Elcometer 456
2
Coating Thickness Gauge
Top Models
Operating Instructions
-
7/27/2019 Elcometer 456 User Manual.pdf
2/76
E
nglish
These instructions apply to the followingElcometer 4562 models:
Ferrous (F)
Non-Ferrous (NF)
Dual Ferrous/Non-Ferrous (FNF)
Equipment described in these instructions iscovered by the following Patents:
FNF UK Patent No: GB2306009B
FNF US Patent No: 5886522F1 2 UK Patent No: 2367135B
F1 2 US Patent No. US 6,762,603
F1 2 German Patent Pending
This product meets the emc directive 89/336/EEC,amended 92/31/EEC and 93/68/EEC
and are registeredtrademarks of Elcometer Instruments Ltd.
All other trademarks acknowledged.
Copyright Elcometer Instruments Ltd. 2004-2005.
All rights reserved. No part of this Document may bereproduced, transmitted, transcribed, stored (in aretrieval system or otherwise) or translated into anylanguage, in any form or by any means (electronic,
mechanical, magnetic, optical, manual or otherwise)without the prior written permission ofElcometer Instruments Ltd.
Doc.No. TMA-0304 Issue 02Text with Cover No: 18614
-
7/27/2019 Elcometer 456 User Manual.pdf
3/76
1
CONTENTS
Section Page
1 About your gauge................................................ 31.1 Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41.2 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41.3 What this box contains . . . . . . . . . . . . . . . . . . . . . 51.4 Conventions in these instructions . . . . . . . . . . . . . 5
1.5 Quick-start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52 Getting started ..................................................... 62.1 Fitting the batteries . . . . . . . . . . . . . . . . . . . . . . . . 62.2 Battery condition. . . . . . . . . . . . . . . . . . . . . . . . . . 62.3 Fitting probes (separate versions only) . . . . . . . . . 72.4 The controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.5 Switching the gauge on . . . . . . . . . . . . . . . . . . . . 92.6 Switching the gauge off . . . . . . . . . . . . . . . . . . . . 92.7 The screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102.8 Selecting a language . . . . . . . . . . . . . . . . . . . . . 112.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3 Taking a reading ................................................ 13
3.1 Before you start . . . . . . . . . . . . . . . . . . . . . . . . . 133.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4 The reading screen and menus........................ 144.1 Reading screen . . . . . . . . . . . . . . . . . . . . . . . . . 144.2 Main MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . . 154.3 Main MENU - Extended menu off . . . . . . . . . . . . 16
4.4 Main MENU - Extended menu on . . . . . . . . . . . . 185 Calibration adjustment...................................... 255.1 Calibration method . . . . . . . . . . . . . . . . . . . . . . . 255.2 Preset calibration methods . . . . . . . . . . . . . . . . . 265.3 Calibration foils and standards . . . . . . . . . . . . . . 275.4 Calibration adjustment procedure . . . . . . . . . . . . 28
6 Statistics............................................................. 346.1 Enlarge stats . . . . . . . . . . . . . . . . . . . . . . . . . . . . 346.2 Stats on LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . 356.3 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356.4 Clear stats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 356.5 Select stats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
-
7/27/2019 Elcometer 456 User Manual.pdf
4/76
2
7 Batching ............................................................. 377.1 Exit batching . . . . . . . . . . . . . . . . . . . . . . . . . . . . 387.2 Open new batch . . . . . . . . . . . . . . . . . . . . . . . . . 387.3 Open existing batch . . . . . . . . . . . . . . . . . . . . . . 427.4 Review batches . . . . . . . . . . . . . . . . . . . . . . . . . 427.5 Set limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 437.6 Free memory . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
8 Transferring readings to a computer............... 45
9 Probes ................................................................ 47
9.1 Ferrous (F) probes . . . . . . . . . . . . . . . . . . . . . . . 479.2 Non-ferrous (N) probes . . . . . . . . . . . . . . . . . . . . 479.3 Dual ferrous/non-ferrous (FNF) probes . . . . . . . 479.4 Probe interchangeability . . . . . . . . . . . . . . . . . . . 489.5 F1 2 probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 499.6 F1 2 High temperature PINIP probes . . . . . . . 499.7 Coatings on galvanised or metallised steel . . . . 49
10 Personalised welcome screen ......................... 50
11 Storage and transit ............................................ 51
12 Maintenance....................................................... 51
13 Statistics terminology ....................................... 52
14 Technical data.................................................... 53
15 Accessories ....................................................... 55
16 Related equipment ............................................ 58
17 Fitting the wrist harness ................................... 58
18 Probe measurement performance ................... 59
19 Probe capabilities.............................................. 60
20 Error messages ................................................. 67
21 Index ................................................................... 69
-
7/27/2019 Elcometer 456 User Manual.pdf
5/76
3
hank you for your purchase of this Elcometer 4562
Coating Thickness Gauge. Welcome to Elcometer.
Elcometer are world leaders in the design, manufacture
and supply of coatings inspection equipment. Ourproducts cover all aspects of coating inspection, fromdevelopment through application to post applicationinspection.
The Elcometer 4562 Coating Thickness Gauge is a worldbeating product. With the purchase of this gauge younow have access to the worldwide service and supportnetwork of Elcometer. For more information visit ourwebsite at www.elcometer.com.
1 ABOUT YOUR GAUGE
The Elcometer 4562 Coating Thickness Gauge is ahandheld gauge for fast and accurate measurement ofthe thickness of coatings on metal substrates.
The gauge is available inthree versions; Basic,Standard and Top. Thismanual describes theoperation of the
Elcometer 4562 Top.
All versions of the gaugefeature an easy-to-usemenu driven graphicalinterface which guides theuser through tasks such as
gauge configuration andcalibration adjustment.
The gauge is available either with a built-in integralprobe or as a separate probe version. A wide range of
probes is available to suit requirements - see page 47.Separate probes may be standard, miniature or Plug inIntegral Probes (PINIP), and must be orderedseparately.
T
R
Figure 1. Elcometer 4562Coating Thickness Gauge
http://www.elcometer.com/http://www.elcometer.com/ -
7/27/2019 Elcometer 456 User Manual.pdf
6/76
4
1.1 Features
A range of smooth and rough surface calibrationadjustments.
Menu driven backlit graphical user interface. Interchangeable probes (separate versions only).
Statistics.
Infrared interface.
RS232 interface.
High/low limits. Memory of up to 50 000 readings in up to 999
batches.
Clock/alarm.
Date and time stamping.
1.2 Standards
The Elcometer 4562 can be used in accordance with thefollowing National and International Standards:
Ferrous (F) Non-Ferrous (NF)
BS 5411 (11) BS 5411 (3)
BS 3900 (C5) BS 3900 (C5)
ISO 2178 BS 5599
ISO 2808 ISO 2360
BS EN ISO 1461 ISO 2808
DIN 50981 DIN 50984
ASTM B 499 ASTM D 1400
ASTM D1186 ASTM B 244
ISO 19840
SSPC-PA2 (2004)
-
7/27/2019 Elcometer 456 User Manual.pdf
7/76
5
1.3 What this box contains
Elcometer 4562 Gauge with integral probe, or
Elcometer 4562 Gauge and separate probe
(probe must be ordered separately) Calibration foils
Gauge carrying pouch
Wrist harness
Batteries
CD containing data collection software PC connection cable
Operating instructions
1.4 Conventions in these instructions
The Elcometer 4562 is controlled using a simple menu
structure which helps you get the most from your gauge- see page 18.
As an example, the LANGUAGES option which is inSETUP from the MAIN MENU would be shown in theseinstructions as MENU/SETUP/LANGUAGES.
These instructions include images of Elcometer 4562
screens with units set to microns (m). Similar screenswill be seen when the gauge is set to other units such asmils or inches.
1.5 Quick-start
To quickly configure the gauge and start taking readings:
1. Fit batteries: see page 62. Fit probe (separate versions only): see page 7
3. Switch on: see page 9
4. Select language: see page 11
5. Try taking a reading: see page 13
6. Adjust calibration: see page 25
The gauge is now configured and ready to use.
To maximise the benefits of your newElcometer 4562, please take some time to read theseOperating Instructions. Do not hesitate to contactElcometer or your Elcometer supplier if you have anyquestions.
-
7/27/2019 Elcometer 456 User Manual.pdf
8/76
6
2 GETTING STARTED
2.1 Fitting the batteries
1. Open battery compartment cover; press down indirection of arrow using thumb nail.
2. Insert 2 x LR03 (AAA), alkaline dry batteries takingcare to ensure correct battery polarity (Figure 2).
3. Close battery compartment cover.
Rechargeable batteries can be used but they will only
have 25% to 30% of the life of alkaline batteries.
Figure 2. Fitting batteries -ensure correct battery polarity
2.2 Battery condition.
Symbol Battery condition/action required
100%
66% to 100%
33% to 66%, replacement recommended.
16% to 33%, replacement required.
-
7/27/2019 Elcometer 456 User Manual.pdf
9/76
7
2.3 Fitting probes (separate versions only)
To ensure correct transfer of data from theprobe and detection of the new probe, the
gauge must be switched off whenseparate probes are fitted.
A probe must be calibrated once it has been fitted -see Calibration adjustment on page 25.
2.3.1 To fit the probe
Align connector keyway and pushin direction shown. The connectorlocks automatically.
Note: The design of the probeconnector allows some movementbetween the probe and the gauge.
This is intentional and does notaffect measurement performance.
2.3.2 To release the probe
Grasp knurled section and pull
gently away from the gauge. Theconnection will unlock and theprobe will release.
2.3.3 To fit the PINIP
Twist the PINIP until theconnector locates. Taking
care not to cross thethreads, lock the probe byturning the locking ring 1times clockwise, or untiltight.
-
7/27/2019 Elcometer 456 User Manual.pdf
10/76
8
2.4 The controls
The gauge is operated by 5 keys (Figure 3).
On/Off key : Switches the gauge on or off. Softkeys: The function of these keys varies and is
described by symbols and writing on the bottom lineof the screen.
LED: Red/green flashes when the gauge is switchedon, green flashes when a reading is taken. Also
indicates when a reading is inside or outside limits -see Set limits on page 43.
Figure 3. Elcometer 4562 control keys
R
Softkeys
On/Off key
LED
-
7/27/2019 Elcometer 456 User Manual.pdf
11/76
9
2.5 Switching the gauge on
Note: Before switching the gauge on for the first timeread Selecting a language on page 11.
2.6 Switching the gauge off
To switch off all gaugetypes, press and hold
key for 3 seconds.
The gauge will beep,two single tonesfollowed by a doubletone.
The Elcometer 4562 switches itself off 60 seconds afterthe last operation unless the Auto Switch Off time is
changed (MENU/SETUP/AUTO SWITCH OFF). TheAuto Switch Off feature can be set to a maximum of 10minutes or can be deactivated - see AUTO SWITCHOFF: on page 24.
Gauges with separateand PINIP probes:
Press key to switchon gauge.
Gauges with integralprobes:
Press key to switchon gauge, or place theprobe on a surface.
R
R
C AL DATA STA TS MENU
09 : 30 09 / 1 0 / 2 000F1
m
F456
R
3 seconds
-
7/27/2019 Elcometer 456 User Manual.pdf
12/76
10
2.7 The screen
Familiarise yourself with the Elcometer 4562 screen. Thescreen displays:
Welcome information Measurement information
Menus to configure the gauge and control functions
Help and miscellaneous information
When the gauge is switched on a welcome information
screen may be displayed briefly (Figure 4).
Figure 4. Typical Elcometer 4562 welcome screen
The screen where readings are displayed is called theReading Screen (Figure 5). The reading screen can alsoinclude statistics if this option is selected.
Figure 5. Typical Elcometer 4562 reading screen
http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
13/76
11
2.8 Selecting a language
The Elcometer 4562 has over 20 built-in languages.When the gauge is switched on for the first time after
dispatch from the Elcometer factory the display will showthe language selection screen (Figure 6).
Figure 6. Language selection screen
2.8.1 At first switch on
1. Press Up/Down softkeys to locate languagerequired.
2. Press SEL softkey to activate the selectedlanguage.
The screen displays an Elcometer 4562 welcomescreen (Figure 4) followed by the reading screen(Figure 5).
The gauge operates in the new language until
changed.
2.8.2 At any time
1. Switch gauge off.
2. Press and hold left hand softkey.
3. Press key to switch on gauge.
The display will show language selection screenwith current language highlighted by cursor.
4. Release left hand softkey.
Follow the instructions given in 2.8.1.
Alternatively, select MENU/SETUP/LANGUAGES - see
Main MENU - Extended menu on on page 18.
http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
14/76
12
2.9 Interfaces
The gauge is fitted with an infrared interface (Figure 7)which allows information to be sent to a suitable printer.
Figure 7. Infrared and RS232 interfaces
The RS232 5-pin connector next to the infrared interfaceis used with the PC connection cable to connect thegauge to a PC. This enables personalised screens to be
downloaded to the gauge (see Personalised welcomescreen on page 50) and measurement data to betransferred to a PC or printer.
For information on the range of Elcometer data collectionsoftware see Transferring readings to a computer onpage 45.
Infrared interfaceRS232 5-pinconnector
http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
15/76
13
3 TAKING A READING
3.1 Before you start
Are you using the correct type of probe?See Probes on page 47.
Is the probe calibrated?See Calibration adjustment on page 25.
Do you require statistics?See Statistics on page 34.
Do you want to save readings in memory?See Batching on page 37.
What units of measurement do you want to use?See UNITS: on page 23.
3.2 Procedure
1. Press key to switch on gauge.
2. Place probe on surface to be measured. Thereading may be inaccurate if the probe is notheld as shown in Figure 8.
Figure 8. Taking a reading
3. Reading is displayed on screen (Figure 9).
Figure 9. Typical reading
R
R
CAL D AT A STA TS MEN U
09 : 30 0 9 / 1 0 / 20 00
F1
mF456
Separate probe Integral probe
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
16/76
14
4 THE READING SCREEN AND MENUS
4.1 Reading screen
The content of the reading screen (Figure 10, Figure 11)depends upon the type of measurement being made andhow the gauge is set up.
Figure 10. Example of reading screen with smoothsurface calibration method selected
CAL. Operates selected calibration adjustment method.MENU. This opens the main MENU of the gauge andprovides access to user-selectable features - seepage 18.
Note: IfCAL softkey symbol is flashing the gauge shouldbe recalibrated. This is due to the calibration adjustment
method having been changed or a probe change - seeCalibration adjustment on page 25. Batches cannot becreated while the CAL softkey symbol is flashing.
Figure 11. Reading screen in extended mode andshowing full set of statistical values.
Automatic-switchingprobe symbol(FNF probes only)
Batterysymbol
Substrate
Units
Softkeyfunction/symbol
CalMethodsymbol
-
7/27/2019 Elcometer 456 User Manual.pdf
17/76
15
4.2 Main MENU
Gauge configuration and measurement functions arecontrolled using menus (Figure 12). The structure of the
menus is shown on page 18.
Figure 12. Typical Elcometer 4562 menu
Some screens allow the status of a feature to bechanged e.g. on to off or select or deselect, etc. A tickbox indicates this type of feature. A tick against amenu item indicates the function is activated or selected.
SEL softkey selects the option displayed and in somecases toggles the status of a tick box off/on.
Up/Down softkeys KL move the cursor to the menu itemrequired. The menus scroll up/down and a line acrossthe screen indicates the start and end of the menu.
BACK softkey returns the gauge to a previous screen.
Holding this softkey down will rapidly exit from any menuand return to the reading screen.
4.2.1 Simple and extended menus
The Elcometer 4562 Top Gauge has two menustructures:
Extended menu off (simple menu mode): The gaugeis shipped from the Elcometer factory with EXTENDEDMENU turned off. In this simple menu mode the gaugecan be calibrated and used to take measurements. Thisis the ideal setting for users who do not require accessto advanced features of the gauge.
Extended menu on (extended menu mode):Additionalitems are automatically added to the MENU and theSTATS softkey and DATA softkey are activated. Thesegive access to more advanced functions such asstatistics, batching, calibration method, print/output,setup, etc.
Menu title
Menu contents
Softkey functions
-
7/27/2019 Elcometer 456 User Manual.pdf
18/76
16
4.3 Main MENU - Extended menu off
Figure 13. Main menu - extended menu off
4.3.1 BACKLIGHT
Switches backlight on and off. Toggle tick box toactivate/deactivate. With BACKLIGHT activated thedisplay is illuminated for approximately 5 seconds whena reading is taken or a key pressed.
Note: The battery life is reduced by about one thirdwhen the backlight is activated.
4.3.2 CALIBATION LOCKED
Protects against inadvertent calibration adjustment.Toggle tick box to activate/deactivate. IfCAL softkey ispressed while CALIBRATION LOCKED is activated thegauge displays CALIBRATION LOCKED USE MENUTO UNLOCK. The message disappears after 3 seconds.
4.3.3 Extended MENU
Provides access to additional features. Toggle tick boxto activate/deactivate. See Main MENU - Extendedmenu on on page 18.
4.3.4 ABOUTProvides information on Gauge, Probe, Contactinformation and Help (Figure 14):
GAUGE INFORMATION: Elcometer 4562 model,software versions, etc.
PROBE INFORMATION: Probe type, range, etc.
CONTACT: Details of Elcometer offices worldwide and,if programmed, the contact details for the Supplier orLocal Distributor.
http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
19/76
17
HELP: Explains symbols used on Elcometer 4562
display screens.
Figure 14. About menu
4.3.5 RESETSelects Factory Calibration or Gauge resets. TheRESET menu option (Figure 15) allows one of threegauge resets to be selected:
FACTORY CAL: Returns gauge to calibrationsettings created at time of manufacture of the probe.
Factory calibration will not necessarily restore precisecalibration values. The calibration of the gauge shouldbe adjusted before use, or at least checked to ensurethat it has been previously adjusted correctly for theconditions of use.
INTL GAUGEa: Resets gauge to International default
settings e.g. DD/MM/YY date format and metric units.
US GAUGEb: Resets gauge to USA default settingse.g. MM/DD/YY date format and imperial units.
Figure 15. Reset menuA confirmation screen will be displayed. Press YES toreset, NO to cancel.
a. International settings can also be activated atswitch on. Press and hold softkey 3 and switchon gauge.
b. US settings can also be activated at switch on.Press and hold softkey 4 and switch on gauge.
http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
20/76
18
4.4 Main MENU - Extended menu on
To toggle EXTENDED MENU on/off selectMENU/EXTENDED MENU/SEL
MENUBacklightCalibration locked
z
z
z
z
zz
Extended menuPrint/OutputDeleteCal methodSetup
AboutReset
CAL METHOD
Smooth2 Point
Rough
Special subZero offsetSet offsetISOSSPC 2004SwedishAustralian
PRINT/OUTPUT
Single batchAll batchesCurrent statistics
DELETE
Last readingSingle batchAll batches
SETUP
StatisticsData
z
z
z
z
ProbeUnitsClock/AlarmOutputBeep volume 3
LanguagesAuto switch off 1Opening screen
ABOUT
Gauge informationProbe informationContactHelp
RESET
Factory calIntl gaugeUS gauge
STATS
Enlarge statsStats on LCDDisplay F1
Clear statsSelect stats
DATA
Exit batchingOpen new batch
Open existing batchReview batchesSet limitsFree memory
CAL
Probe calibration
-
7/27/2019 Elcometer 456 User Manual.pdf
21/76
19
OUTPUT
Baud rate
RS232 bit imageRS232 plain textInfra red printer
z
CLOCK/ALARM
Set clock/dateSet alarmSet date format
z
z
z
DATA
Softkey enabledShow date stampDeleted reading
z
STATISTICS
Softkey enabledSelect statsDisplay F1
z
SELECT STATS
No. of readingsMean
Std deviationCoef of variatnHighest readingLowest reading
DELETED READING
TagDelete
SET CLOCK/DATE
Time/dateDisplay when offDisplay when on
SET ALARM
Alarm onSet time 00:00
DATE FORMAT
DD/MM/YYYYMM/DD/YYYY
BAUD RATE
12002400
4800
96001440019200
-
7/27/2019 Elcometer 456 User Manual.pdf
22/76
20
The following features are added to the MENU whenEXTENDED MENU is active:
4.4.1 PRINT/OUTPUT
Outputs data to a printer or to a PC. A single batch ofreadings, all batches of readings, or the currentstatistical summary can be output via the infraredinterface or the RS232 interface.
To use this function first setup using:
MENU/SETUP/OUTPUT - see OUTPUT: on page 23.If no printers have been setup, PRINT/OUTPUT willdisplay a NOT AVAILABLE message.
4.4.2 DELETE
Deletes last reading only, a single batch of readings orall batches of readings (Figure 16).
Figure 16. DELETE screen
LAST READING
Deletes last reading either in immediate mode or in batchmode.
Gauge displays ARE YOU SURE? (Figure 17).
Figure 17. Delete last reading confirmation screen
Press NO softkey to include reading in statisticalsummary orYES softkey to delete reading.
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
23/76
21
If there is no reading to delete gauge displays LASTREADING NOT AVAILABLE (Figure 18).
Figure 18. Delete last reading not available screen
SINGLE BATCH
Deletes the batch data.
The gauge displays the current batch or the last batchused (Figure 19).
Figure 19. SINGLE BATCH screen
Use the Right/Left softkeys to locate the batch tobe deleted. At either end of the list of batches the gaugebeeps. Press OK when the batch to be deleted isdisplayed. Gauge displays ARE YOU SURE? Press NOsoftkey to leave data unchanged orYES softkey todelete the batch. If there is no data stored in the memorythe gauge displays NOT AVAILABLE DATA MEMORYEMPTY (Figure 20).
Figure 20. Memory empty screen
ALL BATCHESDeletes all batches.
Gauge displays ARE YOU SURE? Press NO softkey toleave data unchanged orYES softkey to delete all thebatches.
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
24/76
22
4.4.3 CAL METHOD
Allows selection of calibration method from list of options- see Calibration method on page 25 for more details.
4.4.4 SETUP
Used to select, change or activate gauge features:
STATISTICS:Activates statistics feature.
Stats softkey on/off.
Select stats - see Select stats on page 36.
Display - see Display on page 35.
DATA:Activates DATA softkey, SHOW DATE STAMPoption and selects the DELETED READING menu(Figure 21).
Figure 21. DATA menu screen
SOFTKEY ENABLED allows DATA softkey to be turnedoff when EXTENDED MENU is active.
SHOW DATE STAMP - toggles date stamp on or off inReview Batches - see Review batches on page 42.
DELETED READING screen (Figure 22) allows readingsto be tagged or deleted.
Figure 22. DELETED READING screen
TAG - readings still appear in the batch (with atag-deleted symbol ), but they are not included instatistical calculations.
DELETE - readings are permanently deleted.
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
25/76
23
PROBE: Only for dual function probes (FNF and F1 2).Changes probe mode.
FNF probes - select from Automatic, F or N.
F1 2 probes - select from F1 or F2.UNITS: Units are automatically set by the probe type,however the user can manually override the automaticsetting. Select from m, mm, mil, thou or inch.
CLOCK/ALARM:Allows setting of time, date and alarmfunctions and setting of the date format DD/MM/YY or
MM/DD/YY (Figure 23).
Figure 23. CLOCK/ALARM screen
Time and date can be set to be displayed on the top lineof the reading screen when the gauge is on.
Time and date can be set to be displayed in large digitswhen the gauge is switched off.
OUTPUT: Selects baud rate (Figure 24) and activatesdata output via the interfaces - see Interfaces onpage 12.
Figure 24. OUTPUT screen
BAUD RATE can be set at values from 1200 to 19200.The default value is 9600 baud.
RS232 BIT IMAGE. Toggle tick box toactivate/deactivate. When activated, readings are sentto the RS232 interface as they are taken. All images andcharacters are output as bit-maps. This allows printingon the Elcometer Miniprinter (see Miniprinter onpage 56).
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
26/76
24
RS232 PLAIN TEXT. Toggle tick box toactivate/deactivate. When activated, readings are sentto the RS232 interface as they are taken. The gaugesends standard ASCII characters from the Courier Newfont setc. This allows printing on devices other than theElcometer Miniprinter, e.g. RS 232 printers or PC viaElcometer software (page 45) or via HyperTerminal.
INFRA RED PRINTER. Toggle tick box toactivate/deactivate. When activated, readings are sentto the infrared portable printer as they are taken. See
Infrared portable printer on page 56 for sales partnumber of this optional accessory.
BEEP VOLUME: Changes volume.
0 = off
5 = loudest
Default = 3LANGUAGES:Allows selection of language.
AUTO SWITCH OFF: Changes delay before gaugeswitches off when displaying Reading Screen.
Minimum = 1 minute
Maximum = 10 minutes Default = 1 minute
Auto switch off may be disabled by selecting off (In thiscase, switch off using On/Off key .)
OPENING SCREEN: Disables the opening (welcome)screens so that the gauge switches on to display the
reading screen.
If a personalised welcome screen has been downloadedinto the gauge (see page 50), Opening Screen must beactivated to display this screen.
c. When RS232 Plain Text is selected the followinglanguages will be output as English: Chinese,Greek, Hebrew, Japanese, Korean, Russian,Lithuanian, Farsi.
-
7/27/2019 Elcometer 456 User Manual.pdf
27/76
25
5 CALIBRATION ADJUSTMENT
Calibration adjustment is the process of setting thegauge to known values of thickness to ensure accuracy
on different substrate types, shapes and surfacefinishes.
Note: When using an FNF probe it must be calibrated inboth the ferrous mode and in the non-ferrous mode toensure accuracy of reading.
5.1 Calibration methodThe calibration of the gauge can be adjusted(MENU/CAL METHOD) using several different methodsdescribed in National and International Standards.
The calibration adjustment method chosen is dependanton the condition of the substrate to be measured and is
indicated on the screen by a symbol (Figure 25):
Figure 25. CAL METHOD screen
SMOOTH. Smooth surface calibration where the gaugeis set to zero on the uncoated surface and a knownthickness above the expected thickness of the coating.
2POINT. Calibration on a thin value and a thick value
either side of the expected thickness. This enhances theaccuracy of the gauge over the thickness range definedby the two values.
ROUGH.A calibration method similar to 2-Point. Thisenhances the accuracy of the gauge over the thicknessrange defined by the two values.
SPECIAL SUB. This method uses the 2-Pointcalibration for unusual substrate materials such as castiron, certain types of stainless steel, high carbon steel,special aluminium alloys, etc.
ZERO OFFSET. This is the method described inISO 19840 for coatings on steel surfaces roughened by
-
7/27/2019 Elcometer 456 User Manual.pdf
28/76
26
blast cleaning. The calibration uses the smooth surfacetechnique, and a correction value (zero offset) is appliedto each reading to account for the effect of theroughened surface; the value depends on the surfaceprofile - see Table 1.
SET OFFSET. This screen sets and changes the offsetfor different surface roughness This value is used onlywith the Zero Offset calibration method.
5.2 Preset calibration methods
The gauge also contains four preset calibration methodswhich follow relevant standards. These set the
calibration method and the data collection method (datacollection method is only set when in batching mode).
ISO: (ISO 19840) This uses the Zero Offset calibrationmethod with counted average data collection set to 5readings.
SSPC: (SSPC-PA2 2004) This uses the 2-Pointcalibration method with counted average set to 3readings.
SWEDISH: (SS 18 41 60) This uses the 2-Pointcalibration method with counted average set to 5readings.
AUSTRALIAN: (AS 3894) This uses the Zero Offsetcalibration method with a correction value of 1/3 thesurface profile peak-to-valley height and a countedaverage data collection set to 5 readings.
Table 1: Correction values as detailed in ISO 19840
Profile accordingto ISO 8503-1
Correction Value(m) (Zero Offset)
Fine 10
Medium 25
Coarse 40
-
7/27/2019 Elcometer 456 User Manual.pdf
29/76
27
Note: When the calibration method is changed, e.g. fromSmooth to Rough, the gauge will display a message(Figure 26).
Figure 26. Recalibration required screen
If the NO softkey is pressed the CAL softkey symbol onthe Readings Screen will flash to warn that calibrationadjustment is still required. While the CAL softkeysymbol is flashing new batches cannot be opened.
If theYES softkey is pressed the calibration adjustmentprocedure is activated - see Calibration adjustment
procedure on page 28.
5.3 Calibration foils and standards
Calibration adjustment should be carried out with theappropriate probe on the same type of metal, the samecurvature and similar finish to the item to be measured.
It is best to use an uncoated sample of the items to betested.
Calibration can be carried out using measured foils orcoated standards.
Foils (shims). These are coating thickness standardswhich have been measured using techniques
independent of the gauge. They are ideal for calibrationbecause they provide a known value of thickness on theactual substrate to be measured. Calibration certificatesfor foils are available upon request.
When using foils care must be taken to keep the foilsclean and free from dust and to avoid damage by
creasing particularly the thinner foils. Always remove afoil from its storage wallet before use.
When calibrating a High Temperature PINIP Probeuse the special thickness standards supplied with theprobe - see Calibrating High Temperature PINIPProbes on page 33.
