Power, Precision, PerformanceOEC® 9900 Elite

gehealthcare.com

Maneuverability and Positioning

• OEC9900platformshave31”offreespaceinbetweentheX-raytubeandtheimagereceptortoprovidealargeimag-ingspace.

• TheC-armdepthis26”ordeepertoprovidealargeimag-ingspaceandC-armclearancearoundthepatientandtheimagingtable.

• TheC-armmodelsprovideaminimumof115°C-armorbitalrotation,90°under-scanand25°over-scan.TheSuper-Cmodelincreasestheover-scantoamaximumof55°.

• ThesystemmechanicsallowtheusertoreversetheX-raytubeandtheI.I.positionsandalsoprovideC-armunder-scanandover-scancapabilities.

• TheC-armisabletorotate±180°toallowtheimagingchaintoaccomplishangledprojections.

• Thesystemhasatleast18”ofverticalC-armtravelcapa-bilitytoadjusttheimagingchainheight.

• TheC-armprovidesside-to-side(wig-wag)andthehorizon-taltravelmovementstoallowpanningduringimaging.

• ThesystemisequippedwithcablepushersontheC-arm wheelssothatthe C-arm can more easily move over the cables.

• ThesystemcanbeupgradedtoprovidemotorizedC-armmovementsinbothorbitalandlateraldirectionsat9°persecond.

Power to Perform

• Thegeneratorisa60kHzhighfrequencyinverter.

• Theoutputpowerratingofthegeneratoris15kW.

• Thesystemoperatesatfullcapacityoutofastandardwalloutleton120VoltsAC,15amps.

• Thegeneratormeetsthefollowingpowerrequirements:-FluoroscopickVprange:40-120kVp-FluoroscopicmArange:0.2-10mA-RadiographickVprange:50–120kVp-RadiographicmArange:upto75mA

• Thegeneratoriscapableofprovidingaboostorahighdosefluoroscopicexposureatupto20mA.

• Thegeneratoriscapableofprovidingpulsedfluoroscopyat1,2,4,or8pulsespersecond.

• ThegeneratorisdesignedtoprovideahighmAdigitalspotordigitalradiographymodewithupto75mA.

• Thegeneratoriscapableofpulsingat1,2,4,or8pulsespersecondinnormalfluoroscopymodetoreduceradiationexposureifneeded.

• Thegeneratoriscapableofpulsingat30pulsespersecondinDigitalCinePulseModetoreduceimaginglagcausedbypatientorC-armmovement.

• ThemaximumX-raytubecurrentinradiographicmodeis75mAfromstandard120VoltsAC,15ampoutlet.

• ThemAsrangeinradiographicmodeisbetween1to300mAs.

It’s All About Imaging

The Tube . . .

• TheX-raytubeusesarotatinganodeX-raytube.

• TheX-raytubehasdualfocalspotssizesof0.3mm&0.6mmnominal.

• Anodeheatstoragecapacityis300,000H.U.

• Theanodecoolingrateis85,000H.U./min.

• Anodetemperatureiscomputermonitoredforprotec-tion.Anon-screenwarningisprovidedbeforetheanodereachesitsmaximumheatstoragecapacity.

• Thetubehousingheatstoragecapacityis1.6millionH.U.

Case 1 Case 2 Case 3 Case 4 Case 5

Wait

Time

Wait

Time

Wait

Time

Wait

Time

• ProprietaryheatpipecoolingtechnologyandaductedfancoolingsystemimproveefficiencyofX-raytubeheatmanagementsothesystemcanmeettherequirementsofhigherpowerinterventions.

• Protectionisprovidedfortubeoverload.

The Image Intensifer . . .

• Thesystemhasa9”tri-modeimageintensifier.

• Theminimumcentralresolutionatthemonitorinthe9”modeis2.2linepairs/mm.

• Thesystemhasa12”tri-modeimageintensifieroptionforvascularapplications.

• Theminimumcentralresolutionatthemonitorin12”modeis1.6linepairs/mm.

• Theimageintensifierisequippedwithacarbonfibergridwith10:1gridratioand150lines/inch.

Display Details

• Thesystemisequippedwithahighresolu-tion1Kx1KCCDcamera.

• Thecameragainandirisarecomputercontrolled.

• Thesystemisequippedwithan automaticABStrackingsystemtooptimizeimagequality.

• Thesystemisequippedwithtwo19”LCDhighresolutionmonitors.

• Thesystemhasanarticulatingmonitorarmcapableofmoving22”inthehorizontaldirectionanda7°up/10°downdegreetiltalongtheverticalplane.Monitorsareviewablefromallfoursidesoftheworkstation.

• Thesystemprovidesanambientroomlightsensortoauto-maticallyadjustthemonitorbrightnessforoptimumimagedisplay.

• Thesystemhasatouchscreenuserinterface.

• ThesystemhasareferencemonitorontheC-armwhich displaysthelivefluoroimage. (Not available on the motorized system.)

Image Management and More

• Thesystemprovidesasimplemethodtoinputpatientinformationintothesystemcomputersothateachimageisproperlyannotated.

• Thesystemallowstheusertoentermultiplepatientsintothesystemviaaworklist.

• Thesystemprovidesalastimageholdcapabilitywherethelastimageisdisplayedontheactivemonitoraftertheterminationofanexposure.

• ThesystemoffersDynamicRangeManagement(DRM)-PrecisionImagingandPre-setImagingProfiles.

