dls help manual

Upload: ximena-puentes

Post on 02-Jun-2018

245 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/10/2019 Dls Help Manual

    1/66

    Table of Contents

    Welcome

    Whats New?

    TECHNICAL QUESTIONSDynamic Light Scattering

    What is a correlation curve?

    How is the correlation curve measured?

    Whats the relevance of the frictional coefficient?

    What is the hydrodynamic radius (RH)?

    Is it better to use longer acquisition times?

    Whats the difference in the intensity, number, and mass distributions?

    Static Light Scattering

    What is the Rayleigh ratio?

    How is the Rayleigh ratio measured?Whats the relevance of toluene in static light scattering?

    Is the Rayleigh ratio of toluene wavelength dependent?

    What are the units of the parameters in static light scattering?

    When is multi-angle data necessary for MW determination?

    Data Interpretation

    What is the baseline?

    What is the SOS?

    What are number fluctuations?

    Whats the difference between the old and the new Regularization?

    Sample Preparation

    Do samples need to be filtered?

    Will filtering perturb the sample?

    Why is the salt content relevant?

    Calculations

    What is the Regularization Method?

    Can you get molecular weight information from DLS data?

    How is the axial ratio calculated?

    How is the mass distribution of radii calculated?

    Are the %Mass calculations valid for all systems and conditions?

    Sample Properties

    Whats the significance of the refractive index () in DLS?

    Is the refractive index increment (d/dC) wavelength dependent?

    Is the solvent refractive index (o) wavelength dependent?

    Are o and d/dC temperature dependent?

  • 8/10/2019 Dls Help Manual

    2/66

    Is it alright to estimate solvent viscosity values?

    Software & Utilities

    What is the _____ value of ______?

    LABORATORY REPORTS

    January 2000

    Summary of Results

    Lysozyme UV Analysis

    Instrumental Error Effects in UV Analysis

    Effects of Acquisition Time on DLS Results

    Effects of Concentration on DLS Results

    Defining a Quality Control Parameter

    Maximum Acceptable CNF Value

    Quality Control Specifications

    May 2000

    Summary of Results

    Cell Comparison UV Measurements

    Cell Comparison Beers Law Region for Lysozyme

    Calculating Minimum Concentrations

    Eppendorf UVettes - QC Analysis

    Bovine Serum Albumin (BSA) UV Study

    Effects of Laser Power on Scattering Intensity

    DLS Study with Polystyrene Standards

    Enhancing DynaPro Sensitivity

    APPLICATION NOTES

    Surfactants & Micelles

    Characterization of Surfactant and Micellar Systems

    Determination of Critical Micelle Concentration

    Measurement of Mixed Micelle Size by DLS

    Vesicle Size Distribution Analysis

    Surfactants as Chaotropic Agents in Protein Systems

    Representative Publications for Surfactant and Micelle Systems

    Polymers & Polypeptides

    Biological and Synthetic Polymer Characterization

    TC Measurements of Recombinant Hydrophobic Polypeptides

    Electrostatic Effects on Polyelectrolyte DLS Measurements

    Polysaccharide Fractionation by Filtration

    Characterization of Low Molecular Weight Peptides

    Representative Publications for Polymeric Systems

  • 8/10/2019 Dls Help Manual

    3/66

    Protein Characterization

    Protein Characterization

    Conformation of Trimeric Form of Migration Inhibitory Factor

    Insulin Quaternary Structure as a Function of pH

    pH Dependent Monomer-Dimer Equilibrium Constants for GARTMonitoring Protein Thermal Stability - Glycosylated Hb Melting Points

    Representative Publication in Protein Characterization

    MANUALS & TUTORIALS

    Utility ReadMe Files

    PSI Archive Setup Instructions

    PSI Books Setup Instructions

    REFERENCESBackground References

    Application References

  • 8/10/2019 Dls Help Manual

    4/66

    What is a correlation curve?

    Photon Correlation Spectroscopy (PCS), also known as Dynamic Light Scattering (DLS), is

    concerned with the investigation of correlation of photons. In a typical light scatteringexperiment single photons are detected with a single photon counting device. This detector

    transforms the signal from a single photon into an electronic signal, basically a 1 or a 0,

    depending on whether a photon was or was not detected in a certain time interval. The objective

    of PCS is to find any peculiar properties of the scattered signal which can be used to

    characterize and describe the seemingly random "noise" of the signal, and the correlation curve

    is used to achieve this objective.

    A typical way to describe a signal is by way of its autocorrelation. The autocorrelation

    function of the signal from the scattered intensity is the convolution of the intensity signal as a

    function of time with itself. In more abstract terms, if the detected intensity is described as a

    function I(t), then the autocorrelation function of this signal is given by the followingexpression, where is the shift time.

    The above function is also called the intensity correlation function, and it is used to describe the

    correlation between the scattering intensities measured at t = 0 and some later time (tn = t0 + ).

    Consider for example, the schematics shown below, where the frame represents the scattering

    volume, the dark blue circle represents the particle, and the light blue region represents the

    volume within which the particle must be contained. The scattering intensity at the detector is

    dependent upon the position of the particle relative to the detector. If we consider very small

    time increments (micro-seconds), then at t0 (t = 0) there is a finite volume of space within which

    the particle can move during the first time increment. For reasons that will become evident

    later, well call this finite volume of space the diffusional volume. At t = t1, the particle has

    diffused a single step. Since the exact path is unknown, the diffusional volume or the volume of

    space within which the particle must reside considering all paths has expanded. The change in

    particle position is accompanied by a change in the measured scattering intensity at the detector.

    However, this new scattering intensity is still strongly correlated with that measured at t0 due to

    the finite diffusional volume and the correlation between the current and initial positions. As

    the shift time increases, the diffusional volume also increases. The increased diffusional volume

    is accompanied by a decrease in the correlation between the current and initial particle

    positions. The correlation is still present however, as long as the diffusional volume is finite.

    At longer times (frame 6), the diffusional volume becomes infinite, and the correlation between

    the initial and current particle position and the respective scattering intensities is lost.

  • 8/10/2019 Dls Help Manual

    5/66

    The correlation in particle position at small shift times is contained within the measured

    intensity correlation function, an example of which is shown below. In the absence of any

    applied forces, the particle position is dictated by the degree of Brownian motion. As such, the

    measured intensity correlation curve is an indirect measure of the particles diffusion coefficient.

  • 8/10/2019 Dls Help Manual

    6/66

    How is the correlation curve measured?

    The intensity correlation function is defined as shown in the following expression.

