fs-s00138 rev d - spi lasers

4
482.0 177.0 101.6 37.3 465.1 8.5 LED INDICATORS RED - FAULT YELLOW - EMISSION INDICATOR WHITE - POWER EMERGENCY STOP RESET BUTTON KEY SWITCH 'X' (SEE ORDER CODE DATA BOX) 3.0 THICK PANEL 95 OVER MAINS TERMINAL OPTION & SOLENOID CABLE SEE NOTE 1 FOR UNIT MASS 170.5 4.0 GENERAL NOTES: 1. FOR ESTIMATED MASS SEE ORDER CODE DATA BOX. 2. 3D MODEL FILES ARE AVAILABLE ON REQUEST. ENVIRONMENTAL NOTES, COMPLETE LASER ASSEMBLY TO BE INSTALLED 1. IN AN ENCLOSURE. CONDITIONS WITHIN THE ENCLOSURE (HUMIDITY 2. AND TEMPERATURE) ARE TO BE MAINTAINED AT A LEVEL THAT PROVIDES A NON-CONDENSING ENVIRONMENT WHEN USED WITH THE APPROPRIATE COOLANT TEMPERATURE, ie DEW POINT WITHIN ENCLOSURE TO BE MAINTAINED AT A TEMPERATURE LOWER THAN THE COOLANT SET POINT. ELECTRICAL SUPPLY & CONNECTION DETAILS, SEE SHEET 3 COOLANT REQUIREMENTS, COOLANT HOSE RECOMMENDATION SMC TYPE TU1208 POLYURETHANE. JOHN GUEST TYPE PE-1209-0100M-* LLDPE MAXIMUM PRESSURE 6 BAR WATER TEMPERATURE 18 - 30 C FLOW RATE 10 l/min/kW ISSUE ECR No. DATE REVISIONS D C18261 - UPDATED MANIFOLD DETAILS & SHEET 3 ELECTRICAL CABLE ATTACHMENT 23/01/2018 C R6039 - ENVIRONMENTAL NOTE 1 - DELETED 'IP52 RATED'. DELETED SERVICE NOTE 18/10/2016 B R6032 - ADDED 1.5kW LENGTH, MASS & ELECTRICAL REQUIREMENTS 17/10/2016 THE LENGHTH OF THE CASE, DENOTED 'X' CAN BE FOUND FROM THE RATED POWER, THE 3rd TO 6th CHARACTERS IN THE ORDER CODE OF THE QUBE FIBRE LASER, AND THE 17th CHARACTER IN THE ORDER CODE SHOWN IN BOLD IN THE REPRESENTATIVE ORDER CODE BELOW SP-1000-C-W-020-10-PIQ-012-001-001 RATED POWER W 17th CHARACTER LENGTH X mm ESTIMATED MASS kg ESTIMATED MASS WITH TRANSPORT FRAME kg COMMENT 300 - 1000 0 681 60 63 MEDIUM CASE 1000 B 832 80 83 LONG CASE 1500 0 832 80 83 LONG CASE 5 A B 3 4 2 C D E MATERIAL TITLE DWG. No 1 A B C D E F FINISH DRAWN SCALE DATE G DWG. No SIZE A1 5 3 4 2 F 1 G FS-S00138 FS-S00138 2:1 ALL OVER UOS 1.6 SHT 1 OF 4 QUBE CASED, 500W TO 1.5kW INSTALLATION DWG SEE ABOVE 07/09/2016 Bob Stevens REMOVE ALL BURRS AND SHARP EDGES THIS DRAWING IS THE PROPERTY OF SPI LASERS UK LTD. IT IS CONFIDENTIAL AND MUST NOT BE COPIED OR REPRODUCED WITHOUT PRIOR PERMISSION IN WRITING. COPYRIGHT C 2016 THIRD ANGLE PROJECTION DIMENSIONS IN MILLIMETRES UNLESS OTHERWISE STATED ALL THREADS TO BE ISO COARSE SERIES 6H OR 6g REMOVE SHARP EDGES AND BURRS DO NOT SCALE THE DRAWING. MATERIAL TOLERANCES UNLESS OTHERWISE STATED X. +/- 0.5 X.X +/- 0.10 X.XX +/- 0.05 ANGLES +/- 0.25 DIMENSIONS APPLY AFTER SURFACE TREATMENT SPI LASERS UK LTD COSFORD LANE, SWIFT VALLEY, RUGBY, WARWICKSHIRE CV21 1QN, ENGLAND, Tel +44 (0) 1788 570321

Upload: others

Post on 18-Dec-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

482.0

177

.0

101

.6

37.

