Service Manual

Diagnostic Ultrasound System

Pro Focus 2202

Valid from Software XPE0118-1353

Page 2: Ecograf System Pro Focus 2202(BI2201-E)

1 General Information

2 Service Notes

3 Checking Procedure

4 Trouble Shooting

5 Adjustment Procedure

6 Mechanical Parts

7 Preventative Maintenance

8 Theory of Operation

9 System Interconnections

10

Page 3: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 General Information

BI2201-E 1-1

Section 1

GENERAL INFORMATION CONTENTS PAGE

1 Safety Aspects...........................................................................................................................................2 1.1 Terms and Symbols used......................................................................................................................2 1.2 WARNINGS and CAUTIONS: ..............................................................................................................3 2 System Overview.......................................................................................................................................4 2.1 Front View..............................................................................................................................................4 2.2 Back View ..............................................................................................................................................5 2.3 System Configuration ............................................................................................................................6 2.4 Models ...................................................................................................................................................6 2.5 Modules and PC Boards........................................................................................................................7 3 Revision System......................................................................................................................................11 4 Special Tools and Equipment ................................................................................................................12 4.1 Checking Procedure (Section 3)..........................................................................................................12 4.2 Troubleshooting (Section 4) ................................................................................................................13 4.3 Adjustment Procedure (Section 5).......................................................................................................13 4.4 Mechanical Parts (Section 6)...............................................................................................................13 4.5 Preventative Maintenance (Section 7).................................................................................................13 4.6 Electrical Safety Test ...........................................................................................................................13

Page 4: Ecograf System Pro Focus 2202(BI2201-E)

General Information System Pro Focus 2202

1-2 BI2201-E

1 Safety Aspects The Pro Focus 2202 System complies with: • EN/IEC 60601-1 • EN 60601-2-37 • UL2601-1 • CSAC22.2 No.601.1. The Pro Focus 2202 System is classified as follows:

• Array Input Module: Type B (Body)

• Single-element Transducer Module: Type B (Body)

• ECG Input Module: Type BF (Body Floating)

1.1 Terms and Symbols used Throughout this manual the following terms are used to indicate a situation where safety precautions are required: "WARNING": Indicates a situation involving risk of injury or loss of life to personnel or

patient. "CAUTION": Indicates a situation involving risk of damage to the instrument or other

equipment connected.

Symbol Name Description

ATTENTION Consult ACCOMPANYING User Guide (BB0340) when this sign is

encountered on the instrument, to avoid reducing its safety

Potential Equalisation

Terminal connected to the chassis. Should be connected to corresponding terminals on other equipment to eliminate potential differences.

Protective Earth

Additional Protective Earth

Type CF CF: Isolated from earth. Maximum Patient Leakage Current under:

Normal Condition ≤10μA, Single Fault Condition ≤ 50μA

Type BF BF: Isolated from earth. Maximum Patient Leakage Current under:

Normal Condition ≤100μA, Single Fault Condition ≤ 500μA

Type BF BF, DEFIBRILLATOR-PROOF

Type B B: Maximum Patient Leakage Current under: Normal Condition

≤100μA, Single Fault Condition ≤ 500μA

IP57 SEALING Dust and immersion protected according to IEC Publication 529

Stand-by Push button for switching the scanner from stand-by to active. (The

power supply cord is the means of separation from the main power supply.)

Off Main power supply off

On Main power supply on

Non-ionising radiation

Ultrasound Scanner emits acoustic radiation

STERILE Device is in a sterile condition

Table 1. IEC safety symbols

Page 5: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 General Information

BI2201-E 1-3

1.2 WARNINGS and CAUTIONS:

For your own and others safety please read the following carefully: Warnings: • Opening the instrument can expose live parts.

• Any work done on the open instrument with power on must only be done by B-K

Medical or their authorised representatives, who are aware of the hazards involved.

• Any repair on the 2202 must be followed by an electrical safety test to verify a continuous safe operation of the system.

• Only the original mains cable must be used NEVER USE EXTENSION CABLES!!

• The 2202 contains a Lithium battery. Under no circumstances must this battery be removed or replaced by the user as there is danger of explosion.

Personal Safety: Be aware that there may be a risk of infection due to contaminated equipment, especially puncture guides/needles and transducers. The following precautions should be taken:

• At the hospital ask the staff to sterilise transducers and puncture guides before

receipt. Consoles must be disinfected as recommended in the User Guide before any repair.

• When working with possible infected equipment, use gloves especially if you have open wounds or scratches.

• Possible infected equipment must be sterilised before handed over to customers. Follow the recommendations in the Transducer User Guide.

• Always wash your hands after working with the equipment.

• If you scratch yourself on contaminated equipment you should immediately contact the hospital staff or see a doctor.

Cautions: • Ensure that the two mains voltage selector switches on the rear panel of the

power supply are set to match the actual mains voltage.

• Always use correct fuses.

• Switch off all equipment before connecting or disconnecting their interfaces. Failure to do so could damage the equipment.

• The power supply cord is the means of connecting the 2202 from the main power supply.

Page 6: Ecograf System Pro Focus 2202(BI2201-E)

General Information System Pro Focus 2202

1-4 BI2201-E

2 System Overview

2.1 Front View

Page 7: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 General Information

BI2201-E 1-5

2.2 Back View

B A

1

2

RS232

5V 0.1 A

R

L

L

R

VGA 1

VGA 2

RGBS

RGBS

! 1 2

VGA 1

VGA 2 With Flat Screen option

Page 8: Ecograf System Pro Focus 2202(BI2201-E)

General Information System Pro Focus 2202

1-6 BI2201-E

2.3 System Configuration

The Diagnostic Ultrasound System 2202 is an ultrasound scanner optimised for use in: • B-mode • B-mode with Tissue Harmonic Imaging • B-mode + M-mode • Doppler mode including: • Spectral Doppler (PW & CW) • Power mode (including Directional Power Doppler) • Color mode (Color Doppler) • Triplex (B-mode plus Spectral Doppler and either Color Doppler or Power Doppler) • Contrast Imaging

2.4 Models

There are 3 versions of the Pro Focus 2202: Yellow, Green and, Blue. The available features for each version can be found in the product catalogue. (BG0143-O)

Page 9: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 General Information

BI2201-E 1-7

2.5 Modules and PC Boards

B-K Item Picture

ZD0774 Digital Beam former Board

ZD0773 / ZD0789 Core Board

ZD0784 Single Board PC

ZV0069 / ZV0070 CRT Monitor

ZV0071 Flat Screen

UL0041 Hard disk with software

Page 10: Ecograf System Pro Focus 2202(BI2201-E)

General Information System Pro Focus 2202

1-8 BI2201-E

UL0036 DVD/CD R/W Combo

ZE0797 / ZE0817 / ZE0821 / ZE0823

Channel Condition Module

ZE0798 / ZE0815 / ZE0822

Front End Board (3 Transducer connectors)

ZE0820 Front End Board (2 Transducer connectors)

ZG0346 Power Supply

ZH0757

Back Plane

ZH0758

Video IF board

ZH0759 HUB controller board

Page 11: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 General Information

BI2201-E 1-9

Keyboard Yellow and Green models

ZN0046 Incl. of TGC/Gain buttons/Cable

8 x DO0912 2 x SN0220 1 x SN0221

ZN0047

Keyboard for Blue model

Incl. of TGC/Gain buttons/Cable 8 x DO0912 2 x SN0220 1 x SN0221

ZN2220 Core Board Assembly ZD0773,UL0041,ZH0758,ZD0784

ZN2230 Core Board Assembly ZD0789,UL0041,ZH0758,ZD0784

Core Board Assembly ZD0773,UL0041,ZH0758,ZD0795ZN2231

Front End Assembly ZE0798 & 16 × ZE0797 ZN2221

Front End Assembly ZN2226 ZE0815 & 16 × ZE0817

Front End Assembly ZN2241 ZE0820 & 16 × ZE0821

(2 Transducer connectors)

Front End Assembly ZN2242 ZE0822 & 16 × ZE0823

Page 12: Ecograf System Pro Focus 2202(BI2201-E)

General Information System Pro Focus 2202

1-10 BI2201-E

ZN2201 / ZN2223 Single element module (option)

ZN2202 / ZN2222

MCM (Mover Control Module) for 3D Professional (option)

Page 13: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 General Information

BI2201-E 1-11

3 Revision System

All PC Boards in the 2202 are described by the board name, Type number, Revision code (ID) and PCB Version. The name of the board indicates the function of the board, for example the Core Board. The name of the board is changed only if the function of the board changes.

Bar Code Label

| | | || | | ||| | | | | | ||| | ||| | ZD0773 001 837 B-K Medical A/S

FGHIJKLMNOPQRSTUVXY Serial Number ABCDE

Revision level in this case: E

Part number

The Part number of the board, for example ZD0773, is the order number of the board. The number is changed if the modification of the existing PCB is found to be too extensive. The serial number of the board is a unique number used for tracking purposes. The revision code (ABCDEFGH.......) is related to the modifications made on the board. The revision letter is marked when modifications are made. When boards are ordered from the B-K Medical stock it is important that the serial number of the scanner is stated. The serial number ensures that the correct revision is shipped. The PCB version (number e.g. 3) is printed on the circuit board. The Version and Revision code can be read electronically. Note! Some modules (assemblies) that consist of a number of PCB’s does only have a Part No. and a serial number but no revision. E.g. Front-end assembly - ZN2226 consists of ZE0815 and 16 × ZE0817. Information about Hardware/ Software compatibility can be found in the B-K Medical Hot Line on the support section of www.bkmed.com (requires password)

Page 14: Ecograf System Pro Focus 2202(BI2201-E)

General Information System Pro Focus 2202

1-12 BI2201-E

4 Special Tools and Equipment

The tools and equipment listed below does not include standard tools and commonly used equipment.

4.1 Checking Procedure (Section 3)

• Transducer Phantom, model 254 (WQ 0973) for checking a 2202 system equipped with Array- and Single-element Transducers.

• Transducer Phantom, type 251 (WQ 0972) for checking a 2202 system equipped with the Endosonic Probe 1850 or 2050.

• Ruler (for checking the size of the image on the monitor)

• Blank CD-R (for testing CD Drive)

Further it is recommended to bring a 150 ml bottle containing preserving fluid for the phantom(s): Prescription for preserving fluid:

94 ml Glycerol (85%)

50 ml Rodalon (10%)

1000 ml purified Water

Page 15: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 General Information

BI2201-E 1-13

4.2 Troubleshooting (Section 4)

For accessing BIOS (new image on Harddisk drive) • BIOS Plug EL1011

4.3 Adjustment Procedure (Section 5)

For adjusting the trackball: • Trackball Adjustment Key QA0228 • Hex key 4 mm for adjusting monitor tilting base friction (CRT monitor only)

4.4 Mechanical Parts (Section 6)

• Static Control Service Kit, type 3M 8501 (WQ 0969) or similar. (when handling the static sensitive PCB’s).

• Torx keys size 10 and 20

• QA0229 Adjustment tool for Pro Focus This tool is needed for adjusting the monitor and keyboard bracket mechanics.

4.5 Preventative Maintenance (Section 7)

• Equipment necessary to perform the Preventative Maintenance is the equipment used in the Checking Procedure and Electrical Safety Test.

4.6 Electrical Safety Test

Testers required:

• Safety Tester

• HV Tester High Voltage test plugs/adopters required:

• WB 1275 HV test adaptor for type BF Transducers

• WJ 0246 HV Test Plug for Mains

• Electrical Safety Test Record (Enclosed)

Page 16: Ecograf System Pro Focus 2202(BI2201-E)

General Information System Pro Focus 2202

1-14 BI2201-E

This page is left blank intentionally

Page 17: Ecograf System Pro Focus 2202(BI2201-E)

2202 Electrical Safety Test Record

1 of 2

Equipment under test Serial no: Location:

For INSTRUCTIONS read the back of this sheet.

Test Date

Test Test Limits Insulation

Resistance Mains to Case 5 MΩ

Protective ground

Console (earth terminal on the back of the

scanner)

0,2 ohm

Protective ground

Monitor shelf (screw underneath the shelf)

0,2 ohm

Protective ground

Monitor (screw next to monitor power inlet)

0,2 ohm

Protective ground

Transducer connector –array 0,2 ohm

Protective ground

Single module connector (metal ring on female connector) 0,2 ohm

Protective ground

MCM connector (metal ring on female connector) 0,2 ohm

Protective ground

(type BF)

ECG (metal part behind the ring on AUX

connector below the ECG connector) 0,2 ohm

Earth Leakage

Normal 500 μA

Earth Leakage

Reverse 500 μA

Earth Leakage

Normal - Single Fault Condition 1000 μA

Earth Leakage

Reverse - Single Fault Condition 1000 μA

2.2 kV (DC) Cassette No Flash over

2.2 kV (DC) Monitor No Flash over

Signature

Page 18: Ecograf System Pro Focus 2202(BI2201-E)

2 of 2

The test complies with IEC 60601-1 regulations for medical equipment, safety class I, Type B (ECG type BF). Important: High Voltage testers deliver hazardous currents. Therefore these testers should only be operated by technicians who are aware of the hazards involved. Disconnect transducers and accessories from the equipment under test. Test limits are valid only for the equipment itself without any applied part(s) connected. The complete test must always follow immediately after a repair made on the 2202 Ultrasound Scanner, and always on the fully assembled unit. General remarks and settings for Protective GND test: Cables: Mains cable for scanner; clip/probe lead (Tester accessories) Tester settings: Class=I, Type: B (For ECG test type BF) Connect the scanners Main cable from IUT POWER (on Tester) to power inlet on the scanner. Connect the clip/probe lead to the test point and keep it firmly pressed against it as the current is 25 Amps when applying this test General remarks and settings for Leakage Current test: Cables: Mains cable for scanner; Tester settings: Class = I, Type: B, Connect the mains cable for the scanner to the IUT POWER outlet (on Tester) to power inlet on the scanner When activating the Leakage test then press the power button on the scanner and wait for it to start up and then record the Current before closing down the scanner as normal turn off. High Voltage Test: General remarks and settings for A and B Tester: HV Insulation Tester Cables: HV probe Adaptors(optional): HV test plug for mains WJ0246 can be used if the HV Tester cannot connect directly to scanner/monitor mains input Test Cassette: Connect HV tester plug for mains into socket for the scanner mains input. Insert tip of HV probe to the metal plate

at the right transducer connector slot. Start the test increasing the HV to 2.2 kV (DC) and apply this voltage for maximum 5 sec. There must be no flash over nor breakdowns.

Test Monitor: Connect the monitor power cable to the HV tester plug for mains. Insert tip of HV probe to screw next to monitor

power inlet. Start the test increasing the HV to 2.2 kV (DC) and apply this voltage for maximum 5 sec. There must be no flash over nor breakdowns.

Page 19: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Service Notes

BI2201-E 2-1

Section 2

SERVICE NOTES Service Notes

To improve the performance of the Pro Focus Ultrasound System 2202, small changes in the hardware and firmware may be made. All information about these changes, and any other change or correction to this Service Manual will be released in the form of Service Information.

Note

All existing and future Service Information concerning the Pro Focus 2202 system should be placed in this section.

Edition Number

If any major changes are made to the Pro Focus 2202 system and/or the Service Manual, a new edition will be released. Any new editions can be recognized by the edition number given in the footer.

