation Manual for details*
H2 Product Port 1/4” CPI™ compression tube fitting, SS
H2 / H2O Vent Port 1/2” CPI™ compression tube fitting, SS
DI Water Port 1/4” tube push-to-lock, polypropylene
Calibration-Gas Port N/A
Coolant Supply Port N/A
Coolant Return Port N/A
Drain Port 1/4” tube push-to-lock polypropylene
Electrical Connect to on-board circuit breaker
Communications RS 232, Ethernet
CONTROL SYSTEMS
Standard Features Fully automated, push button start/stop. E-stop. On-board H2 Leak detection. Automatic fault detection and system depressurization.
Remote Alarm Form C relay 2A/30VDC rated switching
Remote Shutdown Circuit breaker shunt trip
ENCLOSURE CHARACTERISTICS
Dimensions, W x D x H (Product / Est. Shipping)
31” x 38” x 42” (79 x 97 x 107 cm) / 38” x 45” x 52” (97 x 114 x 132 cm)
Weight (Product / Est. Shipping) 475 lbs (216 kg) / 650 lbs (295 kg)
Rating IP22
ENVIRONMENTAL CONSIDERATIONS *Do Not Freeze*
Standard Siting Location Indoor, level ± 1°, 0 to 90% RH non-condensing, Non-hazardous/non-classified environment
Storage/Transport Temperature 5°C to 60°C (41°F to 140°F)
Ambient Temperature Range 5°C to 40°C (41°F to 104°F)
Altitude Range - Sea Level to: 1520 m (5000 ft)
Ventilation Proper ventilation must be provided from a non-hazardous area, at a rate in accordance with IEC60079-10, Zone 2 NE
SAFETY AND REGULATORY CONFORMITY
Maximum On-board H2 Inventory at Full Production
0.016 Nm3 / 0.6 SCF / 0.0014 kg
Cabinet Ventilation with Environment NFPA 69 and EN 1127-1, Clause 6.2. Vent fan draws fresh air up to 28 Nm3/min (1000 ft3/min)
Noise dB(A) at 1 Meter < 70
Approvals cTUVus (UL and CSA equivalent), CE (PED, ATEX, LVD, Mach. Dir. EMC), NYFD Approval
OPTIONS
Proton Onsite offers a wide range of options to tailor your HOGEN hydrogen generation system to meet your specific operational requirements. Please contact your local representative to discuss the current list of options available to best fit your needs.
Consult Proton Onsite Applications Department for proper installation guidelines. Specifications subject to change.
Technical Specifications
HOGEN® H Series
Hydrogen Generation Systems
H2m H4m H6m
DESCRIPTION
On-site hydrogen generator in an integrated, automated, site-ready enclosure.
Load Following operation automatically adjusts output to match demand.
ELECTROLYTE
Proton Exchange Membrane (PEM) - caustic-free
HYDROGEN PRODUCTION
Net Production Rate:
Nm3/hr @ 0°C, 1 bar
SCF/hr @ 70°F, 1 atm
SLPM @ 70°F, 1 atm
kg per 24 hours
2 Nm3/hr
76 SCF/hr
35.8 SLPM
4.31 kg/24hr
4 Nm3/hr
152 SCF/hr
71.7 SLPM
8.63 kg/24hr
6 Nm3/hr
228 SCF/hr
107.6 SLPM
12.94 kg/24hr
Delivery Pressure - Nominal 15 barg (218 PSIG) / 30 barg option (435 PSIG)
Power Consumed per Volume of H2 Gas Produced 7.3 kWh/Nm3
19.2 kWh/100ft3
7.0 kWh/Nm3
18.5 kWh/100ft3
6.8 kWh/Nm3
17.8 kWh/100ft3
Purity (Concentration of Impurities) 99.9995% (Water Vapor < 5 PPM, -65°C (-85°F) Dewpoint, N2 < 2ppm
O2 < 1 PPM, All Others Undetectable)
Turndown Range 0 to 100% net product delivery (Automatic)
Upgradeability Field upgradeable to a maximum of
6 Nm3/hr (228 SCF/hr)
N/A
DI WATER REQUIREMENT
