Page 1: VET-120 Diagnostic Imaging

VET-120Diagnostic

Imaging

Page 2: VET-120 Diagnostic Imaging

Course Set-up

5 Lessons (Webinars) 5 Lesson Exams Final Proctored Exam Required Reading

– Diagnostic Imaging Study Guide– Radiography in Veterinary Technology

(Lavin – 4th edition)

Page 3: VET-120 Diagnostic Imaging

Studying for Exams Make copy of Lesson PowerPoints Take excellent notes during Webinars Read & highlight Assigned Readings

in Study Guide & Lavin book Perform Self Checks in Study Guide Learn the Glossary at the beginning of

each Chapter in Lavin book Study Key Points & Review

Questions at end of each Chapter in Lavin book

Page 4: VET-120 Diagnostic Imaging

Lesson 2 – Producing a

Radiographic Image

Page 5: VET-120 Diagnostic Imaging

Glossary Terms – Page 36

Page 6: VET-120 Diagnostic Imaging

The 4 Exposure FactorsMilliamperage (mA)Time (S)Kilovoltage (kVp)Distance (SID)

Page 7: VET-120 Diagnostic Imaging

The 4 Exposure Factors Exposure factors are THE KEY

to taking good quality diagnostic x-rays

Definition – Machine settings that a technician

can adjust on the machine to take diagnostic x-rays

SIBKIS when it comes to changing exposure factors!

Page 8: VET-120 Diagnostic Imaging

Milliamperage (mA)Figure 4-1, Page 36

Amount of electrons to produce x-rays Fundamental – Use the highest mA

your machine can do– Allows you to use much lower time

settings– Less exposure time to x-rays

The higher the mA, the more exposed your film will be (the DARKER it will be) if all other exposure factors are kept the same

Page 9: VET-120 Diagnostic Imaging

Exposure Time (S)

The longer the exposure time, the greater the number of x-rays produced

The longer the exposure time, the greater the exposure to scatter radiation

Inverse relationship to mA Benefits of shorter exposure times

– Less movement on x-ray (thoracic films)– Technician safety

Page 10: VET-120 Diagnostic Imaging

Kilovoltage Peak (kVp)Figure 4-4, Page 38

Speed of electrons used to create x-rays

Quality of x-ray– Not faster x-rays, but more penetrating

Affects both exposure (darkness) of film, and contrast of film

Santes’ Rule – (2 X thickness of body part) + 40 =

kVp

Page 11: VET-120 Diagnostic Imaging

CalipersFigure 4-5, Page 38

Definition– A tool used to measure the thickness

of a body part to be x-rayed– Use centimeters (cm), not inches– If in between numbers…… latitude

Page 12: VET-120 Diagnostic Imaging
Page 13: VET-120 Diagnostic Imaging

Latitude

kVp range Exposure latitude (kVp)

40-50 +2

70-80 +4

90-100 +6

Page 14: VET-120 Diagnostic Imaging

Kilovoltage Summary

It affects both contrast and exposure

Increased kVp more penetrating x-rays

Increased kVp increased scatter radiation

Increased kVp decreased contrast (leading to more latitude)

Page 15: VET-120 Diagnostic Imaging

Milliamperage and Time Milliamperes Exposure

time Formula

– mAs = mA X s

Examples – same mAs, different mA– Same

darkness (exposure)

Page 16: VET-120 Diagnostic Imaging

Distance (SID) Source-image distance (SID)

– Aka Focal-film distance (FFD)– Should be 40 inches for both table

top technique and grid technique Inverse Square Law –

– Figure 4-6, Page 39– Definition

KISS Fundamental – distance should always stay constant if possible

Page 17: VET-120 Diagnostic Imaging

Greater Distance = Less Shadows

Page 18: VET-120 Diagnostic Imaging

Greater Distance = Less

Intensity (Exposure)Figure 4-6, Page

39

Page 19: VET-120 Diagnostic Imaging

Review of X-ray Production

Figure 4-7 on page 40 X-ray tube parts Primary vs. secondary radiation Exposure factors

