Information presented is subject to change as product enhancements are made. Contact ETS-Lindgren Sales Department for current specifications. Photographs are for illustration purposes only and do not represent all possible configurations. 7/02 - 1k © 2002 ETS-Lindgren REV C USA: Tel +1.512.531.6400 Fax +1.512.531.6500 FINLAND: Tel +358.2.838.3300 Fax +358.2.865.1233 SINGAPORE: Tel +65.65.36.7078 Fax +65.65.36.7093 ONLINE: [email protected] www.ets-lindgren.com UK: Tel +44(0)1438-730700 Fax +44(0)1438-730750 .)+6# 5 Meter EMC Anechoic Chambers Features 26 MHz - 18 GHz Frequency Range Full Compliance Testing for Radiated Emissions: - ANSI C63.4 (2000) Part 15 & 18 - EN 50147-2 - CISPR 11/ EN55011 - CISPR 16/ EN55016 - CISPR 22/ EN55022 - VCCI V-3/97.04 (1997) - SAE J551, RE & RI --Requires chamber modifications to accommodate full vehicle Full Compliance Testing for Radiated Immunity: - IEC 61000-4-3/EN61000-4-3 - ENV 50140 - SAE J-1113, RE &RI Eligible for FCC 3 Meter Class B Facility Filing Available as a Turnkey Package: Chamber, Tower, Turntable, Antennas, Instrumentation, and Software ETS-Lindgren FACT 5 TM Chambers offer the ability to perform radiated emissions testing at a 5 meter distance. While not required by any current standard, measurements made at a 5 meter distance have better correlation to those obtained at a 10 meter distance. While permitting 5 meter testing, these chambers preserve the ability to perform full compliance testing of applicable 3 meter standards. The chamber’s small overall size (only slightly larger than a FACT 3 TM chamber) results from the use of Rantec brand hybrid absorber and ferrite tile, which require less volumetric space than traditional absorbers. ETS-Lindgren FACT 5 chambers will fit in many existing structures, yet are large enough to perform 4 meter antenna scans above the ground plane. When installed with ETS-Lindgren’s LoPro TM or EuroPro TM turntables, pit excavations for motor assemblies are eliminated. ETS-Lindgren FACT 5 chambers are the perfect choice when facility space is limited, but improved 10 meter correlation and better measurement uncertainty are required. Performance ETS-Lindgren FACT 5 chambers achieve their broadband performance using a unique arrangement of hybrid and ferrite tile absorber. The design was modeled using the proprietary numerical electromagnetic software responsible for creating chambers that set new standards for anechoic performance. Radiated Emissions Testing ETS-Lindgren’s FACT 5 chambers can be used to perform full compliance 5 meter testing for ANSI C63.4-2000 (FCC part 15 & 18), EN50147-2, CISPR 11/EN55011, CISPR 16/EN55016, CISPR 22/EN55022, VCCI V-3/97.04 (1997), and SAE J551 requirements. At 3 and 5 meter range lengths, ETS-Lindgren guarantees a Normalized Site Attenuation (NSA) better than of ± 4.0 dB deviation from theoretical NSA within the quiet zone, over the frequency range of 30 MHz to 1 GHz. TM ETS-Lindgren FACT 5 TM EMC Semi-Anechoic Chamber

Upload: vominh

Post on 06-Feb-2018

231 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: EMC Anechoic Chambers 5 Meter - Maxtech · PDF file- CISPR 16/ EN55016 - CISPR 22/ EN55022 - VCCI V-3/97.04 (1997) - SAE J551, RE & RI--Requires chamber modifications to accommodate

Information presented is subject to change as product enhancements are made. Contact ETS-Lindgren Sales Department for current specifications. Photographs are for illustration purposes only and do not represent all possible configurations. 7/02 - 1k © 2002 ETS-Lindgren REV C

USA:Tel +1.512.531.6400Fax +1.512.531.6500

FINLAND:Tel +358.2.838.3300Fax +358.2.865.1233

SINGAPORE:Tel +65.65.36.7078Fax +65.65.36.7093

ONLINE:[email protected]