3

465.1 8.5

LED INDICATORSRED - FAULTYELLOW - EMISSION INDICATORWHITE - POWER

EMERGENCY STOP

RESET BUTTONKEY SWITCH

'X' (

SEE

ORD

ER C

OD

E D

ATA

BO

X)

3.0

THI

CK

PAN

EL

95

OV

ER M

AIN

S TE

RMIN

AL

OPT

ION

& S

OLE

NO

IDC

ABL

E

SEE NOTE 1 FOR UNIT MASS

170.5

4.0

GENERAL NOTES:1. FOR ESTIMATED MASS SEE ORDER CODE DATA BOX.2. 3D MODEL FILES ARE AVAILABLE ON REQUEST.

ENVIRONMENTAL NOTES,COMPLETE LASER ASSEMBLY TO BE INSTALLED1.

IN AN ENCLOSURE.CONDITIONS WITHIN THE ENCLOSURE (HUMIDITY2.

AND TEMPERATURE) ARE TO BE MAINTAINED AT A LEVEL THAT PROVIDES A NON-CONDENSING ENVIRONMENT WHEN USED WITH THE APPROPRIATE COOLANT TEMPERATURE, ie DEW POINT WITHIN ENCLOSURE TO BE MAINTAINED AT A TEMPERATURE LOWER THAN THE COOLANT SET POINT.

ELECTRICAL SUPPLY & CONNECTION DETAILS, SEE SHEET 3

COOLANT REQUIREMENTS, COOLANT HOSE RECOMMENDATION SMC TYPE TU1208 POLYURETHANE. JOHN GUEST TYPE PE-1209-0100M-* LLDPE MAXIMUM PRESSURE 6 BAR WATER TEMPERATURE 18 - 30 C FLOW RATE 10 l/min/kW

ISSUE ECR No. DATE

REVISIONS

D C18261 - UPDATED MANIFOLD DETAILS & SHEET 3 ELECTRICAL CABLE ATTACHMENT 23/01/2018

C R6039 - ENVIRONMENTAL NOTE 1 - DELETED 'IP52 RATED'. DELETED SERVICE NOTE 18/10/2016

B R6032 - ADDED 1.5kW LENGTH, MASS & ELECTRICAL REQUIREMENTS 17/10/2016

THE LENGHTH OF THE CASE, DENOTED 'X' CAN BE FOUND FROM THE RATED POWER, THE 3rd TO 6th CHARACTERS IN THE ORDER CODE OF THE QUBE FIBRE LASER, AND THE 17th CHARACTER IN THE ORDER

CODE SHOWN IN BOLD IN THE REPRESENTATIVE ORDER CODE BELOW

SP-1000-C-W-020-10-PIQ-012-001-001RATED POWER

W17th CHARACTER LENGTH X

mmESTIMATED MASS

kgESTIMATED MASS WITH

TRANSPORT FRAME kg

COMMENT

300 - 1000 0 681 60 63 MEDIUM CASE

1000 B 832 80 83 LONG CASE

1500 0 832 80 83 LONG CASE

5

A B

3

4

2

C D E

MATERIAL

TITLE

DWG.

No

1

A B C D E

F

FINISH

DRAWN

SCALE DATE

G

DWG. No

SIZE A1

5

3

4

2

F

1

GFS

-S00

138

FS-S00138

2:1

ALL OVER UOS1.6

SHT 1 OF 4

QUBE CASED, 500W TO 1.5kW INSTALLATION DWG

SEE ABOVE

07/09/2016

Bob Stevens

REMOVE ALL BURRS AND SHARP EDGES

THIS DRAWING IS THE PROPERTYOF SPI LASERS UK LTD.IT IS CONFIDENTIAL AND MUST NOT BECOPIED OR REPRODUCED WITHOUTPRIOR PERMISSION IN WRITING.COPYRIGHT C 2016

THIRD ANGLE PROJECTION

DIMENSIONS IN MILLIMETRES

UNLESS OTHERWISE STATED ALL THREADSTO BE ISO COARSE SERIES 6H OR 6g

REMOVE SHARP EDGES AND BURRS

DO NOT SCALE THE DRAWING.

MATERIAL TOLERANCES UNLESSOTHERWISE STATEDX. +/- 0.5X.X +/- 0.10X.XX +/- 0.05ANGLES +/- 0.25

DIMENSIONS APPLY AFTERSURFACE TREATMENT

SPI LASERS UK LTDCOSFORD LANE,SWIFT VALLEY,RUGBY, WARWICKSHIRECV21 1QN, ENGLAND,Tel +44 (0) 1788 570321

REVISION D.