Rate at Max Consumption Rate 1.83 L/hr (0.50 gal/hr) 3.66 L/hr (0.96 gal/hr) 5.50 L/hr (1.42 gal/hr)
Temperature 5°C to 50°C (41°F to 122°F)
Pressure 1.5 to 4 barg (21.8 to 58.0 PSIG)
Input Water Quality ASTM Type II Deionized Water required, < 1 micro Siemen/cm (>1 megOhm-cm)
ASTM Type I Deionized Water preferred, < 0.1 micro Siemen/cm (>10 megOhm-cm)
www.protononsite.com | T 203.949.8697 | F 203.949.8016 | Proton OnSite 10 Technology Drive Wallingford, CT 06492 | [email protected]
H2m H4m H6m
HEAT LOAD AND COOLANT REQUIREMENT
Cooling1 Liquid-Cooled; Anti-freeze, non-fouling; 5°C to 35°C (41°F to 95°F)
*25°C cooling water maximum for ambient temperatures above 40°C
Max. Heat Load
(Cooling Requirement)
8.1 kW / 27,368 BTU/hr /
(2.3 tons refrig)
16.1 kW / 54,936 BTU/hr /
(4.6 tons refrig)
23.7 kW / 80,868 BTU/hr /
(6.8 tons refrig)
ELECTRICAL SPECIFICATIONS
Recommended Breaker Rating 22 kVA 40 kVA 58 kVA
Electrical Specification 380 to 480 VAC, 3 phase, 50 or 60 Hz
INTERFACE CONNECTIONS *Consult Installation Manual for details*
H2 Product Port 1/4” Parker CPI™ compression tube fitting, SS
H2/H2O Vent Port 1/2” FNPT, SS
DI Water Port 1/4” FNPT, SS
Calibration-Gas Port 1/8” FNPT, brass
Coolant Supply Port 1” FNPT, brass
Coolant Return Port 1” FNPT, brass
Drain Port 3/8” FNPT, brass
Electrical Connect to on-board circuit breaker
Communications Ethernet
CONTROL SYSTEMS
Standard Features Fully automated, push button start/stop. E-stop. On-board H2 leak detection.
Automatic fault detection and system depressurization
Remote Alarm Form C relay, 5A, 250V, 150W Max. rated switch
Remote Shutdown Safety circuit trip
ENCLOSURE CHARACTERISTICS
Dimensions, W x D x H
(Product / Est. Shipping)
71” x 32 “ x 75” (180 cm x 81 cm x 191 cm) / 81” x 41” x 85” (206 cm x 104 cm x 216 cm)
Note: add 8 cm (3”) to height for installed lifting brackets
Weight
(Product / Est. Shipping)
1500 lbs (682 kg) /
1776 lbs (807 kg)
1600 lbs (727 kg) /
1887 lbs (858 kg)
1700 lbs (773 kg) /
1998 lbs (908 kg)
Rating IP66 for electronics compartment. IP43 for fluids compartment; Upgradeable to IP56.
ENVIRONMENTAL CHARACTERISTICS *Do Not Freeze*
Standard Siting Location Indoor, level ± 1°, 0 to 90% RH non-condensing, Non-hazardous/non-classified environment
Storage / Transport Temperature 5°C to 60°C (41°F to 140°F)
Ambient Temperature Range 5°C to 50°C (41°F to 122°F)
Altitude Range - Sea Level to: 2400 m (7874 ft)
Ventilation Proper ventilation must be provided from a non-hazardous area, at a rate in accordance with IEC60079-10, Zone 2 NE
SAFETY AND REGULATORY CONFORMITY
Maximum On-board H2 Inventory at
Full Production
0.040 Nm3 @ 15 barg; 0.08 Nm3 @ 30 barg
1.5 SCF @ 15 barg; 2.9 SCF @ 30 barg
0.0036 kg @ 15 barg; 0.0069 kg @ 30 barg
Cabinet Ventilation with Environment NFPA 69 and EN 1127-1, Clause 6.2. Vent fan draws fresh air up to 28 Nm3/min (1000 ft3/min)
Noise dB(A) at 1 Meter <83
Approvals cTUVus (UL and CSA equivalent), CE (PED, ATEX, LVD, Mach. Dir., EMC), NYFD
OPTIONS
Proton Onsite offers a wide range of options to tailor your HOGEN hydrogen generation system to meet your specific operational requirements.
