electrochemistry instruments - lambda system · 2016-05-13 · electrochemistry corrosion battery...
TRANSCRIPT
EC-Lab, BT-Lab, MT-Lab, SCAN-Lab products
Generalelectrochemistry
Corrosion Batterytesting Supercapacitors Sensors Coatings Fuel/solarcells Materials Localimaging
Electrochemistryinstruments
Bio-LogicScienceInstruments
ELECTROCHEMISTRY INSTRUMENTS2/3
SAS (founded in 1983) is located near Grenoble in the French Alps. We design and manufacture high performance, research laboratory instruments and software for electrochemistry, battery testing, fuel cell and material testing.
Bio-Logic’s product range is based on a modular and flexible design. We offer electrochemical workstations (EC-Lab division) and battery test equipments (BT-Lab division) from single channel up to 128-channel systems, integrating EIS capability and high current boosters.
MT-Lab division offers a complete solution for material characterization. It provides a complete system including impedance analyzer, furnace, sample holders and software for the control of the instruments and for data acquisition.
In 2012, Bio-Logic SAS extended its choice of instruments to the scanning electrochemical workstations. The Scan-Lab division is dedicated to high resolution electrochemical mapping tools. These products are developed, designed and manufactured by Uniscan Instruments Ltd., a company based in Macclesfield, UK. For more than 20 years, Uniscan Instruments Ltd. have been at the cutting edge of scanning probe electrochemistry technology.
Auniqueandcompleteproductrange-Morethan20differentpotentiostats/galvanostats-Singletomultichannelsystems-StandaloneFRAanalyzer&Built-inEISanalyzer(ElectrochemicalImpedanceSpectroscopy)-Lowcurrentoptions-Highcurrentpotentiostats-Built-inorexternalhighcurrentboosters-Electronicloadbox-Fuelcellteststations-Configurationcanbeadaptedtotheapplication
Modularsystems-Modularchassishavingfrom1to16boardslots-Upto16independentpotentiostats-Highcurrentboosters:2,4,5,8,10,20,80and100A-Extendedvoltagerangeto±15V,±30V,±48V,-Loadbox(50V,150A)-Lowcurrentoptions(downtoaA)-Multipleusersystems(upto16)-LSG(LinearScanGenerator)toperformanalogvoltageramp-Boostersinparallelupto150A(VMP-300technology)-Modularscanningworkstationswithupto9differentlocaltechniques(dc-SECM,ac-SECM,LEIS,SKP,SVET…)-Batterycycler-Fullyfeaturedimpedanceanalyzer
Controlledbyapowerfulsuite of software-Comprehensivesoftwarepackage-Dedicatedtechniquesformoststandardelectrochemicalapplications-Multi-techniqueexperimentbuilder-Externaldevicescontrol-Multiplegraphdisplay-Dataanalysis&fit-Safety&experimental
limits-Softwarecalibration
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Hig
h c
urr
ent
inst
rum
ents
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
atte
ry
test
sta
tions
Boo
ster
s S
oftw
are
pa
ckag
eFu
el c
ell
test
sta
tions
Mat
eria
ls &
liqu
ids
test
sys
tem
sS
can
nin
g
syst
ems
Corrosion
Sensors
Education/training
General electrochemistry
Nanotechnology
Supercapacitors
Modular electrochemical workstations designed to address applications in...
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Hig
h c
urr
ent
inst
rum
ents
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
atte
ry
test
sta
tions
Boo
ster
s S
oftw
are
pa
ckag
eFu
el c
ell
test
sta
tions
Mat
eria
ls &
liqu
ids
test
sys
tem
s
ELECTROCHEMISTRY INSTRUMENTS4/5
Singlepotentiostats/galvanostats
PG-581 6SP-50 6SP-150 7SP-200 7SP-240 8SP-300 8
Multipotentiostats/galvanostats
VSP 10VMP310VSP-300 11VMP-300 11BiStat3200 12SensorStat 12
Boosters VMP-300technology 14VMP3technology 14
Highcurrentinstruments
HCP-803 16HCP-1005 16CLB-2000 16
Batterytest stations
MPG-2xxseries 18BCS-8xxseries 20
Softwarepackage
EC-Lab®/BT-Lab® 22EC-Lab®graphics 23ElectrochemicalImpedanceSpectroscopy 24UiEChem™/UiECorr™ 25
Fuelcelltest stations
FCTseries26FC-Lab 27
Impedanceanalyzer
Conductivitymeter
MTZ-3528
MCS1029
Scanningsystems
Techniquesandfocus 30M370/M470 31
Batteries
Solar cells, photovoltaic systems
Power
Fuel cells
PL
UG
- IN
HY
BR I D E L E C T R I C V
EH
I CL
E
mpg
Coatings
Sca
nn
ing
sy
stem
s
Singlepotentiostats/galvanostats
SP-50 Affordableandeasy-to-useTheSP-50isaneconomical,valueorientedpotentiostat/galvanostatdesignedtoaddressapplicationsingeneralelectrochemistryandcorrosion.Containedinasimple,compactchassis,theSP-50isafixedconfigurationunitwithnooptions.Itistargetedfornewresearchersinthefieldofelectrochemistryandanidealinstrumentforteachingandeducation.Itoffersgoodperformancewithasmallpricetag.
PG-581 On-sitemeasurementWhenyouneedtomakethoseon-sitetrialsorroutinefieldmeasurements,thePG-581goeswithyou.Thecompactdesignprovidesauserconfigurableinstrumentfordemandingresearchapplicationsandin-fielduseviaitsinternalrechargeablebattery.
dI/d
t/m
A/s
0.04
0.02
0
-0.02
-0.04
Ewe/V vs. SCE0.60.40.20-0.2
0.2
0.1
0
-0.1
-0.2
<I>/
mA
Education/training Generalelectrochemistry Corrosion Sensors Electroanalytical
Applications
Options-Floating-±8Vscanvoltage-5-workingelectrodemultiplexer-Computer-programmableexperiment
Education/training Generalelectrochemistry
Applications
ELECTROCHEMISTRY INSTRUMENTS6/7
SP-200 High-end potentiostat/galvanostat SP-200isthenewest,researchgrade,value-orientedpotentiostat/galvanostatintheEC-Lab®productrange.Withexcellentspecificationsinaportablechassis,theSP-200istheperfectinstrumentforanyapplicationinelectrochemistryespeciallycorrosion.TheSP-200offersoutstandingfeaturessuchasafloatingmode,analogfiltering,abuilt-incalibrationboard,and9stabilitybandwidthsforabettercontrolofthecell.Additionally,theSP-200canbepurchasedwithastandardDCpotentiostatoranEIScapableone.ThereisalsoanUltraLowCurrent(ULC)option.On-siteexperimentscanbeperformedthankstoitsportabledesign.
SP-150 Research gradeworkstationTheSP-150isafullfeaturedeconomicalresearchgradepotentiostat/galvanostat.Withitsmodularchassis,thisinstrumentcanbecustomizedtoaddressallapplicationsintheareaofclassicalelectrochemistry.TheSP-150potentiostatconfigurationcanbeselectedatpurchasetoincludeanEIS-capablepotentiostat,andalowcurrentoption(1nAfullscalerange).Itcanalsobeconnectedtoexternalhighcurrentboosters(2,5,10,20,80and100A)or2kWloadbox.
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Generalelectrochemistry Sensors Corrosion Energysources Coatings
Applications
Options-Lowcurrent:fullscalerangedownto1nA-Built-inEISanalyzer:
upto1MHz-Externalcurrentboosters:
2,5,10,20,80and100A-Loadbox:2kW-RDEcontrolkit
Generalelectrochemistry Sensors Corrosion Coatings Materials
Applications
Options-UltraLowCurrent:
fullscalerangedownto1pA-Built-inEISanalyzer:
upto7MHz-LinearScanGenerator:
1MV/s,acquisition1µs-SP-200suitcase
Hig
h c
urr
ent
inst
rum
ents
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
atte
ry
test
sta
tions
Boo
ster
s S
oftw
are
pa
ckag
eFu
el c
ell
test
sta
tions
Mat
eria
ls &
liqu
ids
test
sys
tem
sS
can
nin
g
syst
ems
SP-300 Fast,sensitive,stableandmodularSP-300isastate-of-the-artresearchgradepotentiostat/galvanostat/EISwithremarkablespecifications.ThebasespecificationsarethesameforboththeSP-300andtheSP-200includingfloatingmode,analogfiltering,built-incalibrationboard,andstabilitybandwidths.TheSP-300’smodularchassisacceptanoptionalhighcurrent/highvoltageoptionboard.Alternatively,theSP-300canacceptasecondpotentiostatboard(eitherstandardorEIS)andperformasaBipotentiostat.Itisalsoamultipleusersystemaseachchannelboardcanbeusedindependentlybytwodifferentresearchers.
