hydrogen peroxide – sensitive enzyme sensor based on phthalocyanine thin film
TRANSCRIPT
Hydrogen peroxide ± sensitive enzyme sensorbased on phthalocyanine thin ®lm
T.A. Sergeyevaa,*, N.V. Lavrikb, A.E. Rachkova,Z.I. Kazantsevab, S.A. Piletskya, A.V. El'skayaa
aInstitute of Molecular Biology and Genetics, National Academy of Sciences of Ukraine, 150 Zabolotnogo Str., 252143 Kiev, UkrainebInstitute of Semiconductors Physics, National Academy of Sciences of Ukraine, 45 Prospekt Nauki Str., 252650 Kiev, Ukraine
Received 30 September 1998; received in revised form 10 February 1999; accepted 21 February 1999
Abstract
An enzyme biosensor speci®c for hydrogen peroxide was developed using a new conductometric transducer based on tetra-
tert-butyl copper phthalocyanine (ttb-CuPc) thin ®lms and horseradish peroxidase as sensitive element. This analytical system
is based on detection of molecular iodine produced as a result of the oxidation of the iodide ions by hydrogen peroxide in the
presence of horseradish peroxidase. For the detection of the peroxidase-initiated reaction the ability of the ttb-CuPc thin ®lm
to change its conductivity in response to the appearance of molecular iodine is used. To minimise the interfering effect of the
aqueous electrolyte on the conductometric response of the ttb-CuPc thin ®lm itself, gold interdigitated electrodes bearing ttb-
CuPc layer were covered with a hydrophobic gas-permeable membrane. Thermally evaporated calixarene or plasma
polymerised hexamethyldisiloxane was used as a gas-permeable membrane material. In order to assess the optimum sensor
technology as well as the operating regime, impedance spectroscopy data were analysed. For biosensor creation horseradish
peroxidase was deposited on the sensitive part of the electrodes in a cross-linked bovine serum albumin matrix. The possibility
of hydrogen peroxide detection with the biosensor proposed in the range 5±300 mM was demonstrated. The operational
stability of biosensor was at least 7 h and the relative standard deviation did not exceed 10%. When stored at �48C the sensor
response was stable for more than 90 days. The dependencies of the sensor response on pH, buffer and NaCl concentrations
were investigated. # 1999 Elsevier Science B.V. All rights reserved.
Keywords: Horseradish peroxidase; Biosensor; Hydrogen peroxide detection; Interdigitated planar electrodes; Tetra-tert-butyl copper
phthalocyanine; Langmuir±Blodgett ®lms
1. Introduction
Development of reliable methods of hydrogen per-
oxide determination is of great importance in both
biological and industrial ®elds. Detection of hydrogen
peroxide is important from several points of view:
1. it is widely used in food industry as a sterilising
agent and may be present in ®nal products which
can lead to the loss of nutritional value and
appearance of toxic compounds such as hydro-
peroxides, epoxides and aldehydes [1];
2. hydrogen and organic peroxides can be released
in the environment from industrial processes
and during the ozonation of drinking water
[2,3];
Analytica Chimica Acta 391 (1999) 289±297
*Corresponding author. Tel.: +380-044-266-07-49; fax: +380-
044-266-07-59; e-mail: [email protected]
0003-2670/99/$ ± see front matter # 1999 Elsevier Science B.V. All rights reserved.
PII: S 0 0 0 3 - 2 6 7 0 ( 9 9 ) 0 0 2 0 3 - 2
3. measurements of hydrogen peroxide produced dur-
ing the reaction of many important chemicals in the
presence of oxidases forms the basis for several
biological and medical tests [4±6];
4. hydrogen peroxide is frequently used in pharma-
ceutical and cosmetic formulations [7];
5. horseradish peroxidase is widely used as label in
enzyme-linked immunosorbent assays (ELISA).
