InfinityLab Purification Capillary KitsTechnical Note
This note describes the capillary connections for Agilent preparative systems.
Agilent Technologies
Contents
Abbreviations 3
List of Modules 4
Overview 5General Installation Notes 5Installation Notes for 1260 Infinity II Column Organizer G9328A 6Installation Notes for 1290 Infinity II Column Compartment G7163B 8
Configurations with 1260 Infinity II Column Organizer (G9328A) 10Entry Level System Configurations (with Manual Injector) 10Standard UV/LC System Configurations 16Standard Mass-Based Systems Configuration 24
Configurations with 1290 Infinity II Preparative Column Compartment (G7163B) 28Entry Level System Configurations (with Manual Injector) 28Standard UV/LC System Configurations 34Standard Mass-Based System Configurations 40
Accessories 44UV Accessories 44MS Accessories 44
2
AbbreviationsTable 1 Abbreviations
Modules
VWD Variable wavelength detector
DAD Diode array detector
MI Manual Injector
DC Delay Coil Organizer
RC Recovery collector
FC Fraction collector
PRV Pressure relieve valve
BPR Back pressure regulator
FM Flow modulator
Fittings
R Removable fitting
for flex capillaries: 5067-5966
for rigid capillaries: 5067-5403
S Swagelock fitting
Capillaries and tubings
UV1...5 Capillaries to connect the modules of the liquid chromatographic part of the system
MS1...6 Capillaries to connect the modules of the mass spectrometric part of the system
0.3f 0.3 mm ID flex capillary
0.5r 0.5 mm ID rigid capillary
ESD Electrostatic discharge
Material
SST Stainless steel
PTFE Polytetrafluorethylene
PEEK Polyether ether ketone
3
List of ModulesTable 2 List of modules
G7161A 1260 Infinity II Preparative Binary Pump
G7161B 1290 Infinity II Preparative Binary Pump
G7114A 1260 Infinity II Variable Wavelength Detector
G7115A 1260 Infinity II Diode Array Detector WR
G7165A 1260 Infinity II Multiple Wavelength Detector
G7159B 1290 Infinity II Preparative Open-Bed Fraction Collecor
G7166A 1260 Infinity II Preparative Valve-Based Fraction Collector
G1364E 1260 Infinity II Preparative Fraction Collector
G1328D 1260 Infinity II Manual Preparative Injector
G7157A 1260 Infinity II Preparative Autosampler
G7170B 1290 Infinity II MS Flow Modulator
G9324A 1260 Infinity II Delay Coil Organizer
5067-6857 PREP Pressure Relief Valve Kit
G7163B 1290 Infinity II Preparative Column Compartment
G9328A 1260 Infinity II Column Organizer
G7110A 1260 Infinity II Isocratic Pump
5188-8053 ERI Remote and Leak Module
G6125B MSD
4
OverviewGeneral Installation Notes
Capillaries and Fittings• Capillary UV/LC kits (5067-6858, 5067-6859, 5067-6860, 5067-6809) are flow
dependent. The proper kit needs to be selected based on the desired flow range.
• The MS Upgrade kit (5067-6861) is only one kit. It contains all capillaries for all flow ranges.
• Four capillaries (UV2A, 2B, 2C, 2D) are provided to connect sampler and column.
• All capillaries with an inner diameter (ID) of 0.3 mm and 0.5 mm are flexible.
Only the MS1B capillary from column to flow modulator is always rigid. This capillary is guided between modules and needs to remain in its fixed position.
• The bypass of the flow modulator is part of accessory kit G7170-68705.
• All capillaries going to the ports of the flow modulator need UHP-FF fittings.
• The Multiple Wavelength Detector is not depicted in the drawings. The same principles as for the Diode Array Detector apply.
• Only removable fittings are used for columns.
• Either Quick Turn Fittings for flexible capillaries, or
• UHP-FF for rigid capillaries.
General rules• If the height of the stack configuration exceeds the safety limit (0.89 m), the
solvent cabinet has to be placed next to the stack on the base plate.
• The base plate is needed in case the Solvent Cabinet is placed on a bench by itself, and has to be ordered separately (p/n G1328-44111).
NOTE Do not use swagelock for the flow modulator, the ports can be damaged.
5
Installation Notes for 1260 Infinity II Column Organizer G9328A
Columns and Column Holders• To keep the capillary from column to detector as short as possible, the
column needs to be mounted at the highest possible position, and in the front of the column organizer.
• Two capillaries (UV2A, UV2B) are provided to connect sampler and column:
• UV2A is used for columns shorter than 15 cm,
• UV2B is used for long columns of 15 cm and more.