    The shift time () is often referred to as the delay time, since it represents the delay between the

    original and the shifted signal. In reality, the continuous intensity correlation function can not

    be measured. It can however, be approximated with discrete points obtained by a summation

    over the duration of the experiment. This summation is performed by the correlator, a circuit

    board composed of various logic chips and operational amplifiers which continuously multiplies

    and adds measured intensity values.

    A defining parameter for a correlator is its speed, i.e. its ability to add up numbers fast, or

    inversely expressed, the smallest delay time that can be handled. This delay time () determines

    the smallest accessible time of G(). Hence, G() is the first numerical point in the intensity

    correlation curve that can be experimentally measured for a given correlator. Using a time

    increment of , a grid of experimental data values can be obtained. At the start of the

    experiment, t0 = 0. The first time then is at t1 = , the second is at t2 = 2, and so on. As an

    array, the stream of numbers could be displayed in table format as shown below, with the

    experiment ending at tn = n.

    The correlator now finds out how related these intensity numbers are to each other. It does this

    by comparing each number to its neighbor a defined time interval later. This comparison is

    carried through the duration of the experiment. As a summation, the comparison can be

    expressed as follows, where the upper summation limit is given by the appropriate indexbelonging to the largest available summation term for k.

    The above expression is very general and holds true for both linear and arbitrary delay times. To

    understand the meaning of this summation, lets assume we would like to find G(t1) = G() .

  • 8/10/2019 Dls Help Manual

    7/66

    According to the previous equation, G(t1) is defined as:

    In the same fashion, G(t2) is defined as:

    In table format, the correlation function values can then be written as given below.

    The series of G(t) values is typically normalized, such that the value for very large time (as t

    ) is G() = 1. For standard light (with Gaussian statistics), this normalization imposes a

    theoretically limit of G(t0) = 2 at t = 0. However, only carefully optimized optical systems can

    ever achieve such a high intercept or amplitude.

    For a random signal and a large measurement time, all G values would add up to roughly thesame number. The signal would then be described as uncorrelated, and the resultant intensity

    correlation curve would look like random fluctuations about the baseline, similar to intensity

    correlation curve for H2O shown in the following figure.

  • 8/10/2019 Dls Help Manual

    8/66

    For light scattered from diffusing molecules however, the intensity correlation curve exhibits an

    exponential decay, indicating that the signal is correlated.

    For typical diffusion processes the correlation function has the form of 1 plus an exponential

    decay function. The decay constant, in the following expression, is representative of the

    diffusional properties of the particle under examination. Hence, evaluation of leads to the

    particle diffusion coefficient, which in turn is used to calculate the hydrodynamic radius via the

    Stokes Einstein equation.

    G(t) = 1 + exp(- t )

  • 8/10/2019 Dls Help Manual

    9/66

    What's the relevance of the frictional coefficient?

    In the Stokes model for motion of a hard sphere through a viscous liquid, the resistance to

    motion is related to the shearing force between adjacent solvent planes moving at differentvelocities. The shearing force per unit area (Fs/A) at each slippage plane is described with the

    following differential equation, where m is the mass, A is the surface area, dv/dt is the

    acceleration, is the viscosity of the medium, and dv/dy is the velocity gradient across the

    shearing planes.

    For stick boundary conditions, where the limiting assumption that the fluid layer adjacent to the

    particle has the same velocity as the particle is made, solution of the above equation gives thetotal force of viscous resistance (Fvis) as shown below, where R is the particle radius and vs is

    the limiting steady state velocity of the fluid when y becomes very large.

    The proportionality factor in the above expression is defined as the frictional coefficient (f = 6

    R), and it is the frictional coefficient that is used to calculate the radius of a particle from the

    measured diffusion coefficient (D) in dynamic light scattering studies.

  • 8/10/2019 Dls Help Manual

    10/66

    What is the hydrodynamic radius (RH)?

    In dynamic light scattering experiments, the radius (R) of the particle is calculated from the

    diffusion coefficient (D) via the Stokes-Einstein equation, where k is the Boltzmann constant, Tis the temperature, is the solvent viscosity, and f = 6R is the frictional coefficient for acompact sphere in a viscous medium.

    By definition then, the DLS measured radius is the radius of a hypothetical hard sphere thatdiffuses with the same speed as the particle under examination. This definition is somewhatproblematic with regard to visualization however, since hypothetical hard spheres are

    non-existent. In practice, macromolecules in solution are non-spherical, dynamic (tumbling),and solvated. As such, the radius calculated from the diffusional properties of the particle isindicative of the apparent size of the dynamic hydrated/solvated particle. Hence theterminology, hydrodynamic radius.

    The effects of asymmetry and hydration on the radius calculated from the Stokes-Einsteinequation can be examined using lysozyme. From the crystallographic structure, lysozyme canbe described as a 26 x 45 prolate ellipsoid with an axial ratio of 1.731, a MW of 14.7 kDa,and a partial specific volume of 0.73 mL/g. Considering only the mass and partial specificvolume (Vp), the equivalent radius for a sphere with the same mass as lysozyme (RMass) is 1.62nm, considerably smaller than the 2.25 nm rotational radius (RRot) determined by rotating the

    protein about the geometric center.

    The Perrin or shape factor (F) is defined as the ratio of the apparent frictional coefficient for anellipsoid to that of a sphere of identical volume, and can be used to account for the non-sphericalshape of lysozyme. For a prolate ellipsoid with an axial ratio of 1.731, the Perrin factor is1.022. As shown below, the non-spherical shape of lysozyme leads to an increase in theexpected radius of the particle, designated here as RVol.

    The hydrodynamic radius (RH) includes effects arising from both shape and hydration. For

  • 8/10/2019 Dls Help Manual

    11/66

    lysozyme, the measured RH is 1.90 nm. The 2.4 increase in RH compared to RVol isconsistent with the thickness (2.8 ) of a single layer of water, suggesting that thehydrodynamic radius includes a single hydration layer.

    The various size parameters discussed above are represented in the following figure. As seen

    here, asymmetry and hydration lead to an increase in the measured size of lysozyme compared tothat calculated for a hypothetical hard sphere of the same mass. However, this measuredhydrodynamic radius is still 15 % smaller than the rotational radius.

  • 8/10/2019 Dls Help Manual

    12/66

    Is it better to use longer acquisition times?

    The number of photons striking the detector during the course of the experiment is directly

    correlated with the acquisition or sample time. The more photons collected, the less noise in theautocorrelation curve. Hence, longer acquisition times are synonymous with smoother

    correlation curves and enhanced confidence in the experimental results. Because of the

    correlation with photons collected, longer acquisition times are usually required at low sample

    concentrations, where the signal to noise (analyte to solvent) intensity ratio is inherently small.

