gilian air sampling pump field calibration

17
Gilian Air Sampling Pump Gilian Air Sampling Pump Field Calibration Field Calibration A Step-By-Step Guide

Upload: iola-sosa

Post on 02-Jan-2016

72 views

Category:

Documents


1 download

DESCRIPTION

Gilian Air Sampling Pump Field Calibration. A Step-By-Step Guide. Items required. Sampling Pump Collection Media Gilibrator Calibrator. A Field Calibration Consists of Two Parts. The Pre-Calibration (Pre-Cal) is done before the sample is taken - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Gilian Air Sampling Pump  Field Calibration

Gilian Air Sampling Pump Gilian Air Sampling Pump Field CalibrationField Calibration

A Step-By-Step Guide

Page 2: Gilian Air Sampling Pump  Field Calibration

Items requiredItems required

• Sampling Pump• Collection Media• Gilibrator

Calibrator

Page 3: Gilian Air Sampling Pump  Field Calibration

A Field Calibration Consists of TwoA Field Calibration Consists of Two PartsParts

• The Pre-Calibration (Pre-Cal) is done before the sample is taken

• The Post-Calibration (Post Cal) is done after the sample is taken

Page 4: Gilian Air Sampling Pump  Field Calibration

The Pre-CalThe Pre-Cal

• The Pre-Cal is used to set the desired sample flow rate, and to verify it on an air flow calibrator.

Page 5: Gilian Air Sampling Pump  Field Calibration

ConductingConducting the Pre-Calibrationthe Pre-Calibration

Page 6: Gilian Air Sampling Pump  Field Calibration

The Pre-CalThe Pre-Cal

• Set the Flow Rate• Observe the Flow Rate• Record the Flow Rate• An average of 6 to 10 Gilibrator readings is recommended

Page 7: Gilian Air Sampling Pump  Field Calibration

Run The SampleRun The Sample

• Pump is placed on the belt• Sample media is placed on

the lapel (Breathing Zone) and pointed downward

• Record the Start Time• Record the End Time

Page 8: Gilian Air Sampling Pump  Field Calibration

The Post-CalThe Post-Cal

• The Post-Cal is used to obtain the final flow rate and to verify that the flow rate stayed within +/- 5% of the set flow.

Page 9: Gilian Air Sampling Pump  Field Calibration

The Post-CalThe Post-Cal

• Observe the Final Flow Rate• Record the Final Flow Rate• An average of 6 to 10 Gilibrator readings is recommended

Page 10: Gilian Air Sampling Pump  Field Calibration

Average Flow RateAverage Flow Rate

• The Pre-Cal flow rate and the Post-Cal flow rate may be averaged to obtain an average flow rate for the sample.

Page 11: Gilian Air Sampling Pump  Field Calibration

Pre-Cal Flow Rate + Post Cal Flow RatePre-Cal Flow Rate + Post Cal Flow Rate = = Avg Flow RateAvg Flow Rate 2 2

• Determining the Average Flow Rate

Page 12: Gilian Air Sampling Pump  Field Calibration

Cyclone CalibrationCyclone Calibration

• The Gilian 10mm nylon cyclone has only an exit fitting and must be placed inside a calibration jar to accept the Gilibrator.

Page 13: Gilian Air Sampling Pump  Field Calibration

Low Flow ModulesLow Flow Modules

• Calibrating with low flow modules

Page 14: Gilian Air Sampling Pump  Field Calibration

Constant Flow ModulesConstant Flow Modules

• Set pump flow to 1500cc/min• Turn module switch on• Connect sample tubing to pump• Adjust desired flow rate at module• Read flow rate at Gilibrator

Page 15: Gilian Air Sampling Pump  Field Calibration

Multi-Flow ModulesMulti-Flow Modules

• Set pump to 1500 cc/min• Turn module switch on• Connect sample tubing to pump• Adjust flow rate at holder• Adjust each tube twice• Read the flow rates

Page 16: Gilian Air Sampling Pump  Field Calibration

Further InformationFurther Information

• For detailed information on any of the Gilian products, please consult the product operation manuals.

• Manuals may be accessed through our web site at www.sensidyne.com

Page 17: Gilian Air Sampling Pump  Field Calibration

Thank YouThank You

• Sensidyne, LP.• Clearwater, FL, USA

• www.sensidyne.com

• Fax 727-539-0550

• Phone 727-530-3602 & 1-800-451-9444