force and displacement measurement - department of...

35
ME 144L – Prof. R.G. Longoria Dynamic Systems and Controls Laboratory Department of Mechanical Engineering The University of Texas at Austin Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011

Upload: lamdat

Post on 09-Mar-2018

217 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Force and Displacement

Measurement

Prof. R.G. Longoria

Updated Fall 2011

Page 2: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Simple ways to measure a force

http://scienceblogs.com/dotphysics/2010/02/diy_force_probe.php

Page 3: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Example: Key Force/Deflection

measure low-level forces required to trigger

key-switch on a computer keyboard

keyx

keyx

inx

F k x= ⋅ ∆

Page 4: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Force and torque sensors

Force Sensing

DeviceForce Voltage

Sensitivity = voltage/force

A force or torque sensor provides an output in the form of an

electrical signal (voltage, current).

Examples:

Strain-gauge force sensorsOptical in-line torque sensor

Page 5: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Force sensing in products

Strain gauge bending

beam sensor

Omega Engineering

Economical bending

beam scale, ~$165Postal scale, ~$30

Page 6: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Sensing Mechanism

• To measure force (or torque), it is usually necessary

to design a compliant mechanical structure. This

structure may itself be a sensing material.

• Force will induce stress, leading to strain which can be

detected in various ways, for example:

– using strain gauges (piezoresistive effect)

– using crystals or ceramics (piezoelectric effect)

– optically

• Sometimes force can be measured using other types of

displacement sensing devices.

Page 7: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Lab study: beam sensor

Beam configuration with a

‘full-bridge’ strain gauge

configuration

Experimental setup

1

43

2

Wheatstone

bridge

Page 8: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Strain gauge concepts

Strain gauges exhibit piezoresistive

behavior, and are one of the

most common ways to measure

strain.

Types

• unbonded wire - basically a

wire under strain (c. 1940s)

• foil - type shown to left (c.

1950s) are most common

• semiconductor (c. 1960s)

Terminals

Gage width

Connectingwires

Gage length

Insulating layer and bonding cement

specimen under strain

Matrix length

Matrixwidth

Page 9: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Strain gauge Sensitivity

The gauge Factor, GA measure of the “sensitivity” of a strain gauge is given by the gauge factor,

which is defined as,

Using the derivation in Appendix B,

fractional change in resistance

fractional change in strainG =

“Piezoresistive effect”

1 1(1 2 )

dR dG

R

ρν

ε ε ρ= = + +

Typical values:

80% Ni, 20% Cr, G = 2

45% Ni, 55% Cu, G = 2

Platinum, G = 4.8

95% Pt, 5% Ir, G = 5.1

Semiconductor, G =70 to 135

Page 10: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

More on gauge Types

• Strain gauges come in many specialized forms and

typically include a calibrated gauge factor, G.

• Semiconductor strain gauges have the highest values

of G. These strain gauges can have G values of 70 to

135, and they are typically very small. However, there

are some disadvantages which include:

– output is not linear with strain,

– very temperature dependent,

– usually have a much lower strain limit than metallic type,

– more expensive than metallic type.

Page 11: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Strain DetectionOrder of Magnitude Calculation

Consider a situation where the strain is on the

order of 1 microstrain.

For a metallic foil strain gauge with G = 2, R =

120 ohm,

You need to measure a 0.002% change in R!

How would you detect such a change?

62 1 10 120 0.0024R G Rε −∆ = ⋅ ⋅ = ⋅ × ⋅ = Ω

Page 12: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Beam Sensors

• The beam structure/geometry is used

extensively in designing many types of force

and torque sensors.

• A beam offers certain advantages:

– easy geometry for basic analysis and design

– strain gauges can be mounted easily and configured

in several different ways to achieve different

objectives

See beam configurations

in Appendix C.

Page 13: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Wheatstone Bridge Configuration

1

43

2

( )( )1 4 2 3

1 2 3 4

o s

R R R Rv V

R R R R

−= ⋅

+ +

sV

Output DC voltage

Null condition is satisfied when:31

2 4

RR

R R=

If all the gauges have the same resistance, you can show:

[ ]1 2 3 11 2 3 4

4o

s

dv dR dR dR dR G

V Rε ε ε ε

− − += = − − +

This equation can be used

to guide placement of

gauges on a specimen.