    There is a down side to long acquisition times. At longer acquisition times the likelihood of a

    dust event is significantly increased. If a dust particle enters the scattering volume during the

    course of measurement, the correlation curve is shifted upward proportionally, and baseline

    evaluation becomes problematic. As a rule of thumb then, you should keep your acquisition

    time as short as possible; but keep an eye on the noise in the correlation curve and the baseline

    values.

    On the upside, noise in the correlation curve can also be reduced via statistical averaging, i.e. by

    collecting multiple runs at short acquisition times. This approach to noise averaging is

    particularly useful for samples that arent really as clean as they should be, even though

    significant data outlier editing is typically required. The technical support staff at Protein

    Solutions Inc is currently developing a DYNAMICS software utility that calculates the optimum

    acquisition time and number of measurements for samples of a given molecular weight and

    concentration. Check with your Protein Solutions sales representative for availability.

  • 8/10/2019 Dls Help Manual

    13/66

    What's the difference between the intensity, number, and mass distributions?

    According to static light scattering theory, the scattering intensity of the ith particle (Ii) in a

    distribution of particles is proportional to both the molecular weight (Mi) and the concentration

    as given in the simplified form of the Rayleigh expression shown below, where Ci is the weight

    or mass concentration, Ni is the number concentration, and K is a constant.

    The size or radius (R) of the particle can be estimated from the molecular weight via the partial

    specific volume (Vp) as shown in the following expression, where NA is Avogadros number

    and x is a shape parameter equal to 3 for spheres and 2 for coils.

    After substitution, the simplified Rayleigh expression can be re-written in the following form,

    which relates the scattering intensity to the mass and number concentrations of the ith particle.

    The mass and number distributions are then calculated from the intensity distribution as shown

    below.

    As seen in the following figure, the mean values for the %Mass and %Number distributions are

    shifted towards lower values as a consequence of the Ri dependence in the above equations.

  • 8/10/2019 Dls Help Manual

    14/66

  • 8/10/2019 Dls Help Manual

    15/66

    What is the Rayleigh ratio?

    The Rayleigh ratio is defined as the ratio of the scattered (Is) to the incident (Io) light intensity,

    with the scattered intensity being summed over all space. Because of geometric limitationsimposed by the restriction of measuring the intensity at a given angle () and distance (r) fromthe scattering volume, the working expression for the Rayleigh ratio is that shown below, where

    is is the scattering intensity measured at fixed r and .

    The scattering intensity is dependent upon a number of solvent and particle parameters, principle

    of which are the polarizability () of the particle and the permittivity (o) of the solvent. Hence,

    the Rayleigh ratio can also be written as shown below, where is the particle density, NA is

    Avogadros number, M is the particle molecular weight, and o is vacuum wavelength of theincident radiation.

    Using the Clausius-Mosetti equation to relate the polarizability to the permittivity of a molecule,

    the Rayleigh equation can be reduced to the commonly used form shown below, where C is the

    weight concentration, M is the weight average molecular weight, A2 is the 2nd virial coefficient,

    indicative of solute-solvent interactions, and K is the optical constant.

    The vacuum wavelength, solvent refractive index (o), and analyte specific refractive index

    increment (d/dC) are incorporated into the optical constant as shown below.

    The above two equations are those typically used for molecular weight determinations of small

  • 8/10/2019 Dls Help Manual

    16/66

    particles in static light scattering experiments. All of the variables in the optical constant (K)

    are either known or can be easily measured. As such, a plot of KC/Rvs. C in the absence ofparticle interactions should be linear, with an intercept equivalent to 1/M and a slope

    proportional to the 2nd virial coefficient (A2).

  • 8/10/2019 Dls Help Manual

    17/66

    How is the Rayleigh ratio measured?

    The Rayleigh ratio (R) is defined as the relative scattering intensity measured at an angle and

    a distance r from the scattering center. As shown in the following equation, where is is the r anddependent scattering intensity and Io is the incident intensity, one has simply to measure is, Io,

    r and to determine the Rayleigh ratio for a solution of particles. However, this simplicity is

    strongly attenuated in practical application, where the measurement of Io, r, and cannot be

    accomplished with the precision necessary for accurate determination of the Rayleigh ratio.

    An alternate approach is to compare the scattering intensity of the sample analyte to that of a

    well characterized standard, such as toluene, with a known Rayleigh ratio. The expression for R

    can be rearranged to the form shown below.

    If the scattering intensities of the sample and toluene are measured under the same experimental

    conditions, the Rayleigh ratio of the analyte can then be determined from the ratio of intensities

    as shown below, where the A and T subscripts represent the analyte and toluene respectively and

    the terms are included to account for differences in the refractive indexes of the toluene and

    sample solutions.

    Since it is the photon count rate (CR) rather than the scattering intensity that is measured at the

    detector in typical light scattering instrumentation, the expression for RA is converted to the

    working expression shown below.

  • 8/10/2019 Dls Help Manual

    18/66

    What's the relevance of toluene in static light scattering?

    In static light scattering experiments, the concentration dependent scattering intensity of the

    sample solution can be used to determine the absolute molecular weight (M) and 2nd virialcoefficient (A2) of the analyte under examination.

    In the above expression, K is an optical constant, incorporating wavelength and refractive index

    effects, P() is the shape factor, accounting for angle dependent multiple scattering from large

    particles, and Ris the Rayleigh ratio of the scattered intensity to the incident intensity. For

    particles with diameters more than 10 times smaller than the wavelength of the incident

    radiation, the shape factor goes to 1. The Rayleigh ratio is calculated using the followingexpression, where is is the residual scattering intensity (sample - solvent), Io is the incident

    intensity, is the scattering angle, and r is the distance from the scattering volume to the

    detector.

    The above expression can be rewritten in the form shown below, where the sample dependent

    parameters are grouped on the left hand side of the equality sign and the instrument dependent

    parameters are grouped on the right.

    The instrument dependent parameters (Io, , and r) are difficult to measure with the precision

    required for acceptable molecular weight and 2nd virial coefficient calculations. Rather than

    attempting to measure these parameters directly, the typical approach in static light scattering

    experiments is to calibrate the instrument parameters using a standard with a well defined

    Rayleigh ratio value. Toluene is one of the generally accepted standards for instrument

    calibration in static light scattering experiments. If the scattering intensities of the sample and

    toluene are measured under identical experimental conditions, the Rayleigh ratio of the analyte

    can then be determined from the ratio of intensities as shown below, where the A and T

    subscripts represent the analyte and toluene respectively.

  • 8/10/2019 Dls Help Manual

    19/66

    The terms are included in the above expression to account for differences in the refractive

    indexes of the toluene and sample solutions. There is some debate as to whether or not therefractive index terms should be included. For low concentration aqueous protein solutions, the

    inclusion of the terms functions to scale the toluene Rayleigh ratio by a factor of 0.8.

