easi ecg monitoring _final

19
¡Error! No se pueden crear objetos modificando códigos de campo.

Upload: mariano-villano

Post on 24-Nov-2015

48 views

Category:

Documents


1 download

TRANSCRIPT

  • Error! No se pueden crear objetos modificando cdigos de campo.

  • Centre for Clinical Nursing Research Deakin University Epworth Hospital

    EASI ECG Monitoring _Final

    Author: Bernice Redley 2

    Table of Contents

    What is EASI ECG Monitoring? ................................................................................3

    Advantages of the EASI ECG monitoring system ...................................................4

    Disadvantages of the EASI ECG monitoring System ..............................................4

    Comparisons between the EASI and Mason-Likar 12 lead ECG systems ................5

    Summary of the Research ...........................................................................................6

    Methods..............................................................................................................6

    Detection of acute ischaemia and Myocardial Infarction (MI) .............................6

    Lead amplitudes .................................................................................................7

    Conduction and rhythm disturbances ..................................................................7

    Misdiagnosis ......................................................................................................7

    Limitations .............................................................................................................8

    Summary ................................................................................................................8

    APPENDIX 1. Summary Table by Author ..............................................................9

    APPENDIX 2. Summary Table by Parameter ....................................................... 13

    References ............................................................................................................ 19

  • Centre for Clinical Nursing Research Deakin University Epworth Hospital

    EASI ECG Monitoring _Final

    Author: Bernice Redley 3

    What is EASI ECG Monitoring? EASI is a method of continuous electrocardiogram (ECG) monitoring that is an

    alternative to both the commonly used 5-electrode lead bedside monitoring system

    and the traditional 10-electrode Mason-Likar 12lead ECG system. The EASI lead configuration enables continuous 12 lead ECG ambulatory monitoring using only 5

    electrodes. The EASI 12 lead ECG is derived from this reduced lead set using a

    method described by Dower et al. [1-3].

    The EASI 12-lead ECG is derived from a set of 5 leads; 4 recording electrodes and

    one grounding electrode. The placement of these leads is as follows [4].

    E: Lower extreme of the sternum

    A: Left mid-axillary line, same transverse line as E

    S: Sternal manubrium

    I: Right mid-axillary line, same transverse line as E

    G: Fifth electrode is the ground and can be placed anywhere on the torso

    Diagram of EASI electrode placement

    [4 p. 180]

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 4

    Advantages of the EASI ECG monitoring system

    Easy to apply. The EASI is a 5 electrode system that provides a 12-lead ECG, making it easier to rapidly and accurately apply the leads in stressful situations [4].

    Time saving for staff. As a continuous 12-lead ECG is produced, there is no need to perform traditional intermittent 12-lead ECGs.

    Continuous 12-lead ECG. The reduced set of leads provides continuous 12-lead ECG monitoring as opposed to a snapshot 10-15 second view ECG with the

    traditional ECG [5].

    Better patient comfort. The reduced set of EASI leads is less bulky than a 10-electrode Mason-Likar system, improving patient comfort, simplifying application

    and maintenance for continuous monitoring [5].

    Less waveform interference. With less leads, the EASI system is less susceptible patient movement, signal interference or myeloelectric noise making it more

    suitable and sensitive (than the 10 electrode system) for continuous ambulatory

    ECG monitoring [6-8].

    Better serial comparisons. EASI derived ECGs have better reproducibility for serial comparisons [6, 8, 9].

    More leads viewed continuously. EASI provides a continuous 12-lead ECG allowing detection of transient myocardial ischaemia across multiple leads, while

    the more common 5-lead ambulatory monitoring system does not allow monitoring

    of all the precordial leads at the same time (usually lead II and V1 only)[5, 10-12].

    Disadvantages of the EASI ECG monitoring System

    Unfamiliar to staff. The EASI system is unfamiliar to clinicians so training in this system may be required [13]

    Differs to traditional method. Some leads from the EASI derived ECG have been found to differ slightly to the standard ECG, requiring consideration during

    interpretation. These differences are outlined below.

