Page 1: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Introduction to Measuring JitterDonald B. Sanders

Duke University Medical Center, Durham, NC

Page 2: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Financial Disclosure• Dr. Sanders receives royalties for the book, Single Fiber EMG.

Page 3: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

WarningVideotaping or taking pictures of the slides associated with this presentation is prohibited. The information on the slides is copyrighted and cannot be used without permission and author attribution.

Page 4: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Neuromuscular jitter:what is it, and how isit measured?

Page 5: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Endplate potentials

Page 6: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Endplate potentials & action potentials

Elmqvist et al, 1964

normal MG

Page 7: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

SFEMG:signals from 1 muscle fiber

Stålberg

Page 8: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

The selectivity of SFEMG results from:

• The electrode construction

• High-pass filter (500Hz)

.

Page 9: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

CNEMG: signals from 2 - 15 muscle fibers

Stålberg

Central spike

Slow wave components

Page 10: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Page 11: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

SF vs CN electrodes

500 µ

500 µ

. SFE

Page 12: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

SF vs CN electrodes

500 µ

500 µ

.

CNE 36G

SFE

Page 13: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

SF vs CN electrodes

500 µ

500 µ

.

CNE 36G

SFE

CNE 25G

Page 14: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Jitter with non-SFE

• Payan, 1978 – “Blanket principle”• Weichers, 1985; Clark & Eisen, 1985 – jitter with MNE• Ertas, 2000 – compared SFE & CNE• Sarrigiannis, 2006 – sensitivity of CNE jitter for MG in Orb oculi = 95%

o 36G CNE; 2kHz high pass filter; normal limits determined by ROC in 20 controls; 56 MG patients

• Benatar, 2006 – “ “ “ “ in frontalis = 76%o 36G CNE; 2kHz high pass filter; SFE normal limits; 21 MG patients

Page 15: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

• Re-usable electrodes are not allowed or available in many places.

• CN electrodes don’t need to be serviced.• They’re cheaper. (Though SF electrodes can cost less per use in

the long-term.)• They’re sharper & somewhat less painful.

Why use CN electrodes to measure jitter?

Page 16: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Limitations of using CN electrodes to measure jitter

• Precise measurement of jitter requires great attention to CN signal configuration, thus studies take longer.

• CNEs are less directional, thus harder to position.• CNEs don’t maintain position in the muscle as well.• Stimulation studies are difficult except in facial muscles.• Cannot measure FD with CNEs.

Page 17: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Use smallest CNE available – DCF 25 (orange base)

25mm x 30G, Area = 0.019 mm2

Measuring Jitter with CN Electrodes

Page 18: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

RecordingSurface = 0.019mm2

RecordingSurface = 0.005mm2SFEMG

electrode

DCF 25electrode

Page 19: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Equipment settings for measuring jitter

• LF Filter: to remove distant signalso SFEMG electrode – 500 Hzo CN electrode – 1000 Hz

• Gain - typically 0.2-0.5 mV/div(to see signals clearly)

• Sweep - typically 0.5 ms/div (to see jitter & late components)

• Signal triggered delay – to show components before and after the trigger point

Page 20: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Measuring jitter during axonal stimulation

Page 21: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Measuring jitter during axonal stimulation – frontalis muscle

Recordingelectrode

(SFE)

Stimulatingelectrode

surface stimulation needle stimulation

Page 22: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Electrode position for measuring jitter with voluntary activation

Measuring jitter duringvoluntary activation

Page 23: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Use 500 Hz LF filter for SFE, 1000 Hz for CNEMay use voluntary or stimulation activationRecord widely across the width of the muscle

Measuring Jitter with SF or CN Electrodes

With Voluntary Activation

With Stimulation Activation

Maintain constant activation rateRecord 20 pairs: >15 pairs must be of acceptable quality

Stimulation frequency: 5-10 Hz (10 is preferred)Record 30 APs: > 20 APs must be of acceptable quality

Page 24: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Jitter Measurement Techniques

Voltage Level vs Peak Detection

Page 25: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Acquisitiontrig level

startstop

Voltageleveltechnique

Page 26: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Acquisitiontrig level

startstop

Acquisitiontrig level

calc peakstart

calc peakstop

Stålberg

Voltageleveltechnique

Peakdetectiontechnique

Page 27: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

How to calculate the jitter?• The IPI can be affected by factors

other than neuromuscular variability, e.g., slow trends due to electrode movement, VRF effects of variable firing rate.o Standard deviation of IPI would be

increased by trends.o Mean difference of consecutive IPI’s

(MCD) is relatively unaffected by trends.o Mean sorted difference (MSD) corrects for

affect of firing rate

Page 28: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Inter-Potential Interval (IPI) (jitter)

Page 29: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

IDI1 IDI2 IDI3

Inter-Potential Interval (IPI) (jitter)

Inter-Discharge Interval (IDI) (rate)

Page 30: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Calculating jitter: MCD (Mean difference between consecutive IPIs)

Page 31: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Calculating the jitter

• MCD – corrects for trends• Mean Difference between consecutive

Sorted intervals (MSD )– corrects for effect of variable firing rate

• Use the lower of these as the jitter

Page 32: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Add jitter comparison video

Page 33: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

How does measuring jitter improve your routine EMGs? • Your ear becomes attuned to variability

in the pitch of MUPs during EMG.

• You can then hear unstable MUPso Trigger on them, turn up the LF filtero Increase display speedo Superimpose MUPs & observe waveform variability

Page 34: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Page 35: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Registrants for this course receive a 50% discount. Use

the discount codeSanders50

when ordering.

Page 36: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

and stimulated – Sanders, Kouyoumdjian, Machado

Page 37: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Page 38: Introduction to Measuring Jitter - AANEM...Introduction to Measuring Jitter Donald B. Sanders Duke University Medical Center, Durham, NC. 2019 Financial Disclosure • Dr. Sanders

2019

Share Your Feedback• Please use the 2019 AANEM Annual Meeting app to rate this presentation and the

speaker.

• Your feedback helps us enhance our annual meeting to ensure we are continuing to meet your needs.