Page 20: Ecograf System Pro Focus 2202(BI2201-E)

Service Notes System Pro Focus 2202

2-2 BI2201-E

This page is left blank intentionally

Page 21: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Checking Procedure

BI2201-E 3-1

Section 3

CHECKING PROCEDURE CONTENTS PAGE

3.1. Introduction .................................................................................................................................................. 2

3.1.1. Necessary Equipment to Perform the Checking Procedure ...............................................................2

3.2. Basic Checks................................................................................................................................................ 2

3.2.1. Visual Inspection..................................................................................................................................2

3.2.2. Monitor ................................................................................................................................................2

3.2.3. Before Power Up.................................................................................................................................3

3.2.4. Power Up.............................................................................................................................................3

3.3. Functionality................................................................................................................................................. 3

3.3.1. Monitor Checking ................................................................................................................................3

3.3.2. Transducer Inputs ...............................................................................................................................4

3.3.3. Track Ball ............................................................................................................................................4

3.3.4. Gain/TGC ............................................................................................................................................4

3.3.5. Size .....................................................................................................................................................4

3.3.6. Focusing..............................................................................................................................................4

3.3.7. Frame Rate .........................................................................................................................................4

3.3.8. Image Processing ...............................................................................................................................5

3.3.9. Persistence..........................................................................................................................................5

3.3.10. Puncture ..............................................................................................................................................5

3.3.11. Measure ..............................................................................................................................................5

3.3.12. Image review.......................................................................................................................................5

3.3.13. M-mode ...............................................................................................................................................6

3.3.14. Split Screen .........................................................................................................................................6

3.3.15. Audio/Spectral Doppler .......................................................................................................................6

3.3.16. CFM.....................................................................................................................................................7

Page 22: Ecograf System Pro Focus 2202(BI2201-E)

Checking Procedure System Pro Focus 2202

3-2 BI2201-E

3.1. Introduction

The Checking Procedure is to verify proper operation of the main functions of the 2202 System. The procedure should be used during installation, incoming inspection, regular maintenance and before and after repair. The procedure is arranged in sections and must be followed from the beginning when performed.

Notes If the Checking Procedure cannot be performed successfully or if there is a specific fault in the System, refer to Section 4, TROUBLESHOOTING or Section 5, ADJUSTMENT PROCEDURE. The Checking Procedure does not include a check of the different transducer types. If in doubt about the functions of the 2202 you should consult the user guide.

Important!

Any inside repairs performed on the 2202 must be followed by an Electrical Safety Test.

3.1.1. Necessary Equipment to Perform the Checking Procedure

• Transducer Phantom, model 254 (WQ 0973) for checking a 2400 system

equipped with Array- and Single-element Transducers. • Transducer Phantom, type 251 (WQ 0972) for checking a 2400 system

equipped with the Endosonic Probe 1850.

• Ruler (for checking the size of the image on the monitor)

• CD-R (recordable) (for testing the CD writer) • Array Transducer (Linear or Convex). • Sector Transducer (for systems equipped with Single-element module)

Note: If a phantom is not available the human body is a good alternative.

3.2. Basic Checks

3.2.1. Visual Inspection

Check the overall appearance of the 2202.

Especially check for scratches and stain's on the paint.

3.2.2. Monitor

Check that the monitor is securely fastened and is easy to tilt/rotate.

Page 23: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Checking Procedure

BI2201-E 3-3

3.2.3. Before Power Up

The 2202 System complies with the IEC 60601-1 safety regulations and as such it must be connected to a mains outlet having a safety ground connection. Therefore: Check that the 2202 mains cable is terminated with a three-pole plug that fits a three-pole mains outlet. For safety reasons do not use extension cables!

3.2.4. Power Up

1. Connect a transducer to the scanner.

2. Switch On the 2202. Follow the progress bar in the start-up screen. (It is normal that the start up will take approx. 2 min and 40 sec.)

3. Verify that the monitor opening layouts are shown in Fig. 3.3 -1. Note that the menu shown (settings) differs from transducer type.

Fig. 3.3-1. The default display.

4. Check that the current date and time is displayed on the monitor:

3.3. Functionality The chapter will cover some of the basic functions in the scanner. You can read more about the user interface in the user's guide, as this is not the scope of this document.

3.3.1. Monitor Checking 1. Press ALT+SHIFT+G 2. Press Enter to switch to next image.

3. If you have problems meeting the above requirements, consult chapter 5,

Adjustment Procedure.

Page 24: Ecograf System Pro Focus 2202(BI2201-E)

Checking Procedure System Pro Focus 2202

3-4 BI2201-E

3.3.2. Transducer Inputs

1. Move the transducer connector to one of the other input module(s), Press Freeze and confirm that the transducer can be identified.

2. Start the transducer to verify correct operation of the input module. Note: If available connect a Single-element transducer to the Single element socket

and verify correct operation of the transducer.

3.3.3. Track Ball The track ball should move smoothly. If it too tight or too loose you can adjust it using the adjusting tool QA0228 (shipped with System 2202).

3.3.4. Gain/TGC 1. Press Freeze to start scanning.

2. Adjust the gain using the gain potentiometer and observe the effects on the

B-image.

3. Adjust the TGC curve using the TGC Potentiometers and observe the effect on the B-image.

3.3.5. Size Verify the function of the -/+ button (above the track ball). Press the left side of the button (-) to decrease the size of the objects (higher scale units) and the right side of the button (+) to enlarge the size of the objects (lower scale units).

3.3.6. Focusing Note: This section is only valid for array transducers.

1. Check that there is a focus zone mark on the left side of the ultrasound picture.

2. There are two ways to use the focus functionality. You might check them both.

3. Press Focus and move the track ball up and down and check that the arrow

moves accordingly on the monitor screen. Also check that max. lateral (horizontal) resolution is at the position of the arrow - use the phantom, model 254, to check the focusing. The size of the focus zone can be regulated using the -/+ button.

4. Using the track ball, click on the middle of the focus zone mark. Now using the track ball you can move the focus zone. If you click on the top or bottom piece of the focus zone, you can regulate the size of the focus zone using the track ball.

5. Press Focus again to get back.

3.3.7. Frame Rate 1. Connect a mechanical transducer and press the Res/Hz Up/Down and

observe that the speed of motor change correspondingly.

2. Connect a Array transducer, press the Res/Hz Up/Down and observe that increased frame rate narrows the scanning field.

Page 25: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Checking Procedure

BI2201-E 3-5

3.3.8. Image Processing Use phantom, model 254, to check the following: 1. Flip Dyn Range Up/down and observe the contrast effect on the image.

2. Using the tracker ball, click in the Grey Scale bar in the right top of the US

image. Step trough the different gray scales. Observe the effect on the image.

3.3.9. Persistence

1. If …More is not open, open it clicking on it with the track ball.

2. Roll the track ball arrow over the Persistence.

3. Press Select to see all valid values. Click on desired value to select it. Check the effect on the image.

4. Roll the track ball arrow over the Persistence again.

5. Press the -/+ button to select the desired value. Check the effect on the image.

Note: 0 is no averaging and 8 is max averaging.

3.3.10. Puncture

Press Biopsy to select the default puncture guide for the selected transducer, if applicable, and check that puncture dots appear in the B-image field.

3.3.11. Measure

1. Press Freeze, if not frozen already.

2. If Measurements is not open, click on Measurements or press Measure.

3. You should now be able to test: • Distance • Circumference • Mean Diameter • Area • Volume • Height • Width • Length • H*W*L

4. In some of the choices, you can see an arrow in the right side. By clicking on it, you may find more valuable measurement types here.

3.3.12. Image review

1. Make a scan and press Freeze.

2. Observe that a horizontal Track ball movement scrolls through the most recently recorded B mode images.

Page 26: Ecograf System Pro Focus 2202(BI2201-E)

Checking Procedure System Pro Focus 2202

3-6 BI2201-E

3.3.13. M-mode

Note: This section is only valid for Array Transducers. 1. Press the M-mode potentiometer

2. Start the transducer and check that the echoes now are displayed both as a

normal image and as horizontal lines that reflect the echo information along the M-mode line.

3.3.14. Split Screen

1. Press Split to activate Split Screen.

2. Click on the left side to make a scan "A"

3. Press Split and repeat for "B" Note: Pressing Split will switch between scan "A" and scan "B". Holding Split

pressed for a while will switch of the Split Screen mode

3.3.15. Audio/Spectral Doppler

Note: This section is only valid for Array transducers. 1. Press on the Doppler potentiometer and place the sample volume over a

vessel.

2. Observe that sound can be heard in the speaker and a Doppler spectrum is shown.

3. Audio Volume - Press Volume up/down and check that the volume of the audio signal can be adjusted.

4. Gain - use the smaller topmost Gain Control and adjust the gain up and down. Observe that spectrum noise and sound changes.

5. Filter changes the setting of the wall filter. Observe that the low frequency components close to the base-line disappear.

6. Range – press Dyn Range up/down and observe that the resolution of the spectrum decreases when the range is increased.

7. Baseline - press Baseline up/down and check that the spectrum is offset.

8. Doppler Sample Volume - Press D/M Cursor to change the size of the sample volume, using the trackball.

9. Doppler Angle – press Angle up/down and adjust the angle of the sample volume with the trackball. Check that the scale is changed.

10. Steered Doppler –press Steer up/down to angle the Doppler beam. Press steered to toggle between left angling, right angling and no angling.

Note: Steered Doppler cannot be selected, if not supported by the transducer.

Page 27: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Checking Procedure

BI2201-E 3-7

3.3.16. CFM

1. Press the Colour potentiometer.

2. Unfreeze and place the transducer over a vessel. Move the colour box using the Track ball and verify flow in the colour box.

3. Gain - Increase CFM gain using the smaller topmost Gain Control and observe that the colour box is filled with coloured noise dots.

4. Base line - move the base line up and down with the base line control. Check that the colour scales is offset allowing measuring a larger range of velocities in one direction.

5. Colour resolution- press the resolution control Scale up/down. Increase the resolution and check that the number of lines in the CFM image is increased and frame rate is decreased.

Page 28: Ecograf System Pro Focus 2202(BI2201-E)

Checking Procedure System Pro Focus 2202

3-8 BI2201-E

This page is left blank intentionally

Page 29: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-1

Section 4

TROUBLESHOOTING

CONTENTS PAGE

4.1. Introduction ................................................................................................................................... 3 4.2. Replaceable Parts......................................................................................................................... 3 4.3. Keys to Troubleshoot Specific Problems (Hints)........................................................................... 4

4.3.1. Loss of time.............................................................................................................................. 4 4.3.2. Transducer or scanner............................................................................................................. 4

4.4. Service Mode ................................................................................................................................ 5 4.4.1. Enter Service Mode ................................................................................................................. 5 4.4.2. How to set scanner to Normal (console) Mode - exit Service Mode........................................ 6 4.4.3. How to start console application for service use...................................................................... 7 4.4.4. How to start Windows Explorer................................................................................................ 7 4.4.5. How to start the Control Panel ................................................................................................. 7 4.4.6. How to start the Device Manager............................................................................................. 7 4.4.7. How to shutdown the scanner.................................................................................................. 7 4.4.8. How to restart the scanner....................................................................................................... 7 4.4.9. How to delete patient-related files............................................................................................ 7 4.4.10. Update Hardware..................................................................................................................... 8 4.4.11. Event Viewer .......................................................................................................................... 11 4.4.12. Scandisk................................................................................................................................. 12 4.4.13. Disk Fragment........................................................................................................................ 13 4.4.14. Keyboard Test........................................................................................................................ 14

4.5. Installing a XP Embedded Image................................................................................................ 16 4.5.1. Preparing a clean hard disk for image installation................................................................. 16

4.6. HIPAA (Health Insurance Portability and Accountability Act) ..................................................... 16 4.6.1. Background............................................................................................................................ 16 4.6.2. Enable patient privacy............................................................................................................ 17 4.6.3. Create users........................................................................................................................... 17

4.7. Table of passwords ..................................................................................................................... 18 4.8. Dicom .......................................................................................................................................... 19

4.8.1. Host Setup ............................................................................................................................. 20 4.8.2. Worklist Setup........................................................................................................................ 23 4.8.3. PACS Setup........................................................................................................................... 24 4.8.4. Printers Setup ........................................................................................................................ 25 4.8.5. Troubleshooting ..................................................................................................................... 26

4.8.5.1. Testing errors.................................................................................................................... 26 4.8.5.2. Status................................................................................................................................ 27 4.8.5.3. Ping................................................................................................................................... 28

4.9. How to recover from a crashed main database .......................................................................... 29 4.10. How to recover patient database from a crashed system........................................................... 29

4.10.1. Patient Database Recovery tool ............................................................................................ 29 4.11. Log files ....................................................................................................................................... 30

4.11.1. How to grab the log files ........................................................................................................ 30 4.12. How to send log files via email.................................................................................................... 31

4.12.1. Initialization ............................................................................................................................ 31 4.12.2. Verification (or how to use the feature).................................................................................. 32 4.12.3. Troubleshooting ..................................................................................................................... 32

4.13. How to configure scanner to use an external hard disk to save scans instead of the internal hard disk .............................................................................................................................................. 33

4.13.1. How to configure scanner to use a network drive.................................................................. 35 4.13.1.1. Configure Microsoft Windows-based PC.......................................................................... 35 4.13.1.2. Configure Network ............................................................................................................ 36 4.13.1.3. Configure Scanner ............................................................................................................ 36

4.13.2. How to transfer file from one hard disk to another................................................................. 37 4.13.3. Using IDE (44 pins) to USB adapter ...................................................................................... 37

Page 30: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-2 BI2201-E

4.13.4. Using dual connector IDE (44 pins) Cable............................................................................. 38 4.14. Software Structure ...................................................................................................................... 39

4.14.1. Directory Structure of harddisk (C-drive) ............................................................................... 39 4.14.1.1. CasaEngine ...................................................................................................................... 39 4.14.1.2. Console............................................................................................................................. 39 4.14.1.3. XPE_Backup..................................................................................................................... 41

4.15. How to access the content of the CD in the CD-drive from Windows Explorer in service mode 41 4.16. There is no Recordable CD Drive ............................................................................................... 41 4.17. How to make your own BIOS plug EL1011................................................................................. 42

4.17.1. Tools Required....................................................................................................................... 42 4.17.2. Assembly Procedure.............................................................................................................. 42

4.18. License Codes ............................................................................................................................ 43 4.18.1. Install License Codes from the Back-up CD.......................................................................... 43

4.19. Power Supply Trouble shooting .................................................................................................. 44 4.19.1. Introduction ............................................................................................................................ 44 4.19.2. LED indicators on the Power Supply ..................................................................................... 44 4.19.3. Steps to test the Power Supply.............................................................................................. 45 4.19.4. How to test the Power Supply ZG0346 as standalone .......................................................... 45

Page 31: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-3

4.1. Introduction This section is divided into separate subsections some of which cover specific areas within the troubleshooting others describe the overall test facilities in the 2202. But before you jump to any of the subsections you should first consult Section 4.3 which might give you, if not the solution, then some ideas of how to troubleshoot this particular problem of yours. To prevent injury to persons and damage to the Pro Focus be aware of the following: WARNINGs: * Opening the Power Supply can expose live parts. * Any work done on the open instrument with power On must only be done by B-K Medical or their authorized representatives, who are aware of the hazards involved. * Any repair work done on the 2202 system must be followed by an electrical safety test to verify a continuous safe operation of the system. * The Core board contains a lithium battery. The battery must only be replaced by a person having special knowledge as described in sec. 4.2 Replaceable Parts CAUTIONs: * Switch Off the 2202 before connecting or disconnecting the Pro Focus to any peripheral units. Failure to do so could damage the equipment. * The PC Boards in the 2202 are sensitive for static discharge. Therefore, when handling PC Boards always take steps to prevent static discharge - see Section 1.5, Special Tools and Equipment.