Page 14: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Balanced Bridge

4

3

2

1

R

R

R

R=

4

2

3

1

R

R

R

R=

or,

A B

C

D

1 2

3 4

The bridge is balanced when the ratio of resistances of any two

adjacent arms is equal to the ratio of resistances of the remaining

two arms (taken in the same sense).

Page 15: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Signal Conditioning:

Impedance BridgesImpedance bridges are used to convert the output

of resistive, capacitive, or inductive sensors into

a voltage signal

VOLTAGEIMPEDANCE

Many types of impedance bridges

exist; see examples in Appendix D.

Page 16: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Strain Gauge Measurement System

ovεG R⋅

R∆

Bridge Amplifier

amplifiedv

Sensing

Mechanism

The strain gauge is part of a multi-stage process

that generates a voltage signal proportional to the

strain.

The amplifier used in our lab

experiments is described in Appendix E.

Page 17: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Displacement measurements

• Mechanical (gage blocks, rulers, etc.)

• Strain gauges measure deflection

• Contact sensors

– LVDT

– Inertial sensors (accelerometers, seismometers)

• Non-contact sensors

– Optical

There are MANY ways to

measure displacement…

Page 18: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

LVDT – displacement sensorLinear variable differential transformer (LVDT) monitors

displacement of a core which modulates the mutual inductance

between two coils.

Coil 1(secondary)

Coil 2(secondary)

Primarycoil

Core

Motion

ConstantAC voltage

Difference voltage

vo = v1

-

v2

Insulatingform orbobbin

(a) (b)

Primary

core

Secondary

Secondary

Difference voltage

vo = v1

-

v2

v1

v2

This sensor type is described briefly in

Appendix G.

Page 19: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Page 20: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Summary

• Force measurement takes advantage of the relationship between force, displacement and stiffness.

• Strain gauges are a common basis for sensors that can measure force or torque

• We ‘wrap’ the sensor with signal conditioning to get a measurable signal (voltage or current)

• Strain gauges are core knowledge for mechanical engineers

Page 21: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Appendix A: Example force and torque sensors

Appendix B: Piezoresistivity and gauge factor

Appendix C: Beams as basis for strain-gauge sensors

Appendix D: Summary of impedance bridges

Appendix E: DMD-465W – strain-gauge amplifier

Appendix F: Examples of ‘home-made’ force sensors

Appendix G: Inductive-type sensors

Page 22: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Example force sensors

• Strain-gauge force sensors

• PiezoelectricXYZ force sensor from

PCB Piezotronics

Force and

acceleration

Bruel and Kjaer

Omega

Sensotec

Omega

General purpose

PCB Piezotronics

Appendix A:

Page 23: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Example torque sensors

Reaction torque sensorIn-line torque sensor

Appendix A:

Page 24: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Piezoresistivity (1)

We know that for a conductor of uniform area, the resistance is given by,

where ρ is the resistivity (cm ohm)., l is the length, and A is the cross-sectional

area.

Under strain, the change in R is,

which for uniform A is,

lR

A

ρ≡

For typical conductors, the resistivity values in units of ohm mm2/m are: Aluminum 0.0278, Pure Iron 0.1,

Constantan 0.48, Copper 0.0172, Gold 0.0222, Tungsten 0.059, Manganese 0.423, Nickel 0.087.

