Page 1: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

Elisée™ 150Patient ManualEnglish

Making quality of care easy ResMed Ventilation solutions

Page 2: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

Making quality of care easy ResMed Ventilation solutions

Page 3: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

1.1 Definitions 11.2 User/owner responsibility 1

2 Medical Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

2.1 Purpose of the Elisée 150 22.2 Warnings 2

3 Description of the Elisée 150 . . . . . . . . . . . . . . . . . . . . . . . . . 3

3.1 Components of the Elisée 150 33.2 Front of the Elisée 150 43.3 Touch screen 43.4 Connecting the circuit 53.5 On/Off button 63.6 Rear view 6

4 Connection Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

4.1 Connecting the circuit 74.2 Electrical connection 84.3 Connecting an oxygen supply 104.4 Connecting the mask 114.5 Example of a fully-assembled system 11

5 How to use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

5.1 Starting up the ventilator 125.2 Starting ventilation 125.3 Stopping ventilation 135.4 Shutting down the ventilator 145.5 Viewing ventilation settings and measurements 155.6 Charging the battery 15

6 Alarms and Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . 16

7 Cleaning and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . 17

7.1 Maintenance of the dust filter 187.2 Maintenance of the exterior of the ventilator 187.3 Maintenance of your circuit 18

8 Appendices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

8.1 Technical specifications 198.2 Description of buttons and symbols 198.3 Travelling with your ventilator 218.4 Manual test (while ventilation is stopped) 21

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Contents

iii

Page 4: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

iv

Page 5: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

1 IntroductionThis manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It is not a substitute for the clinical manual.

1.1 Definitions

This manual contains special terms and icons that appear in the margins to draw your attention to specific and important information.

WARNINGAlerts you to possible injury.

CAUTIONExplains special measures for the safe and effective use of the ventilator.

Note: Is an informative or helpful note.

1.2 User/owner responsibility

The owner or user of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from:• The device being set up, operated or cleaned in a manner which does not

comply with the instructions provided• The device being set up, maintained or altered by unauthorised persons and/

or in a manner which does not comply with the instructions.Please read this manual carefully before use.

1Introduction

Page 6: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

2

2 Medical Information

2.1 Purpose of the Elisée 150

The Elisée 150 is a ventilator designed to be used by both adults and children, in the home or clinical environment. The ventilator is used with either a single or a double patient circuit.

2.2 Warnings

• This manual is provided with the ventilator you have received. It does not in any way replace the clinical manual supplied to your physician or clinical care provider.

• The advice contained in this manual does not replace the instructions given by your physician (or clinical care provider), who will already be familiar with the operation of the device through the clinical manual provided.

• The ventilator settings must be made by competent and trained staff under a doctor's supervision.

• The ventilator must be used with the accessories recommended by the manufacturer or your prescribing physician. The use of inappropriate accessories is likely to affect the operation of the ventilator.

• If you have any questions about setting up, operating or maintaining your ventilator or its accessories, contact your HME provider.

• The ventilator must be transported in its travel bag.• In the case of externally visible faults, cease using the ventilator.• In the event of functional problems (e.g. you find it difficult to breathe or

trigger an inspiratory phase), do not hesitate to contact your clinical care provider.

• To avoid the risk of electrocution, do not open the device casing. Repairs and internal servicing should only be performed by an authorised service agent.

• If there is interference on the electrical network, operate the ventilator on battery power.

• Your HME provider must ascertain the electromagnetic characteristics of the environment in which this ventilator will be used. In particular, your HME provider must ensure that:• When the ventilator is operated in proximity to other electrical devices,

including cell phones, there is no interference, and the ventilator performs correctly

• The ventilator is never placed on or under other devices• There is an adequate distance between the ventilator and other electrical

devices in your home.• In accordance with Directive 2002/96/EC concerning waste electrical and

electronic equipment, this ventilator must be sorted and disposed of separately from other types of rubbish. It must not be disposed of with ordinary municipal waste. Contact your HME provider for more information.

The above are general warnings. Other specific warnings and notes will be found throughout the manual.

Page 7: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

3Description of the Elisée 150

3 Description of the Elisée 150

3.1 Components of the Elisée 150

The picture below shows the set of components available from your clinical care provider:• Travel bag (1)• Elisée 150 ventilator (2)• Standard power cord (3)• Patient circuit (double circuit shown here) (4).