SP-240 PowerfulinacompactchassisSP-240isa4Apotentiostat/galvanostat/EISwithexcellentspecificationsinaportablechassis,theSP-240istheperfectinstrumentforanyapplicationinelectrochemistryespeciallyenergydevicestesting.AsapartoftheSP-200/300productrange,theSP-240benefitsinthesamespecificationsandoptions.
0.05 0.1 0.15 0.20
0.02
0.04
0.06
0.08
0.1
Re (Z)/Ohm
-Im(Z
)/Ohm
Singlepotentiostats/galvanostats
Batteries Corrosion Coatings Electrolysis Generalelectrochemistry
Applications
Options-UltraLowCurrent:
fullscalerangedownto1pA-Built-inEISanalyzer:
upto7MHz-LinearScanGenerator:
1MV/s,acquisition1µs-SP-240suitcase
Education/training Generalelectrochemistry
(RRDEmeasurement) Sensors Corrosion Electrolysis/anodizing Coatings
Applications
Options-UltraLowCurrent:
fullscalerangedownto1pA-Built-inEISanalyzer:
upto7MHz-Internalamplifiers:
±1A/±48V,±2A/±30V,±4A/[-3;14]V,±10A/[0;5]V
-Bi-potentiostat-LinearScanGenerator:
1MV/s,acquisition1µs-SP-300suitcase
Single potentiostats/galvanostats
ELECTROCHEMISTRY INSTRUMENTS8/9
ModulesPG-581 SP-50/SP-150 SP-200/ SP-240/SP-300
Std ElS LC* Booster(1) Std ElS ULC** Booster(1) LSG***EducationGeneral electrochemistryAnalytical electrochemistryCorrosionNanotechnology/sensorsBatterySupercapacitorFuel cellsSolar cellsElectrolysisCoatingsMaterialsFor each instrument, modules can be mixed together
suitable recommanded excellent* LC: Low Current ** ULC: Ultra Low Current *** LSG: Linear Scan Generator(1): Booster, more details on page 16
PG-581 SP-50/SP-150 SP-200/SP-240/SP-300
FeaturesMax. number of channels 1 1 1 (SP-300: 2)EIS capability n.a. 10 μHz to 1 MHz (SP-50: n.a.) 10 μHz to 7 MHzLinear Scan Generator n.a. n.a. yes (1 MV/s), sampling rate 1 μsFloating option yes n.a. yesFilters n.a. numeric analog/numericAcquisition time 10 μs 20 μs 12 μs (1 μs with LSG option)Electrodes connections 2, 3 2, 3, 4, 5 2, 3, 4, 5IR compensation n.a. manual, EIS, current interrupt
(software)manual, EIS, current interrupt (software and hardware)
CurrentMaximum ±20 mA ±800 mA SP-200: ±500 mA,
SP-240: ±4 A, SP-300: ±10 A with 10 A booster
Current ranges
with standard board 8: 1 nA to 10 mA 6: 10 μA to 1 A 9: 10 nA to 1 Awith low current option - 9: 1nA to 100 mA (SP-50: n.a.) 13: 1 pA to 1 A
Lowest accuracy
with standard board ±10 pA on 1 nA range ±20 nA on 10 μA range ±100 pA on 10 nA rangewith low current option - ±2 pA on 1 nA range ±100 fA on 1 pA range
Lowest resolution
with standard board 61 fA 0.8 nA on 10 μA range 0.8 pA on 10 nA rangewith low current option - 80 fA on 1 nA range 80 aA on 1 pA range
Current booster
internal n.a. n.a. SP-200: n.a., SP-240: 4 A, SP-300: 1 A, 2 A, 4 A, 10 A
external n.a. 2, 5, 10, 20, 80, 100 A, 2 kW load box, (SP-50: n.a.)
n.a.
Input impedance 0.1 TΩ (//5 pF) 1 TΩ (//20 pF) Standard: 1 TΩ (//10 pF), ULC: 100 TΩ (//6 pF)
VoltageCompliance ±8 V ±10 V ±12 V (±48 V with 48 V booster)Max applied potential ±2 V (±8 V in option) 0-20 V ajustable
SP-50: n.a.SP-200: ±10 V, SP-240: -3;10 V, SP-300: ±48 V with 48 V booster
Resolution 61 μV on ±2 V 5 μV on 200 mV 1 μV on 60 mVAccuracy < 5 mV on ±2 V < 5 mV on ±2.5 V < ±1 mVRange ±2 V (±8 V in option) ±2.5 V, ±5 V, ±10 V ±25 mV, ±250 mV, ±2.5 V, ±5 V, ±10 VMaximum scan rate 10 V/s 200 V/s 200 V/s (1 MV/s with LSG option)
Control amplifierPotentiostat bandwidth 1.6 MHz 1 MHz 8 MHzPotentiostat rise/fall time 1 V/μs into 1 kΩ < 2 μs < 500 ns
GeneralI/O (analog/TTL) n.a./n.a. 3/2 3/2Interfaces USB 2.0 Ethernet, USB 2.0 Ethernet, USB 2.0Dimensions (W x D x H) 84 x 180 x 52.5 mm 136 x 377 x 197 mm SP-200: 167 x 410 x 225 mm,
SP-240/SP-300: 205 x 410 x 225 mmPower requirements 15 W 110 W 350 WWeight (chassis + board) 0.3 kg SP-50: 4.0 kg, SP-150: 4.5 kg SP-200: 6 kg, SP-240/SP-300: 7.5 kgn.a.: not available
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Applications
Specification overview
Hig
h c
urr
ent
inst
rum
ents
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
atte
ry
test
sta
tions
Boo
ster
s S
oftw
are
pa
ckag
eFu
el c
ell
test
sta
tions
Mat
eria
ls &
liqu
ids
test
sys
tem
sS
can
nin
g
syst
ems
ELECTROCHEMISTRY INSTRUMENTS
VSP Affordable research grademultichannelsTheVSPisaresearchgradepotentiostatsystemina5-channelmodularchassis.ThisallowstheVSPtoaddadditionalcapabilityinthewayofplug-in“modules”.ThesemodulesprovideEIScapability(includedonthepotentiostatboard),lowcurrentmeasurement(uses1additionalslot),4Acurrentbooster(twoslots)andadditionalpotentiostatchannels(eachonetakesoneslot).Theinternal4AoptiontakestwoslotsintheVSPchassisandrequiresapotentiostatboardtobeusedwithit.Externalboostersfrom2Aupto100AcanbeusedwitheachchannelintheVSP.
VMP3 16-channelresearch gradepotentiostatsystemTheVMP3isaresearch-grademulti-channelpotentiostat.Withitsmodularchassisdesign,upto16independentpotentiostatchannelscanbeinstalled.TheVMP3canbeequippedwithadditionalcapabilities,includinglowcurrentmeasurement,impedance.EachpotentiostatinstalledintheVMP3canbeconnectedtoanexternalhighcurrentboosterchannel.TheVMP3isalsoamultipleusersystem.ThankstotheEthernetLANconnectioncapability,severalcomputerscanbeconnectedtotheunitatthesametime.