R R RdR dl dA d

l Aρ

ρ

∂ ∂ ∂= + +

∂ ∂ ∂

2

l ldR dl dA d

A A A

ρ ρρ= − +

Appendix B:

Page 25: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Piezoresistivity (2)The fractional change of R is of more interest, so we

find,

= fractional change in length

= fractional change in area

= fractional change in resistivity

dl

l

dA

A

d ρ

ρ

dR dl dA d

R l A

ρ

ρ= − +

Appendix B:

Page 26: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Piezoresistivity (3)

For a linearly elastic body,

where E is the Young’s modulus. Recall

And for an area A = w t, the fractional change is,

Recall that n is Poisson’s ratio. Now the fractional change in R is,

/xx o x

dlF A E E

lσ ε= = ⋅ = ⋅

, , x y z

dl dl dl

l l lε ε ν ε ν= = − = −

2 x

dA dw dt

A w tνε= + = −

(1 2 ) x

dR d

R

ρν ε

ρ= + ⋅ +

Input a

strain

Output a

resistance change

GeometricMaterial

Appendix B:

Page 27: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Compensatedwith dummygage in arm 2 orarm 3

TemperatureBending

SIDE

TOP

SIDE

TOP

1

1

4

1

1

Compensated(with four-armbridge anddummy gages inarms 2 and 3)

SIDE

TOP

2B

1A

1B

2A

1A2A

2 independent bridges

SIDE

TOP

3

1

4

2

12

one four arm bridge

Compensatedwith two-armbridges.

Compensatedthrough four-arm bridge.

Axially-Loaded Beam K

1

1+ν

2

2(1+ν)

Sensitive

NOTES: All axially-loaded beams sensitive to torsion. Requirement for null: R1/R2 = R3/R4K = Bridge constant = (output of bridge)/(output of primary gage)

Compensated withdummy gage inarm 2 or arm 3

TemperatureAxial/

Torsion

SIDE

TOP

SIDE

TOP

1

1

2

1

1

Compensated

SIDE

TOP

a

2

b

1

2

1

2 independent bridges

SIDE

TOP

3

1

2

4

1

one four arm bridge

Cantilevered Beam K

1

4

2

1+bν/a

Sensitive toboth axialand torsionloads

Compensatedfor both axialand torsionloads

Compensated

Compensated

Compensated

4

Compensatedfor both axialand torsionloads

Compensatedfor both axialand torsionloads

Compensated

Compensated

Strain gauge orientations for axially-loaded and cantilevered Beams

A

B

C

D

E

F

G

H

Appendix C:

Page 28: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Impedance BridgesFrom Beckwith,

Buck, and

Marangoni,

“Mechanical

Measurements”,

Addison-Wesley,

3rd ed, 1982.

Appendix D:

Page 29: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

DMD 465WBOmega Engineering

Appendix E:

Page 30: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

‘Home-made’ force sensors

Belt force measurement (Kim and Marshek, c. 1981)

This apparatus used two different types of ‘home-made’ force sensors to measure

the forces induced by a grinding belt on the disk.

Appendix F:

Page 31: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Measuring tangential belt forces

This beam has a

full-bridge of

strain gauges.

Appendix F:

Page 32: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Measuring normal belt forcesAppendix F:

Page 33: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Using displacement to measure low-range forces

data0 0

0.5 30

1 42.5

1.5 45

2 38

2.5 30

2.75 22.5

3 30

3.25 55

3.5 100

:=(mm) (gm)

F (gm)

depress

release

(mm)key

x

Rubber

dome

gm21.6

mmmk =

16.15 gmtm =

tm

0 ?inx≤ ≤

keyx

g

keyx

keyx

inx

Measuring the difference in displacement between the key and

the push tube allows force measurement (assuming stiffness is

known).

Appendix F:

Push tube

Page 34: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

Inductors store magnetic energy

• Inductors store magnetic energy

• An inductive (or variable-reluctance) sensor design takes advantage of how magnetic flux passes through the circuit, and this can be detected as a change in inductance.

Appendix G:

Page 35: Force and Displacement Measurement - Department of …longoria/me344/lab2/L2_Force_Displacement.pdf · Force and Displacement Measurement Prof. R.G. Longoria Updated Fall 2011. ME

ME 144L – Prof. R.G. LongoriaDynamic Systems and Controls Laboratory

Department of Mechanical EngineeringThe University of Texas at Austin

‘Variable-reluctance’ sensors

In all of these

devices, the

inductance, L, is

changing based

on geometry.

You can also have

material change

to affect

reluctance.

Reluctance is inversely related to inductance

Appendix G: