handling interferences ̶ experience report macherey-nagel...macherey-nagel handling interferences...

44
MACHEREY-NAGEL Handling interferences ̶ Experience report Dr. Christian Prokisch, 11.08.2019 Collaborative Efforts to Improve Environmental Monitoring

Upload: others

Post on 09-Mar-2021

14 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

MAC

HER

EY-N

AGEL

Handling interferences ̶ Experience report

Dr. Christian Prokisch, 11.08.2019Collaborative Efforts to Improve Environmental Monitoring

Page 2: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

Agenda

Company Nutrient Removal P‐Tests

N‐Tests

Tox Tests

Green(er) COD

Page 3: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

CompanyMN Water Analysis

Page 4: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

4

MN today

Company 

• 4th Generation family owned• More than 650 employees• More than 25.000 products• Turnover 115 Mio. €

Subsidiaries Distributors

Page 5: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

5

Business units

Company

1911 1959 1961 1970 1989

Page 6: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

6

Get acceptance for COD tube tests

Company – MN Water AnalysisWhy are we here?

• Market share NANOCOLOR®

• Germany: ca. 40%• US: 0%

• Get experience with US regulations

Page 7: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

7

Company – MN Water Analysis

• NANOCOLOR® VIS II• Revolutionary user

experience with 10 inch HD touch screen display

• Intuitive, icon‐based menu guidance

• Integrated turbidity control

Outstanding photometers

Page 8: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

8

Company – MN Water Analysis

• NANOCOLOR® VARIO 4• Touch screen with intuitive

menu navigation• Lockable protective lids and

safety covers• Calibrated sensors to control

digestion temperature optional

Competitive heating blocks

Page 9: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

Nutrient removal

Page 10: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

10

Nutrient removal

• Nitrogen and phosphorus are the primary causes of cultural eutrophication

• Approximately 25% of all water body impairments are due to nutrient‐related causes

• Oxygen depletion• Algal growth• Ammonia

• More stringent effluent limits lead to • P‐removal• N‐removal

EPA Fact Sheet

Page 11: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

P‐Test

Page 12: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

12

Phosphate acc. to EPA 365.3

P‐Test

PO43‐ + 12 MoO4

2‐ + 

24 H+ + 3 NH4+        

(NH4)3[P(Mo3O10)4] + 

12 H2O 

    Phosphor 

molybdenum blue

Reduction

→→

Page 13: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

13

Dissolved ortho‐Phosphate

P‐Test

O ‐O – P – O‐

O‐

ortho‐Phosphat

Page 14: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

14

Total ortho‐phosphate

P‐Test

Page 15: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

15

Phosphate acc. to EPA 365.3

P‐Test

• For total phosphate, suspended solids contribute to total amount

• Make sure to draw a representative sample

• Carefully homogenize sample

Page 16: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

16

“Report results in mg/L P or mg/L PO4‐P”

P‐Test

Know the difference…

Page 17: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

17

Testing NO3‐ and PO4

3‐ in paralle

P‐Test

Danger of carry‐over

Phosphate test WITHOUT changing the pipette tip

Wrong results

Nitrate tests contain phosphoric acid

→ →

Page 18: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

18

When dealing with e.g. paper industry discharges

P‐Test

Be aware of possible silica interference

+ Silica > 150 mg/L

Phosphate test Wrong results

Page 19: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

N‐Test

Page 20: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

20

N‐Test

• Total N seems to be the target valuefor eutrophication

• Available EPA methods (examples)• NH3 acc. To EPA 350.1• Total Kjeldahl acc. to HACH 10242• NO3

‐ acc. to HACH 10206• NO2

‐ acc. to EPA 353.2• Not available EPA method

• Total N

Total N vs. Parameters 40 CFR Part 136

Page 21: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

21

Total N is already part of an approved test

N‐TestHow to detect Total N

TKN NO3‐Approved N‐methods:

TN NO3‐‐ + NO3

org.N + NH3+ + NO3

‐What is total N?

What you then do

+

HACH TKN method: TKN = TN ‐ NO3‐

Page 22: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

22

Total N testing

N‐Test

Norg

NinorgNO3

‐ Determination

Page 23: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

23

Digestion

N‐Test

Page 24: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

24

Determination

N‐Test

JUST invert!

Page 25: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

25

Do not shake

N‐Test

• Sample may develop turbidities

Page 26: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

26

N‐Test

No!