• Fix columns with the following column holders:
• M size holders (5067-6849) for 9.4 mm/10 mm, 21.2 mm and 30 mm columns,
• L size holders (G7163-60002) for 50 mm columns,
L size holders have to be ordered separately.
• The 25 cm columns require two M size holders, shorter columns need only one holder.
• One M holder can be used for pre-column.
Manual Injection Valve G1328D (MI)• The MI must be installed in the first groove of the column organizer
extrusion plate.
• The MI can be installed at any height.
• In combination with columns shorter than 25 cm, the MI can be positioned in the middle of the column organizer.
• In combination with columns 25 cm or longer, the MI needs to be installed at the lowest possible position.
• The MI requires a suitable waste container (bottle or beaker) for waste and siphoning capillaries.
NOTE The kit does not contain a waste container. The waste container must be provided by the customer. The waste container has to be placed on the leak plate.
6
• In the case that a 20 mL stainless steel loop is used, it has to be bent in the direction of the extrusion to leave enough space for the container on the leak plate. None of the other loops limit the space for the container.
• The waste and siphoning capillaries have to be bent. The end of the capillaries have to be at the same height as the position of the injection needle.
Pressure Relieve Valve (PRV, 5067-6857) and Back Pressure Regulator (BPR, 5067-6840)• PRV must be installed with every configuration.
• The PRV and BPR come with the Preparative LC system.
• The PRV is installed at the outlet of the Delay Coil Organizer, if available.
• If the PRV is used as a single item, the following measures are required:
• The PRV has to be connected with system tubing from the detector. This connection requires union 10-32 to ¼-28.
• The PRV has to be fixed to the column organizer using S holders.
• The PRV has to be placed on the upper part of the extrusion.
• The PRV waste tubing is guided to the leak plate by S holders.
• The BPR needs to be installed ahead of the Recovery Collector.
• The use of BPR is recommended only for certain flow rates, check for operating instructions.
• The BPR can be fixed at any position on the extrusion using S holders.
NOTE The PRV is required for safety reasons.
NOTE The length of the capillary from the detector to the PRV is intentionally limited.
7
Installation Notes for 1290 Infinity II Column Compartment G7163B
General• The Delay Coil Organizer must be installed at the highest valve position.
• The Flow Modulator must be installed in the second highest valve position.
• The Recovery Collector can be installed at any available free valve position.
Columns and Column Holders• The column should be mounted at the highest possible position of the
Column Compartment.
• The capillaries UV2A, UV2B, UV2C and UV2D are used to connect the column with the sampler. Select the suitable capillary length depending on the position or length of the column.
• Fix columns with the following column holders:
• M size holders (5067-6849) for 9.4 mm/10 mm, 21.2 mm and 30 mm columns,
• L size holders (G7163-60002) for 50 mm columns,
• The 25 cm columns require two M size holders, shorter columns need only one M size holder.
• One M size holder can be used for pre-column.
Manual Injection Valve G1328D (MI)• The MI should be installed on the bottom part of the small mounting plate of
the Column Compartment in the first groove.
• The MI requires a suitable waste container (bottle or beaker) for waste and siphoning capillaries.
• Columns need to be installed such that they do not collide with MI, loop or siphoning capillaries, usually on the left part of the extrusion.
NOTE The kit does not contain a waste container. The waste container must be provided by the customer. The waste container has to be placed on the plateau of the Column Compartment.
8
• The waste and siphoning capillaries have to be bent. The end of the capillaries have to be at the same height as the position of injection needle and should point down to the waste container.
Pressure Relieve Valve (PRV, 5067-6857) and Back Pressure Regulator (BPR, 5067-6840)• PRV must be installed with every configuration.
• The PRV and BPR come with the Preparative LC system.
• The PRV is installed at the outlet of the Delay Coil Organizer, if available.
• If the PRV is used as a single item, the following measures are required:
• The PRV has to be connected with system tubing from the detector. This connection requires union 10-32 to ¼-28.
• The PRV has to be fixed to the Column Compartment using S holders (5043-1356).
• The PRV has to be placed on the upper part of the extrusion.
• The PRV waste tubing is guided to the leak sensor by S holders.
• The use of BPR is recommended only for certain flow rates, check for operating instructions.
• The BPR can be fixed at any position on the extrusion using S holders.
NOTE The PRV is required for safety reasons.
NOTE The length of the capillary from the detector to the PRV is intentionally limited.