Figure 1: Elisée 150 components.

1

INT.

EXT.

2

3

4

Page 8: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

4

3.2 Front of the Elisée 150

The front has:• a touch screen (1);• an LED panel and an Alarm Silence button (2);• a retractable handle (3).

Figure 2: Front view.

3.3 Touch screen

You can use the buttons displayed on the device's touch screen to navigate through the menus of the Elisée 150.When you start up your Elisée 150, the start screen appears.The following information is displayed:

Figure 3: Start screen.

Note: See Section 8.2, “Description of buttons and symbols” on page 19 for more information on the various screens.

INT.

EXT.

1

2

3

Intern.

Mains

STARTVENTIL �1s

PS

Wednesd. February2811: 28: 31

P1:

P2: ACV

Ventilation programs configured by your physician.

Start ventilation button.

Type of electrical power supply.

Page 9: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

5Description of the Elisée 150

3.4 Connecting the circuit

Your ventilator will be fitted with a circuit support to connect a single or double circuit to, as prescribed by your physician.

Figure 4: Single circuit support.

Figure 5: Double circuit support.

Air outlet

Exhalation and proximal pressure line connections

Air outlet

Exhalation circuit support with valve

Page 10: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

6

3.5 On/Off button

The On/Off button, along with several electrical connection ports, is on the right side of your Elisée 150.

Figure 6: Right side.

3.6 Rear view

At the rear of the ventilator you will see:• The oxygen connection (if prescribed by your physician)• The air inlet.

Figure 7: Rear view.

WARNING• Do not block the air inlet, or the performance of your ventilator may be

compromised.

• Use of oxygen can lead to a risk of explosion. Consult your clinical care

provider on the necessary precautions.

12-28 V

15A max

On/Off button

Oxygen connection

Air inlet

Oxygen sensor

Page 11: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

7Connection Procedures

4 Connection Procedures

4.1 Connecting the circuit

Your system may be configured for either a single circuit or double circuit. For connection instructions, refer to the appropriate section for your air tubing configuration.

WARNINGOnly the circuit supplied by your HME provider may be used with your device. Using a different type of circuit may reduce the effectiveness of your treatment.

Connecting a single circuit

1. Connect the expiratory valve tube which has a connector on one end by pushing it firmly over the longer of the two pressure connectors on the

ventilator (a).2. Then push the tube without a connector firmly onto the shorter pressure

connector (b).

CAUTIONTake great care not to confuse the two tubes.

3. Then fasten the patient circuit firmly to the air outlet.

Figure 8: Connection of a single circuit.

a b

Depending on the type of patient circuit you are using, it may be necessary to cut off the tip of the proximal pressure line.

Page 12: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

8

Connecting a double circuit

1. Connect one limb of your double circuit to the air outlet.2. Secure the other limb of your circuit to the other connector.

Note: Other components in your patient circuit may include a humidification system, an antibacterial filter, a mask or cannulae, and water traps. Contact your clinical care provider if you have any concerns about connecting your patient circuit or accessories. Read the instructions provided with any accessories you are using (humidifier, filters, etc.).

Figure 9: Connection of a double circuit.

4.2 Electrical connection

Your ventilation may have one of two electrical connection configurations:• Standard electrical connectionor• External electrical connection (on the right side of the ventilator).

CAUTIONBe careful not to place the ventilator where it could be knocked over or

where someone may trip over the power cord.

Page 13: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

9Connection Procedures

Standard electrical connectionIf you are using the standard electrical connection located under your Elisée 150 (See Figure 11):

Figure 10: Standard power cord.

1. Insert one end of the cord into the connector underneath the ventilator.2. Push down the fastening clip on the connector.3. Insert the mains plug into the power supply.

Figure 11: Standard mains power connection.

CAUTIONRemember to push down the fastening clip to hold the power cord in place.

This will prevent accidental disconnection from the power supply.

Page 14: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

10

External electrical connectionIf you are using an external electrical connection as shown below:

Figure 12: External electrical connection.

1. On the right side of the device, insert the cord into the connector in the socket marked “ .”

2. Insert the other end of the cord into a power outlet.

Figure 13: External mains power cable connection.

4.3 Connecting an oxygen supply

Your Elisée 150 will be equipped with an oxygen connector should your prescribed treatment require oxygen.Connect the oxygen supply tube to the oxygen connector on the rear of the ventilator.

CAUTIONEnsure that the oxygen supply is switched off whenever oxygen is being

connected or disconnected.

12-28 V15A max

1

2

Page 15: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

11Connection Procedures

Figure 14: Connection of oxygen supply.

4.4 Connecting the mask

Refer to the instructions given by your physician or HME provider.

CAUTIONIt is essential to use only the mask supplied by your physician or HME provider.

4.5 Example of a fully-assembled system

Figure 15: Example of a fully-assembled system (single circuit with expiratory valve and proximal pressure line)

Oxygen supply: Max. pressure: 400 kPaMax. flow: 15 L/min

Page 16: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

12

5 How to useThis ventilator and its accessories must be operated in a dust-free environment and kept away from direct sunlight.The ventilator is a medical device. To ensure smooth operation, keep the ventilator out of reach of pets, and ensure children do not have unsupervised access to it.

5.1 Starting up the ventilator

After making all the required connections (patient circuit, accessories, tracheotomy tube or mask, electrical connections, etc.), press the On/Off button

on the right-hand side of the ventilator.

Figure 16: Switch the ventilator on.

5.2 Starting ventilation

When the ventilator starts up, the start screen is displayed. Press and hold the "START VENTIL" button for one second to start treatment.

Figure 17: Starting ventilation.

12-28 V

15A max

Intern.

Mains

STARTVENTIL �1s

PS

Wednesd. February2811: 28: 31

INT.

EXT.

Page 17: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

13How to use

Selecting a ventilation programYour physician may configure two different ventilation programs on your Elisée 150.To switch from one ventilation program to the other, simply press and hold the appropriate button. The selected program will turn black.

Figure 18: Selecting the ventilation mode.

5.3 Stopping ventilation

1. Press the On/Off button on the right-hand side of the ventilator.2. Press to stop ventilation.

Figure 19: Stopping ventilation.

Intern.

Secteur

ACV

PS

P1:

P2:

Wednesd. February2811: 28: 31

INT.

EXT.

< Ai

0

10

2 0

30

40

50

60

70

80

90

< PE

33.6 H2Ocm

INT.

EXT.

Retour

< Ai

0

10

2 0

30

40

50

60

70

80

90

< PE

33.6 H2Ocm

Pause (stop ventilation)

Stop machine

Page 18: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

14

5.4 Shutting down the ventilator

1. Press the On/Off button on the right-hand side of the ventilator.2. Depending on whether or not your ventilator is delivering ventilation: • During ventilation, press to turn off the ventilator:

Figure 20: Shutting down the ventilator during ventilation.

• When ventilation is not being delivered, press YES to turn off the ventilator:

Figure 21: Shutting down the ventilator when ventilation is not being delivered.

INT.

EXT.

Back

< Ai

0

10

2 0

30

40

50

60

70

80

90

< PE

33.6 H2Ocm

Pause (stop ventilation)

Stop machine

EXT.

Do you want to switch off?

YESYES NONO

INT .

Last test the 06/12/10Double circuit : PASS

TESTTESTDate displayed as Day/ Month/ Year

Page 19: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

15How to use

5.5 Viewing ventilation settings and measurements

To view the ventilation measurements and the settings programmed by your clinical care provider, press the ventilation program button. The following screen will appear:

Figure 22: A typical ventilation measurements and settings screen (while ventilation is being delivered).

5.6 Charging the battery

The battery of your Elisée 150 charges whenever your mains power cord is connected (See Section 4.2, “Electrical connection” on page 8), whether or not the ventilator is switched on.The table below explains the various battery LED displays.

Table 1: Battery LED on the LED panel.

0

10

2 0

30

40

50

60

70

80

90

< PE

┘HP

26 H2Ocm

PEEPPEEP4.2 H2O

cm 26 H2Ocm

VTITI

I/EI/E

500 ml VTETE

500 ml PpeakPpeak

VT

500 ml

PEEPPEEP

F FlowshapeFlowshape

TriggerTrigger3

1

Patient meter : 00000163H

15 bmp 1/2.0

4 H2Ocm

Symbol and LED status Meaning

Steady green light Battery charged.