    Placement error. Lead placement error may lead to discrepancies [4].

    Less sensitive in detecting some conditions. Due to lower amplitudes in some leads, it is difficult to detect chamber enlargement (atrial and ventricular hypertrophy )

    with EASI ECG [7, 12], though similar problems also exist with the standard ECG.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 5

    Comparisons between the EASI and Mason-Likar 12 lead ECG systems

    Duration and amplitude of P, QRS, ST and T waves may differ between the EASI derived and standard ECG systems [5, 7, 9, 14], with the EASI system

    demonstrating less amplitude. It is important that this is considered during analysis

    and serial comparisons are made using the same system.

    Frontal plane transition zone at V2-V3 in EASI and V3-V4 in traditional [7].

    The correlation between EASI ECG and the standard ECG complex interpretations reached less than 90% agreement in some leads (particularly III, aVL, aVF, V4 and

    V5) [9].

    High levels of correlation have been found between the EASI and the standard ECG complexes in leads I, aVR, V1, V2, V3, and V6 [9].

    Greater than 90% agreement for detection of small (100 V) ST segment changes indicating acute ischaemia [12, 13]

    Greater than 90% agreement for detection of prior myocardial infarction (Q wave, isolated ST-T) [9, 12, 13].

    Similar rates of false diagnosis or misinterpretation of ECGs by clinicians with both systems [12, 13]

    100% agreement for rhythm identification [7, 12].

    95-100% agreement for conduction problems (Bundle branch blocks, fascicular blocks) [7, 12].

    Differences in axis of up to 30 degrees between the two systems [7, 12].

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 6

    Summary of the Research Several studies have compared derived EASI 12-lead ECGs with the standard Mason-

    Likar 12 lead ECG. Overall the two systems are similar for many of the parameters

    examined, but some differences also exist. A summary of these studies is provided

    below.

    Methods

    The strongest study design simultaneously collected the EASI derived and standard 12

    lead ECGs continuously across a large population of subjects (540 patients) from

    acute clinical areas (Emergency and cardiac catheter), then used a computer analysis

    program to compare the recordings [12]. The authors then used a clinical expert to

    examine the accuracy of the ECG interpretations in collaboration with clinical data

    such as cardiac enzymes, echocardiography, patient history and clinical examination.

    Use of computer analysis and simultaneous data collection were strengths as slightly

    greater variations within and between EASI and standard ECG recordings have been

    associated with human interpretations [7]; or ECGs that were not collected

    simultaneously [4, 14]. Variations have also been noted between studies that used data

    from healthy subjects in laboratory settings [4, 6] and those conducted in real clinical

    circumstances with patients with variable diagnoses requiring cardiac monitoring [7,

    9, 12-14]. Studies using patient population provide more useful clinically relevant

    information. Few studies incorporated clinical findings to confirm the diagnosis

    derived from the ECG [12, 14].

    Detection of acute ischaemia and Myocardial Infarction (MI)

    The rate of agreement between EASI ECG and the standard 12-lead ECG for the

    detection of acute ischaemia was high in the majority of reviewed studies [9, 11-16].

    Some ECG differences were detected between different types of ischemia. Both

    inflation induced ischaemia during cardiac catheterisation and spontaneous ischaemic

    events have been examined.

    Rautaharju et al. [14] reported 85% agreement between EASI and standard ECGs in

    the detection of inflation induced ischaemia during cardiac catheterisation. These

    authors used 100 V (which is less than usually considered clinically significant) as

    the ST segment threshold. The use of independent electro-cardiographers reading the

    ECGs may have contributed to the lower agreement for ischaemic changes in this

    study. Studies that used computer program analysis reported much higher agreement

    (>99%) between EASI derived and standard ECGs for the detection of ischaemia [12,

    13] than studies that used clinical experts with or without computer assistance [7, 14].

