UW- Madison Geology 777. Electron Probe Microanalysis EPMA. Quantitative Analysis and Matrix Corrections. Revised 10/29/2009. UW- Madison Geology 777. K ratios. Recall Castaing’s approach to quantitative analysis, where specimen intensities are ratioed to standard intensities: - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Electron Probe Microanalysis EPMA

Electron Probe Microanalysis

EPMA

UW- Madison Geology 777

Quantitative Analysis

andMatrix

CorrectionsRevised 10/29/2009

Page 2: Electron Probe Microanalysis EPMA

K ratiosRecall Castaing’s approach to quantitative analysis, where specimen intensities are ratioed to standard intensities:

where K is the “K ratio” for element i, I is the X-ray intensity of the phase and subscript i is the element.

With the counts acquired on BOTH unknowns and standards on the same instrument, under the same operating conditions, we assume that many physical parameters of the machine that would be needed in a rigorous physical model cancel each other out (same in numerator and denominator).

Ki =Iiunk

Iistd

UW- Madison Geology 777

Page 3: Electron Probe Microanalysis EPMA

Castaing’s First Approximation

Castaing’s “first approximation” follows this approach. The composition C of element i of the unknown is the K ratio times the composition of the standard. In the simple case where the standard is the pure element, then, the fraction K is roughly equal to the fraction of the element in the unknown.

Ciunk≈

Iiunk

IistdCi

std =KiCistd

UW- Madison Geology 777

Page 4: Electron Probe Microanalysis EPMA

...close but not exact

However, it was immediately obvious to Castaing that the raw data had to be corrected in order to achieve the full potential of this new approach to quantitative microanalysis.

The next two slides give a graphic demonstration of the need for development of a correction procedure.

Ciunk≈

Iiunk

IistdCi

std =KiCistd

UW- Madison Geology 777

Page 5: Electron Probe Microanalysis EPMA

Raw data needs correction

This plot of Fe Ka X-ray intensity data demonstrates why we must correct for matrix effects. Here 3 Fe alloys show distinct variations. Consider the 3 alloys at 40% Fe. X-ray intensity of the Fe-Ni alloy is ~5% higher than for the Fe-Mn, and the Fe-Cr is ~5% lower than the Fe-Mn. Thus, we cannot use the raw X-ray intensity to determine the compositions of the Fe-Ni and Fe-Cr alloys.

(Note the hyperbolic functionality of the upper and lower curves)

UW- Madison Geology 777

Page 6: Electron Probe Microanalysis EPMA

Absorption and Fluorescence

• Note that the Fe-Mn alloys plot along a 1:1 line, and so is a good reference.

• The Fe-Ni alloys plot above the 1:1 line (have apparently higher Fe than they really do), because the Ni atoms present produce X-rays of 7.278 keV, which is greater than the Fe K edge of 7.111 keV.Thus, additional Fe Ka are produced by this secondary fluorescence.

Z el Ka keV K edge keV24 Cr 5.415 5.98925 Mn 5.899 6.53826 Fe 6.404 7.11127 Ni 7.478 8.332

• The Fe-Cr alloys plot below the 1:1 line (have apparently lower Fe than they really do), because the Fe atoms present produce X-rays of 6.404 keV, which is greater than the Cr K edge of 5.989 keV. Thus, Cr Ka is increased, with Fe Ka are “used up” in this secondary fluorescence process.

UW- Madison Geology 777

Page 7: Electron Probe Microanalysis EPMA

In his 1951 Ph.D. thesis, Castaing laid out the two approaches that could be used to apply matrix corrections to the data:

• an empirical ‘alpha factor’ correction for binary compounds, where each pair of elements has a pair of constant -factors representing the effect that each element has upon the other for measured X-ray intensity, and

• a more rigorous physical model taking into account absorption and fluorescence in the specimen. This later approach also includes atomic number effects and became known as ZAF correction.

Two approaches to corrections

UW- Madison Geology 777

Page 8: Electron Probe Microanalysis EPMA

Z A F

In addition to absorption (A) and fluorescence (F), there are two other matrix corrections based upon the atomic number (Z) of the material: one dealing with electron backscattering, the other with electron penetration (or stopping). These deal with corrections to the generation of X-rays.C is composition as wt% element (or elemental fraction).

We will now go through all these corrections in some detail, starting with the Z correction, which has two parts, the stopping power correction, and the backscatter correction.