4.2. Replaceable Parts Apart from a few exceptions listed below all PC Boards and modules in can be replaced without further actions made. Any work that requires opening the 2202 must however be followed by an electrical safety test. Before replacing a subassembly it is necessary to verify that the ID (revision) of the new part is compatible with the ID of the defective part. Item to be replaced: Type of unit

effected: Necessary action after replacement:

QB0041 Lithium battery All Set system clock, PIN codes

ZN2220 ZN2230 ZN2231

Core Board Assembly / All Set system clock Restore Patient database User settings PIN codes (if new MAC address)

ZD0784 ZD0795

ETX Module All Set system clock + PIN codes (if new MAC address)

UL0041

Hard disk All Restore Patient database, User settings, PIN codes

ZD0757 Back Plane From 1864900 PIN codes

Page 32: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-4 BI2201-E

4.3. Keys to Troubleshoot Specific Problems (Hints)

4.3.1. Loss of time If you remove the ETX board from the Core board, the time settings will be lost. The same is valid, if you remove the lithium battery on the Core board. Check the lithium battery on the Core board (also check for bad contact).

4.3.2. Transducer or scanner • First of all: Try another transducer! • Try the other Array Input Connector • If possible, connect a single-element transducer to bypass the FE64+ board and the beam

former board. Radial or Vertical lines The problem is most likely caused by the FE64+, Digital beam former, or the transducer.

Concentric rings or Horizontal bands Noise pattern in the B-image A noise pattern in the far field is often caused by external noise entering the scanner trough the air or via the mains cable. To isolate the problem check/try the following: • Connect the scanner to another mains outlet, preferably in another room. A change in the

noise pattern indicates noise on the mains. • Check that the wall outlet has a proper grounding. • Compare different transducer types and frequencies (MFI). External noise is often

frequency specific. Note: Before contacting the support group in Denmark for help with noise problems it is recommended that you try out the suggestions above. Use the image storage facility of the 2202 to save a couple of images showing the problems. The images can be attached to an e-mail.

Page 33: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-5

4.4. Service Mode The operative system of the 2202 is Windows XP Embedded. Windows XP Embedded is a componentized version of Windows XP Professional that contains all of the features, functionality, and familiarity of Windows XP Professional. In our case this means that all not needed software, device drivers and other files are removed. This also means that support for non-generic USB storage keys, USB hard disks and USB CD/DVD drive is non-existing. The way Windows XP Embedded is build makes is impossible to install a specific driver after the scanner has left the factory. In this new environment, the usual Windows Desktop has been replaced by the B-K Medical Ultrasound application: Console. This way there is no Windows Desktop behind the application. This is very good when we want to prevent unauthorized access. When we want to get behind the console application, so we have access to Windows, we must switch to Service Mode. In service mode the console-based desktop is replaced by a service menu. Service mode can be activated if you want to: • Upgrade software • Check of Hard Disk • Access Windows Explorer • and other service related issues

4.4.1. Enter Service Mode 1. Start up the scanner and wait for complete boot-up

2. Press ALT + F4 (if you don’t see the below dialog then first click anywhere at the Ultrasound

image field and the press ALT + F4)

3. Enter password: viking

and the scanner will then turn OFF

Page 34: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-6 BI2201-E

4. Turn on the scanner and wait for it to boot up in service mode.

4.4.2. How to set scanner to Normal (console) Mode - exit Service Mode 1. Double click on the “Start in Console Mode” icon on the desktop.

2. Click on Yes

3. Click on OK 4. Double click on the “Reboot Scanner” icon on the desktop.

Page 35: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-7

4.4.3. How to start console application for service use In many cases it is desired to run the console application from service mode to test different issues. Double click on “Start Console” icon on desktop When you run the console application in service mode, you can close it using ALT+F4, and you can switch to other windows/applications using the ALT+TAB key combination.

4.4.4. How to start Windows Explorer Double click on “explorer” icon on desktop The Window Explorer is a very powerful tool, as you can get access to almost everything in Windows using this tool.

4.4.5. How to start the Control Panel Double click on “Control Panel” icon on desktop The control panel can also be started from the Windows Explorer.

4.4.6. How to start the Device Manager Double click on “Device Manager” icon on desktop The device manager can also be started from the Windows Explorer.

4.4.7. How to shutdown the scanner Double click on “Shutdown scanner” icon on desktop

4.4.8. How to restart the scanner Double click on “Reboot scanner” icon on desktop

4.4.9. How to delete patient-related files Double click on “Delete Console Documentation Directory” icon on desktop Doing this all saved images/clips will be deleted as well as the patient database will be cleared.

Page 36: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-8 BI2201-E

4.4.10. Update Hardware

The Update Hardware can reprogram the flash memory located on the Core board. Usually this feature is used in connection with software upgrade where e.g. improved functionality can be added to the board as part of the software upgrade. Warning!! If programming the Flash fails then it must be reprogrammed until it succeed otherwise the console application will not work and if you set the scanner to normal mode you will not be able to go back into service mode without reinstalling the hole software (image install). If you have a system with Coreboard ZD0773 version 4 we recommend not to run the Update Hardware!

1. Click on Update Hardware

2. Click on Report Status

3. If the status field contains any “needs update” then click on Program otherwise quit the Coreboard Flash Utility

Needs update

Page 37: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-9

Note! The screen will be blank for several minutes during this programming - this is quite normal you just have to wait!

4. Wait for the Core board Flash Utility to finish programming i.e. until Working field is no longer blue and the last lines in the Output/Progress window is :

Updating Flash Table in AVR…

Flash Table is updated

5. When the Coreboard Flash Utility has finished programming you should check for error messages in the Output/Progress field. Errors will have been marked out by a line of “****************” - in that case you have to programme again. Note that you must scroll up to see all text from the Output/Progress

Page 38: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-10 BI2201-E

If an error has occurred then it will look like the below example in that case you should click on Clear List and then Program as described in step 3.

Failed!!!

6. When the programming has not failed then click on Quit

7. Click on Shutdown Scanner

conds and then turn on the scanner again. This will complete the Update Hardware function

8. After it has turned off wait 30 se

Page 39: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-11

4.4.11. Event Viewer Event Viewer is the first place to look in for information when you are troubleshooting any problem with your computer. It is used to view events and manage them. There are three categories of events: System, Application and Security events. The System Log records events related to Windows XP system components such as failure of a driver or other system components to load during start up. Usually at start up there is a warning message indicating that a driver or other system component failed, check Event Viewer for more details. The Application Log records events logged by applications, such as errors generated by certain applications. The Security Log records events related to system security. This helps track changes or attempts to break system security.

1. Left click on Event Viewer and select a log to view. The log will appear in the right hand

pane.

2. If you want to look at an event, double click on it. A dialog box will open showing information about what caused the error.

3. You can navigate from error to error in the same log by clicking the up and down arrows.

4. If you need to record an error so you can email it to a technical support person, click on the

button right below the down arrow. This will copy the error to the clipboard. It is now available to be pasted into a text file, you can send to Service at B-K Medical.

5. Close the dialog box when you are done reviewing the event log. Event Viewer displays these types of events:

Page 40: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-12 BI2201-E

Error: A significant problem, such as loss of data or loss of functionality. For example, if a service fails to load during start-up, an error will be logged. Warning: An event that is not necessarily significant, but may indicate a possible future problem. For example, when disk space is low, a warning will be logged. Information: An event that describes the successful operation of an application, driver, or service. For example, when a network driver loads successfully, an Information event will be logged. Success Audit: An audited security access attempt that succeeds. For example, a user's successful attempt to log on to the system will be logged as a Success Audit event. Failure Audit: An audited security access attempt that fails. For example, if a user tries to access a network drive and fails, the attempt will be logged as a Failure Audit event. Note: Event Viewer is a part of Microsoft Windows XP. Further help can be gained using the Event Viewers built-in help.

4.4.12. Scandisk Scandisk is the program that verifies the logical integrity of a file system on Windows. When Scandisk encounters logical inconsistencies it does not take actions to repair file system data, but let you know that there is problem: Errors found. CHKDSK cannot continue in read-only mode. 1. The scandisk will run automatically in a command prompt box. You are not supposed to

enter anything. When the scandisk has completed, you see this output.

2. Press any key to continue 3. Click on Yes or No depending on the outcome of the test.

C:\CasaEngine>C:\WINDOWS\system32\Chkdsk.exe The type of the file system is NTFS. Volume label is 2202 XP Embedded. WARNING! F parameter not specified. Running CHKDSK in read-only mode. CHKDSK is verifying files (stage 1 of 3)... File verification completed. CHKDSK is verifying indexes (stage 2 of 3)... Index verification completed. CHKDSK is verifying security descriptors (stage 3 of 3)... Security descriptor verification completed. 39070048 KB total disk space. 2922396 KB in 3790 files. 1060 KB in 347 indexes. 0 KB in bad sectors. 71324 KB in use by the system. 65536 KB occupied by the log file. 36075268 KB available on disk. 4096 bytes in each allocation unit. 9767512 total allocation units on disk. 9018817 allocation units available on disk. C:\CasaEngine>Pause Press any key to continue . . .

Page 41: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-13

4.4.13. Disk Fragment Analyzing and defragmenting disk volumes can help to preserve the performance and general operation of the system. File fragmentation can negatively affect operating system speed and performance. To maintain peak performance when using Windows XP, the condition of your disks should be analyzed on a regular basis - preferably once a year - and defragmentation performed as needed. Note: Prior to running Defrag, you should run a Scandisk to correct any drive errors. Note: Please don’t touch anything running this test, as it may disturb the process. Because defragmenting a disk volume can take a long time (depending on the size of the volume, the number of files, the percentage of fragmentation, and the availability of system resources), you should analyze volumes before defragmenting them, to decide whether or not it is worthwhile to take the time to run the defragmentation process.

How to Analyze a Disk Volume To check for fragmented files and folders on a volume: 1. Click the volume that you want to analyze. 2. Click Analyze to begin the analysis. 3. Review the results of the analysis after it is complete by clicking View Report. If the analysis

tool recommends that the volume be defragmented, follow the steps in the next section. How to Defragment a Disk Volume 1. Click the volume that you want to defragment. 2. Click Defragment to begin the operation.

3. Review the progress of the operation in the Defragmentation Display window. Fragmented

files on the disk appear in red, contiguous files are blue, and system files are green. The goal is to eliminate most of the red in this window.

Page 42: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-14 BI2201-E

4.4.14. Keyboard Test 1. Double click on “Start Keyboard Tester” icon on desktop

1. Setup: Select Comport: COM4 and press Reset. If you test everything, you should be able to get 0 remaining in the field in the upper left corner. How to test the Sound Test method: Click on the various buttons one by one. Expected Result: You should hear different sounds. How to test the Legend LEDs Test Method: Click on Legend LEDs. Expected Result: You should see light in all legend LEDs. The light should be uniform, i.e.

nearly same intensity in all LEDs. (You can alter the intensity using the slider below the button)

How to test the Indicator LEDs Test Method: Click on Indicator LEDs. Expected Result: You should see light in all 28 Indicator LEDs. The light should be uniform,

i.e. nearly same intensity in all LEDs. (You can alter the intensity using the slider below the button)

How to do the Ping test Test Method: Click on Ping 1 sec. Expected Result: You should see the ping result (d1) in the Received Key Codes field each

second. When done, click on Stop Ping. How to test the Tracker ball Test Method: Turn tracker Ball up/down and left/right. Expected Result: Check that the cursor (arrow) follows tracker ball.

Page 43: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-15

How to test the PC Keyboard Test Method: Place cursor in the ASCII Keyboard Test field and 'left click. Press on the

desired ASCII Keyboard button to test. Expected Result: Check character in the ASCII Keyboard Test field. How to test the User Buttons Test Method: Press any button. Expected Result: Check that there is a mark in the field belonging, and that the value in the

upper left corner has decreased by one. Test Method: Move the TGC to the right side. Expected Result: Check the value in the field belonging. Max value = 127 Test Method: Move the TGC to the left side. Expected Result: Check the value in the field belonging. Min value = -127 Test Method: Turn any knob clockwise. Expected Result: Check that the value in the field belonging is a positive number Test Method: Turn any knob counterclockwise. Expected Result: Check that the value in the field belonging is a negative number Test Method: Press each of the three foot switches. Expected Result: Check that there is a mark in the field belonging.

Page 44: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-16 BI2201-E

4.5. Installing a XP Embedded Image. With each software release a new installation procedure can be downloaded from the service webpage.

4.5.1. Preparing a clean hard disk for image installation A clean / brand new hard disk needs to be prepared to use the B-K Medical image installation CD. Basically what we need to do is to create an active partition on the hard disk. 1. Boot from B-K Medical image installation CD

2. When dialog window comes, click "Cancel" to come to the DOS-prompt.

3. Write "DISKPART" and press enter (without ")

4. In Diskpart you should enter these commands:

5. select disk 0

6. clean

7. create partition primary

8. select partition 1

9. active

10. assign letter=C

11. exit

12. Start "ImageUnzip" again

13. choose "clean installation"

14. remove the checkmark in "Quick Format"

15. Now follow rest of the procedure

4.6. HIPAA (Health Insurance Portability and Accountability Act)

4.6.1. Background The American Health Insurance Portability and Accountability Act of 1996 (HIPAA) requires security and privacy of health information. The Privacy Rule is a US federal regulation defining administrative steps, policies, and procedures to safeguard individuals' personal, private health information (protected health information or PHI). The Privacy Rule is designed to empower patients by guaranteeing them access to their medical records, giving them more control over how their PHI is used and disclosed, and providing a clear avenue of recourse if their medical privacy is compromised. The rule is designed to protect medical records and other personal health information maintained by certain health care providers, hospitals, health plans, health insurers and health care clearinghouses. The rule defines "protected health information" as health information that 1. identifies an individual and 2. is maintained or exchanged electronically or in hard copy.

Page 45: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-17

If the information has any components that could be used to identify a person, it would be protected. The protection would stay with the information as long as the information is in the hands of a covered entity or a business associate. The protections apply to individually identifiable information in any form, electronic or non-electronic. The paper progeny of electronic information is covered (i.e. the information would not lose its protections simply because it is printed out of a computer), and oral communications are also covered.

4.6.2. Enable patient privacy 1. Start the scanner 2. Click on Setup 3. Click on the System-tab 4. Click on the Password-tab 5. Mark Password Protection Enabled

6. Click on Close.

4.6.3. Create users You have enabled password protection, but you must now create users to allow access. 1. Click on Login 2. Username is admininstrator 3. Password is superuser (This is the default password from the factory. If you reset it, the

password is blank, and you may have set it to something special.) 4. Click on OK 7. Click on Setup 8. Click on the System-tab

Page 46: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-18 BI2201-E

users, reset the password for the existing user and enable/disable

assword protection.

4.7. Table of passwords

9. Click on the Password-tab

Here you can to the user administration. Only the administrator user have rights to add newusers, remove existingp

Purpose Username Password Reference Service Mode - viking 4.4

BIOS (Installing a new HD image) - BKASTRA 4.5

Password Protection / HIPAA dministrator uperuser 4.6.2 a

s

Network Configuration - nw 4.8.1

Page 47: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-19

4.8. Dicom DICOM is a global Information Technology standard that is used in virtually all hospitals worldwide. Its current structure, which was developed in 1993, is designed to ensure the interoperability of systems used to: • Produce • Store • Display • Process • Send • Retrieve • Query • Print Medical images and derived structured documents as well as to manage related workflow. To be able to use DICOM features on 2202, make sure that the DICOM license has been entered. Then second try to get the Network Administrator from the hospital to fill out the "DICOM network description" questionnaire. In the following chapters we'll cover the DICOM features on 2202 and how to setup the scanner to work in a DICOM network.

