Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School Monday, September 23 rd 2013 Harvard Medical School Exchange Clerkship Program Primary Care Radiology Clerkship Gillian Lieberman, M. D. 1

Upload: vanxuyen

Post on 16-Feb-2018

220 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School

Monday, September 23rd 2013

Harvard Medical School Exchange Clerkship Program Primary Care Radiology Clerkship Gillian Lieberman, M. D.

1

Page 2: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Identify the sources of various types of artifacts in Computed Radiography images.

Orient on how they can be remediated.

MAH = Mount Auburn Hospital, Cambridge, MA

2

Page 3: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Wednesday, October 16, 2013 3

First, let’s see an example of an artifact: index patient 1.

Page 4: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Wednesday, October 16, 2013

4

Page 5: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Wednesday, October 16, 2013

5

Easy to tell – the opacity is outside of the patient.

Easy to tell – the opacity is outside of the patient.

Easy to tell – the opacity is outside of the patient.

Wait, what is this?

Page 6: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Wednesday, October 16, 2013 6

The patient was sent to the fluoroscopy room for evaluation and extraction of a possible foreign body located in the soft tissue anterior to the distal left tibia.

Page 7: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

7

Fluoroscopy

Page 8: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

8

There’s no foreign body. It was an artifact.

Fluoroscopy

Page 9: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Wednesday, October 16, 2013

9

New x-ray taken after the fluoroscopy, with a different x-ray cassette.

Page 10: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Wednesday, October 16, 2013

10

New x-ray taken after the fluoroscopy, with a different x-ray cassette.

There’s no foreign body. It was an artifact.

Page 11: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

11

Now let’s see another patient with a more serious case.

Page 12: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

52 year-old woman who had just underwent a pelvic surgery. After the surgery, there was a needle missing (incorret needle count). An x-ray was performed to check whether the needle had been forgotten inside the patient. All information provided to the radiologist was: “Pelvic surgery. Incorrect needle count.”

12

Page 13: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

13 PACS, MAH

Page 14: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

14 PACS, MAH

Page 15: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

15

The linear opacities in the previous slide do not have exactly the density we would expect from a metal object, but their shape and location (specially the pelvic opacity) raise enough suspicion. If the opacity could not be confirmed to be an artifact, the patient would need to be re-operated, exposing her to more risk, and exposing the hospital and staff to malpractice litigation.

Page 16: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

16

The attending radiologist then ordered two new x-rays, one of the abdomen, one of the pelvis, to re-evaluate for both linear opacities.

Page 17: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

17 PACS, MAH

Page 18: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

18 PACS, MAH

The linear opacity between the 11th and 10th rib has dissapeared. It was an artifact.

Page 19: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

19 PACS, MAH

Page 20: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

20 PACS, MAH

The linear opacity at the pelvis is still there. This greatly increases the likelihood of it being an actual needle.

Page 21: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

21

Before telling you the end of this case, let’s talk about the technology called “Computed Radiography”.

Page 22: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

22 Mount Auburn Hospital – Cambridge, MA

Page 23: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

COMPUTED

Uses the same equipment used for film x-rays, except for the film, which is replaced by an imaging plate that can be erased and reused for several years.

Has technical advantadges over film radiography, as well as some extra possible pitfalls which must be properly watched for.

DIGITAL

Fully digital: the x-rays are directly sensed by an electronic sensor.

Technically, it is the superior technology, achieving better images at lower radiation levels and with less dependency on the skills of the radiographer.

23

Page 24: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

COMPUTED

The majority of hospitals do not have a portable digital x-ray, so computed radiography is the choice for whenever it is not possible to bring the patient into the x-ray room:

Patients unable to move or be moved.

Patients in the emergency room.

DIGITAL

There is instant feedback from the sensors to the x-ray generator, so the machine is able to analyse the image and optimize the imaging parameters in real time. Because of this, better radiographs are obtained while exposing the patient to lower radiation levels.

24

Page 25: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

COMPUTED

The image is delivered to the PACS after being read by the CR scanner.

Is cheaper than digital radiography and its results are already good enough and better than conventional film.

DIGITAL

The image is delivered to the PACS instantly.

Is more expensive and its better results, in practice, are only of marginal benefit to either patients or radiologists.

25

Page 26: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

26

Let’s learn how an image is obtained through Computed Radiography.

Page 27: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011. 27

Page 28: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

The cassette is the box which contains an imaging plate instead of a conventional film.

The imaging plate absorbs the x-rays and acquires a latent image. If you just look at the imaging plate, you won’t see anything.

28

Page 29: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

After exposure to the x-rays, the operator brings the cassette to the CR scanner machine. Before doing that, it chooses the proper settings in the CR scanner for each image. Different body parts require different scanning settings.

The CR scanner machine opens the cassette and releases the imaging plate, which is rolled inside the machine by the rollers.

29

Page 30: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011. 30

Page 31: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

The imaging plate is exposed to a laser beam, causing the plate’s electrons to release photons which the photomultiplier tube is capable of sensing.

The light guide is similar in technology to an

optic fiber cable and leads the photons from the imaging plate to the photomultiplier tube.

31

Page 32: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

The photomultiplier tube is the actual sensor which produces an electric signal corresponding to the photons it receives. It can be imagined as a camera with only one pixel (do not confuse with “1 megapixel”).

In order to read millions of pixels from the imaging plate, a mirror redirects the laser beam in scan lines while the plate moves perpendicularly to those scan lines.

32

Page 33: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

The electric signals generated by the photomultiplier tube are electronically processed to a digital signal, which is then sent to the PACS – picture archiving and communications system.

33

Page 34: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

After the scanning is done, the CR scanner exposes the imaging plate to the light of an halogen lamp. This causes it to be erased.

Different types of x-rays require different exposures to the halogen lamp to ensure the image is completely erased. This is one of the reasons why the operator must configure the CR scanner according to what body part was x-rayed before inserting the cassette.

34

Page 35: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

Finally, the imaging plate is inserted back into the cassette and it CR scanner releases it for reuse.

35

Page 36: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

36

Now let’s see some examples of artifacts in Computed Radiography and how they can be addressed.

Page 37: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

The imaging plate

The CR scanner (plate reader)

Electronic processing of the image

Operator errors

37

Page 38: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

38 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Page 39: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

39

Dirt in the light guide

Remedy: Professionally service

the CR reader.

L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The British Journal of Radiology, 74 (2001), 195–202

Perfectly straight linear opacity extending all through the image.

Page 40: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

40 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 41: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

41

Dirt in the light guide and malfunctioning

rollers

Remedy: Professionally service

the CR reader.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Perfectly straight linear opacity extending all through the image. – Dirt in the light guide. Missing part of the

image, but notice that the missing part does interact with the image itself (the dragged black shadow below). – Malfunctioning of the rollers.

Page 42: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

42 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Page 43: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

43 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Patterned opacities extending through most or all of the image – they are the imprinting of the shape of the back of the cassette, which was exposed to the x-rays from the back an not from the front.

Exposure through the back of the cassette

Remedy: Ensure proper

education of the radiographers.

Page 44: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

44 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 45: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

45

Exposure through the back of the cassette

Remedy: Ensure proper

education of the radiographers.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Patterned opacities extending through most or all of the image – they are the imprinting of the shape of the back of the cassette, which was exposed to the x-rays from the back an not from the front.

Page 46: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

46 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 47: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

47

Delayed scanning

Remedy: Ensure plates are

scanned as soon as possible after exposure.

In 8 hours, 25% of

signals stored in the imaging plate are lost,

but this does not mean that the image will be

25% less visible. In this example, the cassette was inserted in the CR scanner 48 hours after

having been exposed to the x-rays.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Fading of the periphery of the image while the center seems adequately exposed.

Page 48: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

48 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 49: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

49

Scatter radiation

Remedy: Erase the cassette and

protect it from radiation.

Any cassette found to

be lying unattended in an x-ray room should be

erased before being used.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

The imaging plate is very sensible and will be affected by scattered radiation even if it is not directly under the source of x-rays. This image was taken from a cassette that was forgotten inside a room where x-rays were being performed. There was a pen over the cassette and its shape got imprinted on the imaging plate.

Page 50: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

50 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 51: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

51

Double exposure

Remedy: Proper knowledge on

the usage of x-ray equipment.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Two double-J stents are seen in the left kidney, because the imaging plate was exposed two times -- one during inspiration, another during expiration, making the kidney be at different heights at each exposure.

Page 52: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

52 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Page 53: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

53

Improper erasure

Remedy: Correctly select the

imaged body part before inserting the

cassette in the CR reader,

or replace the halogen bulb in the plate reader.

L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The British Journal of Radiology, 74 (2001), 195–202

Confusing opacities are seen around the image. The imaging plate wasn’t properly erased between two uses. The problem can be easily identified by noticing there are two identification stamps in the image, one in the upper right corner, another in the lower middle.

Page 54: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

54 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 55: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

55

Damage to the imaging plate caused

by the rollers

Remedy: Clean the rollers and replace the imaging

plate.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Damaged areas on the imaging plate become opaque in the image. The alignment of the linear opacities allow you to identify the rollers as the cause of that damage to the imaging plate.

Page 56: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

56 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 57: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

57

Dust on the rollers

Remedy: Clean the rollers.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

The scattered opacities easily mimick real findings and can be difficult to identify as artifacts. The opacity pointed by the black arrow could be a gallstone. A new x-ray, with a different cassete (and thus different imaging plate), should help identify the findings as artifacts.

Page 58: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

58 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 59: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

59

Dust over the imaging plate

Remedy: Clean the imaging plate.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Again, the scattered opacities might mimick real findings and can be difficult to identify as artifacts. A new x-ray, with a different cassete (and thus different imaging plate), should help identify the findings as artifacts.

Page 60: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

60 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Page 61: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

61

Improper use of grid

Remedy: Use grids with no less than 60 lines

per cm, and positioned perpendicular to the plate reader’s

scan lines.

