Hydrophilic Interaction ChromatographyHydrophilic Interaction Chromatography(HILIC)

Bill Champion Agilent Technologies, Inc.

800-227-9770 opt 3/opt3/opt [email protected]

Page 1 April 27, 2010

HILIC - Agilent Restricted

Page 2: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

HILICHILIC“A method of recent attention”

Bill Champion Agilent Technologies, Inc.

800-227-9770 opt 3/opt3/opt [email protected]

Page 2 April 27, 2010

HILIC - Agilent Restricted

Page 3: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

HILICHILIC… but not the universal answer.

Bill Champion Agilent Technologies, Inc.

800-227-9770 opt 3/opt3/opt [email protected]

Page 3 April 27, 2010

HILIC - Agilent Restricted

Page 4: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Objective - Outline

• Overview - Why HILIC is used Advantages/Disadvantagesg g

• How HILIC works

• Examples

• Method developmentp

• Comments concerning use of HILIC

Page 4 April 27, 2010

HILIC - Agilent Restricted

Page 5: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Why HILIC Is Used

• Retention of polar components

Similar to NPLC but with aq mobile phase (General• Similar to NPLC but with aq mobile phase (Generalreversal of elution order from RPLC)

• Several different stationary phases available• Several different stationary phases available

• MS compatible

• May simplify sample prep

Page 5 April 27, 2010

HILIC - Agilent Restricted

Page 6: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

… but there is always a price

• Slower equilibration than RPLC

• Peak distortion with mp-sample solvent mis-match (i.e., too much water in sample solvent)

• Poor retention of anions on silica

• Mechanism not well understood

Page 6 April 27, 2010

HILIC - Agilent Restricted

Page 7: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Additional Benefits, Comments

• ACN has low viscosity – allows high flow rates

• Van Deemter plot similar to RPLC• Van Deemter plot similar to RPLC

• Not truly “orthogonal” to RPLC, but dissimilar not just inverse separation– not just inverse separation

– complementary technique

• Type B Silica high purity less acidic• Type B Silica, high purity, less acidic – less active but more reproducible

Page 7 April 27, 2010

HILIC - Agilent Restricted

Page 8: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Parameters to Choose

• Type of stationary phase

O i l t t ti• Organic solvent concentration

• Type of buffer

• Buffer (salt) concentration

• pH

• Temperature

April 27, 20108

HILIC - Agilent Restricted

Page 9: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Typical Stationary Phases - Polar

• Silica

• Amine

• DiolDiol

• Amide

• Zwitterionic

Page 9 April 27, 2010

HILIC - Agilent Restricted

Page 10: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Some Typical Polar Stationary Phases

• Silica, Si-OH Si-O H(+)(-)

• Amine, -(CH2)3-NH2

Di l (CH ) O CH (CHOH)CH OH• Diol, -(CH2)3-O-CH2-(CHOH)CH2OH

• Amide, -(CH2)n-(CO)NH22 n 2

• Zwitterionic, -(CH2)n-N(Me)2-(CH2)n-SO3(+) (-)

Page 10 April 27, 2010

HILIC - Agilent Restricted

Page 11: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Typical Mobile Phase

• Water /ACN

• Buffer

• pH

Page 11 April 27, 2010

HILIC - Agilent Restricted

Page 12: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Typical Mobile Phase

• Water (at least 2- 3%, ~ 25%)/CANSolvent strength:gWater > MeOH > EtOH > IPA > ACN

• Buffer AmOAc, 5 to 20 mMBuffer AmOAc, 5 to 20 mMBuffer must be soluble in high CAN concentration

• pH 3 8• pH 3 – 8

Page 12 April 27, 2010

HILIC - Agilent Restricted

Page 13: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Mobile Phase

• Need some water for hydrationNeed some water for hydration

• Water is “strong solvent”

