To call personnel to the place where the call was triggered from
To search for personnel or groups of people
To forward information (e.g. from medical devices)
A call system is a visual/acoustic system that can be used:
What are call systems ?
ELSO GmbH 4- Mai 2008
DIN 0834 areas of application:●Old people's homes●Assisted living●Medical practices●Dialysis stations●Hospitals●Adult day-care centres
Patient call system for hospitals and nursing facilities
ELSO GmbH 5- Mai 2008
Call systems -> DKE subcommittee "General signalling systems and signalling devices" is responsible
Responsible for the DIN VDE 0834 standard "Light call systems; installing and operating systems in hospitals, nursing homes and
penal institutions"
Electrical engineering standards are created by the DKE (German Electrical Engineering Commission) -> Subsidiary of the VDE and the German Standards Committee
Standards and regulations for call systems
ELSO GmbH 6- Mai 2008
Systems for domestic use (home emergency call systems) are subject to the EN 50134 standard
Emergency call systems used for ensuring the safety of persons living at home
Excepted from the DIN VDE 0834 standard
ELSO GmbH 7- Mai 2008
Presence buttons with acoustic signallers (call forwarding) must be installed in rooms where nursing staff are usually to be found
Triggering of emergency calls when present with appropriate signalling
Visual display of triggered call -> reassurance light
Call button clearly marked and visible in the darkEvery bed has a call button -> easily reachable
A room signal light (RSL) must be installed outside every room with a red + green display and an optional white display (WC call)
Key features of the DIN VDE 0834 standard
ELSO GmbH 8- Mai 2008
Nursing staff must receive information within 5 s
Non-visible rooms (wet rooms) require a separate cancel button
Text displays must be recognisable from 20 m at 500 lux
A call must be displayed on the RSL within 1 second
The light must be clearly visible at 500 lux
No common covers for call / operating devices and high voltage devices
The presence button is used for cancelling the call
Key features of the DIN VDE 0834 standard
ELSO GmbH 9- Mai 2008
System voltage may only be used for the intended purpose Exception:
An impulse relay switch for a reading lamp or similar.
This must be permanently connected and have its own overvoltage protection.
Non-intercom systems -> no remote cancel -
but the acoustic call forwarding may be cancelled.
For call systems with intercoms, calls may be remotely cancelled
whenvoice communication has occurred
Transmission paths for other systems may not be used for call systems (separate cable)
System voltage less than 30 V AC or 60 V DC
Key features of the DIN VDE 0834 standard
ELSO GmbH 10- Mai 2008
All protective grounds connected to the call system must be connected to the same main potential equalisation point. If this is not possible then the individual areas must be galvanically isolated.
Cables connecting buildings must be equipped with overvoltage protection at the exit points
Call systems require emergency power no later than 15 s after the normal power has failed. Must be maintained for 1 h / calls must
be
stored for at least 30 s after a power failure.
System voltage less than 30 V AC or 60 V DC
Key features of the DIN VDE 0834 standard
ELSO GmbH 11- Mai 2008
●Operating device
(call button): 0.7 –
1.5 m
●Operating elements
for wheelchairs: 0.85 m (as per DIN 18024-2 Barrier-free construction)
●Pull button: must be
at
least 20 cm above
the highest
spray head position
●Pull cord
must be
reachable
by a person
lying
on the floor
●Operating devices
with
text
display: 1.5 m –
1.8 m not subjected
to direct sunlight
●Lights / Hall displays: 1.3 m –
2.2 m
Installation heights
ELSO GmbH 12- Mai 2008
●An operating log must always
be
available
at
the system
●All faults
must be
documented, with
date/time, cause, author
and repair measures
●Repairs
must begin
within
24 h of a fault
occurring
and this
must also be
documented
in the operating log
●Unless
otherwise
defined
by the manufacturer, maintenance must be performed
quarterly
and documented
in the operating log
Maintenance / Repair
ELSO GmbH 13- Mai 2008
●Conceived
for large and small
applications●Max. 59 stations, each
with
99 rooms
= 5841 rooms
●Consists
of different
independent
microprocessor
controlled
devices, connected
via a bus system
●Every
module has a microcomputer●The modules operate
autonomously, which
