Page 1: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-(MS) methods

Pim Leonards

Page 2: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Outline

• Principles of GCxGC

• Selection of detectors

• PCB, PBDE analysis

• Other examples

Page 3: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Injector Detector

Modulator

1st column 2nd column

•Selection of modulator •6 different modulators:

•SWEEPER •LMCS •Quad N2(l) jet •Dual CO2 jet •Loop CO2

•CFT

Selection of proper column combination

Principles of GCxGC

Page 4: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Selection 1st and 2nd dimension GC columns

• 1st dimension

– Apolar type phases:

DB-1, DB-5, CP-Sil 8 type of columns

2nd dimension

-Polar type and phase selective phases:

DB-XLB (proprietary) fused-silica

007-65HT (65% phenyl-methylpolysiloxane)

LC-50 (50% liquid crystalline-methylpolysiloxane)

Page 5: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

How does it work? Cryogenic modulator

Page 6: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

How does GCxGC work ?

Page 7: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

How does GCxGC work ?

Page 8: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

1

1

0 s 10 s

How does GCxGC work ?

Separation on the 2nd dimension column

Page 9: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

How does GCxGC work ?

1

2 3

4

5

6

1

6

0s 10 s

Page 10: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Principle and contour plots

0 s 10 s

43

46 min

43 46

10 s

0

1

6

0 10 s

1st

2nd

2nd

1st

Page 11: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:
Page 12: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:
Page 13: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Contour plot

Page 14: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Detectors for GCxGC systems

• Fast scanning detectors

• Acquisition rate >20 Hz

• Low dead volume detectors

Commercial equipment

• GCxGC-FID

• GCxGC-ECD

• GCxGC-qMS

• GCxGC-ToF-MS

• GCxGC-AED not commercial available

Page 15: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-μECD

Page 16: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

0

1

2

3

4

5

6

7

8

41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73

0

1

2

3

4

5

6

7

8

9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41

0

1

2

3

4

5

6

7

8

9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41

0

1

2

3

4

5

6

7

8

9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41

0

1

2

3

4

5

6

7

8

9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41

0

1

2

3

4

5

6

7

8

41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73

0

1

2

3

4

5

6

7

8

41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73

0

1

2

3

4

5

6

7

8

41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73

0

1

2

3

4

5

6

7

8

41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73

0

1

2

3

4

5

6

7

8

10 15 20 25 30 35 40

2n

d d

imen

sio

n r

eten

tio

n t

ime

(s)

1st dimension retention time (min)

0

1

2

3

4

5

6

7

8

45 50 55 60 65 70

155

105

126

85

114 106

127

116

124

99

104

103

121

102 100

101 120

125

88

109

98,119

40,77

81

55

78

42,66

58,79

74

47

67

80

49

68

71

46,48

72 6F-47

62

51

53

69,73

50

22,37

20,35

38

30

27 19

18

34

32

26

29

36

17

39

16,33

28

10

4

9

14

12,13

15 11

8

2

3

115

76

7

6

25,31

97,118

108,123

86 87

4MeO-49

4OH-42

3F-100

6F-66 4’F-69

2OH-28

BB52

BB49

4’F-27

BB15

1 4’F-25 3’F-28

BB101

6OH-47

4OH-49

154

me-TBBP-A

BB153

6OH-99

144 161

168

153

TBBP-A

139

131,158,140

141

160

142

159 167

HBCD

138

166

156

184

128

BB169

183

182

185

192

191

181

173,190 204

4’,6F-199 2,4’F-198

198

203 205

208 207

206

4’F-208

0

1

2

3

4

5

6

7

8

41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73

209

80

4’F-160

3F-119

6MeO-47

203

198

183 181

PCAs

HBCD

138

139

153

TBBP-A

144

154

85

116

99 101

100

4’F-27

27 BB153

154

me-TBBP-A

GCxGC-μECD 125 PBDE’s DB-1x007-65HT

Korytar et al., J Chrom A, 1100 (2005) 200–207

Page 17: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

PCDD/Fs

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

PCBs

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

PBBs

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

PBDEs

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

PCDEs

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

OCPs

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

Toxaphenes

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

PCNs

0

1

2

3

4

5

6

7

8

9

10

11

10 20 30 40 50 60 70

PCDTs

PCAs

GCxGC-μECD various contaminants

20 30 40 50 60 70 0

1

2

3

4

5

6

7

8

10

9

10

11

2n

d d

imen

sio

n r

eten

tio

n t

ime

(sec

)

1st dimension retention time (min)

Korytar et al., J. Chromatogr. A, 2005. 1086, 29-44

DB-1 LC-50

Page 18: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-μECD different column combination

20 30 40 50 60 70 0

1

2

3

4

5

6

7

8

10

9

10

0

1

2

3

4

5

6

7

8

9

10

10 15 20 25 30 35 40 45 50 55 60 65 70 75 80

PBDEs

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

PCBs

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

PCNs

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

PCDTs

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

PCDEs

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

PBBs

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

PCDD/Fs

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

Toxaphene

2n

d d

imen

sio

n r

eten

tio

n t

ime

(sec

)