• Increasing water, decreases retention

April 27, 201013

HILIC - Agilent Restricted

Page 14: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Buffer

• Controls Ionization of Silica

• Controls Ionization of Analyte

• Ammonium acetate, formate – good solubility, “MS , g y,friendly”

• Phosphate - poor solubility at high % ACNPhosphate poor solubility at high % ACN

April 27, 201014

HILIC - Agilent Restricted

Page 15: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

pH

• Controls Ionization of SilicaControls Ionization of Silica

• Controls Ionization of Analyte

• Organic solvent affects the actual [H+]

April 27, 201015

HILIC - Agilent Restricted

Page 16: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Temperature

• Higher Temperature Decreases Ret’n Timeg

• Higher Temperature Increases Column Efficiency (increases N)( )

• Lower Temperature May Increase Selectivity

April 27, 201016

HILIC - Agilent Restricted

Page 17: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Separation Mechanism

Page 17 April 27, 2010

HILIC - Agilent Restricted

Page 18: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Retention in Reversed-Phase Separations:Mechanism Similar to Liquid-Liquid Extraction

Heptane Heptane

In RPLC decrease retention by decreasing polarity of mobile phaseMechanism Similar to Liquid Liquid Extraction

HeptaneNon-Polar

HeptaneNon-Polar

H20Polar

H20/MeOHLess Polar

Relatively non-polar Decreasing polarity of aqueousSolvent Polarity: H2O > MeOH, ACN > EtOH > IPA >> Heptane

analyte (e.g., toluene)favors non-polar phase.

phase increases affinity for analyte in H2O/MeOH phase.

Page 18 April 27, 2010

HILIC - Agilent Restricted

Page 19: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Retention in Reversed-Phase Separations:Mechanism Similar to Liquid-Liquid Extraction

Heptane Heptane

In RPLC decrease retention by decreasing polarity of mobile phaseMechanism Similar to Liquid Liquid Extraction

HeptaneNon-Polar

HeptaneNon-Polar

H20Polar

H20/MeOHLess Polar

Relatively non-polar Decreasing polarity of aqueousMobile Phase Strength: H2O < MeOH, ACN < EtOH < IPA << Heptane

analyte (e.g., toluene)favors non-polar phase.

phase increases affinity for analyte in H2O/MeOH phase.

Page 19 April 27, 2010

HILIC - Agilent Restricted

Page 20: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Comparison of HILIC and RPLC RPLC• Non-polar stationary phase (e.g., C18)• Polar mobile phase (i.e., H20/MeOH, H20/ACN, etc.)• Decrease retention by decreasing polarity of mobile phase

(e.g., increase ACN in mobile phase to decrease retention)(e.g., increase ACN in mobile phase to decrease retention)

HILICP l t ti h ( ili )• Polar stationary phase (e.g., silica)

• Polar mobile phase (i.e., H20/ACN) • Decrease retention by increasing polarity of mobile phasey g p y p

(i.e., increase H2O in mobile phase to decrease retention)

Page 20 April 27, 2010

HILIC - Agilent Restricted

Page 21: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Typical Elution Order for Test Compounds on HILIC Column – RPLC Column Test MixtureColumn RPLC Column Test Mixture

4-chloronitrobenzenemAU

35

40

Phenol

Column 4.6 x 150 mmMobile phase 75% ACN: 25% H2OFlow Rate 0.5 ml/ml Col Temp 25 °C

20

25

30 Injection Volume 5 µl

1 Notice Uracil

10

15Naphthalene

Uracil1. Notice Uracil

•normally unretained in RPLC

•retained in HILIC

i1 2 3 4 5 6 7 8 9

0

5

eta ed C

2. Notice completely different retention order than in RPLC

April 27, 201021

min1 2 3 4 5 6 7 8 9

HILIC - Agilent Restricted

Page 22: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

How Does HILIC Work on Silica Based Columns?

A water layer must be adsorbed onto the stationary phase.

The polar analyte partitions into and out of this adsorbed layerThe polar analyte partitions into and out of this adsorbed layer.