Multipotentiostats/galvanostats
Electroanalytical Generalelectrochemistry
Corrosion Sensors
Energysources Energystorage
Batteries Coatings
Applications
Options-Lowcurrent:
fullscalerangedownto1nA-Built-inEISanalyzer:
upto1MHz-Externalcurrentboosters:
2,5,10,20,80and100A-ExternalloadboxLB-2000(50V/-150A)-Internal±4AboosterkitforVSPonly-SAM50forstack(50V)measurement
(upto30elementswith3SAM-50)forVMP3only-Additionalpotentiostat/galvanostat/EIS
ELECTROCHEMISTRY INSTRUMENTS10/11
VSP-300Versatile high-end multipotentiostatTheVSP-300multichannelpotentiostat/galvanostat/EISisaversatileinstrumentoffering6slotstosetfrom1to6channelboards.EachchannelboardcanaccomodateanUltraLowCurrentcableandcanbeassociatedwithoneorseveralboosterkits.Upto4boosterkitscanbepluggedinparalleinoneVSP-300chassis.
VMP-300 UltimatemultichannelelectrochemicalworkstationTheVMP-300isthemostmodularchassisofourproductrangeoffering16slotsforpotentiostats/galvanostats/EISboardsand/orboosterboards.Thechannelsandboosterboardscanbecombinedinthechassiseithertohavemanychannelsortoreachhighcurrents.EISmeasurementscanbeaddedasanoption.Thebuilt-inEIShasawidefrequencyrangeupto7MHz.LowcurrentsensitivitycanbeimprovedusingtheUltraLowCurrentoption.
Ewe/V-0.5-1-1.5
log
(|<I>
/mA|
)
0
-2
-4
-6
-8
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
s
Batteries/supercapacitors Fuel/photovoltaiccells Generalelectrochemistry Corrosion Sensors Materials Energystorage
Applications
Options-UltraLowCurrent:fullscalerangedownto1pA(resolutionof80aA)-EISmeasurement:
upto7MHz-LinearScanGenerator:
1MV/s,acquisition1µs-Currentboosters:
±1A/±48V,±2A/±30V,±4A/[-3;14]V,±10A/[0;5]V-Additionalpotentiostat/galvanostat/EIS
Hig
h c
urr
ent
inst
rum
ents
Bat
tery
te
sts
tatio
nsB
oost
ers
Sof
twar
e
pack
age
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
BiStat 3200High quality bipotentiostatfordemandingresearchapplicationsWithitstrue2-channelmeasurementcapability,theBiStat 3200providesthefeaturesfoundinatraditionallaboratorybasedinstrument,inadualdesign.TheBiStat 3200canoperateastwoindependentisolatedinstrumentsorasabipotentiostat/bigalvanostat.Thelowestfullscalerangeis1nA(resolutionof61pA).
SensorStatDedicatedtosensorapplicationsThemodulardesignprovidesauserconfigurablesystemfordemandingresearchapplications.TheSensorStatinconfigurablefor8to14channels.Byconnectinga14-channelSensorStattootherSensorStatunits,anextendedsystemiscreated.A98-electrodesensorwaycanbeindividuallyaddressedbychainingtogetherseven14-channelsystems.
Multipotentiostats/galvanostats
Ewe/V0.50-0.5
I/nA
0
-20
-40
-60
-80
-100
-120
Electroanalyststudies Dualelectrodevoltammetry Ring-diskelectrodestudies Electrochemicalsensors Biomedical Electroanalytical Corrosion
Applications
Options-RC10Vrotatorcontroller-±8Vscanrange-Floating-±8Vcompliance
Multi potentiostats/galvanostats
ELECTROCHEMISTRY INSTRUMENTS12/13
BiStat 3200/SensorStat VSP/VMP3 VSP-300/VMP-300Modules Std ElS LC* Booster(1) Std ElS LC* Booster(1)
EducationalGeneral electrochemistryElectroanalyticalCorrosionNanotechnology/sensorsRRDE/catalysisBattery/supercapacitorFuel cells/stacksSolar cellsElectrolysisCoatingsFor each instrument, modules can be mixed together
suitable recommanded excellent* LC: Low Current (1): Booster, more details on page 16
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
s
Applications
BiStat 3200/SensorStat VSP/VMP3 VSP-300/VMP-300
FeaturesMaximum number of channels BiStat 3200: 2,
SensorStat: 14VSP: 5, VMP3: 16
VSP-300: 6, VMP-300: 16
EIS capability n.a. 10 μHz to 1 MHz 10 μHz to 7 MHzLinear Scan Generator n.a. n.a. yes (1 MV/s), sampling rate 1 μsFloating option yes CE to Ground yesFilters n.a. numeric analog/numericAcquisition time 10 μs 20 μs 12 μs (1 μs with LSG option)Electrodes connections 2, 3 2, 3, 4, 5 2, 3, 4, 5IR compensation n.a. manual, EIS, current interrupt
(software)manual, EIS, current interrupt (software and hardware)
CurrentMaximum current ±20 mA ±400 mA ±500 mA (VMP-300: 150 A with
fifteen 10 A boosters)Current ranges
with standard board 8: 1 nA to 10 mA 6: 10 μA to 1 A 9: 10 nA to 1 Awith low current option - 9: 1nA to 100 mA (SP-50: n.a.) 13: 1 pA to 1 A
Lowest accuracy
with standard board ±10 pA on 1 nA range ±20 nA on 10 μA range ±100 pA on 10 nA rangewith low current option - ±2 pA on 1 nA range ±100 fA on 1 pA range
Lowest resolution
with standard board 61 fA 0.8 nA on 10 μA range 0.8 pA on 10 nA rangewith low current option - 80 fA on 1 nA range 80 aA on 1 pA range
Current booster
internal n.a. 4 A for VSP only 1 A, 2 A, 4 A, 10 Aexternal n.a. 2, 5, 10, 20, 80, 100 A,
2 kW load boxn.a.
Input impedance 0.1 TΩ (//5 pF) 1 TΩ (//20 pF) 1 TΩ (//10 pF), ULC: 100 TΩ (//6 pF)
VoltageCompliance ±8 V ±10 V ±12 V (±48 V with 48 V booster)Max applied potential ±2 V (±8 V in option) 0-20 V ajustable ±10 V (±48 V with 1 A/48 V booster)Resolution 61 μV on ±2 V 5 μV on 200 mV 1 μV on 60 mVAccuracy < 5 mV on ±2 V < 5 mV on ±2.5 V < ±1 mVRange ±2 V ±2.5 V, ±5 V, ±10 V ±2.5 V, ±5 V, ±10 V, ±25 mV, ±250 mVMaximum scan rate 10 V/s 200 V/s 200 V/s (1 MV/s with LSG option)
Control amplifierPotentiostat bandwidth 1.6 MHz 1 MHz 8 MHzPotentiostat rise/fall time 1 V/μs into 1 kΩ < 2 μs < 500 ns
GeneralI/O (analog/TTL) n.a./n.a. 3/2 3/2Interfaces USB 2.0 Ethernet, USB 2.0 Ethernet, USB 2.0Dimensions (W x D x H) BiStat 3200: 213 x 110 x 240 mm,
SensorStat: 450 x 320 x 135 mmVSP: 435 x 335 x 95 mm,VMP3: 495 x 465 x 260 mm
VSP-300: 254 x 517 x 337 mm,VMP-300: 534 x 565 x 315 mm
Power requirements BiStat 3200: 15 W, SensorStat: 130 W
VSP: 300 W,VMP3: 650 W
VSP-300: 650 W,VMP-300: 1500 W
Weight (chassis + pstats) BiStat 3200: 2.1 kg, SensorStat: 14 kg
VSP: 8 kg,VMP3: 20 kg
VSP-300: 20 kg,VMP-300: 30 kg
n.a.: not available
Specification overview
Hig
h c
urr
ent
inst
rum
ents
Bat
tery
te
sts
tatio
nsB
oost
ers
Sof
twar
e
pack
age
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
Boosters
VMP3 technology Deliver more powertoyourapplicationwithahighcurrentboosterSP-150,VSPandVMP3potentiostatscanbeinterfacedtoaseparatecurrentboosterunit.Thesemodularboosterunitscanhavedifferentboostermodulesplacedinsidethem(2,5,10,or20A).Inthestandardboosterchassisthereare8availableboosterslotsandthecurrentlimitis20A.Eachboosterchannelisconnectedtoapotentiostatboard.Thereisan80or100Aboosterunit.The80Aand100Aboosterexistalsoasastandalonesystem(HCP-803,HCP-1005seepages15/16).One4AboosterkitcanbedirectlyinsertedintheVSPchassis.
