H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO- 15? James Picker, Ph.D. That is the question.

Upload: jada-hood

Post on 26-Mar-2015

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

TO-15 or not TO-15?

James Picker, Ph.D.

That is the question.

Page 2: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Objective

Used for the analysis of VOCs in soil vapor (gas) to measure health risk

Compare/Contrast

EPA Method 8260B vs. EPA Method TO-15/TO-14A

Page 3: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

OR

Perceived (advertised?) Difference

Page 4: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

OR

Actual difference

Page 5: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Summary EPA 8260 vs. TO-15/14A

SAME IDENTICAL

Data Reduction GC/MS System Sample Preparation

SLIGHTLY DIFFERENT

Page 6: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Similarities

GC/MS Method (SCAN or SIM)

GC with FSCC (20-60m X 0.18-0.32mm id)

LRMS (quadrupole, ion trap, TOF)

*Concentration of compounds on multiadsorbent trap*

Uses Internal Standards/Surrogate(s) for accuracy/QC

Multipoint Calibration ICAL, linearity checks, Daily CCV

Varying sample size permitted

Quant by RFs

Page 7: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Differences Standards and Sample Preparation

EPA 8260B (modified)

*Standards in liquid (MeOH)*

Purge and trapPurged from headspace and H2O

using He or dry purgingWater managementDesorption direct to GC/MS

EPA TO-15/14A

*Standards in Air (N2)*

Vacuum (or Pressure) flow and Trappulled from cylinder, using pump or pressure

Desorption to cryofocus or directly to GC/MS.

Page 8: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Standards in Water or Air

-

Page 9: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Compounds Analyzed equally well by EPA 8260 or TO-15

• BTEX, Trimethylbenzenes, Freons

• PCE, TCE, vinyl chloride, DCEs, 1,2- DCPE, 1,1,1,2- TCA

• 1,1,1-TCA,

Page 10: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Compounds Analyzed with a slight high bias by EPA 8260 versus TO-15

• MTBE, other oxygenates• 1,1,2,2-TCA, 1,2- DCA, 1,1,2-TCA• Naphthalene• rest of common 8260 list

Page 11: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Compounds Analyzed with a high bias by EPA 8260 versus TO-15

Ethanol, TBA1,4 DioxaneAcetone & other ketonesAny water soluble low volatility compound

Page 12: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Differences II Calibration and Range

EPA 8260B (modified)

“Typical” Calibration 500 – 50000 ug/m3 (160-16000 ppbv for benzene)

MDLs 100 ug/m3 (31 ppbv benzene) For 8260 SIM 10 ug/m3 (3 ppbv benzene

EPA TO-15/14A

“Typical” Calibration 3-120 ug/m3 benzene (1-40 ppbv)

MDLs 0.3 ug/m3 benzene( 0.1ppbv)

For TO-15 SIM 0.03 ug/m3 benzene (0.01ppbv)

Page 13: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Calibration RangeMDL to highest standardassumes 20 cc sample

10-2 10-1 1 10 102 103 104 105

ug/m3

TO-15/14A SIM

8260 SCAN

TO-15/14A SCAN

8260 SIM

Normal Range

Extended by dilution

Page 14: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Fail LevelsTCE @ subslab TCE @ 5’bgs

10-2 10-1 1 10 102 103 104 105

ug/m3

TO-15/14A SIM

8260 SCAN

TO-15/14A SCAN

8260 SIM

EPA R9 CA-EPA

Page 15: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

10-2 10-1 1 10 102 103 104 105

ug/m3

TO-15/14A SIM

8260 SCAN

TO-15/14A SCAN

8260 SIM

Fail LevelsBenzene @ subslab Benzene @ 5’bgs

CA EPA EPA R9

Page 16: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Project BL 4FEB05results in ug/L

Sample TO-15

PCE

8260

PCE

TO-15

TCE

8260

TCE

TO-15

c12DCE

8260

c12DCE

TO-15

t12DCE

8260

t12DCE

TO-15

F-113

8260

F-113

SV1 128 88 42 31 <1 <1 <1 <1 1.6 1.1

SV2 399 280 134 100 4.4 3.7 1.7 1.4 3.4 2.7

SV3 95 46 40 31 <1 <1 <1 <1 1.7 1.6

SV4 541 330 75 60 <1 <1 <1 <1 <1 <1

SV5 169 110 70 65 1.5 1.7 <1 <1 3.4 3.1

SV6 234 150 138 110 3.6 3.2 <1 <1 4.8 4.1

SV7 108 54 25 17 <1 <1 <1 <1 3.1 2.1

SV8 250 120 120 74 4.7 3.5 <1 <1 9.9 4.8

AVG

%Diff.52 28 13 19 28

Average % Diff across all compounds = 38%

Page 17: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Project MC 1FEB05results in ppbv

Sample TO-15

PCE

8260

PCE

TO-15

TCE

8260

TCE

TO-15

c12DCE

8260

c12DCE

TO-15

111TCA

8260

111TCA

TO-15

F-113

8260

F-113

SG1 190 250 <16 <25 32 <25 <16 <25 NA NA

SG3 59 83 110 93 <16 <25 40 36 NA NA

SG8 <16 <14 <16 64 <16 <25 540 370 NA NA

SG12 <16 <14 1300 910 91 65 360 261 NA NA

SG13 29 <14 98 81 <16 <25 28 <20 NA NA

Page 18: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Conclusions

1. For most vapor intrusion compounds of interest, methods yield identical results

2. For more water soluble compounds 8260 is within factor of 2. Fine for VI calculations

3. A few compounds may require TO method. Some compounds may require 8260 method

4. On-site analysis assists in decision making, regulatory QA requirements and assurance of tracer compound integrity

Page 19: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Recommendations

Choose either or a combination of methods, depending on compounds and DLs required

Select on-site analysis for decision making, regulatory QA requirements (assurance of tracer compound integrity, purge volume testing, etc.)

Do onsite TO-15 where levels are low Do onsite 8260 where levels are moderate to high Follow with offsite TO-15 for Quality Assurance

Page 20: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Page 21: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P Mobile GeoChemistry Escondido, CA

Page 22: H&P Mobile GeoChemistry Escondido, CA TO-15 or not TO-15? James Picker, Ph.D. That is the question

H&P’s Solutions

Option 1 – Where 8260B Meets Project DQOs:On-site, full QC 8260B fulfilling DTSC/LA-RWQCB requirements with off-site TO-15 confirmation on a sub-set of samples.

Option 2 – Where TO-14/15 is required to meet Project DQOs:On-site screening-level 8260B of all samples with of-site analysis of ALL samples by TO-15.

Option 3 – For larger projects with lower DLs:On site TO-15

All Options:

Fulfill all DTSC soil gas advisory requirements

Allow on-site analysis for leak/tracer compound test

Guarantee sample integrity

Provide real-time data for real-time decisions