allows
easy
expansion of
the system●Data traffic
is
coded
with
addresses
●Permanent independent
self-monitoring of the microcomputers, which allows
faults
and/or dropouts
to be
individually
recognised
and signalled
●Programming
without
special
software, simple cabling● In the case of power cuts, the last information state of the modules is
retained●With
and without
room intercoms
System characteristics
ELSO GmbH 14- Mai 2008
●Power supply
/ UPS●Station transfer
point●Wireless transmitter
/ receiver●PC with
SW for logging●Call-forwarding
to DECT●Telephone
dialler
SMScontroller
●Connection
to monitoring device
Input devices Display devices Additional devices
●Call
cancel
button●Cancel
button
/WC●Pull button●Pear
button
/ light button●Call
button
with
socket●Socket●Pneumatic
button
●Designer lights●Single sided
corridor
display●Double sided
corridor
display●Duty
room
display●Room
display●Room
intercom
Product range
ELSO GmbH 15- Mai 2008
Room bus:
Station bus:
Floor bus:
Connects all devices in a room (incl. bathroom) e.g.: Call cancel button, WC cancel button, pull button pear button / light switch, call button with socket socket, pneumatic button
Connects all room modules
Connects all station transfer points
Bus types
ELSO GmbH 16- Mai 2008
Basic structure
Ground
floor
1st floor
Floor
bus
Station bus
Station bus
Installation
Room
bus
Room
bus
Station transfer
point
Station transfer
point
ELSO GmbH 17- Mai 2008
●These
4 conductors
are to be
connected
to the terminals
of the same
designation
on the room module (different
versions possible) in exactly
the manner
shown
here. ●This connection
can
be
a room bus, implemented
in a star configuration or mixed. The cable
length
(sum
of all cables) must not exceed
100m.●A cable
of type: JY (St) Y 2x2x0.8 must be
used
for the room bus. ●The shield
is
to be
connected
to the GND terminal of each
room module. To avoid
short circuits at
the device, the shield
must be
insulated.
All light call system devices
in a room are connected
to the room bus. The room bus has 4 conductors.
"d" = data"t" = clock"+24" = + 24VDC"GND"= ground
Room bus description
Room
bus
Station bus
Room
bus
Shield
Room
data
bus
24 V DC
WC
Station bus
ELSO GmbH 18- Mai 2008
All the room modules for the station must be connected to the station bus. Stub cables are to be avoided. The station bus should be laid as a ring cable. ELSO GmbH recommends closing the ring. This results in double data reliability. Please make absolutely sure that the conductors ("a"; "b"; "+24VDC" and "GND") are not swapped. The station bus cable (subsequently referred to as SYK=system cable) must be of type: Lapp Kabel UNITRONIC Bus Combi EIB 2x2x0.8 + 3x1.5, or equivalent quality. The shield must be connected to each of the GND terminals. To avoid short-circuits in the device, the shield must be insulated.The data cable must not exceed a length of 500m.
Station bus description
RM = room moduleST = station transfer pointPS = power supply
Further stations
ELSO GmbH 19- Mai 2008
Room modules are contained in:
Duty room module Room module with room indicator light
Room module with call cancel button
(also available with a display)
Installation
A room module is required for every room.
ELSO GmbH 20- Mai 2008
Room module
2x2x0.8mm2
with room signal light
Room bus
All devices are connected to the room bus. The room bus has 4 conductors.
"d" = data"t" = clock"+24" = + 24VDC"GND"= ground
Installation Room installation using a room module with room signal light
WC
ELSO GmbH 21- Mai 2008
+ 24VDC Ground
Data Clock
BusPower supply
Connection example:
Installation
ELSO GmbH 22- Mai 2008
Terminal Pin FunctionX1 Station bus
1 +24V 2 a (data "a" station bus)3 b (data "b" station bus)4 GND
X2 Room bus
1 +24V2 +24V3 d, room bus data
(terminal designation: BL-B/d)
4 t, room bus clock(terminal designation: R-B/t)
5 GND6 GND
X3 Inputs
1 +24V2 R-Z, room call3 S1 not used4 R-W, not used5 A-W, not used
X10 Outputs
1 TG acoustic alarm, call forwarding
2 ZL-R, (red) room call3 ZL-W, (white)WC call 4 ZL-S, (green) nurse5 GND
For quiescent current monitoring of the call cables, all external call inputs (R-Z and R-W) must be terminated with a 3.6 k termination resistor. This terminator is installed at the factory and must not be removed.
Room module with room indicator light
InstallationRoom module 735040
ELSO GmbH 23- Mai 2008
All light call system devices in a room are connected to the room bus. The room bus has 4 conductors -> Terminal strip X2.
