electrocardiogram

23
ELECTROCARDIOGRAM DEFENITION Electrocardiogram is a graphic recording of the time invariant voltage produced by the myocardium during the cardiac cycle. P, QRS and T waves – de-polarization & Re polarization Clinical use – Diagnosis of diseases & conditions associated with heart.

Upload: drtarek-selim

Post on 03-Jun-2015

423 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Electrocardiogram

ELECTROCARDIOGRAM

DEFENITION

Electrocardiogram is a graphic recording of the time invariant voltage produced by the myocardium during the cardiac cycle.

P, QRS and T waves – de-polarization

& Re polarization

Clinical use – Diagnosis of diseases & conditions associated with heart.

Page 2: Electrocardiogram

ECG

1. Atrial depolarization

3. Ventricular

repolarization

Page 3: Electrocardiogram

Electrocardiography (ECG)

1. Atrial depolarization

2. Ventriculardepolarization

3. Ventricular repolarization

Page 4: Electrocardiogram

Sum of all cardiac action potentials Amplitude: 1-5 mV Bandwidth: 0.05-100 Hz

Page 5: Electrocardiogram

Cardiologists looks critically At various time intervals Polarities and amplitudes

Page 6: Electrocardiogram

Normal values of ECG parameters

Amplitude

P 0.25 mv

R 1.6 mv

Q 25 % of R wave

T 0.1 to 0.5 mv

Page 7: Electrocardiogram

Duration P-R interval 0.12 to 0.20 sec Q- T interval 0.35 to 0.44 sec SW- T segment 0.05 to 0.15 sec P wave interval 0.11 sec QRS interval 0.09 sec

Page 8: Electrocardiogram

Diagnosis of cardiac diseases

First the the heart rate – 60- 100 beats/ min ( Normal value)

Slower than this ( bradycardia) Higher than this ( Tachycardia ) Check whether the cycle is evenly spaced – If not

Arrhythmia

Page 9: Electrocardiogram

If PR interval > 0.2 sec ---BLOCKAGE OF AV node

If one or more of the basic features of ECG are missing -----indicate HEART BLOCK

Page 10: Electrocardiogram

Device – Electrocardiograph Advantage

Can provide several diagnostic information

Disadvantage

Can not provide information about the disorders involving heart valves.

Angiography and Echocardiography

Page 11: Electrocardiogram

Electrodes and leads

Lead --- The particular electrodes selected and the way in which they are connected to the amplifier are called lead

For the individual lead wire, as well as the physical connection to the body of the patient, the term electrode will be used.

Page 12: Electrocardiogram

3- Bipolar Limb lead --- Einthoven

Page 13: Electrocardiogram

Einthoven Postulated 1. At any instant of the cardiac cycle, the frontal

plane representation of the electrical axis of the heart is a two dimensional vector.

2.Einthoven also made the assumption that the heart is near the centre of an equilateral triangle, the apexes of which are the right and left shoulders and the crotch.

Page 14: Electrocardiogram

3. By assuming that the ECG potentials at the at the shoulders are same as the wrists and the potential at the crotch differ little from those at either ankle, he let the points of this triangle represents the electrode positions for the 3- limb leads.

THIS TRIANGLE IS KNOWN AS EINTHOVEN TRIANGLE

Page 15: Electrocardiogram

EINTHOVEN TRIANGLE

Page 16: Electrocardiogram

The sides of the triangle represents the lines along which the three projections of the ECG vector are measured.

Based on this Einthoven showed that , the instantaneous voltage measured from any one of the three limb lead positions is approximately equal to the algebraic sum of the other two.

THE VECTOR SUM OF THE PROJECTIONS ON ALL THREE LINES IS EQUAL TO ZERO.

Page 17: Electrocardiogram

12-Lead ECG MeasurementMost widely used ECG measurement setup in clinical environment

Signal is measured non-invasively with 9 electrodesLots of measurement data and international reference databases

Well-known measurement and diagnosis practicesThis particular method was adopted due to historical reasons, now it is already rather obsolete

Page 18: Electrocardiogram

12 Lead ECG Measurement

Einthoven leads: I, II & III Goldberger augmented leads: VR, VL & VF

Precordial leads: V1-V6

Page 19: Electrocardiogram

Voltage generated by the pumping action of the heart

Is a vector – Magnitude as well as spatial orientation changes with time.

ECG signal is measured from electrodes applied to the surface of the body.

Waveform of ECG signal is depended on the placement of electrode.

Page 20: Electrocardiogram

ELECTRODES

Einthoven – Record ECG from electrodes placed vertically as well as horizontally to the body.

Experiments with immersion electrodes---

Placed electrodes not only on the arm but also on one leg.– Left Leg.

Electronic amplifiers ----Ground reference. ( RL )

Page 21: Electrocardiogram

ECG Recording set up

Page 22: Electrocardiogram

SPECIFICATIONS OF ECG

Frequency response – Upper 100Hz, Lower 50 Hz Sensitivity – 20 mm/mV Input impedance 5 Mega ohms Output impedance Less than 100 ohms Standardization signal 1 mv CMRR 10,000: 1 Recording – heated stylus and heat sensitive paper

Page 23: Electrocardiogram

Paper speed 25 mm/sec or 50 mm/sec Frequency response 0.1 to 60 Hz.