Ian Treviranus from HORIBA Scientific (http://www.horiba.com/particle) discuss methods for confirming particle size results. This presentation will be useful for anyone using laser diffraction, dynamic light scattering, or dynamic image analysis instruments. The information will be suitable for any instrument from any manufacturer of these technologies. Topics to be covered include: * Checking sizes smaller than 1 micron * Checking sizes larger than 1 micron * Orthogonal techniques * Filtration * Sieves * Optical and electron microscopy

TRANSCRIPT

Page 1: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Help! How Can I Trust My Particle Size Results?

Ian [email protected]

www.horiba.com/particle

Page 2: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

What we’ll talk aboutGeneral concerns

Orthogonal techniques

Confirming particles < 1 micron

Confirming particles > 1 micron

Q&A

Page 3: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Starting point

Instrument is fully functionalSystem verification webinar TR004

Suitable sample for technologyCounters are classic example

Boot Camp webinars free on websiteParticle size basics, RI selection, sampling

& dispersion advice, method development for maximum precision

Page 4: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

General concerns

Is my distribution really this broad?Light scattering broadens distributionsLS resolution < IA resolution

How to handle multi-mode results?Discuss how to test different peaks todayUse multi-modal report for better metrics

Data matchingDifferent instruments or different

techniques will provide different data

Page 5: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Method validation

Have existing method, does it produce “good” data? Test robustness of methodMeasurement lengthALWAYS! collect multiple samplings

– Reproducibility > challenge than repeatability– Evaluate with COV of chosen metric(s)

Check with orthogonal technique

Page 6: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

What we’ll talk aboutGeneral concerns

Orthogonal techniques

Confirming particles < 1 micron

Confirming particles > 1 micron

Q&A

Page 7: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Featured technologies

LA-950Laser Diffraction

SZ-100Dynamic Light Scattering & Zeta Potential

CAMSIZER & CAMSIZER XTDynamic Image Analysis

PSA300Static Image Analysis

SA-9600Flowing Gas BET Surface Area

Page 8: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

OPTICAL MICROSCOPY / IMAGE ANALYSIS

ELECTROFORMED MESH SIEVES

CENTRIFUGAL SEDIMENTATION

ELECTRICAL CONDUCTIVITY

DYNAMIC LIGHT SCATTERING

ACOUSTIC SPECTROSCOPY

LIGHT OBSCURATION / ELECTRICAL SENSING ZONE

LASER DIFFRACTION

10 nm 100 nm 1 µm 10 µm 100 µm 1 mm

Size Range by Technique

Page 9: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

LA-950: Laser Diffraction

Particle size performance leaderNinth generationUltra durableLowest total cost of ownershipSuspension, emulsion, powder,

paste, gel10 nanometer – 3 mm

Page 10: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Particle size: 0.3 nm – 8 µmZeta potential: -200 – +200 mVMolecular weight: 1x103 – 2x107 DaPatented ultra long-life

graphite electrodesLowest total cost of

ownershipOptional autotitrator

SZ-100: Dynamic Light Scattering

Page 11: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Image AnalysisDynamic:particles flow past camera

Static:particles fixed on slide,stage moves slide

Page 12: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

CAMSIZER Series

High resolution size & shape Intelligent sieve correlationPatented dual captureCAMSIZER30 µm – 30 mmFree-flowing powders

CAMSIZER XT 1 µm – 3 mmCohesive or free flowing

Page 13: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

PSA300

High resolution size & shapeReferee technique for micronized powdersTurnkey, automated image analysis1 µm – 1,000 µmCohesive or free flowing

powdersOptional Powder

Disperser accessory

Page 14: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

What we’ll talk aboutGeneral concerns

Orthogonal techniques

Confirming particles < 1 micron

Confirming particles > 1 micron

Q&A

Page 15: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Likely Technologies

Laser diffraction

Dynamic light scattering

Ultramicroscopy

Electron microscopy

Page 16: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Laser Diffraction < 1 µm Option 1: Filter

Remove large particles Re-measure with laser

diffraction and/or DLS Reliable

measurement?

Page 17: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Laser Diffraction < 1 µm Option 2: Sediment

Create/sample suspension

Let sit overnight or centrifuge

Carefully sample supernatant

Re-measure with laser diffraction or DLS

Reliable measurement?