Green light, flashing slowly Battery discharging.

Orange light, flashing slowly Battery discharging (battery low).

Red light, flashing slowly Battery discharging (battery flat).

Orange light, flashing rapidly Battery charging interrupted (battery temperature too high or too low).

Green light, flashing rapidly Battery charging.

Page 20: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

16

6 Alarms and TroubleshootingThe Elisée 150 is fitted with alarms to alert you to changes that will affect your treatment. Check that the patient circuit is correctly connected to your ventilator.Some alarms can be deactivated temporarily (for 2 minutes) by pressing the button.The alarm will be reactivated if the problem persists.

CAUTIONService and repairs should only be performed by an authorised service

agent. Do not attempt to service or repair the ventilator yourself.

The table below will help you identify the cause of the problem. If the fault persists or cannot be identified, do not try to open the device. Instead, contact your HME provider.

Table 2: Alarm, cause and solution.

Alarm icon Cause Solution

Internal battery power low. Connect ventilator to mains supply.

External battery power low. Connect ventilator to mains supply.

A component in your patient circuit is wrongly connected.

Check that your patient circuit is set up correctly as demonstrated by your clinical care provider, or replace it. Contact your clinical care provider if the alarm persists.

A technical fault has occurred in your ventilator.

Contact your clinical care provider.

HP A component in your patient circuit (tubing, mask, tracheotomy tube, etc.) may be blocked.

Clean, empty or replace the components of your patient circuit. Contact your clinical care provider if the alarm persists.

LP A component in your patient circuit (tubing, mask, tracheotomy tube, etc.) may be wrongly connected.

Check that your patient circuit is set up correctly as demonstrated by your clinical care provider, or replace it. Contact your clinical care provider if the alarm persists.

INT.

EXT.

Page 21: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

17Cleaning and Maintenance

7 Cleaning and MaintenanceYour clinical care provider will maintain your ventilator and its accessories regularly.You should also maintain your ventilator and accessories regularly as described below. Maintenance instructions for the various components are shown in this table.

Table 3: Maintenance frequency.

O2 A component in your oxygen circuit may be wrongly connected.

Check that the oxygen supply equipment is securely connected. Contact your clinical care provider if the alarm persists.

V Your breathing has changedORA component in your patient circuit (tubing, mask, tracheotomy tube, etc.) may be wrongly connected.

Check that your patient circuit is set up correctly as demonstrated by your clinical care provider, or replace it. Contact your clinical care provider if the alarm persists.

F Your breathing has changedORA component in your patient circuit (tubing, mask, tracheotomy tube, etc.) may be wrongly connected.

Check that your patient circuit is correctly set up as demonstrated by your clinical care provider, or replace it. Contact your clinical care provider if the alarm persists.

The power supply is disconnected.

If the disconnection is intentional, or due to a power outage, acknowledge by pressing the Alarm Silence button .Otherwise, reconnect the power supply.Contact your clinical care provider if the alarm LED remains lit.

Alarm icon Cause Solution

Component Frequency

Patient circuit. Weekly.

Exterior of ventilator. Weekly.

Dust filter (clean). Monthly.

Dust filter (replace). Every 6 months, or when worn or soiled.

Page 22: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

18

CAUTION• For maintenance of the following accessories: mask, humidifier,

antibacterial filter and water traps, refer to the manuals provided with

these items.

• Do not use bleach, chlorine, alcohol, aromatic-based solutions

(including all scented oils), or moisturising or antibacterial soaps. These

solutions may cause hardening and reduce the life of the product.

WARNINGBeware of electric shock. Do not immerse the ventilator or power cord in

water. Always unplug the ventilator before cleaning and be sure that it is dry

before plugging it back in.

7.1 Maintenance of the dust filter

Inspect the filter every month to check if it is blocked by dirt or contains holes.With normal use, the dust filter of your Elisée 150 needs to be replaced every six months.To replace or clean the filter, use a pair of tweezers to pull it out of its housing at the rear of the ventilator.

Figure 23: Removing the dust filter.

7.2 Maintenance of the exterior of the ventilator

Clean the exterior of your Elisée 150 with a damp cloth and a mild detergent.

CAUTIONNever place the ventilator in direct contact with liquid.