Note that all these corrections require close attention to exactly what feature’s value is being input: the target (matrix), or the X-ray in question.

Ciunk=

Iiunk

Iistd

ZAFiunk

ZAFistdCi

std

UW- Madison Geology 777

Page 9: Electron Probe Microanalysis EPMA

Stopping Power CorrectionIncident electrons lose

energy in inelastic interactions with the inner shell electrons of the target. The “stopping power” (energy lost by HV electrons per unit mass penetrated) is not constant but drops with increasing Z. A higher number of X-rays will be produced in higher Z targets. Thus, if the mean Z of the unknown is higher than that of the standard, a downward correction in the composition must be applied. The stopping power correction factor is “S”, and can be approximated by:

Stopping power of pure elements for 20 keV electrons

where J=11.5+ Z and Emean=

(E0+Ec)/2(J is the mean ionization energy; J, Z and A are of the target, Emean is of the X-ray)

S=ZA

ln1.166×Emean

J

⎛ ⎝ ⎜

⎞ ⎠ ⎟

Reed, 1996, Fig. 8.6, p. 135

UW- Madison Geology 777

Page 10: Electron Probe Microanalysis EPMA

Backscatter Correction

As we discussed earlier, the fraction of high energy incident elections that are backscattered () increases with atomic number. There then will be relatively less incident electrons penetrating into higher Z specimens, resulting in a smaller number of X-rays. Thus, if the mean Z of the unknown is higher than that of the standard, a upward correction in the composition must be applied. The backscatter correction factor is “R”.

R can be approximated by

where W = Ec/E0 (the inverse of

overvoltage), and Z is of the target, and W is of the X-ray

R=1

1+[0.008×(1−W)×Z]

Reed, 1996, Fig. 2.11, p. 17

UW- Madison Geology 777

Page 11: Electron Probe Microanalysis EPMA

Z correctionThe total atomic number correction is formed by multiplication of the R and S of the unknown and standard thusly:Z = Rstd/Runk * Sunk/Sstd

Overall the backscatter and the stopping power corrections tend to cancel each other out.But if there is a (small) correction, it is usually in the direction of the backscatter correction.

UW- Madison Geology 777

Page 12: Electron Probe Microanalysis EPMA

Beers LawThe intensity I of X-rays that pass through a substance are subject to attenuation of their initial intensity I0 by the material over the distance they travel within the material. The attenuation follows an exponential decay with a characteristic linear attenuation length 1/, where is the (linear) absorption coefficient. Beers Law can also be expressed in terms of mass, using density terms:

I = I0 exp -

(/)( Z)

where () is the mass absorption coefficient (cm2/g), is the material density (g/cm3), and Z is the distance (cm) Als-Nielsen and McMorrow, 2001, Fig

1.10, p. 19

UW- Madison Geology 777

Page 13: Electron Probe Microanalysis EPMA

Mass Absorption Coefficients

Mass absorption coefficients (MACs) have been tabulated* for many X-rays through many substances (though some are extrapolations). They exist as a matrix of numbers: absorption of a particular X-ray line (emitter, e.g. Ga k) by a absorber or target (e.g. As) will have one value (51.5). Note that the absorption of As K by Ga is a totally different phenomenon with a distinct MAC (221.4) .

Emitter = X-ray (here,

K)

Absorber = matrix material

Goldstein et al, 1992, p. 750.

UW- Madison Geology 777

*See following discussion

Page 14: Electron Probe Microanalysis EPMA

Mass Absorption Coefficients

Emitter = X-ray (here,

K)

Absorber = matrix material

Goldstein et al, 1992, p. 750.

UW- Madison Geology 777

Terminology:

“the mass absorption of Ga Ka by As”

Question for you: Is the mass absorption of As Ka by Ga the same as the mass absorption of Ga Ka by As?

Why or why not?

Page 15: Electron Probe Microanalysis EPMA

AbsorptionX-rays produced within the

material will be propagated in all directions, and will suffer attenuation in the process. Note that the path length of travel of the X-ray to the spectrometer is z cosec, where psiis the takeoff angle (cosec = 1/sin).

Castaing’s approach was to integrate the Beer’s Law equation over the depth at the given , producing the absorption correction factor f() where is defined as cosec where is the MAC.

The absorption (A) correction is then defined as

A= f()std / f()sample

Reed, 1993,, p. 219

f (χ) =emergent intensitygenerated intensity

UW- Madison Geology 777

Page 16: Electron Probe Microanalysis EPMA

Absorption

To be able to correct for this absorption of the measured X-rays, we need to know how the production of X-rays varies with depth (Z) in the material.

