Software Installation Guide

HAMILTON-T1

Software version 2.2.10

624476/27  |  2021-03-03

Valid for devices with SNxxxx > SN3000

Page 2: HAMILTON-T1-SW v2.2.10-Inst Guide-PN 624476

2 English | 624476/27

© 2021 Hamilton Medical AG. All rightsreserved. Printed in Switzerland.

No part of this publication may be reproduced,stored in a database or retrieval system, ortransmitted in any form or by any means, elec-tronic, mechanical, or by photocopying, record-ing, or otherwise, without prior written permis-sion of Hamilton Medical AG.

This document may be revised, replaced, ormade obsolete by other documents by Hamil-ton Medical AG at any time and withoutnotice. Ensure that you have the most currentapplicable version of this document; if indoubt, contact the technical support depart-ment of Hamilton Medical AG, Switzerland.While the information set forth herein isbelieved to be accurate, it is not a substitute forthe exercise of professional judgment.

Nothing in this document shall limit or restrictin any way Hamilton Medical AG’s right torevise or otherwise change or modify theequipment (including its software) describedherein, without notice. In the absence of anexpress, written agreement to the contrary,Hamilton Medical AG has no obligation to fur-nish any such revisions, changes, or modifica-tions to the owner or user of the equipment(including software) described herein.

The equipment must be operated, serviced, orupgraded only by trained professionals. Hamil-ton Medical AG’s sole responsibility withrespect to the equipment and its use is asstated in the limited warranty provided in thedevice Operator’s Manual.

Hamilton Medical AG shall not be liable for anyloss, cost, expense, inconvenience, or damagethat may arise out of misuse of the product, orif non-Hamilton Medical AG parts were usedwhen replacing parts, or if serial numbers wereamended, deleted, or removed.

If returning parts to Hamilton Medical AG, besure to use the standard Hamilton Medicalreturned goods authorization (RGA) procedure.Disposal of parts shall follow all local, state,and federal regulation with respect to environ-mental protection.

For all proprietary as well as third-party trade-marks used by Hamilton Medical AG, seewww.hamilton-medical.com/trademarks.Product and/or company names marked with a§ symbol may be the trademarks and/or regis-tered trademarks of their respective owners,including but not limited to Aerogen§, NihonKohden§, Masimo§, and Capnostat§.

Manufacturer

Hamilton Medical AGVia Crusch 8, CH-7402 BonaduzSwitzerlandPhone: (+41) 58 610 10 20Fax: (+41) 58 610 00 [email protected]

Page 3: HAMILTON-T1-SW v2.2.10-Inst Guide-PN 624476

SW update from version 2.x.x to 2.2.10

3Hamilton Medical | HAMILTON-T1 Software Installation Guide

1   SW update from version2.x.x to 2.2.10

Note that the software update fromversion 2.x.x to 2.2.10 is only applicable toHAMILTON-T1 with SNxxxx > SN3000.

CAUTION

• Make sure that the device is con-nected to an AC power source beforeyou do a software update/upgrade.

• Make sure that the mandatory back-up battery or batteries are installed inthe battery compartment and are fullycharged before you do a softwareupdate/upgrade.

• Do not turn off the device during thefirst startup after performing a soft-ware update/upgrade.

• During the first startup, the buzzerwill sound for several minutes.

• Do not turn off the device during thesoftware update. This will help pre-vent possible damage to the embed-ded system module (ESM) board.

NOTICE

• Record the ventilator settings for ref-erence. The HAMILTON-T1 ventilatorsettings are defined in the Operator'sManual Appendix, Configuration.

• Configuration settings generated withsoftware version 2.2.4 or lower arenot compatible with software version2.2.10.

• Configuration settings generated withsoftware versions 2.2.6 to 2.2.10 arecompatible with software version2.2.10.

• Configuration settings generated withsoftware version 3.0.0 or higher arenot compatible with software version2.2.10.

• A software upgrade from softwareversion 2.x.x to 2.2.10 will reset theconfigurations to their defaultsettings.

