infrared thermographic system

Upload: achu-balachandran

Post on 14-Apr-2018

235 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/30/2019 Infrared thermographic system

    1/40

    INFRARED THERMOGRAPHIC SYSTEMS AND THEIR

    MECHANICAL APPLICATIONS.

  • 7/30/2019 Infrared thermographic system

    2/40

    ABSTRACT

    Infrared thermography is the science of detecting and measuring variations in heat emitted by an

    object and transforming them into visible images. Infrared Thermal Imaging is an electronic

    technique which quite literally allows us to see thermal radiations. With this new capability,

    plant maintenance personnel have recognized Infrared Thermal Imaging as one of the most

    versatile and effective condition monitoring tools available today. Infrared thermography can

    detect numerous conditions in which an anomaly is characterized by an increase or decrease in

    surface temperature or retained residual heat. These include high resistance heating of

    connection points in electrical systems; heat generated by abnormal friction in bearing surfaces;

  • 7/30/2019 Infrared thermographic system

    3/40

    differences in heat transfer across surfaces, such as a wall or boiler; and moisture retained in

    insulation in a roofing system. Infrared imaging, when combined with software analysis tools,

    may also be used to quantify these types of thermal differences.

    Infrared Thermal Imaging is an excellent condition monitoring tool to assist in the reduction of

    maintenance costs on mechanical equipment. The technique allows for the monitoring of

    temperatures and thermal patterns while the equipment is online and running under full load.

    Most mechanical equipment has allowable operating temperature limits that can be used as

    guidelines. Unlike many other test methods, Infrared Thermal Imaging can be used on a wide

    variety of equipment including pumps, motors, bearings, pulleys, fans, drives, conveyors etc.

    Advantage of these methods is contactless measurement, usually very fast and without any

    destructive influence on tested/measured objects

    CONTENTS

    List if figures..5 Notations....6

  • 7/30/2019 Infrared thermographic system

    4/40

    Introduction..7 History..8 Theory..9 Common infrared thermometer construction.13 Surface temperature in infrared thermometers...14 Elements in infrared thermometer..15 Temperature measurement calculations.19 Types of infrared thermometers.22 Advantages and disadvantages of IR thermometry....25 Non-destructive evaluation.26 Application.29 Conclusion..40 Future of IR imaging...41 References...42

    LIST OF FIGURES

    Fig 3.1: Splitting of incident energy.10

    Fig3.2: Spectral Distribution....11

  • 7/30/2019 Infrared thermographic system

    5/40

    Fig 4.1: Optical arrangement in IR detector......13

    Fig 4.2: Block diagram of IR thermographer.13

    Fig 5.1: Hemispherical IR thermometer15

    Fig 6.1: Field of view..17

    Fig 10.1: Fast transient NDE27

    Fig 10.2: Steady state NDE.....28

    Fig 11.1: Thermographs for comparing working of motors30

    Fig 11.2: Thermograph of belt and pulley.30

    Fig 11.3: Thermograph of a faulted bearing......31

    Fig 11.4: Thermograph of a steam trap.32

    Fig 11.5: Thermograph of a defective steam trap...33

    Fig 11.6: Thermograph of a rotary kiln.....34

    Fig 11.7: Thermograph of a blower system...35

    NOTATIONS

    WI = incident energy received by the object, W

    WR = energy reflected off the objects surface, W

    WT = energy transmitted by the object, W

    WA = energy absorbed by the object, W

  • 7/30/2019 Infrared thermographic system

    6/40

    E = emissivity

    = Stefan-Boltzmann Constant = 5.6703 10-8

    , W/m2K

    4

    T = absolute temperature,oK

    A = emitting area, m2

    TIR = Surface temperature,K

    esurf = Emissivity of the surface

    Tback = Background Temperature,K

    TIR, Ref = Emitter temperature.K

    TIR, smpl = sample surface temperature,K

    TFOV = The uncertainty in temperature due to variations in field of view,K

    TLin = The uncertainty in temperature due to non-linear performance of the

    imager.K

    T = The total uncertainty in the measured surface temperature,K

    1.0 INTRODUCTION

    Temperature is one of the important parameters to tell the condition of internal process, material

    and even quality of the desired output. A qualitative but accurate conclusion can be drawn by

    observing the temperature profile of any surface. On the other hand higher temperature also

    indicates obvious loss of energy in the form of heat. Therefore temperature monitoring would

    give ample indication of the condition of the material, process quality and explore the possibility

    of energy conservation avenues. In recent years, many temperature monitoring techniques have

    been in use in our industries. These techniques find their application based on the measurable

    temperature range, sensitivity and ease in application. With the on-line condition monitoring

  • 7/30/2019 Infrared thermographic system

    7/40

    technology becoming an inevitable part of maintenance strategy in todays scenario, non -contact

    type temperature monitoring methods have become more popular. Infrared Thermography is

    such a non-contact type technique which provides a fast, reliable and accurate temperature

    profile of any material surface. This paper deals with various application of infrared

    thermography including the justification of energy saving potential in many cases.

    Infrared (IR) energy is part of the electromagnetic spectrum and behaves similarly to visible

    light. It travels through space at the speed of light and can be reflected, refracted, absorbed, and

    emitted. The wavelength of IR energy is about an order of magnitude longer than visible light,

    between 0.7 and 1000 m. All objects emit infrared radiation as a function of their temperature.

    Infrared energy is generated by the vibration and rotation of atoms and molecules. The higher the

    temperature of an object, the more the motion and hence the more infrared energy is emitted.

    This is the energy detected by infrared cameras. The cameras do not see temperatures, they

    detect thermal radiation. At absolute zero (273.16C, 459.67F), material is at its lowest

    energy state so infrared radiation is minimized.

    Infrared thermography is the technique for producing an image of invisible (to our eyes) infrared

    light emitted by objects due to their thermal condition. An image produced by an infrared camera

    is called a thermogram or a thermograph. Infrared (IR) thermographic systems, or IR imagers,

    provide images that represent surface temperatures, by measuring the magnitude of infrared

    radiation emanating from the surface of an object. Because IR imagers see the radiation naturally

    emitted by objects, imaging may be performed in the absence of any additional light source.