The distribution of X-rays as a function of depth is known as the (z) [phi-rho-z] function, where a “mass depth” parameter is used instead of simple z (bottom right).

The (z) function is defined as the intensity generated in a thin layer at some depth z, relative to that generated in an isolated layer of the same thickness.

Reed, 1993, p. 219

UW- Madison Geology 777

Page 17: Electron Probe Microanalysis EPMA

Absorption

One commonly used simplified form (Philibert 1963) is

where = cosec , is a measure of electron absorption and depends on effective electron energy, where

The Philibert approximation breaks down, however, at the near surface, creating errors when dealing with low energy light elements, and we need to go to more complicated and accurate forms of the (z) function.

Reed, 1993, p. 219

f (χ) =(1+h)

1+ χσ( )[ ]× 1+h× 1+ χ

σ( )[ ]{ }

σ =4.5×105

E01.65 −Ec

1.65h=1.2×

AZ2

UW- Madison Geology 777

Page 18: Electron Probe Microanalysis EPMA

(z) [phi-rho-z] Curves

To be able to correct properly for absorption -- particularly for light elements, the exact shape of the (z) [phi-rho-z] curve must be known. Each X-ray has its own curve. There are 3 main parameters that affect the shape of the curve:

Reed, 1993, p. 220

•E0 (accelerating voltage)

• Ec (critical excitation energy of a particular element line

• mean Z of the material

UW- Madison Geology 777

Page 19: Electron Probe Microanalysis EPMA

Tracer Method

UW- Madison Geology 777

The (z) [phi-rho-z] curves are usually determined by the “tracer method”, where successive layers are deposited by vacuum evaporation. The tracer layer B is deposited atop substrate A, with successive layers of A deposited on top.

Characteristic X-rays from the tracer element are measured (“emitted”) and then a generation curve is calculated by correcting each step for absorption and fluorescence effects

Page 20: Electron Probe Microanalysis EPMA

Fluorescence Correction

The X-rays produced within a specimen have the potential for producing a second generation of X-rays: this is secondary fluorescence, generally shortened to fluorescence. This occurs when the characteristic X-ray has an energy greater than the absorption edge energy of another element present in the specimen.

As we saw earlier, Ni Ka (7.48 keV) is able to fluoresce Fe Ka (Ec 7.11 keV). This effect is maximized when there is a small amount of the fluoresced element present, e.g. Fe in a Ni-Fe alloy.

Reed, 1996, Fig. 8.10, p. 139

Reed gives an example where the Fe intensity is 142% of what it “should be”. Also, the continuum above an absorption edge also causes fluorescence, although this is generally weak.

UW- Madison Geology 777

Page 21: Electron Probe Microanalysis EPMA

Fluorescence Correction

The form of the correction F is

where If/Ip is the ratio of emitted X-rays from fluorescence, compared to the X-ray intensity from inner shell ionization. In a compound, this term is summed overall all the elements that could fluorescence the element of interest.

F =1

1+I fIp

UW- Madison Geology 777

Page 22: Electron Probe Microanalysis EPMA

Fluorescence ProblemsSecondary fluorescence

is an important issue that must be appreciated. Generated X-rays are not scattered nearly as much as incident electrons, and thus the generated X-rays can travel relatively long distances (50 um in Fig 3.49) within the specimen and produce a second generation of X-rays. If the specimen (and standards) are relatively large (=homogeneous), this is not a problem. However, if minor or trace elements are being analyzed in small grains (Phase 1 in Fig 16.10) and the host phase (2) has high abundance, an error may be made in the EPMA analysis.

Goldstein et al. p. 142; Reed 1993, p. 258

UW- Madison Geology 777

Page 23: Electron Probe Microanalysis EPMA

Fluorescence across boundariesSecondary fluorescence is a

potential source of analytical error across linear boundaries, either horizontal (e.g. thin films) or vertical (e.g., diffusion couples).

In the example here of a vertical interface between untreated Cu and Co, there is NO diffusion. However, the resulting EPMA profiles clearly imply there is diffusion. There is NO diffusion – there is only secondary fluorescence across the boundary. Cu Ka X-rays can excite Co, to the extent that there is apparently 1 wt% Co about 15 um away from the boundary within the Cu. But Co Ka cannot excite Cu, so only the continuum X-rays can create secondary fluorescence, which is less – but certainly distinguishable, an apparent 0.5 wt% Cu at 10 um from the boundary in the Co.