Conventional methods for the determination of
hydrogen peroxide including spectrophotometry [8],
colorimetry [9] and chemiluminescence [10] often
involve complicated methods and expensive equip-
ment. A commonly used electrochemical method for
the detection of hydrogen peroxide is its oxidation at
high overpotentials at a wide variety of electrodes [11±
17]. The main disadvantage of this approach is its poor
speci®city because the high voltage applied to the
working electrode can lead to unacceptable interfer-
ences. To reduce the interfering effect caused by
substances which can be oxidised at used potentials,
alternative systems which can operate at much
reduced voltages are being developed for hydrogen
peroxide detection. A number of amperometric
enzyme sensors combining the speci®city of enzy-
matic reaction with the high sensitivity of electroche-
mical transducer have been proposed for this purpose
[18±22]. An amperometric sensor based on horserad-
ish peroxidase (HRP) modi®ed platinised carbon par-
ticles which are capable of reducing hydrogen
peroxide at 0.0 V (vs. SCE) has been developed by
Cardosi [23]. A new scheme for fabricating bulk-
modi®ed amperometric biosensors based on gra-
phite±PTFE electrodes which give a possibility of
hydrogen peroxide detection at 0.0 V (vs. SCE) with
a detection limit of 2.5 mM, and a linear dynamic
range of 2.5±150 mM has been reported [24]. A
potentiometric sensor for this purpose has been pro-
posed by Zul®car et al. [25]. The potentiometric
response of this sensor in the ¯ow system was linear
in the concentration range 0.75±50 mM.
As one can conclude considerable progress and
practical achievements in the ®eld of hydrogen per-
oxide determination have been obtained by using
amperometric and potentiometric biosensors. Until
now no conductometric biosensors for this purpose
have been reported. The main reason for this is that
because of the relatively low impedance of aqueous
medium signi®cant dif®culty is created in proper
operation of the conductometric biosensors. Never-
theless, conductometric biosensors can be more easily
integrated since they do not need a reference electrode
and the conductometric transducers can be manufac-
tured using simple thin ®lm technology [26±28].
In the present study, the possibility of hydrogen
peroxide monitoring was investigated with a perox-
idase enzyme sensor based on a conductometric trans-
ducer having tetra-tert-butyl Cu-phthalocyanine thin
®lms as a sensitive element. The construction of the
transducer is simple, of low cost and has good pos-
sibilities for mass production.
2. Experimental
2.1. Materials
Tetra-tert-butyl copper phthalocyanine (ttb-CuPc)
of high purity was purchased from the Research
Institute of Organic Dyes (Moscow, Russia). Chloro-
form and benzene of a reagent grade were obtained
from Sigma and were distilled prior to use. Hexa-tert-
butyl-calix[6]arene was synthesised at the Institute of
Organic Chemistry (Kiev, Ukraine) in Dr. Kalchenko's
laboratory as described elsewhere [29]. Hexamethyl-
disiloxane was obtained from the microelectronics
plant `̀ Kvazar'' (Ukraine). Glutaraldehyde and bovine
serum albumin (BSA) were purchased from Sigma.
HRP (EC 1.11.1.7) with an activity 257 U/mg (guaia-
col as a reducing substrate) was purchased from
Biozyme. Hydrogen peroxide standardised by iodi-
metric titration was freshly prepared in deionised
water. Iodine (I2) was purchased from Sigma as
crystals of 99% purity. I2 crystals were dissolved in
96% (v/v) ethanol to prepare a 0.2 M stock iodine
solution for further experiments. Potassium iodide
(KI) was obtained from Sigma and a 0.5 M stock
solution of KI was prepared by dissolving in
0.02 M K-phosphate buffer, pH 6.0 immediately
before use.
2.2. Iodine-sensitive transducer and measurements
of the I2 concentration
Interdigitated microelectrodes were purchased from
Emokon (Kiev, Ukraine). The ceramic chips of
290 T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297
5 mm�30 mm size and 0.2 mm thickness contained
two identical pairs of gold interdigitated electrodes.
The electrodes were deposited onto a ceramic sub-
strate by using conventional thin-®lm technology
which includes the following main stages:
1. vacuum deposition of 0.1 mm chromium adhesive
layers onto ceramic substrates;
2. vacuum deposition of 1 mm gold on the top of the
chromium layers;
3. patterning of the interdigitated microstructures by
photolithography.
Each finger in the electrode arrays was 20 mm wide
and approximately 1 mm long. The separation
between the adjacent fingers was 20 mm. The total
sensitive area of each electrode array was about
1.5 mm2.