7.3 Maintenance of your circuit

If your patient circuit is reusable, you can clean and disinfect it. Please follow recomandations provided by your home care provider regarding maintenance of your patient circuit.

Page 23: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

19Appendices

8 Appendices

8.1 Technical specifications

Dimensions (in mm)L x W x H = 290 x 250 x 130.

WeightWeight of the ventilator: 3.7 kg (8.16 lbs).Weight of the ventilator with the mains power pack: 4.4 kg (9.7 lbs).Weight of the ventilator with an external battery pack: 4.5 kg (9.9 lbs).

Power supplyMains power: 110-240 V AC ; 50/60Hz ; 0.67-1.33 A.External power: 12-28 VDC, 15 A max.Batteries: Li-Ion; 14.4 VDC; 6.3 AH.Internal battery life: Minimum of 6 hours.External battery life: Minimum of 6 hours.Battery charging time: 6 hours.

WARNINGThe above battery life applies only for the following precise settings: Type: adult/invasive, PCV mode, P Insp. = 20 cmH2O, PEEP = 0 cmH2O, F = 15 bpm, Ti = 1.2 s, Triggers = No, Slope = 3, Patient circuit CIR009727,

Maquet 190 test lung, VTe = 425 mL, Screen brightness: Automatic, Altitude:

100 metres, Temperature: 20°C.

8.2 Description of buttons and symbols

Table 4: Buttons and symbols on the Elisée 150

Button / symbol Function

Start ventilation Pressing and holding this button for 1 second starts the selected ventilation program.

Ventilation program This button is used to select one of the ventilation programs configured by your physician, as well as your main prescribed treatment settings.

STARTVENTIL �1s

ACV

Page 24: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

20

Manual test This button is used to carry out a manual test, so that the ventilation can record the configuration of your patient circuit.Date displayed as Day / Month / Year.

CAUTIONPress this button only on the recommendation

of your clinical care provider, and only if you

have been trained in how to perform a manual

test.

Type of battery and power supply

These symbols show the type of battery (internal or external) and the type of power supply (mains or external).

Date and time This is where the date and time are displayed.

Alarm Silence button This backlit button lights up when an alarm condition is present, and can be used to silence the audible alarm temporarily (for two minutes).

Setting These are the ventilation settings prescribed by your physician and entered by your clinical care provider.

Measurement These are the measurements taken by your Elisée 150 during ventilation.

Button / symbol Function

Last test the 18/06/08Double circuit : PASS

TESTTEST

Mains

>

Intern.

Wednesd. February2811: 28: 31

VT

390 ml

Ppeak

33.6 H2Ocm

Page 25: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

21Appendices

For more information on the ventilation modes, parameters or measurements, contact your HME provider.

8.3 Travelling with your ventilator

For long journeys, it is advisable to carry your device in its travel bag.Also remember to pack: • the power cord,• the patient circuit and accessories,• the connector for the oxygen tubing (if you are using the ventilator's

oxygen option).If you intend to travel by air with the ventilator, ask your clinical care provider about the required formalities.

8.4 Manual test (while ventilation is stopped)

If your clinical care provider recommends that you do so, you may perform a manual test. This test lets your ventilator record the configuration of your patient circuit after you have made changes to the circuit (replaced the air tubing or added accessories).

WARNINGS• The patient must not be connected to the ventilator during this test.

• If your ventilator is connected to an oxygen supply, disconnect it.

• Make sure that when you perform this test, all the accessories that will

be used during ventilation are connected. Repeat the test every time an

accessory is added or removed.

• If the test fails, turn off the ventilator and contact your clinical care

provider.

Pressure indicator This area shows the pressures being delivered by your Elisée 150 (inspiratory and expiratory pressure).The figure at the top is the maximum pressure.

Back This button takes you back to the previous screen.

Button / symbol Function

< Ai

0

10

20

30

90

< PE

14.6 H2Ocm

Back

Page 26: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

22

1. On the start screen (See Section 5.2, “Starting ventilation” on page 12), press the button on the right-hand side of the device, then press the “Test” button for 3 s. If no other buttons are pressed for 8 s, the ventilator will return to the start screen.

2. The results of the last test are displayed. Press the “Restart” button to start a new manual test. If no other buttons are pressed for 30 s, the ventilator will return to the start screen.

