Sealed offers are invited for & on behalf of President of India for "Design, fabrication and
installation of 3-axis manipulator for holding Flat panel detector for Real-time radiography
as per tec/rnical specification given below in Description."
S.No.
1.
DISCRIPTION
Technical Specification for XYZ-axis manipulator for Radiography
A high precision automation system has to be designed, fabricated and
installed for holding and positioning of Flat panel detector. This system
consists of following components:
Qty.
1. Detector Holding Fixture: O t no. orThis fixture to hold detector shall be made up of SS sheet of 1 manipulatormm thickness. Plates shall be screwed to form a rectangular
holder with necessary cut outs on the side wall to allow cables
to pass. Front side shall have openable shutter type of
arrangement. Shutter front door shall be made up of lead sand-
witched with SS sheet. Lead sheet of thickness 2 mm shall be
filled into the cavity made between SS sheets. The exact
dimension of the detector will be provided to the selected
vendor with the work order.
Fig.1: A typical Flat panel