Configurations-External:±2A,±5A,±10A,±20Aon±10Vadjustablefrom–20to+20V
±80Aon±3V±100Aon(0.6-5)V
-Internalkit: ±4Aon20VonlyforVSP
VMP-300 technologyHigh current and high voltage boosters. From1 pA to 150 Aandfrom5 V to 48 VArangeofinternalboostershasbeendesignedtoincreasethecurrentandthevoltagespecificationsoftheSP-300,VSP-300,andVMP-300.Fourmodelsareavailable:±1A/±48V,±2A/±30V,±4A/[-3;14]V,±10A/[0;5]V.Dependingontheinstrumentchassis,severalsimilarboosterscanbeconnectedinparalleltoexpandthemaximumcurrentofthesystem.Thisfeatureisavailableforthe2A,4Aand10Aboosters.
Configurations±1A/±48V±2A/±30V±4A/[-3;14]V±10A/[0;5]V
-Boosterrangeincludedintheautoranging
-EIScapabilityupto2MHz-Plug-inmodules-Plug-and-play-5-leadconnectiontype-Parallelboosterstoincrease
themaxcurrent
General features
-Boosterrangeincludedintheautoranging(forboostersupto20A)
-EIScapability-Plug-inmoduleorexternalchassis-Plug-and-play-5-leadconnectiontype
General features
Boosters high current and high voltage
ELECTROCHEMISTRY INSTRUMENTS14/15
±1 A/±48 V ±2 A/±30 V ±4 A/[-3;14] V ±10 A/[1;6] V
CurrentCompliance ±1 A ±2 A ±4 A ±10 AAccuracy < 2 mA on 1 A range < 4 mA on 2 A range < 8 mA on 4 A range < 60 mA on 10 A range
VoltageCompliance ±49 V ±30 V -3 V ; +14 V -1 ; +6 VControl ±48 V ±30 V -3 V ; +10 V -1 ; +6 V
FeaturesEIS frequencies 10 μHz - 2 MHz 10 μHz - 1 MHz 10 μHz - 1 MHz 10 μHz - 1 MHzBandwidth (-3 dB) > 2 MHz > 3 MHz > 4 MHz > 8 MHzSlew rate (no load) > 15 V/μs 50 V/μs 50 V/μs 50 V/μsRise/fall time (no load) < 250 ns < 200 ns < 200 ns < 200 nsFloating mode yes yes yes yesParallel ability no yes yes yesConnection 2, 3, 4, 5 leads 2, 3, 4, 5 leads 2, 3, 4, 5 leads 2, 3, 4, 5 leads
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
oost
ers
VMP-300 boosters
2/4/5 A 10/20 A 80/100 A
Cell controlConnection 2, 3, 4, 5 terminal leads 2, 3, 4, 5 terminal leads 2, 3, 4, 5 terminal leadsCompliance adjustable ±10 V range adjustable ±10 V range 80 A: ±3 V
100 A: 0.6 - 5 VMaximum current 2 A: ±2 A,
4 A: ±4 A,5 A: ±5 A
10 A: ±10 A,20 A: ±20 A
80 A: ±80 A,100 A: ±100 A
Maximum potential ±20 V ±20 V ±3/5 VRise time and fall time
potentiostat 15 μs 25 to 60 μs 95 μs to 1.7 msgalvanostat 40 μs 50 to 120 μs 150 μs to 4.0 ms
MeasurementCurrent accuracy 2 A: < 4 mA on 2 A range,
4 A: < 8 mA on 4 A range,5 A: < 10 mA on 5 A range
10 A: < 20 mA on 10 A range,20 A: < 40 mA on 20 A range
80 A: < 160 mA on 80 A range, 100 A: < 200 mA on 100 A range
Potential noise (peak to peak 0-100 kHz)
0.6 mV 0.6 mV 0.6 mV to 0.15 mV (0-10 kHz)
Current noise (peak to peak 0-100 kHz)
1 mA @ 2/4/5 A 10 A, 6 mA @ 20 A 20 mA @ 80 A, 5 mA (0-10 kHz)
ElectrometerBandwidth 1 MHz 1 MHz 1 MHz
EISMax frequency (accuracy 1%, 1°) 2 A: up to 150 kHz,
4 A: up to 130 kHz,5 A: up to 120 kHz
10 A: up to 80 kHz,20 A: up to 80 kHz
80 A: up to 15 kHz,100 A: up to 10 kHz
Amplitude 0.5 mV to 0.5 V - 0.1% to 50% of the current range
0.5 mV to 0.5 V - 0.1% to 50% of the current range
0.5 mV to 0.5 V - 0.1% to 50% of the current range
General1 external input security to open circuit (TTL level) security to open circuit (TTL level) security to open circuit (TTL level)
Emergency push buttonPower 1,000 W 1,000 W 1,000 WChassis dimensions (W x D x H) 495 x 465 x 280 mm 495 x 465 x 280 mm 495 x 465 x 280 mmChassis weight 24 kg 24 kg 24 kgBooster board weight 2 A: 1 kg, 4 A: 0.85 kg, 5 A: 1 kg 10 A: 2 kg, 20 A: 4.2 kg 80 A: 4.8 kg, 100 A: 9 kg
VMP3 boosters
Hig
h c
urr
ent
inst
rum
ents
Bat
tery
te
sts
tatio
nsS
oftw
are
pa
ckag
eFu
el c
ell
test
sta
tions
Mat
eria
ls &
liqu
ids
test
sys
tem
sS
can
nin
g
syst
ems
HCP-803 Highcurrentforsupercapacitorsandfuel cellsTheHCP-803isaHighCurrentPotentiostatcapableofhandling±80Awithavoltagerangeof±3V.Itisprimarilydesignedforapplicationsinthefuelcellandsupercapacitorareas.Itisacombinationofaresearchqualitypotentiostatandan80Aboosterbuiltintothesamechassis.ThepotentiostathasthesamespecificationsastheVMP3potentiostatboards(withEISoption)whennotconnectedtotheboosterportionoftheunit.
HCP-1005 Morepowerforbattery testingTheHCP-1005isacompactHighCurrentPotentiostatspeciallydesignedtostudysecondarybatterieswithahighcapacity.Withavoltagerangeof0.6to5Vandacurrentrangeof±100A,thisunitcanbeusedtotestindustrialLi-ionbatterymodules.TheEIScapabilityintegratedinthechassisisidealforageingtests.TheHCP-1005structureissimilartotheHCP-803.Itisacombinationofaresearchpotentiostatanda100Aboosterbuiltintothesamechassis.
CLB-2000 150A/50Velectronicloadforstack testingTheCLB-2000isa150A/50Vcontrolledelectronicloaddesignedtostudystacksoffuelcellsandsolarcells.TheCLB-2000couplesa2kWelectronicloadwithanEIScapablepotentiostatinthesamecompactchassis.ThisunitcanbeusedinconnectionwithaVMP3toperformsimultaneousmeasurements(ACandDC)onindividualelementsofthestacks(upto30elements).
PL
UG
- IN
HY
BR I D E L E C T R I C V
EH
I CL
E
mpg
Highcurrentinstruments
Fuelcells Photovoltaicsystems
Applications Supercapacitors Electroplating Battery
Lithium-ion Nickel-Cadmium
Applications Nickel-Metalhydride Battery
Fuelcells Solarcells Photovoltaicsystems Stacksofcells
Applications-SAM-50:voltageadapter(50V)tobeusedwithVMP3
Options
High current instruments
ELECTROCHEMISTRY INSTRUMENTS16/17
Modules HCP-803 HCP-1005 CLB-2000Batteries/intercalation compoundsSupercapacitorsFuel cellsPhotovoltaic cellsStack of fuel & photovoltaic cellsElectrolysisCoatings
suitable recommanded excellent
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Hig
h cu
rren
tin
stru
men
tsM
ult
ipo
tent
iost
ats/
galva
nost
ats
Applications
HCP-803 HCP-1005 CLB-2000
FeaturesEIS capability 10 μHz to 15 kHz 10 μHz to 10 kHz 10 μHz to 10 kHzAcquisition time 20 μs 200 μs 200 μsElectrodes connections 2, 3, 4, 5 2, 3, 4, 5 2, 3, 4, 5IR compensation manual, EIS, current interrupt manual, EIS, current interrupt manual, EIS, current interrupt
CurrentMaximum current ±80 A ±100 A -150 ACurrent ranges 7: 10 μA to 80 A 7: 10 μA to 100 A 7: 10 μA to 150 AAccuracy < 160 mA on 80 A range < 200 mA on 100 A range < 750 mA on 150 A rangeBest resolution (meas) 6.4 mA on 80 A range 8 mA on 100 A range 6 mA on 150 A rangeInput impedance 10 MΩ (//20 pA) 10 MΩ (//20 pA) n.a.