Page 48: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-20 BI2201-E

4.8.1. Host Setup

This is setup of the scanner. 1. Click on the DICOM tab

2. Enter the AE Title for the scanner and then click on Apply.

3. Click on Configure Network

4. Password is nw

Page 49: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-21

5. Click on OK. This will open this window:

6. Right click on Local Area Connection and choose Properties

7. Select Internet Protocol (TCP/IP) and click on Properties.

Page 50: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-22 BI2201-E

8. Default setting is that the scanner will obtain an IP address from the DHCP server in the network. If you want to use a fixed IP address, click on Use the following IP address and fill out the fields for IP address, subnet mask and default gateway according to the data collected from hospital.

9. Click OK until all windows are closed.

10. Finally close the Network Connections window.

Page 51: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-23

4.8.2. Worklist Setup Note! Scanner AE Title must be present before testing can be done.

1. Click on Configure next to Worklist field.

2. Enter Name. Note, this can not be changed later on. Enter AE Title, IP Address and Port

Number. Select the desired value in the search filter and then click on Test. Search filter can have these values: This system only: The AE Title of the scanner will then be used to differentiate Ultrasound Systems only:

(0008,0060)-tag must be US in the record in the worklist

All Systems: No filtering

3. Click on OK

4. Click on OK

Page 52: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-24 BI2201-E

4.8.3. PACS Setup Note! Scanner AE Title must be present before testing can be done.

1. Click on Add next to PACS field.

2. Enter Name. Note, this can not be changed later on. Enter AE Title, IP Address and Port Number. Select the desired archiving method and then click on Test. Each time a new PACS server is added, it will be default. In case you want to modify that, change the Default PACS setting.

Archiving Methods are: Manual: You have send the images manually. Send complete examination at end of examination:

If you press and hold the patient id button for a couple of seconds, the current examination will end and all images and clips (except 3D) will be sent. If you start on a new patient, the send behavior is the same.

3. Click on OK 4. Click on OK

Page 53: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-25

4.8.4. Printers Setup Note! Scanner AE Title must be present before testing can be done.

1. Click on Configure next to Printers field.

2. Enter Name. Note, this can not be changed later on. Enter AE Title, IP Address and Port Number. Select Printer Type and click on Test. Each time a new Printer is added, it will be default. In case you want to modify that, change the Default Printer setting. Other settings can be modified on request.

3. Click on OK

4. Click on OK

Page 54: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-26 BI2201-E

4.8.5. Troubleshooting

4.8.5.1. Testing errors When you click on test following will happen: 1. scanner will ping target host. 2. scanner will send out a Dicom echo to target host If the first part fails, you may get a message like this:

Corrective actions can be: Check that the cable between scanner and network is OK Check scanners IP address, netmask and default gateway. Check the target host IP address.

Check that the target host will reply on a ping. Talk with hospital IT.

When pinging is done successfully, the Dicom echo will be send towards the same target host. If the AE title of the scanner has not been set, you may get this message: The solve this, enter an AE title on the scanner and click Apply and reboot scanner and then try again. It’s important that scanner is rebooted. If AE title is set and the Dicom echo still fails, you may get a message like this

Corrective actions can be: Check target host AE Title and Port Number on scanner Check that the target host will reply on an echo. Talk with hospital IT. Check that the target host has granted access to scanner. Talk with hospital IT.

Page 55: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-27

4.8.5.2. Status .8.5.2. Status When Dicom is enabled and configured, you will notice this: When Dicom is enabled and configured, you will notice this: Green LED Green LED Everything is OK. Transferal of images was OK. The “button” will disappear. Everything is OK. Transferal of images was OK. The “button” will disappear. Yellow LED Transfer is going on Red LED Transfer failed. The “button” will stay.

If you click on Dicom…, you get a pop up with Dicom Status. You can also click on Customize, Dicom and then Status.

This will bring up the same pop up with Dicom Status:

In left you see the device list. These are the Dicom servers/printers the scanner has tried to send something to. You also see the status indicated by the colored LED. If you select one of the items, you can see more details in right area. These are the documents that are either being send or could not be sent. Clicking on Cancel will remove the document from the list and thus stop sending of this document. Clicking on Cancel All will clear the list. In case you regret, click on No otherwise click on Yes to delete the item. You also have the option to test each of the Dicom servers/printers in the device list. Clicking on Details can give an indication on what went wrong.

Page 56: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-28 BI2201-E

4.8.5.3. Ping Click on Customize, Dicom and then Ping…

Enter a target IP address and click on OK. In case you are running with a DNS in the network, you can also try a name. If you have success with pinging the target, you will get this message:

A failure looks like this:

Page 57: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-29

4.9. How to recover from a crashed main database Console software saves different kind of information in different databases. The main database (2202.mdb) is located in the c:\console\database directory. Databases can crash independently. This mean that you can have a working patient database while you main database is corrupt, so it might be helpful if you start with one database at a time. If you start a scanner and you get an error message with either the word “database” or “jet”, the main database has crashed. Examine the c:\console\database directory in service mode and if you find lock-file 2202.ldb, you must delete it. You can now recover the scanner by copying an empty database from c:\console\database\org to c:\console\database. All settings are now lost, except DICOM settings. Note: If you are able to start scanner and are able to make a “restore to factory” then the effect would be the same as the above.

4.10. How to recover patient database from a crashed system The patient database (PatientDB.mdb) is located in the c:\console\documentation directory. If you experience problems with the archiving system, it could be a corrupt patient database. You can recover the scanner by copying an empty database from c:\console\database to c:\console\documentation. But by doing this all files in the c:\console\documentation directory can not be accessed from the console software any more. The solution is this case is to copy all files to CD’s or DVD’s, and then access them from the archiving system.

4.10.1. Patient Database Recovery tool There is a slight chance for you to recover some part of the patient database if you have a archive CD’s. On the latest archive CD, find the patient database. This patient database can be used to recover. But it must be converted first: 1. Bring scanner to service mode.

The console application MUST NOT be running.

2. Double Click on c:\Console\ PatientDbRecovery.exe.

3. Click on Start.

4. Click on OK.

5. Double click on Start Console on the desktop/or start desktop from the Scanner service menu.

6. Click on View Archive and see if you can find the patients in the record.

Only patients with scans on the hard disk will be found now. All patients you have archived and deleted from the hard disk will not be in the recovered patient database. Finally all scans after the time, the archive CD was created, must be copied to a CD/DVD, as they will not be reachable from the console software.

e hard disk will be found now. All patients you have archived and deleted from the hard disk will not be in the recovered patient database. Finally all scans after the time, the archive CD was created, must be copied to a CD/DVD, as they will not be reachable from the console software.

Page 58: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-30 BI2201-E

4.11. Log files The log files are a collection of files which are saved when clicking Export in the Version tab (Customize - System - Version)

4.11.1. How to grab the log files 1. Start scanner and wait for it to get ready. 2. Click Customize… 3. Click System 4. Click Versions 5. Insert USB Flash Key 6. Click Export (USB Device) 7. Enter fault description in the Comment / Description field and click Export. 8. A folder will be created on your USB Flash Key: Logs_<Date>_<Time>. In this

folder, a number of files will be copied. Attach all these files to a mail and send to [email protected].

Page 59: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-31

4.12. How to send log files via email As a new feature the Pro Focus is able to send log files via email to a pre-defined email address. This feature requires: 1. SMTP server available on the network 2. Permission to send email over the network Now, the requirements can only be fulfilled with help from the hospital IT department, as we would be using their SMTP server. Before you continue make sure that these requirements have been fulfilled.

4.12.1. Initialization 1. Bring scanner into service mode

2. Change to networks IP address & net masks according to hospital IT advice.

3. Start Explorer

4. Browse to c:\console.

5. Make a copy of c:\console\console.ini to c:\console\console.ini.org

6. Open c:\console\console.ini (in Notepad).

7. Scroll down to the [Behaviour]-section.

8. Find this line:

;SMTPToAddress=

9. Remove the ; sign and enter the address to receive the log-files here. Fx: [email protected] (This address is fictive.)

10. Scroll down to the [Settings]-section.

11. Find this line: SMTPServer=

12. Enter the name or IP address of the SMTP server here. Fx: SMTPServer=smtpserver.hospital.com (This address is fictive.) or SMTPServer=10.0.0.1 (This address is fictive.)

13. Save this file and exit Notepad.

14. Close Windows Explorer.

Page 60: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-32 BI2201-E

4.12.2. Verification (or how to use the feature) 1. Start Console

2. Click on Setup

3. Click on the System tab

4. Click on the Version tab

5. Click on the down arrow next to Export and choose Send Log Files as E-mail

6. Enter a comment in the field and click on Export. This comment will be written in the mail

and the log files with be attached to it. (If there is any log file of any kind it will be possible to select the files in the two dropdown lists.)

7. You will then receive an email on the address you provided as the receiver address. Subject of this mail will be "Support call from scanner '<Identification number (MAC address)>'".

4.12.3. Troubleshooting Question: When I click on Send Log Files as E-mail, I get an error message saying: There is

no mail receiver configured (SMTP To-address). What should I do? Answer: Either you did not enter a valid email address, or you forgot to remove the ; sign

from the beginning of the line. Please check your console.ini file to correct the line. Question: When I click on Send Log Files as E-mail, I get an error message saying: Failed to

connect to SMTP server "<name>" (). What should I do? Answer: This means the SMTP server is not valid. You could try to ping it as the first step.

Next step would be to make sure that the SMTP service on the specified server is alive. This can be done by the hospital IT department only.

Page 61: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-33

4.13. How to configure scanner to use an external hard disk to save scans instead of the internal hard disk Many situations demand that you use an external hard disk on the scanner. The support for external hard disk exists already. You can use the external hard disk as an extended USB Flash key to save scans on the hard disk. On the hard disk scans (2D / 3D / cine) are stored in this directory: c:\console\documentation. This folder also contains a patient database (PatientDB.mdb). The database is vital for the archive system to work, as it contains information about patients that has been entered manually or fetched from a Dicom worklist. This patient database also has the links for each of the image-files from the documentation directory. This link is a hard link, which means that the link consists of the partition data, directory data and file name data. A link could look like this: c:\console\documentation\20051114_142222_NoID.bmp. The database is a Microsoft Access database and it has been password protected, so unauthorized access can be prohibited. When the console software is running, the patient database you will be able to see an additional file: PatientDB.ldb. This file is a lock-file for the PatientDB.mdb-file. When the console software is running, you will not be able to copy or view the PatientDB.mdb-file, because of the lock file. When you examine the directory in service mode you may not find this lock-file anymore, but if you do, you can delete it if needed. Next time the console application is started, a new lock file will then be generated. Finally there is a backup of the PatientDB.mdb in the directory as well, but at time being it is useless. If your PatientDB.mdb has crashed, the backup has been damaged as well. (We are looking into this). The standard solution when you use the external hard disk as an extended USB flash key does not save the patient database on the hard disk. The tweak below makes it possible to save the patient database to the external hard disk as well. More precisely this tweak will make the c:\console\documentation directory almost obsolete. Here is the procedure on to get it work. (If you are not sure about what you are doing, please do not continue. A mistake could make the scanner unstable.) Your external hard disk has been partitioned (NTFS) and formatted before you plug it into the scanner. Only external hard disk types supported by the scanner will work. 1. Bring the scanner into service mode

2. Connect you external hard disk and wait for approx 2 min.

3. Right click on My Computer and select Manage

4. Click on Disk Management

5. Examine in the right side on the window that the external hard disk has been found. If you do not see the external hard disk either it’s not working properly or it’s not supported. Do not continue if you do not find your external hard disk in the window.

6. Make sure that the DVD drive has drive letter D.

Page 62: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-34 BI2201-E

7. Now select your external hard disk and make sure that it has NOT any of these drive letters: A, B, C, D, E. (E is reserved for the USB flash key, so even if the USB flash key is not connected, we still want to keep the reservation.)

8. Exit Disk Management

9. Double click on explorer

10. Navigate to your external hard disk

11. Make a directory (fx Documentation). (You can use the root of you external hard disk, as this will not work).

12. Navigate to the c:\console directory

13. Make a backup of the console.ini file

14. Edit the console.ini file

15. Find this section: [Paths]

16. Make a new line just below this line: ;;;;Path for documentation directory

17. Write this in the new blank line: DocumentationDirPath1=<drive letter>:\<directory> (In my case I wrote:

DocumentationDirPath1=F:\Documentation, as my external hard disk had drive letter F, and my directory was called Documentation.)

18. Save console.ini

19. Start console (from service mode)

20. Create a test patient and save a couple of images

21. Switch back to explorer and examine that you find the scans and the patient database on

your external hard disk. Default behavior for console software is that it would look in the console.ini file for the documentation paths. First it will look DocumentationDirPath1, if this is not available, it will look for DocumentationDirPath2. Default values for DocumentationDirPath1 is q:\documentation and for DocumentationDirPath2 is c:\console\documentation. This works out as a safety line for this tweak. In case the external hard disk is not connected, console software will use the second path. If the console software does not find a patient database, a new empty one will be created. It can be confusing having two different patient databases and two different set of images etc.

Page 63: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-35

4.13.1. How to configure scanner to use a network drive It is possible to configure scanner to use a network drive to archive examinations. There are 3 steps in this configuration. 1. Configure Microsoft Windows-based PC 2. Configure Network 3. Configure Scanner

4.13.1.1. Configure Microsoft Windows-based PC Configuration of Microsoft Windows-based PC is a very big topic. Bottom-line is that we need a shared folder on the hard drive, where scanner can save files. Basic step for a Microsoft Windows XP: 1. Start Explorer 2. Create a folder on your hard drive.

Make sure you have enough space.

3. Right click on the folder and select Sharing and security…

4. Select Share this folder

5. Enter a Share name as the field will now be available.

6. Click on Permissions as the button is also available now.

7. Make sure the user have Change rights.

8. Click on OK to close Permission Pop up 9. Click on OK

Page 64: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-36 BI2201-E

4.13.1.2.

4.13.1.3.

Configure Network You can connect the Microsoft Windows-based PC and the scanner different way. It’s not the scope of this document to cover that. Main idea is that you need to configure network for the Microsoft Windows-based PC first. Then configure network on the scanner. This can be done by entering Service mode, if Dicom is not available on the scanner. Just make sure the network connection between scanner and PC is working.

Configure Scanner Click on Customize, System and then Network Drive

Enter the UNC path in the field. The UNC path is the path to the folder you created in the first step. So fx. If your PC has this IP Address: 192.168.1.10 and the folder name is 2202 then you enter this: \\192.168.1.10\2202 Now enter username and password for the user that is allowed to save files in the folder created in the first step. If you are on a Domain you may have to write the username like this: Domain\username. (eg bkmed\sla, where bkmed is our domain name and sla is my username.) Click on Test Connection to test the connection. A success full test looks like this:

A failing test will keep asking for a correct set of username/password: When you have success fully tested the connection, click on Close. (Click on Yes to save, if asked.)

Page 65: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-37

4.13.2. How to transfer file from one hard disk to another If the scanner hard disk crashes du to Microsoft Windows or the console application, in most cases, you are able to rescue the images. First you should read chapter 4.9 and chapter Error! Reference source not found.. There are two easy ways to transfer file from one hard disk to another. First one needs to realize which files are going to be transferred. Usually we are interested in the particularly this directory: c:\console\documentation, as all exams and the patient database is stored in this directory. Both of the ways require special tools. None of these parts are sold by B-K Medical. You must buy them at your local store.

a. IDE (44 pins) to USB adapter b. Dual connector IDE (44 pin) cable + 1 jumper

This adapter can be bought in many different types. This is just one of them.