UK:Tel +44(0)1438-730700Fax +44(0)1438-730750

�������5 Meter

E MC An e c h o i c C h am b e r s

Features● 26 MHz - 18 GHz Frequency Range

● Full Compliance Testing forRadiated Emissions:- ANSI C63.4 (2000) Part 15 & 18- EN 50147-2- CISPR 11/ EN55011- CISPR 16/ EN55016- CISPR 22/ EN55022- VCCI V-3/97.04 (1997)- SAE J551, RE & RI --Requires chamber modifications

to accommodate full vehicle

● Full Compliance Testing forRadiated Immunity:- IEC 61000-4-3/EN61000-4-3- ENV 50140- SAE J-1113, RE &RI

● Eligible for FCC 3 Meter Class BFacility Filing

● Available as a Turnkey Package:Chamber, Tower, Turntable, Antennas,Instrumentation, and Software

ETS-Lindgren FACT 5TM Chambers offerthe ability to perform radiated emissionstesting at a 5 meter distance. While notrequired by any current standard,measurements made at a 5 meterdistance have better correlation tothose obtained at a 10 meter distance.While permitting 5 meter testing,these chambers preserve the ability toperform full compliance testing ofapplicable 3 meter standards.

The chamber’s small overall size (onlyslightly larger than a FACT 3TM chamber)results from the use of Rantec brandhybrid absorber and ferrite tile, whichrequire less volumetric space thantraditional absorbers. ETS-LindgrenFACT 5 chambers will fit in many

existing structures, yet are largeenough to perform 4 meter antennascans above the ground plane. Wheninstalled with ETS-Lindgren’s LoProTM

or EuroProTM turntables, pit excavationsfor motor assemblies are eliminated.

ETS-Lindgren FACT 5 chambers are theperfect choice when facility space islimited, but improved 10 metercorrelation and better measurementuncertainty are required.

PerformanceETS-Lindgren FACT 5 chambersachieve their broadband performanceusing a unique arrangement of hybridand ferrite tile absorber. The design wasmodeled using the proprietary

numerical electromagnetic softwareresponsible for creating chambers thatset new standards for anechoicperformance.

Radiated Emissions TestingETS-Lindgren’s FACT 5 chambers canbe used to perform full compliance 5meter testing for ANSI C63.4-2000(FCC part 15 & 18), EN50147-2,CISPR 11/EN55011, CISPR 16/EN55016,CISPR 22/EN55022, VCCI V-3/97.04(1997), and SAE J551 requirements.

At 3 and 5 meter range lengths,ETS-Lindgren guarantees a NormalizedSite Attenuation (NSA) better than of± 4.0 dB deviation from theoreticalNSA within the quiet zone, over thefrequency range of 30 MHz to 1 GHz.

��

ETS-Lindgren FACT 5TM EMC Semi-Anechoic Chamber

Page 2: EMC Anechoic Chambers 5 Meter - Maxtech · PDF file- CISPR 16/ EN55016 - CISPR 22/ EN55022 - VCCI V-3/97.04 (1997) - SAE J551, RE & RI--Requires chamber modifications to accommodate

Information presented is subject to change as product enhancements are made. Contact ETS-Lindgren Sales Department for current specifications. Photographs are for illustration purposes only and do not represent all possible configurations. 7/02 - 1k © 2002 ETS-Lindgren REV C

USA:Tel +1.512.531.6400Fax +1.512.531.6500

FINLAND:Tel +358.2.838.3300Fax +358.2.865.1233

SINGAPORE:Tel +65.65.36.7078Fax +65.65.36.7093

ONLINE:[email protected]

UK:Tel +44(0)1438-730700Fax +44(0)1438-730750

�������5 Meter

E M C An e c h o i c C h a m b e r s

The quiet zone at the 3 and 5 meter testdistance is a cylinder up to 2.0 meter indiameter, following the volumetric testprocedure of ANSI C63.4-2000.