  • 8/10/2019 Dls Help Manual

    20/66

    Is the Rayleigh ratio of toluene wavelength dependent?

    The wavelength dependence of the Rayleigh ratio (R) is embodied in the definition shown

    below, where is the particle density, NA is Avogadros number, M is the particle molecularweight, o is the wavelength of the incident radiation, is the polarizability of the particle, and o is the permittivity of the solvent.

    As evident in the above expression, the Rayleigh ratio should vary linearly with -4. Thistheoretical linear dependence is consistent will experimental results, exemplified in the

    following figure, which shows the measured Rayleigh ratio of toluene at 25 oC and a 90o

    scatting angle as a function of -4.

    For the DynaPro line of molecular sizing instruments, the laser wavelength is around 830 nm.

    The Rayleigh ratio for toluene at 830 nm and 25 oC is 4.23 x 10-6 cm-1. For other wavelengths,

    the toluene R90 value can be predicted using the following expression, where the units for the

    Rayleigh ratio and the wavelength are cm-1 and nm respectively.

  • 8/10/2019 Dls Help Manual

    21/66

    What are the units in static light scattering?

    The key to simplifying the units in static light scattering experiments is to recognize that the

    units must be the same on both sides of the equality sign. When the Rayleigh equation is writtenin the standard form shown below, it is easily seen that the collection of units on the left hand

    side of the equation must ultimately reduce to units of inverse molecular weight, i.e. mol/g.

    The reduction to mol/g can be accomplished using the units listed in the following table.

    The dimensional analysis is shown below. As seen in the final expression, incorporation of the

    units from the above table reduces the units in the Rayleigh equation to those of inverse

    molecular weight on both sides of the equality sign.

  • 8/10/2019 Dls Help Manual

    22/66

    When is multi-angle data necessary for MW determination?

    The expression used to describe the scattering of a dilute solution of particles is shown below,

    where K is an optical constant, C is the particle concentration, Ris the Rayleigh ratio ofscattered to incident light intensity, M is the weight average molecular weight, A2 is the 2nd

    virial coefficient, Rg is the radius of gyration, is the vacuum wavelength of the incident

    radiation, and is the scattering angle.

    The angular dependent portion of the second term arises from interference effects due to

    multiple scattering from a single particle. For particles much smaller than the wavelength of the

    incident radiation, this term goes to zero, and the angular dependence of the scattered light

    vanishes. Under these conditions, the absolute molecular weight is determined from the

    concentration dependence of the Rayleigh ratio, and angular dependent data is redundant. For

    larger particles, it is still the concentration dependence that leads to the molecular weight, but

    interference effects must be accounted for. It is at this point that multi-angle instruments

    become necessary. In preface to the question of what is much smaller, the figure below, which

    shows the dependence of the interference factor on the wavelength to particle size ratio at

    different scattering angles is included here. As a rule of thumb, the size cutoff for angle

    independent Rayleigh scattering is Rg 20*. Typical wavelengths for the DynaPro series of

    instruments are on the order 800 nm, corresponding to an upper size limit for single angle

    particle MW determination of circa 80 nm diameter.

  • 8/10/2019 Dls Help Manual

    23/66

    What is the baseline?

    The baseline in the Cumulants Datalog is the measured value of the last coefficient in the

    correlation curve. For ideal, noise free dynamic light scattering experiments the baseline shouldbe 1.000, indicating that there has been sufficient diffusion during the acquisition such that the

    location of the particle is no longer correlated with its initial position. The DynaPro line of

    molecular sizing instruments utilizes mono-modal fiber technology to minimize the introduction

    of stray light or background noise. As such its not difficult to achieve baseline tolerances on

    the order of 0.005.

    Since the baseline value reported in the Cumulants Datalog is a measured, rather than a

    calculated value, it is independent of the algorithm used to deconvolute the size and distribution

    from the correlation curve. It is therefore a useful parameter for filtering out bad or outlying

    correlation curves. One should remember however, that a good baseline value doesnt

    necessarily mean the correlation curve is also good. Consider for example the following figure,which shows a 3 second acquisition time correlation curve for a dusty sample. The discontinuity

    in the curve is indicative of number fluctuations caused by a dust particle passing through the

    beam during the course of the acquisition. Even though this correlation curve is an obvious

    outlier, the measured baseline is 1.000. Hence data filtering based solely on the measured

    baseline value would leave this correlation curve unmarked, and lead to erroneous data

    interpretation.

  • 8/10/2019 Dls Help Manual

    24/66

    What is the SOS?

    The SOS is the sum of squares difference between the measured correlation curve and the best

    fit curve calculated using the Cumulants method of analysis, where a dust and noise freemonomodal (single distribution) low polydispersity (narrow distribution) sample is assumed.

    The SOS then is a quantitative value representative of the residual or the goodness of fit

    between the measured and theoretical data (see figures below).

  • 8/10/2019 Dls Help Manual

    25/66

    In the DynaPro line of light scattering instruments, the SOS value for each correlation curve is

    reported in the Datalog within the DYNAMICS software. Any deviation of the measured curve

    from the theoretical will lead to increased SOS values. Sources of deviation could be noise,

    dust, multi-modal samples, large polydispersity, etc. Hence, there is no absolute SOS value thatone should look for, particularly since the SOS value is also dependent upon the number of

    correlation channels included in the calculations (see correlation channel cutoffs in the

    DYNAMICS Data Filtering window).

    The best way to use the SOS values in the Cumulants Datalog is to look for inconsistencies. As

    an example: Because of background noise at low acquisition times, the SOS value for a dust

    free sample might be on the order of 10 - 50. If dust passes through the beam during the

    collection of a single correlation curve (during 1 acquisition time), the calculated SOS value will

    jump by as much as 1 or 2 orders of magnitude. So the SOS value is a good indicator of

    whether of not a dust free correlation curve has been collected. Depending upon how dirty, i.e.

    dusty, the sample is, filtering the SOS values at around 500-1000 is typically sufficient to filterout bad correlation curves arising from the presence of dust during the acquisition. But these

    numbers should not be taken as absolute or even as a rule of thumb. For longer acquisition

    times, the cutoff could be much lower. In a like fashion, multi-modal samples could lead to

    very large SOS values, as evident in the following figures.

    The figures below show the Cumulants Datalog, the correlation curves for Measurements #1 and

    #2, and the Regularization histograms for a PEGylated hemoglobin sample, before and after data

    filtering based on inconsistencies in the SOS values. A discontinuity is evident in the

    correlation curve for Measurement #2 (the same discontinuity is seen in Measurement #3),

  • 8/10/2019 Dls Help Manual

    26/66

    suggesting the presence of dust. The elevated SOS values in the Datalog for Measurements #2

    and #3 are a consequence of the discontinuities in the correlation curves. Prior to data filtering,

    the Regularization histogram indicates a bi-modal system. After Marking the dusty

    measurements (#2 and #3), the Regularization histogram shows a mono-modal system with a

    relatively low polydispersity.