The dual connector IDE (44-pins) cable is needed.

You actually only need 1 jumper.

4.13.3. Using IDE (44 pins) to USB adapter 1. You must replace the scanner hard disk first. 2. Then you must start the scanner and see that everything is working perfect. 3. Bring the scanner into service mode

4. Now you take the old hard disk and mount it on the adapter and connect the adapter to

the USB connector (using the cable coming with the adapter).

5. Wait until Windows has found the hardware and the hardware is ready to be used.

6. Start Windows Explorer and use the hard disk (which now is your USB hard disk)

Con’s Pro’s 1. On the first core boards ZD0773 the USB

standard is 1.1. This gives a very limited transfer speed.

2. Not all the adapters will be detected by Windows. If your adapter is not detected, only way out is to replace it.

1. On the latest core boards ZD0773 the USB standard is 2.0. This gives a very high transfer speed.

2. It’s more convenient to work with. 3. You can use the adapter to connect the hard

disk almost everywhere.

Page 66: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-38 BI2201-E

4.13.4. Using dual connector IDE (44 pins) Cable 1. You must dismantle the scanner. 2. Take out the core board assembly. 3. Remove the hard disk and the cable. 4. Connect the dual connector cable. 5. Insert carefully the core board assembly in the cassette. 6. Connect the new hard disk as Device 0 7. Connect the old hard disk as Device 1. 8. Put on the jumper as shown on the picture on the hard disk. 9. Then you must start the scanner and see that everything is working perfect. 10. Bring the scanner into service mode 11. Start Windows Explorer and use the hard disk.

Con’s Pro’s 1. It’s not convenient to work with. 2. You must assemble and disassemble a

couple of time

1. Transfer speed is very high, as the IDE bus is being used.

Page 67: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-39

4.14. Software Structure In this chapter we will look at the directory structure and then on some of the important files. The hard disk has been partitioned into one single partition. This is mapped to drive letter C. The CD drive has been mapped to drive letter D. Drive letter E is reserved for an optional USB flash key.

4.14.1. Directory Structure of hard disk (C-drive) Some of the directories and files are Microsoft Windows standard. These will not be addressed any further here. The important directories have been highlighted in the image above. We will in the following concentrate on these five directories.

4.14.1.1.

4.14.1.2.

CasaEngine All hardware configuration files are located in this directory. There are two sub directories.

Console This is the main software directory. There are 10 subdirectories which all require a short presentation. In the console directory you find the important console.ini file, which is the software configuration file.

Bodymarks Each of the body marks is represented by an image file. The default naming policy is bodym_<number>.bmp (fx. Bodym_001.bmp). If you find any files which do not follow the above rule, they are custom made. In case of file rescuing these might be needed.

CasaWare This directory includes supporting software for the console application.

Page 68: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-40 BI2201-E

Codec, Lagarith In this directory you can the find the Lagarith Codec. This codec is used in the Cine clips. First time you want to view a clip on a PC, you must install the codec. This directory is copied to the CD every time you copy exams (depending on your configuration.)

Database This is a very important directory, as (almost) all databases are stored here. The main database is 2202.mdb. Then there is the backup of this file: 2202_backup.mdb. The Dicom setup is stored in DICOMSetupDB.mdb. PatientDB.mdb is blank copy of the patient database. The sub directory org contains an empty 2202.mdb

Documentation This is where all exams are stored. All 3D scans, all US images and all clips. Furthermore you find the actual patient database PatientDB.mdb and a backup of this here.

Report Files needed to create a report printout are stored here.

StagingArea This is the staging area directory. All exams that have been ordered to burn on a CD will be here until the CD has been burned. Then this directory will be empty. If you have something in this directory, each time the console application starts you will see a message about it.

Dicom This directory stores both the Dicom program files as well as links to documents that are being send to a Dicom device.

Page 69: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-41

4.14.1.3. XPE_Backup This directory is important after software upgrade. If you make a clean installation, this directory is not used. Inside this directory a collection of different files can be found. These are copied from different places on the hard disk. Here is the directory listening:

File / Directory name Comment Copied from 2202.mdb Main setup database C:\Console\Database Console.ini Main configuration file C:\Console Lbk.bin License File C:\ PatientDB.mdb Empty Template C:\Console\Database Viking.ini Actual configuration C:\ Bodymarks [DIR] All bodymarks C:\Console Documentation [DIR] All patient scans and

Patient database C:\Console

4.15. How to access the content of the CD in the CD-drive from Windows Explorer in service mode It’s not possible to read the content of the CD in the CD-drive in Windows Explorer, due to some Microsoft setting. You can access the content using a Command Prompt.

4.16. There is no Recordable CD Drive Somehow Windows could not recognize the CD drive as a recordable CD drive. We must make Windows use the correct driver for the drive. 1. Start scanner in service mode 2. Remove all USB flash keys (only the CD should remain) 3. Open Device Manager (icon on right top of screen)

4. Select DVD/CD-ROM drives 5. Click on the CD-rom device (without any brand name) and click on Action and then

Uninstall

6. Click on OK. 7. Select DVD/CD-ROM drives

CD-Rom drive

Page 70: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-42 BI2201-E

8. Click on Action and select "Scan for hardware changes" 9. Examine the CD-drive. It should be: fx. TEAC DW-224E-C USB Device

10. Exit Device Manager

4.17. How to make your own BIOS plug EL1011

The BIOS plug EL1011 is used in the 2202 keyboard whenever BIOS access is needed. The plug short circuits pin 5 & 6 on the RJ 12 connector in the keyboard.

4.17.1. Tools Required

• 1 pcs RJ12 6/6 line plug (Farnell no. 106309

• 4 cm cable 26 AWG (used in 100Mb LAN cable)

• Tool for RJ12 (Paladin Tools 1302)

4.17.2. Assembly Procedure On the RJ12 plug short Pin 5 and Pin 6 with the 4 cm 26 AWG cable. Use the crimping tool to fasten the cable.

Page 71: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-43

4.18. License Codes The software options on the scanner are controlled by a license enabling a specific option. The license(s) are kept in a file on the hard disk. The license combined with the Scanner Identity which is a unique number for each scanner opens up for the option so a specific license will only work on one scanner. The license(s) and Scanner Identity can be found under Customize - System - License. The scanner identification is the MAC address of the ETX board on scanners with serial no. below 1864900. Scanners with a higher serial no. have the serial number programmed into the Backplane board so the scanner identification is the serial no. in Mac address format. This means that if changing the EXT board or Core board assembly in a scanner with a serial no. below 1864900 then new licenses have to be entered. If replacing a backplane board in a scanner with serial no higher than 1864900 then the MAC address will be the new Scanner Identification unless the back plane board has been preprogrammed from the factory. If installing a hard disk from another scanner then the license will be inactive because the Scanner Identification does not match. In that case the licenses have to be changed using the license CD that came with the system see 4.18.1 or by re-entering the license(s) manually. If entering the license manually then the file C:\LBK.BIN must be deleted first. Licenses cannot be entered on a system which has one or more invalid licenses in the LBK.BIN file. If the LBK.BIN file contains an invalid license then all licenses will become invalid and the scanner will behave as a Blue Pro Focus without any options enabled. Note! The “Delete License” is only for demo licenses and not permanent licenses.

4.18.1. Install License Codes from the Back-up CD

Note! Changing the time and date after the licenses has been applied may result in deactivation of the license(s)

Scanner Identification Either serial no. or MAC address

Page 72: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-44 BI2201-E

1. Enter Service Mode

2. Click on “Command Prompt”

Microsoft Windows XP [Version 5.1.2600] (C) Copyright 1985-2001 Microsoft Corp.

C:\WINDOWS\system32>

3. Insert the license CD in the CD-drive. 4. Type d: 5. Type install.bat

When the line: “1 file copied.” appears the license(s) have been installed

4.19. Power Supply Trouble shooting

4.19.1. Introduction The Power Supply On/Off control is placed on the Core board so if you have a problem starting up the Power Supply it could be the Power Supply itself or the Core board. Note : A number of things on the Core board Assembly could result in start up problems (Core board it self, Lithium battery being completely flat, defective EXT board)

4.19.2. LED indicators on the Power Supply There are some LED’s on the Power Supply indicating the status of it. If you look behind the holes near the mains input connector you will find 3 green LEDs. These should all be lit when the system is turned on. Further inside the Power Supply you find a green and a Yellow LED. These can be difficult to see as you have to look through a hole in the PCB that is placed above the LEDs, see the below drawing. The Green LED is the 5V standby voltage and this should be on whenever mains are fed to the Power Supply. The Yellow Led indicates that the system is OFF but has previously been turned ON.

Page 73: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Troubleshooting

BI2201-E 4-45

4.19.3. Steps to test the Power Supply 1. Check that the Green 5V Standby is lit. If not check mains supply and fuses

2. When pressing the Power ON switch on the front of the scanner do you then

hear any reaction from the Power Supply (Fans starts running, light in the 3 Green LEDs, Light in the Power ON switch). If not first remove the mains for 1 minute and the reconnect it and try turning on the scanner again - if that does not help check the Power ON Switch and also check the voltage on the Lithium Battery on the Core board

3. Test the Power Supply as described in 4.19.4

4.19.4. How to test the Power Supply ZG0346 as standalone In section 9, chapter 3 you find the power supply connector diagram. The power supply behaves just like a PC ATX Power supply, but provides additional voltages. In the stores you can find different PC power supply testers. We have tested a product from Coolmax: PS-124. (http://www.coolmaxusa.com/powertester.asp?item=powertester). This product is a simple tester for switching power supply. It makes it easy to test the power supply output. (This is not an advertisement for Coolmax. We do not sell this product, so you have to buy it or some other equivalent from your local store.) Next step is to create connector cable, as the pin layouts of the ZG0346 connector are not direct mapped to the pin layouts of the tester. Using the table, you make the “24-pins to 20-pins adapter” cable. Please use AWG 16 and AWG 20 cables.

5V Standby OK

Page 74: Ecograf System Pro Focus 2202(BI2201-E)

Troubleshooting System Pro Focus 2202

4-46 BI2201-E

Tester/ATX

connector Pin

Size AWG Name Color ZG0346 connector

Pin 1 16 +3.3V Orange 5 3 16 COM (GND) Black 12 4 16 +5V Red 8 8 20 PWR_OK Gray 2 9 20 +5Vsb Violet 3 10 16 +12V Yellow 4 16 20 PS_ON Green 13

When your adapter cable is ready, you can make the test: 1. Plug in your 24-pins “24-pins to 20-pins adapter” connector to tester

2. Plug in your 20-pins “24-pins to 20-pins adapter” connector to the power supply ZG0346

3. Connect the power cord to the power supply

4. Turn on the power switch in the wall outlet.

5. Check LED lights on the tester: (+5V. +12V. +3.3V. +5VBS. PG)

• If the power output is working, the LED will light and you will hear a sound • If the power output fails, you won't see the LED light & no sound

6. Turn off the power switch in the wall outlet.

7. Remove the connectors Warning: This tester is designed only for power supply output tests. Do not leave it connected to the power supply unattended. It generates heat and could possibly damage your power supply. The pin layout for a standard ATX power connector looks like this:

Page 75: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Adjustment Procedure

BI2201-E 5-1

Section 5

ADJUSTMENT PROCEDURE CONTENTS PAGE

5.1. Introduction................................................................................................................................................2

5.2 Trackball Friction Adjustment .......................................................................................................................2

5.3 Adjusting the monitor friction...................................................................................................................3

5.4 Adjusting the Brightness of the Screen...................................................................................................6

5.5 Adjusting the Reading Lights....................................................................................................................6

Page 76: Ecograf System Pro Focus 2202(BI2201-E)

Adjustment Procedure System Pro Focus 2202

5-2 BI2201-E

5.1. Introduction All of the adjustable components in the 2202 are factory adjusted and need no further adjustment. Do not attempt to make other adjustments than stated in this chapter.

5.2. Trackball Friction Adjustment The trackball should move smoothly. If it is too tight or too loose it can be adjusted using the adjusting tool (QA0228) which is placed in the compartment underneath the keyboard. Turn clockwise to tighten and counter clockwise to loosen the trackball. The trackball can be removed completely for cleaning.

+-

Page 77: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Adjustment Procedure

BI2201-E 5-3

5.3. Adjusting the monitor friction (CRT monitor only) 1. Loosen the two screws holding the CD/DVD Drive. (Use a screwdriver if

necessary)

2. Eject the CD tray by inserting a paper clip or similar in the small hole on the front and pull out the CD/DVD drive completely

12

3

Page 78: Ecograf System Pro Focus 2202(BI2201-E)

Adjustment Procedure System Pro Focus 2202

5-4 BI2201-E

3. Using a 4 mm hex key the monitor friction can now be adjusted.

+-

4.7- 5 Kg.

Page 79: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Adjustment Procedure

BI2201-E 5-5

4. Reinstall the CD/DVD drive by pushing it back into the monitor shelf. Make sure that it is connected properly - i.e. you should hear a click when pushing it the last ½ cm.

5. Secure the CD/DVD drive by tightening the two thumbs screws

21

2

Page 80: Ecograf System Pro Focus 2202(BI2201-E)

Adjustment Procedure System Pro Focus 2202

5-6 BI2201-E

5.4. Adjusting the Brightness of the Screen

Your scanner may be used with different levels of room lighting. Whenever the lighting changes significantly, you should recalibrate the screen brightness. To calibrate the screen brightness: 1. Freeze the image.

2. Press ALT+SHIFT+G on the keyboard. A series of gray bars appears on a

black background.

3. Use the contrast button on the monitor to adjust the bars so that the 100% bar is bright white and the 90% to 10% bars seem to decrease progressively in brightness.

4. Use the brightness button on the monitor to increase the brightness until the 1%, 2% and 3% bars (above the 10% bar) become visible.

5. Use the brightness button to decrease the brightness until the background becomes black and the 1% bar just disappears.

5.5. Adjusting the Reading Lights After you have turned on the scanner, you can adjust the brightness of the 3 small "lamps" under the CD drive - the Reading Lights. To adjust the reading light brightness: Click Light under Image on the right side of the screen and drag the slider (or point at Light and press [+/-]).

Reading "lamps"

Page 81: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-1

Section 6

Mechanical parts CONTENTS PAGE

6.1. Dismantling and Reassemble Procedures ....................................................................................2 6.1.1. Removing the Electronic Cassette......................................................................................................2 6.1.2. Getting access to the PC boards ........................................................................................................5 6.1.3. Removing the Front End board...........................................................................................................6 6.1.4. Removing the Delay board..................................................................................................................7 6.1.5. Removing the Core board ...................................................................................................................8 6.1.6. Removing the Power Supply...............................................................................................................9 6.1.7. Replacing the Single Element Module ..............................................................................................11 6.1.8. Replacing the Mover Control Module (3D MCM)..............................................................................13 6.1.9. Replacing the ECG Module...............................................................................................................15 6.1.10. Replacing the Flat screen .................................................................................................................18 6.1.11. Replacing the monitor ZV0070..........................................................................................................19 6.1.12. Replacing the ZV0065/ZV0069 monitor............................................................................................20 6.1.13. Replacing the Gas springs ................................................................................................................25 6.1.14. Replacing the CD drive .....................................................................................................................27 6.1.15. Replacing monitor brackets GV2563 ................................................................................................28 6.1.16. Replacing keyboard brackets............................................................................................................29 6.2. Part list .............................................................................................................................................30 6.2.1. Exploded view Cassette....................................................................................................................30 6.2.2. Exploded View Monitor .....................................................................................................................32 6.2.3. Exploded view Keyboard, Trolley......................................................................................................33

Page 82: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-2 BI2201-E

6.1. Dismantling and Reassemble Procedures

6.1.1. Removing the Electronic Cassette

1. Slide up the monitor

and keyboard to top position

2. Remove all

transducers from the transducer connectors and disconnect the keyboard.