    Modern IR imagers resolve surface temperature differences of 0.1C or less. With this high

    sensitivity, they can evaluate subtle thermal phenomena, which are only revealed in the form of

    slight temperature gradients. Some applications that employ IR thermography include:

    inspections for predictive maintenance, non-destructive evaluation of thermal and mechanical

    properties, building science, military reconnaissance and weapons guidance, and medical

    imaging.

  • 7/30/2019 Infrared thermographic system

    8/40

    2.0 HISTORY

    The first patent for a total radiation thermometer was granted in 1901. The instrument used a

    thermoelectric sensor; it had an electrical output signal and was capable of unattended operation.In 1931, the first commercially-available total radiation thermometers were introduced. These

    devices were widely used throughout industry to record and control industrial processes. They

    are still used today, but mainly used for low temperature applications. The past 20 years, non-

    contact infrared thermometer in the rapid development of technology, performance, continuous

    improvement, function continuously enhanced, increasing variety, scope also expanded its

    market share year by year.

    3.0 THEORETICAL BASIS FOR RADIATION

    MEASUREMENTS

    Heat is transferred from one body to another through conduction, convection, or radiation.

    Radiation is a process where heat energy in a form of electromagnetic waves is emitted by a hot

    object and absorbed by a colder object. Most of this radiation is in the infrared (IR) region of the

    electromagnetic spectrum, but some also spreads into the visible light band. The IR wavelength

    band stretches from 0.7 to 1000 microns, however practical IR measurement systems use only

  • 7/30/2019 Infrared thermographic system

    9/40

    certain wavelength bands between 0.7 and 14 microns because the radiation is the strongest in

    this range.

    If an object is exposed to IR energy radiated by a heat source, such as an electric heater, light

    bulb, sun, or other source, the energy reaching the object is called incident energy. Part of this

    energy is reflected off the object surface. Theoretically, the objects coefficient of reflectivity can

    vary from 0 (no reflection) to 1.0 (100% reflection). Rough, matt surfaces have low reflectivity.

    Polished and glossy surfaces, especially metals, have high reflectivity.

    Depending on the object material, thickness, and the radiation wavelength, part of the radiation

    can go through the object or be transmitted. The coefficient of transmission can vary from 0 (no

    energy transmitted through object) to 1.0 (100% energy transmitted through object). High

    transmittance examples include glass, quartz, plastic film, and various gasses. Materials opaque

    in the IR spectrum have close to zero transmission coefficients.

    The remaining energy is absorbed by the object and raises its temperature. A hypothetical body

    that has no reflection or transmission and absorbs all incident energy across the entire spectrum

    has a coefficient of absorption equal to 1.0 and is called a blackbody. Real-life objects, referred

    to as gray bodies, have coefficients of absorption that fall between 0 and 1.0.

    Incident energy, WI, is defined as:

    WI = WR+ WT + WA

    Fig 3.1Splitting of incident energy

  • 7/30/2019 Infrared thermographic system

    10/40

    When incident heat energy reaches an object, part of this energy is reflected, part passes through

    the object, and the rest is absorbed. The coefficients of reflection, transmission, and absorption

    depend on the object material and surface finish and on the wavelength spectrum of the incident

    energy.

    As the object absorbs energy and heats, it also emits energy. When an object is in a state of

    thermal equilibrium, the amount of energy it absorbs (WA) equals the amount of energy it emits

    (WE): WA = WE. When an object absorbs more energy and its temperature increases, the amount

    of radiation it emits also increases.

    IR thermometry is based on the fact that any body (solid, liquid, or gaseous) that has a

    temperature above absolute zero (0oK or -273

    oC) emits radiant energy. This energy is

    proportional to the forth power of the body temperature, and the bodys ability to absorb and

    emit IR energy is called emissivity. Energy radiated by a body can be expressed as follows:

    W = E A watts.

    Fig 3.2 Graph showing relation between wavelength and temperature

  • 7/30/2019 Infrared thermographic system

    11/40

    When the temperature of a hypothetical blackbody increases, the radiated IR energy also

    increases. Temperature T2 is several times greater than temperature T1. The rise between 1 and

    10 microns is most pronounced.

    Emissivity can range from 0 to 1 for various bodies. A hypothetical blackbody emits and absorbs

    all energy and thus has an emissivity equal to 1. Real-life objects have an emissivity between 0

    and 1.

    When an IR thermometer measures an objects temperature, consider the energy that actually

    enters the lens. That is, in addition to emitting energy related to its own temperature, the object

    may reflect energy coming from another source, or transmit energy passing through it from a

    source behind it. Thermal radiation is largely a surface phenomenon, because most materials arenot transparent to IR. As a result, it is the material properties of the surface of an object that

    determine emittance. Polished metals have low emittance, but a thin layer of paint can transform

    them to high emittance. For materials that are opaque to IR radiation, emittance can be

    considered the complement of reflectance (the amount of incident radiation that is reflected by a

    surface), expressed by 1-e. It is important to realize that because no real surface has an emittance

    e=1, the radiation that is viewed as coming from an object is always a combination of emitted

    and reflected radiation, termed radiosity, and it contains information regarding both the

    temperature of the object and its surroundings. Special techniques described within are necessary

    to distinguish the two components and obtain accurate surface temperatures for the object of

    interest. When imaging outdoors, it should be taken into account that the sun is a significant

    source of IR radiation, particularly in the shorter wavelengths of the IR band.

    For accurate measurements, survey the surrounding area for possible sources of extraneous IR

    radiation and choose the thermometer position and aiming angle to minimize the effects of those

    sources

  • 7/30/2019 Infrared thermographic system

    12/40

    4.0 COMMON INFRARED THERMOMETER CONSTRUCTION

    The most basic design consists of a lens to focus the infrared thermal radiation on to a detector,

    which converts the radiant power to an electrical signal that can be displayed in units of

    temperature after being compensated for ambient temperature.

  • 7/30/2019 Infrared thermographic system

    13/40

    Fig 4.1 Optical Arrangement

    Fig 4.2 IR imager flow diagram

    This configuration facilitates temperature measurement from a distance without contact with the

    object to be measured. As such, the infrared thermometer is useful for measuring temperature

    under circumstances where thermocouples or other probe type sensors cannot be used or do not

    produce accurate data for a variety of reasons.