L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The British Journal of Radiology, 74 (2001), 195–202

Special grids are used to cover the cassette during exposure to increase the image sharpness by filtering scattered rays while allowing collimated rays to pass to the imaging plate. However, if the grid lines are paralel to the scan lines in the plate reader, and of lower density than the scan lines of the reader, the artifacts shown here can appear.

Page 62: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

62 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 63: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

63

Damage to the lead foil in the back of the cassete

Remedy:

Replace the cassette.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

The back of the cassette is covered in lead foil to prevent scattered radiation from hitting the imaging plate from the back. If there is damage to that lead foil, the scattered radiation will enter and produce lucencies in the image. In this example it is easy to tell the lucent lines are artifacts, but in other cases they might get misdiagnosed as fractures.

Page 64: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

64 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Page 65: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

65

Inadequate image processing

Remedy: Select special processing settings

for images with wide dynamic ranges.

Re-imaging might be necessary.

L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The British Journal of Radiology, 74 (2001), 195–202

The prosthesis’ (metal) density is too high, making it difficult to see the interface between the prosthesis and the bones. Special processing settings may be able to compensate for that and provide visible detail in both prosthesis’ and bones’ density ranges.

Page 66: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

66 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Page 67: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

67

Memory or transmission problems

Remedy: Professionally service the electronic

equipments.

L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The British Journal of Radiology, 74 (2001), 195–202

Notice that the artifacts seem to hold no relation to the image itself: the black stripes over the dark background are as black as the ones over the whiteness of the bones. This allows the identification of the electronic nature of the problem.

Page 68: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

68 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 69: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

69

Damage to the imaging plate

Remedy: Discard the imaging

plate.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

These are the most difficult to identify. The opacity over the scapula could be mistaken as a parasitic calcification. A new x-ray, with a different cassette and thus different imaging plate, should help identify the findings as artifacts.

Page 70: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

70 L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The

British Journal of Radiology, 74 (2001), 195–202

Page 71: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

71

Damage to the imaging plate

Remedy: Replace the imaging

plate.

L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The British Journal of Radiology, 74 (2001), 195–202

This time, the location of the artifact incidentally makes it easier to suspect it is an artifact. However, a new x-ray, with a different cassette and thus a different imaging plate, should be ordered for proper confirmation.

Page 72: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

72 Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Page 73: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

73

Damage to the imaging plate

Remedy: Replace the imaging

plate.

Shetty et al. Computed Radiography Image Artifacts Revisited. AJR:196, January 2011.

Can be misdiagnosed as extravasation of contrast at the right renal pelvis.

Can be misdiagnosed as a calculus in the ureter.

A new radiograph, using a different cassette and thus a different imaging plate, will help identify the opacities as artifacts. Physical buckles in the imaging plate are unrecoverable – the plate will need to be discarded.

Page 74: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

74

Now let’s go back to our index case.

Page 75: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

75

“52 year old woman, pelvic surgery, incorrect needle count.”

Page 76: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

76 PACS, MAH

Page 77: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

77 PACS, MAH

This is already the second x-ray and the opacity is still there, in the same place.

Page 78: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

78

The attending radiologist then ordered a third x-ray, but with the following requests:

1. Remove the cushioning from under the pelvis of the patient.

2. Use a new cassette that wasn’t used before.

3. Do not use any grid to cover the cassette.

Page 79: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

79 PACS, MAH

Page 80: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

80 PACS, MAH

The linear opacity has disappeared from the pelvis. It was an artifact. No re-operation is needed.

Page 81: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

81

Fabrício Sampaio Péres Kury Federal University of Rio de Janeiro Medical School

September 23rd 2013, Beth Israel Deaconess Medical Center

Flamengo Park and Sugarloaf Mountain, Rio de Janeiro, Brazil Image source: http://www.flickr.com/photos/luliweb/2362204197/

Page 82: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

1. L J Cesar, B A Schueler, F E Zink et al. Artefacts found in computed radiography. The British Journal of Radiology, 74 (2001), 195–202.

2. Shetty et al. Computed Radiography Image Artifacts Revisited. American Journal of Roentgenology:196, January 2011.

3. Charles E. Willis, PhD, DABR; Stephen K. Thompson, MS, DABR; S. Jeff Shepard, MS, DABR. Artifacts and misadventures in digital radiography. Paper from the presentation at the Society of Computer Applications in Radiology Meeting. SCAR University Course 305, 20th Symposium. Boston, MA, June 7–10, 2003

4. Rowlands , J A, The physics of computed radiography. Phys. Med. Biol. 47 (2002) R123–R166

82

Page 83: Fabrício Sampaio Péres Kury Federal University of Rio de ...eradiology.bidmc.harvard.edu/LearningLab/gastro/Kury.pdf · Fabricio Sampaio Peres Kury

1. Gillian Lieberman, M.D, from Harvard Medical School. 2. J. Pierre Sasson, M.D., Ralph L. Reichle, M.D., George L. Cushing, Jr.,

M.D., and the people at the Radiology department of Mount Auburn Hospital.