VoltageMax applied potential ±3 V 0.6-5.0 V 50 VResolution 5 μV on 200 mV 5 μV on 200 mV 400 μV on 0-50 V rangeAccuracy 10 mV on ±5V range mV on 0-5 V range 50 mV on 0-50 V range
Control amplifierBandwidth with boosters 10 kHz 10 kHz 10 kHzPotentiostat rise/fall time 95 μs 1.7 ms n.a.
GeneralI/O (analog/TTL) 3/2 3/2 3/2Temperature control n.a. with analog I/O With analog I/OTemperature probe n.a. PT100 n.a.Interfaces Ethernet, USB 2.0 Ethernet, USB 2.0 Ethernet, USB 2.0Dimensions (W x D x H) 495 x 465 x 280 mm 495 x 465 x 280 mm 495 x 465 x 280 mmPower requirements 1000 W 1000 W 100 WWeight (chassis + pstats) 23 kg 25 kg 28 kg n.a.: not available
Specification overview
Boo
ster
s B
atte
ry
test
sta
tions
Sof
twar
e
pack
age
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
Batteryteststations
Accordingtoyourapplications,yourrequirementsforbattery/supercapacitorinvestigationaredifferent(referenceelectroderequiredornot,sampling,maximumfrequencyforEIS,channelnumber…).
That’swhy,Bio-Logicisofferingseveralsolutiontomeetyourneed.
VMP3/VMP-300ismoreforadvancedmeasurement(researchlevel)andBCS-8xxoffersthestandardofthecyclerwithresearchEISmeasurement.MPG-2xxisasolutionin-between.
Forhigh-throughputmeasurement,BCS-8xx wouldberecommended(upto128channelscanbeputinasinglecabinet).NoteEC-Lab®allowstheusertocontrolseveralVMP3orMPG-2xxinstrumentsfromonesession.
VMP3/VMP-300 MPG-2xx series BCS-8xx seriesCurrent ranges 1 pA - 150 A 10 μA to 40 A 10 μA – 120 AVoltage range ±10 V -2;9 V 0;9 VEIS frequency range 7 MHz – 10 μHz 20 kHz – 10 μHz 10 kHz-10 mHzIn-built EIS Yes on each channel Yes on each channel Yes on each moduleAcquisition time 200 μs/1μs with LSG option 200 μs 2 msElectrode connection 2,3-electrode/4-points
measurement with CE measurement
2-electrode/4-points measurement
2-electrode/4-points measurement
HPC measurement Yes. Down to 6 ppm Yes. Down to 6.3 ppm Yes. Down to 6 ppmMax Channel number 16 16 128Application High end research Research development High-throughput battery/material
screening
Specification overview
MPG-2xx series R&D gradebatteryteststationsTheMPG-2xx seriesisproposedasarangeof5units(infixedconfigurations)withorwithoutEISoneverychannels:-MPG-2: 16channels/100mAeach,-MPG-205: 8channels/5Aeach,-MPG-210: 4channels/10Aeach,-MPG-220: 2channels/20Aeach,-MPG-240: 1channel/40A.
TheMPG-2xx seriescanbeprovidedinarackcapableofsupportingupto5units.OnlyonecomputerisnecessarytocontrolalltheunitsthankstotheEthernetconnection.Withthisconnection,theMPG-2xx seriescanbeinstalledonaLocalAreaNetworktoallowmultipleuserstoaccesstheinstrumentsandfollowthebatterycyclingfromanywhere.TheMPG-2xx seriesoffersatemperaturemeasurementandthreeoptionalconnectionmodestothebattery(batteryholder,shortorlongcables).Eachchannelhastwoanaloginputsandoneanalogoutputtoallowinterfacingwithexternalinstruments.TheMPG-2xx seriesissuppliedwithEC-Lab®software,developedforbatteryandsupercapacitorapplications.Mostofthetechniquesaredesignedspecificallyforbatteries.Specificanalysistoolsarealsoavailable.
ELECTROCHEMISTRY INSTRUMENTS18/19
MPG-2
MPG-205
MPG-210
MPG-220
MPG-240
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
atte
ry
test
sta
tions
-Currentranging:10µAuptomaxcurrentwitharesolution0.004%oftherange
-0-9Vreferencevoltage-Resolutionof300µVprogrammable
downto5µVbyadjustingthedynamicrange(100µVresolutionon5Vrange)
-Acquisitiontime:200µs-Nolimitintimeanddatarecording
Specifications
Boo
ster
s H
igh
cu
rren
t in
stru
men
ts
Options-EIS-Rack(5units)-Short(25cm)orlongcables(2.5m)-Temperatureprobe-BH-1iBatteryHolder
Sof
twar
e
pack
age
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
-HighqualityEIS:Fullscanfrom10kHzto10mHz-18-bitA/Dconverter(40µVresolution)-HPCmeasurementdownto6.3ppm-ModularityfromfewµAto120A-Voltagemeasurementfrom0Vto9V-Modulemixing(BCS-805/810/815)-Powerfulinterface-2msacquisitiontime-Severalcabinetsizes-Plugandplaymoduleinstallation
Specifications
m
easurementEIS
Batterycyclers
BCS-8xx seriesBatterycyclingteststations
EachBCS-8xxmoduleiscomposedof8channels.
Toensurebetteraccuracyincurrentcontrolandmeasurement,5 current rangesareavailabledependingonthemodel.
ChannelsofBCS-815modulescanbeconnectedinparalleltoincreasethemaximumcurrentupto120 A.
Withan18-bitanalogtodigitalconverterforthevoltagemeasurement,theresolutionoftheBCS-8xx isanimpressive40 µV.
Toaddevenmorecapabilityandvalue,everychannelinamoduleisEIS-capableoverafrequencyrangefrom10 kHz to 10 mHzforaccurateandfastdeterminationofthebatteryinternalresistance.
BCS-8xxmodulescanbeaddedtoasinglecabinet.Severalsizesareoffered(38U,24U,12Uand6U).
Connection:-Cellcablefrom25cmto10m-CCH-1xxCoincellholder-BH-1iCylindricalbatteryholder-CC8Currentcollectortosetparallelmode(upto±120A).
Cabinet:-Rollingcabinet(38U,24U)-Benchtopcabinet(12U,6U)
Options
ELECTROCHEMISTRY INSTRUMENTS20/21
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
atte
ry
test
sta
tions
Boo
ster
s H
igh
cu
rren
t in
stru
men
ts
Impedance measurement for battery testing, why...?
BCS-8xx series,withitsfastbandwidthcontrollers,offersbothpotentio-andgalvano-controlforEISmeasurementsin10 kHz to 10 mHzfrequencyrange,withuser-specifiednumberofdatapointsandfrequencyrangetoberecorded.Userisalsoincontroloftheexcitationamplitude.Impedancespectroscopycanbeusedasastandalonetechnique,butitcanbealsoseamlesslyincorporatedintoourModuloBattechniquetorecordtheimpedancespectrumofthebattery,eitherinequilibriumorsteady-stateconditions.Thatmeansyounotonlycanrecordaspectrumofthecellataconstantvoltage,butalsounderagivengalvanostaticload.
Iftherearesourcesofuncompensated inductanceorcapacitanceinthecell,onlyfullimpedancespectrumcanbeareliablemeasureorinternalresistancethroughanautomatedspectrumfittingprocedure.