  • 7/30/2019 Infrared thermographic system

    14/40

    The majority of IR detectors are either single-wavelength (also called single-color), or dual-

    wavelength (also called two-color) type. The single-wavelength detectors measure IR energy

    within a certain wavelength band, and the instrument calculates object temperature based on the

    detector output and the preset emissivity. Some thermometers have adjustable emissivity, and

    most simple units have fixed emissivity.

    Dual-wavelength detectors measure energy at two different wavelength bands, and the

    instrument calculates temperature based on the ratio of the two readings. If emissivity or the

    energy changes by the same amount at both bands, the measurement accuracy is not affected.

    Emissivity or the amount of radiated energy may change due to object change or movement, lens

    contamination or misalignment, or view obstruction. The dual-wavelength detectors drawback

    is higher cost and lower accuracy under certain conditions

    5.0 SURFACE TEMPERATURE INFRARED THERMOMETERS

    We have seen that a radiation source with characteristics close to that of a blackbody could beconstructed by assuring multiple reflections within a cavity. Fig. 8 illustrates a device for

    measuring the temperature of the surface of an object based on this principle

    Fig 5.1 Surface temperature infrared thermometer

  • 7/30/2019 Infrared thermographic system

    15/40

    Radiation from the target is multiply reflected from the hemispherical mirror. A detector receives

    radiation through a small opening in the reflector. The radiation multiply reflected between the

    mirror and the target appears to the detector to be from a blackbody. A commercial thermometer

    using this principle can read the temperature of targets with emittance as low as 0.6 without

    correction. The reflector must be placed close to the surface being measured to exclude

    extraneous radiation and prevent radiation losses.

    6.0ELEMENTS OF INFRARED THERMOMETERS

    6.1 DETECTORS

    Based on the principle of operation, IR detectors fall into one of two categories: thermal

    detectors and photo detectors (photodiodes). Thermal IR detectors absorb the incident energy,

    raise the sensing element temperature, and change the detectors electrical properties:

    thermopiles generate thermoelectric voltage, bolometers change resistance, and pyroelectric

    devices change their polarization. In general, they are slower than photo detectors.

    A thermopile is made by connecting several thermocouples in series and placing their hot

    junctions in contact with a black body that absorbs the incident IR energy and heats the hot

    junctions. The cold junctions are placed in the area of the detector with adequate heat sinking.

    These detectors have fast response, broad band, large dynamic range, and are frequently used in

    general-purpose, automotive, air conditioning, and human-body thermometers.

    Bolometers use a slab of material that changes its resistance in response to a change of

    temperature. The circuit converts resistance change to a voltage change, which is further

    processed by the instrument. Bolometers are frequently used for measuring low-level IR energy,

    often as an attachment to a telescope.

    Pyroelectric devices become electrically charged when their body temperature changes. To

    produce a usable signal, the incident IR energy has to pulse. The output peak-to-peak AC

    signal is proportional to the pulse energy. Since energy emitted by measured objects is usually

    steady, thermometers that use pyroelectric detectors have a mechanical or optical chopper in

    front of the sensor. These sensors are used in many home security systems.

  • 7/30/2019 Infrared thermographic system

    16/40

    Photo detectors are built on a silicon substrate with an IR sensitive area that releases free

    electrons when impacted by the photons. The flow of electrons produces electrical signals

    proportional to the incident energy. These detectors are often used as arrays in thermal imaging

    systems.

    A detector needs protection from the environment, and the selected window material must allow

    the correct wavelength band to pass through with minimum attenuation. A zinc sulfide or

    germanium window is best for the long-wavelength detectors, glass is suitable for short-

    wavelength detectors, and quartz for the mid-wavelength spectrum. Some instruments use a

    fiber-optic light guide to direct the radiation to the detector.

    6.2 OPTICAL SYSTEMS

    The optical system of an IRT may be composed of lenses or mirrors or combinations of both.

    Mirror systems have been common mainly in fixed focus optical systems. To vary the focus of

    an optical system requires that at least one element be movable. Providing for this motion in

    mirror systems is often more complicated than in lens systems. The selection of lens and window

    materials is always a compromise between the optical and physical properties of the material and

    the desired wavelength response of the pyrometer.

    6.3 DATA ACQUISITION AND IMAGE PROCESSING

    Collecting image data is an important part of operating a real-time imager. The data are

    commonly recorded for later analysis, assembly into reports, etc. Continuous video signals may

    be recorded on tape or with the aid of a computerized video digitizer. In steady-state thermal

    situations a single IR image would be generated by averaging a number of frames over a period

    of time. Transient thermal situations are usually recorded continually, and a series of images are

    stored over time.

  • 7/30/2019 Infrared thermographic system

    17/40

    6.4 FIELD OF VIEW

    Fig 6.1 Field of view

    The key specifications and considerations for any IR thermometer application are field of view

    (FOV) and distance; spectral band; response time; accuracy and repeatability; emissivity of the

    object or media being measured; media between the object and IR thermometer, such as vacuum,

    air, steam, gas, glass, or other; object temperature range; mounted or hand-held application; and

    type of output signal or display.

    The FOV characterizes the diameter of a circle (target) that the IR detector will see at a certain

    distance from the measured surface. However, there is always a minimum target diameter that

    depends on the optical system and detector size. The detector measures and averages

    temperatures of all objects within the target area. FOV is commonly called distance-to-spot size

    ratio and is a ratio of the distance between the meter and the target to the diameter of the

    target.For example, a 10:1 distance-to-spot size ratio means that if a measured surface is located

    10 inches from the thermometer it will measure and average the temperature of a circle with 1

    in.diameter. Move the thermometer away to 20 in. and the target will increase to 2 in., and so on.

    A thermometer with a 1:1 ratio will measure within a one-foot diameter circle when held onefoot away from the target.Thermometers designed for measurements of small areas have a very

    narrow FOV and measure temperatures of objects less than a tenth of an inch. For example, such

    a thermometer held near a component on a pc board will measure the temperature of just that

    component and ignore the components around it.