-
7/27/2019 Elcometer 456 User Manual.pdf
30/76
28
Note: To calibrate 5 mm (200 mils) and 13 mm(500 mils) range gauges it will be necessary to stack thefoils (Figure 27). Care must be taken to avoid errors dueto placing the foil labels between the foils.
Figure 27. Stacking foils to increase thickness
Coated standards. Thickness standards using typicalsubstrate materials coated with hardwearing materialsand measured using techniques independent of thegauge.
Coated standards are most often used to confirm that thegauge meets its specifications if it is not possible to use
foils (shims).
5.4 Calibration adjustment procedure
Calibration adjustment can be carried out at any time bypressing CAL softkey from the reading screen. Toprevent inadvertent calibration adjustment the CAL
softkey can be locked (MENU/CALIBRATIONLOCKED).
The user is guided through the operation of the chosencalibration procedure by means of instructions andillustrations on the graphics screen. Audible warningsare also provided when action is required, e.g. when the
probe must be placed down to get a reading.If the routine is interrupted in any way the previoussettings will be restored until after the full calibrationroutine has been completed or the reset has beencompleted.
Serial no.Inspected by
micron mils/thou
502 19.75
TK1009
Serialno.
Inspectedby
micron
mils/thou
176.1
6.9
3
TK1010
-
7/27/2019 Elcometer 456 User Manual.pdf
31/76
29
The screen detail depends on the calibration methodchosen, but the calibration is in two steps.
The following example is for a Smooth Calibration
adjustment.
Step 1
1. Hold probe in air and press CAL softkey (Figure 28).
Figure 28. Step 1 - on thickness standard
ESC softkey returns the gauge to the ReadingScreen from the Calibration Procedure without
making any changes.
TEST softkey allows the user to take readings toverify the accuracy of the current calibration. Thesereadings do not affect statistical calculations and arenot added to batch memory.
2. Place probe on calibration standard. The gauge will
display a reading.
3. Lift probe and then replace on calibration standard.Gauge displays the average ( ) of these readingsand the last reading. Repeat this action until astable reading is obtained.
Figure 29. Step 1 - Calibration adjustment onthickness standard
First reading Second reading
Average
Last
-
7/27/2019 Elcometer 456 User Manual.pdf
32/76
30
To reject the displayed reading and start thecalibration procedure again, press both the Up andDown softkeys at the same time.
To adjust the displayed reading until it is correctrelative to the thickness standard use the Up/Downsoftkeys.
4. Press SET softkey to accept the value.
Note: - - - indicates over-range (Figure 30). Taking areading within range clears this screen.
Figure 30. Over-range reading
Step 2
1. Place probe on uncoated standard or zero plate.The gauge will take and display a reading.
2. Lift probe and then replace on uncoated standard orzero plate. Gauge displays the average ( ) of these
readings and the last reading. Repeat this actionuntil a stable reading is obtained.
Figure 31. Step 2 - Calibration adjustment onuncoated sample
To reject the displayed reading and start Step 2 of
the calibration procedure again, press the Resetsoftkey .
First reading Second reading
-
7/27/2019 Elcometer 456 User Manual.pdf
33/76
31
3. Press ZERO softkey to zero the display (Figure 32).
Figure 32. Zero the display
4. Press SET softkey to accept this value.
The gauge will display the option to test thecalibration of the gauge.
Figure 33. TEST READINGS screen
5. Either press NO softkey to complete the calibrationadjustment procedure and return the gauge to thereading screen, or proceed to take test readings -
see Taking test readings on page 31.
Taking test readings
Press YES softkey (see previous section) to take testreadings. This allows the calibration of the gauge to betested without adding readings to data memory or
contributing to the statistical calculations.
Figure 34. TEST READINGS screen
CAL softkey returns the gauge to Step 1 of thecalibration adjustment procedure.
OK softkey completes the calibration adjustmentprocedure and returns the gauge to the reading screen.
-
7/27/2019 Elcometer 456 User Manual.pdf
34/76
32
Other calibration methods
For the 2-POINT, ROUGH and SPECIAL SUBSTRATEcalibration methods Step 2 requires readings to be taken
on a thin standard value instead of on an uncoated base.
Figure 35. Step 2 - On thin standard
Repeating the reading will display the average. This isparticularly useful for rough surfaces as it allowsvariations in the surface to be accounted for in thecalibration adjustment, therefore improving the accuracyof the gauge.
Figure 36. Step 2 - Calibration adjustment on thinstandard
First reading Second reading
-
7/27/2019 Elcometer 456 User Manual.pdf
35/76
33
Calibrating High Temperature PINIP Probes
Special thickness standards are supplied with F1 2 HighTemperature PINIP Probes - see page 49. These
thickness standards should be used in place of thecalibration standard in Step 1 ofCalibration adjustmentprocedure on page 28.
1. Place appropriate thickness standard over end ofPINIP probe.
2. Press CAL softkey.
3. Place probe on hotd surface and take a reading.
4. Lift probe and then replace on hot surface to takesecond reading.
5. Repeat as necessary until reading is stable.
6. Press SET to accept value.
7. Remove thickness standard from end of PINIPprobe.
8. Proceed with Step 2 - see page 30.
d. The temperature of the surface used forcalibration should be equal to the temperature ofthe substrate being measured.
-
7/27/2019 Elcometer 456 User Manual.pdf
36/76
34
6 STATISTICS
The Elcometer 4562 Top has a Statistics feature(STATS) which calculates and displays a statistical
analysis of readings as they are taken. The statisticalcalculations are also applied to the readings storedwithin a single batch in memory.
When EXTENDED MENU is active (MENU/EXTENDEDMENU/SEL), press STATS softkey to access STATSMENU (Figure 37).
Figure 37. Stats menu
The statistical values available are:
Number of Readings
Mean
Standard Deviation
Coefficient of Variation
Highest Reading
Lowest Reading
See Select stats on page 36 and Statisticsterminology on page 52.
6.1 ENLARGE STATSDisplays the chosen statistical values as double-heightcharacters. The example screens (Figure 38) appearwhen all the statistical values are selected. The
-
7/27/2019 Elcometer 456 User Manual.pdf
37/76
35
Up/Down softkeys can be used to move through the list.OK softkey returns to the Reading Screen.
Figure 38. Enlarged statistics
6.2 STATS ON LCDActivates the presentation of the chosen statisticalvalues on the reading screen. If STATS ON LCD isactivated while CLOCK ON LCD is active, the STATS aredisplayed and the CLOCK is not displayed.
Figure 39. Reading screen with all statistics
6.3 DISPLAY
Only applies when using dual function probes. Allowsselection of the types of readings used in the statisticalcalculation when a dual function probe is connected.
Probe OptionsFNF F, N or F and N combinede
F1 2 F1, F2 or F1 and F2 combinede
6.4 CLEAR STATS
Resets to zero all statistical values selected inSTATS MENU/DISPLAY.
e.When readings are combined a symbol will bedisplayed on the Reading Screen (Figure 39).
Combinedstatisticssymbol
-
7/27/2019 Elcometer 456 User Manual.pdf
38/76
36
6.5 SELECT STATS
Allows the user to chose which statistical values aredisplayed. The default condition is all values (Figure 40).
Figure 40. Select stats menu
Use Up/Down softkeys to move cursor and SEL softkeyto select or deselect the statistical values.
-
7/27/2019 Elcometer 456 User Manual.pdf
39/76
37
7 BATCHING
The Elcometer 4562 operates in one of two modes;immediate or batch.
Immediate mode. The gauge takes readings andcalculates statistics but does not store any readings inmemory (Figure 41).
Figure 41. Reading screen - Immediate mode
Batch mode. The gauge takes readings and calculatesstatistics and stores readings in memory (Figure 42).
The gauge stores readings in one of up to 999 separatebatches.
Batch mode (batching) allows reading data to becollected in groups to allow easier analysis of largestructures or complex assemblies.
Figure 42. Reading screen - Batch mode
The Elcometer 456
2
Top has memory capacity for upto 50 000 readings in up to 999 batches.
Batching is configured using the DATA MENU.
To access the DATA MENU (Figure 43) press the DATAsoftkey (this softkey is only displayed with EXTENDEDMENU on).
Figure 43. DATA MENU screen
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
40/76
38
7.1 EXIT BATCHING
This option returns the gauge to immediate mode and nofurther readings are stored in memory. The gauge
returns to the Reading Screen.
7.2 OPEN NEW BATCH
This option opens (creates) a new batch using the nextavailable batch number.
To create a batch with new calibration settings use
OPEN NEW BATCH in immediate mode - see 7.2.1. To create a batch with a copy of the calibration
settings from another batch use OPEN NEWBATCH in batch mode - see 7.2.2. This is known ascloning a batch.
Cloning is typically used when several products are
being tested, each with different measurement andcalibration settings. Cloning allows the settings for aproduct to be quickly and easily duplicated. To clonea batch, first open the batch to be cloned (seesection 7.3) and then OPEN NEW BATCH.
Note: While the CAL softkey symbol is flashing, new
batches cannot be opened. Calibrate the gauge beforeopening a new batch. FNF probes should be calibratedin F mode and in N mode.
7.2.1 Opening new batch in immediate mode
When a new batch is opened the following settings arecopied into the batch from immediate mode:
Calibration method
Offset (if applied)
Calibration adjustment
Note: Calibration method and offset must be set beforeopening the new batch. Calibration adjustment can be
changed after the batch has been created - seeCalibration adjustment procedure on page 28.
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
41/76
39
The initial OPEN NEW BATCH screen (Figure 44) stayson while the gauge sets up the batch, as indicated by theprogress bar.
Figure 44. Initial OPEN NEW BATCH screen -
Immediate mode
ESC takes the gauge back to DATA MENU.
The second OPEN NEW BATCH screen (Figure 45)shows the current batch settings.