A charged polar analyte can also undergo ion exchange with the charged silica molecules (i.e., cation exchange with amines) g ( g )

The combination of these mechanisms drives retention in HILIC.

Retention/elution is from least to most polar – the opposite of p ppreversed-phase LC.

HILIC offers more retention than reversed-phase for very polar bbases.

April 27, 201022

HILIC - Agilent Restricted

Page 23: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

How Does HILIC Work on Silica Based Columns?Potential MechanismsPotential Mechanisms

CH CHCHACN

ACNACN

ACN ACN

3

CH 2CHCH 3

NH+ACN

ACN

ACN ACNACNACN 1.

O OO OH2OH2O

H2O H2O H2O

H2O

H2OH2O

- - - -

H2O

3

CH 2CHCH 3

NH+2 O OO O -32.

1. Partitioning in and out of adsorbed water layer2. Ion exchange with silanols

April 27, 201023

HILIC - Agilent Restricted

Page 24: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Examples

Page 24 April 27, 2010

HILIC - Agilent Restricted

Page 25: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Some Typical Analytes for HILIC• Amino acids (when only a few are of interest) • Nucleobases (purines and pyrimidines,

adenine, guanine, thymine, th mine c tosine racil)thymine, cytosine, uracil)

• Nucleosides (Adenosine, cytosine etc.)• Alkaloids• Carbohydrates• Carbohydrates• Polar compounds, small basic compounds

NH2

Cytosine

NH2

NNMelamineCH CHCH

NH2NH2N

Melamine

Amphetamine

CH2CHCH3

NH3+

April 27, 201025

Amphetamine

HILIC - Agilent Restricted

Page 26: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Comparison of Reversed-Phase and HILIC

Structures of Drug Compounds Studied

NF

O

a) ParoxetineAntidepressantMW 329 36

OO

MW 329.36

OO

N NOb) Ranitidine

SN NO

N

NO2

AntiulcerativeMW 314.41

April 27, 201026

Basic portion of molecule, impacts HILIC retention

HILIC - Agilent Restricted

Page 27: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Chromatographic ConditionsRPLCRPLC

Instrument: Agilent Series 1100 HPLC

Column: ZORBAX Eclipse XDB-C18 (2.1 mm × 150 mm, 5 μm)

Mobile phase: A: 8-mM HCO2NH4 in water; B: 8-mM HCO2NH4 in 95% ACN/5% water

Gradient: 5% B to 90% B in 10 min

Column temp: 40 °C

Sample volume: 5 μL

Flow rate: 0.3 mL/min

HILIC

Everything same as for RPLC except for column and gradient conditions:

Column: ZORBAX Rx-SIL (2.1 mm × 150 mm, 5 μm)

Gradient: 100% B to 50% B in 10 min

April 27, 201027

HILIC - Agilent Restricted

Page 28: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

LC/MS/MS Separation of Paroxetine & Ranitidine on ZORBAX Eclipse XDB-C18 Column (RPLC Mode)

RPLC separationZORBAX Eclipse XDB-C18 (2.1mm×150mm)

ZORBAX Eclipse XDB C18 Column (RPLC Mode)

Paroxetinep ( )

100 ppb5% B→ 90% B (10 min) gradient

ce

m/z=330→192

Abu

ndan

c

RPLC has poor peak shape and limited sensitivity for Ranitidine!