• A downgrade to a lower softwareversion is not possible.

To update from software version 2.x.x to2.2.10

The following file is required for theinstallation:

1. Touch SW update button to access theupdate page.

2. Insert the USB drive containing theupgrade versions as listed in Table 1.

NOTICE

• The software update version is shownon the display if the USB drive con-tains the correct file. If not, the mes-sage No Update Tarball File is avail-able... is shown on the display.

• Make sure that you select the correctsoftware version if the USB drive con-tains more than one software updatefile.

Page 4: HAMILTON-T1-SW v2.2.10-Inst Guide-PN 624476

SW update from version 2.x.x to 2.2.10

4 English | 624476/27

CAUTION

• During the installation of the softwarefrom the USB drive to the HAMILTON-T1 ventilator, some technical faultsand audible alarms can occur. Thesetechnical faults can be ignored.

• Do not tun off the device during soft-ware update process until the statusmessage shows Update successfullycompleted on the display.

NOTICE

The software installation can take up to10 minutes to complete.

3. Touch Start button to start the SWupdate. This comprises extracting, verify-ing and installing the new software.

NOTICE

A progress bar is shown to indicate thecurrent status of the software update.When the software update is completed,the status message Update successfullycompleted is shown on the display.

4. Restart the HAMILTON-T1 ventilator.

NOTICE

• The firmware update process takesapproximately 2 minutes. The buzzersounds for approximately 45 secondsand cannot be silenced.

• The firmware update is completedafter the device has restarted.

Table 1. Firmware update sequence (for HAMILTON-T1 ventilator with SNxxxx > SN3000)

Approx. timeelapsed after restart(Min:Sec)

Procedure description

0:00 Updating boot loader

0:35 The buzzer sounds for ashort time and the devicerestarts automatically.

1:00 The buzzer sounds contin-uously.

1:45 Startup completed (opera-tional mode).

5. Restart the HAMILTON-T1 ventilator inoperational mode if the device starts up inservice software mode.

6. Restart the HAMILTON-T1 ventilator inthe service software mode.

7. Verify that the software version is now2.2.10.

Steps 8 thru 10 are only applicable toHAMILTON-T1 Mil ventilators (PN 161009)delivered with SW v < v2.2.7

NOTICE

We have introduced new part numbersfor the different blower modules in theTechnical state for SW v2.2.7. This allowsthe software to identify the differentblower module types and blower mod-ules that are NBC compatible and thosethat are not NBC compatible.

Page 5: HAMILTON-T1-SW v2.2.10-Inst Guide-PN 624476

Configure the ventilator

5Hamilton Medical | HAMILTON-T1 Software Installation Guide

8. Change the Part No. of the blowermodule from 161170 to 161300 in theTechnical state: start the ventilator in Ser-vice software mode > touch VentilatorInfo. tab > Technical state tab > Modifytab > Part No. button, then select the cor-responding Part No. for the blowermodule. See Figure 1.

Figure 1. New blower module Part No. selectionin Technical state—applicable to HAMILTON-T1Mil with SW v2.2.10 (Page No. 1302)

9. Touch Save button to save the Part No.for the new blower module.

10. Restart the HAMILTON-T1 ventilator inservice software mode to apply thechanges to the ventilator.

11. Perform all calibrations and tests asdescribed in the HAMILTON-T1 ventilator'sService Manual.

12. Restart the HAMILTON-T1 ventilator inthe operational mode and check the confi-guration settings and if necessary reset orimport the settings. The configurationsettings are defined in the HAMILTON-T1ventilator's Operator's Manual Appendix,Configuration.

Figure 2. SW update from version 2.x.x to2.2.10 (Page No. 5101)

Table 2. File required for the software upgrade

Device Installation file

HAMILTON-T1with SNxxxx >SN3000

T1_EM10A_V2.2.10.tar

2   Configure the ventilator

CAUTION

• The following functions may not beapproved in all countries. It is yourresponsibility to verify that the func-tions which you will activate with theactivation and deactivation keys areapproved in your country.

