  • 7/31/2019 Fluke India Temperature Presentation

    1/13

    Infrared Digital

    Thermometers

  • 7/31/2019 Fluke India Temperature Presentation

    2/13

    Infrared thermometers are designed for applications

    requiring users to instantly measure temperature inhard-to-reach areas

    http://www.fluke.com/fluke/inen/products/Thermometers.htm

  • 7/31/2019 Fluke India Temperature Presentation

    3/13

    http://www.fluke.com/fluke/inen/products/Thermometers.htm

    Temperature measurement

    Contact Type Thermometers Uses temperature probes / thermo-couples

    Highly accurate, good for Lab / calibration work

    Non-Contact Thermometers Uses Infra Red beams to measure surface temperature

    Moderate accuracy for day to day work

  • 7/31/2019 Fluke India Temperature Presentation

    4/13

    http://www.fluke.com/fluke/inen/products/Thermometers.htm

    Why IR thermometers - 3 main reasons

    Safety From high temperature objects

    From high voltages (temperature at circuit

    breakers terminals, bus-bars)

    Moving objects - Rotating motors, conveyors,

    belts

    Hard to Reach Object mounted at heights

    Object mounted inside safety perimeter

    Objects at hard to reach places

    Convenience IR thermometer is easy to use

    Just point-shoot-measure

  • 7/31/2019 Fluke India Temperature Presentation

    5/13

    http://www.fluke.com/fluke/inen/products/Thermometers.htm

    Emissivity

    A numerical value between 0 and 1 which indicates the ability of an

    object to emit infrared energy.

    Emissivity is determined primarily by the material from which an

    object is constructed and its surface finish.

    Examples:

    Polished brass: 0.03

    Roughly polished copper: 0.07

    Black lacquer paint: 0.96

    Circuit boards

    Important Parameters IR Thermometers

  • 7/31/2019 Fluke India Temperature Presentation

    6/13

    http://www.fluke.com/fluke/inen/products/Thermometers.htm

    Important Parameters IR Thermometers

    DISTANCE TO SPOT RATION

    Ratio of the distance from the measured object vs. sampling

    spot size

    EXAMPLES:

    - 4:1 optical resolution: 4 away from measured object1

    diameter

    - 12:1 optical resolution: 24 away from measured object2

    diameter

    OBJECT SIZE

    Know your application! An IR device with a 4:1 optical

    resolution

    can not effectively be used to measure a diffuser temperature

    15 away.

  • 7/31/2019 Fluke India Temperature Presentation

    7/13http://www.fluke.com/fluke/inen/products/Thermometers.htm

    FIELD OF VIEW

    - Field of view refers to the target size relative to the IR thermometers spot size .

    - Ideally, the target surface to be measured should be twice the size of the spot .

    - If the spot size is larger than the surface to be measured, measurement errors are

    increased .

    - Spot size calculator www.Fluke.Com

    Important Parameters IR Thermometers

    http://www.fluke.com/http://www.fluke.com/
  • 7/31/2019 Fluke India Temperature Presentation

    8/13http://www.fluke.com/fluke/inen/products/Thermometers.htm

    HOW DOES IR THERMOMETER WORK

    Infrared thermometers measure the surface

    temperature of an opaque object

    The thermometers optics sense emitted,

    reflected, and transmitted energy, which is

    collected and focused onto a Detector

    The units electronics translate the information

    into a temperature reading which the unit

    displays

  • 7/31/2019 Fluke India Temperature Presentation

    9/13http://www.fluke.com/fluke/inen/products/Thermometers.htm

    IMPORTANT SPECIFICATIONSFOR SELECTION OF RIGHT IRTHERMOMETER

    Temperature range

    Measurement accuracy

    Distance to Spot Ratio

    Rugged & Reliable

    Response time

    Display resolution

    Battery life

  • 7/31/2019 Fluke India Temperature Presentation

    10/13

    OPERATING THE IR THERMOMETER

    To measure temperature, point the thermometer at an object and pull the trigger.

    Laser pointer to help aim the thermometer

    To find a hot or cold spot, aim the thermometer outside the desired area.

    As the distance from the object being measured increases, the spot size of the area

    measured by the unit becomes larger

    For accurate measurements, make sure that the target is larger than the units spot

    size

    The display retains its last infrared measurement for twenty seconds when the trigger

    is released, with HOLD appearing on the display

    The stored data can be downloaded to a personal computer (PC) via the

    included USB cable and the included Fluke View Forms Documenting Software

    http://www.fluke.com/fluke/inen/products/Thermometers.htm

  • 7/31/2019 Fluke India Temperature Presentation

    11/13

    SAFETY PRECAUTIONS

    Before using the thermometer inspect the case. Do not use the thermometer if it appears

    damaged. Look for cracks or missing plastic. Pay particular attention to the insulation

    around the connectors

    Disconnect the thermocouple(s) from the thermometer before opening the case

    Replace the batteries as soon as the battery indicator appears. The possibility of false

    readings can lead to personal injury

    Do not use the thermometer if it operates abnormally. Protection may be impaired. When

    in doubt, have the thermometer serviced / calibrated at authorized labs.

    Reflective objects result in lower than actual temperature measurements. These objects

    pose a burn hazard

    Do not operate the thermometer around explosive gas, vapor, or dust

    http://www.fluke.com/fluke/inen/products/Thermometers.htm

  • 7/31/2019 Fluke India Temperature Presentation

    12/13http://www.fluke.com/fluke/inen/products/Thermometers.htm

    Company Name : Fluke Corporation - India

    Website: http://www.fluke.com/fluke/inen/home/default.htm

    Address: Division of DHR Holding India Pvt. Ltd.10th Floor, Sigma

    Hiranandani Business Park, Powai

    Mumbai, Maharashtra 400076

    ContactNumber: 1 800 209 9110

    http://www.fluke.com/fluke/inen/home/default.htmhttp://www.fluke.com/fluke/inen/home/default.htmhttp://www.fluke.com/fluke/inen/home/default.htm
  • 7/31/2019 Fluke India Temperature Presentation

    13/13

    Thank You