Please contact your local representative to discuss the current list of options available to best fit your needs.
PD-0600-0062 Rev. D
© Proton Energy Systems, Inc. All Rights Reserved.
Proton, Proton OnSite, Proton Energy Systems and HOGEN are
trademarks of Proton Energy Systems, Inc. d/b/a Proton OnSite
1Consult Proton Onsite Applications Department for proper installation guidelines. Specifications subject to change.
C Series Hydrogen Generation Systems
www.protononsite.com | T 203.949.8697 | F 203.949.8016 | Proton OnSite, 10 Technology Drive, Wallingford, CT 06492 | [email protected]
Model C10 C20 C30
DESCRIPTION
On-site hydrogen generator in two integrated, automated, site-ready enclosures.
Dual-mode Operation (Selectable):
Load Following mode automatically adjusts output 0-100% to match demand.
Tank Filling mode operates with power-conservation mode during standby.
ELECTROLYTE
Proton Exchange Membrane (PEM) - caustic-free
HYDROGEN PRODUCTION
Net Production Rate:
Nm3/hr @ 0° C, 1 bar
SCF/hr @ 70° F, 1 atm
SLPM @ 70° F, 1 atm
kg per 24 hours
10 Nm3/hr
380 SCF/hr
179 SLPM
21.6 kg/24hr
20 Nm3/hr
760 SCF/hr
359 SLPM
43.3 kg/24hr
30 Nm3/hr
1140 SCF/hr
538 SLPM
65.0 kg/24hr
Delivery Pressure - Nominal 30 barg (435 PSIG)
Power Consumed per Volume of Mass H2 Gas
Produced*
6.2 kWh/Nm3
16.3 kWh/100 ft3
68.9 kWh/kg
6.0 kWh/Nm3
15.8 kWh/100 ft3
66.7 kWh/kg
5.8 kWh/Nm3
15.2 kWh/100 ft3
64.5 kWh/kg
Purity (Concentration of Impurities)* ISO 14687-1 Type 1 grade C ISO 14687-2 Type 1 grade D
99.9998% Water Vapor < 2 PPM, -72°C (-98°F) Dewpoint, N2< 2 PPM, O2< 1 PPM, All others undetectable
Turndown Range 0 to 100% Net Product Delivery (Automatic)
Upgradeability Field Upgradeable to a maximum of 30 Nm3/hr (1140 SCF/hr) N/A
DI WATER REQUIREMENT
Rate at Max. Consumption Rate 9 L/hr (2.4 gal/hr) 17.9 L/hr (4.7 gal/hr) 26.9 L/hr (7.1 gal/hr)
Temperature +5°C to 40°C (41°F to 104°F)
Pressure 1.0 to 4.1 barg (10 to 60 PSIG)
Input Water Quality ASTM Type II Deionized Water required, < 1 micro Siemen/cm (>1 MegOhm-cm)
ASTM Type I Deionized Water recommended, <0.1 micro Siemen/cm (> 10 MegOhm-cm)
HEAT LOAD AND COOLANT REQUIREMENT*
Coolant Liquid Cooled; Non-Fouling*
Max. Heat Load 33.5 kW (114,307 BTU/hr)
(9.6 tons refrigeration)
66.8 kW (227,931 BTU/hr)
(19.0 tons refrigeration)
100.2 kW (341,897 BTU/hr)
(28.6 tons refrigeration)
Coolant Temperature 5°C to 40°C (41°F to 104°F)
Coolant Flowrate Up to 110 L/min (24 gal/min) Up to 229 L/min (50 gal/min) Up to 411 L/min (90 gal/min)
Pressure Drop (at full flow) Up to ~3.0 barg (~45 PSIG) Up to ~3.0 barg (~45 PSIG) Up to ~3.0 barg (~45 PSIG)
* Consult Proton Applications Engineering Department for specific requirements and options.