Reed 1993, p. 259-260

UW- Madison Geology 777

Z 27 29Co Cu

Ka 6.93 keV 8.05 keVK edge 7.71 keV 8.98 keV

“false Co” “false Cu”

Cu Co

Page 24: Electron Probe Microanalysis EPMA

An SF real storyAnother lab reported 10 wt%

Nb (below) in what should have been Nb-free phase (by EDS at 30 kV). The issue was small grain size and nearby Nb (right image)

UW- Madison Geology 777

When I ran WDS (18 kV) I found essentially zero Nb -- what is the problem?

The original researchers used Nb Ka because, with EDS, it is impossible to resolve Nb La (it sits between Al Ka, Hf Ma and Pd La).

Fournelle, Kim and Perepezko (2005)

Above: EDS spectrum on Pd2HfAl 5 um away from Nb. Pd Ka is very efficient at traveling across the border and exciting the Nb

Page 25: Electron Probe Microanalysis EPMA

Secondary Fluorescence Correction

A recent article (below left) reports an innovative approach to correcting the secondary fluorescence (SF) in diffusion couples and from adjacent phases. This utilizes a complex Monte Carlo program called PENELOPE (Penetration and Energy Loss of Positrons and Electrons) that permits complicated geometric models of electron and X-ray behavior in materials. SF can be simulated in a model that represents the actual specimen (e.g. Fig 1 below), and then subtracted from the observed data (right figure).

UW- Madison Geology 777

Page 26: Electron Probe Microanalysis EPMA

Matrix Correction ProgramsThe raw X-ray intensities are first corrected for:

• background contribution

• beam drift (i.e. counts are normalized)

• deadtime

• interferences* (if appropriate)

and then the K-ratios are input into an automated matrix correction program.

To run, the correction calculations must assume an initial composition for the unknown -- because the magnitude of each factor is proportional to the abundance of the element times its correction in a pure end member. The assumed composition is a normalized (to 100%) value of the K-ratio. Based upon the first iteration with this assumed composition, the result gives a more truer composition, which then is the input for the second iteration. The process is iterated until convergence, usually 3-5 times.

* Probe for Windows does the interference correction within the matrix correction, a far better approach compared to the normal (antiquainted) procedure of correcting the data after the matrix correction is completed.

UW- Madison Geology 777

Page 27: Electron Probe Microanalysis EPMA

ZAF optionsOne currently widely used matrix correction program is CITZAF, developed by John Armstrong (then CIT, now at American Univ.) and implemented in our Probe for Windows software. There are several options, which we elucidate here, but that generally we do not modify them from the default values. Probably the only parameter you would ever modify would be mass absorption coefficients (there are different ones for the light elements).

UW- Madison Geology 777

Page 28: Electron Probe Microanalysis EPMA

Alpha correction In the early decades of probing when computer power was negligible, the alpha correction technique was widely used, as it required less number crunching and relied mainly on empirical data and less on complex physical models and physics. Today, however, there may be a rekindled interest in this approach, as it may “work better” in many cases.

UW- Madison Geology 777

Page 29: Electron Probe Microanalysis EPMA

Ziebold and Ogilvie -

binary -factors

Geology 777

In 1963-4, Ziebold and Ogilvie* developed corrections for some binary metal alloys, with an equation in the form

where 12 is the-factor for element 1 in the binary with element 2, K is the K-ratio, and composition (fractional) is C. This equation can be rearranged in the form

If experimental data exist for binary alloys, then a plot of C1/K1 versus C1 is a straight line with a slope of (1- 12), leading to determination of 12. Such a hyperbolic relationship between C1 and K1 was shown to be correct for several alloy and oxide systems, but it was difficult to find appropriate intermediate compositions for many binary systems.

(1−K1)K1

=α12

(1−C1)C1

*Quantitative Analysis with the Electron Microanalyzer, Analytical Chemistry, Vol 35, May 1963, p. 621-627; An Empirical Method for Electron Microanalysis, Analytical Chemistry, Vol 36, Feb. 1964, p. 322-327.