  • 7/30/2019 Infrared thermographic system

    18/40

    7.0 Infrared Temperature Calculations

    In order to obtain quantitative surface temperatures, raw thermographic data may need to be

    extracted from the imaging system and processed in a separate mathematical program. The

    image-processing component of the imaging system is used to extract data from the thermal

    image and export them in appropriate arrays of text values to be synchronized with spatial

    coordinates. The result for temperature at each data point should be calculated with the correct

    emittance and a location-specific background level. The calculated IR temperature for the

    reference emitter may be compared to direct contact measurements, and deviations may be used

    to scale the rest of the IR data. Separate data sets may be merged together to combine data from

    different views.

    Equation given below, shows an expression for calculating a surface temperature, TIR, from the

    total thermal radiation represented by the variable Te=1, the emittance of the surface, esurf, and the

    background radiation level represented by the variable Tback. Te=1 is an equivalent blackbody

    temperature for the surface being measured. Tbackis the equivalent blackbody temperature for the

    background thermal radiation level as measured at each location with the aid of an applied

    mirror. Values for both Te=1 and Tback are obtained from the thermographic system by setting

    emissivity to unity. Conversely, setting emittance to less than one should cause the IR image-

    processing component of the imaging system to process the image data using the calculation

    using the equation,

    This equation is used to determine both the apparent reference emitter temperature, TIR, Ref, (as

    measured by IR imaging) and the apparent sample surface temperature, T IR, smpl. The difference

  • 7/30/2019 Infrared thermographic system

    19/40

    between TIR,Refand the direct contact measured value, TDC,Ref, may be applied to correct TIR,smpl

    and to arrive at the final IR surface temperature result, T. This correction should be made for

    each temperature datum and for each IR image, to produce arrays of temperature values.

    Final data sets merge spatial location coordinates with temperature values. The real distances

    between location markers on the specimen are measured, and a coordinate system is used to

    create temperature/location data pairs. The temperatures may be distributed linearly.

    Temperature data are then made a function of spatial coordinates, mapped to the appropriate

    coordinate system.

    7.1 Infrared Temperature Uncertainty

    The thermographer usually wants to estimate the uncertainty in temperature data. Errors in esurf

    should be treated as a source of undefinable systematic uncertainty, esurf, and may be analyzed

    by propagating uncertainty. Errors in Te=1 and Tbacklead to random uncertainties, Te=1 and Tback.

    Te=1 and Tback are usually closely related to equipment specification for Noise Equivalent

    Temperature Difference (NETD). To analyze error propagation, the sum of the squares of partial

    differentials of Equation for TIR are calculated with respect to variables Te=1, Tback, and esurf,

    which leads to one possible solution,as shown. The uncertainty in T IR, TIR, is then calculated

    using Equation:

    Values for both TIR,smpl and TIR,Ref are obtained using this equation. The uncertainty in the

    direct contact measurement of the reference emitter surface temperature, T DC,Ref, is determined

    from the overall system accuracy of the direct contact sensor combined with any errors

    associated with adjustments that correct for gradients in the surface material. The uncertainty

    arising from variations across the field of view, TFOV, is determined from the magnitude of

    deviations in heavily averaged data for an isothermal plate that fills the section of the field of

  • 7/30/2019 Infrared thermographic system

    20/40

    view being used. The uncertainty arising from non-linear performance or calibration of the

    imager, TLin, is determined by evaluating TIR for a temperature controlled and separately

    measured target for a range of temperatures and evaluating how deviations change over that

    range. The total uncertainty in the measured surface temperature, T, is obtained from Equation :

    8.0 TYPES OF INFRARED THERMOMETERS

    The following is a convenient classification of IRTs:

    Broadband thermometers

    Band-pass thermometers

    Narrow band thermometers

    Ratio thermometers

    Fiber optic thermometers

    8.1 BROADBAND THERMOMETERS:

    Broadband thermometers have usually been the simplest IRTs, with spectral responses from 0.3

    microns wavelength to an upper limit of 2.5 to 20 microns (m), determined by the lens or

    window material. They have been termed total radiation thermometers because, in the

    temperature ranges of normal use they measure a significant fraction of all the thermal radiation

    emitted by the object whose temperature is being measured.

    8.2 BAND-PASS THERMOMETERS:

    Band-pass thermometers were initially derived from simple, broadband thermometers. Lens,

    window, or filter characteristics were selected to view only a selected portion of the spectrum.

  • 7/30/2019 Infrared thermographic system

    21/40

    The 5 0.2m band-pass was used to measure glass surface temperature because glass emits

    strongly in this region, but poorly below or immediately above this band.

    8.3 NARROW BAND THERMOMETERS:

    Narrow band thermometers operate over a narrow range of wavelengths. The spectral response

    of many narrow band thermometers is determined by the optical filter used. Such narrow band

    thermometers are used as general purpose instruments over the temperature range of interest. For

    example, a thermometer using a silicon cell detector will have a response that peaks at 0.9m.

    The upper limit of usefulness is about 1.1m. Such a thermometer can only be used at

    temperatures above 600C (1100F).

    8.4 RATIO THERMOMETERS:

    A ratio thermometer measures radiated energy in two narrow bands and calculates the ratio of

    the two energies. This ratio is a temperature-dependent function. The temperature measurement

    is not primarily dependent on the energy in the two bands, only on the ratio of the two energies.

    Therefore, any influence that affects the amount of energy in each band by the same percentage

    has no effect on the temperature indication. Changes in target size have no effect. If the

    emissivity of the target is the same at both wavelengths, the indicated temperature is not changed

    by changes in emissivity. Unfortunately, this condition is not fulfilled by all oxidizable materials.

    For other materials, the ratio techniques may reduce or eliminate changes in indicated

    temperature caused by changes in surface finish. The ratio technique may reduce the effect of

    energy absorbing materials such as particulates or C02 between the target and the thermometer,

    if the percentage of absorption is the same at each wavelength.

    8.5 OPTICAL PYROMETERS

    Although optical pyrometers are no longer manufactured in significant numbers, there areprobably hundreds of thousands in everyday use, and they are sufficiently unique in design and

    use to warrant a separate discussion. An optical pyrometer measures the radiation from the target

    in a narrow band of visible wavelengths, centered at about 0.65 m in the red/yellow portion of

    the spectrum. The most commonly used optical pyrometers are manually operated. The operator

    sights the pyrometer on the target. At the same time he can see in the eyepiece the image of an

  • 7/30/2019 Infrared thermographic system

    22/40

    internal tungsten lamp filament. The operator matches the filament color to the target by varying

    the current through the filament with a rheostat, and when the target image and the tungsten

    filament image are the same color, the target temperature can be read from a scale on the rheostat

    knob. When the target and filament colors are the same, the filament image apparently vanishes,

    so these pyrometers have also been called disappearing filament pyrometers.