LowfrequencyrangeofthespectrumisaninvaluablesourceofinformationaboutLi+ diffusionwithintheelectrolyte(separator),aswellasporositywithinelectrodesandinsidetheactivematerialparticles..
18-bit resolutionallowsrecordingsmall-amplitudeEISontopofsubstantialDCcurrentswithverygoodaccuracy.
Highfrequency Lowfrequency
Diffusionprocess
Internalresistance
Inductionbehavior
Sof
twar
e
pack
age
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
AcomprehensivesoftwarepackageEC-Lab®isanadvancedsoftwarepackageforperformingelectrochemistrymeasurements,whereasBT-Lab®isdedicatedtobatterycyclingsystem(BCS-8xxseries).Withmorethan15yearsofdevelopmentandconstantimprovementintechniquesandfeatures,EC-Lab®softwarehasbecomethebenchmarkinpotentiostatcontrolsoftware.
DisplaymodeTwoviewmodesareavailableinEC-Lab® /BT-Lab®.Thesetupofexperimentscanbedonewithatechniquemenu,orbyinputtingparametervaluesintoaspreadsheet.Mostoftheexperimentalparameterscanbemodified“onthefly”duringtheexperiment,withthechangesstoredintotherawdatafile.Theappearanceofthesoftwareinterfaceisadjustabletocreatethebestworkingenvironmentfortheuser.
ExperimentalsequencebuilderEC-Lab®softwarecontainsmorethan80techniques.Thesetechniquescanaddressapplicationsinvoltammetry,EIS,corrosionandenergysourcedevelopment.Apowerfultechniquebuildercanexecuteaseriesofdifferentmodulartechniques,waitandloopfeaturestocreatecomplexexperimentalsequences.Evenwithineachtechnique,theusercancreateupto100linkablesequencesofthatexperimentwithdifferentparameters.Whenacertainleveloftheexperimentisreached,anemailcanbesenttotheuser.
EISmeasurementsEISmeasurementscanbemadeinbothcontrolledpotentialandcontrolledcurrentmodesfrom10µHztoaremarkable7MHz.Thepatented“driftcorrection”algorithmandmultiplestabilityparametersallowuserstoacquirehighqualitydatafromtheirEISmeasurements(moreinformationinpage22).
EC-Lab® /BT-Lab®
Powerfulandcomprehensiveadvancedsoftware
Impedance spectroscopy
General electrochemistry
Electro-analytical
Corrosion
Battery
Super-capacitor
Fuel cell
Photovoltaic cell
Singlepotentiostats/galvanostats:SP-50,SP-150,SP-200,SP-240,SP-300Multipotentiostats/galvanostats:VMP-300,VSP,VMP3,VSP-300Highcurrentpotentiostats/galvanostats:HCP-803,HCP-1005,CLB-2000Batteryteststations:MPG2,MPG-205,MPG-210,MPG-220,MPG-240
EC-Lab® provided with...
Softwarepackage
Batterycyclingsystems:BCS-805,BCS-810,BCS-815
BT-Lab® provided with...
ELECTROCHEMISTRY INSTRUMENTS22/23
Display…EC-Lab®’sgraphicpackageprovidedwiththesoftwareincludesapowerful3-Dimensionplotfeatureandatooltocreateuniquegraphtemplates.Usingouradvanced“Process”function,theusercancreatenewvariablesforeachaxis.Thisenablesmathematicalfunctionstobeperformedondataplottedonanyaxis(x,y1andy2).
…fitting…Powerfulelectro-analyticalanalysistools(suchaspeakfind/height,convectionwave,integral,Tafelfit,Rpdetermination)areavailableinEC-Lab®.Theseanalysesincorporateclassicalfitroutines(linear,polynomial,multi-exponential)andalgorithms.Alltheanalysisresultsarestoredinaseparatefile.EC-Lab®’sEISmodelingpackage,ZFit,utilizestheequivalentcircuitapproach.Thereareover150standardcircuitsandtwominimizationsalgorithmsavailableforuseinunderstandingimpedanceplotinformation.Theusercandefineandbuildhisowncircuitmodelusingarangeofthirteensimpleelements(R,C,L,Q,W,G,Wd,M,Ga,Gb,La,Ma,Mq).Thelastelementscanbeassimilatedtoatransmissionlines.Abatchprocessingfeatureallowsthefittingofmultiplecyclesinanimpedanceexperiment.
…&simulationSeveraltoolsareavailabletosimulateCVcurve,TafelplotorEISdata.Theycanbeusedastrainingtools.CVSimallowtheusertocreatedatawithdifferentmechanismssuchassingle(E)ormulti(uptoEEEEE)electrontransfer.ElectrontransferreactioncanalsobemixedwithchemicalreactionelectrontransfertosimulateanECmechanism.
EC-Lab® graphicsAn extended range of analysis
Log (<I>/mA)-1-2-3-4-5-6
E we/V
0-0.1
-0.2-0.3
-0.4-0.5
-0.6
Ewe/V0.60.40.20-0.2
<I>/
mA
0.8
0.6
0.4
0.2
0
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sS
oftw
are
pa
ckag
e
PeakAnalysis...
CVFit...
WaveAnalysis...
General electochemistry
RPFit...
ElectrochemicalNoise...
CorrSim...
CASPFit...
VASPFit...
TafelFit...
Corrosion
Mott-Schottky...
ZFit...
Kramers-Kronig...
ZSim...
EIS
Processdata(capacity,efficiency,energy...)
Battery process
Photovoltaicanalysis(fillfactor,efficiency...)
Photovoltaic
LinearInterpolation...
Integral...
SubtractFiles...
LineFit...
PolynomialFit...
MinMax...
Multi-ExponentialFit...
Filter...
FourierTransform...
Math
Re(Z)/Ohm 40 01
23020
10
-Im(Z
)/Ohm
151050
Time/h
Boo
ster
s B
atte
ry
test
sta
tions
Hig
h c
urr
ent
inst
rum
ents
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
Electrochemical Impedance Spectroscopy Built-in EIS capabilitywithequivalentcircuitfitting
MeasurementsManyunitsintheelectrochemistryproductrangedevelopedbyBio-Logiccanbeequippedwithabuilt-inEISanalyzer.ThismeasurementisbasedontheDSPtechnologythatmixesaccuracyandfastsampling.Thisfeatureresultsinseveralmajortoolsdevelopedinourcontrolsoftware:singlesineandmultisineEIStechniques,FastFourierTransform(FFT)basedintegration,cycleaveragingupto16frequenciesandapatenteddriftcorrection.
ControllingtechniquesACsinewavescanbesuperimposedonaDCpotentialoraDCcurrentvalue.Additionaltechniquesareavailablethatimposethesinewaveonapotential(SPEIS,alsoknownasMottSchottkytechnique)oracurrent(SGEIS)sweep.FinallyourEIStechniquescanbesequencedtoapplydifferentdefinedconditionsduringthefrequencysweep.
AccuracyAccuracyisgivenbytheerrorobservedonthephaseandonthemagnitude.Foralmostallofourpotentiostats,wespecify1%,1°accuracybelow500kHz,whichisreducedwithabooster.ForSP-200/300technology,theperformanceishigherandreaches0.3%,0.3°accuracyinthesamearea.
ModelingEC-Lab®softwareoffersZFit,amodelingtoolforequivalentcircuitfitting.Over150standardcircuitscombining13elementsandtwominimizationalgorithms(SimplexDownHillandLevenberg-Marquardt)areavailabletoinvestigateimpedancedata.ZFitandZSimtoolsareavailableinEC-Lab®demoversion.