  • 8/10/2019 Dls Help Manual

    27/66

    What are number fluctuations?

    Number fluctuations are caused by a non-constant particle concentration in the scattering

    volume. In light scattering experiments, the scattering intensity is proportional to both the MWand the concentration of the particles in the solution. As a consequence of Brownian motion,

    the particles within the scattering volume are moving, and the recombined scattering light from

    these moving particles leads to constructive and destructive interference effects or short time

    scale total intensity fluctuations at the detector. In dynamic light scattering, these intensity

    fluctuations are correlated and then transformed to give the diffusion coefficient and

    subsequently the hydrodynamic radius. During the acquisition of the correlation curve however,

    the particle concentration within the scattering volume must be constant. If the particle

    concentration fluctuates, then the average total intensity will be skewed or weighted by the

    variation in concentration, and the resultant correlation curve will be discontinuous (see figure

    below).

    Number fluctuations are particularly problematic with high sensitivity light scattering

    instruments such as the DynaPro-MS800. In order to produce a high power density in the

    scattering volume, the laser beam in the DynaPro-MS800 is focused down to a diameter of 16

    m. While the increased power density increases the sensitivity of the instrument to sampleconcentration, the smaller scattering volume increases the likelihood of number fluctuations if

    larger particles are present (see figures below). As a direct consequence of the smaller

    scattering volume, the upper size limit for the DynaPro-MS800 is around 100 nm (diameter).

    For the larger size range DynaPro instruments such as the MS/X and LSR, the laser beam

    diameter is focused to 100 m. This larger beam diameter increases the scattering volume, and

    while some sensitivity to concentration is lost, a much larger size range is allowed, with the

  • 8/10/2019 Dls Help Manual

    28/66

    MS/X and LSR instruments being able to accurately measure particle sizes up to 2 m diameter.

  • 8/10/2019 Dls Help Manual

    29/66

    What's the difference between the `old' and the `new' Regularization?

    In DYNAMICS software versions 5.26.36 and greater, the Regularization algorithm has been

    replaced with a new and improved Regularization algorithm. The fundamental difference in thetwo algorithms is that the regularization parameter in the new algorithm has been diminished

    slightly to be consistent with the smoothing parameters in other correlation curve analysis

    algorithms such as DynaLS and/or CONTIN.

    In addition to improving the algorithm, Protein Solutions, Inc. has also added a resolution slider

    to the Regularization histogram display in DYNAMICS. If the user has apriori information

    concerning the number of distributions, the resolution slider can be used to diminish the

    smoothing parameter even further and perhaps resolve multiple distributions from a single

    peak. The figures below show the resultant Regularization histograms at various resolution

    levels, using DLS data derived from a mixture of PEGylated and non-PEGylated chymotrypsin

    in 0.1 M NaCl after incubation for 3 hours at pH 8.0. At the lowest resolution level, thecomponents are unresolved and the histogram shows a single, highly polydisperse peak

    (ignoring the noise peak at 0.06 nm). The second figure shows the distribution at the resolution

    level for the old Regularization algorithm, and indicates two populations, one with mean radius

    of 3.58 nm and another with a mean radius of 25.5 nm. The former is consistent with the size of

    the PEGylated protein measured in the absence of the non-PEGylated component. The third

    figure shows the histogram distribution at the optimal resolution for the new Regularization

    algorithm. For this system, the results are the same, regardless of which Regularization

    algorithm is used. In the fourth figure, the resolution slider is set to the maximum level, and the

    resultant histogram indicates three different populations. Subsequent chromatographic analysis

    verified the presence of all three components, with the 1st peak being composed of

    oligio-peptide by-products of chymotrypsin self-digestion, the 3rd peak being aggregates of thesame, and the middle peak being the non-digested PEGylated chymotrypsin.

  • 8/10/2019 Dls Help Manual

    30/66

    While the above example indicated no significant difference between the old and newRegularization algorithms, there are occasions when the two algorithms can yield what looks to

    be different results. The figure below shows an example of this. The old Regularization

  • 8/10/2019 Dls Help Manual

    31/66

    algorithm indicates four populations. Because of the reduction in the smoothing parameter, the

    new Regularization algorithm is able to resolve the third peak into two distinct populations,

    yielding five peaks. However, if the resolution in the new algorithm is reduced by one

    increment on the slider (Unresolved New in figure below), the results from the new algorithm

    are consistent with those from the old.

  • 8/10/2019 Dls Help Manual

    32/66

    Do samples need to be filtered?

    The scattering intensity of a sample is proportional to both the concentration and the molecular

    weight of the particles in the solution. For high molecular weight particles, the contribution tothe total sample scattering intensity can be quite large. Hence, light scattering experiments are

    extremely sensitive to dust. As a general rule of thumb then, samples should always be filtered.

    The effects of dust on light scattering results are dependent upon a number of factors, primary of

    which are the concentration and the molecular weight of the analyte. The figures below show

    the DLS results and sample correlation curves for lysozyme solutions at 0.07 g/L (1st figure) and

    0.7 g/L (2nd figure), both prepared from unfiltered 0.1 M NaCl. In the first figure, for the low

    concentration sample, two measurements are marked as outliers (red). The correlation curves

    for these two runs are shifted upward, indicative of a small number of very large particles, i.e.

    dust, entering the scattering volume during the course of the measurement. In the second figure,

    for the high concentration sample, no outliers are noted. While the dust particles are stillpresent, their contribution to the total scattering intensity is insignificant, compared to that of the

    concentrated analyte.

  • 8/10/2019 Dls Help Manual

    33/66

    The figure below shows the DLS results for a high molecular weight 40 nm polystyrene

    microsphere standard at low concentration in the same unfiltered 0.1 M NaCl. As with the

    higher concentration lysozyme sample, the scattering contribution from the dust is insignificant

    and no outliers are noted. In this sample however, it is the high molecular weight analyte that

    diminishes the importance of removing dust via filtration.

  • 8/10/2019 Dls Help Manual

    34/66

  • 8/10/2019 Dls Help Manual

    35/66

    Will filtering perturb the sample?

    Depending upon the size of the analyte and the size of the filter pore, filtering can indeed perturb

    the sample. The figure below shows the DLS radius by Cumulants and DynaLS size distributionresults for an unfiltered poly(-cyclodextrin) sample. As evident in the DynaLS histogram, thesample is bi-modal (2 peaks) and polydisperse.