Filtration possible

Page 27: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

27

Test for Chloride

N‐Test

Yes!• Chloride determination with QUANTOFIX© Chloride• Dilute• Chloride elimination cartridges

Page 28: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

Tox Tests Inhibitors for denitrifaction

Page 29: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

29

Processes

Tox Tests for denitrification

NH4+

NO2‐

NO3‐

N2

Ammonium Oxidation

Nitrite Oxidation

Denitrification (Reduction)

Inhibitors

A‐Tox

N‐Tox

Page 30: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

30

How Tox‐tests work

Tox Tests for denitrification

FeedOxygen

Nothing happens

Nothing happens

Feed

Oxygen

Page 31: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

31

Tox Tests for denitrificationControl and sample

Inhibitors

Control

Inhibited sample

ConsumptionControl: 4 O2

ConsumptionSample: 3 O2

Inhibition = ∆O2 : ConsumptionControl x 100 = [1 : 4] x 100 = 25 %

∆O2 = 1

Page 32: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

32

How the test is done

Tox Tests for denitrification

• Use oxygen probe• Measure

• Control• Sample

• Calculate inhibition from differenceSyringe

Oxygen Probe

Reaction vessel

Page 33: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

33

ISO 9509 vs. BioFix® Tox Tests

Tox Tests for denitrification

• Only for POTW with nutrient removal• Easily detects nitrification‐inhibitors• Lab test, result in 12 minutes

Page 34: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

Green(er) COD?

Page 35: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

35

Green(er) COD?

• Tube tests for chemical oxygendemand

• Principle• Cr2O7

2‐ + “Dirt” 2 Cr3+

• NANOCOLOR® COD HR 1500• NANOCOLOR® COD LR 150

Method summary

Cr2O72‐, 436 nm

e.g. COD 160 

Cr3+, 620 nme.g. COD 1500 

Page 36: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

36

Green(er) COD?

• Cr2O72‐ is actice reagent

• Test is based on oxidation potential• So far not replacable for real samples

• Hg2+ masks Chloride• allows use of Silver as catalyst• Prevents false to high results caused by Cl‐

‐ Oxidation

More detail

Ag+

Cr2O72‐ + “Dirt” + Cl‐ + Hg2+ Cr3+ +HgCl2

Page 37: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

Cr2O72‐ + “Dirt” 2 Cr3+

37

Green(er) COD?Mercury is used to overcome Chloride interference

Ag+

Cr2O72‐ + Cl‐ 2 Cr3+ (false too high result)

Ag+ + Cl‐ AgCl (turbidity)

Cr2O72‐ + 2 Cl‐ + Hg2+ Cr2O7

2‐ +HgCl2

Ag+ + 2Cl‐ + Hg2+ Ag+ + HgCl2

Chloride interference Part I

Chloride interference Part II

Page 38: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

38

Green(er) COD?

• POTW inflow• Expected “Chloride error” at 800 ppm

COD: approx. 3%• Most substances easy to crack

• POTW effluent• Expected “Chloride Error” at 40 ppm

COD: approx. 60%• Ag+ catalyst needed to crack

refractary substances

Hg‐free COD Part I: Chloride Compensation

Use test strip to measure Chloride and substract from result

Cr2O72‐ + Cl‐ + Dirt Cr3+Ag+

Page 39: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

39

Green(er) COD?

• POTW inflow• Filtration not suitable

because loss of particles

• POTW effluent• Filtration no problem (little

particles in effluent)• Ag+ catalyst available to

crack refractary substances

Hg‐free COD Part II: Chloride Removal

Cr2O72‐ + Cl‐ + Dirt Cr3+Ag+

Use cartridge to remove Chloride(involves some kind of filtration)

Page 40: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

• Chloride compensation

• Inflow: reasonably OK

• Outflow: useless

• Cl- teststrip would need improvement

• Chloride removal

• Inflow: useless

• Outflow: reasonably OK (expected)

• Handling would need improvement

40

Hg‐free COD could work – if the demand was there…

Green(er) COD?Result

Page 41: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

Summary

Page 42: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

42

Summary

• P‐Tests• Take care when Nitrate is tested at the

same time• N‐Test

• Total Nitrogen test is almost there• Tox‐Tests for denitrification

• Easy detection of inhibitors the reducePOTW performance

• Green(er) COD• Improved Mercury‐free test could be

available when needed

Page 43: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

MAC

HER

EY-N

AGEL

Thank you for your attention!

Dr. Christian Prokisch I cprokisch@mn‐net.com I +49‐2421‐969‐166

Page 44: Handling interferences ̶ Experience report MACHEREY-NAGEL...MACHEREY-NAGEL Handling interferences Experience report Collaborative Efforts to Improve Environmental Monitoring Dr. Christian

Image credits©© Erwin Wodicka (2, 19‐27), SG‐design (10), Zauberhut (15), Darknightsky (20), Arnd Drifte (21) , oriori (25), sirirak (27), KaeslerMedia (29) , sirirak (40) / Fotolia