º

Page 83: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-3

3. Disconnect the cables from the monitor assembly and the mains cable (+ printer or other connected accessories)

4. Unscrew the two

screws and remove the back cover

Page 84: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-4 BI2201-E

5. Disconnect the earth

wire from the Cassette, loosen the four screws holding the four metal plates and turn them so the cassette can be pulled out.

6. Pull out the cassette

Page 85: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-5

6.1.2. Getting access to the PC boards

Remove the electronic cassette as described in section 6.1.1. 1. Unscrew the 4 screws

marked at the picture

2. Loosen the three

screws holding the lid and pull off the lid.

Page 86: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-6 BI2201-E

6.1.3. Removing the Front End board

Pull out the cassette and get access to the PC boards as described in section 6.1.1 and 6.1.2.

1. Unscrew the two screws holding the cover plate for the single module and remove the plate by turning it.

2. Unscrew the 13 screws

holding the cover plate for the transducer connectors

3. Disconnect the 3

connectors on the flat cable by pushing the lock and then pulling the cable

Page 87: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-7

4. Pull out the Front-End board

6.1.4. Removing the Delay board Pull out the cassette and get access to the PC boards as described in section 6.1.1 and 6.1.2.

1. Disconnect the 3 connectors on the flat cable by pushing the lock and then pulling the cable

2. Pull out the Delayboard

Page 88: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-8 BI2201-E

6.1.5. Removing the Core board Pull out the cassette and get access to the PC boards as described in section 6.1.1 and 6.1.2.

1. Unscrew the screw at the top and the screw at the bottom of the video connector plate

2. Loosen the four screws

at the right side of the video connectors and pull out the Core board

Page 89: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-9

6.1.6. Removing the Power Supply Pull out the cassette as described in section 6.1.1.

1. Unscrew the two screws holding the cover plate for the single module and remove the plate by turning it.

2. Unscrew the two screws

3. Unscrew the screw on side of

the Power Supply

Page 90: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-10 BI2201-E

4. Place the cassette on the side and unscrew the last screw holding the Power Supply fixed to the cassette

5. Slide out the Power Supply 6-7 cm and disconnect the two connectors from the Back Plane

Page 91: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-11

6.1.7. Replacing the Single Element Module

Pull out the cassette as described in section 6.1.1 .

1. Unscrew the four screws holding the cover plate for the single module and remove the plate by turning it.

2. If the scanner has a 3D

Mover module (MCM) that is connected to the Single module then this must also be slide out to disconnect the internal cable.

Internal cable between Single Element Module and MCM

3. Unscrew the screw

securing the Single Element module (top screw). If the Mover Control Module bee taken out then also unscrew this (bottom screw)

Page 92: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-12 BI2201-E

4. If necessary then slide out the Single module and the MCM and then disconnect the internal connection cable

( )

5. Pull out the Single

Element Module and disconnect the Cable from the Back Plane

Page 93: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-13

6.1.8. Replacing the Mover Control Module (3D MCM)

Pull out the cassette as described in section 6.1.1 .

6. Unscrew the four screws holding the cover plate for the single module and remove the plate by turning it.

7. If the scanner has a 3D

Mover module (MCM) that is connected to the Single module then this must also be slide out to disconnect the internal cable.

Internal cable between Single Element Module and MCM

Page 94: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-14 BI2201-E

8. Unscrew the screw securing the Mover Control Module (bottom screw). If the Single Element module has to bee taken out then also unscrew this (top screw)

9. If necessary then slide

out the single module and MCM and then disconnect the internal connection cable

( )

10. Pull out the MCM and

disconnect the cable from the Back Plane

Page 95: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-15

6.1.9. Replacing the ECG Module

1. Slide up the monitor

and keyboard to top position

2. Remove all

transducers from the transducer connectors and disconnect the keyboard.

º

Page 96: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-16 BI2201-E

3. Disconnect the cables from the monitor assembly and the mains cable (+ printer or other connected accessories)

4. Unscrew the two

screws and remove the back cover

5. Unscrew the three

screws holding the cable cover plate and remove the plate

Page 97: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-17

6. Disconnect the flat

cable

7. Slide up the module

and then pull it off the cassette. Turn it so the connectors can be accessed and disconnect them. 1

2

3

4

Page 98: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-18 BI2201-E

screen

nitor and

the power cable on the Flat panel

6.1.10. Replacing the Flat 4.1. Disconnect the mo

signal cable (VGA)

4.2. Loosen two

thumbscrews while supporting the flat screen

4.3. Lift up the flat screen

Page 99: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-19

6.1.11. Replacing the monitor ZV0070 ttached handle so this must be moved to the replacement monitor

. Unplug the VGA and

Note the monitor comes without the a

8mains connector and remove the Earth wire.

9. Pull the two thumbsc

1

2

rews out and rotate them 45° so that the pin is in the position where the monitor foot can be released from the tilting base and lift off the monitor

Place the monitor upside down Mount the handle on the monitor with the four screws, washers and spacers

×4

Page 100: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-20 BI2201-E

onitor 6.1.12. Replacing the ZV0065/ZV0069 m 1.1. Slide up the monitor

and keyboard to top position

tes 1.2. Remove the cover

pla

Page 101: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-21

1.3. Loosen the two screws

Drive (a screw driver can be used if they are

holding the CD/DVD

very tight)

1.4. Eject the CD tray by

inserting a paper clip or similar in the small hole on the front and pull out the CD/DVD drive completely

12

3

1.5. Disconnect the cables

from the monitor assembly

Page 102: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-22 BI2201-E

s from the legs by pressing at the bottom part while pulling up the plastic cover.

1.6. Remove the monitor assembly

3.1. Remove the two black

cable cover

1

2

1

2

3.2. Unscrew the eight

screws holding the plate with the USB-hub

Page 103: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-23

b as shown and cut the two strips

3.3. Place the plate with the USB-hu

releasing the cables from the two arms.

3.4. Unscrew the screw

holding the ground wire e and remove the wir

from the plate.

3.5. Loosen the 6 screws

holding the two brackets and slide the brackets towards the centre to remove the monitor shelf from the monitor

Page 104: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-24 BI2201-E

g ing

it

3.6. Remove the tilting

base from the old monitor by unscrewinthe four screws hold

Page 105: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-25

Gas springs

holding the top cover plate and remove the plate

6.1.13. Replacing the 11. Unscrew the four screws

1

2 1

12. Unscrew the screw holding the spring and lift up the hole assembly with the spring in both sides

4

3

Page 106: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-26 BI2201-E

.3. Remove the gas springs by turning the top and

2

bottom assemblies.

three strengths reen ZV0071 is mounted

Nm when CRT monitor ZV0065 is mounted 150 Nm when CRT monitor ZV0070 is mounted Replace the spring and reverse the reassemble procedure for the springs

2.4. The spring comes in100 Nm when flat sc120

Page 107: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-27

drive

1.7. Loosen the two screws holding the CD/DVD Drive (a screw driver can be used if they are very tight)

6.1.14. Replacing the CD

12

3

1.8. Eject the CD tray by inserting a paper clip or similar in the small hole on the front and pull out the CD/DVD drive completely

Page 108: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-28 BI2201-E

ets GV2563

ke off

t do the screws, so they

are loose.

3. Put the adjustment tool (QA0229) in between the two brackets.

6.1.15. Replacing monitor brack 1. Loosen the knobs and ta

the monitor. 2. Replace the backets, bu

not tighten

4. Tighten the screws in the left

side and then in the right side, nt tool to

ave

and use the adjustmemake sure the brackets hthe correct alignemt.

5. Slide out the adjustment tool after having tighten all screws.

Page 109: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-29

nobs and take off the keyboard.

6.1.16. Replacing keyboard brackets

º

1. Remove the keyboard 2. Loosen the k

3. Replace the backets, but do

not tighten the screws, so they are loose.

4. Put the adjustment tool

(QA0229) in between the two brackets

5. Tighten the screws in the left

side and then in the right side, and use the adjustment tool to make sure the brackets have the correct alignemt.

6. Push the adjustment tools

inwards and slide out the adjustment tool after having tighten all screws.

Page 110: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-30 BI2201-E

.2. Part list

6

6.2.1. Exploded view Cassette

6

15 14

213

5

2

1

3

13 20

11

4

7

8

22

16 9

19

17

(12)

18

12

10

Page 111: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-31

tem no. Part no. Description Quantity 1 ZN2242 Front-end Assembly (Front-end + QCCM) 1

243 Front-end Assembly (Front-end + QCCM) for Blue 1 798 Front-end board 1

1 ZE0815 Front-end board 1 1 ZE0822 Front-end board 1 1 ZE0820 Front-end board for Blue 1 2 ZE0797 Channel Condition Module (QCCM) - only for ZE0798 16 2 ZE0817 Coded Quad Channel Condition Module (CQCCM)only for ZE0815 16 2 ZE0823 Coded Quad Channel Condition Module (CQCCM)only for ZE0822 16 2 ZE0821 Quad Channel Condition Module (CQCCM)only for ZE0820 16 3 ZD0774 Digital Beam former 1 4 ZN2220 Core board assembly (Core - , Video - , ETX board and HD) 1 4 ZN2231 Core board assembly (Core -, Video - , ETX board and HD) Green 1 4 ZN2230 Core board assembly (Core - , Video - , ETX board and HD) Blue 1 4 ZD0773 Core board 1 4 ZD0789 Core board for Blue 1 5 ZH0758 Video IF board 1 6 ZD0784 ETX Single board PC 1 6 ZD0795 ETX Single board PC for Green and Blue 1 7 UL0041 Hard Disk Incl. SW 1 8 ZG0346 Power Supply 1 9 ZN2201 Single module (incl. of AQ1725) 0-1 9 ZN2223 Single module (incl. of AQ1725) with splitter 0-1

10 ZN2202 MCM module 0-1 10 ZN2222 MCM module with splitter 0-1 11 ZH0757 Back Plane 1 12 ZN2203 ECG Module 0-1 14 AQ1728 Flat cable Hard disk to Core board 1 15 AQ1725 Flat cable Video IF to Core board 1 16 AQ1725 Flat cable Single Module to Back plane 0-1 17 AQ1729 Flat cable MCM to Back plane 0-1 18 AT2314 Power ON/OFF switch 1

SC3157 Cover no options 1 SC3158 Cover Single module 1

19 SC3159 Cover 3D Mover 1 SC3160 Cover Single module + 3D Mover 1 SC3161 Cover ECG 1

20 DM0295 Spring for transducer connector 30 21 QB0041 Lithium battery for Core board 1 22 VF0118 Fuse 10 A T 2

I

1 ZN21 ZE0

Page 112: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6-32 BI2201-E

6.2.2. Exploded V ito

Item no. Part no. Descr rt no. Description ntity

iew Mon r

iption Quant. Item Pano.

Qua

101 ZV0065 Replacement kit ZN2225 GU1792 Monitor fixing plate 1 N/A 116 101 ZV0069 15” CRT Monitor 117 GU1799 Monitor fixing plate 1 1 101 ZV0070 15” CRT Monitor 1 118 YT7410 Screw M4×10 mm 6 102 ZV0071 17” Flat Screen 1 119 DP0811 Tilt Stop 2 103 DH0927 Handle for ZV0070 1 120 YO0687 Washer 4 104 YS1320 Screw for handle M3×20mm 4 121 YO0800 Spring washer 2 105 YO0402 Washer 4 122 YQ2635 Screw M6×35 mm 2 106 DB3527 Bushing 4 123 HP0025 Loudspeaker 2 107 YB0007 Securing bolt for monitor 2 124 UL0036 CD R/W - DVD 1 108 EA0201 Base Plate for monitor 1 125 ZH0759 HUB controller board 1 109 YS9352 Screw M3×8 mm 4 126 GV2561 Plate 1 110 DZ9766 Tilting Base 1 127 YT4408 Screw M4×8 mm 8 111 YJ3133 O-ring 2 128 DD0581 Cable cover 2 112 DP0822 Slide 2 DH0459 Handle for flat screen 1 113 YS1412 Screw M4×12 mm 4 114 DZ9767 Tilting Seat 1 115 YO1626 Sliding ring 1

101 1

102 127 124

23

125 126

108

109

122

103 104 105

106

107

128

113

111

112

114

115

116

118 119

120 121

110

117

Page 113: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Mechanical Parts

BI2201-E 6-33

, Trolley

Item no. Quant. Quantity

6.2.3. Exploded view Keyboard

Part no. Description Item no.

Part no. Description

201 UD0152 Arm assembly for flatscreen base 1 217 DH0912 Handle for keyboard 1 204 DH0871 Thumbscrew 4 218 DS0967 Hand rest 1 205 YT1408 Screw M4× 8 mm 2 219 DZ9001 Top plate 1 206 DZ9003 Back Cover 1 220 YS0823 Screw M8 × 12 mm 4 207 AO0462 USB cable 1 221 DZ9002 Side bar cover 2 208 AO0455 Cable for USB HUB 1 222 DS0971 Gas spring (ZV0071)100 NM 2 209 DF0257 Twin Wheel with brake 2 222 DS0964 Gas spring (ZV0065)120 NM 2 210 DF0256 Twin Wheel 2 222 DS0968 Gas spring (ZV0070)150 NM 2 211 ZN0046 Keyboard comp. - Green/Yellow de profile 1 223 DK1300 Blue si 2 211 ZN0047 Keyboard complete - Blue 1 223 DK1301 Green side profile 2 212 SN0220 Dial - small 1-2 223 DK1291 Yellow side profile 2 213 SN0221 Dial - Gain 1 224 DH0678 Keyboard cable relief 1 214 DP0912 Potentiometer button 225 DZ9751 ront cover 8 F 1 215 DZ9755 Transducer holder 4 216 DZ9754 Gel holder 2

213

211

212 214 215

216

217

218 219 220

221

204

222 223

225

224

201

204 207

205 208

206

209

210

Page 114: Ecograf System Pro Focus 2202(BI2201-E)

Mechanical Parts System Pro Focus 2202

6 BI2201-E

This page is left blank intentionally

-34

Page 115: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Preventative Maintenance

BI2201-E 7-1

Section 7

PREVENTATIVE MAINTENANCE CONTENTS PAGE

7.4. Introduction........................................................................................................................................ 7-2

7.5. Necessary Equipment ....................................................................................................................... 7-2

7.6. What to do .......................................................................................................................................... 7-3

Page 116: Ecograf System Pro Focus 2202(BI2201-E)

Preventative Maintenance System Pro Focus 2202

7-2 BI2201-E

7.1. Introduction The purpose of the Preventative Maintenance is to ensure the performance and stability over the years of use. Note: The Preventative Maintenance should be performed once every year. Preparation of the Preventative Maintenance is essential as it in most cases will take place at the hospital as part of a Continuity Agreement. The Preventative Maintenance consists of eight steps. Some of these steps refer to specific sections in the Service Manual. If errors occur during the procedure go to TROUBLESHOOTING, Section 4.