The FerroSorbTM hybrid and FT seriesferrite tile absorbers used in this chamberto also guarantee NSA performance ofbetter than ± 4.0 dB deviation fromtheoretical NSA using the test methoddefined in ANSI C63.4-2000 at theextended frequency range of 1 to 40 GHz.

Radiated Immunity TestingETS-Lindgren’s FACT 5 chambers canalso be used to perform full compliancetesting for IEC 61000-4-3/EN61000-4-3,ENV 50140, and SAE J-1113 requirements.

At 3 meter range lengths, fielduniformity of 0, + 6 dB is achieved inthe test aperture over the frequencyrange of 26 MHz to 1 GHz. The testaperture is a vertical plane 1.5 m x 1.5 mat an elevation of 0.8 meter to 2.3 mabove the ground plane, following thetest procedure of IEC 61000-4-3.

Anticipating tomorrow’s requirements,an optional absorber treatmentpackage is available which permitsradiated immunity testing atfrequencies up to 18 GHz with 0,+ 6 dB field uniformity, following thetest procedure of IEC 61000-4-3.

Additionally, Rantec brand FerroSorbTM

can safely withstand continuous fieldintensity of up to 200 V/m andintermittent field intensity of up to500 V/m. This safely exceeds the fieldintensity requirements of mostcommercial RI tests.

Turnkey SystemsETS-Lindgren is an integratedmanufacturer producing such well-known brands as Rantec absorber,Series 81 and Euroshield shielding andRF doors, EMCO antennas, towers,turntables, and test accessories. All ofour products are designed to worktogether for maximum efficiency andperformance.

In addition to providing performance,our “Total Solution, Single Source”TM

approach simplifies complex decisionsand gives you peace of mind with asingle source of responsibility.

Baseline Configurations for2 m and 3 m Quiet Zones● 2 m Quiet Zone:

Design and fabricate RF-shieldedenclosure, interior nominal shield-to-shield dimensions of 10.36 m x 6.09 mx 5.94 m (34’ L x 20’ W x 19.5’ H) includes15.24 cm (6”) raised floor

● 3 m Quiet Zone:Design and fabricate RF-shieldedenclosure, interior nominal shield-to-shield dimensions of 11.58 m x 7.31 mx 6.09 m (38’ L x 24’ W x 20’ H) includes30 cm (12”) raised floor

● One single-leaf, ETS-Lindgren Series 201,recessed contact mechanism (RCM),manually operated, RF-shielded door,1.22 m x 2.13 m (4’ x 7’) clear opening

● One door maintenance kit

● Dielectric vapor barrier and masoniteunderlayment

● Raised, reflective ground plane, 15.24 cm(6”) or 30.84 cm (12”) high, depending onquiet zone size selected (2 m or 3 m)

● Four access hatches in raised floor, 30 cmx 30 cm (12” x 12”)

● Six high-hat light fixtures with integralwaveguide air vents (electrical distributionnot included) and two flood lights at theturntable area

● Six 2 x 30 amp, 60 Hz power-line filtersfor lights and EUT (wiring not included)

● Two connector panels, 15.24 cm (6”) x60.96 cm ( 24”) clear opening

● Three “N” type connectors

● One 3.81 mm (1.5”) pipe penetration withflange nuts and cap

● Two fiber optic ST-type feed-through kits

● One RF-shielded penetration for air lineto MiniMast™ (compressed air source notincluded)

● One threaded brass ground stud 1.27 cm(.5”) diameter x 12.7 cm (5”) long

● Shield test per MIL-STD-285, onefrequency test at 1 GHz

● Rantec brand PS-600 on the chamberceiling and two side-walls

● Rantec brand PS-400 or PS-600 on the twoend-walls

● Light reflective covers for PS-600 -absorber on the walls

● Light reflective covers over bare ferrite tile

● 16 pieces Rantec brand FT-100c ferrite tilepanels for an area of the floor forradiated immunity testing up to 1 GHz

● 16 pieces Rantec brand EHP-18PCL on anarea of the floor for radiated immunitytesting from 1 GHz to 18 GHz.