    8.6 FIBER OPTIC THERMOMETERSIn fiber optic thermometers, the infrared radiation from the target is guided to the detector by a

    light guide. The first such sensors used a 1/8 in. diameter sapphire rod to pick up energy from the

    target and transmit it to a detector. Contemporary fiber optics pyrometers use a flexible bundle of

    glass fibers with or without a lens. The spectral response of these fibers extends to about 2 m,

    though some exotic materials such as fluorides have a wider band pass. Some are useful at target

    temperatures as low as 100C (212F). Fiber optic thermometers are especially useful where it is

    difficult, dangerous or impossible to obtain or maintain a clear sighting path to the target, as in

    pressure or vacuum chambers. Fiber optic thermometers have also been used to measure

    temperatures of turbine blades in gas turbines, and the temperature of small objects in induction

    heating coils.

    9.0 ADVANTAGES AND DISADVANTAGES OF IR THERMOGRAPHY

    The following points spell the advantages and disadvantages of this technique.

  • 7/30/2019 Infrared thermographic system

    23/40

    9.1 Advantages:

    It is a non-contact type technique. Fast, reliable & accurate output. Minimize preventive maintenance and troubleshooting time. Detect problems quickly, without interrupting service. A large surface area can be scanned in no time Presented in visual & digital form. Software back-up for image processing and analysis. Requires very little skill for monitoring.

    9.2 Disadvantages:

    Cost of instrument is relatively high. effects of thermal losses (convective, radiative, conductive) perturbing thermal contrasts Unable to detect the inside temperature if the medium is separated by

    glass/polythenematerial etc.

    10.0 Non-Destructive Evaluation (NDE)

    10.1 Fast Transient NDE

  • 7/30/2019 Infrared thermographic system

    24/40

    Fast transient thermal imaging is based on the time-evolution of surface temperatures on test

    specimen during heating or cooling. Variations in temperature depend on the heat flow through

    the object; heat flow is sensitive to cracks, delaminations, voids, and changes in material

    properties. In a typical setup, flash equipment is used to generate a heat pulse, and the surface

    temperature of the test specimen is monitored with an IR imager. The pulse duration and the total

    energy delivered over the length of the pulse depend on the material properties and the nature of

    the defect to be detected. A series of approximately 10 to 30 frames is recorded with frame rates

    up to 80 Hz with scanners or 1 kHz with FPA imagers. In the absence of thermal pulse

    equipment and a high speed imager, transient NDE can be successfully performed on some

    specimens, particularly those with low thermal conductivity. Figure 6 shows several frames

    taken at 10 second intervals after a steady IR heat source was applied to the back side of a

    composite truck bed panel. The well bonded portions warm faster on the opposite, imaged side.

    An image sequence like this can already be very informative when visually inspected by a

    technician experienced in interpreting thermograms. However, image post-processing can greatly

    enhance images and reveal features that are hidden in the unprocessed image. Post-processing

    includes averaging of image sequences, coloration with different palettes, image subtraction and

    edge enhancement. Complex processing is possible, including Fourier transformation of time

    sequence images and tomography, if sufficient computer power is available to perform these

    tasks. Tomography analysis translates the time image sequence into a three-dimensional object

    sequence; defects can then be located in depth as well as in the image plane even though data are

    collected only at the surface.

  • 7/30/2019 Infrared thermographic system

    25/40

    Fig 10.1 Fast Transient NDE

    The, white circle, horizontal stripes, and dots are the adhesive glue pattern between two layers of

    a composite truck bed panel assembly. The pattern is relieved by the transient heat flow induced

    by exposing the panel to radiant heat source on one side and imaging it from the opposite side

    (10 seconds between each frame). The vertical lines are due to corrugated surface of the panel.

    10.2 Steady-state NDE

    Fast transient thermography can provide a more robust non-destructive evaluation technique than

    steady-state thermography, but its use has been limited because it requires expensive thermal-

    pulse equipment and storage of a large number of data for each specimen. Steady-state IR

    thermography, by contrast, is much less expensive, requiring minimal equipment and data

    storage. Near steady-state thermography could be used for NDE during production phases where

    temperature cycles have been applied to parts, for example the cool-down period after paint

    baking, adhesive curing, or injection molding of plastic parts. The required temperature

    difference between the part to be evaluated and the environment depends on the material,

  • 7/30/2019 Infrared thermographic system

    26/40

    geometry, and size of the part; the size and nature of the defects being sought; and the minimum

    resolvable temperature difference of the IR imager being used. This difference can be as small as

    a few degrees Celsius. For example, the figure shows an IR thermogram of a composite truck

    bed panel after reaching steady state with a radiant heat source. Gradients are not as sharp in fast

    transient thermography when imaging materials with high thermal conductivity. Post processing

    can improve image quality as shown in the image to the right.

    Fig 10.2 Steady state NDE

    11.0 APPLICATIONS

    Infrared Thermal Imaging is an excellent condition monitoring tool to assist in the reduction of

    maintenance costs on mechanical equipments. The technique allows for the monitoring of

  • 7/30/2019 Infrared thermographic system

    27/40

    temperatures and thermal patterns while the equipment is online and running under full load.

    Most mechanical equipment has allowable operating temperature limits that can be used as

    guidelines. Unlike many other test methods, Infrared Thermal Imaging can be used on a wide

    variety of equipment including pumps, motors, bearings, pulleys, fans, drives, conveyors etc.

    This section will explain the benefits of Infrared Thermal Imaging as a condition monitoring tool

    for plant mechanical systems.

    11.1 IMPROVED TROUBLESHOOTING

    Infrared Thermal Imaging definitely should be one of the tools that are selected for motor and

    rotating equipment inspection. Thermographic examination can help technicians use the other

    tools, such as vibration analysis, more effectively. If a thermal anomaly is found, then the other

    tools can be employed to help isolate the cause of the problem.