Page 20: VET-120 Diagnostic Imaging

Radiographic Quality

Page 21: VET-120 Diagnostic Imaging

Glossary Terms – Page 44

Page 22: VET-120 Diagnostic Imaging

Radiographic Quality

Definition– How easily details can be seen on an

x-ray– How sharp the details are

Page 23: VET-120 Diagnostic Imaging

Radiographic Quality Factors

Radiographic DensityRadiographic ContrastExposure FactorsScatter RadiationGrids

Page 24: VET-120 Diagnostic Imaging

Radiographic Density

Definition – amount of darkness (exposure) in x-ray film

When looking at a film for exposure, look at the body part you are interested in

What causes density, you ask? ---

Page 25: VET-120 Diagnostic Imaging

Densities of Various Structures

Subject densities: 1, Air. 2, Fat. 3, Water. 4, Bone. 5, Metal. Air is least dense, allowing x-rays to penetrate and expose the film. Metal is the most dense, absorbing most of the x-rays and allowing only a few to penetrate, exposing the film.

Page 26: VET-120 Diagnostic Imaging
Page 27: VET-120 Diagnostic Imaging

Factors Affecting Radiographic Density

All 4 exposure factors Figure 5-2 on page 46

– Same mAs (exposure), different thickness of body part

White structures on x-rays– Radio-opaque….. Why?

Black structures on x-rays– Radio-translucent….. Why?

Page 28: VET-120 Diagnostic Imaging

Density on X-raysFigure 5-3, Page 47

Page 29: VET-120 Diagnostic Imaging

Subject Density/Exposure

Page 30: VET-120 Diagnostic Imaging
Page 31: VET-120 Diagnostic Imaging
Page 32: VET-120 Diagnostic Imaging

How is the Exposure on This X-ray?

“Overexposed” (Figure 5-7 on page 49)– Body parts too dark– Why?

“Under-exposed” (Figure 5-6 on page 48)– Body parts too light– More common than overexposed

Correct exposure (“just right”!)

Page 33: VET-120 Diagnostic Imaging

Overexposed Film (Too Dark)

Page 34: VET-120 Diagnostic Imaging

Underexposed Film(Too Light)

Page 35: VET-120 Diagnostic Imaging

“Just Right, Goldilocks!”

Page 36: VET-120 Diagnostic Imaging

Underexposed or Overexposed?

Page 37: VET-120 Diagnostic Imaging

Underexposed or Overexposed?

Page 38: VET-120 Diagnostic Imaging

Underexposed or Overexposed?

Page 39: VET-120 Diagnostic Imaging

Radiographic Contrast

Definition– The density differences between 2

adjacent areas of an x-ray High contrast (short scale) Low contrast (long scale) Just like exposure, look at

body parts (not background) on film

Page 40: VET-120 Diagnostic Imaging

High Contrast vs. Low Contrast

Table 5-1, Page 46

Page 41: VET-120 Diagnostic Imaging

High Contrast X-rays

Lots of black and white on the film with very little shades of gray

Bone x-rays should be high contrast

Page 42: VET-120 Diagnostic Imaging
Page 43: VET-120 Diagnostic Imaging
Page 44: VET-120 Diagnostic Imaging
Page 45: VET-120 Diagnostic Imaging
Page 46: VET-120 Diagnostic Imaging

Low Contrast X-rays

The film has black, white, and lots of shades of gray in between

Usually soft tissue x-rays Latitude

Page 47: VET-120 Diagnostic Imaging
Page 48: VET-120 Diagnostic Imaging
Page 49: VET-120 Diagnostic Imaging
Page 50: VET-120 Diagnostic Imaging

Factors Affecting Radiographic Contrast

kVp – Relationship of contrast and kVp

(Table 5-1 on page 46)***– Latitude

Subject contrast– Difference in density between 2 body

parts– Depends on thickness & density of

part (Table 5-2 on page 46)

Page 51: VET-120 Diagnostic Imaging

Exposure Factors and X-ray Quality

mAskVpSID

Page 52: VET-120 Diagnostic Imaging

Digital Control Panel with Exposure Factors

Page 53: VET-120 Diagnostic Imaging

Milliamperage-Seconds (mAs)

Affects density (darkness, exposure) only

Does not affect contrast

Page 54: VET-120 Diagnostic Imaging

Kilovoltage (kVp)

Affects both contrast and density

Increased kVp causes increased scatter radiation

Increased kVp causes decreased contrast, which leads to more latitude– Soft tissue x-rays

Page 55: VET-120 Diagnostic Imaging

Distance (SID)

Affects density (exposure, darkness) only

Does not affect contrast

Page 56: VET-120 Diagnostic Imaging

Scatter Radiation

Definition– Secondary radiation formed as a result

of objects in the path of the primary beam

Mainly originates from the patient, but could come from increased kVp

Fogs the film Potentially damages RVT!!!