Model C10 C20 C30
ELECTRICAL SPECIFICATIONS
Recommended Breaker Rating 100 kVA 200 kVA 275 kVA
Electrical Specification 380,400,415 VAC, 3 phase, 50 Hz (+/- 10% from nominal voltage)
480VAC, 3 phase, 60 Hz (+/- 10% from nominal voltage)
INTERFACE CONNECTIONS - Consult Mechanical Interface Diagram drawing, PD-9900-0014 for details -
H2 Product Port 3/8” Parker CPI™ compression tube fitting, SS
H2 Vent Port 1” Parker CPI™ compression tube fitting, SS
DI Water Port 1/2” FNPT, SS
Coolant Supply and Return Ports Electrolyzer Enclosure: 1-1/2” MNPT, brass (Cell Stack); 1/2” FNPT, brass (Hydrogen Dryer)
Power Supply Enclosure 3/4” MNPT, brass (Power Supply Cooling)
Drain Port 1/2” FNPT, brass
O2 Vent Port 1” Parker CPI ™ compression tube fitting, SS
Electrical Electrical terminals inside supplied fused power Disconnect Box
Communications Ethernet
CONTROL SYSTEMS
Standard Features - Fully automated - Remote start/stop
- E-Stop - On-board H2 leak detection
- Automatic fault detection and system depressurization
Remote Alarm Form C relay, 5A, 250V, 150W Max. rated switching
Remote Shutdown Safety circuit trip
ENCLOSURE CHARACTERISTICS
Dimensions, W x D x H (Product) Electrolyzer Enclosure: 2388 mm x 914 mm x 2007 (94” x 36” x 79”)
Power Supply Enclosure: 1880 mm x 914 mm x 2007 mm (74” x 36” x 79”)
Weight (Product) 2041 kg (4500 lbs) 2449 kg (5400 lbs) 2812 kg (6200 lbs)
IP Rating Overall unit rating of IP56
ENVIRONMENTAL CONSIDERATIONS - Do Not Freeze -
Standard Siting Location Indoor/Sheltered; level ±1°, 0 to 100% RH non-condensing;
Non-hazardous/non-classified environment
Storage/Transport Temperature +5°C to 60°C (41°F to 140°F)
Ambient Temperature Range +5°C to 40°C (41°F to 104°F)
Altitude Range - Sea Level to: 2000 m (6562 ft)
Ventilation Proper ventilation must be provided from a non-hazardous area at a rate in accordance with IEC60079-
10, Zone 2 NE
SAFETY AND REGULATORY CONFORMITY
Maximum On-board H2 Inventory at
Full Production
0.10 Nm3 4 SCF
0.009 kg
0.14 Nm3 5.4 SCF 0.013 kg
0.16 Nm3 6 SCF
0.014 kg
Cabinet Ventilation with Environment Vent fan draws fresh air up to 8.5 m3/min (300 ft3/min)
Noise dB(A) at 1 Meter < 75
Approvals cTUVus, Tmark (ISO22734-1, UL508A) CE (PED, ATEX, LVD, Mach. Dir., EMC)
OPTIONS
Factory Matched RO/DI Water System
Factory Matched Cooler/Chiller
Dew Point Monitoring
Low Ambient Temp. Package
(-10 to 40°C)
Equipment Orientation
High Ambient Temp. Package
(+5 to 50°C)
Cold Ambient Temp. Package
(-20 to 40°C)
Proton OnSite offers a wide range of options to tailor your HOGEN Hydrogen Generation System to meet your specific operational requirements.
Please contact your local representative to discuss the current list of options available to best fit your needs.
Proton OnSite reserves the right to make changes to this product without further notice.
*Consult Proton OnSite Applications Engineering Department for proper installation guidelines.
PD-0600-0068 Rev. C
© 2011-2013 Proton Energy Systems, Inc. All Rights Reserved.