7.2. Necessary Equipment The equipment, necessary for a preventative visit is divided into 3 packages. A: Parts necessary to ensure the performance and stability of the system. Check in the Service Informations (Hot-Line) if there are any modifications to be done. B: Parts necessary to meet customers complaints. Before arranging the visit, check if the customer has any complaints and bring the necessary spare parts and tools to solve the problem. C: Tools and equipment necessary to perform the Preventative Maintenance. The only equipment necessary to perform the Preventative Maintenance is the equipment used in the Electrical Safety Test and phantoms for the Checking Procedure: For the Checking Procedure:

Transducer Phantom, model 254 (B-K no.WQ 0973) for checking a 2202 system equipped with Array- and Single Element Transducers. Transducer Phantom, model 251 (B-K no.WQ 0972) for checking a 2202 system equipped with the Endosonic Probe 1850.

For the Electrical Safety Test:

Testers required: Safety tester HV Tester

Note: It is recommended to bring additionally the Static Control Service Kit, type 3M 8501 (B-K no. WQ0969) or similar in case one of the static sensitive PC Boards is to be removed from its location in the cassette.

Page 117: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Preventative Maintenance

BI2201-E 7-3

7.3. What to do 1. Users Comments and Corrections

Check with the user that the system is operating satisfactorily. A number of malfunctions can be corrected by a set-up adjustment! If an error is obvious correct it at this state.

2. Modifications

In accordance with the Service Informations (HOT-LINE) perform necessary hw/sw modifications to ensure performance and stability of the system.

3. Mobility

Check that the keyboard/monitor can be moved up/down Check that the trackball moves smoothly Caution - Never add oil or grease to any other movable parts.

4. Fans

Remove the rear cover from the 2202. Connect a mains cable and turn on the unit. Check that both fans in the power supply rotates. Note that the fans are controlled by electronics inside the power supply. Reassemble the scanner.

5. Checking Procedure

To verify proper operation of the main functions perform the CHECKING PROCEDURE, Section 3.

6. Peripherals

This is to verify the function of the video output and the peripherals used with the 2202 System.

Make an image/print and check the quality. If necessary adjust the peripheral using the original instruction manual.

7. Adjustment Procedure

If necessary, perform an adjustment of the Trackball using the ADJUSTMENT PROCEDURE, Section 5.

8. Electrical Safety Test

Verify that the system complies with IEC 60601-1 using the Electrical Safety Test, Section 1

9. Software Maintenance Check available HD space on the system and if necessary clean-up the system or inform the customer to back-up patient data to free space on the HD

Page 118: Ecograf System Pro Focus 2202(BI2201-E)

Preventative Maintenance System Pro Focus 2202

7-4 BI2201-E

This page is left blank intentionally

Page 119: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Theory of Operation

BI2201-E 8-1

Section 8

THEORY OF OPERATION CONTENTS PAGE

8.1. Introduction................................................................................................................................................ 2

8.2. General Description .................................................................................................................................. 2

8.2.1. Block Diagram........................................................................................................................................... 3

8.3. Board Description ..................................................................................................................................... 4

8.3.1. Front End, FE............................................................................................................................................ 4

8.3.2. Digital Beamformer, DBF.......................................................................................................................... 5

8.3.3. Core Board................................................................................................................................................ 6 Real-time Scan Controller (RTSC):.................................................................................................... 6 Vector Processor (VECP): ................................................................................................................. 7 Scan Converter (SCON): ................................................................................................................... 8 Video Merge and Capture (VIME):..................................................................................................... 8 Embedded PC: ................................................................................................................................... 9

8.3.4. Analog Video Interface Board, AVIB...................................................................................................... 10

8.3.5. Analog Video Connector Board, AVCON .............................................................................................. 10

8.3.6. USB Hub, HUB ....................................................................................................................................... 11

8.3.7. External Devices ..................................................................................................................................... 11

8.3.8. Optional Devices..................................................................................................................................... 11

Page 120: Ecograf System Pro Focus 2202(BI2201-E)

Theory of Operation System Pro Focus 2202

8-2 BI2201-E

8.1. Introduction This section describes the theory of operation of the Diagnostic Ultrasound System 2202. The description is divided into two separate sub-sections: General Description Supported by a complete block diagram, this section aims to give you an understanding of the interaction between modules, Boards and major function blocks. It explains briefly the main signal flow through the 2202 system. Board Description This section provides an overall description of each Board, explaining the main function of the function blocks on the board.

8.2. General Description

The basic hardware architecture consists of three active PCBs connected to a passive backplane. The active boards are: 1. A 64-channel analog Front End (FE) board with transmitters, 3 array transducer

connectors, relays, receivers, Harmonic Imaging filters, and TGC amplifiers. 2. A 64-channel Digital Beam former (DBF) with signal conditioning circuitry and

A/D converter per channel and 16 beam former ASICs plus control logic and profile memory.

3. A Core board with real-time scan controller, vector processor, Doppler system

and scan converter/display system. Further, an industry standard PC module and a hard disk drive is attached to the Core board.

Several other units connect to the Backplane. These are: • Power Supply • CD-RW Disk Drive* • Single Element Transducer Module** (field option) • Mover Control Module** for 3D application (field option) • Keyboard assembly • Foot switch (accessory) • Internal speakers* * Connects through a hub. ** The Single Element and Mover Control Modules connect to the Backplane via an interface board. The Core board interfaces to an Analog Video Interface Board (AVIB) with all video and graphics i/o, including the main monitor output. The keyboard is a self-contained unit including keys, track ball, gain dials, TGC potentiometers and a small keyboard controller which interfaces to the built-in PC via a single USB serial interface.

Page 121: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Theory of Operation

BI2201-E 8-3

8.2.1. Block Diagram

VECP FPGA

ETXCard-PC

P1

FE64+ ZE0798 XMIT RAM

RAM Test

Dig Rotate64 32

RAM Test

32HVMux192 64

64

Relays

192conn.

192conn.

192

64

ElementBus

ATGC

SingleElementProbe

PMEM 1

Filter8

A/D8

D/A8

ASIC# 8

PMEM 8

PMEM

Filter1

A/D1 ASIC

# 1

PMEM 1RAM Test

ASIC# 16

64 Ch. Beamformer

DisplayMemory

RAM Test

ImageFormat

Combiner

Scan ConverterVideo Merge/ RT Interface

CLOCKGENERATORSHARC DSP

32SHARC Bus

ATGC

Mouse

Parallel

Keyb.

Powe

r

ResetOn/Off

IsolationTrans-former

SVGA/LCD

6 pairLVDS lines

BEAM

2202Engine

File:Date:19 Jan, 2005

Real-Time Scan Controller

ReviewBuffer

RAM Test

192conn.

Ethernet,DICOM

P2

P3

CORE ZD0773 PC Back-end

Filter64

A/D64

SEMI/F

ZH0756SGLM ZH0651

Backplane ZH0757

VGA

PCI tohostbus

NTSC/PAL

VCR VCR

RGBS out

CVBSCVBS B/W

NTSC/PAL

CVBSS-Video outExt.VGA

24

Line out

Line in

15" VGAMonitor

Videodecoder

TripleDAC

Videoencoder

PCI

18/24

IDE

Seria

l

USB

Analog TCG

AVIB ZH0758

Digital Audio

B.P. I/F

SEQUENCERFPGA

SDRAMMEMORY

BUSBUFFER

HOSTIFFPGA

Local Bus

Timing Bus

SCON FPGA

DSC

Video/MergeFPGA

16/24

RGBS

RGBS

FrameBuffer

I2C

CFMDSP

InputBuffer

SlidingFilter

DownCon-verter

Amplitudedetector

LUTRAM Test

DopplerIntegra-

tor

2x16

16

CFMInputBufferRAMTest

16

SpectralDoppler

DSP

CFMDSP

CFMLUT

RAM Test

32

TGC24 16 16

ContrastLUTRAMTest

9 Resamp-ling8 Contour

Enhance8

8

2x16

88

CFMoutputbufferRAMTest

SpectralDoppleroutputbuffer

8

8

8

Digital Audio

VGA

SPI

Vector Processor

Host bus

SHAREDRAM

RAM Test

AVR uCI2C, WDT

48

USB

ID,Temp.

ID,Temp.

Fans

LVDSserializer

DBF ZD0774Size: 330 x 320 mm

IEEE1394

2 x Firewire

Audio out Audio in

AudioCodec

ID

ID,Temp.

Size: 330 x 320 mm

Size: 330 x 320 mm

CD

RW

USB HUB

US

B

HUBZH0759

HAR

DD

ISK

HARDDISK

PCU

Doc.Trig

2

Serial

16 x QCCM ZE0797

S-VideoZV0065

US

B

AC Output

AlphanumericKeyboard

ControlPanel

Trackball

NP0166

USBHUB

USB

USB

Footswitch

RS23

2

LAN100Base-TX

iLink

BUSCTRL

Powe

r

MCM

ZH0752/0753/0754/0784

AC MainsInput

ID

ID

PMEM 2RAM Test

ID

Power Supply ZG0346

I2C

ID

IPC

BipolarXmitters

Receivers

BipolarXmitters

Receivers

64

Analog Video-I/O board

I2C

Mover

2202 BLOCK DIAGRAM

Relays

AnalogOsc

D/A

AnalogOsc,

Tr code

Signal file

Beamf. Control

Dig. Generator

Light

Fig. 8.2-1 - Block Diagram Including location of test oscillators/generators.

Page 122: Ecograf System Pro Focus 2202(BI2201-E)

Theory of Operation System Pro Focus 2202

8-4 BI2201-E

8.3. Board Description

8.3.1. Front End, FE

XducerConn

A

XducerConn

B

XducerConn

C

Relays 64:192HV Mux

Xmit DigitalRotate

BoardControlFPGA

XmitRAM

Rcv TGC

Backplane

Beamformer64

3264

64192192

192

192

6464

16 QCCM Modules

FE64Block.vsd

64

Fig. 8.2-2 - FE64+ Block Diagram Transducer Connectors: Three 260 pole ZIF CANNON connectors for array probes. Xmit Timing (XMIT RAM & Dig. Rotate): 32 channels transmit timing control with the option for pulse width apodization for controlling the transmit beam profile. The beam profile is unfolded and rotated digitally, and the 64 output channels (bits) drive the bipolar transmitters of the 2202. Transmitters: 64 bipolar current driving transmitters. Output regulation is common to all channels, individual channel apodization may be achieved by pulse width modulation. In CW mode, the output stages are supplied by ±12V instead of the ±HV supply in order to reduce power dissipation. HVMUX: High voltage switches for connecting the 64-channel element bus to the 192 transducer elements. The switches are normally reprogrammed in the time interval between receive end and transmit start, but may also be reprogrammed immediately after transmit end to support synthetic aperture. Relays: 384 bi-stable switches for connecting the element bus to one of three array transducer connectors. Receiver: 64 low-noise preamplifiers with 19dB gain. These are followed by selectable high-pass filters for improved Harmonic Imaging. TGC: 64 TGC amplifiers which handles the first 0 to 48dB TGC gain. Each outputs a differential voltage signal, which is converted into a single ended current output in the transconductance output stage. There is an additional digital TGC gain stage in the Vector Processor on the Core board.

Page 123: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Theory of Operation

BI2201-E 8-5

8.3.2. Digital Beam former, DBF

Input Channel 1

Input Channel 64

ASIC 1

ADC

ADC

ASIC 16

Input Channel 2 ADC

Input Channel 3 ADC

Input Channel 4 ADC

FPGA

ProfileMemory

Beam Sum (TTL)

Backplane

Front End

10

10

10

10

10

64

192

DBFblock.vsd

12

12

Beam Sum (LVDS)

Control (LVDS)

Configuration

Fig. 8.2-3 - DBF Block Diagram Channel Conditioning: 64 input channel strips. Each consists of a current mode receiver, an anti-aliasing filter bank, a filter select switch and a buffer amplifier. A/D Converters: 64 10-bit pipelined ADCs capable of 60Msps conversion. The inputs to the ADCs are AC-coupled (TBD). The outputs are fed to the digital beam former ASICs. Beam former ASICs: 16 Analogic ASICs, each processing four channels with 1, 2 or 4 beams per channel. The ASICs dynamically apply individual delay and apodization to each channel. Profile Memory & Control: Local SDRAM bank for storing the receive profiles needed by the ASICs to process data. The controller FPGA provides host access for uploading the memory, even during scanning, and handles the dynamic download of parameters from memory to the ASICs during scanning. To minimize download time (and increase PRF), all 16 ASICs are loaded in parallel, so the memory is 192 bits wide.

Page 124: Ecograf System Pro Focus 2202(BI2201-E)

Theory of Operation System Pro Focus 2202

8-6 BI2201-E

8.3.3. Core Board

RTSC VECP

SCON

DOP

VIME

EmbeddedPC

B a c k p l a n e I n t e r f a c e s

Misc I/O Static and Real-timeControl Beam Sum

AVIB

Overlay

B/W Vectors

FFT + CFMVectors

Local BusHost BusGraphics

CoreBlock.vsd

SEM

Fig. 8.2-4 - Core Board Block Diagram

The Core Board consists basically of six blocks, as shown in Fig. 8.2-4. These blocks are described in the following:

Real-time Scan Controller (RTSC):

HostInterface

FPGA(XC2S150E)

(FG456)

Local BusSharc Bus

RTSCDSP

(ADSP21161N)(MBGA225)

ClockGen.

SharedMemory

BusBuf-fers

RTSCSequencer

FPGA(XC2S150E)

(FG456)

Backplane

TGCDAC

SEMIF FE DBF

VECP,SCON

FE

RTSCDSP

Memory

RTSCDSP

Memory

RTSCDSP

Memory

SharedMemory

240MHz

Mosc

Host Bus

CoreBlock_RTSC.vsd

Fig. 8.2-5 - Core Board RTSC Sub-system

The RTSC takes care of the hard real-time tasks in the scanner. A local processor (Sharc DSP) is located here to perform a number of tasks: • Calculation of beam profiles for the DBF • Communication with the Embedded PC host processor • Dynamic transfer of parameters to the FE and DBF during scanning • Dynamic transfer of parameters to VECP, DOP and SCON sub-systems during

scanning • Interfacing to the ECG Module • SPI communication with the transducers and

Page 125: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Theory of Operation

BI2201-E 8-7

• Configuration of FPGAs during boot.

A Host Interface provides a memory area, which is shared between the Embedded PC and the Sharc DSP; it is through this memory only that they communicate and exchange data. The Sequencer part of the RTSC generates all timing signals for the acquisition of ultrasound data to the FE and DBF and also internally to VECP, DOP and SCON. It also contains the physical backplane interface to the front-end boards (FE, SEMIF and DBF). Finally, a master clock generator provides all system clocks (except for the Embedded PC), both fixed and variable. The variable clock frequency is determined by the Sequencer (ultimately controlled by the Sharc DSP).

Vector Processor (VECP):

CFMDSP

Clusters& LUT

FFTDSP

(ADSP21161N)(MBGA225)

CFMOutput Buffer

RF InputBuffer

B-modeLUTs

Backplane

SEMIF

VECPFPGA

(XC3S1500)(FG676)

BusBuf-fers

DBF CFMInput Buffer

Local Bus

to Audio Codec

CoreBlock_VPDOP.vsd

to SCON

Fig. 8.2-6 - Core Board VECP and DOP Sub-systems

The Vector Processor performs down-conversion, envelope detection, dynamic range compression, sampling frequency conversion and image enhancement on RF data from the Beam Former. The signal is then delivered to the Scan Converter. In Doppler modes, the VECP delivers a down-converted signal to the Doppler module. Doppler (DOP): The Doppler module comprises the two flow detection modules: Spectral Doppler and CFM, see Fig. 8.2-6. The Spectral Doppler is a CW/PW Doppler system. A Floating-point DSP performs FFT on a time signal from one volume element and outputs a spectrogram to the Scan Converter. The Doppler signal is also output as a stereo audio signal. The processing includes: 1. Input from the Doppler Input Buffer. 2. Wall Filter for removing low frequency tissue-generated motion artifacts. 3. Missing Signal Estimation in block interleaved duplex/triplex modes. 4. FFT: motion detection. 5. Real Time Curves: statistical values for each spectrum (min, mean, max). 6. Audio output. 7. Motion Noise Cancellation: removing transducer movement artifacts. 8. Output to the Scan Converter

Page 126: Ecograf System Pro Focus 2202(BI2201-E)

Theory of Operation System Pro Focus 2202

8-8 BI2201-E

The CFM module maps flow in a section of the scanned area. It processes input data up to a rate of 2.5M CFM cells/second. Output data are velocity, variance and power. Velocity and variance or velocity and power can be combined in one image. The CFM processing includes: 1. Input from the CFM Input Buffer 2. Wall Filter for removing low frequency tissue-generated motion artifacts. 3. Correlation: motion detection. 4. Signal power detection. 5. Parameter extraction in look-up tables. 6. Median Filtering for reducing image noise. 7. Discrimination: eliminating tissue motion artifacts. 8. Output to the Scan Converter.