    11.1.1 MOTORS AND GENERATORS

    When considering electric motors and generators, operating temperatures and thermal patterns

    can be a valuable key in a predictive maintenance program. All motors have a normal thermal

    pattern as well as given maximum operating temperature. This temperature is usually stated on

    the nameplate of the motor and is normally given as a rise in degrees C above the ambient air

    temperature. Most motors are designed to operate in ambient temperatures that do not exceed 40

    C. Conditions such as inadequate air flow, partial discharge, unbalanced voltage, bearing failure,insulation failure and degradation in the rotor or stator can be identified with an Infrared Thermal

    Imaging monitoring program. Abnormal thermal patterns can also identify misalignment in

    couplings when these devices are used in conjunction with motors. Figure shows the thermal

    patterns of two motors, where there is a marked overall temperature rise on the motor on the left,

    indicating a possibility of abnormal operation.

  • 7/30/2019 Infrared thermographic system

    28/40

    Fig 11.1 IR imager of motor

    11.1.2 BELTS AND PULLEYS

    Belts and pulleys are good candidates for Infrared Thermal Imaging inspection. The interaction

    of the pulley wheel and the belt generates friction as the belt contacts and then leaves the pulley

    surface. Additionally, the continuous tension and compression of the belt causes internal friction.

    Both of these processes result in heat being generated which can be seen with the infrared

    camera. Comparing the thermal patterns of several pulley belt systems can provide clues to

    improper operation. The temperature distribution across a pulley sheave should be uniform if

    everything is working as intended.

    Fig 11.2 IR Image of pulley

    11.2 PROVIDE EVIDENCE OF PROPER INSTALLATION

    11.2.1 BEARINGS

  • 7/30/2019 Infrared thermographic system

    29/40

    Bearing problems are generally found by a comparison of surface temperatures; comparing one

    bearing to another working under similar conditions. Overheating conditions are documented as

    hot spots within the infrared camera and are usually found in comparing equipment to

    equipment, end bell to end bell (for the same type of bearings) and stator to end bell temperatures

    (determined by motor design and configuration).

    Fig 11.3 IR Image of bearing

    Vertical motor pump combination displaying elevated temperature in lower thrust bearing.

    11.3 REDUCE OPERATING COSTS(Leaking or Blocked Pipes and Valves)

    11.3.1 STEAM TRAPS

    Steam is an efficient way to transport heat energy. The latent heat needed to transform water into

    the gaseous state, steam, is extremely large. This means that large amounts of heat energy can be

    transported with a minimum temperature differential to the environment. This equates into lower

    energy and insulation costs.

    When steam arrives at the site where heat energy is needed, it condenses, thereby releasing the

    large latent heat it was carrying. The condensate formed by this process must be removed by the

  • 7/30/2019 Infrared thermographic system

    30/40

    steam system and returned to the boiler where it is turned back into the steam and the cycle is

    repeated.

    The steam trap performs this function; to hold back live steam while allowing gases and

    condensate to pass through. This allows more energy to be obtained from the steam, thus raising

    a steam system's overall efficiency.

    Fig 11.4 IR image of steam trap

    In these two photos of steam traps, the Infrared Thermal Imaging photo on

    the left shows the point of leakage.

    Steam traps, like any mechanical device, eventually fail. Most are designed to fail in the open

    position to maintain steam system operation. When they fail in the open position, they "blow"

    live steam. This costs energy dollars which the steam trap was installed to conserve.

    Occasionally, steam traps fail in the closed position. This causes condensate back up in the steam

    system with potentially disastrous results.

    Fig11.5 IR image of faulty steam trap

  • 7/30/2019 Infrared thermographic system

    31/40

    Thermogram and visible image of a failed inverted bucket steam trap.

    Properly performed Infrared Thermal Imaging can identify steam traps that are blowing steam as

    well as those that may fail closed. Steam traps are sometimes located in inconvenient areas

    where they are hard to reach using other NDT technologies. This fact makes trap diagnosis by IR

    especially useful.

    11.3.2 FURNACE & REFRACTORY

    A refractory material is basically a high temperature insulator. They are usually a non-metallic

    ceramic type of material and are commonly supplied in brick form. Refractory materials are used

    inside furnaces, ovens, boilers, hot storage tanks and other equipment that produces or contains

    extremely hot temperatures.

    Using Infrared Thermal Imaging to inspect refractory begins with the assumption that a uniform

    temperature inside a vessel will result in uniform warming of its exterior as heat is conducted to

    the vessel's walls.

    An ideal vessel would have a perfectly uniform temperature on its external surface. If a crack or

    other defective conditions existed in the insulating medium, the exterior surface would increase

    in temperature in direct proportion to and in the exact location of the defect. Thus, the defect

    would be obvious based on the non-uniform conductance of heat into the exterior surface. If the

    exterior metal is threatened or a safety hazard may exist, absolute temperature measurements ofthe hot areas become important.

    In normal practice, there are other structural variations which cause non-uniform heat patterns in

    addition to the defects; these include grid work, refractory type variations, port holes, catwalks,

    etc.

    11.3.3 ROTARY KILNS

    In the manufacture of cement the raw materials are ground together, the mixture is heated until it

    fuses into a clinker, and the clinker is ground into a fine powder. The heating is usually

    accomplished in rotary kilns which look like huge rotating pipes, sometimes hundreds of feet

    long. The kilns are slightly tilted from the horizontal, and the raw material is introduced at the

    upper end, either in the form of a dry rock powder or as a wet paste composed of ground-up rock

    and water. As the charge progresses down through the kiln, it is dried and heated by the hot gases

  • 7/30/2019 Infrared thermographic system

    32/40

    from a flame at the lower end. After it leaves the kiln, the clinker is cooled quickly and ground,

    and then conveyed by a blower to packing machinery or storage silos.

    Fig 11.6 IR image of rotary klin

    Thermal patterns indicate refractory damage in this lime kiln.

    The refractory lining of the kilns eventually wears out and must be replaced. If it is not, the steel

    shell can overheat and fail. Infrared Thermal Imaging is ideally suited for locating and measuring

    the temperature of areas where refractory has failed. Continuous tracking allows the kiln

    operators to maximize the amount of time they can operate until a shutdown for refractory repair

    is necessary.