    Drew et al. [12] reported the EASI system accurately detected 93% of acute transient

    myocardial ischaemic events, while the leads routinely used for bedside monitoring

    (Leads II and V1) detected only 42%. In this study, discrepancies between the EASI

    and standards ECG were found in 17 of 238 ST events. Review by an expert clinician

    revealed the EASI system was less sensitive in detecting low voltage ST segment

    changes. None of these 17 discrepancy events were associated with chest pain, none

    involved large changes (>200 V) and none involved the need for clinical

    intervention. These findings suggest that these differences were not clinically

    significant.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 7

    Similar findings were experienced in the detection of ECG changes suggesting old

    MI. Comparisons of EASI derived and standard ECGs revealed 95% agreement for

    anterior and 92% agreement for inferior Q-wave MI in a large study sample [12]. The

    EASI system had greater sensitivity than the standard system in detecting these

    changes (59% vs 55%) but slightly less specificity (95% vs 99%).

    Lead amplitudes

    Differences in the amplitudes of several leads have been consistent findings in studies

    comparing the EASI derived ECG with the standard 12-lead. Most commonly

    discrepancies were detected in V3-V4 [4, 7, 9, 13] or isolated ST-T segment changes

    of low amplitude [12, 14]. Only minor differences of 14-30 degrees in frontal plane

    axis were detected [7, 12]. These small differences are not usually clinically

    significant.

    Low amplitudes were also associated with low sensitivity in detecting chamber

    enlargements using the both methods of ECG acquisition. These difficulties were

    similar with both the EASI derived and standard ECG, except in the cases of left

    ventricular hypertrophy where the EASI derived ECG had higher specificity than the

    standard ECG [12].

    EASI derived ECG have shown lower amplitude Q waves in V6, smaller amplitude R

    wave in V1, V2 and V4, and less ST amplitude in V4 and V5 [14]. Horacek [9]

    compared ECGs collected from clinical settings revealing less than a 90% correlation

    between EASI derived and standard ECG in leads III, aVL for normal and post MI

    ECGs, with Q waves, non-Q waves and history of VT.

    Conduction and rhythm disturbances

    Comparisons between the EASI derived and standard ECGs consistently demonstrate

    high levels of agreement (95%-100%) in the detection of conduction problems such as

    pre-excitation patterns, bundle branch blocks, fascicular blocks [7, 12, 16] and rhythm

    disturbances such as VT and SVT [9, 12, 16]. Minor deviations of 0-2ms in the

    cardiac intervals for the PR, QRS, QT and QTC have been detected. While the

    observed difference of 1ms in QRS was statistically significant it was not clinically

    significant [12]. Again the clinical significance of these findings must be questioned.

    Misdiagnosis

    Similar rates of misdiagnosis have been reported for the two types of ECGs with the

    majority related to old MI changes. Drew et al. [7] reported two false positives and

    four false negative in EASI derived ECGs from patients with prior myocardial

    ischaemia, while six false negatives were reported from the standard ECGs from

    patients with a prior MI. Similarly Rautaharju [14] reported 10% of ECGs had

    significant differences between the EASI derived and standard method with 3.4%

    involving Q-wave MI or ST changes in lead II. However, in this study none of the

    new changes were identified as old, and none of the Q-wave MIs were missed using the EASI derived ECG for interpretation. Drew [13] found only one false negative

    from 207 derived ECGs and in her 1999 study, two ECGs were misdiagnosed using

    both the EASI derived and standard ECG methods.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 8

    Limitations

    There are several variables to consider that pose limitations to these study findings.

    The clinical significance of these limitations must be considered when deciding about

    the appropriateness of the EASI methods for continuous ECG monitoring.