1st dimension retention time (min)

0

1

2

3

4

5

6

7

8

9

10

10 20 30 40 50 60 70 80

OCPs

PCAs

DB-1 007-65HT

Korytar et al., 2005. J. Chromatogr. A, 1086, 29-44

Page 19: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-μECD dust sample PBDEs

Polychlorinated alkanes P. Korytar et al. / J. Chromatogr. A 1086 (2005) 29–44

DB-1×007-65HT

Page 20: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Background matrix separation from dioxins

4F1 4D1 5F1

5F2

5D1

6F1

6F2

6F3

6D1

6D2

6D3

7F1

6F4

7D1

7F2

OCDD

OCDF

1st dimension retention time [min]

2nd d

imen

sion r

eten

tion t

ime

[s]

GCxGC-ECD, DB-XLB x LC-50

P. Korytar et al. / J. Chromatogr. A 1100 (2005) 200–207

Background matrix

Page 21: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-MS

Page 22: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC with qMS

• Requirements

fast acquisition

spectral quality: no skewing

Page 23: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Modulator control

Single quadrupole MS

Dual-jet CO2 modulator

GC

Set-up

Page 24: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Quadrupole MS

Rf setup time: 10.4 ms

Scan rate: 10000 amu/s

Scan width (amu): @ 50Hz: 96

@ 33 Hz: 195

@ 25 Hz: 296

High acquisition rates • short Rf setup time • fast scanning

Page 25: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

No skewing mass ratio plot (m/z 330/326)

0

0.2

0.4

0.6

0.8

1

0 2 4 6 8 10

Data point number

Mas

s R

atio

Shimadzu qMS

Page 26: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Quality of spectra

P. Korytar et al. / J. Chromatogr. A 1067 (2005) 255–264

Page 27: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Mass range scanned

P. Korytar et al. / J. Chromatogr. A 1067 (2005) 255–264 259

Page 28: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-qMS

• Fast enough

• Selective enough – separation of coeluters

• Limited scanning range (300 Da)

• Not sensitive enough in EI – NCI required

Page 29: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Eel extract GCxGC-qMS PBDE analysis

P. Korytar et al. / J. Chromatogr. A 1067 (2005) 255–264

Page 30: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC ToFMS

• Fast scan speed: 100-500 Hz

• No mass skewing

• Automated search

Page 31: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-ToF-MS, Tern egg Western Scheldt

PCBs

PCAs

Brominated compounds

1st dimension retention time (min)

2n

d d

imen

sio

n r

eten

tio

n t

ime

(sec

)

Page 32: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Screening chemicals in household dust GCxGC-ToFMS of an hexane extract of dust

Hilton et al., 2010. J. Chrom A, 1217 (2010) 6851–6856

Page 33: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Chlorinated and brominated compounds in house dust

• >10,000 peaks detected

• 145 compounds contain chlorine or bromine

Hilton et al., 2010. J. Chrom A, 1217 (2010) 6851–6856

Cl and Br compounds

Page 34: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Other compounds identified in household dust

PAHs Phthalates

Hilton et al., 2010. J. Chrom A, 1217 (2010) 6851–6856

Page 35: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

GCxGC-AED oil

Sulphur

Carbon

van Stee et al. / J. Chromatogr. A 1019 (2003) 89–99

Page 36: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Bioassay

Bioassay

Bioassay

Effect-directed analysis (EDA)

x Y

x

x

x x

Y

Y

Y

complex mixture of compounds

x x Y

1st Fractionation

2nd Fractionation

Chemical identification

Bioassay confirmation

Page 37: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Combined sample treatment scheme

chemical and bioassay analysis

Houtman et al. 2006 •Sediment sample from a harbour •Bioassays

•Dioxin like compounds (DR-CALUX) •Estrogenic compounds (ER-CALUX)

•Chemical identification •GC-MS •GCxGC-ToFMS

Houtman et al., 2006, Chemosphere 65, 2244–2252

Page 38: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Bioassay activity of the sediment

Houtman et al., 2006, Chemosphere 65, 2244–2252

Page 39: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Active fraction GCxGC-ToFMS

Houtman et al., 2006, Chemosphere 65, 2244–2252

Fraction 4-B Fraction 4-B

Fraction 5-B Fraction 2

PAHs

Me-PAHs

PAHs Hormones

Page 40: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Identified compounds and explained bioassay activity

• 76% of the estrogenic activity was explained by 17α-, 17β-estradiol and estrone

• 38% of dioxin-like activity explained by PAHs

Page 41: GCxGC-(MS) methods · PDF filespectral quality: no skewing . Modulator control Single quadrupole MS Dual-jet CO 2 modulator GC Set-up . Quadrupole MS R f setup time: 10.4 ms Scan rate:

Summary

• GCxGC is a powerful separation technique

• Separation of interfering compounds and matrix

• GCxGC is especially suitable for the separation of complex mixtures

• GCxGC combined with mass spectrometry can be used for the identification of “new” flame retardants