Page 57: VET-120 Diagnostic Imaging

Backscatter

Definition – When primary beam strikes patient, table, tray, or floor…. scatter radiation bounces back to patient, film, and you!

Therefore… GLOVE UP, OR ELSE!

Page 58: VET-120 Diagnostic Imaging
Page 59: VET-120 Diagnostic Imaging

Grids

Definition – Device placed between patient & x-ray film– Designed to absorb scatter radiation

Composed of lead strips Aligned so that most of primary

beam can get through them, but all secondary radiation is absorbed

Found under table, usually permanently mounted

Page 60: VET-120 Diagnostic Imaging

Grid Under Table

Page 61: VET-120 Diagnostic Imaging

Figure 5-11, Page 50

Page 62: VET-120 Diagnostic Imaging
Page 63: VET-120 Diagnostic Imaging

Grid Ratio

Definition – Relation of height of the lead strips to distance between them

Figure 5-13 on page 51 Example – 10:1 So what does it mean?

Page 64: VET-120 Diagnostic Imaging

Grid Factor

Definition– Amount that mAs must be increased

if a grid is used– Example – grid factor of 2

Page 65: VET-120 Diagnostic Imaging

Grid PatternsFigures 5-14 & 5-15 on Pages 51 & 52

Linear grids Crossed grids Focused grids Grid lines Movable grids

– Potter-Bucky Diaphragm (Figure 5-18 on page 53)

Page 66: VET-120 Diagnostic Imaging
Page 67: VET-120 Diagnostic Imaging

Potter-Bucky Diaphragm

Page 68: VET-120 Diagnostic Imaging

Care of Grids

Leave them alone Expensive and delicate

Page 69: VET-120 Diagnostic Imaging

To Use a Grid or Not?

Grid TechniqueTable Top Technique

Page 70: VET-120 Diagnostic Imaging

Grid Technique

Capture of scatter radiation Captures some of primary beam

also Less fogging Higher exposure factors needed Generally used for larger dogs &

cats

Page 71: VET-120 Diagnostic Imaging

Table Top Technique

Animal closer to the film Lower exposure factors needed Less shadowing

– Figures on page 55– Object-Film Distance (OFD) lower

Generally used for smaller animals/extremities

Page 72: VET-120 Diagnostic Imaging

Table Top – Less Shadowing

Page 73: VET-120 Diagnostic Imaging

Image Receptors

Page 74: VET-120 Diagnostic Imaging

Glossary Terms – Page 60

Page 75: VET-120 Diagnostic Imaging

What Are Image Receptors?

Definition – Tools used to capture invisible x-rays in such a way that they can be seen with the naked eye

Page 76: VET-120 Diagnostic Imaging

Types of Image Receptors

RadiographyXeroradiographyComputed TomographyNuclear Scintigraphy

Page 77: VET-120 Diagnostic Imaging

Radiography

Light-sensitive film in a light-proof cassette

Film is sandwiched between 2 intensifying screens in this cassette– 95% of exposure on film due to

light emitted on these screens!– Only 5% from x-rays themselves

Page 78: VET-120 Diagnostic Imaging

The Cassette

Definition – Lightproof encasement designed to hold x-ray film, with intensifying screens in close contact with film

Page 79: VET-120 Diagnostic Imaging

Types of Cassettes

Page 80: VET-120 Diagnostic Imaging
Page 81: VET-120 Diagnostic Imaging

Cassette Care Handle with care! Or ELSE!

Some physical abuse in large animal practice

Film-screen contact a MUST Cleaning cassettes – regularly with

mild soap & water

Page 82: VET-120 Diagnostic Imaging

Intensifying Screens

Definition – Sheets of luminescent phosphor crystals– When these phosphor crystals are

struck by x-rays they fluoresce 2 screens per cassette Requires less radiation exposure

(mAs)– ~95% of exposure on film due to

light from intensifying screens!