NF

O

SN NO

N

Ranitidine

A

OO

N

NO2m/z=315→176

April 27, 201028

min2 4 6 8 10 12HILIC - Agilent Restricted

Page 29: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

LC/MS/MS Separation of Paroxetine and Ranitidine on ZORBAX Rx-SIL Column (HILIC Mode)- 100 ppb Level

HILIC separationZorbax RX SIL Silica column

ZORBAX Rx SIL Column (HILIC Mode) 100 ppb Level

Paroxetine m/z=330→192

Zorbax RX-SIL Silica column (2.1mm×150mm)100 ppb100% B→ 50% B (10 min) gradient

Ranitidine / 315 176

NFHILIC Separation has better peak m/z=315→176

SN NO

NOO

Ohas better peak shape and more sensitivity for ranitidine! N

NO2

April 27, 201029

62 4 8 10 12minHILIC - Agilent Restricted

Page 30: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

HILIC Separations are Ideal for LC/MSAnalysis of MelamineAnalysis of Melamine

HILIC HILIC Separation Using ZORBAX Rx-Sil

MelamineMelamineMelamine 50pptMelamine 50pptNH2

NN

NH2NH2N

HPLC system : Agilent 1200 RRLC Column : Agilent ZORBAX Rx-Sil, 2.1×150 mm, 5 um Injection Volume : 10 μLInjection Volume : 10 μL Temp : 40Flow rate : 0.2 mL/min Mobile phase : A - 5 mM Ammonium acetate in Water

: B - 5 mM Ammonium acetate in ACN I ti 95%B

April 27, 201030

Isocratic : 95%B

HILIC - Agilent Restricted

Page 31: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of Injection Solvent & Volumeon Melamine Retentionon Melamine Retention

@ACN 1uL@ACN 1uL

@ACN 10 uL@ACN 10 uL

@ACN 20 uL@ACN 20 uL

@ACN 50 uL@ACN 50 uL

@Water 1 uL@Water 1 uL@Water 1 uL@Water 1 uL

@Water 10uL@Water 10uL

Don’t inject in strong solvent!!!!

April 27, 201031

HILIC - Agilent Restricted

Page 32: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Mobile Phase

Page 32 April 27, 2010

HILIC - Agilent Restricted

Page 33: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of Increasing % ACN on Cyanuric Acid RetentionCyanuric Acid Retention

60%60% Decreasing strong solvent (increasing AACN) increases retention

70%70%AACN) increases retention.

Same behavior as reversed-phase but strong and weak solvents are opposite!

80%80%strong and weak solvents are opposite!

90%90%

95%95%

April 27, 201033

HILIC - Agilent Restricted

Page 34: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Retention Time Changes with Changes in Mobile Phase CompositionPhase Composition

me,

min

entio

n Ti

mR

ete

Mobile phase: A= 5 mM NH4OAc/H20B= 5 mM NH4OAc/ACN

% Acetonitrile

4

April 27, 201034

HILIC - Agilent Restricted

Page 35: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of Ammonium Acetate Concentration on Cyanuric Acid RetentionCyanuric Acid Retention

Buffer strength can dramatically affect retention and resolution and5 mM5 mM affect retention and resolution and should be a key parameter in method development and optimization on HILIC columns.

10 mM10 mM

optimization on HILIC columns.

20 M20 M20 mM20 mM

April 27, 201035

HILIC - Agilent Restricted

Page 36: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of pH, Buffer on Retention on Silica16.0

14.0 StrongAcids

10.0

12.0

me 2-nsaAmines

6 0

8.0

Ret

'n T

im p-xsa

phenol

caffeine

nortrip

diphhyd

4.0

6.0 p y

procain

NeutralNeutral

0.0

2.0

%15 M15 M

Ref: McCalley, J. Chromatogr. A 1171 (2007) 46

Page 36

0 1 2 3 41.5 mMpH 3.0

0.1%TFA

15 mMpH 3.0

15 mMpH 3.5..

April 27, 2010

HILIC - Agilent Restricted

Page 37: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of pH, Buffer on Retention on Silica16.0

14.0

procainamide

10.0

12.0

me Amines

6 0

8.0

Ret

'n T

im

nortrip

diphhyd

procaindiphenhydraminebenzylamine

4.0

6.0

A B C - HCOONH4/85% ACN; D 95% ACN

0.0

2.0

%C 15 MA 15 M

Ref: McCalley, J. Chromatogr. A 1171 (2007) 46

A, B, C HCOONH4/85% ACN; D 95% ACN

Page 37

0 1 2 3 4B - 1.5 mMpH 3.0

D - 0.1%TFA

C - 15 mMpH 3.0

A -15 mMpH 3.5..