● Installation of the enclosure and absorber

● Field uniformity calibration perIEC 61000-4-3 from 26 MHz to 1 GHz(optional testing to 18 GHz)

� �

Page 3: EMC Anechoic Chambers 5 Meter - Maxtech · PDF file- CISPR 16/ EN55016 - CISPR 22/ EN55022 - VCCI V-3/97.04 (1997) - SAE J551, RE & RI--Requires chamber modifications to accommodate

Information presented is subject to change as product enhancements are made. Contact ETS-Lindgren Sales Department for current specifications. Photographs are for illustration purposes only and do not represent all possible configurations. 7/02 - 1k © 2002 ETS-Lindgren REV C

USA:Tel +1.512.531.6400Fax +1.512.531.6500

FINLAND:Tel +358.2.838.3300Fax +358.2.865.1233

SINGAPORE:Tel +65.65.36.7078Fax +65.65.36.7093

ONLINE:[email protected]

UK:Tel +44(0)1438-730700Fax +44(0)1438-730750

Recommended Test Equipment

AntennasImmunity Only3109 Biconical, 3148 Log Periodic,3142B BiConiLogTM with optional endcaps3140 BiConiLogTM

Emissions Only3110B Biconical3124 BiCalTM Precision Calculable Biconical3142B BiConiLogTM

3148 Log Periodic

Immunity and Emissions3104C Biconical (50 W max.)3142B BiConiLogTM with optional endcaps3148 Log Periodic

Turntables2081-2.03, Heavy Duty, 2.0 m diameter*2081-3.03, Heavy Duty, 3.0 m diameter*2088-2.03, EuroPro, 2.0 m diameter* Model 2081 requires 18” raised floor

Tripods and TowersImmunity Only4-TR Tripod7-TR Tripod7-TR/POL Tripod

Immunity and Emissions2070-1 Antenna Tower2070-2 Antenna Tower2071-2 Antenna Tower

�������5 Meter

E MC An e c h o i c C h a m b e r s

● Normalized site attenuation test perANSI C63.4-2000, from 30 MHz to 1 GHz(optional testing to 18 GHz)

● Turntable for 2 m Quiet Zone:ETS-Lindgren Model 2088 electricallypowered, 2 m turntable with cable andfiber optic feed-through

● Turntable for 3 m Quiet Zone:ETS-Lindgren Model 2087-3.03electrically powered, 3 m turntable withcable and fiber optic feed-through

● ETS-Lindgren Model 2090 dual devicecontroller

● ETS-Lindgren Model 2075-2 MiniMast™electrically powered air polarizationboom tower with 10 m cable and fiberoptic feed-through.

● Guaranteed performance and a five-yearlimited warranty; one-year warranty ondoors, filters and moving parts; two-yearwarranty on optional ETS-Lindgrenequipment.

� �

Options:● Shielded control room, 4.87 m x 3.05 m

x 2.44 m (16 ft x 10 ft x 8 ft)

● Compact-sized shielded control room,3.66 m x 3.05 m x 2.44 m (12 ft x 10 ft x 8 ft)

● Shielded amplifier room, 2.44 m x 2.44 mx 2.44 m (8 ft x 8 ft x 8 ft)

● 30.48 cm (12”) and 45.75 cm (18”) raisedfloors

● Low profile door sill for Series 201 door

● Fire detection and suppression system

● Electrical distribution

● Heating, ventilating, and air conditioning(HVAC) system

● Seismic structural design calculations andcertification

● Radiated Immunity testing 1 GHz to 18 GHz

● NSA testing to 1 GHz to 40 GHz

● CCTV monitoring system

● Intercom system

● Fire detection and suppression system

● Low profile door sill for Series 201 door

● Other door options available

● Anti-static vinyl floor tile

● Extra filtering

● Company logo screen printed on white caps

● Immunity interlock switch

● Assistance in preparation of FCC filingpaperwork (class B)