12.0

COOLANT WATERFLOW CONNECTION

12mm PIPE

BD0MINIMUM BEND RADIUSOF 100mm

COOLANT WATERRETURN CONNECTION

12mm PIPE

ELECTRICAL SUPPLY CONNECTIONVARIOUS , SEE SHEET 3FOR CONNECTOR TYPES. SEE MANUAL SM-S00500SECTION 7.6 FOR POWER& WIRING REQUIREMENTS

SOLENOID

COOLANT DISTRIBUTIONMANIFOLD

ELECTRICAL CONNECTIONS

ID FUNCTION CONNECTOR DESCRIPTION MATING CONNECTOR DESCRIPTION

PL1 RS-232 SERIAL PORT CONTROL INTERFACE 9-WAY D- TYPE MALE CONNECTOR

9-WAY D-TYPE FEMALE CONNECTOR

PL5 MACHINE INTERFACE 25-WAY D-TYPE MALE CONNECTOR

25-WAY D-TYPE FEMALE CONNECTOR

PL81 PROCESS CABINET DOOR INTERLOCK SWITCHES

4 WAY 5.08mm PITCH CONNECTOR

PHOENIX CONTACT PART NUMBER 1778001

PL82 POWER SUPPLY REMOTE CONTROL 8 WAY 5.08mm PITCH CONNECTOR

PHOENIX CONTACT PART NUMBER 1778043

PL91 EXTERNAL INTERLOCK AND REMOTE RESET 6 WAY 5.08mm PITCH CONNECTOR

PHOENIX CONTACT PART NUMBER 1778027

PL96 COOLER INTERLOCK 2 WAY 5.08mm PITCH CONNECTOR

PHOENIX CONTACT PART NUMBER 1777989

PL98 REMOTE INDICATORS 10 WAY 5.08mm PITCH CONNECTOR

PHOENIX CONTACT PART NUMBER 1778069

SK2 ETHERNET RJ-45 SOCKET

SK95 FCMS 9-WAY D-TYPE FEMALE CONNECTOR

9-WAY D-TYPE MALE CONNECTOR PROVIDED BY SPI LASERS

SK99 SPI LASERS' USE ONLY -

5

A B

3

4

2

C D E

MATERIAL

TITLE

DWG.

No

1

A B C D E

F

FINISH

DRAWN

SCALE DATE

G

DWG. No

SIZE A1

5

3

4

2

F

1

GFS

-S00

138

FS-S00138

1:5

ALL OVER UOS1.6

SHT 2 OF 4

07/09/2016

Bob Stevens

REMOVE ALL BURRS AND SHARP EDGES

THIS DRAWING IS THE PROPERTYOF SPI LASERS UK LTD.IT IS CONFIDENTIAL AND MUST NOT BECOPIED OR REPRODUCED WITHOUTPRIOR PERMISSION IN WRITING.COPYRIGHT C 2016

THIRD ANGLE PROJECTION

DIMENSIONS IN MILLIMETRES

UNLESS OTHERWISE STATED ALL THREADSTO BE ISO COARSE SERIES 6H OR 6g

REMOVE SHARP EDGES AND BURRS

DO NOT SCALE THE DRAWING.

MATERIAL TOLERANCES UNLESSOTHERWISE STATEDX. +/- 0.5X.X +/- 0.10X.XX +/- 0.05ANGLES +/- 0.25

DIMENSIONS APPLY AFTERSURFACE TREATMENT

SPI LASERS UK LTDCOSFORD LANE,SWIFT VALLEY,RUGBY, WARWICKSHIRECV21 1QN, ENGLAND,Tel +44 (0) 1788 570321

QUBE CASED, 500W TO 1.5kW INSTALLATION DWG

REVISION D.

REMOVE THIS COVER TO GAIN ACCESS TO ELECTRICAL SUPPLY CABLE CONNECTION TERMINALS.4 POSITIONS FOR CABLE ENTRYBY ROTATION OF COVER THRU' 90INCREMENTS.