"d" = data = "BL-B/d""t" = clock =
"R-B/t""+24" = + 24VDC=
"+24V""GND"
= Ground
=
"GND"
Installation
BUS data cable
Room bus
Room module with RSL
Station bus
ELSO GmbH 24- Mai 2008
Room module
2x2x0.8mm2
with call cancel button
Room bus
Installation Room installation using a room module with call cancel button
(also available with a display)
WC
ELSO GmbH 25- Mai 2008
Installation
DOUBLE SWITCH SOCKET
ELSO GmbH 26- Mai 2008
All light call system devices in a room are connected to the room bus. The room bus has 4 conductors -> Terminal strip X2.
"d" = data "t" = clock "+24" = + 24VDC"GND"
= Ground
A distribution box is necessary when using an SYK system cable. This is provided with delivery (-> Terminals with connection cable and fuse -
> slow blow fuse 250V / 2.5A ).The power supply cables of the SYK are connected to the +24V and
GND distribution box terminals. The red output cable of the distribution box is connected to +24V, the blue to GND.The supplied system cable 2x2x0.8 must be used for connecting to
the station bus.
Installation Room module with call cancel button and display
Room bus
data
Station bus
data
Twisted-pair cable
To station bus
From station bus
ELSO GmbH 27- Mai 2008
Installation Room module with call cancel button and display
White
WC call
Red
Room or bed call
Green
Nurse present
ELSO GmbH 28- Mai 2008
Room module
2x2x0.8mm2
Duty room module with call cancel button and display
Room bus
Installation Duty room module with call cancel button and display
WC
ELSO GmbH 29- Mai 2008
Description
Traine
e nurs
e pres
ence
butto
nNurs
e call
/
Nurse e
mergen
cy ca
ll
butto
n
Doctor
emerg
ency
call
butto
n
Functi
on bu
tton 2
Speak
er
Functi
on bu
tton 1
Microp
hone
Nurse p
resen
ce bu
tton
Functi
on bu
tton 4
Doctor
pres
ence
butto
n Display
Functi
on bu
tton 3
ELSO GmbH 30- Mai 2008
Representation of floor busForeign system
Station transfer point
Station transfer point
Shield Data aData b
Speak a1/b1
Speak a2/b2
Floor bus
ELSO GmbH 31- Mai 2008
All presence markings are summarised as a collective message for
forwarding, the display of the presence occurs at all buttons simultaneously in parallel. (6 times due to the bed-wise installation) The WC call cancelsimultaneously affects all triggered WC calls for the room. The acoustic alarm is only activated for devices having an address of 0.
Address Function 0 Presence S1-1 + acoustic call
forwarding1 Presence S1-22 Presence S1-33 Presence S1-44 Presence S1-55 Presence S1-66 Presence S1-77
8 WC call cancel 19 WC call cancel 2A WC call cancel 3B WC call cancel 4C
D
E
F
Presence cancel button
ELSO GmbH 32- Mai 2008
All calls are summarised as a collective message for forwarding. The reassurance light only illuminates at the device from which the call was triggered. All presence markings are summarised as a collective message for forwarding, the display of the presence occurs at all buttons simultaneously in parallel. The WC call cancel simultaneously affects all triggered WC calls for the room. The acoustic alarm is only activated for devices having an address of 0.
Address Function 0 Room call 1 and presence S1-1 +
acoustic call forwarding
1 Room call 2 and presence S1-22 Room call 3 and presence S1-33 Room call 4 and presence S1-44 Room call 5 and presence S1-55 Room call 6 and presence S1-66
7 WC call 1 with cancel8 WC call 2 with cancel9 WC call 3 with cancelA WC call 4 with cancelB
C
D
E
F
Call presence / cancel button
ELSO GmbH 33- Mai 2008
All calls except the bed calls are summarised as a collective message for forwarding.
The reassurance light only illuminates at the device from which the call was triggered.