    11.4 IMPROVE THE EFFICIENCY OF A PROCESS

    11.4.1BLOWER SYSTEMS

    Blower systems can be used to move materials through pipes from one process or location to

    another. An example might be a system to deliver coal dust to a boiler. The condition of the

    pipes and process itself can often be monitored with an infrared camera. Buildup of product

    debris in a pipe will usually show up as a temperature differential in a section of pipe. After the

    buildup has been removed, Infrared Thermal Imaging can be used to verify that the work hasbeen completed satisfactorily.

  • 7/30/2019 Infrared thermographic system

    33/40

    Fig 11.7 IR image of blower pipe

    Product buildup can be clearly seen as the cold area

    on the bottom of the pipe.

    11.5 OTHER APPLICATIONS

    11.5.1 Automobile industries

    Infrared thermometers are used extensively in the automotive industry to diagnose problems

    ranging from air conditioning, brakes, to engines. They also help in avoiding dangerous car

    interiors by convenient temperature measurement with these non-contact devices. They are

    however most effective in diagnosing temperature-related engine problems. This is because, as

    with other traditional thermometers, the user doesn't have to dissemble the whole engine, as it

    reaches even inaccessible areas. Among the various uses of these infrared thermometers in

    engine and exhaust troubleshooting are for scanning manifold temperature to determine low

    compression, for misfire condition, measuring and comparing exhaust system components for

    correct fuel/air mixture and for scanning both sides of the catalytic converter for temperature

    differences

    These are also used for cooling system diagnostics like checking for overheating problems,

    scanning radiators for core restrictions, measuring thermostat temperature and above all for

    measuring coolant temperature sensors and manifold air temperature sensors

  • 7/30/2019 Infrared thermographic system

    34/40

    Even for measuring climate control, in automobiles infrared thermometers are used most of the

    times. With them you can check heaters, air conditioners, and vents at a convenient pace for any

    kind of malfunction or system output. These are equally powerful like digital multimeters, as

    gauging instruments.

    Further, infrared thermometers are used for checking the condition of braking system in

    automobiles. This is very necessary because if the temperature is significantly higher than

    ambient or varied temperatures of rotors, drums or bearings, it would indicate excessive wear.

    Uneven braking can also be diagnosed with this amazing instrument.

    11.5.2 Steel industry

    Steel industry use thermometer because products are moving, the temperature is very high.

    Ordinary steel industry application is temperature is a constant state of molten steel is beginning

    to change into pieces. With the same temperature to heating steel is preventing it from the

    deformation of the key, infrared thermometeris used to measure the internal temperature of the

    regenerator. In high temperature spin rolling grinding machine, the infrared thermometer was

    used to identify of product temperature is in rotating limits. In cooling grinding machine, infrared

    thermometer rolling in steel cooling process monitoring steel temperature.

    11.5.3 Glass industry

    In glass industry, products are in the dynamic state, which is to be heated to a high temperature.

    Infrared thermometer used to monitor the temperature of furnace. Portable sensors are used to

    detect by measuring the external temperature points. Measurement of the temperature during

    melting of glass is used to determine proper mouth furnace temperature. On the flat glass,

    sensors in each processing phases are used for testing temperature. Wrong temperature or

    excessive temperature change can cause uneven expansion or contraction. For bottles and

    container products, it would undergo melt flow to keep the glass at the same temperature before

    furnace. Infrared thermometer is used to detect the glass before furnace temperature. So it in

    export place should be the proper state. In fiberglass products, infrared sensor is used in

    http://www.temperature-controller.info/categories.php?id=6http://www.temperature-controller.info/categories.php?id=6
  • 7/30/2019 Infrared thermographic system

    35/40

    processing furnace detection of glass furnace temperature before. Infrared sensor in glass

    industry another use is used in windscreen products process.

    11.5.4 Plastic industry

    In the plastic industry, infrared thermometer is used to avoid defiled product, by measuring

    temperature profiles of dynamic objects. In blown film ejective process, temperature

    measurements can be successfully performed with ease using an infrared imager to adjust

    heating and cooling which can help to keep plastic tension completely distributed along its

    thickness. Also in the case of film made in cast out process, sensors help control temperature to a

    large extent. In chip pressure out, the sensor can let operator to adjust extinguished heater and

    cold roll to guarantee the quality of its products. Infrared thermometer in preventive

    maintenance, uses portable heat display system, maintain personnel can identify potential or

    already existing problems. For example: the engine overheat, transformer winding of plugged

    cooling fins, capacitance contact undesirable, heat in a compressor cylinder lid rally. Any

    problems are accompanied with temperature, or temperature curve with surrounding temperature

    completely not, so can use portable heat display system positioning. In most cases, when asked

    to stop process before, can be timely find out problems, and timely correcting.

    11.5.5 Chemical industry

    In petrochemical industry, refineries in routine preventive maintenance program, uses

    temperature display system. These programs include furnace process monitoring and several

    confirmations. In furnace processes, infrared heating tube displays are used to detect rally carbon

    ratio. These are called coking rally, which may lead to a melting pot of higher DianHuoLv. The

    high temperature condition will reduce the life of the pipes used. This condition can be

    controlled to a large extent by incorporating infrared thermometer detectors in the vulnerable

    sites.

    There are many more instances besides those mentioned previously for which thermography can

    be utilized. Here are a few:

    11.5.6 Aerospace Applications

  • 7/30/2019 Infrared thermographic system

    36/40

    Water ingress in airplane control surfaces and radomes Tire and brake system diagnosis Windshield and wing surface deicing system diagnosis Stress crack and corrosion identification and location Jet and rocket engine analysis Composite materials delamination and disbanding location Target signature analysis

    11.5.7 Petrochemical Applications

    Refinery process line insulation loss or leak detection Refinery process evaluation Heat exchanger quality and efficiency evaluation Furnace refractory (insulation) inspections Furnace internal flame evaluation and tube inspections Flame propagation explosion analysis