    Various computer programs (Phillips, Marquette, Montana) and equipment were used

    to collect ECG data across the studies. The types of electrode, the method used for

    placement of ECG electrodes, the expertise of the person applying the electrodes and

    the timing of data collected are all independent variables not addressed in many

    studies. It has been established that slight variations in the placement of the

    electrodes may increase error when making ECG comparisons [8]. The influence of

    these factors on the quality of the recordings and subsequent comparisons is unknown.

    The clinical conditions under which data were collected also varied. Laboratory

    conditions and healthy subjects differed to clinical conditions and ill patients. Finally,

    few studies confirmed the ECG diagnosis using other clinical diagnostic criteria,

    suggesting that the error in patient diagnosis may have been higher with both methods

    of ECG data acquisition.

    Summary

    EASI derived ECGs are comparable to the standard ECG for the diagnosis of wide

    complex tachycardias [9, 12, 16] and myocardial ischaemia [9, 11-16]. The

    reproducibility of the EASI derived ECG makes it more useful in serial comparisons

    over time. The EASI derived system for continuous ambulatory monitoring has

    advantages over the standard 5-lead monitoring systems and continuous 12 lead

    monitoring systems with the benefits of both. The EASI system allows continuous 12

    lead monitoring instead of the limited II and V1 capability of the traditional 5-

    electrode method. The EASI system is less susceptible to noise and interference than

    the bulky 10-electrode 12-lead system. These benefits enable significant

    improvements in the detection of spontaneous ischemia during continuous ambulatory

    monitoring.

    While there are differences between ECG data collected using the EASI derived and

    standard 12-lead ECG methods, the clinical significance of these changes is

    questionable. The lower amplitude of some leads in the EASI derived ECG may be

    overcome by adjusting the thresholds used to determine clinical significance and

    training staff in interpretation. Similar limitations in terms of clinical diagnosis and

    variation in interpretation by clinicians affect both methods of 12-lead ECG

    collection.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 9

    APPENDIX 1. Summary Table by Author

    AUTHOR TITLE STUDY DESIGN Number of

    Comparisons

    FINDINGS

    Drew [12] Accuracy of the

    EASI 12 lead ECG

    compared to the

    standard 12 lead

    ECG for

    diagnosing cardiac

    abnormalities

    Two methods of

    continuous ECG

    monitoring

    simultaneously from

    ED admission or from

    catheter based

    intervention. Computer

    analysis program used

    to compare ECGs.

    Disagreements

    examined and diagnosis

    made by specialist and

    clinical and echo

    criteria.

    All ED patients over 2

    year period. 540

    patients enrolled, 426

    with acute coronary

    syndromes. Some not

    diagnosed with acute

    ischaemia and ruled

    out.

    EASI and standard ECG compared for cardiac

    rhythm, cardiac intervals, QRS axis, chamber

    enlargement-hypertrophy, BBB and fascicular

    blocks and prior MI. 100% agreement on

    rhythm identification 84-89% agreement on

    chamber enlargement, 997% agreement on R &

    L BBB, 97-98% agreement on fascicular

    blocks, and 92-95% agreement on prior MI.

    Drew [13] Comparison of

    standard and

    derived 12 lead

    ECG for diagnosis

    of coronary

    angioplasty

    induced myocardial

    ischaemia

    Compared standard and

    derived continuous

    ECG

    Threshold of >1 lead of

    >100 V, 80 ms post j

    point used to define

    ischaemia

    207 patients with timed

    measurements (derived

    analysed every 20 secs,

    standard continuously),

    151 during procedure

    Agreement regarding ischaemia in 150 of 151

    patients with both methods.

    Different time ratio for measurement may have

    led to error in one care but may also have led to

    agreement in others.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 10

    AUTHOR TITLE STUDY DESIGN Number of

    Comparisons

    FINDINGS

    Drew [16] Comparison of a

    new reduced lead

    set ECG with

    standard ECG for

    diagnosing cardiac

    arrhythmias and

    myocardial

    ischaemia

    Compared EASI with

    Interpolated standard

    limb leads and V1 and

    V5

    649 patients (CP in ED

    = 509 and Tachycardia

    in EP Lab = 140).