Iodine-sensitive transducers were prepared by
deposition of thin ttb-CuPc ®lm onto the active area
of the interdigitated electrodes by one of the following
methods: casting from chloroform/benzene solutions,
Langmuir±Blodgett (LB) deposition, and thermal eva-
poration in vacuum. The details of the thermal eva-
poration and LB deposition procedures are given
elsewhere [30,31]. Hydrophobic gas-permeable mem-
branes (HGPMs) deposited on the ttb-CuPc ®lms were
formed by either thermal evaporation of hexa-tert-
butyl-calix[6]arene (HGPM-1) or plasma polymerisa-
tion of hexamethyldisiloxane (HGPM-2) as described
elsewhere [32]. The cross-section of the resulting
structures is schematically shown in Fig. 1. The thick-
ness of each of the deposited organic layers was
de®ned by ellipsometric measurements using Si-
monocrystal satellites.
To measure impedance spectra and conductometric
responses of the sensor, the differential pair of the
interdigitated electrodes with the ttb-CuPc/HGPM
layer was connected to the electronic units as
described earlier [33]. The output voltage was
recorded with a chart recorder and the values corre-
sponding to the steady-state response were taken as a
sensor response. These values were also used for
impedance analysis. To plot impedance spectra, the
absolute values of the complex impedance |Z| of the
sensor were calculated using the following equation:
jZj � Ro�Ug ÿ Ux�=Ux; (1)
where Ro is the resistance across the a.c. nanovolt-
meter input, Ug the a.c. generator output voltage
(60 mV) and Ux is the voltage measured with the
a.c. nanovoltmeter.
2.3. Enzyme immobilisation and measurements of
H2O2 concentration
Horseradish peroxidase was immobilised onto the
transducer surface in a cross-linked BSA matrix
according to the following procedure. Peroxidase
and BSA were dissolved separately in 0.1 M K-phos-
phate buffer solutions, pH 8.0. The resulting protein
concentrations were 100 mg/ml in both solutions.
Equal aliquots of these solutions were mixed with
inositol (Sigma, St. Louis, USA) and ZnSO4 so that
the concentrations of the inositol and ZnSO4 in the
resulting mixture were 5% (w/v) and 0.9 mM, respec-
tively. Approximately 1 ml of this mixture was depos-
ited on a sensitive area of the electrodes. To complete
the polymerisation of the membrane, the sensor chip
was exposed to glutaraldehyde vapours for 30 min and
then dried at room temperature for 15 min.
For H2O2 determination, the prepared enzyme sen-
sor was placed in a 2.5 ml glass cell with gently stirred
0.02 M K-phosphate buffer (pH 6.0) containing
0.01 M KI and 0.15 M NaCl. The measurements were
carried out at room temperature. The sensor responses
were recorded after successive additions of the stock
H2O2 solution into the buffer.
3. Results and discussion
The conductometric method of H2O2 detection is
based on the reaction of iodide ions oxidation by H2O2
Fig. 1. The scheme of a conductometric horseradish peroxidase-
based enzyme sensor with a chemoresistive layer separated from
the region where the enzymatic reaction occurs.
T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297 291
in the presence of HRP:
H2O2 � 2Iÿ � 2H� !peroxidase2H2O� I2 (2)
Therefore, the amount of hydrogen peroxide used in
this reaction can be monitored by measurement of
either iodide consumption or free iodine release.
Iodide-sensitive potentiometric electrodes have been
used previously in order to measure the iodide con-
centration [34,35], but integration of potentiometric
biosensors implies quite sophisticated technology. On
the other hand, the reaction (2) can be monitored by
measurement of the free iodine released. In particular,
we propose to monitor the iodine concentration using
a conductometric transducer with a phthalocyanine
thin ®lm as a sensitive element. To detect the mole-
cular iodine concentration, the ability of ttb-CuPc ®lm
to change its conductivity in response to appearance of
I2 is used in the present study. Chemoresistors includ-
ing phthalocyanins are widely recognised as one of the
most attractive basis for chemical and biosensors due
to their direct electrical responses and good compat-
ibility with electronics [36]. However, high electrical
conductivity of aqueous electrolyte prevents realisa-
tion of biosensors on this basis, therefore properties of
chemoresistors were mainly studied in gaseous envir-
onments only [30,31]. A speci®c feature of the con-
ductometric transducer designed is that ttb-CuPc ®lm
was covered by a hydrophobic gas-permeable mem-
brane (HGPM) to suppress interfering effect of an
aqueous electrolyte (Fig. 1).