Page 83: VET-120 Diagnostic Imaging

Cassette with Screens

Page 84: VET-120 Diagnostic Imaging
Page 85: VET-120 Diagnostic Imaging

Desired Properties of Screens

High level of x-ray absorption High x-ray to light conversion Little or no “afterglow”

Afterglow– Definition – tendency of a phosphor to

give off light after x-ray production has stopped

Page 86: VET-120 Diagnostic Imaging

Intensifying Screen Anatomy

Figure 6-5 on page 63– 2 screens per cassette

4 layers– Plastic base – support layer, tight to

cassette– Reflective layer – thin – Phosphor crystal layer

• Layer that fluoresces

– Protective coat – thin, tight to film

Page 87: VET-120 Diagnostic Imaging

Cassette/Screen/Film Anatomy

Page 88: VET-120 Diagnostic Imaging

Phosphor Types

Thomas Edison – 1896– Calcium tungstate

• High x-ray absorption• Still used today

Rare earth phosphors – 1972– Expensive– Great x-ray to light conversion

• Needs a special type of film

Page 89: VET-120 Diagnostic Imaging

Rare Earth Screens

Page 90: VET-120 Diagnostic Imaging

Screen Speed – 2 FactorsFigure 6-8 on page 64

Phosphor crystal size Thickness of crystal layer

Page 91: VET-120 Diagnostic Imaging

Screen Speed Ratings

Slow – small crystals– Increased exposure time– High definition; more detail; sharp– Not grainy

Medium High – large crystals

– Decreased exposure time– More grainy– Less detail

Page 92: VET-120 Diagnostic Imaging

Screen Care Critical, as

they are so expensive

Inspect and clean regularly

Use commercial cleaner only

Chemical spills on the screen

Page 93: VET-120 Diagnostic Imaging

Don’t Develop the Screens!!!

Page 94: VET-120 Diagnostic Imaging

Fluoroscopy

Figure 6-11 on page 67 Definition – “Live x-rays” Main feature is the screen, as there is

no film Image viewed is exactly opposite of an

x-ray Recorded on TV screen Old hand-held units illegal in USA

Page 95: VET-120 Diagnostic Imaging

“Me & My Dog Tighe”

Button from 1910!

Page 96: VET-120 Diagnostic Imaging
Page 97: VET-120 Diagnostic Imaging

X-ray Film

Purpose of film– Provide a permanent record

Creates a latent image until developed

Page 98: VET-120 Diagnostic Imaging

Film AnatomyFigure 6-14 on page 69

Base Emulsion layer

– Gel layer– Phosphors located here

•Silver halide crystals

– Easily penetrated by developing solutions

Protective coating

Page 99: VET-120 Diagnostic Imaging

Layers of X-ray Film

Page 100: VET-120 Diagnostic Imaging

Latent Image

In emulsion layer Physical change in film When developed, chemical change

takes place to produce visible image Unexposed film that is developed

– Clear after processing Film totally exposed to light or x-

rays– Black after processing

Page 101: VET-120 Diagnostic Imaging

Film Types

Screen film– Silver crystals more sensitive to light

than x-rays– Matching film with screens

Non-screen film– Exposed by direct action of x-rays– No loss of detail due to screens– Dental radiography

(Video from St. Petersburg VTP)

Page 102: VET-120 Diagnostic Imaging

Non-screen Film

Page 103: VET-120 Diagnostic Imaging

Cassette for Screen Film

Page 104: VET-120 Diagnostic Imaging

Film SpeedPage 70

Same concept as screen speed Slow film (High detail) Medium film (Standard speed)

– Most widely used in veterinary practice today

Fast film (Ultraspeed film) Film latitude

– Narrow latitude film – high contrast film

Page 105: VET-120 Diagnostic Imaging

Film Care Fundamentals

Gentle with film, especially loading cassettes in dark room

Film boxes – store on end– No pressure creases in film

Storage temperature– Cool, low humidity

Expiration date – rotate film boxes so oldest film used first

Page 106: VET-120 Diagnostic Imaging

Loading Film in Cassettes

When loading a cassette, use both hands to avoid kink marks, and carefully place the film into the cassette. The cassette must be closed and latched gently.