    Identical for BBB, LAH, RVH, prior MI. 99.2%

    agreement of ischaemia in the ED. Appears to

    have higher agreement than EASI model.

    Applied to patients by expert.

    Drew [7] Comparison of a

    vectorcardiography

    derived 12-lead

    ECG with the

    conventional ECG

    during wide QRS

    complex

    tachycardia, and its

    potential

    application for

    continuous bedside

    monitoring.

    Compare during EP

    studies. Simultaneous

    recordings and two

    expert investigators

    independently

    compared using

    standard criteria

    including QRS pattern,

    sequence, width,

    morphology and

    voltages.

    64 episodes of wide

    QRS complex

    tachycardia in 49

    patients

    9 pts had LVH, only 6 with evidence on derived

    ECG. 6 pts with WPW identical delta wave

    morphology, 45 pts with prior MI 92%

    agreement Lower voltages present in V3 and

    V4 of derived ECG. QRS morphology

    dissimilar in 35% of pts in V3 and 51% in V4.

    (early transition of the RV rS to the LV qR in

    derived). Transitional zone generally appeared

    in V3 or V4 or the ECG appeared in V2-V3 of

    the derived ECG. QRS voltage less in the

    precordial leads of the derived ECG affecting

    Dx of LVH. Derived ECG had false positive in

    2 cases prior MI in young person with no

    history, 2 false negative of inferior MI.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 11

    AUTHOR TITLE STUDY DESIGN Number of

    Comparisons

    FINDINGS

    Horacek [9] Diagnostic

    accuracy of derived

    compared to

    standard 12-lead

    ECG

    290 normal subjects

    and 497 with prior MI

    (36 with non Q, 282 Q-

    wave, 179 with history

    of VT)

    V6 best derived lead with highest agreement

    closely followed by V1 and V2. V3-V5 not

    reproduced well in the MI subgroups as they are

    in normal subjects. Worse as the structural

    damage to the myocardium increases the

    diagnostic performance markedly increases.

    Rautaharju [14] Comparability of

    12-lead ECGs

    derived from EASI

    leads with standard

    12-lead ECG in

    classification of

    AMI and old MI

    Read by two

    independent expert

    ECG readers and

    analysis program.

    ECGs printed side by

    side in random order.

    Readers blinded.

    Coding forms used to

    classify the ECGs

    40 patients recorded

    prior to and at peak

    inflation during PCTA,

    382 with old MI ECGs

    472 with non MI.

    No significant differences between ECG

    readers in acute ischaemia. Suggest need for

    modified prior MI criteria and ST thresholds for

    AMI specific for EASI. Significant differences

    within readers and the program

    10% with clinically significant differences,

    3.4% classed as q-MI and 11 as isolated ST-T

    evolution

    Sejersten [4] The relative

    accuracies of ECG

    precordial lead

    waveforms derived

    from EASI leads

    and those acquired

    from paramedic

    applied standard

    leads

    Compared waveforms

    from Gold Standard ECG with paramedic

    applied and EASI

    derived ECGs. First

    acquired from

    paramedic the two

    simultaneously by

    technician. Difference

    threshold of 10 V.

    20 paramedics

    collected data on each

    other. Paired to act as

    experimental technician

    and study participant. 3

    ECGs on each. 720

    comparisons of

    precordial leads.

    EASI and paramedic waveforms were equally

    accurate in 47%, paramedic more accurate in

    31% and EASI in 22% when compared to Gold

    Standard. Significant difference in paramedic

    placement of leads mean 30mm misplacement.

    Review of discrepancies revealed that EASI

    more accurate for ST changes in V4 & V5. and

    T in V6. Paramedic more accurate Q amplitude

    in V6, R amplitude in V1, V2, V4 and ST

    deviation in V1

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 12

    AUTHOR TITLE STUDY DESIGN Number of

    Comparisons

    FINDINGS

    Welinder [6] Comparison of

    signal quality

    between EASI and

    Mason-Likar 12-

    Lead ECG during

    physical activity

    Compared baseline

    wander and myoelectric

    noise amplitudes of

    EASI and traditional.