To de®ne the optimum technological procedures
and operating frequency, impedance spectra of a set of
the prepared structures were measured in air (Fig. 2,
curve 1) and in various electrolytes, with and without
molecular iodine (Fig. 2, curves 2±4). As can be seen,
at the lowest frequency, the increase of the electrolyte
conductivity has no signi®cant effect on the impe-
dance (Fig. 2, curves 2 and 3), especially in the case of
LB-deposited ttb-CuPc ®lms. At the same time, low-
frequency impedance decreases when 20 mM of free
iodine is present in the electrolyte (Fig. 2, curve 4). It
was shown that the shift of the impedance spectra due
to the presence of iodine is much lower in the case of
the casted ®lms. Therefore, samples with the latter
type of ttb-CuPc ®lms were eliminated from further
experiments. The structures with the LB-deposited
and thermally evaporated ®lms exhibited quite similar
characteristics. Thermal evaporation of the ttb-CuPc
®lms provides all-dry processing of the transducers up
to biomaterial immobilisation. On the other hand, the
LB deposition uses more effectively the material to be
deposited. To minimise the interfering effect of the
electrolyte, the working frequency of 1 Hz was
selected for monitoring the sensor impedance in all
subsequent experiments. The chemoresistors with a
LB-deposited and thermally evaporated ttb-CuPc
layer demonstrated similar sensitivity to the appear-
ance of free iodine in solution. Similar results were
obtained in the case of the thermally evaporated active
layers. The response was rather slow and at least
10 min was needed to reach the steady-state response.
As noted before [29,30] such a slow kinetics is typical
for conductivity changes in phthalocyanine ®lms at
room temperature.
Similar responses on iodine were obtained for both
types of hydrophobic gas permeable membranes,
HGPM-1 and HGPM-2, of equal thickness
(100 nm). However, HGPM-2 was chosen for the
further experiments because of its better adhesion
and mechanical stability.
HRP was immobilised on a surface of the designed
iodine-sensitive transducer in a cross-linked BSA
matrix. To compensate non-speci®c reactions of the
biosensor, an enzyme-free BSA matrix was deposited
on a reference transducer. All measurements were
carried out in a differential pair mode. The calibration
curve of the biosensor obtained in 20 mM potassium-
phosphate buffer, pH 6.0 was found to be linear within
the range of the hydrogen peroxide concentration from
5 to 300 mM (Fig. 3), each point represents the aver-
age of three measurements and the sensor standard
deviation did not exceed 10%. However, no sensor
response was observed when H2O2 was added to the
buffer solution in the absence of KI. Addition of the
iodide in the absence of the substrate, H2O2, did not
cause any noticeable response as well.
The dependence of the enzyme sensor response on
the pH of the sample solution was investigated. The
greatest response takes place at a pH between 5.0 and
6.5 (Fig. 4). This corresponds approximately to the
region where the peroxidase activity has a maximum.
One can therefore conclude that the observed
responses are indeed controlled by the rate of the
enzymatic reaction (2).
The generally recognised main drawback of con-
ductometric biosensors is their poor speci®city exhib-
292 T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297
ited particularly in measurements in buffers with a
high capacity and ionic strength. The dependencies of
the buffer capacity and ionic strength on the magni-
tude of the peroxidase enzyme sensor response have
been investigated. It has been shown that neither ionic
strength nor buffer capacity affected the magnitude of
the sensor responses signi®cantly (Figs. 5 and 6). This
fact seems to be of great importance because it gives a
Fig. 2. Impedance spectra of iodine-sensitive conductometric transducer with casted (a) and Langmuir±Blodgett deposited (b) tetra-tert-butyl
copper phthalocyanine films. The impedance spectra were measured for: 1 ± dry electrodes, 2 ± electrodes immersed in distilled water,
3 ± 0.15 M NaCl solution, 4 ± 0.15 M NaCl solution containing 20 mM I2.