9
Configurations with 1260 Infinity II Column Organizer (G9328A)Entry Level System Configurations (with Manual Injector)Entry Level System ConfigurationsTable 3 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
UV5 VWD/DAD
PRV 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
10
Table 4 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
Table 5 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
11
12
Entry Level System Configurations with Delay Coil Organizer (G9324A)Table 6 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
UV5 VWD/DAD
PRV 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
Table 7 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
13
Table 8 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
14
15
Standard UV/LC System ConfigurationsConfiguration with Preparative Open-Bed Fraction CollectorTable 9 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
UV5 VWD/DAD
PRV 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
Table 10 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
16
17
Configuration with Preparative Open-Bed Fraction Collector and Delay Coil Organizer (G9324)Table 11 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
UV5 VWD/DAD
PRV 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
Table 12 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
18
19
Configuration with Preparative Fraction CollectorTable 13 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
UV5 VWD/DAD
PRV 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
Table 14 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
20
21
Configuration with Preparative Fraction Collector and Delay Coil Organizer (G9324)Table 15 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
UV5 VWD/DAD
PRV 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
Table 16 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
22
23
Standard Mass-Based Systems ConfigurationTable 17 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
UV5 VWD/DAD
PRV 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
24
-
-
N
Table 18 Upgrade Capillary kit MS-Based Prep LC
Upgrade Capillary kit MS-Based Prep LC Part No. 5067-6861
1 kit 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
1 kit only
From module
To module
Length /cm
ID / mm
Part No.
ID / mm
Part No.
ID / mm
Part No.
ID / mm
Part No.
Fittings
MS1B Column Flow modul.
90 0.3r 5500-1321
0.5r 5500-1322
0.6r 5500-1323
0.94r 5500-1403
RemovableRemovable
MS2 Flow modul.
VWD/DAD
55 0.3f 5500-1397
0.5f 5500-1400
0.6r 5500-1402
0.94r 5500-1405
RemovableSwage
MS3 Make up Flow modul.
120 0.12 5500-1398
FLOW INDEPENDENT CAPILLARIES Swage-Swage
MS4 Flow modul.
MS 120 0.12 5500-1398
Swage-Swage
MS5 MS restriction capillary
7 0.12 5500-1329
Swage-Swage
MS6 MS waste capillary
70 0.25 5500-1330
Swage-NO
Table 19 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
25
Table 20 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
Length/cm ID/mm Part No. Fittings
FM1 8 0.25 5067-6820 Removable-Swage
FM2 10 0.12 5067-6821 Swage-Swage
26
27
Configurations with 1290 Infinity II Preparative Column Compartment (G7163B)Entry Level System Configurations (with Manual Injector)Entry Level System ConfigurationTable 21 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV2C Sampler Column 80 0.3f 5500-1406
0.5f 5500-1384
0.6r 5500-1387
0.94r 5500-1392
Swage-Removable
UV2D Sampler Column 100 0.3f 5500-1407
0.5f 5500-1385
0.6r 5500-1388
0.94r 5500-1394
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
28
Table 22 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
Table 23 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
29
30
Entry Level System Configuration with Delay Coil Organizer (G9324A)Table 24 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV2C Sampler Column 80 0.3f 5500-1406
0.5f 5500-1384
0.6r 5500-1387
0.94r 5500-1392
Swage-Removable
UV2D Sampler Column 100 0.3f 5500-1407
0.5f 5500-1385
0.6r 5500-1388
0.94r 5500-1394
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
Table 25 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
31
Table 26 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
32
33
Standard UV/LC System ConfigurationsStandard UV/LC System ConfigurationsTable 27 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV2C Sampler Column 80 0.3f 5500-1406
0.5f 5500-1384
0.6r 5500-1387
0.94r 5500-1392
Swage-Removable
UV2D Sampler Column 100 0.3f 5500-1407
0.5f 5500-1385
0.6r 5500-1388
0.94r 5500-1394
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
34
Table 28 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
Table 29 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
35
36
Standard UV/LC System Configurations with Delay Coil Organizer (G9328A)Table 30 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV2C Sampler Column 80 0.3f 5500-1406
0.5f 5500-1384
0.6r 5500-1387
0.94r 5500-1392
Swage-Removable
UV2D Sampler Column 100 0.3f 5500-1407
0.5f 5500-1385
0.6r 5500-1388
0.94r 5500-1394
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
Table 31 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
37
Table 32 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
38
39
Standard Mass-Based System ConfigurationsTable 33 Capillary Kit Prep UV/LC Overview
System Capillary Kit Prep LC/UV 5067-6858 5067-6859 5067-6860 5067-6809 1 kit for each flow rate
Flow range 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
From module
To module
Length/cm
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
ID/mm
Part no.