Figure 45. Second OPEN NEW BATCH screen -
Immediate mode
Upper and lowerLimits for the batch can be changed atany time once the batch has been created - see Setlimits on page 43.
Data collection method must be set before OK softkeyis pressed.
Data Collection Method
The Data Collection Method can be changed by pressingthe Data Collection Method softkey or . Thedisplay toggles between two options, normal andcounted average:
Lower andupper limits
Number ofreadings inbatch (0)
Batchnumber
Data collection
method
Data collectionmethod softkey
Calibrationmethod
Offset
Limitsselected/
deselectedsymbol
Calibrationtype
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
42/76
40
NORMAL - Each reading is added to the number ofreadings and contributes to the statisticalcalculation.
COUNTED AVE: n - Readings are taken in presetgroupsf, the default being 5 readings, n=5 (Figure46). At the end of each group the gauge beeps andcalculates the average for the group and stores thisaverage value, which is used for the statisticalcalculation. The individual readings in the group arenot stored. The value for n can be changedg by
pressing the n=5h softkey.
Figure 46. Setting data collection methodCounted average n = 5
f. When using an FNF probe in AUTO mode the
first reading in each group will lock the probe tothat substrate for the whole of the group.g. If a preset calibration method (see Preset
calibration methods on page 26) is selected,the value for n cannot be adjusted.
h. The value for n shown on the softkey can beany number between 2 and 255.
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
43/76
41
7.2.2 Opening new batch in batch mode
Opening a new batch in batch mode duplicates (clones)an existing batch.
Cloning copies the following settings to the new batch: Limits
Calibration method
Data collection method
Offset (if applied)
Calibration adjustmentNote: Calibration method, data collection method andoffset cannot be changed once the cloned batch hasbeen created. Calibration adjustment and limits can bechanged after the batch has been created - seeCalibration adjustment procedure on page 28 and Set
limits on page 43.The initial OPEN NEW BATCH screen displaysCURRENT BATCH.
Figure 47. Initial OPEN NEW BATCH screen -Batch mode
The second OPEN NEW BATCH screen (Figure 48)shows the new batch settings.
Figure 48. Second OPEN NEW BATCH screen -Batch mode
Press OK to create batch.
Cloned batchsymbol
http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
44/76
42
Batch numbering
The gauge automatically allocates the next batchnumber when a new batch is created. Up to 999 batches
can be active at any time. Deleted batches are removedfrom the list so that batch numbers above 999 are valid.A deleted batch number cannot be reused until all thebatches are deleted, when the batch number count startsagain at 1.
7.3 OPEN EXISTING BATCH
This allows any batch to be opened to add readings. Ifthe gauge is switched off with the batch open, the currentbatch will re-open when the gauge is switched back on.
Note: If the probe is changed to another probe of thesame type the gauge will display a warning (Figure 49).
Figure 49. Changed probe warning screen
IfNO is selected, the serial number and calibration dataof the new probe are saved into the current batch as adata entry.
If YES is selected, the batch calibration details areretained and the user can locate and fit the originalprobe, if they wish.
7.4 REVIEW BATCHES
This allows the content of any batch to be displayed(Figure 50).
Figure 50. REVIEW BATCHES screen
Status
area
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
45/76
43
The Up/Down softkeys move the cursor through the listof readings.
The Status softkey toggles the status area on/off.
The Review Batches screen can contain the followinginformation:
Readings (including symbols indicating whetherthey are above or below limits if LIMITS ON isactivated).
Date stamp (entered whenever the batch is
opened). Probe change (includes date stamp and probe serial
number).
Recalibration (includes date stamp, recalibrationstamp and, for dual function probes, the probemode).
The status area in the bottom half of the screen displaysthe batch number and the type and serial number of theprobe used when the batch was created.
7.5 SET LIMITS
Upper and lower values can be set by the user to monitor
specification values.
To activate limits tick the LIMITS ON box (Figure 51).
Figure 51. SET LIMITS screen
To set the UPPER and LOWER limit values move thecursor to the limit required and press SEL.
http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
46/76
44
The gauge will display the current settings (Figure 52).Values shown are for illustration only.
Figure 52. Setting upper and lower limits
Adjust the values with Up/Down softkeys. When thecorrect value is displayed, press OK to enter.
With LIMITS ON activated the out-of-limits reading willbe indicated by a triple beep and a red light on thekeypad LED. A reading within limits is indicated by asingle beep and a green light. Out-of-limits symbolsand will be stored in the batch and can be seen when
the batch is reviewed.
7.6 FREE MEMORY
This option displays the amount of free memoryavailable for storage of readings (Figure 53).
Figure 53. FREE MEMORY screen
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
47/76
45
8 TRANSFERRING READINGS TO ACOMPUTER
Software is available which allows data to be transferred
from an Elcometer 4562
Top to a PC using the suppliedPC connection cable. Elcometer supplies three types ofdata transfer software:
Elcometer Data Transfer Software (EDTS+ ExcelLink)
Elcometer Data Collection Software (EDCS)
Elcometer Data Collection Software (EDCS+)
The Elcometer 4562 Top gauge is supplied with a CDcontaining EDTS+ Excel Link, EDCS, and ademonstration version of EDCS+. This software can alsobe downloaded from the Elcometer websitewww.elcometer.com.
8.1 EDTS+ Excel Link
EDTS+ Excel Link allows the user to transfer data fromthe memory of the gauge into Microsoft Excel. The datacan then be processed in software such as Word orExcel.
8.2 EDCS and EDCS+
EDCS+ allows the user to transfer data from the memoryof the gauge to a PC. It also includes all the charts thatyou may need together with a report designer to let you
design the report the way you wish to see it. Datamanagement and archiving is also included.
EDCS is a limited functionality version of EDCS+.
EDCS+ is available from your local Elcometer supplier ordirectly from Elcometer.
8.3 Software installationEDTS+ Excel Link, EDCS and EDCS+ are compatiblewith computers operating on Windows 95 (servicepack 1), 98, ME, NT, 2000 and XP. The installationconforms to the standard Windows method usingSetup.exe.
-
7/27/2019 Elcometer 456 User Manual.pdf
48/76
46
8.4 Connecting the data transfer cable
1. Plug the 9-pin female D-type connector on one endof the data transfer cable into the COM port on the
PC.2. Plug the 5-pin connector on the other end of the
data transfer cable into the RS232 connector on theside of the gauge - see Interfaces on page 12.
3. Switch on the gauge and ensure that the ReadingScreen is displayed.
8.5 Setting-up the software
1. Start the communications software (EDTS+, EDCS
or EDCS+).
2. Select gauge type 456.
8.6 Transferring data
Follow the on-screen instructions.
-
7/27/2019 Elcometer 456 User Manual.pdf
49/76
47
9 PROBES
An extensive range of probes is available for theElcometer 4562 Coating Thickness Gauge. Probes for
ferrous (F), non-ferrous (N) and dual ferrous/non-ferrous(FNF) operation are available as either integral (built-in)or separate options.
Separate probes are fully interchangeable and areavailable in standard, PINIP and miniature formats.
PINIP format (Plug-In Integral Probe) is an integral
style probe which plugs in to a separate gauge. Thisprovides all the benefits of an integral gauge and theflexibility of a separate gauge in a single unit.
Miniature separate probes permit measurements wherespace is restricted.
9.1 Ferrous (F) probesF probes measure the thickness of non-magneticcoatings on magnetic substrates. They can be used onpaint, plastic, galvanising, enamel, powder paint, hardchrome and other coatings such as electro-less nickelapplied to steel or iron.
9.2 Non-ferrous (N) probesi
N probes measure the thickness of non-magneticcoatings on non-magnetic metals. They can be used onanodising, paint, plastic coatings, powder paint, etc.applied to aluminium, brass, non-magnetic stainless
steel, etc.
9.3 Dual ferrous/non-ferrous (FNF) probes
FNF probes are dual function, F and N in one probe. FNFgauges will automatically detect the type of substrateand set the mode accordingly. Alternatively the mode
can be set manually - see PROBE: on page 23 andCoatings on galvanised or metallised steel on page 49.
i. Using an N probe (or an FNF probe manually setto N1) on a ferrous substrate will give a reading,but the reading will be incorrect.
-
7/27/2019 Elcometer 456 User Manual.pdf
50/76
48
9.4 Probe interchangeability
The following table shows which probes can be used inthe three types of Elcometer 4562 Gauge.
* indicates probes available for integral type gauges.
Probe type Gauge Type
F N FNF
FERR
OUS
F1 * 9 8 9
F2 * 9 8 9
F1 2 * 9 8 9
F3 * 9 8 9
F1 right angle 9 8 9
F2 right angle 9 8 9
F1 2 right angle 9 8 9
F1 telescopic 9 8 9
F2 telescopic 9 8 9
F1 PINIP 9 8 9
F2 PINIP 9 8 9
F1 2 PINIP High Temp 9 8 9
F3 PINIP 9 8 9
F3 miniature straight 9 8 9
F3 miniature 45 9 8 9
F3 miniature 90 9 8 9
NON-FERROU
S
N1 * 8 9 9
N2 8 9 9
N1 right angle 8 9 9
N1A anodisers 8 9 9
N1 PINIP 8 9 9
N3 miniature straight 8 9 9
N3 miniature 45 8 9 9
N3 miniature 90 8 9 9
DUAL
FNF1 * 8 8 9
FNF1 right angle 8 8 9
FNF PINIP 8 8 9
-
7/27/2019 Elcometer 456 User Manual.pdf
51/76
49
9.5 F1 2 probes
The F1 2 scale combines the F1 scale with the F2 scalein a single probe. The user selects the appropriate range
for the work in hand. The resolution of the gauge isdependent on the scale selected on the gauge.
9.6 F1 2 High temperature PINIP probes
These probes are capable of measuring onsurfaces up to 250C (480F). Wear
appropriate protective clothing and takecare to avoid bodily contact with the hotsurface during measurement.