• ThesystemisequippedwithabacklitX-raycontrolpanelthatallowstechnologiststooperatethesysteminadimmedoperatingroom.

• Thesystemallowstheusertochangetheimageorienta-tiononthedisplayscreenduringaliveexposureorusingthelastimagehold.Thosefunctionsincludeimagerota-tion,lefttorightandtoptobottomimagereversals.

• ThesystemisequippedwithaniriscollimatorandshutterstolimittheimagingfieldandreducetheX-rayexposureareatoimproveimagequality.

• Thesystemprovidesanon-screencollimatorpositionindi-catorwhichcanbeappliedonthelastimageholdsothattheusercanadjustcollimationwithoutX-rayexposure.

• Thesystemallowstheusertochangetheimagefieldofviewinreal-timetomagnifyimagesandimproveimageresolutionforbettervisualizationofsmallanatomicalparts.

• Thesystemprovidesameanstomonitoraccumulatedexposuretimeduringaprocedureandanalarmtowarntheuseroflongexposuretime.

• Thesystemprovidesmeanstorecordanddisplayradiationexposuredoseinformationforaprocedure.

• Thesystemautomaticallyselectsproperimagingtech-nique,kVpandmA,duringimaging,butalsoprovidestheuserwiththeabilitytooverridethesesettingsmanually.

• Thesystemallowstheusertoreviewadynamicimagingrunframe-by-frameandtosetcuesintherun.

• Thesystemprovidesdigitalsubtractionandroadmap imagingmodesforvascularapplications.Theseimaging modescanbeaccessedandoperatedbytheuser.

• Thesystemprovidespixelshift,landmark,peakopacifica-tionandre-maskingcapabilitiesinthevascularmode.

• Thesystemprovidesintegratedhighqualityhardcopyfeaturestoathermalprinter,laserqualityfilmprinter,orinterfacetoalasercamera.

• ThesystemprovidesaDICOM3.0interfacecapabilitythatcanbeconnectedtothehospital’snetworksystemtotransferimagesforimagearchiveandprintpurposes.

• Thesystem’sDICOM3.0interfaceprovidesstorageclass,printclassandquery/retrieveworklistfeatures.

• ThesystemallowstheoptionforwirelessDICOMwhichuti-lizesaWirelessLAN802.11b/g/n,10/100MBdevice,runsacustomembeddedWindowsimplementation,andcurrentlysupportsWEP-64,WEP-128,WPA-PSK(TKIP),WPA2-PSK(AES-CCMP)andPEAPwithEAP-MS-CHAPV2securitymeth-ods.

Training Advantages

• Thepurchaseofthisproductincludesin-housetrainingofstafftechnologistsbyanapplicationsspecialist(certifiedtechnologist)fromthemanufacturer.

• Thetrainingincludescontinuingeducationcreditsforstafftechnologistswhocompletethecoursework.

LR-990xxx-01

©2014GeneralElectricCompany–Allrightsreserved.

GeneralElectricCompanyreservestherighttomakechangesinspecificationsandfeaturesshownherein,ordiscontinuetheproductdescribedatanytimewithoutnoticeorobligation.ContactyourGERepresentativeforthemostcurrentinformation.

GE,GEMonogramandOEC9900ElitearetrademarksofGeneralElectricCompany.

GEOECMedicalSystems,Inc.

GE Healthcare, Surgery — Americas: Phone801-328-9300Fax801-328-4300

GE Healthcare — Europe: Paris, FranceFax33-1-30-70-94-35

GE Healthcare — Asia:Tokyo,Japan—Fax:+81-452-85-5490HongKong—Fax:+852-2559-3588

• Theimageprocessingmatrixis1,000x1,000x16bit.

• Thesystemiscapableofprovidingdigitalcinemodesateitherfifteen(15)orthirty(30)pulsespersecond.

• Thesystemprovidesnoisereductioncontroltoimproveim-agequality.

• Real-timeautomaticbrightnessandcontrastcontrolareprovidedtooptimizethedisplayedimage.

• Thesystemprovidesareal-timeandpostprocessingedgeenhancementcapabilitytoimprovebonesandvasculardetails.

• Anelectroniczoomfunctionisprovidedtomagnifyacap-turedimageforimprovedvisualization.

• Thesystemprovidesanautomaticsavefunctionsothatthelastimagetakenisautomaticallysavedtotheharddisk.

• Thesystemprovidesmeanstoputmarkersorcommentsonanimage.

• Thesystemprovidesasimplemethodtoretrievestoredimages(staticanddynamic),displayacollageofstoredim-ages,andallowtheusertoreviewstoredimagesatalatertime.

• Thesystemiscapableofsaving63-1000staticimagestotheinternalharddisk,dependingontheproductconfigura-tion.Storedimagescanberetrievedatalatertime.

• ThesystemhasanintegratedCD/DVDburnerandallowssavinginDICOMorOECformat.

• ThesystemhasaUSB2.0massstoragewrite-onlycapa-bilitytosaveimagesin.jpg/.bmp/.avefileformats.USBstoragedevicesmustbeunencryptedandunprotected.

• Thesystemiscapableofstoringangiographicruns(dy-namicimages)totheinternalharddiskatvariousframerates.Themaximumframerateis30framespersecond.

• Thesystemprovidesdynamicimagestorageofupto10minutesat30framespersecond.

• Thesystemprovidesanautomaticplaybackcapabilitysoimagescanbeplayedbackimmediatelyafterthecomple-tionofadynamicimagingrun.