3. Follow the instructions on the screen, then press “Continue”.

4. Block the insufflation outlet on the ventilator, then press “Continue”.

Ctot = 1,0 ml/cmH2ORi = 0,2 cmH2O (30 l/min)Ri = 0,7 cmH2O (60 l/min)Re = 1,5 cmH2O (30 l/min)Re = 2,8 cmH2O (60 l/min)

Back Last test the

6/12/10 at 9 : 42 : 08Circuit test : PASS

Kind of circuit : DOUBLE

RESTARTRESTART

Cancel

Disconnect oxygen!!

Unplug patient circuit!

When done, press :

CONTINUECONTINUE

CONTINUECONTINUE

Cancel

Obstruct the inspiratory circuit support

When done, press :

Page 27: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

23Appendices

5. Connect one end of the patient circuit, leaving the other end free, then press “Continue”.

6. Block the free end of the circuit and press “Continue”.

The test results tell you:• Whether the test failed, passed or was interrupted;

• What type of circuit is connected;• Any additional information.

CONTINUECONTINUE

Cancel

When done, press :

Attach your circuit to the machine.

Leave the other end of the circuitloose.

CONTINUECONTINUE

Cancel

Obstruct the end of the patient circuit.

When done, press :

Ctot = 1,0 ml/cmH2ORi = 0,2 cmH2O (30 l/min)Ri = 0,7 cmH2O (60 l/min)Re = 1,5 cmH2O (30 l/min)Re = 2,8 cmH2O (60 l/min)

Circuit test : INTERRUPTEDBack

Last test the15/11/10 at 10 : 42 : 08

Kind of circuit : DOUBLE

RESTARTRESTART

Ctot = 1,0 ml/cmH2ORi = 0,2 cmH2O (30 l/min)Ri = 0,7 cmH2O (60 l/min)Re = 1,5 cmH2O (30 l/min)Re = 2,8 cmH2O (60 l/min)

Back Last test the

6/12/10 at 9 : 42 : 08Circuit test : PASS

Kind of circuit : DOUBLE

Passed screen test Interrupted screen test

Page 28: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

24

Page 29: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

Index

A

accessoriesmanual test 21

air inlet 6

B

bar graph 21battery life 19battery type 20button

Alarm Silence 20On/Off 6, 12, 13, 14

D

date and time 20dimensions 19dust filter

maintenance 18

E

electrical connectionexternal 8, 10standard 8

exterior of devicemaintenance 18

external batteryicon 16

I

internal batteryicon 16

M

mains power cordexternal 10standard 9

maintenance 17manual test 20measurements 15, 20

O

oxygen 10connection 6

P

patient circuit 3double circuit 5, 7, 8maintenance 18single circuit 5, 7

power cordfastening clip 9standard 3

power supply 19power supply type 20

R

retractable handle 4

S

settings 15, 20start screen 4starting ventilation 19

T

touch screen 4travel bag 3tubes 7

V

ventilation program 19

W

warnings 2weight 19

25Index

Page 30: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

6

2
Page 31: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

Making quality of care easy ResMed Ventilation solutions

Page 32: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

0123

ResMed Paris, 240 rue de la Motte 77 550 Moissy-Cramayel, France.

See www.resmed.com for other ResMed locations worldwide. Protected by the following patents: AU 004224573, CA 2242882, CN 200480007797.2, FR 2766568, FR 2812203, FR 2852853, FR 2889812, EP 0893673, EP 1177810, JP 4060449, JP 4591443, JP 4689893, NZ 543179, NZ 566148, US 6164141, US 6722356, US 7905231. Others patents pending. Elisée is a trademark of ResMed Paris and is registered in the U.S. Patent and Trademark Office. The technical specifications may be changed without notice. © 2012 ResMed Paris.

Global leaders in sleep and respiratory medicine www.resmed.com

NOT014229-7 2012-01Elisée 150

USER

ENG

Page 33: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

DOC 511 1 /2 MP_E150_ENG_NOT014229-7_SPECSHEET Revision 1 Revision 7

(Document de spécification, ne pas imprimer ni publier avec le document principal)

Manuel Patient Elisée 150 (ENG)

1. Designation Référence Oracle (Sydney)/

Référence Produflex (Paris) : NOT014229-7

Nombre de pages total (hors spécification) : 32

2. Détails

COUVERTURE Format : A4 (210 X 297+-2mm) A5 (148 X 210+-2mm) autre : ………..……………. Style : couverture imprimée recto-verso (2 premières pages et 2 dernières pages) Type d’impression : 4 couleurs Papier et grammage: 250gr silk Illustration: Voir document ci-joint Pour une réimpression conserver les mêmes style et même tailles Dans le cas où les couleurs utilisées dans le document diffèrent des couleurs spécifiées, la

spécification doit être prioritaire.