1e+9
3% 3° 1pF
Frequency (Hz)
|Z| (
Ohm
)
1% 1°0.3% 0.3°
1e+8
1e+7
1e+6
1e+5
1e+5
1e+3
1e+2
1e+1
11e-1
1e-2
1e-3
101 100 1e+3 10e+3 100e+3 1e+6 10e+6
EIScontourplotofSP-300withstandardcable
Softwarepackage
-0,208 -0,0325 0,142 0,317 0,492 0,667
1,42
0,76
0,11
-0,54
-1,19
E/V
Curr
ent/m
A
ELECTROCHEMISTRY INSTRUMENTS24/25
UiEChem™:-CyclicVoltammetry,-LinearSweepVoltammetry,-Chronoamperometry,-Chronopotentiometry,-ChronoOCV,-Chronoauxiliarypotential,-SquareWaveVoltammetry,-NormalPulseVoltammetry,-DifferentialPulseVoltammetry
UiECorr™:-EcorrvsTime,-LinearPolarizationResistance,-TafelPlot,-PotentiostaticPolarization,-GalvanostaticPolarization,-PotentiodynamicPolarization,-ZeroResistanceAmmetry
UiEChem™/UiECorr™ TheUiEChem™/UiECorr™softwaresuppliedwiththePG-581,BiStat3200,SensorStat,isacomprehensiveandfullyfeaturedpackagewhichencompassesawideselectionofstandardelectrochemicaltechniques.Allthisiscombinedwithaneasy-to-usegraphicsinterface.Afteratechniquehasbeenselectedawaveformdiagramisdisplayedshowingthestructureoftheexperiment.Atthesametimetheusercanenterintheexperimentalparameters,setthesamplingrate,numberofcyclesandcurrentautoranging.Thesoftwareautomaticallyconfiguresitselftothenumberofpotentiostatchannelsinstalledintheunit.
AutosequencingTheUiEChem™/UiECorr™softwareprovidesaflexible,intuitiveinterfaceforrunningmultipleexperimentsinauser-definedsequence.Oncethesequenceisstartedtheexperimentsarerunautomatically,one-at-a-time,intheorderinwhichtheyappearinthesequence.
Graphicdisplay&analysisAstandardgraphicsdisplayisprovidedforeachtechniqueandwehaveaddedtheabilityfortheusertodefinetheirpreferredformatforeachtechniquefromasimpletemplate.TheUiEChem™softwarefeaturesapeakanalysisroutinewhichiseasytouse.
AdvancedwaveformgenerationTheinstrumentsarecapableofgeneratingcomplexandrepetitivewaveformswithprecisiontiming.Theinstrumenthasasequencedwaveformgeneratorwhichgenerateswaveformsbasedonasequenceofsteps.Eachstepmaycompriseawaveformwhichincludespulses,linearsweepsandarbitrarywaves.
-0.184 0.008 0.2 0.392 0.584
-31.76-42.18
-21.34-10.91
0
E/V
Curr
ent/m
A
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
s
Singlepotentiostats/galvanostats: PG-581Multipotentiostats/galvanostats: BiStat3200,SensorStat
Provided with...
Boo
ster
s B
atte
ry
test
sta
tions
Sof
twar
e
pack
age
Hig
h c
urr
ent
inst
rum
ents
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
FCT seriesTwomodelsfor 25 and 50 cm²PEMFC
TheFCT-50S/ZandFCT-150S/ZarecompactfuelcellteststationsdesignedforPEMFC(ProtonExchangeMembraneFuelCell)testing.Basedonfeedbackfromourcustomers,theyaredesignedtobethemostflexibleteststationsonthemarket.Thecompactchassisintegratesaprogrammableelectronicload(250W),gasandwatercontrolcircuitry(flow,pressureandtemperature)anddataacquisition.Theelectronicloadusesstate-of-the-arttechnologyandiscapableofreachingverylowvoltagelevels(virtually0V)at150A(forFCT-150S/Z)and50A(forFCT-50S/Z).Anumberofsafetyfeatureshavebeenincorporatedinthehardwareandsoftware.Toconductlong-termexperiments,suchasfuelcellageing,Bio-Logichasdevelopedanautomatedhumidifier,waterfillingandcondenserwaterpurging,allmanagedbyFC-Labcontrolsoftware.
SpecificationsFCT-50S/Z FCT-150S/Z
Max current 50 A 150 AMax voltage 5 V 5 VElectronic Load 250 W 250 WMinimum load resistor 0.7 mΩ 0.7 mΩBuilt-in EIS analyzer 10 μHz up to 10 kHz 10 μHz up to 10 kHzCurrent measurement accuracy 0.5% FSR 0.5% FSRCurrent resolution 4 mA 11 mAPotential resolution 76 μV 76 μVFuel gas flow control 0 - 40 L/h 0 -120 L/hOxidant gas flow control 0 -100 L/h 0 - 300 L/hBack pressure control 0 - 5 bars 0 - 5 bars5 temperature control cell, humidifiers, heaters cell, humidifiers, heatersInterfaces Ethernet EthernetSafety Alarms with automatic N2 purge under voltage,
over current, over temperature, over water
under voltage, over current, over temperature, over water
Fuelcellteststations
PEMFCdevelopmentandtesting Singlecellorsmallstacks
Applications
ELECTROCHEMISTRY INSTRUMENTS26/27
FC-LabPowerfulmonitoringsoftwareFC-Labistheoperatingsoftwareusedtocontrolandanalyzedatafromtheinstrument.Withaneasy-to-useinterface,itoffersallcontrolsandfunctionsforfuelcelltestingwithindividuallytabbedmenus.Thegraphicfunctionsofferareal-timedisplayofperformanceofthefuelcellwithrespecttoeachofthemeasuredorcontrolledvariables.DatacanbeanalyzedbyEC-Lab®fittingtools(ZFit...).
Re(Z)/mOhm10080604020
-Im(Z
)/mOh
m
40
30
20
10
0
-10
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sFu
el c
ell
test
sta
tions
Boo
ster
s
SpecificationsFC-Lab
Gas control and display independent for each cell and each gas
temperatures, pressures, flows
Facilities
current density display,on the fly parameter changes,multivariable graphic display
Loal Control
galvanostatic/galvanodynamic,potentiostatic/potentiodynamic,power/load
Techniques
open circuit voltage,voltage pulses,current pulses,voltage scan,current scan,potentio EIS,galvano EIS,constant load discharge,constant power discharge,macrotechniques (SGEIS, SPEIS),loop
Analysis EC-Lab®
Bat
tery
te
sts
tatio
nsS
oftw
are
pa
ckag
eFu
el c
ell
test
sta
tions
Hig
h c
urr
ent
inst
rum
ents
Mat
eria
ls &
liqu
ids
test
sys
tem
sS
can
nin
g
syst
ems
Impedanceanalyzer
MTZ-35 35MHzImpedanceAnalyzerMTZ-35isanimpedanceanalyzerdedicatedtomaterialcharacterizationoverthefrequency10 µHz to 35 MHz.TheMTZ-35canbecoupledwiththeHighTemperatureFurnaceHTF-1100inordertoinvestigatematerialspropertiesoverawidetemperaturerange(ambient to 1100°C).Twomainsampleholdersareproposed:HTSH-1100forhightemperatureuseandCESHforambienttemperatureusewithmaximumtemperatureof150°C. Specifications
Ceramics Sensors Polymers Nanomaterials Solar/photovoltaiccells Rubbers Glues Epoxies Liquidcrystals Ferroelectrics Biologicalcells Polarliquids....
Applications
Furnace and sample holders
HTSH-1100
CESHonitsbase
Generator
Frequencyrange Frequency Range: 10 μHz to 35 MHzVoltagerangeDC ±100 μV to ±10 V VoltagerangeAC 100 μV to 10 V Measurement ranges
Inductance 10 nH to 10 kHCapacitance 1 pF to 1000 μFResistance 1 mΩ to 500 MΩBasicaccuracy 0.1%
Product model name / part number Operating temperature FeaturesFurnace HTF-1100: 097-110 From Ambient to 1100 °C Controlled through MT-Lab software/MTZ-35
Temperature Scan: adjustable from 0.1 °C/min to 20 °C/min
High Temperature Sample Holder (HTSH-1100)
HTSH-1100: 097-120 From Ambient to 1100 °C Controlled atmosphere with quartz tube (Tmax= 900 °C)Two or four points measurements set-up
Controlled Environmental Sample Holder (CESH)
Surface Sample Holder: 097-150/1Bulk Sample Holder: 097-150/2
From -40 °C to 150 °C Controlled atmosphereTwo or four points measurements set-up
HTF-1100
ELECTROCHEMISTRY INSTRUMENTS28/29
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
sB
oost
ers
Bat
tery
te
sts
tatio
nsS
oftw
are
pa
ckag
eH
igh
cu
rren
t in
stru
men
ts
Conductivitymeter
MCS 10 FullyIntegratedMultichannelConductivitySpectroscopyMCS 10isamultichannelconductivityspectrometerdedicatedtostudyionic liquids and liquid electrolytesoverawidetemperaturerange.