T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297 293
Fig. 3. Dependence of horseradish peroxidase-based enzyme sensor response on substrate concentration. Measurements were carried out in
0.02 M potassium phosphate buffer, pH 6.0, containing 0.15 M NaCl and 0.01 M KI.
Fig. 4. The pH dependence of the activity of horseradish peroxidase immobilised on the surface of the iodine-sensitive transducer.
Measurements were carried out in 0.02 M potassium phosphate buffer containing 0.15 M NaCl, 50 mM H2O2 and 0.01 M KI.
294 T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297
Fig. 5. Dependence of horseradish peroxidase-based enzyme sensor response on NaCl concentration in working solution. Measurements were
carried out in 0.02 M potassium phosphate buffer containing 50 mM H2O2 and 0.01 M KI.
Fig. 6. Dependence of horseradish peroxidase-based enzyme sensor response on buffer concentration of working solution. Measurements
were carried out in potassium phosphate buffer containing 0.15 M NaCl, 50 mM H2O2 and 0.01 M KI.
T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297 295
possibility to avoid the main drawback of the con-
ductometric detection method. The results obtained
prove that the proposed transducers have high and
speci®c sensitivity to the iodine generated by the
enzymatic reaction. Thus, despite some slight effect
of the electrolyte on the measured sensor impedance,
the hydrophobic membrane is thick enough to provide
stable operating of the sensor in the range of physio-
logical pH and ionic strength.
Stability of a biosensor system is a very important
characteristic for their commercial application. The
operational stability test demonstrated that the steady-
state response was not decreased for at least 7 h
(approximately 30 measurements). When stored at
�48C, the biosensor response was stable for at least
three months.
The main characteristics of the horseradish perox-
idase-based biosensor are summarised in Table 1.
Both the response and recovery times could be sig-
ni®cantly reduced if the initial rate of the conductivity
increases was measured instead of steady state
responses. However, this operation mode would also
lead to an increased error due to a much lower signal/
noise ratio in the beginning of the response kinetics.
The range of the detectable concentrations can be
extended to the upper end by adding a diluting system
based, for instance, on a micro¯uidic device. At the
same time, the threshold sensitivity of the proposed
sensor has to be still improved to make it useful for
biodiagnostic analysis of hydrogen peroxide. One of
the possible ways to achieve this is optimisation of
diffusion properties of the matrix for the enzyme
immobilisation. The stability of a biosensor over a
long period of time is one of the most critical char-
acteristics that de®nes the potential for commercial
applications. Although 90 day shelf-lifetime requires
refrigerating at�48C, this can be considered as appro-
priate for most of the practical applications.
4. Conclusions
A chemoresistor based on the organic semiconduc-
tor material, phthalocyanine, has been used in aqueous
medium as a transducer for biosensors. Both the
thermal evaporation and Langmuir±Blodgett deposi-
tion of the chemoresistive material provided transdu-
cers with satisfactory and quite reproducible
parameters. Due to the presence of hydrophobic
gas-permeable membrane covering the iodine-sensi-
tive ®lm, neither ionic strength nor buffer capacity has
a signi®cant in¯uence on the measured responses. The
separation of the chemoresistive material and the
aqueous medium effectively reduces electrolyte inter-
ference, and thus, practically eliminates the main
disadvantage of the conductometric method as applied
to biosensors. It has been shown that free iodine
generated due to the peroxidase-initiated reaction
can be effectively monitored with the designed trans-
ducers.
Acknowledgements
Financial support from Ministry Ukraine for
Science and Technology (Grant 05.41.07/005-92) is
gratefully acknowledged. The authors thank Tammy
Calvert for linguistic advice.
References
[1] M.G. Simic, M. Karel (Eds.), Autoxidation in Foods and
Biological systems, Plenum Press, New York, 1979.
[2] IARC Monographs on the evaluation of the carcinogenic risk
of chemicals to humans. Allyl compounds, aldehydes,
epoxides and peroxides, IARC, Lyon, France 36 (1985) 267.
[3] W.H. Glaze, Environ. Sci. Technol. 21 (1987) 224.
[4] L. Braco, J.A. Daros, M. de la Guardia, Anal. Chem. 64
(1992) 129.
[5] G.F. Hall, A.P.F. Turner, Anal. Lett. 24 (1991) 1375.
[6] R.Z. Kazandjian, J.S. Dordick, A.M. Klibanov, Biotechnol.