    Simultaneous

    recordings. Inter-rater

    not addressed. Range

    thresholds (?valid)

    established for noise.

    20 healthy volunteers

    with simultaneous

    measurements

    EASI winner significantly more often in treadmill and supine to L in precordial.

    Traditional winner significantly more often in Supine to R limb and precordial leads

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 13

    APPENDIX 2. Summary Table by Parameter

    Parameter

    Examined

    Drew (1992) Rautaharju,

    (2002)

    Welinder,

    (2004)

    Drew (1997) Sejersten,

    (2003)

    Drew (1999) Horacek

    (2000)

    Acute

    Ischaemia

    Inflation

    induced

    ischaemia using

    100 V as

    threshold 85%

    agreement,

    27/40 positive

    and 7/40

    negative

    3.4%

    differences in

    New Q MI

    6.1% difference

    in ST-T

    segments

    99% agreement

    in Cx artery

    peak ST in V3

    for derived and

    V3 or V4 for

    standard

    138 patients with 238 ST

    events. 26 had acute ST

    elevation patterns. 63

    angioplasty induced

    ischaemia and 150

    spontaneous ischaemia. 89%

    agreement, the (14 standard

    and 3 EASI) events with

    discrepancy were not

    associated with chest pain

    and none involved large

    changes (>200 V). EASI

    detected 93% of ischaemic

    events while the leads

    routinely used for bedside

    monitoring (II and V1)

    detected 42%.

    EASI low amplitude ST

    changes ? clinical

    significance, none resulted

    in clinical interventions or

    poor outcomes..

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 14

    Parameter

    Examined

    Drew (1992) Rautaharju,

    (2002)

    Welinder,

    (2004)

    Drew (1997) Sejersten, (2003) Drew (1999) Horacek

    (2000)

    Prior MI (Q-

    wave)

    92%

    agreement

    85% agreement

    with computer,

    57% agreement

    for isolated ST-

    T in old MI

    Agreement 95% for

    anterior and 92% for

    inferior. In prior

    inferior MI, EASI had

    greater sensitivity than

    standard (59% vs 55%)

    but lower specificity

    (95% vs 99%)

    Correlations

    between

    EASI and

    standard less

    than 90% in

    leads II, III,

    aVL, aVF,

    V4 and V5

    for q wave

    MI.

    Normal ECG 80% agreement

    with computer

    98%

    agreement

    Similar deviations

    between paramedic

    and EASI.

    Lead

    differences

    Transition

    zone V2-V3

    in EASI and

    V3-V4 in

    traditional

    EASI-Gold

    compared with

    Paramedic-Gold

    revealed significant

    differences; inc Q

    amp in V6, smaller

    R amp in V1, V2

    and V4. Less ST in

    V1, inc ST in V4

    and V5 and inc T

    amp in V6

    Correlations

    less than

    90% in III,

    aVL for

    normal and

    post MI

    patients with

    no Q, Q and

    VT

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 15

    Parameter

    Examined

    Drew (1992) Rautaharju,

    (2002)

    Welinder,

    (2004)

    Drew (1997) Sejersten, (2003) Drew (1999) Horacek

    (2000)

    Axis Frontal plane

    Axis in

    Precordial

    leads only

    Difference 19

    deg. 95% CI

    of 14-24 deg.

    Mean differences in P

    wave and QRS and T-

    wave axis were 21

    degrees, t-wave axis

    differed more between

    the two ECG types

    mean 30 degrees.

    Conduction 100%

    agreement on

    BBB and

    fascicular

    blocks

    Agreement for BBB

    and fascicular blocks

    ranged from 95-99%.

    EASI closer to expert

    in RBBB.