Proton, Proton OnSite, Proton Energy Systems and HOGEN are trademarks
www.protononsite.com | T 203.949.8697 | F 203.949.8016 | Proton OnSite 10 Technology Drive Wallingford, CT 06492 | [email protected]
Technical Specifications
StableFlow®
Power Generator Hydrogen Monitoring
and Optimization System
SYSTEM FUNCTIONS
Continuous Monitoring Hydrogen Purity (H2 in Air) and Dew Point
Measurement and Control Hydrogen Purity and Dew Point while maintaining specified pressure
Re-Gas Monitoring H2 in Air, H2 in CO2, Air in CO2
SYSTEM REQUIREMENTS
Electrical Specification 110 to 240 VAC, single phase, 5 amps, 50 or 60 Hz
Inlet H2 Pressure 1.7 to 5.2 bar (25-75 psig)
Electrical Cabinet Purge Gas Instrument Air - 1/4” Parker CPI™ Compression fitting, <16 SLPM
(≈ 0.6 SCFM, 5.5-8.2 bar (80-120PSIG))
Purge Gas Quality Clean, dry air per ISO 8573.1 Class 5.4.5
Operating Temperature 4° to 50°C (40° to 120°F)
Communications Ethernet - RJ45
Hydrogen Inlet, Outlet Connections 1/4” Parker CPI™ Compression fitting
SYSTEM FEATURES
Dimensions (W x D x H) (Product / Est. Shipping) 39” x 23” x 82” (99 cm x 58 cm x 208 cm) / 47” x 40” x 90” (119 cm x 102 cm x 229 cm)
Weight (Product / Est. Shipping) 375 lbs (170 kg) / 730 lbs (332 kg)
Standard Features Fully automated. Data Logging. Remote Telemetry. Local operator interface.
StableFlow software control. Integrated System Controller.
Standard Siting Location Indoor. Floor mount. ± 3° of level
DEW POINT SENSOR
Dew Point Range -80°C to +20°C, (0.5 ppmv min.)
Accuracy ± 3°C or better, (± 0.015 ppmv)
Repeatability ± 0.5°C or better, (± 0.002 ppmv)
Response Time < 10 seconds for 63% step change in moisture content (wet-up or dry-down cycle)
HYDROGEN PURITY MONITOR
Measurement Technique Vibrating Element
Response Time < 10 sec
Linearity ± (0.001 + 1% of set range)
Repeatability ±0.001
Long Term Stability ±0.003/month
PRESSURE TRANSDUCER
Operating Pressure Range 0 to 6.9 bar (0 to 100 PSIG)
Operating Temperature -30°C to +50°C
Accuracy ±0.25% Full Scale
4 - 20 MA OUTPUTS
Signals Pressure, Purity, Dew Point, Bleed Valve Position
Output 4-20 mA DC
Customer Load ≤ 600 ohm Max.
4-20 mA Scale Pressure: 0 to 6 bar (0 to 100 PSIG) H2 Purity in Air: 85 to 100%
Dew Point: -80 to 20°C Valve Position (Open): 100 to 0%
DESCRIPTION
Monitors and controls the electric power generator’s hydrogen atmosphere to generator OEM specified levels.
APPLICABILITY
StableFlow Hydrogen Control system is applicable to all hydrogen-cooled electric power generators.
SAFETY AND REGULATORY CONFORMITY
CSA/EN/IEC/UL 61010-1, NFPA 70 (NEC), NFPA 79, NFPA496; EN/IEC 60204-1, EN/ISO 12100-1, EN/ISO 12100-2, EN 1127-1, EN 50014, EN 5020, IEC 60079-0,
EN/IEC 60079-1, EN60079-2, IEC 60079-11, IEC 60079-14
Area Classification: Class 1, Div 2 / Zone 2
Consult Proton Onsite Applications Department for proper installation guidelines. Specifications subject to change.
PD-0600-0059 Rev. D
© 2011 Proton Energy Systems, Inc. All Rights Reserved.
StableFlow, StableFlow Hydrogen control system and Design
Proton, Proton OnSite, Proton Energy Systems, the Proton symbol, and HOGEN are
trademarks of Proton Energy Systems, Inc. d/b/a Proton OnSite.