April 27, 2010

HILIC - Agilent Restricted

Page 38: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of pH, Buffer on Retention on Silica16.0

14.0

A, B, C - HCOONH4/85% ACN; D 95% ACNStrongAcids

p-xylenesulfonic acid

10.0

12.0

me

2-naphthalenesulfonic acid

6 0

8.0

Ret

'n T

im 2-nsa

p-xsa

phenol

caffeine

4.0

6.0

NeutralNeutralcaffeine

h l

0.0

2.0

%C 15 MA 15 M

Ref: McCalley, J. Chromatogr. A 1171 (2007) 46

phenol

Page 38

0 1 2 3 4B - 1.5 mMpH 3.0

D - 0.1%TFA

C - 15 mMpH 3.0

A -15 mMpH 3.5..

April 27, 2010

HILIC - Agilent Restricted

Page 39: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of Buffer, pH

• Controls Ionization of Silica

• Controls Ionization of Analyte

• Organic solvent affects the actual [H+]g [ ]

• Acids can be repulsed (reduces ret’n time) by anionic silicasilica

• Buffer can mask anionic sites (reduce ret’n time of amines increase for acids)amines, increase for acids)

April 27, 201039

HILIC - Agilent Restricted

Page 40: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Effect of Temperature

• Higher temperature – improves kinetics

sharper peaks, shorter ret’n times

• Lower temperature – improves selectivity

• Affects buffer equilibrium (effective pH)

Page 40 April 27, 2010

HILIC - Agilent Restricted

Page 41: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Columns

Page 41 April 27, 2010

HILIC - Agilent Restricted

Page 42: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Comparison of Silica and Amino HILIC Columns

HPLC system Agilent 1200

Column 1 Agilent Zorbax-NH2, 2.1×50 mm, 5 um g

Column 2 Agilent Zorbax-Rx Sil, 2.1×150 mm, 5 um

Injection Volumn 2 uL

T 25Temp 25

Flow rate 0.4 mL/min

Mobile phase A - 5 mM Ammonium acetate in Water

B - 5 mM Ammonium acetate in ACN

Isocratic 95%B

April 27, 201042

HILIC - Agilent Restricted

Page 43: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

HILIC Separation Using Zorbax NH25 mM Ammonium acetate in 5% H2O/ACN

2x10

0.9

0.95

+ MRM (127.10000 -> 85.20020) C_M_-mrm-reversed-NH2-11.d

1 1 2 2

5 mM Ammonium acetate in 5% H2O/ACN

M l i C i A id

0 65

0.7

0.75

0.8

0.85 Melamine Cyanuric Acid

0.45

0.5

0.55

0.6

0.65

0.2

0.25

0.3

0.35

0.4

0

0.05

0.1

0.15

0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5

April 27, 201043

Counts (%) vs. Acquisition Time (min)

HILIC - Agilent Restricted

Page 44: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

HILIC Separation Using Zorbax Rx-Sil5 mM Ammonium acetate in 5% H2O/ACN

2x10

0 9

0.95

1+ MRM (127.10000 -> 85.20020) C_M_Si_01.d

1 1 2 2

C i A id M l i

5 mM Ammonium acetate in 5% H2O/ACN

0.7

0.75

0.8

0.85

0.9 Cyanuric Acid Melamine

0.45

0.5

0.55

0.6

0.65

0.25

0.3

0.35

0.4

0.45

0

0.05

0.1

0.15

0.2

April 27, 201044

0

Counts (%) vs. Acquisition Time (min)0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 5.6 5.8 6 6.2 6.4 6.6