Do not exceed the maximum measurement speed whenusing these probes - see page 53.
Note: Refer to special calibration procedure - see
Calibrating High Temperature PINIP Probes onpage 33.
9.7 Coatings on galvanised or metallised steel
The FNF probe in fixed N1 mode may be used tomeasure paint coatings on galvanised, aluminium (Al) or
zinc (Zn) sprayed steel substrates.1. Set the gauge to the N1 mode
(MENU/SETUP/PROBE).
2. Zero and calibrate the gauge on a sample of thecoated steel - see Calibration adjustment onpage 25.
Care must be taken to ensure that the calibrationconditions are not affected by changes in the zinc oraluminium coating thickness. This can bedetermined by checking the zero over an area of thegalvanised or metal-coated steel. Metal coatings onsteel above 50 m (2 mil/thou) should be consistentenough to obtain a stable zero on the layer of metal.
3. Take readings.
-
7/27/2019 Elcometer 456 User Manual.pdf
52/76
50
10 PERSONALISED WELCOME SCREEN
A personalised welcome screen can be created anddownloaded into the gauge.
Screen dimensions are 128 pixels x 64 pixels. Thewelcome screen is typically used to personalise thegauge with a logo, serial number, user name, etc. This isthe first screen displayed when the gauge is switched on.
10.1 Creating the screen
1. Download Elcometer Welcome Screen Wizardsoftware. This software is available free of chargefrom the downloads section of the Elcometerwebsite, www.elcometer.com.
2. Connect gauge to PC using 456 to PC connectioncable.
3. Press key to switch on gauge.
4. Ensure Reading Screen is displayed.
5. Run Welcome Screen Wizard software and followthe on-screen instructions.
10.2 Deleting the screen1. Run the Welcome Screen Wizard software.
2. Click Next.
3. Select Create a new screen setup.
4. Click Next.
5. Select Disabled.6. Click Next.
Follow the remaining on-screen instructions to delete thewelcome screen.
http://www.elcometer.com/http://www.elcometer.com/ -
7/27/2019 Elcometer 456 User Manual.pdf
53/76
51
11 STORAGE AND TRANSIT
This gauge incorporates a Liquid CrystalDisplay (LCD). If the display is heated
above 50C (120
F) it may be damaged.
This can happen if the gauge is left in a carparked in strong sunlight.
Always store the gauge in its carrying pouch when it isnot being used.
Remove the batteries from the gauge and store them
separately if the gauge is to remain unused for a longperiod of time. This will prevent damage to the gauge in theevent of malfunction of the batteries.
12 MAINTENANCE
You own one of the finest hand-held coating thickness
gauges in the world. If looked after, it will last a lifetime.The gauge does not contain any user-serviceablecomponents. In the unlikely event of a fault, the gaugeshould be returned to your local Elcometer supplier ordirectly to Elcometer. Contact details are stored in thegauge - MENU/ABOUT/CONTACT.
Worldwide: [email protected] USA/Canada: [email protected]
Note: Probes will eventually wear. Probe life willdepend on the number of measurements taken andhow abrasive the coating is. Probe life can beprolonged by careful positioning of the probe on the
surface.Replacement separate and PINIP probes can befitted by the user without the need to return thegauge for service.
Gauges with an integral probe have to be returnedfor re-programming or replacement if the probe
becomes worn or damaged.
-
7/27/2019 Elcometer 456 User Manual.pdf
54/76
52
13 STATISTICS TERMINOLOGY
Term Meaning
COEF OF VARIATN Coefficient of Variation. Thestandard deviation divided bythe mean for a group ofreadings, expressed as apercentage.
HIGHEST READING The value of the maximumthickness in a group ofreadings.
LOWEST READING The value of the minimumthickness in a group ofreadings.
MEAN The average of a group ofreadings; the sum of theindividual readings divided bythe number of readings.
NO. OF READINGS Number of Readings. The
running value for the number ofreadings taken in a group. Inthe case of the averaging orcounted average, the Numberof Readings is the number ofvalues recorded, not the total
number of readings taken.
STD DEVIATION Standard Deviation. Astatistical measure of thespread of values in a group ofreadings.
-
7/27/2019 Elcometer 456 User Manual.pdf
55/76
53
14 TECHNICAL DATA
14.1 Measurement speed
>60 readings per minute.This rate is reduced to >40 readings per minute whenInfrared printer output is active - see OUTPUT: onpage 23.
When measuring high temperature materialsmeasurement speed must be reduced to prevent
overheating of the probe. The maximum measurementspeed of the High Temperature PINIP probe at 250C(480F) is 4 readings per minute.
14.2 Minimum substrate thickness
Ferrous: 300 m (12 mils)
Non-ferrous: 100 m (4 mils)Measurements can be taken on thinner substrates if2-point calibration is carried out either side of therequired substrate thickness, however gauges will havereduced range when adjusted for thin substrates.
14.3 Probe operating temperatureSeparate ferrous probes: 150C (300F)
High temperature PINIP probes: 250C (480F)
Miniature probes without outer sleeve: 150C (300F)
All other probes: 80C (176F)
14.4 Physical
Weight(includingbatteries):
Separate Probe (FNF1), 190g (6.7oz)Separate Probe (PINIP), 155g (5.5oz)Integral Probe, 130g (4.6oz)
Dimensions: 130 mm x 70 mm x 35 mm
(5.12" x 2.76" x 1.38")Gaugeoperatingtemperature:
0C to 50C (32F to 120F)Operation outside these limits dependsupon climatic conditions.
Case: High impact ABS
http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
56/76
54
14.5 Power supply
Internal batteries, 2 x LR03 (AAA), alkalinej dry batteriesor rechargeablek equivalents.
Battery life
30l hours to 40 hours continuous use with alkaline drybatteries. (15 000 to 20 000 readings at an average of 8readings per minute.) Battery life is reduced by one thirdwhen using the backlight.
14.6 Packaging
The gauge is packed in cardboard and plasticpackaging. Please ensure that this packaging isdisposed of in an environmentally sensitive manner.Consult your Local Environmental Authority for furtherguidance.
j. Alkaline batteries must be disposed of carefully toavoid environmental contamination. Please consultyour local environmental authority for informationon disposal in your region.Do not dispose of any batteries in fire.
k.Rechargeable batteries can be used if they are
charged outside the gauge.
l. Battery life is reduced to approximately 25% of drybattery life when using rechargeable batteries.Follow the instructions provided by the batterymanufacturer when charging and disposing ofrechargeable batteries.
-
7/27/2019 Elcometer 456 User Manual.pdf
57/76
55
15 ACCESSORIES
The Elcometer 4562 is complete with all the itemsrequired to get started and take measurements.
Many of the following accessories are optional.However, some are consumable items that may need tobe replaced over the lifetime of the gauge.
All these accessories are available from Elcometer, oryour local Elcometer supplier. At time of ordering pleasequote the sales part number which follows the
description of each accessory.
15.1 Consumable items
15.2 ProbesFull details of the extensive range of 456 probes can beobtained from Elcometer, your local Elcometer supplieror the Elcometer website, www.elcometer.com.
15.3 Foil sets
Individual foils in the range 12.5 m to 20 mm (0.5 mil to
790 mils) and customised sets chosen from this rangeare also available. Consult your local Elcometer supplier.
15.4 Calibration certificates for foils
Certificates traceable to National Standards includingUKAS and NIST are available on request.
15.5 Test certificates
A certificate with results of a standard test on known foilvalues over the full range of the probe. Order using salespart number TEST-456.
Alkaline LR03 (AAA) Battery(2 required):
T9999329-
2.2 mm (85 mils) 8 pieces: T9904199F1.3 mm (51 mils) 3 pieces: T9904199G
5.5 mm (220 mils) 4 pieces: T9904199J
15 mm (595 mils) 4 pieces: T9904199K
http://www.elcometer.com/http://www.elcometer.com/ -
7/27/2019 Elcometer 456 User Manual.pdf
58/76
56
15.6 Coated thickness standards includingcertificate
15.7 Probe placement jig
To aid probe positioning on small components a probe
placement jig is available and an adapter suitable for usewith the full range of miniature probes is also availableas an accessory.
15.8 Probe adapters
15.9 Infrared portable printer
15.10 Miniprinter
42 column, rechargeable battery powered Miniprinter
complete with charger. Three charger options:
Miniprinter spares
Ferrous Standard (4 Values): T995111261
Non-Ferrous Standard (4 Values): T995111271Ferrous Standard (2 Values): T995166001
Non-Ferrous Standard (2 Values): T995166011
Probe placement jig: T95012880
Jumbo Hand Grip (F and N probes): T9997766-
Jumbo Hand Grip (FNF probes): T99913225
V Adapter for pipes (F & N probes): T9997381-
V Adapter for pipes (FNF probes): T99913133
Infrared Portable Printer: X45613877
Thermal Printer Paper: T99913878
230V (UK Plug): X4569964B
230V (European Plug): X4569964C
110V (US Plug): X4569964D
456 to printer connection cable(25-pin):
T45616267
Ribbon Cassettes (Pack of 5): T9769992-
Paper Rolls (Box of 20): T9999993-
-
7/27/2019 Elcometer 456 User Manual.pdf
59/76
57
15.11 Bench Stands
15.12 PC Connection Cable
Note: A 9-pin to 25-pin adapter may be required for certainPC RS232 ports.
Integral Probe Version: T45616161
Integral/Separate Probe Version: T45616162
456 to PC Connection Cable (9-pin): T45616217
-
7/27/2019 Elcometer 456 User Manual.pdf
60/76
58
16 RELATED EQUIPMENT
Elcometer produces a wide range of coating thicknessgauges and associated paint inspection equipment.
