x3 FilmArray ® Meningitis / Encephalitis Panel Quick Guide BioFire Diagnostics, LLC | 390 Wakara Way, Salt Lake City, Utah 84108, USA | 1-801-736-6354 Step 1: Prepare Pouch — Insert pouch into Pouch Loading Station. — Place Sample Injection Vial into red well. — Place Hydration Injection Vial into blue well. Step 2: Hydrate Pouch — Twist off Hydration Injection Vial, leaving cap in Pouch Loading Station, and insert into pouch hydration port. — Forcefully push down to puncture seal and wait as Hydration Solution is drawn into pouch. Step 3: Prepare Sample Mix — Add Sample Buffer to Sample Injection Vial : • Invert Sample Buffer Ampoule so that tip is facing up. Note: Do not touch the tip of the ampoule. • Firmly pinch textured plastic tab on side of ampoule until seal snaps. • With the tip facing down, dispense Sample Buffer into Sample Injection Vial using a slow, forceful squeeze, followed by a 2 nd squeeze. Avoid generating excessive bubbles. Using transfer pipette, draw cerebrospinal fluid (CSF) specimen to 2 nd line. — Add to Sample Injection Vial. — Tightly close lid of Sample Injection Vial. — Mix sample by gently inverting Sample Injection Vial 3 times. — Return Sample Injection Vial to red well of Pouch Loading Station. Warning: The Sample Buffer is harmful if swallowed, can cause serious eye damage and/or skin irritation. Step 4: Load Sample Mix — Unscrew Sample Injection Vial from cap. — Pause for 3-5 seconds, then remove Sample Injection Vial, leaving cap in Pouch Loading Station. — Insert Sample Injection Vial into pouch sample port. — Forcefully push down to puncture seal and wait as Sample Mix is drawn into pouch. Step 5: Run Pouch — Follow instructions on computer for initiating a test. To avoid contamination always wear gloves and work behind a protective shield. 0:05

Upload: others

Post on 29-Sep-2020

9 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FilmArray Meningitis / Encephalitis Panel Quick Guide · 2020. 3. 13. · FilmArray ® Meningitis / Encephalitis Panel Quick Guide BioFire Diagnostics, LLC | 390 Wakara Way, Salt

x3x3

x3

x3

x3

x3

FilmArray® Meningitis / Encephalitis Panel Quick Guide

BioFire Diagnostics, LLC | 390 Wakara Way, Salt Lake City, Utah 84108, USA | 1-801-736-6354

Step 1: Prepare Pouch — Insert pouch into Pouch Loading Station. — Place Sample Injection Vial into red well. — Place Hydration Injection Vial into blue well.

Step 2: Hydrate Pouch — Twist off Hydration Injection Vial, leaving cap in Pouch Loading Station, and insert into pouch hydration port.

— Forcefully push down to puncture seal and wait as Hydration Solution is drawn into pouch.

Step 3: Prepare Sample Mix — Add Sample Buffer to Sample Injection Vial:• Invert Sample Buffer Ampoule so that tip is facing up. Note: Do not touch the tip of the ampoule.• Firmly pinch textured plastic tab on side of ampoule until seal snaps.• With the tip facing down, dispense Sample Buffer into Sample Injection Vial using a slow, forceful squeeze, followed by a 2nd squeeze. Avoid generating excessive bubbles.

— Using transfer pipette, draw cerebrospinal fluid (CSF) specimen to 2nd line.

— Add to Sample Injection Vial. — Tightly close lid of Sample Injection Vial. — Mix sample by gently inverting Sample Injection Vial 3 times. — Return Sample Injection Vial to red well of Pouch Loading Station.

Warning: The Sample Buffer is harmful if swallowed, can cause serious eye damage and/or skin irritation.

Step 4: Load Sample Mix — Unscrew Sample Injection Vial from cap. — Pause for 3-5 seconds, then remove Sample Injection Vial, leaving cap in Pouch Loading Station.

— Insert Sample Injection Vial into pouch sample port. — Forcefully push down to puncture seal and wait as Sample Mix is drawn into pouch.

Step 5: Run Pouch

— Follow instructions on computer for initiating a test.

To avoid contamination always wear gloves and work behind a protective shield.

x3

0:05

Page 2: FilmArray Meningitis / Encephalitis Panel Quick Guide · 2020. 3. 13. · FilmArray ® Meningitis / Encephalitis Panel Quick Guide BioFire Diagnostics, LLC | 390 Wakara Way, Salt

FilmArray® Meningitis / Encephalitis Panel Quick Guide

BioFire Diagnostics, LLC | 390 Wakara Way, Salt Lake City, Utah 84108, USA | 1-801-736-6354

FilmArrayMeningitis / Encephalitis (ME) Panel - IVD

www.BioFireDx.com

Run SummarySample ID: ATCC-VR-1467-2X-r1 Run Date: 12 Jun 2015

4:41 PMDetected: Human herpesvirus 6 Controls: Passed

WARNING: The FilmArray ME Panel does not distinguish between latent and active CMV and HHV-6 infections. Detection of theseviruses may indicate primary infection, secondary reactivation, or the presence of latent virus. Results should always be interpretedin conjunction with other clinical, laboratory, and epidemiological information.