TheConductimetrysystemcanaccommodateup to 10 cellsandworksinthetemperaturerangefrom-40 °C to +150 °Cwithahighaccuracylevelof±0.2°C.AnEIS diagram(NyquistandBode)canbedisplayedforeachmeasurementpoint.
Ionicliquidstudy Organicelectrolytestudy Liquidconductimetry
Applications
Fuel
cel
l te
st s
tatio
nsM
ater
ials
& li
quid
s te
st s
yste
ms
Sca
nn
ing
sy
stem
s
SpecificationsMCM 10
Conductivityranges 7 ranges from 0.2 - 2 μS/cm, to 0.2 - 2 S/cmConductivityaccuracy <4 % of the 2 μS/cm and the 2 S/cm range
<2 % of the 20 μS/cm and the 0.2 S/cm range<1 % of all other ranges
WTSH 10
Temperaturerange -40 °C to +150 °CTemperatureAccuracy(typ.) ± 0.1 °CExternalinputAuxiliary PT1000Numberofslots 10ExternalDimension 290 x 340 x 375 mm (W x D x H)HTCC
Celltype Parallel plates platinized platinum on glass holderConnections 2-wire connectionsCellfactor K = 1 (nominal)Minimalsamplevolume 0.5 - 1.0 mL
TheMCS10consistsofthreeintegratedcomponents:
MCM 10:FRA-basedimpedanceanalyzerwithtenchannelsWTSH 10:WideTemperatureSampleHolderwithtenslotsHTCC:HighTemperatureConductivityCells–Platinized/
PtParallelPlateElectrode
Scanning techniques Information Resolution ApplicationsSECM Scanning ElectroChemical
Microscopyreactivity probe size bio electrochemistry,
corrosion, fuel cells, batteries, catalysts
ac-SECM alternating current Scanning ElectroChemical Microscopy
topography or reactivity (no mediator required)
probe size bio electrochemistry, corrosion, fuel cells, batteries, catalysts
ic-SECM intermittent contact Scanning ElectroChemical Microscopy
topography and reactivity
probe size bio electrochemistry, corrosion, fuel cells, batteries, catalysts
LEIS Localised Electrochemical Impedance Spectroscopy
local impedance of the sample
tens of μm corrosion, coatings, catalysts
SVP (SVET) Scanning Vibrating Electrode Technique
electrochemical activity 10-15 μm corrosion, coatings, catalysts
SDS Scanning Droplet System dc electrochemistry in a droplet of electrolyte
hundreds of μm
corrosion, coatings
ac-SDS alternating current Scanning Droplet System
impedance in a droplet of electrolyte
hundreds of μm
corrosion, coatings
SKP Scanning Kelvin Probe work function difference / topography
probe size corrosion, coatings, semi-conductors, catalysts
OSP Optical Surface Profiler topography 20 nm any field
Modularsystems fortheelectrochemistry at the local scale
SECMareascan
Thepotentiostat/galvanostatgiveanaverageresponseoftheelectrodematerialanditisconsideredthatitishomogeneous.However,foroptimizationpurposes,itmightbeinterestingtoconfirmthishomogeneityandtostudythespatialdependenceoftheelectrodeproperties.Ourlocalelectrochemistryplatformisamodularsystemthatcanincludeupto9techniques.Thetablebelowsummarizesthetechniques,thetypeofinformationthatcanbeobtained,thesmallestsizeofthefeaturesthatcanbedetectedandthetypicalapplications.
57
46
-0.75
0.73
0.00
mm
µm
mm
1.80
Scanningsystems
M470 and SP-300 are now compatibleTheM470isnowavailablewithachoiceofpotentiostat/galvanostats.SupporthasbeenaddedforthehigherspecificationSP300standalonepotentiostat/galvanostataswellastheoriginalmodel3300,whichisintegratedintotheM470chassis.SelectingtheSP300optionoffersfurtherdccurrentmeasurementsensitivityandincreasedEISbandwidth.AsanSP300useryoumayconsiderpurchasingtheM470systemwithahardwareinterfaceforusewithyourexistingpotentiostat/galvanostat.TheSP300hardwareinterfacemoduleplugsdirectlyintotheM470modularchassisandprovidesalltherequiredconnectionstointerfacetheanaloguesignals,andcontrolfunctionalityrequiredbythesystem.TofacilitatemountingoftheSP300 lowcurrentamplifierslocallytothescanningprobecellconnections,weprovidebespokeadjustablestandsandbracketswhichmountdirectlyonthe M470baseplate.
Stand for ease of use
hardware interface
22
11
0.75-0.751.50
mm
nA
mm
0.00
OSPareascan
ELECTROCHEMISTRY INSTRUMENTS30/31
Sin
gle
po
tent
iost
ats/
galva
nost
ats
Mu
lti
pote
ntio
stat
s/ga
lvano
stat
s
Awidevarietyofoptionalaccessoriesarealsoavailable,includingvariousprobeoptions,celloptions(environmentalTriCell™andµTriCell™),longworkingdistanceopticalvideomicroscope(VCAM3)and3Dshadedsurfacerenderingsoftware(3DIsoPlot™).
Tools
M470M470isthe4thgenerationofscanningprobesystems,whichincludesahigh-resolutionscanningstageandadditionalprobetechniques.TheM470achievestheperfectbalanceofscanspeed,resolutionandaccuracytodeliveranewstandardinspatiallyresolvedelectrochemicalmeasurements.Thefastpreciseclosedlooppositioningsystemisdesignedspecificallyforthedemandsofscanningprobeelectrochemistrywithnanometerresolution.
9availabletechniques-SECM-ac-SECM-ic-SECM
-LEIS-SVET-SDS
-ac-SDS-SKP-OSP
Ultra high resolution scanning stagewithadditionaltechniques
M370 TheM370isamodularconfigurablesystemwhichwillperformallthekeyscanningprobeelectrochemicaltechniquesandlaser-basednon-contactsurfaceprofiling.TheM370isavailableinanyorallofthesixconfigurations(SECM,SVET,SKP,LEIS,SDS,OSP)andmaybeupgradedatalaterdatebysubsequentpurchaseofanycombinationoftheavailabletechniques.
6availabletechniques-SECM-LEIS
-SVET-SDS
-SKP-OSP
Versatilityandprecision
SpecificationsM470 M370
Number of techniques 9 techniques 6 techniquesPositioning resolution 20 nm 100 nmScanning range 110 mm 70 mmMax. scan speed 10 mm/s 2 mm/sPiezo positioning yes n.a.Potential range ±10 V ±2 VCurrent ranges 1 A to 1 pA 20 mA to 1 nAAnalog to digital converter resolution 24-bit 16-bitImpedance ability up to 1 MHz n.a.n.a.: not available
Boo
ster
s B
atte
ry
test
sta
tions
Sof
twar
e
pack
age
Fuel
cel
l te
st s
tatio
nsH
igh
cu
rren
t in
stru
men
ts
23
12
-3.753.65
mm
0
10
kΩ c
m²
mm
LEISareascan
Mat
eria
ls &
liqu
ids
test
sys
tem
sS
can
nin
g
syst
ems
Sales&technicalcontacts
Bio-LogicScienceInstrumentsispresentinmostcountriesthroughanextensivedistributionnetworkHeadquartersBio-Logic SAS1, rue de l’Europe - 38 640 Claix FrancePhone: +33 476 98 68 31 Fax: +33 476 98 69 09
Affiliate officesBio-Logic USA, LLCP.O.Box 30009 - 37 930 Knoxville USAPhone: +1 865 769 3800Fax: +1 865 769 3801
Bio-Logic Science Instruments Pvt Ltd304,Orion Business Park, Next to Cine Wonder, G. B. Road, Thane(W), 400 607 Mumbai IndiaPhone: +91 2225842128
www.bio-logic.info
Broc
hure
rele
ased
on
july
201
5