Users of the Elcometer 4562
may also benefit from thefollowing Elcometer products:
Uncured powder thickness gauges
Coatings analyser
Inspection management software
Mechanical coatings thickness gauges
Appearance testers
Adhesion testers
For further information contact Elcometer, your localElcometer supplier or visit www.elcometer.com
17 FITTING THE WRIST HARNESS
1. Pass harnessround pin
2. Pass harnessthrough loop
3. Pull tight
-
7/27/2019 Elcometer 456 User Manual.pdf
61/76
59
18PROBE
ME
ASUREMENT
PERFORMAN
CE
Scale
Totalrange
Accuracy
a
a.Whicheveristh
egreater.Lower
valueachievedwhencalibratedclo
setothethicknes
stobemeasured.
Resolution
inrange
F1F12(F1mode)
FNF1
N1,N1A
0mto1500m
1%to3%
or2.5m
0.1m
1.0m
0
mto99.9m
1
00mto1500m
0milto60mils
1%to3%
or0.1mil
0.01mil
0.1mil
0
milto4.99mils
5
milsto60mils
F2F12(F2mode)
N2
0mmto5.0mm
1%to3%
or0.02mm
1.0m
10m
0
mmto0.99mm
1
.0mmto5.0mm
0milto200mils
1%to3%
or1mil
0.1mil
1mil
0
milto49.9mils
5
0milsto200mils
F3
0mmto13mm
1%to3%
or0.05mm
1.0m
10m
0
mmto1.99mm
2
mmto13mm
0milto500mils
1%to3%
or2.0mils
0.1mil
1mil
0
milto99.9mils
1
00milsto500mils
FM3
NM3
0mto500m
1%to3%
or2.5m
0.1m
1.0m
0
mto99.9m
1
00mto500m
0milto10mils
1%to3%
or1.0mil
0.01mil
0.1mil
0
milto3.99mils
4
milsto10mils
-
7/27/2019 Elcometer 456 User Manual.pdf
62/76
60
19PROBECA
PABILITIES
19.1Integralp
robes
Probetype
Minimum
convexsurface
diameter
Minimum
concave
surfacera
dius
Headroom
Minimum
sample
diamete
r
Calfoil
valuea
a.Thisistherec
ommendedmax
imumcalibration
foilvaluetoach
ievethespecifie
daccuracyunde
rthese
measurementsconditions
F1(orF12setfor
F1
operation)
4mm(0.16)
25mm(0.98)
130mm(
5.1)
4mm(0.16)
250m
(10mil)
F2(orF12setfor
F2
operation)
4mm(0.16)
25mm(0.98)
135mm(
5.3)
8mm(0.32)
1mm(4
0mil)
F3
15mm(0.59)
40mm(1.57)
150mm(
5.9)
14mm(0.55)
2.5mm
(100mil)
N1(N)
35mm(1.38)
25mm(0.98)
130mm(
5.1)
6mm(0.24)
250m
(10mil)
FNF1(N)
38mm(1.50)
25mm(0.98)
135mm(
5.3)
8mm(0.32)
250m
(10mil)
FNF1(F)
4mm(0.16)
25mm(0.98)
135mm(
5.3)
4mm(0.16)
250m
(10mil)
-
7/27/2019 Elcometer 456 User Manual.pdf
63/76
61
19.2Separate
ferrousprobe
s
Probetype
Minimum
convexsurf
ace
diameter
Minimum
concave
surfacera
dius
Headroom
Minimum
sample
diameter
Calfoilvaluea
a.Thisistherec
ommendedmax
imumcalibration
foilvaluetoach
ievethespecifie
daccuracyunde
rthese
measurementsconditions
F1(orF12settoF1)
4mm(0.16)
25mm(0.98)
85mm(3
.35)
4mm(0.
16)
250m(10mil)
F2(orF12settoF2)
4mm(0.16)
25mm(0.98)
89mm(3
.50)
8mm(0.
32)
1mm(40mil)
F1RightAngle
(orF12settoF1)
4mm(0.16)
25mm(0.98)
28mm(1
.10)
4mm(0.
16)
250m(10mil)
F2RightAngle
(orF12settoF2)
4mm(0.16)
25mm(0.98)
32mm(1
.26)
8mm(0.
32)
1mm(40mil)
F1Telescopic
4mm(0.16)
25mm(0.98)
32mm(1
.26)
4mm(0.
16)
250m(10mil)
F2Telescopic
4mm(0.16)
25mm(0.98)
36mm(1
.42)
8mm(0.
32)
1mm(40mil)
F3
15mm(0.59)
40mm(1.57)
102mm(
4.02)14mm(0
.55)
2.5mm(100mil)
-
7/27/2019 Elcometer 456 User Manual.pdf
64/76
62
19.3Separate
non-ferrousp
robes
Probetype
Minimum
convexsurface
diameter
Minimum
concave
surfacera
dius
Headroom
Minimum
sample
diameter
Calfoil
valuea
a.Thisistherec
ommendedmax
imumcalibration
foilvaluetoach
ievethespecifie
daccuracyunde
rthese
measurementsconditions
N1
35mm(1.38)
25mm(0.9
8)
85mm(3.35)
6mm(0.24)
250m(10mil)
N1RightAngle
35mm(1.38)
25mm(0.9
8)
28mm(1.10)
6mm(0.24)
250m(10mil)
N1A
Anodisers
Probe
35mm(1.38)
25mm(0.9
8)
85mm(3.35)
6mm(0.24)
250m(10mil)
N2
100mm(3.97)
150mm(5.90)
85mm(3.35)
14mm(0
.55)
1mm(40mil)
-
7/27/2019 Elcometer 456 User Manual.pdf
65/76
63
19.4Separate
dualFNF
Probetype
Minimum
convexsurface
diameter
Minimum
concave
surfacera
dius
Headroom
Minimum
sample
diameter
Calfoil
valuea
a.Thisistherec
ommendedmax
imumcalibration
foilvaluetoach
ievethespecifie
daccuracyunde
rthese
measurementsconditions
FNF1(N)
38mm(1.50)
25mm(0.98)
88mm(3
.46)
8mm(0.32)
250m(10mil)
FNF1(F)
4mm(0.16)
25mm(0.98)
88mm(3
.46)
4mm(0.16)
250m(10mil)
FNF1RightAng
le
(N)
38mm(1.50)
25mm(0.98)
34mm(1
.34)
8mm(0.32)
250m(10mil)
FNF1RightAng
le
(F)
4mm(0.16)
25mm(0.98)
34mm(1
.34)
4mm(0.16)
250m(10mil)
-
7/27/2019 Elcometer 456 User Manual.pdf
66/76
64
19.5PINIPp
robes
Probetype
Minimum
convexsurface
diameter
Minimum
concave
surfacera
dius
Headroom
Minimum
sample
diameter
Calfoil
valuea
a.Thisistherec
ommendedmax
imumcalibration
foilvaluetoach
ievethespecifie
daccuracyunde
rthese
measurementsconditions
F1(orF12settoF1)
4mm(0.16)
60mm(2.36)
155mm(
6.10)4mm(0.16)
250m(10mil)
F2(orF12settoF2)
4mm(0.16)
60mm(2.36)
159mm(
6.25)8mm(0.32)
1mm(4
0mil)
F3
15mm(0.59)
45mm(1.77)
169mm(
6.65)14mm(0
.55)
2.5mm(100mil)
N1
35mm(1.38)
50mm(1.97)
155mm(
6.09)6mm(0.24)
250m(10mil)
FNF1(N)
38mm(1.50)
55mm(2.17)
156mm(
6.15)8mm(0.32)
250m(10mil)
FNF1(F)
4mm(0.16)
55mm(2.17)
156mm(
6.14)4mm(0.16)
250m(10mil)
-
7/27/2019 Elcometer 456 User Manual.pdf
67/76
65
19.6Separate
miniatureferrousprobes
ProbeType
M
inimum
convex
surface
diameter
M
inimum
concave
surface
ra
dius
M
inimum
sa
mple
diameter
Minimum
access
heig
ht
Minimum
acce
ss
width
Overal
llength
(headr
oom)
F,Straight,
45mm(1.77)
1.5mm(0.06)6.5mm(0.26)
3mm(0.12)
6mm(0.24)
150mm
(5.91)
F,Straight,
150mm(5.9)
1.5mm(0.06)6.5mm(0.26)
3mm(0.12)
6mm(0.24)
260mm
(10.24)
F,45,
45mm(1.77)
1.5mm(0.06)6.5mm(0.26)
3mm(0.12)
18m
m(0.71)
7mm
(0.28)
145mm
(5.71)
F,45,
150mm(5.9)
1.5mm(0.06)6.5mm(0.26)
3mm(0.12)
18m
m(0.71)
7mm
(0.28)
250mm
(9.84)
F,90,
45mm(1.77)
1.5mm(0.06)6.5mm(0.26)
3mm(0.12)
16m
m(0.63)
7mm
(0.28)
140mm
(5.51)
F,90,
150mm(5.9)
1.5mm(0.06)6.5mm(0.26)
3mm(0.12)
16m
m(0.63)
7mm
(0.28)
245mm
(9.64)
-
7/27/2019 Elcometer 456 User Manual.pdf
68/76
66
19.7Separate
miniaturenon
-ferrousprob
es
ProbeType
M
inimum
convex
diameter
M
inimum
concave
ra
dius
M
inimum
sa
mple
diameter
Minimum
access
heig
ht
Minimum
acce
ss
width
Overal
llength
(headr
oom)
NF,Straight,
45mm(1.77)
3
mm(0.12)
25
mm(0.98)
4mm(0.16)
6mm(0.24)
150mm
(5.91)
NF,Straight,
150mm(5.9)
3
mm(0.12)
25
mm(0.98)
4mm(0.16)
6mm(0.24)
260mm
(10.24)
NF,45,
45mm(1.77)
3
mm(0.12)
25
mm(0.98)
4mm(0.16)
18m
m(0.71)
7mm
(0.28)
145mm
(5.71)
NF,45,
150mm(5.9)
3
mm(0.12)
25
mm(0.98)
4mm(0.16)
18m
m(0.71)
7mm
(0.28)
250mm
(9.84)
NF,90,
45mm(1.77)
3
mm(0.12)
25
mm(0.98)
4mm(0.16)
16m
m(0.63)
7mm
(0.28)
140mm
(5.51)
NF,90,
150mm(5.9)
3
mm(0.12)
25
mm(0.98)
4mm(0.16)
16m
m(0.63)
7mm
(0.28)
245mm
(9.64)
-
7/27/2019 Elcometer 456 User Manual.pdf
69/76
67
20ERROR
ME
SSAGES
Undercertainco
nditionsthegaugewilld
isplayerrormessages(Figure54).Thesemessagesare
normally
clearedbypress
inganyoneofthesoftkeys.The
causeoftheerrorwillbeindicatedbythemessageand
shouldbecorrec
tedbeforeproce
eding.