    False

    diagnosis

    EASI False

    positive in 2

    and false

    negative in 4

    with prior MI

    and traditional

    False negative

    in 6 pts with

    prior MI

    10% clinically

    significant

    differences B/W

    EASI and

    traditional 3.4%

    q-MI and 11 ST

    changes. None

    of normal

    classed as old

    MI and none of

    Q-wave missed.

    No false

    positives and

    1 false

    negative with

    derived ECG

    2 misdiagnosed by

    both the EASI and

    standard 12 lead due to

    low amplitude p

    waves.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 16

    Parameter

    Examined

    Drew (1992) Rautaharju,

    (2002)

    Welinder,

    (2004)

    Drew (1997) Sejersten,

    (2003)

    Drew (1999) Horacek

    (2000)

    Cardiac

    rhythm

    100% agreement

    in diagnosis and

    misdiagnosis for

    Wide complex

    tachycardia

    Difference in axis

    17 deg with CI

    95% 12-22 deg.

    SVT Vs VT

    criteria agreement

    in V1, V2, V6

    100% agreement Correlations

    less than

    90% for

    leads II, III,

    aVL, aVF,

    V3, V4, V5

    between

    EASI and

    patients with

    VT.

    Chamber

    enlargement

    Lower V3-V4

    voltages, leading

    to inability to

    detect LVH in 1/3

    of patients with

    wide complex

    tachycardia 39%

    in V3 and 55% in

    V4

    177 evaluated for atrial

    enlargement using Echo.

    RAH in 17 patients and

    LAH in 46 patients, RVH

    in 15 and LVH in 68.

    Agreement was between

    84-98%. Both methods

    had low sensitivity but

    high specificity. For LVH

    EASI had higher

    specificity than standard

    (98% vs 88% respectively)

    Both methods insensitive

    to chamber enlargement.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 17

    Parameter

    Examined

    Drew (1992) Rautaharju,

    (2002)

    Welinder,

    (2004)

    Drew (1997) Sejersten,

    (2003)

    Drew (1999) Horacek

    (2000)

    Others Ventricular pre-

    excitation

    patterns 100%

    agreement

    Cardiac intervals for PR,

    QRS, QT and QTC varied

    by 0-2ms. The observed

    difference of 1ms in QRS

    was statistically significant

    but not clinically

    significant.

    Artifact Less muscle

    artifact after

    DCCR and on

    arm movement.

    Similar

    baseline

    wander,

    EASI less

    noise for

    treadmill

    and supine

    to L.

    No MI 80%

    agreement

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 18

    Other Variables to consider:

    Parameter

    Examined

    Drew (1992) Rautaharju,

    (2002)

    Welinder,

    (2004)

    Drew (1997) Sejersten,

    (2003)

    Drew (1999) Horacek

    (2000)

    Electrodes

    used

    Not identified Not identified Identified Not identified No specified

    Method of

    deriving

    traditional

    ECG

    (machine

    used)

    Marquette

    computer

    program.

    Simultaneous

    measurement

    Standard

    measure

    ECGs

    collected in

    time sequence

    Identified

    Simultaneou

    s collection

    of ECGs

    Described

    ECGs

    collected

    simultaneousl

    y during

    procedure

    Identified

    Three

    independent

    recordings on

    same subject.

    Identified

    Simultaneous monitoring

    Simultaneou

    s recording

    supine for 15

    seconds

    Application

    of electrodes

    by whom/

    position

    In the ED and

    PCTA procedure

    Not identified

    Method of

    application

    Not identified,

    position id.

    Identified I V4 used in

    study and

    EASI for

    latter

    Paramedic

    and expert

    (Traditional

    and EASI)

    Position

    described

    Not identified, presumed

    to be in the ED/ cath lab

    Nor

    specified.

    Setting ED & EP studies, PCTA, AMI

    pts with

    enzyme ?

    from location

    suggests

    clinical

    situations.