    11.5.8 Research and Development Applications

    Design prototyping evaluation

    11.5.9 Pulp and Paper

    Detect uneven heat distribution in Fourdrinier steam boxes Identify wet streaks, nonuniformity that can have adverse effects on paper quality Identify basis weight variations Monitor size press performance Analyze dryer temperatures to look for nonuniformity in dryers Monitor coating to see that it is being applied uniformly to surface of the paper Analyze reel to find anomalies that may be induced by pieces of process equipment

    connected to the paper machine

    Inspect chip piles for hot spots

    11.5.10 Mechanical Systems

  • 7/30/2019 Infrared thermographic system

    37/40

    Boilers Inspect burners for flame impingement and burner management Look at combustion patterns of fuel Detect thermal patterns on boiler tubes and measure tube skin temperature during normal

    operation or when boiler is on standby

    Scan and record temperatures in areas of boiler not monitored Scan the exterior of boiler for refectory damage or locate warmer areas where potential

    refectory damage may occur

    Detect coke buildup in crude oil furnaces Power plant boiler flue gas leak detection Mechanical bearing inspections Heat, ventilation, and air-conditioning equipment evaluation Cold storage cooling losses Detect insulation leaks in refrigeration equipment

    12.0 CONCLUSION

    This seminar provides a summary of infrared thermographic systems. The theory of operation,

    which is based on measuring blackbody thermal radiation, and the fourth power temperature

    dependence of emitted IR energy, demonstrates the strong signal inherent to this sensing

    technology and hence, the high temperature resolution that is achievable. There are various types

    of IR imaging systems. All of them generate images that represent the surface temperatures of an

    object, but they can be comprised of a variety of detector materials, each with unique spectral-

    physical properties, which makes it important to select the proper detector technology for a

    particular application. There are two major families of optical design, focal plane array imagers

    and motorized scanning mirror based imagers. For successful IR imaging, it is essential to be

    aware of the surface property emittance. Knowledge of surface emittance, in the spectral band of

    the imager being used, allows the thermographer to distinguish the emitted portion of the

    measured radiation from the reflected portion. The thermal setup and environment also require

    special considerations for successful thermography. The thermographer often manipulates the

  • 7/30/2019 Infrared thermographic system

    38/40

    overall heat transfer situation that leads to the temperature differences imaged on the specimen of

    interest in order to produce the most accurate and informative thermal data. Scientific

    temperature mapping is a methodology that improves upon the absolute accuracy of quantitative

    thermography. The procedures employed include external referencing targets and background

    mirrors. Typical applications of IR imaging include: thermographic inspection for predictive

    maintenance, medical imaging, non-destructive materials testing, and building component

    testing.

    13.0 Future of IR imaging

    The future of IR imaging will likely see many more applications emerge, as this powerful

    technology becomes more accessible with less expensive and more compact equipment. New

    detector types such as microbolometers, QWIPS or direct sensing optical systems (change of

    refraction due to temperature) will become more prominent. IR imagers will be substantially

    smaller and more portable with reduced power requirements, as a result of compact focal plane

    array optics and not being burdened by providing cryogenic cooling for sensors. Increasing

    computing power will make image analysis and processing possible for online usage. Image

    resolution of focal plane arrays is expected to rise. Mega-pixel arrays will likely be available for

    thermal imaging. Faster image acquisition speeds will open a new world of diagnostic

    possibilities for very fast thermal phenomena such as combustion or blasting technology. The

    various detection mechanisms may make IR imagers even more specialized for their intended

    application. Selecting the appropriate technology for the job will continue to be important. For

    instance, an imager with a very high frame rate may lack the ability to make absolute

    temperature measurements as accurately as an imager designed for slower phenomenon, where

    absolute accuracy is most crucial. Overall, there is much to be learned by seeing temperatures

  • 7/30/2019 Infrared thermographic system

    39/40

    with an IR imager, and this capability is likely to grow more powerful, and become more

    accessible, in the future.

    14.0 References

    Carlomagno, G. M., Heat Transfer Measurements by Means of Infrared Thermography,In Measurement Techniques, von Karman Institute, Lect. Series 2005-09, pp. 1-114,

    Rhode-Saint-Gense, Belgium, 2009

    The Encyclopedia of Imaging Science and Technology by Joseph P. Hornak, Editor-in-Chief, John Wiley and Sons.

    De Luca, L. and Cardone, G., MTF Cascade Model for a Sampled IR Imaging System,Applied Optics, vol. 30, pp. 1659-1664.

    H. Kaplan, Practical Applications of Infrared Thermal Sensing and Imaging Equipment,Second Edition, Proc. Soc. of Photo-Opt. Instrumentation Eng. (SPIE), TT34, 160, 2003.

    D. Turler, B.T. Griffith, D. Arasteh. Laboratory procedures for using infraredthermography to validate heat transfer models. Insulation materials: Testing and

    Applications: Third Volume ASTM STP 1320. R. S. Graves and R. R. Zarr, ed.

    Philadephia, PA, American Society for Testing and Materials, 2001

    B.T. Griffith, H. Goudey, D. Arasteh, Infrared Thermography Measurements ofWindow Thermal Test Specimen Surface TemperaturesASHRAE Transactions ?(not yet

    published) American Society of Heating, Refrigeration and Air-Conditioning Engineers

    Inc., 2002

    V. Vavilov, X. Maldague, J. Picard, R.L. Thomas, L.D. Favro, Dynamic thermaltomography: New NDE technique to reconstruct inner solids structure by using multiple

  • 7/30/2019 Infrared thermographic system

    40/40

    IR image processing, Review of Progress in Quantitative NDE, D.O. Thompson and

    D.E. Chimenti ed., Brunswick, Maine, 1991, 11A, Plenum Press, 1999, pp. 425-432

    X. Maldague, P.O. Moore, Infrared and Thermal Testing. American Society for Non-destructive Testing, 3

    rdEd. Columbus Ohio, 2006

    Internet:http://www.infrared institute.com

    http://www.sensortips.com/2009/09/14/understanding-infrared-thermometry

    http://www.ces-world.com/InfraredTech/mechanical1.html

    http://www.eutechinst.com/index.html

    http://www.sensortips.com/2009/09/14/understanding-infrared-thermometry/http://www.ces-world.com/InfraredTech/mechanical1.htmlhttp://www.ces-world.com/InfraredTech/mechanical1.htmlhttp://www.eutechinst.com/index.htmlhttp://www.eutechinst.com/index.htmlhttp://www.eutechinst.com/index.htmlhttp://www.ces-world.com/InfraredTech/mechanical1.htmlhttp://www.sensortips.com/2009/09/14/understanding-infrared-thermometry/