C1

K1

=α12 +(1−α12)C1

= C2= K2

Page 30: Electron Probe Microanalysis EPMA

Ziebold and Ogilvie -

ternary -factorsZiebold and Ogilvie showed that corrections could be developed for some ternary metal alloys, with an equation in the form

where 123 is the-factor for element 1 in the ternary with elements 2 and 3, and is defined as

This equation can be rearranged

Similar relationships can be written for elements 2 and 3, and used to calculate a-factors for the 3 binary systems of the ternary.These a-factors were limited to a particular E0 and takeoff angle.

(1−K1)K1

=α123

(1−C1)C1

C1

K1

=α123+(1−α123)C1

α123 =α12C2 +α13C3

C2 +C3

UW- Madison Geology 777

Page 31: Electron Probe Microanalysis EPMA

Bence-Albee -multicomponent

systemsBence and Albee* in 1968 showed that this approach could be extended to silicates and other minerals, i.e. a system of n components, where for the nth component a -factor could be found

where

where n1 is the-factor for the n1 binary.

These factors were determined for a limited set of conditions, i.e. 15 and 20 keV, and take off angles of 52.5° and 38.5°.

The 1968 Bence and Albee paper is one of the most highly cited papers in the geological literature (over ~20,000 citations).

* Empirical correction factors for the electron microanalysis of silicates and oxides, J. Geology, Vol. 76, p. 382-403; also see Albee and Ray, Correction Factors for Electron Probe Microanalysis of Silicates, Oxides, Carbonates, Phosphates, and Sulfates, Analytical Chemistry, Vol 42, Oct 1970, p. 1408-1414.

Cn

kn=βn

βn =k1αn1 +k2αn2 +k3αn3 +⋅⋅⋅+knαnn

k1 +k2 +k3 +⋅⋅⋅+kn

UW- Madison Geology 777

Page 32: Electron Probe Microanalysis EPMA

Evaluating matrix corrections

Armstrong also reviewed† ZAF and (z) corrections and suggested that some of these correction algorithms “produce poorer results in the analysis of silicate and oxide minerals than some of the earlier corrections”. He specifically was referring to various corrections that were optimized for metal alloys† Quantitative analysis of silicate and oxide materials: comparison of Monte Carlo, ZAF and (z) procedures, in Microbeam Analysis-1988, p. 239+

*Bence-Albee after 20 years: review of the accuracy of -factor correction procedures for oxide and silicate minerals, in Microbeam Analysis-1988, p. 469-76.

In 1988, John Armstrong* reviewed the Bence-Albee (-factor) correction scheme for EPMA of oxide and silicate minerals. He evaluated the old factors, and revised some, using -factors calculated from newer ZAF and (z) algorithms, and showed “that with some modifications the -factor corrections can be as accurate as any other correction procedure currently available and much easier and quicker to process.”

UW- Madison Geology 777

Page 33: Electron Probe Microanalysis EPMA

Before we forget....

The matrix corrections assume that all elements present (and interacting with the X-rays) will be included. There are situations, however, where either an element cannot be measured, or not easily, and thus the analyst must make explicit in the quantitative setup the presence of unanalyzed element/s -- and how they are to be input into the correction.

Unanalyzed elements

Typically oxygen (in silicates) is calculated “by stoichometry”. Elements can also be defined in set amounts, or relative proportions, or “by difference” – although this later method is somewhat dangerous as it assumes that there are no other elements present.

UW- Madison Geology 777

Page 34: Electron Probe Microanalysis EPMA

Oxygen is a major part of many materials (e.g. silicate minerals and glasses). Carbon is a major part of carbonates. Oxygen and carbon are typically not “acquired” (measured directly) in many EPMA procedures, BUT THEY MUST BE SOMEHOW INCLUDED IN THE MATRIX CORRECTION.

If oxygen (and say C in carbonates) is not included, there will be errors in the matrix corrections of some elements, as the presence these elements affects the electron (and x-ray behavior), e.g., the backscattered % of incident HV electrons will be different, and there may be absorption of those x-rays by the oxygen (and C) present.

In the next slides, we see the effect on a measurement of a carbonate sample: first, with just the cations; then with just Oxygen added; finally with both Oxygen and Carbon.