Fittings
UV1 Pump Sampler 65 0.3f 5500-1308
0.5f 5500-1309
0.6r 5500-1318
0.94r 5500-1390
Swage-Swage
UV2A Sampler Column 50 0.3f 5500-1316
0.5f 5500-1317
0.6r 5500-1320
0.94r 5500-1391
Swage-Removable
UV2B Sampler Column 65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Swage-Removable
UV2C Sampler Column 80 0.3f 5500-1406
0.5f 5500-1384
0.6r 5500-1387
0.94r 5500-1392
Swage-Removable
UV2D Sampler Column 100 0.3f 5500-1407
0.5f 5500-1385
0.6r 5500-1388
0.94r 5500-1394
Swage-Removable
UV3 Column VWD/DAD
65 0.3f 5500-1314
0.5f 5500-1315
0.6r 5500-1319
0.94r 5500-1393
Removable-Swage
UV4 VWD/DAD
DC Box/PRV
55 0.3f 5500-1408
0.5f 5500-1386
0.7r 5500-1389
0.94r 5500-1395
Swage-Swage
40
-
-
N
Table 34 Upgrade Capillary kit MS-Based Prep LC
Upgrade Capillary kit MS-Based Prep LC Part No. 5067-6861
1 kit 4 – 8 mL/min
15 – 40 mL/min
40 – 80 mL/min
80 – 200 mL/min
1 kit only
From module
To module
Length /cm
ID / mm
Part No.
ID / mm
Part No.
ID / mm
Part No.
ID / mm
Part No.
Fittings
MS1A Column Flow modul.
50 0.3f 5500-1396
0.5f 5500-1399
0.6r 5500-1401
0.94r 5500-1404
RemovableRemovable
MS2 Flow modul.
VWD/DAD
55 0.3f 5500-1397
0.5f 5500-1400
0.6r 5500-1402
0.94r 5500-1405
RemovableSwage
MS3 Make up Flow modul.
120 0.12 5500-1398
FLOW INDEPENDENT CAPILLARIES Swage-Swage
MS4 Flow modul.
MS 120 0.12 5500-1398
Swage-Swage
MS5 MS restriction capillary
7 0.12 5500-1329
Swage-Swage
MS6 MS waste capillary
70 0.25 5500-1330
Swage-NO
Table 35 Preparative Open-Bed Fraction Collector tubing kits (G7159B)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Open-Bed Fraction Collector tubing kit
Tubing Kit 8 (G9321-60953)
Tubing Kit 50 (G9321-60952)
Tubing Kit 150 (G9321-60954)
Tubing Kit 200 (G9321-60951)
Suitable for the flow up to
1 – 8 mL/min 15 – 50 mL/min
50 – 150 mL/min
150 – 200 mL/min
41
Table 36 Preparative Fraction Collector tubing kits (G1364E)
P/N 5067-6858 5067-6859 5067-6860 5067-6809
Preparative Fraction Collector tubing kit
G1364-68603 G1364-68604 G1364-68605
Comment 4 – 8 mL/min 15 – 40 mL/min
40 – 100 mL/min
Length/cm ID/mm Part No. Fittings
FM1 8 0.25 5067-6820 Removable-Swage
FM2 10 0.12 5067-6821 Swage-Swage
42
43
AccessoriesUV Accessories
MS Accessories
Table 37 Accessories for kits 5067-6858, 5067-6859, 5067-6860, 5067-6809
A-Line Quick Turn LC Fitting for Column 2 pcs for kits 5067-6858 and 5067-6859 5067-5966
UHP-FF FITTING for Column 2 pcs for kits 5067-6860 and 5067-6809 5067-5403
3m Tube ESD PTFE OD2.5 ID1.6 5023-3086
ESD-Fitting OD-2.5mm 5023-2871
Union Flat-Bottom 1/4-28 ESD 5023-2508
Plastic and PEEK tubing cutter 8710-1930
Fitting-Handle 5043-1471
Sealtight-Handle 5043-0915
Table 38 Accessories
3 m Tube ESD PTFE OD 2.5 ID 1.6 5023-3086
ESD-Fitting OD-2.5 mm 5023-2871
Fitting adapter 10-32 to 1-4-28 SST 5023-2875
TEE-LOW DEAD VOLUME 0100-1818
Fitting-Handle 5043-1471
Sealtight-Handle 5043-0915
NOTE The flow modulator bypass is part of the G7170-68705 Flow Modulator Accessory Kit.
*G7161-90201**G7161-90201*G7161-90201
Part Number:G7161-90201