INTERIEUR Format : A4 (210 X 297 +- 2mm) A5 (148 X 210 +- 2mm) autre : ………..……………. Style et mode d’assemblage : Impression recto-verso, piqûre à cheval. Des pages vierges doivent être ajoutées à la fin du

document si le nombre de pages de texte n’est pas un multiple de 4.

Papier et grammage : 80gr offset Type d’impression : 4 couleurs Fabricant : Imprimeur extérieur Conditionnement : Film. Produits emballés dans un carton assurant la protection pendant le transport. Une étiquette doit

être collée sur le carton mentionnant le nom du produit, le code du produit, le numéro de lot, le numéro de version de la spécification et la quantité contenue dans le carton. Toutes ces informations doivent apparaître sous forme de code barre 39 Full Ascii.*

3. Contrôle Qualité Paris : Doc 515 (PV de contrôle documents imprimés) – valable pour validation BAT et contrôle qualité série Sydney : AWI203-002 – Printed Materials/Labels/CDs/Videos Section 19.1 & 19.2

Version autorisée consultable sur OBJECTIVE. Copie non gérée en l’absence d’un tampon « Document contrôlé ». Ce document confidentiel est la propriété de ResMed. Ne pas reproduire ni permettre son utilisation par un tiers sans autorisation préalable.

Ne pas utiliser sans autorisation. © ResMed Ltd 2009

SPECIFICATION

DOCUMENTS IMPRIMES

Rev Date Author Change note

4 07/10/2009 Mylène Goude M01330

5 31/03/2011 S. Pruchon M02014

6 17/08/2011 A. Piera M02150

7 06/06/2012 A. Piera M02485

Page 34: Elisée™ 150 - ResMed · 2020-01-07 · Introduction 1. 1 Introduction. This manual is provided with your Elisée™ 150 ventilator (with software version 2.54 NIV+or higher). It

DOC 511 2 /2 MP_E150_ENG_NOT014229-7_SPECSHEET Revision 1 Revision 7

(Reference only, not for printing or web publishing)

Patient Manual Elisée 150 (ENG)

1. Description Oracle reference (Sydney) :

Produflex reference (Paris) : NOT014229-7

Total number of pages (spec sheet excluded): 32 2. Details

COVER PAGE Format : A4 (297H X210W +- 2mm finished) A5 (210H X 148W +- 2mm finished other : ………..…………. Style: covers printed double sided (two first pages and the two last ones) Colour of printing: 4 colour Process Stock: 250 gsm Silk Art Work: As shown on following pages If re-typeset, the same styles and sizes must be maintained. Where the colours indicated differ between the printing specification and contents, the printing

specification should be followed

TEXT Dimensions: A4 (297H X210W +- 2mm finished) A5 (210H X 148W +- 2mm finished other : ………..…………. Style and binding Printed doubled-sided, saddle stitch. Additional blank page may be added to the back of the

document if the current text page count is not a multiple of 4 Stock: 80 gsm offset Colour of printing: 4 colour process Manufacturer: Outside printer Packaging: Film. Products must be packed in a cardboard box for transportation. A label must be sticked on the

cardboard box, and must mention quantity, product name and code, batch number and spec sheet version number. This information must appear on a 39 Full Ascii bar code format.

3. ResMed QC Inspection Paris: Doc 515 (PV de contrôle documents imprimés) – valid for passed for press and quality control Sydney : AWI203-002 – Printed Materials/Labels/CDs/Videos Section 19.1 & 19.2

Authorized version available on OBJECTIVE. Non controlled copy if no “Document contrôlé” stamp. ResMed Confidential Proprietary Information, not to be reproduced or made available to third parties without prior consent from ResMed

and not to be used in any unauthorised way. © ResMed Ltd 2009

PRINTED MATERIALS SPECIFICATION