Bioeng. 28 (1986) 417.
[7] J. Wang, Y. Lin, L. Chen, Analyst 118 (1993) 277.
[8] R.M. Sellers, Analyst 105 (1980) 950.
[9] Y. Ito, Y. Tonogai, H. Suzuki, J. Assoc. Off. Anal. Chem. 64
(1981) 1448.
[10] G.M. Kok, T.P. Holler, M.B. Lopez, Environ. Sci. Technol. 12
(1978) 1073.
[11] R.M. Iannielo, A.M. Yacynych, Anal. Chem. 53 (1981)
2090.
Table 1
Characteristics of HRP-based biosensor
Operation mode Steady state
Optimum pH 6.0
Linear dynamic range 5±300 mM
Baseline recovery time 15 min
Relative standard deviation 10%
Storage stability >90 days
Operational stability 7 h
296 T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297
[12] J. Wang, N. Naser, L. Angnes, W. Hui, L. Chen, Anal. Chem.
64 (1992) 1285.
[13] L.C. Clarc, Meth. Enzymol. 56 (1979) 448.
[14] H.P. Benetto, D.R. Keyzer, G.M. Delaney, Analyst 8 (1987) 22.
[15] X.H. Cai, B. Ogorevc, G. Tavcar, J. Wang, Analyst 120
(1995) 2579.
[16] M. Somasundrum, K. Kirtikara, M. Tanticharoen, Anal.
Chim. Acta 319 (1996) 59.
[17] R. Toniolo, N. Comisso, G. Bontempelli, G. Schiavon,
Electroanalysis 8 (1996) 151.
[18] M. Corsegove, G.J. Moddy, J.D.R. Thomas, Analyst 113
(1988) 118.
[19] P.D. Sanchez, P.T. Blanco, J.M.F. Alfarez, Electroanalysis 3
(1991) 281.
[20] G.J. Peterson, Electroanalysis 3 (1991) 741.
[21] M.S. Lin, S.Y. Tham, G.A. Rechnitz, Electroanalysis 2 (1990)
511.
[22] L. Charpentier, N.E. Murr, Analysis 23 (1995) 265.
[23] M.F. Cardosi, Electroanalysis 6 (1994) 89.
[24] J. Wang, A.J. Reviejo, L. Angnes, Electroanalysis 5 (1993) 575.
[25] D. Zulficar, B. Hibbert, P.W. Alexander, Electroanalysis 7
(1995) 722.
[26] S.V. Dzyadevich, V.N. Arkhipova, A.P. Soldatkin, A.V.
El'skaya, A.A. Shul'ga, Anal. Chim. Acta 374 (1998) 11.
[27] A.A. Shulga, S.V. Dzyadevich, A.P. Soldatkin, S.V. Patskovs-
ky, V.I. Strikha, A.V. El'skaya, Biosensors and Bioelectronics
9 (1994) 217.
[28] M.S. De Silva, Yu. Zhang, P.J. Hesketh, G.J. Maclay, S.M.
Gendel, J.R. Steller, Biosensors and Bioelectronics 10 (1995)
669.
[29] J. Vicens, V. Bohmer (Eds.), Calixarenes, a Versatile Class of
Macrocyclic Compounds, Kluwer Academic Publishers,
Dordrecht, 1991.
[30] C.C. Lesnoff, Phthalocyanins: Properties and Applications,
VCH, Weinheim, 1989.
[31] A.V. Nabok, Z.I. Kazantseva, N.V. Lavrik, B.A. Nesterenko,
Int. J. Electronics 78 (1995) 129.
[32] G. Kampfrath, R. Hintche, Anal. Lett. 22 (1989) 2423.
[33] T.A. Sergeyeva, N.V. Lavrik, S.A. Piletsky, A.E. Rachkov,
A.V. El'skaya, Sensors and Actuators B 34 (1996) 283.
[34] J.L. Boitieux, G. Desmet, D. Thomas, Clin. Chem. 25 (1979)
318.
[35] R.L. Lenado, G.A. Rechnitz, Anal. Chem. 45 (1973) 826.
[36] T.M. Swager, M.G. Marcella, Adv. Mater. 6 (1994) 595.
T.A. Sergeyeva et al. / Analytica Chimica Acta 391 (1999) 289±297 297