Result SummaryBacteria

Not Detected Escherichia coli K1Not Detected Haemophilus influenzaeNot Detected Listeria monocytogenesNot Detected Neisseria meningitidisNot Detected Streptococcus agalactiaeNot Detected Streptococcus pneumoniae

VirusesNot Detected CytomegalovirusNot Detected EnterovirusNot Detected Herpes simplex virus 1Not Detected Herpes simplex virus 2Detected Human herpesvirus 6Not Detected Human parechovirusNot Detected Varicella zoster virus

YeastNot Detected Cryptococcus neoformans/gattii

Run DetailsPouch: ME Panel v1.4 Protocol: CSF v3.0

Run Status: Completed Operator: karen senne (karen)Serial No.: 02549366 Instrument: FA4046

Lot No.: 200415

FilmArray®

Meningitis / Encephalitis (ME) Panelwww.BioFireDx.com

Run SummarySample ID: 0417_137 Run Date: 21 Apr 2014

5:34 PMDetected: Epstein-Barr virus

Varicella zoster virusControls: Passed

WARNING: The FilmArray ME panel does not distinguish between latent and active herpesvirus infections (e.g., CMV, EBV,HHV6). Herpesvirus results should be used in conjunction with other clinical, laboratory, and epidemiological data.

Result SummaryBacteria

Not Detected Escherichia coli K1Not Detected Haemophilus influenzaeNot Detected Listeria monocytogenesNot Detected Neisseria meningitidisNot Detected Streptococcus agalactiaeNot Detected Streptococcus pneumoniae

VirusesNot Detected CytomegalovirusNot Detected EnterovirusDetected Epstein-Barr virusNot Detected Herpes simplex virus 1Not Detected Herpes simplex virus 2Not Detected Human herpesvirus 6Not Detected Human parechovirusDetected Varicella zoster virus

YeastNot Detected Cryptococcus neoformans/gattii

Run DetailsPouch: ME Panel v1.4 Protocol: CSF v2.0

Run Status: Completed Operator: Cody Oswald (cody)Serial No.: 01422247 Instrument: ITI FA "FA2430"

Lot No.: 140411A

FilmArrayMeningitis / Encephalitis (ME) Panel - IVD

www.BioFireDx.com

Run SummarySample ID: ATCC-VR-1467-2X-r1 Run Date: 12 Jun 2015

4:41 PMDetected: Human herpesvirus 6 Controls: Passed

WARNING: The FilmArray ME Panel does not distinguish between latent and active CMV and HHV-6 infections. Detection of theseviruses may indicate primary infection, secondary reactivation, or the presence of latent virus. Results should always be interpretedin conjunction with other clinical, laboratory, and epidemiological information.

Result SummaryBacteria

Not Detected Escherichia coli K1Not Detected Haemophilus influenzaeNot Detected Listeria monocytogenesNot Detected Neisseria meningitidisNot Detected Streptococcus agalactiaeNot Detected Streptococcus pneumoniae

VirusesNot Detected CytomegalovirusNot Detected EnterovirusNot Detected Herpes simplex virus 1Not Detected Herpes simplex virus 2Detected Human herpesvirus 6Not Detected Human parechovirusNot Detected Varicella zoster virus

YeastNot Detected Cryptococcus neoformans/gattii

Run DetailsPouch: ME Panel v1.4 Protocol: CSF v3.0

Run Status: Completed Operator: karen senne (karen)Serial No.: 02549366 Instrument: FA4046

Lot No.: 200415

1. 2.

FilmArrayMeningitis / Encephalitis (ME) Panel - IVD

www.BioFireDx.com

Run SummarySample ID: ATCC-VR-1467-2X-r1 Run Date: 12 Jun 2015

4:41 PMDetected: Human herpesvirus 6 Controls: Passed

WARNING: The FilmArray ME Panel does not distinguish between latent and active CMV and HHV-6 infections. Detection of theseviruses may indicate primary infection, secondary reactivation, or the presence of latent virus. Results should always be interpretedin conjunction with other clinical, laboratory, and epidemiological information.

Result SummaryBacteria

Not Detected Escherichia coli K1Not Detected Haemophilus influenzaeNot Detected Listeria monocytogenesNot Detected Neisseria meningitidisNot Detected Streptococcus agalactiaeNot Detected Streptococcus pneumoniae

VirusesNot Detected CytomegalovirusNot Detected EnterovirusNot Detected Herpes simplex virus 1Not Detected Herpes simplex virus 2Detected Human herpesvirus 6Not Detected Human parechovirusNot Detected Varicella zoster virus

YeastNot Detected Cryptococcus neoformans/gattii

Run DetailsPouch: ME Panel v1.4 Protocol: CSF v3.0

Run Status: Completed Operator: karen senne (karen)Serial No.: 02549366 Instrument: FA4046

Lot No.: 200415

Ashley Hunter (ah)

4.3.

5.R

FIT-

PR

T-02

75-0

3

The Run Summary Section displays information about the sample and a summary of the control and test results.

1. Detected: — Names of any detected pathogens — If 'None', no pathogens were de-tected

— If ' Invalid', RETEST SAMPLE 2. Controls:

— If 'Passed', results are valid — If ' Failed', RETEST SAMPLE — If ' Invalid', RETEST SAMPLE

3. ' Detected', pathogen was detected4. 'Not Detected', pathogen was not

detected — If ' Invalid', RETEST SAMPLE as shown in the run summary section above

Note: If repeated “Invalid” results are obtained contact BioFire Diagnostics, the local bioMéri-eux sales representative, or an authorized distributor.

The Results Summary Section lists the test results for each organism targeted by the Meningitis / Encephalitis Panel.

5. Run Status: — If Completed, run is complete. — If Incomplete, Aborted, Instrument Com-munication Error, Instrument Error, or Software Error, RETEST SAMPLE

Note: If repeated “Error” messages are obtained contact BioFire Diagnostics, the local bioMérieux sales representative, or an authorized distributor.

The Run Details Section displays information about the pouch, instrument, run status, and operator.

®