electric field probe: e-field, 100 khz ÷ 6.5 ghz probe 06e electric field probe: e-field, 100 khz...

2
1 PROBE 06E Electric Field Probe: E-field, 100 kHz ÷ 6.5 GHz Information subject to change without prior notice Piazza delle Azalee, 13/14 05018 – Orvieto (TR) - Italy Tel. +39 0763 393291 /Fax. +39 0763 394423 [email protected] - www.microrad.it Main features: Frequency range: 100 KHz ÷ 6.5 GHz Dynamic range: >65 dB Max measured level 650 V/m (CW) Typical applications: Industrial ovens, welding systems, RF heating, tempering and drying equipment. Diathermy equipment and medical devices RF generating, NMR machines. Power plants and related systems maintenance. Sensitive sites (Hospitals). Measurements for railways and ground transport systems. Wireless telecommunication systems such as mobile phone base stations, satellite communication equipment, Broadcasting equipment, Wi-Fi, Wi-Max and LTE systems.

Upload: lykiet

Post on 27-Mar-2018

235 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Electric Field Probe: E-field, 100 kHz ÷ 6.5 GHz PROBE 06E Electric Field Probe: E-field, 100 kHz ÷ 6.5 GHz! Information subject to change without prior notice Piazza delle Azalee,

!

!1!

PROBE 06E Electric Field Probe: E-field, 100 kHz ÷ 6.5 GHz!

Information subject to change without prior notice

Piazza delle Azalee, 13/14 05018 – Orvieto (TR) - Italy Tel. +39 0763 393291 /Fax. +39 0763 394423 [email protected] - www.microrad.it

!!

Main features: • Frequency range: 100 KHz ÷ 6.5 GHz

• Dynamic range: >65 dB

• Max measured level 650 V/m (CW)

Typical applications:

• Industrial ovens, welding systems, RF heating, tempering and drying equipment.

• Diathermy equipment and medical devices RF generating, NMR machines.

• Power plants and related systems maintenance.

• Sensitive sites (Hospitals). • Measurements for railways and ground

transport systems. • Wireless telecommunication systems such as

mobile phone base stations, satellite communication equipment, Broadcasting equipment, Wi-Fi, Wi-Max and LTE systems.

Page 2: Electric Field Probe: E-field, 100 kHz ÷ 6.5 GHz PROBE 06E Electric Field Probe: E-field, 100 kHz ÷ 6.5 GHz! Information subject to change without prior notice Piazza delle Azalee,

!!!

!2!

Piazza delle Azalee, 13/14 05018 – Orvieto (TR) - Italy Tel. +39 0763 393291 /Fax. +39 0763 394423 [email protected] - www.microrad.it

Information subject to change without prior notice

PROBE 06E Electric Field Probe: E-field 100 kHz ÷ 6.5 GHz!

!

!TECHNICAL SPECIFICATIONS Frequency range 100 kHz ÷ 6.5 GHz Type of frequency response Flat

Measurement range 0.35 ÷ 650 V/m (CW) Overload protection up to 1kV/m (CW)

Dynamic range > 65 dB Sensor type Diode dipoles Directivity Isotropic Pulse response speed < 500 uS

Flatness frequency response +/- 1 dB (100 KHz ÷ 1 GHz) +/- 2.5 dB (1 GHz ÷ 6.5 GHz)

Linearity +/- 0.5 dB (2 ÷ 350 V/m) Isotropic response ( @100 MHz ) +/- 0.5 dB

!

DESCRIPTION The probe 06E can measure CW , OFDM and Radar signals, with fast reaction time at the pulse down to 500 uS becoming the best solution for measurements on digital and pulsed modulations. The probe is able to analyze electric fields in the frequency range between 100kHz and 6,5GHz starting from 0,35V/m up to 650 V/m. Main applications are mobile telecommunications, broadcasting, Medical Scientific (ISM) and Radar.

The high sensitivity of this probe makes it ideal for the measurement of human exposure to electric fields in both public and professional environments. The probe includes a calibration certificate according with IEEE std.1309:2005.

GENERAL SPECIFICATIONS

Calibration frequencies

0.1-0.5-1-5-10-27.12-50-100-200-300-400-500-600-700-800-900-1000-2000-2500-3500-4000-4500-5000-5500-6000-6500 (MHz)

Calibration interval (suggested) 24 months Operating temperature 0°C ÷ 50°C Relative humidity From 5 to 95% @ 25 °C no wet Dimensions 327 x 60 (mm) Weight 120g Country of origin Italy !