HILIC - Agilent Restricted

Page 45: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

HILIC Separation of Sugars Using Zorbax Rx-NH2Comparison 70/30 and 75/25 ACN/H2Op 2

glucose

April 27, 201045

HILIC - Agilent Restricted

Page 46: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Method Development

Page 46 April 27, 2010

HILIC - Agilent Restricted

Page 47: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Method Development/OptimizationMethod Development/OptimizationSystematic Approach to Method Development

• Stationary phase

• Mobile PhaseMobile Phase

• Buffer, Buffer concentration

pH• pH

April 27, 201047

HILIC - Agilent Restricted

Page 48: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Method Development/OptimizationMethod Development/OptimizationSystematic Approach to Method Development

• Separations are too complicated to expect to find optimum conditions using “Random Walk”

I ti t ff t f b ff H b ff t ti• Investigate effects of buffer, pH, buffer concentration

• Recommend use of experimental design

April 27, 201048

cf: B. Dejaegher, D. Mangelings, Y. Vander Heyden, J. Sep. Sci., 31 (2008) 1438

HILIC - Agilent Restricted

Page 49: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Method Development/OptimizationMethod Development/OptimizationStarting Conditions

• Silica

• Ammonium formate, 5, 10, 20 mMAmmonium formate, 5, 10, 20 mM

• pH 3.0, 3.5, 4.0 (5, 6 ?)

ACN/H O 97% 95% 90% 85% 75%• ACN/H2O 97%, 95%, 90%, 85%, 75%

• TFA?

April 27, 201049

HILIC - Agilent Restricted

Page 50: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Typical Stationary Phases - Polar

• Silica

• Amine

Di l• Diol

• Amide

• Zwitterionic

Page 50 April 27, 2010

HILIC - Agilent Restricted

Page 51: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Elution Order Inversely Related t RPLC BUT N t O itto RPLC, BUT Not Opposite

M h i R t ti i t tl it f RPLC• Mechanism, Retention is not exactly opposite of RPLC

• RPLC cannot be used as predictive model

• Not truly “orthogonal”, but dissimilar separation, complementary technique

• Acids often poorly retained

Page 51 April 27, 2010

HILIC - Agilent Restricted

Page 52: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

Summary - Advantages of HILICRetention of polar compounds

Good peak shape for basic compounds where RPC may give tailing and/or low efficiencyy

Higher flow rates and/or long columns can be used due to low viscosity mobile phases with high organic content; greater efficiency

Enhanced detection sensitivity with MSEnhanced detection sensitivity with MS• Efficient spraying and de-solvation in electrospray MS• As much as 3X sensitivity

Can directly inject extracts from C18 SPE cartridges or acetonitrile precipitated plasma supernatant

Dissimilar to RPLC (2D separations); elution order reversal, ComplementaryDissimilar to RPLC (2D separations); elution order reversal, Complementary separations

April 27, 201052

HILIC - Agilent Restricted

Page 53: Hydrophilic Interaction ChromatographyHydrophilic …...HILIC “A method of recent attention” Bill Champion Agilent Technologies, Inc. 800-227-9770 opt 3/opt3/opt 2 william_champion@agilent.com

References

A.J. Alpert, J. Chromatogr. 499 (1990) 177

B A Bidlingmeyer et al Anal Chem 54 (1982) 442B.A. Bidlingmeyer, et. al., Anal. Chem. 54 (1982) 442

D.V. McCalley, J. Chromatogr. A 1171 (2007) 46

D V McCalley J Chromatogr A 1217 (2010) 858D.V. McCalley, J. Chromatogr. A 1217 (2010) 858

B. Dejaegher, D. Mangelings, Y. Vander Heyden, J. Sep. Sci., 31 (2008) 1438

B. A. Olsen, J. Chromatogr. A 913 (2001) 113

R.Li, Junxiong Huang, J. Chromatogr. A 1041 (2004) 163

T. Langrock, P. Czihal, R. Hoffmann, Amino Acids 30 (2006) 291

Page 53 April 27, 2010