Unanalyzed oxygen, carbon, etc

UW- Madison Geology 777

Page 35: Electron Probe Microanalysis EPMA

First: how you do it

UW- Madison Geology 777

For a carbonate

1. Declare C and O as elements in the analysis (but not measured)

2. Tell it to calculate with stoichiometric oxygen

3. Create a rule so that there is 1 atom of C to every 3 atoms of O

4. Check: display results on basis of 3 atoms of oxygen (perfect result would be 5.000 total atoms, and grand total of 100 wt%

Page 36: Electron Probe Microanalysis EPMA

UW- Madison Geology 777

Good, Bad, Ugly …. Carbonate by EPMAFirst, if only cations

used for ZAF …

Next, oxygen only included …

If both O and C included in the ZAF

Notice the differences in the final values of the elements (wt%)

Page 37: Electron Probe Microanalysis EPMA

One complication for oxygen is variable valence states of elements such as Fe. Robust software will allow you to enter case by case different valence states.

Unanalyzed oxygen

UW- Madison Geology 777

* Tingle, T.N., Neuhoff, P., Ostgergren, P., Jones, R.E. and Donovan, J.J. (1996) The effect of “missing” unanalyzed oxygen on quantitative electron probe microanalysis of hydrous silicate and oxide minerals. GSA Abstracts, 28, 212.

In some cases, if oxygen is not included, there can be errors in the matrix corrections of some elements, as the presence of O, OH, and H2O can affect the actually measured elements, as there may be significant absorption of those x-rays by the oxygen present*.

Page 38: Electron Probe Microanalysis EPMA

Impact of unaccounted for oxygen

Consider: Apophyllite -- KCa4Si8O20(F,OH)•8 H2O

Which has LOTS of oxygen which typically is “unanalyzed” and therefore not involved in the

matrix correction

UW- Madison Geology 777

Solution: Iterate a fixed amount of H2O (16 atoms of H = 1.76 wt% H plus stoichometric O) per formula to achieve good results.

As shown in the bottom analysis where the H2O is missing, there is up to 3% relative error for cations.

Page 39: Electron Probe Microanalysis EPMA

Physical Parameters Needed

• Electron stopping power

• Mean ionization potentials

• Backscatter coefficients

• X-ray Ionization cross sections

• Mass absorption coefficients

• Surface ionization potentials

• Fluorescent yields

The ZAF corrections require accurate and precise knowledge about many physical parameters, such as

UW- Madison Geology 777

Page 40: Electron Probe Microanalysis EPMA

“State of EPMA parameters”

As David Joy points out in his 2001 article “Constants for Microanalysis”, there are problems in our knowledge of many parameters:

• there are experimental stopping power profiles for 12 elements and 12 compounds, which raise questions about the traditional Bethe equation

• only half of the elements whose K lines are used for EPMA have measured K shell ionization cross-sections ; only 6 elements have measured L shell cross-sections; there are zero M shell cross-sections

• K shell fluorescent yields are the best documented parameters; there are gaps in the data for L shell yields; there are only 5 measured M shell yields

• despite the fact that backscatter coefficients have been measured for 100 years, the data has many gaps and is of poor precision (i.e. 30%)

UW- Madison Geology 777

Page 41: Electron Probe Microanalysis EPMA

At the Eugene EPMA workshop in September 2003, John Armstrong reviewed the state of EPMA matrix corrections

• Big problem with software/manufacturers, not documenting which corrections used. Some have picked "improved" parameters which do not fit with the other parameters, e.g. in some, where no formal fluorescence correction, the absorption correction was tweaked to take fluor into account, and then when later fluorescence corrections developed, to use this in addition to absorption correction, has an overcorrection for fluorescence.• Problem with researchers not stating in their publications which correction they used; NIST is trying to develop some protocols which people can reference (brief notation with pointer to NIST for full description).• There are a few errors/typos in the long accepted X-ray tables (i.e., Bearden) – 3 are major errors.• Actually measured mass absorption factors are rare! Measurements exist for Na Ka by Al; Si Ka by Al and Ni; Mg Ka by O, Al, Ti and Ni; and Al Ka by O, Na .....• There is over 30% variation in published values of some macs for geologically relevant elements; they can’t all be correct!

UW- Madison Geology 777

Page 42: Electron Probe Microanalysis EPMA

“So what do we do?”We have discussed various ways to correct the raw data, the goal being to come up with the most accurate and precise analytical procedures to give us the most trustworthy data.

We have just mentioned that everything is not as rosy as one would hope.

So, can we trust the numbers we get out of the probe? In many/most cases, given care, yes. But we cannot blindly look at the electron probe and computer as a black box!

Stay tuned for an upcoming installment, where we discuss standards, accuracy and precision in EPMA.

UW- Madison Geology 777