INW Sensor Technology
…Continued
What we will cover
• Installation Practices• Security• Changing Batteries• Aqua4Plus Operation• Calibration Overview
Installation Practices
Installation PracticesAbove Ground Monument
Sensor
CableWell casing
Strain relief
Desiccant
Service port
Well capMonument
• Steel or fiberglass monument
• Steel or fiberglass flush mounted lid
Installation PracticesBelow Ground Monument
Sensor
CableWell casing
Strain relief
Desiccant*
Service port
Well cap
Monument Lid
* Prone to flooding!!Keep desiccant dry or use absolute sensors.
Installation Practices
Do not drop into well
– Lower gently to prevent damage to end of sensor.
Installation Practices
Do not scrape cable over edge of well
– May nick or fray cable.
Installation Practices
Do not bend cable sharply
– This may close off vent tube*.
*Gauge (vented) pressure sensors contain a vent tube inside the cable. If this vent tube is closed off, readings will be wrong.
Installation Practices
Install in vertical position
– Off vertical may introduce an offset.
Installation Practices
Do not lower into mud or silt
– This may clog or blind sensor.
Use a strain relief kit to prevent excess pulling on the cable.
Installation Practices
Install so that service connector stays dry.
Installation Practices
On vented pressure sensors…
Install so as to keep desiccant dry.
Installation Practices
Replace desiccant when it turns pink.
Installation Practices
Pink
=
Bad
Blue
=
Good
Security Issues
Security Issues
Use locking monuments
Security Issues
Paint to match surroundings
plain painted plain painted
Security Issues
Surround with a fence
Security Issues
Install caution signs or alarms
Security Issues
Use passworded software
Security Issues
Proprietary data formats and software
Changing Batteries
Changing Batteries
Should be done in a clean, dry environment to avoid contamination or moisture damage to the circuitry.
Changing Batteries
Top- cap
Cable
O-rings
AA batteries
1. Twisting gently, unscrew the top-cap.
Note: Two O-rings provide a water-tight seal and often seal tightly. Pulling hard can cause the O-rings to release suddenly and the top-cap to disengage with enough force to pull the insides out of the sensor.
Changing Batteries
2. Gently separate the top-cap from the body of the sensor. Top-cap remains attached to body via several colored wires.
3. Gently slide batteries out.
4. Insert new batteries – positive terminals towards top-cap.
5. Replace and retighten top-cap.
Aqua4Plus Software Operation
Aqua4Plus Operation
Connect the sensor to your computervia Serial Port
Aqua4Plus Operation
Connect the sensor to your computervia USB Port
Aqua4Plus Operation
Step 1: Select a COM port
Select the COM port to which your sensor is connected.
Aqua4Plus Operation
Step 2: Select a sensor and take real time readings
Open the sensor window by clicking on the tool button.
Aqua4Plus Operation
Step 2: Select a sensor and take real time readings
On the sensor map, click on the desired sensor.
General information about that sensor will display to the right.
Aqua4Plus Operation
Step 2: Select a sensor and take real time readings
Click on the Start button to view real time readings from the sensor.
Aqua4Plus Operation
Step 3: Create a recording session
Open the session window by clicking on the tool button.
Aqua4Plus Operation
Step 3: Create a recording session
Assign a name and then describe your recording session.
Click the Start button to begin the session.
Aqua4Plus Operation
Step 4: Retrieve data from sensor
Start the data upload by clicking on the tool button.
Aqua4Plus Operation
Step 5: View the data
View your data by clicking on either the table tool button or the graph tool button.
Aqua4Plus Operation
Step 5: View the data
Table Format Graph Format
Aqua4Plus Operation
Step 6: Export data to .csv
Export data to a .csv (comma separated value) file by clicking on the tool button.
Aqua4Plus Operation
Step 6: Export data to .csv
Export data can easily be imported into and viewed by many different analysis programs.
MS Excel
MS Access
Modeling programs
Data analysis programs
Calibration
Overview
Calibration Overview
Aqua4Plus’ built in calibration calculator allows you to easily perform two-point calibrations on each channel.
Calibration Overview
Using the interactive calculator, enter known points and take measurements.
Aqua4Plus determines the needed calibration values.
Calibration Overview
Step-by-step instructions display on the calibration window for ease of reference.
What we covered
• Installation Practices• Security• Changing Batteries• Aqua4Plus Operation• Calibration Overview
Summary