Errormessage
Causes
A
ctiontotake
#1-PROBE
Probe-to-gaugecommunicationfailure.
In
tegralgauge-re
turntoElcomete
r*.
Separategauge-
removeprobea
ndrefit.
If
errorpersists,re
turntoElcomete
r*.
#2-PROBE
Corruptdata
outputfromprobe.
In
tegralgauge-re
turntoElcomete
r*.
Separategauge-
trynewprobe.
If
errorpersists,re
turntoElcomete
r*.
#3-PROBE
Internalerror.
ReturntoElcomet
er*.
PROBE
UNSUPPORTED
a)Probeisd
ualFNF,butgau
geisferrous
only,ornon-
ferrousonly.Seepage47.
b)Oldgaug
esoftwaredoes
notsupport
newprobe.
a)Gaugemodel
FandgaugemodelN
ca
nnotbeusedwithadualFNFprobe.
b)
Return
toE
lcometer*
for
software
upgrade.
PROBE
CHANGED
Probehasbeenchanged.
Calibrationadjustm
entrequired.
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/27/2019 Elcometer 456 User Manual.pdf
70/76
68
*ContactElcometeroryourlocalElcometerSupp
liertoarrangereturn.
Figure54.Exampleerrormessage-noprobe
isconnectedtogauge
UNSTABLE
READING
a)Externale
lectro-magnetic
interference.
b)Gaugefaulty.
a)Isolatefromele
ctro-magnetic
in
terference.
b)ReturntoElcom
eter*.
VALUETOO
LARGE
Numericalerror.
Switchgaugeoffthenonagain.
If
errorpersistsco
ntactElcometer.
CALIBRATION
Incorrectcalibrationcalculation.
Re-calibrate.
If
errorpersists,re
turntoElcomete
r*.
LANGUAGE
MEMORY
Softwareerror.
ReturntoElcomet
er*.
Errormessage
Causes
A
ctiontotake
-
7/27/2019 Elcometer 456 User Manual.pdf
71/76
69
21 INDEX
Numerics2 Point calibration method 25
AAccessories 55
AS 3894 26ASTM B 244 4ASTM B 499 4
ASTM D 1400 4ASTM D1186 4Australian calibration method
26
BBatch mode 37
Batch numbering 42BatchesCopying (Cloning) 38Reviewing 42
Batching 37Creating new batch 38Data collection method 39
Exiting 38Opening existing batch 42Batteries
Fitting 6Life of 54Ordering 55
Precautions 51
RechargeableCharging 54Life of 6
Specification 54Baud rate 23
BeepChanging volume 24
Switching off 24
BS 3900 (C5) 4BS 5411 (11) 4
BS 5411 (3) 4BS 5599 4BS EN ISO 1461 4
CCable
PC to gauge 57CAL softkey
Flashing 27Calibration
Adjustment 27Coated standards 27
Coated thicknessstandards 28FNF probes 25
Foils 27Locking 16Methods 25Shims 27
Calibration Adjustment 28Procedure, 2Point 32
Procedure, Rough 32Procedure, Smooth 29Procedure, Special
substrate 32Calibration certificates 55Calibration methods
2 POINT 25AUSTRALIAN 26ISO 26
ROUGH 25SET OFFSET 26SPECIAL SUB 25
SSPC 26
SWEDISH 26ZERO OFFSET 25
Clock and alarm 23Cloning a batch 38Coated thickness standards
28Spares 56
-
7/27/2019 Elcometer 456 User Manual.pdf
72/76
70
CoatingsOn galvanised steel 49On metallised steel 49
Coefficient of Variation 52
ComputerConnection cable 57
Computer, transferring data to45
Copying a batch 38
DData collection method 39
Counted average 40Normal 40
Data transfer 45Data, activating 22Date stamp 22
Delete all batches 21Delete last reading 20Delete single batch 21
Deleting readingsPermanent or tagged 22
Dimensions 53
DIN 50981 4DIN 50984 4
EEDCS 45
EDCS+ 45EDTS+ Excel Link 45Elcometer 456
Features 4Overview 3Error messages 67
ESC 29Extended menu
Contents 20
FFactory calibration 17FNF 47Foils 27
Ordering 55
GGalvanised steel 49Gauge information 16
HHigh temperature 49Highest reading 52
IImmediate mode 37
Infrared interface 12Effect on measurement
speed 53Interface 12
Infrared 12
RS232 12International gauge settings
17ISO 19840 4, 26ISO 2178 4ISO 2360 4
ISO 2808 4ISO calibration method 26
KKeypad 8
LLanguage, selecting 11Limits
Setting 43
Lowest reading 52
MMaintenance 51Mean 52Measurement speed 53Memory
Amount available 44Menu items
2-POINT 25ABOUT 16
AUSTRALIAN 26AUTO SWITCH OFF 24BACKLIGHT 16
-
7/27/2019 Elcometer 456 User Manual.pdf
73/76
71
BAUD RATE 23BEEP VOLUME 24CAL METHOD 22
CALIBRATION LOCKED
16CLEAR STATS 35
CLOCK/ALARM 23CONTACT 16DATA 22
DATA MENU 37DELETE 20
DELETED READING 22DISPLAY 35ENLARGE STATS 34EXIT BATCHING 38
EXTENDED MENU 18FACTORY CAL 17FREE MEMORY 44
GAUGE INFORMATION16
HELP 17INFRA RED PRINTER 24INTL GAUGE 17ISO 26
LANGUAGES 24
OPEN EXISTING BATCH42
OPEN NEW BATCH 38OPENING SCREEN 24OUTPUT 23PRINT/OUTPUT 20PROBE 23
PROBE INFORMATION16
RESET 17REVIEW BATCHES 42
ROUGH 25RS232 BIT IMAGE 23RS232 PLAIN TEXT 24SELECT STATS 36SET LIMITS 43SET OFFSET 26
SETUP 22SHOW DATE STAMP 22SMOOTH 25
SOFTKEY ENABLED 22SPECIAL SUB 25SSPC 26
STATISTICS 22
STATS MENU 34STATS ON LCD 35
SWEDISH 26TAG 22UNITS 23
US GAUGE 17ZERO OFFSET 25
MenusExtended off (simple) 16Extended on 18Overview 15
Structure 18Metallised steel 49Minimum substrate thickness
53Mode
Batch 37Immediate 37
NNumber of readings 52
OOn/off 9
Opening screenCreating 50Disabling 24
PPackaging 54PINIP probes
Fitting of 7
High temperature 49Power supply 54
PrinterInfrared 56Miniprinter 56
Printing
All batches of readings 20Batch of readings 20Statistical summary 20
-
7/27/2019 Elcometer 456 User Manual.pdf
74/76
72
Printing readings 23To infrared printer 24
Probe positioning jig 56
Probes 47
Adapters 56Capabilities 60-66
Extending life of 51F (Ferrous)
What used for 47
F1 2High Temperature
PINIP 49Substrate selectionmodes 23
F1 2 overview 49
Fitting of 7FNF (Ferrous/Non-ferrous)
Substrate selection
modes 23, 47What used for 47
Maximum temperature 53N (Non-ferrous)
What used for 47Performance 59
PINIP
What used for 47Replacement of 51
Wear of 51Which one to use? 48
QQuick-start 5
RReading screen 10
Typical 14Reading, taking 13
Resetting the gauge 17
Rough calibration method 25RS232 interface 12
SScale selection, F1 F2 23Screen
Welcome information 10
Screen symbolsBattery condition 6Definition of 17
Screen symbols, explanation
of 17Set offset 26
Setting up the gauge 22Shims 27
Ordering 55
Simple menuTurning on/off 16
Smooth calibration method 25Softkey
BACK 15CAL 14, 31
Flashing 14Procedure 28
DATA COLLECTION
METHOD 39ESC 29MENU 14OK 31SEL 15SET 30, 31
TEST 29
UP/DOWN 15ZERO 31
SoftkeysDATA 37
Software 45Special sub calibration method25
SS 18 41 60 26SSPC calibration method 26SSPC-PA 2 26SSPC-PA2 4
Standard deviation 52Standards 4
AS 3894 26Coated thickness 56ISO 19840 26SS 18 41 60 26
SSPC-PA 2 26Statistics
Activating 22
-
7/27/2019 Elcometer 456 User Manual.pdf
75/76
73
Choice of 36Clearing 35Displaying on screen 35
Dual function probes 35
Increasing size on display34
Overview 34Terminology 52
Substrate selection
Automatic 23, 47Manual 23, 47
Substrate thicknessMinimum 53
Swedish calibration method26
Switching on/off 9Automatically 9, 24
TTemperature
Effect on measurementspeed 53High, measuring of 49
Operating 53Probes, maximum 53
Test certificates 55
Test readings 31Testing calibration of gauge31
Tick box 15Transferring data to computer45
UUnits
Automatic setting of 23Manual setting of 23
Up/Down 15
US gauge settings 17
WWeight 53
Welcome screenCreating 50Disabling 24
Wrist harness, fitting of 58
ZZero offset calibration method
25
-