    The concentration dependence of the measured hydrodynamic radius for the large particle isshown below. The solid and open symbols are used to indicate filtered (100 nm Anotop) and

    unfiltered samples respectively. For concentrations 10.5 g/L, the measured radii are 61 and 75

    nm respectively for the filtered and unfiltered samples. The elevated values at 20.95 g/L are

    likely a result of interparticle interactions, which slow down the diffusion and lead to an

    increase in the apparent hydrodynamic radius.

  • 8/10/2019 Dls Help Manual

    36/66

    The 20% difference in radii between the filtered and unfiltered samples is a consequence of the

    removal of higher molecular weight particles during the filtering process. The figure belowshows the effects of filter pore size on the distribution of diameters for the poly(-cyclodextrin)sample. The dashed lines represent the pore cutoffs for the filters. As the filter pore size is

    decreased, the larger polymers are clipped out of the solution. The result is a decrease in the

    mean diameter of the larger particles and an increase in the contribution to the scattering

    intensity attributed to the smaller particles at circa 6 nm. Hence the better resolution of the two

    particle sizes with filtration.

  • 8/10/2019 Dls Help Manual

    37/66

    The size distributions shown previously are intensity distributions. If a flexible coil model,

    where the molecular weight is proportional to R2, is assumed with regard to particle shape, the

    intensity distributions can be converted to a %Mass distributions. The %Mass distribution for

    the filtered 10.5 g/L poly(-cyclodextrin) sample (200 nm filter) is shown below.

  • 8/10/2019 Dls Help Manual

    38/66

    As evident in the above figure, only 7% of the total mass of the poly(-cyclodextrin) sample canbe attributed to the large particle. This system exemplifies the effects of large particle scattering

    in dynamic light scattering studies. Even though there is only a small amount of the large

    particle present, 97% of the total scattering intensity is attributed to that particle. In the absence

    of filtering, information would have been lost, due to the inability to resolve the small, low

    polydispersity particle from the larger and more polydisperse one.

  • 8/10/2019 Dls Help Manual

    39/66

    Why is the salt content relevant?

    In dynamic light scattering experiments, it is the diffusion coefficient (D), rather than the

    hydrodynamic radius, that is directly measured. The hydrodynamic radius is calculated fromthe diffusion coefficient via the Stokes-Einstein equation. Any factors influencing the

    translational diffusion coefficient, whether or not they have any direct influence on the radius,

    will have an apparent effect on the calculated radius. At high sample concentrations for

    example, particle interactions can lead to a decrease in the diffusion coefficient and a subsequent

    increase in the calculated radius. Electrostatic interactions between the particles can lead to

    similar effects, even at low sample concentration. Electrostatic interactions can be shielded on

    the other hand, by the addition of salt. Hence the relevance of salt content in dynamic light

    scattering experiments.

  • 8/10/2019 Dls Help Manual

    40/66

    What is the Regularization Method?

    Protein Solutions, Inc. would like to thank Dr. Maria Ivanova and Dr. Ulf Nobbmann for their

    contribution of the following article.

    In Photon Correlation Spectroscopy (PCS), single photons are detected and auto-correlated.

    The autocorrelation function is the convolution of the intensity signal as a function of time with

    itself. If the detected intensity is described as a function I(t), then the autocorrelation function of

    this signal is given by

    The above function is also called the intensity correlation function. The shift time is oftenreferred to as the delay time since it represents the delay between the original and the shifted

    signal. In reality, the continuous intensity correlation function can not be measured. It can

    however, be approximated with discrete points obtained by a summation over the duration of the

    experiment, where the upper limit of the summation is given by the appropriate index belonging

    to the largest available summation term for k.

    The above general expression holds true for both linear and arbitrary delay times. Here, we arenot concerned with the specific technicality of obtaining the correlation function. That job is

    left to the digital correlator, usually a plug-in card or an external USB device connected to a

    personal computer. The correlator produces the normalized intensity correlation function and

    makes the measured values available for further interpretation. The remainder of this article

    then, focuses on the interpretation of the numerical values obtained from the correlator, namely

    an array of intensity correlation function values Gk and the corresponding delay time values k.

    If the intensity statistics of the measured signal are gaussian (which is true for all diffusion and

    for most random processes) then the Siegert relation holds true. The Siegert relation states that

    the normalized intensity autocorrelation function can be expressed as the sum of 1 and the

    square of the field autocorrelation function g (scaled with a coherence factor expressing theefficiency of the photon collection system).

    This equation can equivalently be written in the discrete form with the index k. For ideal

    detection, the coherency factor () = 1. As a first approximation then, we can ignore the

  • 8/10/2019 Dls Help Manual

    41/66

    coherency factor, and develop a table of gk and k values.

    Here the negative root value is only used for the unphysical case of G < 1 to approximate the

    experimental noise of real data. This method imposes less bias than either restricting data to the

    only the positive root or completely eliminating the outliers.

    The ideal field correlation function of hypothetical identical diffusing spheres is given by a

    single exponential decay function with decay rate determined by the diffusion coefficient and

    the wave vector of the scattered light. The main objective of the data inversion consists offinding the appropriate distribution of exponential decay functions which best describe the

    measured field correlation function. In mathematical circles this problem is known as a

    Fredholm equation of the first kind with an exponential kernel. It is also known as an ill-posed

    problem, since relatively small amounts of noise can significantly alter the solution of the

    integral equation.

    The fitting function for gk then consists of a summation of single exponential functions which is

    constructed as a grid of exponentials with decay rate i .

    The factor Ai is the area under the curve for each exponential contribution, and represents the

    strength of that particular ith exponential function. The best fit is found by minimizing the

    deviation of the fitting function from the measured data points, where a weighing factor k isincorporated to place more emphasis on the good, rather than the noisy, data points.

    The weights are proportional to the intensity correlation function values, i.e. correlation functionvalues at small times have a higher weight than those at large times. Consider for example, the

    following sets of figures. In the upper figure, the correlation curve baseline for measurement #1

    is noisy. This noise is absent in the lower curve (measurement #12). In the absence of a

    weighing factor, the noise could be interpreted as decays arising from the presence of very

    large particles. As a direct consequence of the weighting factor however, the subtle variations in

    the baseline for the upper correlation curve are correctly interpreted as noise by the

    Regularization algorithm, and no significant differences are observed in the resultant size

  • 8/10/2019 Dls Help Manual

    42/66

    distributions for the two curves.

  • 8/10/2019 Dls Help Manual

    43/66

    The normal procedure to find the solution of Ais in the grid of fitted gk expressions is to

    minimize the deviation 2 with respect to each Ai, and then find the solutions from the resultingsystem of equations. In other words, if we construct the solution out of N exponential functions

    there will be N differentiations of the following form.