• Make sure that each configurationfunction(s) are fully described in theoperator‘s manual and will be pro-vided to the customer. Activating thefunction(s) constitutes your accep-tance of our terms and conditions.

Page 6: HAMILTON-T1-SW v2.2.10-Inst Guide-PN 624476

Configure the ventilator

6 English | 624476/27

Activation and deactivation keys

Suctioning Tool

Ask the clinicians whether they are usingopen suctioning or closed suctioning systems.In case the hospital is using open suctioningsystems and would like to benefit from thesuctioning tool enable it by entering the Con-figuration menu (access code 7132) andentering the corresponding code under theOptions menu.

ON: 65DD2FEA1F98DA

OFF: D2E36F2DEF55D2

PSync-PSIMV+

To enable or disable the PSync-PSIMV+ func-tionality, enter the Configuration menu(access code 7132), then enter the corre-sponding code under the Options menu.

ON: 4336568F45E787

OFF: 3792A8D526FD50

Default I:E Ratio 1:2

The ventilator is preconfigured with an I:Eratio of 1:4 for adult/pediatric. To change theI:E ratio, enter the Configuration menu (accesscode 7132), then enter the correspondingcode under the Options menu.

ON: EE19AC76C97BAC

OFF: 5F7841CF79C4B5

O2 consumption

To enable or disable the O2 consumptiondisplay, enter the Configuration menu (accesscode 7132), then enter the correspondingcode under the Options menu.

ON: D31AE249406F78

OFF: 383590FBCEA740

Activation and deactivation keys

ASV 1.1

To enable or disable ASV 1.1, enter the Confi-guration menu (access code 7132), thenenter the corresponding code under theOptions menu.

ON: 5884DA444B4ED5

OFF: 92DB94E3730E8D

When ASV1.1 is enabled, the user can togglebetween the standard ASV and ASV1.1 inthe Configuration menu (access code 7132)under the Modes menu.

Note that ASV 1.1 will be set as default afterthe software upgrade/update.

Adjustable HPO alarm limit

This option allows you to manually adjust theO2 alarm limits when using HPO. To enableor disable the functionality, enter the Configu-ration menu (access code 7132), then enterthe corresponding code under the Optionsmenu.

ON: A747A44F0E05A5

OFF: 6264FCF185EF43

Note that when activating this option , youmust also activate the Extended O2 calibrationtext option.

Extended O2 calibration text

To enable or disable the status informationon the O2 cell calibration, enter the Configu-ration menu (access code 7132), then enterthe corresponding code under the Optionsmenu.

ON: 02F159E368AB2F

OFF: F86FA1A15539C4

Page 7: HAMILTON-T1-SW v2.2.10-Inst Guide-PN 624476

Configure the ventilator

7Hamilton Medical | HAMILTON-T1 Software Installation Guide

Activation and deactivation keys

Vt/IBW monitoring

It is well established that you must closelymonitor the tidal volume per weight, espe-cially for ARDS patients. The HAMILTON-T1ventilator now offers direct monitoring of thetidal volume per IBW (adult/pediatric) or tidalvolume per weight (neonatal). To enable ordisable the functionality, enter the Configura-tion menu (access code 7132), then enter thecorresponding code under the Options menu.

ON: 69D96CEE8D3F05

OFF: 0C94A873583846

O2 sensor deactivation memory

To enable or disable the O2 sensor deactiva-tion memory, enter the Configuration menu(access code 7132), then enter the corre-sponding code under the Options menu.

ON: 306C7986D4CFF2

OFF: 878F5B2A085F90

Cflow (Hi Flow) O2 restricted (Only USA market)

CAUTION

This information is only applicable todevices sold in the USA:

• Hi flow O2 restricted ON setting is amandatory requirement in the USA.

• You must set the Hi flow O2restricted to ON for the HAMILTON-T1ventilator after initial installation or asoftware update/upgrade.

The flow is reduced to a maximum of 60 l/min. Note that there is no restrictionon the blower back pressure. The blowerhas a pressure limit of 50 cmH2O.

ON: 32CC3B4C486D37