Implantable MOSFET dosimeter response to 192Ir HDR radiation
Jessica Fagerstrom
University of Wisconsin – Madison, Department of Medical PhysicsMedical Radiation Research Center
North Central Chapter AAPM meetingMayo Clinic, Rochester, MN
April 25, 2008
![Page 2: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/2.jpg)
Dose Verification System®
• Implantable dosimeters can verify the sum contribution of many factors.
• Sicel Technologies, Inc. has developed a compact, implantable MOSFET Dose Verification System® (DVS) (SiO2 active area).
• DVS detectors are surgically implanted in or near irradiated areas of patient.
• Dose is read telemetrically by a handheld antenna, which inductively powers the dosimeter.
• DVS received FDA 510(k) clearance in 2006 for external beam breast and prostate treatments.
• DVS is currently characterized for 6 and 18 MV linear accelerators.
![Page 3: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/3.jpg)
Outline• The goal of this study was to characterize the
DVS response to 192Ir HDR sources– Dose rate dependence– Energy dependence
– Accumulated dose effect
– Rotational angular dependence
– Longitudinal angular dependence
![Page 4: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/4.jpg)
Dose Verification System
Photograph courtesy of Sicel Technologies, Inc.
G. P. Beyer et al. , “An implantable MOSFET dosimeter for the measurement of radiation dose in tissue during cancer therapy,” IEEE Sensors J., Vol. 8, No. 1, 2008.
![Page 5: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/5.jpg)
Virtual Water phantoms: Phantom 1
30 cm
30 cm
15 cm
![Page 6: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/6.jpg)
Virtual Water phantoms: Phantom 2
30 cm
30 cm11 cm
![Page 7: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/7.jpg)
Virtual Water phantoms: Phantom 2
Longitudinal testing Rotational/accumulated dose testing
![Page 8: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/8.jpg)
Dose rate dependence• Dosimeters irradiated in Phantom 1 with 60Co
source• SDD varied to achieve dose rates varying
between 22 cGy/min to 84 cGy/min• Dose rates chosen to cover range of those
delivered by 192Ir in subsequent tests• Delivered dose determined using a NIST-
calibrated Exradin A12 ion chamber inserted into the phantom
• 200 cGy/fraction for nine fractions, 1 fraction/day
![Page 9: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/9.jpg)
Dose rate dependence
![Page 10: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/10.jpg)
Energy dependence• Dosimeters irradiated in Phantom 1 with 60Co
and 192Ir• Fractions alternated between 60Co and 192Ir• 60Co delivered dose determined using a NIST-
calibrated Exradin A12 ion chamber inserted into the phantom
• 192Ir delivered dose was determined using measured SK and TG-43U11 methodology
• 600 cGy/fraction for a total of 10 fractions[1] M.J. Rivard, B.M. Coursey, L.A. DeWerd, W.F. Hanson, M.S. Huq, G.S. Ibbott, M.G. Mitch, R. Nath, and J.F. Williamson, “Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations,” Med. Phys. 31, 633-674 (2004).
![Page 11: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/11.jpg)
Energy dependence
![Page 12: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/12.jpg)
Accumulated dose effect
• Dosimeters irradiated in Phantom 2 with 192Ir• All inserts oriented at 0°• Dosimeters placed at 3 distances
from source:– 3 cm (1150 cGy/fraction)– 4 cm (650 cGy/fraction)– 5 cm (415 cGy/fraction)
• 10 fractions delivered to all detectors, 1 fraction/day for 2 weeks
![Page 13: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/13.jpg)
Accumulated dose effect
ΔVth/Ddelivered = a + b·Dacc + c·Dacc2 + d·Dacc
3
R2>0.997
![Page 14: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/14.jpg)
Rotational angular dependence• Dosimeters irradiated in Phantom 2 with 192Ir• Varied rotational radiation incidence (about the
long axis of DVS)• Dosimeters placed 4 cm from source• 215 cGy/fraction for 10 fractions,
1 fraction/day for 2 weeks• Irradiations performed at 0°, 45°,
90°, 225°, and 270°• Fractions alternated between 0°
and other angles
![Page 15: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/15.jpg)
Rotational angular dependence
![Page 16: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/16.jpg)
Longitudinal angular dependence
• Dosimeters irradiated in Phantom 2 with 192Ir• Varied longitudinal radiation incidence (about
the short axis of DVS)• Dosimeters placed 10 cm from source• 190 cGy/fraction for 10 fractions,
1 fraction/day for 2 weeks• Irradiations performed at 0°, 30°,
60°, 90°, and 270°• Fractions alternated between 0°
and other angles
![Page 17: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/17.jpg)
Longitudinal angular dependence
![Page 18: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/18.jpg)
Uncertainty summaryTest Isotope Sensitivity uncertainty
Dose rate dependence 60Co ±4.6% to ±6.3%60Co ±2.8% to ±5.1%192Ir ±4.1% to ±6.1%
3 cm 192Ir ±3.2% to ±3.5%
4 cm 192Ir ±4.0% to ±7.2%
5 cm 192Ir ±3.3% to ±6.6%
Rotational angular dependence 192Ir ±4.3% to ±7.7%
Longitudinal angular dependence 192Ir ±4.2% to ±5.6%
Accumulated dose effect
Energy dependence
![Page 19: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/19.jpg)
Conclusions• Dose rate changes at 60Co did not have a statistically
significant effect on detectors• DVS was more sensitive to 192Ir than 60Co
– Higher interaction cross section in SiO2 at an average energy of 397 keV vs 1.25 MeV
• Accumulated dose effect can be characterized with third order polynomials
• Little angular dependence except where detector is irradiated with coil/electronics assembly between MOSFETs and incident radiation– Problem minimized if implanted with coil facing away from
nearest catheter
![Page 20: DVS MOSFET dosimeter response to kilovoltage radiationchapter.aapm.org/.../z_meetings/2008-04-25/JessicaFagerstrom2008… · 2008-04-25 · Group No. 43 Report: A revised AAPM protocol](https://reader035.vdocuments.mx/reader035/viewer/2022071102/5fdb4e94db76bc07df5c849d/html5/thumbnails/20.jpg)
Acknowledgements
• John Micka• Larry DeWerd• Sicel Technologies• UW MRRC students and staff• UW ADCL customers