    Each of the terms in the above differential contains a sum over k. The whole equation system

    can therefore conveniently be re-expressed in matrix form as shown below, where the Y-vector

    turns out to be a convolution of the experimental data with the kernel (= the grid matrix or the

    exponential decays) and the matrix W consists of a convolution of the kernel with itself.

    The standard procedure to solve this equation is to find the eigenfunctions and eigenvalues, and

    construct the solution as a linear combination of the eigenfunctions. However, when the

    eigenvalues are small, a small amount of noise can make the solution extremely big. Hence the

    previous ill-posed problem classification. To overcome the problem, a stabilizer () is added tothe system of equations. This parameter is called a regularization parameter, and with the

  • 8/10/2019 Dls Help Manual

    44/66

    incorporation of this term we are performing a first order regularization.

    The above expression is called a first order regularization because the first derivative (in Ai) isadded to the equation system. The regularization parameter determines how much emphasis

    we put on this derivative or the degree of smoothness in the solution. If is small, it has little

    influence and the solution can be quite choppy; whereas larger will force the solution to be

    very smooth. It is this value or degree of smoothness that the user controls with theresolution slider in the new Regularization algorithm in the DYNAMICS version 6 software.

    In addition to the smooth solution constraint, it is also required that the solution be physical; that

    is, all Ai be non-negative and uniformly constrained. In short, these additional constraints

    require that the summation of all Ai yields a finite number.

    With the above constraints, Z is minimized by requiring that the first derivatives with respect to

    Ai be zero. As indicated previously, this minimization corresponds to solving a system of linear

    equations in Ai. The solution of Ai values is found using an iterative approach called the

    gradient projection method.

    A spherical particle with radius Ri will produce a correlation function with decay rate iaccording to the following expression, where D is the translational diffusion coefficient, q is the

    scattering vector, kB is the Boltzmann constant, T is the absolute temperature, is the viscosity, is the solvent index of refraction, is the laser wavelength, and is the scattering angle.

    The normalized display of Ai vs. Ri is the %Intensity distribution of the Regularization

    histogram. The mean radius of a particular sub peak is the intensity weighted average, and is

    obtained from the histogram using the following expression.

    The polydispersity () (or standard deviation), indicative of the distribution in the sub peak, isalso obtained directly from the histogram.

  • 8/10/2019 Dls Help Manual

    45/66

  • 8/10/2019 Dls Help Manual

    46/66

    Can you get molecular weight information from DLS data?

    The molecular weight of a particle can be estimated from dynamic light scattering data using

    MW vs. RH calibration curves developed from standards of known molecular weight and size.However, because of the correlation between the specific volume (1/) of a particle and its

    tertiary conformation, a universal calibration curve has yet to be developed, although numerous

    calibration curves, each particular to a specific family or type of macromolecule are available.

    Hence, in order to select an appropriate MW vs. RH calibration curve, some a prioriinformation

    regarding the conformation of the particle must be assumed. The calibration curves included in

    the latest development version of the Dynamics software package are shown below. For

    reference purposes, it is noted that all of the proteins in the Protein family are globular;

    Pullulans are linear polysaccharides; Ficolls are densely branched polysaccharides; and

    Dendrimers are CO2 starburst type polymers, described as spherical, with a density that

    increases with radial distance from the core.

  • 8/10/2019 Dls Help Manual

    47/66

    How is the axial ratio calculated?

    The volume of a hypothetical sphere of homogeneous density can be calculated as shown below,

    where Rsph is the spherical radius, MW is the molecular weight, V (bar) is the partial specific

    volume (1/), and NA is Avogadros number.

    The radius of the sphere can be written as follows.

    Assuming stick boundary conditions, the frictional coefficient of this sphere (fo), can then be

    expressed as

    The above equation is also valid for non-spherical systems, with the exception that the measured

    hydrodynamic radius (RH) is used, rather than the hypothetical spherical radius.

    The Perrin or shape factor (F) then is defined as the ratio of the measured frictional coefficient

    to the frictional coefficient for a hypothetical sphere of the same radius.

    The above equation is that used in the Dynamics axial ratio calculator, where RH, Mw, and V

    (bar) are input values and F is an output value.

  • 8/10/2019 Dls Help Manual

    48/66

    Once the frictional ratio is determined, the axial ratios for oblate and prolate ellipsoids of

    revolution are interpolated from the data given in the following figure (see Cantor and

    Schimmel reference).

  • 8/10/2019 Dls Help Manual

    49/66

    How is the mass distribution of radii calculated?

    The mass distribution is calculated directly from the intensity distribution of radii via a

    simplified form of the Rayleigh equation developed from static light scattering theory.According to static theory, the scattering intensity of the ith particle is proportional to both the

    molecular weight (Mi) and the weight concentration (Ci) of the particle.

    The molecular weight of the ith particle can be estimated from the expression shown below,

    where vp is the partial specific volume and x is a shape parameter equal to 3 for spherical

    particles and 2 for coils.

    After substitution, the Rayleigh expression can be rewritten in the following form.

    The %Mass of the ith particle is then defined as:

    As evident in the above equations, the calculated %Mass is strongly dependent on the radius of

    the particle. This dependence is exemplified it the following figures, which show the

    Regularization distribution of radii displayed by %Intensity and %Mass for an 8000 molecular

    weight poly(ethylene glycol) sample in 0.1 M NaCl. Comparison of the figures reveals a 10%

    difference in the mean radius values for the two profiles, indicative of the 1/R2 dependence in

    the %Mass calculations for coils.

  • 8/10/2019 Dls Help Manual

    50/66

  • 8/10/2019 Dls Help Manual

    51/66

    Are the %Mass calculations valid for all systems and conditions?

    The complete expression describing the scattering intensity of a dilute solution of particles is

    shown below, where Io is the intensity of the incident radiation with a vacuum wavelength of o, o is the solvent refractive index, d/dC is the particle refractive index increment, A2 is the

    2nd virial coefficient, P() is a shape factor that ranges from 0 to 1, is the scattering angle, and

    r is the distance from the scattering center to the detector.

    In the %Mass calculations within the DYNAMICS software package, this equation is simplified

    to the following form, where it is assumed that d/dCi is independent of particle size, A2Ci

  • 8/10/2019 Dls Help Manual

    52/66

    What's the significance of the refractive index () in DLS?

    In typical dynamic light scattering studies, it is the autocorrelation function, an example of

    which is shown below, that is measured during the course of an experiment. The correlationfunction can be fit to an exponential form, or more correctly, a summation of exponentials, with

    the decay constants (i) containing the information needed to calculate the diffusion coefficients

    (Di) of the system of particles under examination.