Small particles will stay suspended by Brownian motion

supernatant + sediment

Page 18: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Laser Diffraction < 1 µm Option 3: Electron Microscopy

Sampling is a concern Can confuse primary particles

and aggregates Very high resolution

Page 19: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Laser Diffraction < 1 µm Option 4: Change RI

Possible secondary scattering

Not as definitive as other options

Use Method Expert software to speed up

1.85 – 0.01i

1.57 – 0.01i

Large particles scatter intensely at low angles

Small particles scatter weakly at high angles

Page 20: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Dynamic Light Scattering < 1 µm Option 1: Laser Diffraction

DLS very sensitive to presence of large particles (scattering d6)

Diffraction less sensitive (d3), more likely to see small particles which may be “hidden” by large

Most advanced diffraction analyzers have detection limit of 10 nm

Need greater sample quantity than DLS (min. dispersed vol. 5 mL)

Need greater sample concentration than DLS

Page 21: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Dynamic Light Scattering < 1 µm Option 2: Electron Microscopy

Sampling is a concern Can confuse primary particles

and aggregates Very high resolution

Page 22: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Dynamic Light Scattering < 1 µm Option 3: Filter

Remove the large particles to eliminate their scattering

Page 23: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

What we’ll talk aboutGeneral concerns

Orthogonal techniques

Confirming particles < 1 micron

Confirming particles > 1 micron

Q&A

Page 24: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Likely Technologies

Laser diffraction

Dynamic light scattering

Image analysis

Sieves

Page 25: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Laser Diffraction > 1 µm Option 1: Static Image Analysis

Quickly verify large particles Provide shape info Automated or manual? Reliable measurement?

Page 26: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Laser Diffraction > 1 µm Option 2: Sampling & dispersion

Large particles are more difficult to sample accurately

Try different sampling techniques like spin riffling (powders) or magnetic stirring (suspensions)

Large particles may actually be agglomerates

Try different surfactants Use internal ultrasound See webinar TR005 on horiba.com

Page 27: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Dynamic Light Scattering > 1 µm Option 1: Laser Diffraction

Not as sensitive to large particles as DLS though doesn’t suffer from sedimentation problem

Most advanced diffraction analyzers have detection limit > 1mm

Need greater sample quantity than DLS (min. dispersed vol. 5 mL)

Need greater sample concentration than DLS

Page 28: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Option 2: Static Image Analysis May have concentration problem

for typical DLS applications Quickly verify large particles Provide shape info Automated or manual? Reliable measurement?

Dynamic Light Scattering > 1 µm

Page 29: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Image Analysis > 1 µm Option 1: Save an image

Straightforward, convincing Manual microscopy subjective Automated IA software will save

only those particles which meet certain size and shape criteria

Page 30: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Image Analysis > 1 µm Option 2: Sampling & dispersion

Large particles are more difficult to sample accurately

Try different sampling techniques like spin riffling (powders)

Large particles may actually be agglomerates

Try different dispersing pressures See webinar TR005 on horiba.com

Page 31: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Expectations

What contributed to the size expectation?Apples to apples?

What is the size of this particle?

Page 32: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Equivalent Volume Distributions

Page 33: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Equivalent Volume Distributions

Page 34: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Equivalent Number Distributions

Page 35: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Equivalent Number Distributions

Page 36: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Same material shown as volume, number and area distribution

VolumeArea

Number12

4

8

10

6

2

0

Per

cent

in C

hann

el

0.34

4.47

8.82

34.2

5

0.58

1.15

2.27

17.3

8Particle Size

Volume DistributionMean = 12.65µmMedian=11.58 µmSA=13467 cm2/cm3

Std Dev.=8.29

Number DistributionMean = 0.38µmMedian=0.30 µm

SA=13467 cm2/cm3

Std Dev.=0.40

Comparing Distribution Bases

Page 37: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Default Bases by TechDynamic light scattering

Laser diffractionSieving

Microscopy (all)Image analysis

Single particle optical sensingElectrical sensing zone

Sedimentation (all)Acoustic spectroscopy

Scattering intensityVolumeMass (vol. if density constant)NumberNumberNumberNumberVolumeVolume

Page 38: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Correlation?

Make sure data from both techniques is compatibleCheck both tools working correctly

Is the spec or expectation fair?

Used sieve mesh, no longer reliable

Page 39: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

x [mm]0.2 0.4 0.6 10

10

20

30

40

50

60

70

80

Q3 [%]

Sample A_BZ_0.2%_xc_min_001.rdfSample A_.ref

Digitale Imaging Sieving

Page 40: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Sieve Correlation

Page 41: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

What we’ll talk aboutGeneral concerns

Orthogonal techniques

Confirming particles < 1 micron

Confirming particles > 1 micron

Q&A

Page 42: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

Danke

Gracias

Большое спасибо

GrazieاُشْكرΣας ευχαριστούμε

감사합니다Obrigado

Tacka dig

谢谢ขอบคุณครบั

ありがとうございました

धन्यवादநன்ற

Page 43: Help! How can I trust my particle size results?

© 2013 HORIBA, Ltd. All rights reserved.

www.horiba.com/particle

Receive news of updatesView application & technical notes (170+), webinars (60+), white papers.

Talk to us, ask [email protected]

Ian TreviranusProduct Line Manager

[email protected]