Scan Converter (SCON):

SCONFPGA

(XC3S1000)(FG456)

LUT

ReviewMemory

DisplayMemory

Xform Coeff.Memory

DSCASIC

ReviewMemory

Local Bus from VECP

to VIME

CoreBlock_SCNV.vsd

Fig. 8.2-7 - Core Board SCON Sub-system

The SCON is integrated and implemented in a single FPGA. It has the same functionality, but given the general physical implementation, see Fig. 8.2-7, it can be changed (even radically) later without changes to the Core board itself. The SCON comprises a frame averaging function with an Image Review Buffer for a large amount of frames (64MBytes), a Scan Line Buffer, and a Scan Processor function, which works in tandem with B-K’s coordinate transformation ASIC. The converted images are buffered in the Display Memory from which they are transferred to the VIME through a color look-up table.

Video Merge and Capture (VIME):

Host Bus

CaptureBuffer

MergeBuffer

Video Mergeand Capture

FPGA(XC3S1000)

(FG456)

to/from AVIB

merged graphics

digital video

from SCNV

from PC CoreBlock_VIME.vsd

Fig. 8.2-8 - Core Board VIME Sub-system

Page 127: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Theory of Operation

BI2201-E 8-9

The VIME sub-system receives image streams from the Embedded PC, the SCON and from external video equipment and merges the streams together, de-interlacing the video input on the way, for display on the main monitor. This merging includes scaling of either SCON or external video for picture-in-picture function. The Embedded PC sets a window size and a transparency color for each of the images in order to define the overlay priority. The VIME system also captures the merged frames, thus allowing the Embedded PC to access screen snapshots or real-time sequences and transfer them to documentation units such as disk drives, DICOM or digital VCRs. It also captures SCON images for the benefit of the 3D application.

Embedded PC:

Embedded PC(ETX Module)

PCI Bridge(PCI9056)(PBGA256)

IEEE1394Link & Phy

Free PMC Slot

PCI Bus

Host Bus

Backplane

AudioCodec

digital audio

analog audioanalog audio

analog audioVideo I/O Board

4 x I2C

Board IDEEPROM

+ Backplane

graphics

analog audio

LAN USB COMSpeaker

Backplane Backplane

1394

CoreBlock_PC.vsd

Video I/O Board

2½"HDD

E-IDE

AVR

to VIME from FFT

Fig. 8.2-9 - Core Board Embedded PC Sub-system The Embedded PC runs the user application and generally controls the entire scanner. It is an off-the-shelf module with an 1400MHz Pentium M processor, chipset, memory, etc. The Embedded PC interfaces physically to the Core board through high density connectors, no cables. A PCI-to-Local Bus bridge interfaces the PC module’s PCI bus to the Core board hardware. The Embedded PC further provides two USB ports and two COM ports for attaching external devices and a parallel port for an internal boundary scan chain. Hubs are provided for increasing the number of USB connections and an additional COM port. An IDE interface connects to a 2.5” HDD, which is also mounted on the Core board. This disk holds the operating system (embedded Windows XP), the firmware configuration files and the ultrasound application files. An audio codec is used as a combined DAC/switch to convert the digital audio from the Doppler FFT and select between FFT audio and audio from an external source, typically a VCR. The audio output goes to a 2W power amplifier in the HUB module. Finally, the module includes an Atmel AVR microcontroller, which serves as an I2C bridge; it translates RS-232 commands from the Embedded PC to four I2C system busses. The AVR may also poll the various temperature sensors in the system and control the PSU fans for optimal noise reduction.

Page 128: Ecograf System Pro Focus 2202(BI2201-E)

Theory of Operation System Pro Focus 2202

8-10 BI2201-E

8.3.4. Analog Video Interface Board, AVIB

The AVIB connects to the Core board and provides the interface for connection of system monitor as well as miscellaneous video equipment. The physical connectors are placed on a separate board, the AVCON (see Section 8.3.5).

VGA-to-TVDigital Video

EncoderAL128

24

VIME(I2C)

2

FieldStoreAL422

Y/C + CVBS

Digital VideoDecoder

VPX3220A

Y/C CVBS

VideoADC

AD9884

RGB

Mux

VIME(Video)

8/16

8/16 24

Core Board Interface

I/O Connectors (AVCON)

Switch

VIME(VGA)

PC(analog VGA)

VideoDAC

2 x VGA

FieldStoreAL422

Board IDEEPROM

GPIORegister

AVIBblock.vsd

AnalogEncoder

RGBS

Fig. 8.2-10 - AVIB Block Diagram

Video DAC: Triple DAC for conversion of digital RGB stream to analog (S)VGA. Digital Video Encoder: Translates (a selected part of) the non-interlaced digital output from the VIME sub-system on the Core board into interlaced RGB for viewing on a TV monitor. Analog Video Encoder: Translates the interlaced RGB output from the Digital Video Encoder into S-video and CVBS for connection of analog VCRs, video printers and low-end TV monitors. Digital Video Decoder: Converts incoming S-video and CVBS signals to digital component video as input to the VIME sub-system on the Core board. Video ADC: Triple ADC for conversion of analog RGB input to digital component video as input to the VIME sub-system on the Core board.

8.3.5. Analog Video Connector Board, AVCON The AVCON board contains all connectors for audio and video I/O. It interfaces to the AVIB. There are no active components on the board.

Page 129: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 Theory of Operation

BI2201-E 8-11

8.3.6. USB Hub, HUB The USB hub board provides the two front USB connectors as well as the USB interface to the CD-RW unit, see below. A single USB connects the Backplane to the hub board. A separate power cable is also provided which includes 12V for the CD-RW unit; the 500mA limit of USB 5V power is insufficient. Due to the placement of the internal speakers, the Hub Board also contains a 2W class-D audio amplifier. In this way, the speaker wires can be kept very short, and an output filter is not required.

8.3.7. External Devices External devices are part of the scanner and can be mounted directly on the central scanner unit (also known as the Engine) containing the main PCBs and Power Supply. However, they may also be placed away from it at places convenient to the user. Keyboard Assembly: Assembly consisting of an alphanumeric keyboard, a pointing device (track ball), and an ultrasound control panel with dedicated keys, gain dials and TGC slide potentiometers. A microcontroller handles the key scanning, gain dial decoding and A/D conversion of the TGC pots. A USB hub in the keyboard connects the various modules and provides a single USB connection to the Embedded PC. Monitor: A standard 15" SVGA CRT or 17” flat panel monitor. Mass Storage: CD-RW with USB interface. Power supply (non removable): The power supply is a SMPS type. Four isolated outlets for external equipment are provided. One is allocated for the monitor and one for a built-in video printer. The unit supplies 3.3VD, 5VD, 12VD, ±5VA, ±12VA, ±HVMUX, ±HV. The HV output for the transmitters is not programmable. A standard ATX handshake allows the Windows© powered PC to control start-up and shut-down. The unit also contains circuitry for controlling the two fans below the card cage and its own fan.

8.3.8. Optional Devices Other devices may be connected. An accessory is a device, which the customer can purchase and use without further requirements. It attaches to an external connector. A field option is a device or module, which a customer purchases after the scanner, but it may only be installed inside the scanner by service personnel. Foot Switch (accessory): Existing foot switch UA1215 or a similar one with three pedals (B-K Number TBD), which is recommended for use with the 3D application. The foot switch is attached to the Backplane (rear panel). Some pedal functions are user programmable. Palm Control Unit (accessory): Existing Palm Control Unit UA1270. Connects to the Backplane (COM1 on the rear panel) using existing adapter UA1271. Requires a 5VDC supply of 100mA. Magnetic Tracking Freehand (accessory): Existing MTF Upgrade kit UA0576, can be connected to the Backplane (rear panel) COM1 port.

Page 130: Ecograf System Pro Focus 2202(BI2201-E)

Theory of Operation System Pro Focus 2202

8-12 BI2201-E

Single Element Module, SEM (field option): Existing SEM, ZH0651. A separate module with motor control, transmitter, receiver, TGC and transducer code identification. Connects internally to the SEMIF. SEM Interface, SEMIF (field option): Must be installed with the SEM and MCM options. Contains a 14-bit ADC in order to interface the analog output from the SEM to the digital input port of the VECP and a TGC “splitter” circuit for generating the two TGC control voltages for the SEM. There is also a USB-to-RS232 converter for the MCM. Connects internally to the Backplane. Mover Control Module, MCM (field option): Existing MCM, consisting of modules ZH0752/53/54 or a new compatible single-board solution. Controls the external mover units UA0550/51/52. Connects internally to the SEMIF. Whenever both SEM and MCM are installed, the control signals from the MCM are connected to the transducer LEMO connector in the SEM so that seamless integration with 205x transducers is possible. Additionally, it will be possible to connect USB based drive types to the external USB ports, provided that drivers are included in the embedded operating system.

Page 131: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 System Interconnections

BI2201-E 9-1

Section 9

SYSTEM INTERCONNECTIONS CONTENTS PAGE

9.1. Introduction.............................................................................................................................................. 2 9.1.1. System Interconnection Diagram........................................................................................................3 9.2. Multi-pin Connections (external) ........................................................................................................... 4 9.2.1. Footswitch ...........................................................................................................................................4 9.2.2. Monitor ................................................................................................................................................4 9.2.3. USB .....................................................................................................................................................4 9.2.4. LAN .....................................................................................................................................................4 9.2.5. S-VHS .................................................................................................................................................4 9.3. Power Supply Connectors ..................................................................................................................... 5

Page 132: Ecograf System Pro Focus 2202(BI2201-E)

System Interconnections System Pro Focus 2202

9-2 BI2201-E

9.1. Introduction By showing the cables between PCB's, and external cable connections this section aims to help you keep track of important signals internally in the 2202 and to other equipment.

Page 133: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 System Interconnections

BI2201-E 9-3

9.1.1. System Interconnection Diagram FE

BP

Cor

e

DB

F

SE

MIF

SE

M

MC

M

PS

U

HU

BE

CG

2 x

US

B

10/1

00 L

AN

2 x

IEE

E13

94

CO

M1

Arr

ay X

duce

r

SE

Xdu

cer

Mov

er U

nit

Arr

ay X

duce

r

Arr

ay X

duce

r

2 x SVGA

RGBS

S-video

3 x CVBS

2 x Audio

RGBS

S-video

CVBS

2 x Audio

SVGA

2 x Doc Trig

O u

t p

u t s

I n p

u t

s

2202 Signal Interconnect Diagram

InterfaceDiagram.vsd

Key

boar

d

2 x

FAN

2 x

Spe

aker

ECG

I2C + SPI Bus

FEou

t

PS

U I2 C

PS

U H

ands

hake

Tach

Sta

tus

I2 C +

SP

I Bus

SP

I Bus

Con

trol B

usR

eal-t

ime

Ctrl

SE

M O

utpu

t

RS

-232

Fsw

PW

RB

utto

nP

WR

BTN

AV

IBA

VC

ON

Vid

eo

2 x USB

CD

-RW

US

B

SEM: Single Element ModuleMCM: 3D Mover Control ModuleSEMIF: Single Element Module InterfaceFE: Front End BoardDBF: Digital Beamformer Board

BP: Backplane

PSU: Power Supply UnitAVIB: Analog Video Interface Board

CD-RW: CD-RW Disk Drive

ECG: ECG Module

Units:

Core: Digital Back End & RTC Board

AVCON: AV Connector Board

HUB: USB Hub Board

FE Real-time ControlFE Control Bus

DBF Control Bus

DBF FPGA ConfigurationDBF I2C Bus

DBF Real-time Control

Beam SumTAP

TAP

ECG Control & Data

SEMIF Control Bus

HUB USB

SEMIF Real-time Control

SEMIF Beam Data

SEMIF USB

Foot

Sw

itch

AVIB I2C BusLight Pen

AudioVideo

Aud

ioD

oc T

rigLi

ght P

en

Light Pen

IDE

Em

bedd

ed P

C2.

5" D

isk

Driv

eHUB Audio

Fig. 9.1-1 Signal Interconnection Diagram

Page 134: Ecograf System Pro Focus 2202(BI2201-E)

System Interconnections System Pro Focus 2202

9-4 BI2201-E

9.2. Multi-pin Connections (external) The following section describes the multi-pin connectors situated on the rear of the 2202.

9.2.1. Footswitch

9.2.2. Monitor This is a standard 15-pin DSUB socket.

9.2.3. USB This is a standard USB Type A socket.

9.2.4. LAN This is a standard RJ45 10/100 Mbps socket.

9.2.5. S-VHS This is a standard S-VHS socket.

Page 135: Ecograf System Pro Focus 2202(BI2201-E)

System Pro Focus 2202 System Interconnections

BI2201-E 9-5

9.3. Power Supply Connectors

The DC outputs and control signals are delivered from the power supply as 2 sets of wires with a 16-pin connector (J1) and a 20-pin connector (J2) mounted. (Mini-Fit, Jr.™ type 5557 with High Current Terminals 44476. Output Header for external fans is Molex KK® Header type 6410. Length of the wire must be so, that at least 120mm of the connectors are outside the power supply box. J1 - Mini-Fit, Jr.™ 16-pin Receptacle:

Pin Name Colour Size AWG

UL Style

Colour Name Pin

1 +5VA Red 16 1007 Black AGND 9 2 +5VA Red 16 1007 Black AGND 10 3 -5VA White 16 1007 Black AGND 11 4 +12VA Yellow 16 1007 Black AGND 12 5 -12VA Blue 16 1007 Black AGND 13 6 AGND Black 16 1007 Black AGND 14 7 +100VA Red 20 1007 White -100VA 15 8 +80VA Yellow 20 1007 Gray -80VA 16

J2 - Mini-Fit, Jr.™ 20-pin Receptacle: Pin Name Colour Size

AWG UL Style

Size AWG

Colour Name Pin

1 BPPSUSCK Brown 20 1007 20 Brown BPPSUSDA 11 2 BPPWR_OK Gray 20 1007 16 Black GND 12 3 BP5VSB Brown 20 1007 20 Brown BPPS_ON_L 13 4 12VD Yellow 16 1007 16 Black GND 14 5 3V3D Orange 16 1007 20 Brown BPSTOP_L 15 6 3V3D Orange 16 1007 16 Black GND 16 7 3V3D Orange 16 1007 16 Black GND 17 8 5VD Red 16 1007 16 Black GND 18 9 5VD Red 16 1007 16 Black GND 19 10 5VD Red 16 1007 16 Black GND 20

Page 136: Ecograf System Pro Focus 2202(BI2201-E)

System Interconnections System Pro Focus 2202

9-6 BI2201-E

This page is left blank intentionally