Page 4: EMC Anechoic Chambers 5 Meter - Maxtech · PDF file- CISPR 16/ EN55016 - CISPR 22/ EN55022 - VCCI V-3/97.04 (1997) - SAE J551, RE & RI--Requires chamber modifications to accommodate

Information presented is subject to change as product enhancements are made. Contact ETS-Lindgren Sales Department for current specifications. Photographs are for illustration purposes only and do not represent all possible configurations. 7/02 - 1k © 2002 ETS-Lindgren REV C

USA:Tel +1.512.531.6400Fax +1.512.531.6500

FINLAND:Tel +358.2.838.3300Fax +358.2.865.1233

SINGAPORE:Tel +65.65.36.7078Fax +65.65.36.7093

ONLINE:[email protected]

UK:Tel +44(0)1438-730700Fax +44(0)1438-730750

Specifications for 5 m Chamber with 2 m Quiet Zone

2 m ● PS-600 on two 26 MHz ANSI C63.4-2000, U.S. FCC (L) 9.6 m 10.4 m 11.10 m 26 MHz - 18 GHz 30 MHz - 40 GHzQuiet side-walls to EN50147-2 parts 15&18 31.4 ft 34.0 ft 36.42 ft 1.5 m x 1.5 mZone and ceiling 40 GHz VCCI V-3/97.04 (1997) (W) 4.9 m 6.1 m 6.68 m 0 +6 dB

● PS-400 on two CISPR 11/EN55011 16.1 ft 20.0 ft 21.92 ft 75% 16 Pointsend-walls CISPR 16/EN55016 (H) 5.3 m 5.9 m 6.41 m (Field Uniformity

● Light reflective covers CISPR 22/EN55022 17.5 ft 19.5 ft 21.03 ft above 18 GHzfor PS-600 IEC 61000-4-3/EN61000-4-3 Includes Includes may be limited

● 16 pieces ea. of FT-100c and Bell Core GR-1089 15.24 cm (6”) 15.24 cm (6”) by the antennaEHP-18PCL on area of the raised floor raised floor gain pattern)floor for immunity testing

MODEL ABSORBER FREQ APPLICABLE INTERNAL SHIELD OVERALL IEC 61000-4-3 ANSI C63.4# TYPE RANGE STANDARDS WORKING ROOM STRUCTURE FIELD NORMALIZED SITE

AREA INNER DIM DIM UNIFORMITY ATTENUATION

Specifications for 5 m Chamber with 3 m Quiet Zone

3 m ● PS-600 on two 26 MHz ANSI C63.4-2000, U.S. FCC (L) 10.8 m 11.6 m 12.2 m 26 MHz - 18 GHz 30 MHz - 40 GHzQuiet side-walls to EN50147-2 parts 15&18 35.4 ft 38.0 ft 40.42 ft 1.5 m x 1.5 mZone and ceiling 40 GHz VCCI V-3/97.04 (1997) (W) 6.1 m 7.3 m 7.90 m 0 +6 dB

● PS-400 on two CISPR 11/EN55011 20.1 ft 24.0 ft 25.92 ft 75% 16 Pointsend-walls CISPR 16/EN55016 (H) 5.5 m 6.1 m 6.56 m (Field Uniformity

● Light reflective covers CISPR 22/EN55022 18.0 ft 20.0 ft 21.53 ft above 18 GHzfor PS-600 IEC 61000-4-3/EN61000-4-3 Includes Includes may be limited

● 16 pieces ea. of FT-100c and Bell Core GR-1089 30.5 cm (12”) 30.5 cm (12”) by the antennaEHP-18PCL on area of the raised floor raised floor gain pattern)floor for immunity testing

MODEL ABSORBER FREQ APPLICABLE INTERNAL SHIELD OVERALL IEC 61000-4-3 ANSI C63.4# TYPE RANGE STANDARDS WORKING ROOM STRUCTURE FIELD NORMALIZED SITE

AREA INNER DIM DIM UNIFORMITY ATTENUATION

�������5 Meter

E M C An e c h o i c C h a m b e r s� �

1 In critical area around EUT.