EARTH CONTINUITY WIRETO CASE MUST BE ATTACHED

IEC 6O309 TYPE 6H CONNECTOR OPTION MAINS EXTERNAL TERMINAL BLOCK OPTION

IEC 60320 TYPE C20 CONNECTOR OPTION

500W 750W 1000W 1500W

SUPPLY SINGLE PHASE SINGLE PHASE SINGLE PHASE SINGLE PHASE

RATED VOLTAGE 200 - 240V AC 200 - 240V AC 200 - 240V AC 200 - 240 V AC

RATED VOLTAGE TOLERANCE 10% 10% 10% 10%

FREQUENCY 47 - 63Hz 47 - 63Hz 47 - 63Hz 47 - 63Hz

MAXIMUM SUPPLY CURRENT - NOTE 1 11 - 9A 16 - 14A 22 - 18A 32 - 27A

SUGGESTED MINIMUM WIRE SIZE mm^2 -NOTE 2 1.5 1.5 2.5 4

SUGGESTED MINIMUM WIRE SIZE AWG - NOTE 2 14 14 12 10

MAXIMUM WIRE SIZE mm^2 10 10 10 10

MAXIMUM WIRE SIZE AWG 8 8 8 8

5

A B

3

4

2

C D E

MATERIAL

TITLE

DWG.

No

1

A B C D E

F

FINISH

DRAWN

SCALE DATE

G

DWG. No

SIZE A1

5

3

4

2

F

1

GFS

-S00

138

FS-S00138

1:5

ALL OVER UOS1.6

SHT 3 OF 4SEE ABOVE

07/09/2016

Bob Stevens

REMOVE ALL BURRS AND SHARP EDGES

THIS DRAWING IS THE PROPERTYOF SPI LASERS UK LTD.IT IS CONFIDENTIAL AND MUST NOT BECOPIED OR REPRODUCED WITHOUTPRIOR PERMISSION IN WRITING.COPYRIGHT C 2016SPI Lasers UK Ltd

THIRD ANGLE PROJECTION

DIMENSIONS IN MILLIMETRES

UNLESS OTHERWISE STATED ALL THREADSTO BE ISO COARSE SERIES 6H OR 6g

REMOVE SHARP EDGES AND BURRS

DO NOT SCALE THE DRAWING.

MATERIAL TOLERANCES UNLESSOTHERWISE STATEDX. +/- 0.5X.X +/- 0.10X.XX +/- 0.05ANGLES +/- 0.25

DIMENSIONS APPLY AFTERSURFACE TREATMENT

SPI LASERS UK LTDCOSFORD LANE,SWIFT VALLEY,RUGBY, WARWICKSHIRECV21 1QN, ENGLAND,Tel +44 (0) 1788 570321

QUBE CASED, 500W TO 1.5kW INSTALLATION DWG

REVISION D.

REMOVAL OF TRANSIT/LIFTING GEAR

1

2

3 2x

150mm APPROX

POSITION UNIT ONTO CABINET RUNNERS AS INDICATED.LIFTING GEAR ITEMS 1, 2 & 3 CAN THEN BE REMOVED WHILE STILL SUPPORTING THE FRONT END.

10 mm

SLIDE UNIT FURTHER INTO THE ENCLOSUREUNTIL LIFTING GEAR IS CLOSE TO THE RUNNER AS SHOWN. REMAINDER OF TRANSIT/LIFTINGGEAR CAN THEN BE REMOVED.THE FRONT END OF THE UNIT SHOULD CONTINUETO BE SUPPORTED UNTIL THE WEIGHT IS FULLYLOADED ONTO THE RUNNERS

5

A B

3

4

2

C D E

MATERIAL

TITLE

DWG.

No

1

A B C D E

F

FINISH

DRAWN

SCALE DATE

G

DWG. No

SIZE A1

5

3

4

2

F

1

GFS

-S00

138

FS-S00138

1:3

ALL OVER UOS1.6

SHT 4 OF 4

07/09/2016

Bob Stevens

REMOVE ALL BURRS AND SHARP EDGES

THIS DRAWING IS THE PROPERTYOF SPI LASERS UK LTD.IT IS CONFIDENTIAL AND MUST NOT BECOPIED OR REPRODUCED WITHOUTPRIOR PERMISSION IN WRITING.COPYRIGHT C 2016

THIRD ANGLE PROJECTION

DIMENSIONS IN MILLIMETRES

UNLESS OTHERWISE STATED ALL THREADSTO BE ISO COARSE SERIES 6H OR 6g

REMOVE SHARP EDGES AND BURRS

DO NOT SCALE THE DRAWING.

MATERIAL TOLERANCES UNLESSOTHERWISE STATEDX. +/- 0.5X.X +/- 0.10X.XX +/- 0.05ANGLES +/- 0.25

DIMENSIONS APPLY AFTERSURFACE TREATMENT

SPI LASERS UK LTDCOSFORD LANE,SWIFT VALLEY,RUGBY, WARWICKSHIRECV21 1QN, ENGLAND,Tel +44 (0) 1788 570321

QUBE CASED 500W TO 1.5kW INSTALLATION DWG