Address Function 0 Room call 11 Room call 22 Room call 33 Room call 44 WC call 15 WC call 26 WC call 37 WC call 48
9
A Bed call 1-3B Bed call 2-3C Bed call 3-3D Bed call 4-3E Bed call 5-3F Bed call 6-3
Call button
ELSO GmbH 34- Mai 2008
Address Function 01 Bed call 1-12 Bed call 2-13 Bed call 3-14 Bed call 4-15 Bed call 5-16 Bed call 6-17 Room call 28 Room call 39 Room call 4A Bed call 1-2B Bed call 2-2C Bed call 3-2D Bed call 4-2E Bed call 5-2F Bed call 6-2
A button address must never be assigned twice, i.e. the address "Room Call 2" must only exist once per room. This also applies to all other cases. Failure to do this will result in system malfunctions.However, a single bed can be assigned a maximum of 3 buttons with different sub-addresses. A call button with system socket provides two sub-addresses "-1" and "-2" (for example: "Bed call 1-1" and "Bed call 1-2"). Note: The call button also provides the sub-address "-3" (for example: "Bed call 1-3"). The pull button and pneumatic button also provide the sub-address "-1" (for example: "Bed call 1-1")
Call button with system socket
ELSO GmbH 35- Mai 2008
Power supply: 24 VDCCurrent consumption: 30 mA Degree of protection: IP 20Output loading:
max. 24VDC / 0.2A / cos = 1
Call button with system socket
Load 2 Load 1
Ext. Rel. 2
Ext. Rel. 1
ELSO GmbH 36- Mai 2008
Setting for jumper X4
Left "ON": (factory setting) 24 Volt supply switched on
Right "OFF": 24 Volt supply for the room module and the entire room installation is switched off. This should always be switched over for servicing / de-installation purposes!
Settings for jumpers X5a and X5b
Down: Factory setting: Both jumpers must connect the terminals BL-B/d and R-B/t in order to implement connection of devicesvia the room bus.
Up: Thus jumper setting must not be selected ->
Malfunction!
Room module with room indicator light
ELSO GmbH 37- Mai 2008
No. 1 2 3 4 5 6 7 81 Off On On On On On On On2 On Off On On On On On On3 Off Off On On On On On On4 On On Off On On On On On5 Off On Off On On On On On6 Off Off Off On On On On On7 Off Off Off On On On On On8 On On On Off On On On On9 Off On On Off On On On On10 On Off On Off On On On On11 Off Off On Off On On On On12 On On Off Off On On On On13 Off On Off Off On On On On14 On Off Off Off On On On On15 Off Off Off Off On On On On16 On On On On Off On On On17 Off On On On Off On On On18 On Off On On Off On On On19 Off Off On On Off On On On20 On On Off On Off On On On21 Off On Off On Off On On On
Room module with room indicator light Extract from the address table
ELSO GmbH 38- Mai 2008
DCF radio receiver
ELSO GmbH 39- Mai 2008
The radio clock is connected to the floor bus via the data cables "at" and "bt". It is essential that the conductors ("a“"; "b"; "+24VDC" and "GND") are not swapped.
DCF radio receiver
SignalErrorSends
ELSO GmbH 40- Mai 2008
DCF radio receiver
1: Ground2: + Operating voltage 5..15 V-3: DCF output
ELSO GmbH 41- Mai 2008
Corridor
display
ELSO GmbH 42- Mai 2008
The hall display (KDIS8 -> Hall display 8 characters) is
used
for displaying
calls and call requests
in light call systems. The display is
connected
to the station bus and reports the information from
station transfer
points and room modules. 8 characters
are available
on the display for showing
call information. The call type is
displayed
using
2 characters
and the call location using
6 characters. The call location is
represented
by the electrical
station and room address.Electrical
addresses
are displayed
in the following
format:
SS RR ZZSS : Station address
RR : Call type
ZZ : Room address
Corridor
displayP
lug
Text
dis
play
ELSO GmbH 43- Mai 2008
Level of urgency Call type
Displayed as (program version KDIS8-XX)Language Display
typeGerman Italian Hungarian Polish English1 Doctor call A M O L M Blinking
Diagnostic call D D D D D Blinking
WC call W W W W W Static
Room emergency call Z S S P R Blinking
Bed emergency call B L B M B Blinking2 Room call Z S S P R Static
Bed
call B L B M B Static
Telephone
call T T T T T Static3 Doctor
call
request Aà Mà Oà Là Mà Blinking
S1 call
request S1à I1à N1à S1à N1à Blinking
S2 call
request S2à I2à N2à S2à N2à BlinkingDoctor
call
request, foreign
station ßA ßM ßO ßL ßM Blinking
S1 call
request,
foreign
station ßS1 ßI1 ßN1 ßS1 ßN1 Blinking
S2 call
request, foreign
station ßS2 ßI2 ßN2 ßS2 ßN2 Blinking
Corridor display
ELSO GmbH 44- Mai 2008
Room intercom
...addition
al devices
Hall display
Duty
room
Foreign
system
Floor
bus
Foreign
system
Station transfer
point
Station transfer
point
Shield Data
aData
b
Speak
a1/b1
Speak
a2/b2
Room
101
Room
102
Room
bus
Station bus
2
Station bus
1
ELSO GmbH 45- Mai 2008
Many
thanks
for
your
attention