Page 1: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

GC-HR-ToFInstrument overview

7/16/2015JDC

Page 2: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Instrument Systems

Page 3: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Windows computer

• Primary user interface to instrument• Run QNX virtual machine – interface to QNX system• Control TPS via web interface (TCP/IP)• Collect ion data from MCP via preamp and ADC

(ADQ 1600)

Page 4: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Mass flow controllers• Ion source N2 (500 sccm FS)• Ion gas CF3OOCF3/N2 (5 sccm FS)• Dilution N2 (3 slm FS)• Ambient zero (1 slm FS)• Two types: digital STEC, analog STEC/MKS both powered

by +/-15V, digital are set and read via RS-485 which can be daisy-chained from single serial port. Analog are set by DAC-controlled analog voltage, and readback is recorded by reading an analog voltage via ADC.

Page 5: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Pressure controller• STEC digital pressure controller to control pressure of

gas in the cal system to ~2 bar.• ~500 sccm maximum flow rate• Pressure readback digitally via RS485 (same chain as

MFCs)• Pressure set with hardwired analog voltage stepped

down from +15V power source, so the setpoint is always applied with there is power to the controller.

Page 6: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Hot Calibration system• On the flight instrument we’ve used a eutectic metal controlled to a

temperature just above the melting point (~48C) in an insulated system with a large heat capacity, so that temperature is maintained for ~24hrs after power is turned off.

• For this system which will not fly on airplane, power should be on all the time, thus we don’t need the metal bath.

• Suggest designing an insulated oven which can hold ~4-5 permeation tubes, with temperature control from (~amb-70C) using heater, RTD and Watlow temperature controller. We now have driver for PM6 digital watlow which allows setpoint to be set and read via RS485 (separate port from STEC)

• Traditional calibrants have been formic and acetic acids, and HNO3. Hope not to use HNO3 in future, but rather the hydroxy nitrate of propene.

Page 7: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Pressure gauges• MKS 750B model – compact absolute (capacitance)

sensors• 1000 torr FS (flow-tube)• 100 torr FS (flow-tube)• 100 torr FS (flowtube pump)• These are powered by +24VDC, give analog voltage output

which is read by ADC. To improve stability these have been insulated and heated to above ambient ~40C using heater and heaterstat. This is particularly important for the 100 torr flowtube gauge used to control the ambient flowrate.

• Differential pressure gauge (MKS) used with pitot tube to measure flow velocity in the inlet.

Page 8: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Athena III SBC (http://www.diamondsystems.com/products/athenaiii)• Single board computer with serial ports (4), digital IO

(24),analog in (16), analog out (4).• We have put this into the Pandora enclosure system for

protection• Powered by +5VDC• Seamless communication with primary Hercules board

via TCP/IP

Page 9: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Athena BB board

• Signal routing/connector board for Athena I/O• Combines analog IO, digital IO, and power into

connectors for individual devices.

Page 10: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Ion reagent gas bottle• On aircraft instrumentation this had to be enclosed

inside a pressure vessel vented to the outside of the airplane. For this instrument this is not nessisary, but mounting for 3 bottles inside this box will be good, then these will not need to be switched out over a deployment.• Ultraclean regulator for CF3OOCF3 bottles• CF3OOCF3 shutoff valve• Electronic pressure gauges (high and low) for regulator• External mounting of bicarb filter for the CF3OOCF3 line

so that it can be easily changed

Page 11: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Manual shut-off valve for N2 upstream of MFCs

Page 12: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Gas Handling/Cal box

• Valve driver board: 5-channel cooldrive board to switch the Nresearch valves on/off• Power +5V, +24V• Signals +5V digital signals, note Athena digital lines are

+3.3V, so these need to be converted to +5V• These are to control:

• Zero valve• Cal – bypass valve• Cal – scrub valve• Individual calibration gasses if needed

Page 13: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

GC-Control Box

• Mass flow controllers• Dilution flow• Bypass flow• ?• STEC digital MFCs

Page 14: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

GC-Control Box

• Athena III SBC (same as described above)

Page 15: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

GC-Control Box

• Watlow temperature controllers• PM6 – single setpoint control (optional pre-trap, or

transfer line)• F2 – ramping controller (main column ramping control)

Page 16: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

GC-Control Box

• Valve driver board (cooldrive):• CO2 cooling valve• Nresearch Teflon valves (collection / elution, others?)

Page 17: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

GC-Control Box

• Temperature sensors:• Thermocouples? Perhaps we should move to RTD?

Page 18: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

GC:

• Column• Oven

• Resistive Heater• CO2 cooler• Temp sensors

• Valve to switch between collection and elution• Suck back valve or MFC to capture gas during collection• Pump – what is needed?

• Bypass flow• Suck/backflow

Page 19: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

Analog digital converter:

• SP devices ADQ 1600 ~ $24,000.00• Power: 12V• Input voltage range: +0.2V to -2.0V: Do not exceed.• Speed: 1.6 GSPS• 14-bit• Control and data transfer by ToFWerk DACRecorder via

USB signals

Page 20: GC-HR-ToF Instrument overview 7/16/2015 JDC. Instrument Systems

ToF Power Supply - bipolar (Tofwerk)• Lens voltages• Pulser voltages• Reflecton voltages• MCP voltages• Power: 120VAC/60Hz• Ethernet control