Page 1: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

Filmless MLC Quality Assurance in Robotic Radiosurgery

Jacob A. Gersh, PhD DABR

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 2: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

Disclosures

Speaker is the owner and president of Spectrum Medical Physics, LLC, which has contracts or financial interest with: Standard Imaging Inc. Siemens Medical Integrated Medical Technologies

Speaker received financial reimbursement for travel

associated with the presentation presented herein (Standard Imaging Inc).

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 3: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

QA StereoCheckerTM

Filmless CyberKnife® QA System

High-Resolution Flat-Panel Imager

200 micron Resolution High Frame Rate Portable and Compact

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 4: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

QA StereoCheckerTM

Aligned using implanted high-Z markers Automated alignment using treatment

localization system Fast and consistent setup No manual manipulation of robot No attachment to collimator housing Designed to be used by physicists and therapists

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 5: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

The InCiseTM MLC

Available for CyberKnife M6 26 leaf pairs 0.385 mm wide at 800 mm Full over-travel Allows for treatment of larger lesions 100 mm x 115 mm Leaf Position Accuracy ≤0.95 mm

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 6: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

Less than 5 minutes Replaces visual film-based

tests Replaces monthly Garden

Fence Tests Capable of detecting 0.2

mm leaf variations

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 7: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 8: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 9: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 10: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 11: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 12: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 13: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

MLC QA – The Garden Fence Test

0.1mm 0.2mm 0.3mm 0.4mm 0.6mm 0.8mm 1.0mm

Induced Individual leaf errors 0.1 mm to 1.0 mm Increments of 0.1 mm

“Detection” occurs if a cluster of pixels is simultaneously above a threshold with SNR above 10.

Mean pixel difference values are used as “Warning” and “Error Thresholds” as well as for trend analysis.

Presented at the 2016 Radiosurgery Society Scientific Meeting

Page 14: Filmless MLC Quality Assurance in Robotic Radiosurgery · 2017. 4. 12. · MLC QA – The Garden Fence Test Less than 5 minutes Replaces visual film -based tests Replaces monthly

Thank You

J. David Myers, CMD Michael Starnes, CMD Ryan S. Dubose, MS DABR Kevin L. Shay, MS DABR Daniel B. Fried, MD PHD Melanie Rhyne, RT Abby Sammut, RT Lauren Miles, RT LinLee Khoune, RT

Presented at the 2016 Radiosurgery Society Scientific Meeting