Page 107: VET-120 Diagnostic Imaging

Film Processing

Chapter 7

Page 108: VET-120 Diagnostic Imaging
Page 109: VET-120 Diagnostic Imaging

Glossary Terms – Page 74

Page 110: VET-120 Diagnostic Imaging

Film Processing

A chemical reaction Latent (invisible) image on film

converted to the visible image that is seen on developed film

Page 111: VET-120 Diagnostic Imaging

The Darkroom

Most important place for turning out consistent, high quality x-rays

Where the greatest number of errors are made in producing a “diagnostic x-ray”

KISS PPPPP

Page 112: VET-120 Diagnostic Imaging

Qualities Of A Good Darkroom

Clean Organized

– Dry side– Wet side– All things back in their proper place

Lightproof – no cracks of light– Fogged film

Climate control – cool with low humidity

Page 113: VET-120 Diagnostic Imaging

Organization Of The Darkroom

Figure 7-1 on page 75

Dry side – Film exchange in cassettes– No splashing here!

Wet side – Chemical processing of exposed films– Drying of films

Page 114: VET-120 Diagnostic Imaging
Page 115: VET-120 Diagnostic Imaging

Dry Side

No “wet” stuff here Cassette unloading & reloading Empty film hangers on wall Film storage bin

Page 116: VET-120 Diagnostic Imaging

Wet Side

Developing tank– Thermometer for determining

temperature of developer Rinse bath – usually 4X the size of

other tanks Fixing tank Film drying

Page 117: VET-120 Diagnostic Imaging

Light-proofing

Film fogging without this – Film phosphors react

to room light– Light-colored walls… Why?

• More safelight reflection

Safelight – < 15 watts, 4 feet away from work

area– Filter to eliminate blue & green

light (shorter wavelengths)

Page 118: VET-120 Diagnostic Imaging

Safelights

Page 119: VET-120 Diagnostic Imaging

Dark Room with Safelight

Page 120: VET-120 Diagnostic Imaging

Film Exposed to Light

Page 121: VET-120 Diagnostic Imaging

Film-Processing Techniques

Manual processing– 10 steps (pages 79-82)

Automatic processing– Figure 7-17 on page 83– Processor maintenance

Digital x-rays! KOOL!– Latent image developed inside

computer

Page 122: VET-120 Diagnostic Imaging

Film-Processing Solutions

Developer Rinse Bath Fixer Wash bath Drying

Reticulation – mottling of film from different temperatures in the solutions

Page 123: VET-120 Diagnostic Imaging
Page 124: VET-120 Diagnostic Imaging

Preparing Film For Developing

Opening cassettes– Keep closed always!

Hanging film Reloading cassettes

– When?– How?

Be humble if you don’t know something! (OR ELSE!)

Page 125: VET-120 Diagnostic Imaging

Unexposed Film Stored in Dark Room

Figure 7-2, Page 75

Page 126: VET-120 Diagnostic Imaging

Reloading Cassettes

Page 127: VET-120 Diagnostic Imaging

Developing Solution

Developing AgentsAcceleratorsPreservativesHardenersRestrainersSolvent

Page 128: VET-120 Diagnostic Imaging

Developing Solution

Purpose – to convert latent image to visible image– Done by converting exposed silver

crystals on film to BLACK metallic silver

How long here? – 5 minutes at 68 degrees

Fahrenheit– KISS!

Page 129: VET-120 Diagnostic Imaging

Developing Solution

Developing agents – think emulsion layer of film– Hydroquinone or phenidone (alkaline

pH)– Converts exposed silver grains to

black metallic silver on the film

Accelerators – KCO3 or NaCO3

– Increase activity of developer– Increase pH

Page 130: VET-120 Diagnostic Imaging

Developing Solution

Preservatives – prevent oxidation Hardeners – in auto-processors

– Prevent swelling of emulsion layer Restrainers – limit developer

action on exposed silver grains Solvent – water

Page 131: VET-120 Diagnostic Imaging

Rinse Bath

Purpose – to rinse excess developer off before fixing film– If film is not rinsed, alkaline

developer will neutralize acid fixer No “backwash”! Stops developing process Rinse in circulating water for 30

seconds Not found in auto-processors

Page 132: VET-120 Diagnostic Imaging

Fixing Solution

Clearing (Fixing) AgentsPreservativesHardenersAcidifiersBuffersSolvent

Page 133: VET-120 Diagnostic Imaging

Fixing Solution

Purposes – To remove unexposed silver crystals,

and undeveloped exposed silver crystals

– To harden (preserve) emulsion layer of film

How long in fixer?– Ideally, 2X developing time (10

minutes)