    Laboratory

    Healthy

    participants.

    PCTA. Experimental

    conditions at

    conference

    Healthy

    participants.

    ED and catheter lab Not specified

    but presumed

    to be a

    clinical

    environment

    and clinical

    diagnostic

    criteria used.

  • EASI ECG Monitoring _Final

    Author: Bernice Redley 19

    References

    1. Dower, G.E., The ECGD: a derivation of the ECG from VCG leads. Journal of

    Electrocardiology, 1984. 17(2): p. 189-91.

    2. Dower, G.E., H.B. Machado, and J.A. Osborne, On deriving the

    electrocardiogram from vectorcadiographic leads. Clinical Cardiology, 1980.

    3(2): p. 87-95.

    3. Dower, G.E., et al., Deriving the 12-lead electrocardiogram from four (EASI)

    electrodes. Journal of Electrocardiology, 1988. 21 Suppl: p. S182-7.

    4. Sejersten, M., et al., The relative accuracies of ECG precordial lead

    waveforms derived from EASI leads and those acquired from paramedic

    applied standard leads. Journal of Electrocardiology, 2003. 36(3): p. 179-85.

    5. Drew, B.J., et al., Practice standards for electrographic monitoring in hospital

    settings. Circulation, 2004. 110: p. 2721-2746.

    6. Welinder, A., et al., Comparison of signal quality between EASI and Mason-

    Likar 12-lead electrocardiograms during physical activity. American Journal

    of Critical Care, 2004. 13(3): p. 228-234.

    7. Drew, B.J., M.M. Scheinman, and G.T. Evans, Jr., Comparison of a

    vectorcardiographically derived 12-lead electrocardiogram with the

    conventional electrocardiogram during wide QRS complex tachycardia, and

    its potential application for continuous bedside monitoring. American Journal

    of Cardiology, 1992. 69(6): p. 612-8.

    8. Adams, M.G. and B.J. Drew, Body position effects on the ECG: implication

    for ischemia monitoring. Journal of Electrocardiology, 1997. 30(4): p. 285-91.

    9. Horacek, B.M., et al. Diagnostic accuracy of derived compared to standard

    12-lead electrocardiograms. in International Society for Computerized

    Cardiology. 2000. Yosemite, California: Journal of Electrocardiology.

    10. Drew, B.J., et al., Value of a derived 12-lead ECG for detecting transient

    myocardial ischemia. Journal of Electrocardiology, 1995. 28 Suppl: p. 211.

    11. Drew, B.J. and B. Ide, Diagnosing ischemia from the bedside monitor.

    Prognostic Cardiovascular Nursing, 1996. 11(1): p. 45-6.

    12. Drew, B.J., et al., Accuracy of the EASI 12-lead electrocardiogram compares

    to the standard 12-lead electrocardiogram for diagnosing multiple cardiac

    abnormalities. Journal of Electrocardiology, 1999. 32: p. 38-47.

    13. Drew, B.J., et al., Comparison of standard and derived 12-lead

    electrocardiograms for diagnosis of coronary angioplasty-induced myocardial

    ischemia. American Journal of Cardiology, 1997. 79(5): p. 639-44.

    14. Rautaharju, P.M., et al., Comparability of 12-lead ECGs derived from EASI

    leads with standard 12-lead ECGS in the classification of acute myocardial

    ischemia and old myocardial infarction. Journal of Electrocardiology, 2002.

    35 Suppl: p. 35-9.

    15. Drew, B.J., et al., Derived 12-lead ECG. Comparison with the standard ECG

    during myocardial ischemia and its potential application for continuous ST-

    segment monitoring. Journal of Electrocardiology, 1994. 27 Suppl: p. 249-55.

    16. Drew, B.J., et al., Comparison of a new reduced lead set ECG with the

    standard ECG for diagnosing cardiac arrhythmias and myocardial ischemia.

    Journal of Electrocardiology, 2002. 35 Suppl: p. 13-21.