Topics to be covered Repeat Analysis Analysis Of Reject Versus
Repeated Films Determination Of Rejection
Rate Distribution Of Rejected Film Guide To Good Practice
The desired outcome of a diagnostic x-ray study is the
creation of an acceptable diagnostic image, a correct
diagnosis, and the satisfaction of all internal and external customer.
Quality control testing can only minimize the risk of obtaining sub-quality images but cannot prevent them entirely.
Quality management program must also include looking at the final outcome of a diagnostic x-ray procedure and determining the quality of outcome to see if further improvement can be achieved.
Outcome assessment
Repeat analysis of images Artifact analysis of images Accuracy, sensitivity and
specificity analysis of diagnosis.
Repeat analysisIt is a systematic process of
cataloging rejected films and determining the nature of the repeat so that repeat images can be minimized or eliminated in the future
Importance of repeat analysis
It provides knowledge of data about:Equipment and accessory performanceDepartmental proceduresSkill level of the technical staff
With this knowledge, solutions can be found to minimize repeats and also document the effectiveness of quality control and quality assurance protocols
Advantages
Repeat analysis can show how many repeats in the depart w/c in turns address the party concerns of what are the problems.
With lower department repeat rates:Improved department efficiencyLower department costLower patient doses
Improved department efficiency
With number of repeats are kept low, the amount of time that patient must spend undergoing diagnostic procedures decreases.
This in turns increases patient satisfaction and allows the department to service more patients in the same period.
Lower department costs
When the number of repeats are reduced, the costs associated with film, processing, labor and depreciation of the equipment decrease significantly.
Lower patient doses
Images that are unacceptable results in the repeat of the particular view, which means that the patient must be re-exposed to ionizing radiation.
Causal Repeat rate
In repeat analysis studies performed on departments without quality control procedures for the darkroom, processor and equipment, 75% of all repeats were caused by improper optical density of film.
With quality control protocol in place, studies shows that most repeats are results of positioning errors.
National average of repeat causes for departments with quality control proceduresCategory % Repeat
Positioning 30
Light films 14
Misc. (artifacts) 14
Darkrooms 11
Black films 9
Tomography scouts 8
Fog 5
Patient motion 5
Mechanical problems 4
Repeat analysis For a study to be done, a
worksheet for radiographic imaging departments should be used, so that the proper statistical information is obtained and recorded.
Worksheet
A worksheet should include the radiographic procedures performed in the department, along with the possible causes of rejection, such as positioning, overexposure, under exposure, motion, artifacts, and miscellaneous causes
Causal repeat rate
The causal repeat rate is the percentage of repeats from a specific cause and is calculated with the following equation:
Causal repeat =
rate
Number of repeats for a specific cause
Total Number of repeatsX 100
Example
A total repeat of a department has 200 films in a month, and 90 out of 200 films are the result of positioning errors.
So the percent of repeats caused by positioning error is 45% of the total repeat films.
Total repeat rate
The total department repeat rate is determined with the following equation:
Total repeat rate =
Number of repeat films
Total number of views takenX 100
Example
If a department has a total of 2300 views during a month and has incurred a 300 rejected film.
A 13% percent the department has incurred in a month
Department repeat rates should not exceed 4% to 6% and should be less that 2% for mammographic procedures.
Any department exceeding 10% to 12% should be examined seriously, because of inefficiency and contribute to high patient dose.
For a department to evaluate its efficiency a worksheet should be used for any repeat film.
As a QA/QC officers a worksheet can aide in identifying problems if it is mechanical/technical or by human errors.
Laboratory Exercise
Instruction Each group are assigned with a number of
films.Analyze the films.After analyzing the films, the film should be
rotated to other groups for analysis.
Survey period to location
repeat category examinati
on
Position Over-exposed
Under-expose
Motion Artifacts Other Total %
UGI 12 4 3 1 5 8 33 23%
Ba. Enema
15 1 1 20 3 1 41 28%
IVP 5 2 3 1 7 1 19 13%
OTHER 10 1 3 11 25 1 51 35%
TOTAL 42 8 10 33 40 11 144
% 29% 6% 7% 23% 28% 8%
Total films
viewed
1320 Total repeat
144 Total Repeat Rate 11%