Page 134: VET-120 Diagnostic Imaging

Fixing Solution

Clearing (Fixing) Agents – sodium thiosulfate or ammonium thiosulfate– “Clears” film– “Fixes” film

Preservatives – sodium sulfite

Page 135: VET-120 Diagnostic Imaging

Fixing Solution

Hardeners – aluminum salt– Prevent excessive swelling of

emulsion layer– Shortens drying time

Acidifiers – accelerate action of other chemicals– Neutralize developer

Buffers – maintain pH of solution Solvent – water

Page 136: VET-120 Diagnostic Imaging

Wash Bath

Purpose– Remove ALL chemicals from film

Circulating water – 20 to 30 minutes

Prevents fading and discoloration

Page 137: VET-120 Diagnostic Imaging

Drying Films

It takes time! Careful with shortcuts

– Water streaks

Page 138: VET-120 Diagnostic Imaging

Solution “Issues”

Solution replenisher– One 14X17 radiograph takes 60 cc’s

of developer into rinse bath– “Fill it up!”

Solution replacement – about every 30 days– How to tell when?– Change the fixer solution first! Why?

Page 139: VET-120 Diagnostic Imaging

Biologic Growth

Bacteria, fungi, algae can grow in rinse bath

Slime on sides of tanks? Treat this or metal can corrode

Page 140: VET-120 Diagnostic Imaging

Manual Processing – Key Points

KISS PPPPP Unloading cassette Loading film on hanger Developing film – lid on! Rinsing film – no “backwash”! Drying film – water marks on film Filing film – cut the edges

Page 141: VET-120 Diagnostic Imaging

Auto Processing – Key Points

Figure 7-17 on Page 83 Same procedure, but higher

temperatures and specialized chemicals. Why?

No rinse between developing & fixing Chemical replenishment Processor maintenance

– Have a reliable service company

Page 142: VET-120 Diagnostic Imaging
Page 143: VET-120 Diagnostic Imaging
Page 144: VET-120 Diagnostic Imaging

Silver Recovery – Fixer & Films

Why?– Environmental– Economic – 1 ounce of silver in 40-45

small films Where is the silver?

– Fixer solution– Old x-ray films

Page 145: VET-120 Diagnostic Imaging

Methods for Fixer Solution Harvesting

Metallic replacement– Iron (steel wool)

Electrolyte recovery– 2 electrodes into fix tank

Chemical precipitation– Adding more chemical compounds

Page 146: VET-120 Diagnostic Imaging

Film Identification Every x-ray needs to be properly

identified– Medical reasons– Legal reasons

What is the minimal information?– Name & address of veterinary

practice– Date (including year)– Patient identification (owner name,

patient name, etc.)

Page 147: VET-120 Diagnostic Imaging

Film Identification Tools

Lead MarkersLead-impregnated Tape Photo-imprinting Label System

Page 148: VET-120 Diagnostic Imaging

Lead MarkersFigure 7-20 on Page 86

Easiest tool to use Letters and numbers used Lead absorbs x-ray…… are letters

and numbers black or white on film?

Page 149: VET-120 Diagnostic Imaging

Lead Markers

Page 150: VET-120 Diagnostic Imaging

Lead-Impregnated TapeFigure 7-21 on Page 86

Disposable after use Write on tape with ball point pen

– This displaces the lead when making letters & numbers

What color are the displaced letters & numbers – black or white? Why?

Page 151: VET-120 Diagnostic Imaging
Page 152: VET-120 Diagnostic Imaging

Photo-Imprinting Label System

Figure 7-22 on Page 87

Uses lead blocker on outside of cassette along with a typed-up ID card with information for identifying x-ray

Also uses a photo-imprinter in dark room

Lead blocker keeps a small portion of x-ray film clear of radiation for the imprinter

Page 153: VET-120 Diagnostic Imaging
Page 154: VET-120 Diagnostic Imaging

Label for Photo-Imprinting System

Page 155: VET-120 Diagnostic Imaging

Film Filing & Storage

Cut the corners (manual processing)

Side by side, chronologically– Alphabetical, by owner– By case number– By date

Store for 7 years legally– Or until patient dies?