Page 1: Mindray DP-50 Quick Guide

7/27/2019 Mindray DP-50 Quick Guide

http://slidepdf.com/reader/full/mindray-dp-50-quick-guide 1/2www.bcftechnology.com

Help

Exit currentstatus tothe previousstatus

Press to openor close themeasurement andcalculation reports

Press toopen/closethe set-upmenu

Time Gain Compensationincreases/decreasesbrightness of returning signalat varying depths

4. Freq/THI – Rotate tochange frequency.

7. Focus – Press toactivate then rotate tomove focal point up/downthe depth scale.

5. Depth ofview. Rotateto alterscanning

depth. Pressto zoom.

Two imageson split screen

Press for measurementpackages

Four imageson splitscreen

M mode

B mode

Press to movebetweenactive imageswhen in split

screen mode

12. Press toend the exam

2. Enterthe patientdetails

Press toshow/hideon-screenarrow

Calipers formeasuringdistance andarea

6. Gain –Increases/decreasesoverall

brightness

3. Change presetor probe(if two attached)

Press to addtext to image

8. Pressto activatemenu

Set button foruse in menus

9. Freeze image.Use trackballto scroll backthrough pausedimages to selectrequired image

10. Save 1stores stillimage,Save 2 stores

a cine-loop

F1-F4 – Assign afunction to each ofthese keyes throughset up menu

11. iStation –Access storedimages and clipsfor all patients

Press to displayor hide thebiopsy guide line

Press to deletecomments 1. On/Off button

Press toreview storedimages ofcurrent patient

Your ultrasound and X-ray people

Mindray DP-50 Quick Guide

Page 2: Mindray DP-50 Quick Guide

7/27/2019 Mindray DP-50 Quick Guide

http://slidepdf.com/reader/full/mindray-dp-50-quick-guide 2/2© BCF Technology 2013

Your ultrasound and X-ray people

1. Attach relevant probe before turning on the scanner. Do not remove the probe whilst thescanner is turned on.

2. Patient – Enter patient details.

3. Probe – Ensure you are in the correct application setting (preset) for the type of exam beingperformed.

4. Frequency – Ensure that your frequency is at the correct setting. A higher frequency settingprovides higher resolution to the image, however penetration is reduced. A lower frequencysetting improves penetration ability, however resolution is diminished.

5. Depth – Adjust so that the region of interest occupies approximately ¾ of the screen.A depth scale (in centimetre increments) is displayed next to the image.

6. Gain – Increase or decrease the overall brightness of the image.

7. Focus – Move the focal point(s) up and down the depth scale until it/they are pointing atthe region of interest.

8. Menu – Press the menu button to access further image optimisation parameters. Use thetrackball to highlight individual parameters and use the set and back buttons (located tothe left/right of the trackball) to alter each parameter.

9. Measurements and calculations – Freeze the image and scroll back through the pausedimages (using the track ball) to select an appropriate image. Then press the caliper ormeasure button.

10. Save 1 or 2 – Save relevant images and cine-loops.

11. iStation – Access stored information for each exam performed, review images/cine-loops,transfer images to USB memory stick.

12. End exam – Finishes the exam of the current patient. You are then ready to repeat forscanning the next patient.

Note: All BCF Technology products are to be used for animal applications only.