    The diffusion coefficient is calculated using the expression shown below, where o is thesolvent refractive index, is the vacuum wavelength of the incident radiation, (= 1/) is the

    inverse of the exponential decay time, and is the scattering angle. As seen here, the solvent

    refractive index is introduced into the dynamic light scattering calculations through the

    scattering vector (K).

    A common misconception is that the subscript in o indicates that the refractive index to be

    used is that for the pure/base solvent. Such is not the case. The scattering vector is dependentupon the dielectric properties of the medium, with o entering the expression via the

    Clausius-Mosetti equation, which relates the polarizability to the permittivity of a molecule.

    Hence, the refractive index that should be used is that for the medium within which the particle

    is diffusing, i.e. the base solvent + any additives.

    The CRC Handbook of Chemistry and Physicscontains a significant collection of temperature

    and wavelength dependent refractive index values for aqueous mixtures of various

  • 8/10/2019 Dls Help Manual

    53/66

  • 8/10/2019 Dls Help Manual

    54/66

    Is the refractive index increment (d/dC) wavelength dependent?

    In Chapter 6 ofLight Scattering From Polymer Solutions(Ed. M.B. Huglin; Academic Press:

    New York, 1972), Huglin includes an extensive list of d/dC values (100 pages). According toHugglins empirical observations, the specific refractive index increment varies directly with -2. In order to verify this prediction, ten proteins were randomly selected from Huglins table

    and checked. The figure below exemplifies the d/dC dependence on -2 for all of the proteins,with the BSA and Lactoglobulin data being the outer limits of the entire set of ten. As evident

    below, the slope [(d/dC)/-2] is independent of protein type.

    The effects of ionic strength on d/dC for BSA and -lactoglobulin at pH 7 are indicated in the

    figure shown below. As seen here, (d/dC)/-2 is also independent of the ionic strength,although the addition of salt does lead to a decrease in the absolute d/dC value. Similar trends

    were observed for all of the proteins in the original set of ten. (It is noted, that all of the proteins

    are reported to be stable with regard to conformational changes under the cited conditions.)

  • 8/10/2019 Dls Help Manual

    55/66

    The independence of (d/dC)/-2 on protein type and solution salt content is likely a result of

    the globular nature of the proteins examined. The refractive index is correlated with the densityof the media. For globular proteins, the specific volume (1/) is relatively independent of

    protein type (vsp = 0.73 cm3/g). Hence the similarity between the d/dC and (d/dC)/-2values for the globular proteins examined here. As shown in the figure below however,

    significant changes in both d/dC and (d/dC)/-2 are observed when the macromoleculesare coiled, rather than globular.

  • 8/10/2019 Dls Help Manual

    56/66

    Ideally, d/dC should be measured under the appropriate solution conditions and at the specified

    wavelength, prior to beginning any static light scattering experiment. In the absence of theideal, one could estimate the protein d/dC value using the equation shown in the following

    figure, where the symbols are the same as those used in previous figures. The equation is that of

    the best fit line (black) for all ten of the protein data sets.

  • 8/10/2019 Dls Help Manual

    57/66

    The dashed lines are included in the above figure for estimating error. If solution conditions and

    protein type are ignored, i.e. if the equation is used, the error in d/dC is 3%. If thewavelength dependence is ignored, i.e. via the use of a single d/dC value for all proteins underall solution conditions and wavelengths, the error in d/dC can jump to upwards of 10-15%.

    With those observations in mind, it is instructional to consider the relative error in molecular

    weight arising from d/dC error in static light scattering measurements. The Rayleigh equation

    for static light scattering can be written as shown below, where is the solvent refractive index,

    Ris the Rayleigh ratio of scattering intensities, is the wavelength of the incident radiation,NA is Avogadros number, A2 is the 2nd virial coefficient, and C is the analyte weight

    concentration.

    The derivative with respect to d/dC of the above expression can be written as follows.

  • 8/10/2019 Dls Help Manual

    58/66

    After rearrangement, the relative error in the molecular weight can then be expressed in the form

    shown below. As evident in this equation, the relative d/dC error is amplified by a factor

    proportional to the 2nd virial coefficient, the concentration, and the molecular weight, when

    calculating the molecular weight from total intensity light scattering data.

    Static light scattering measurements using ovalbumin, with M 43 kDa and A2 0.0005, havebeen conducted at Protein Solutions, Inc. The highest concentration was 0.02 g/L. Using these

    parameters, the propagation factor (1 + 2A2CM) in the relative error equation is 1.0009 1.Under these conditions then, the relative error in calculated molecular weight is equivalent to the

    relative error in the refractive index increment, implying that the 3% difference in d/dC for

    globular proteins, due to different solvent conditions, is negligible. The 10 - 15% error in

    d/dC arising from wavelength effects however, could be problematic, depending upon the

    exact solution conditions employed and the properties (M and A2) of the particle under

    examination.

    Editors Note (10/13/00)

    Please note that earlier versions of this document contained a unit error in the calculations of the relative error inovalbumin molecular weight and in the empirical equation for the wavelength dependence of d/dC. Wed like to

    thank our friends at Wyatt Technology for bringing these mistakes to our attention (Thanks Ron - as always your

    reviews and comments are greatly appreciated). The mistakes has been corrected, and the conclusions modified to

    be consistent with the corrected value of the propagation factor in the relative error calculations. We apologize for

    any inconvenience this oversight may have caused.

  • 8/10/2019 Dls Help Manual

    59/66

    Is the solvent refractive index (o) wavelength dependent?

    The refractive index () of a medium is defined as the ratio of the speed of light in a vacuum to

    that in the medium. As the light passes through an interface into a medium of higher density,say for example from a vacuum into a solvent, the decreased velocity of the radiation in the

    solvent causes the light to be refracted or bent. According to Snells law, the magnitude of the

    refraction, and hence the velocity of the radiation in the solvent, is dependent upon the

    wavelength of the incident beam. Ergo, since the velocity of light in the solvent is wavelength

    dependent, the measured solvent refractive index will be dependent upon the same.

    On a molecular level, the decrease in the velocity of the radiation in the solvent arises from the

    electromagnetic interactions of the light wave with the solvent molecules. If the solvent

    molecules are more densely packed, the magnitude of the decrease in light velocity is

    proportionate. Hence, the refractive index of the solvent is strongly correlated with the solvent

    density. The dependence of the refractive index on density is described by the Lorentz-Lorenzformula shown below, where R is the molar refraction, M is the molecular weight, and is thedensity.

    The figure below shows a compilation of refractive index values, extracted from the 77th edition

    of the CRC Handbook for Chemistry and Physics, for pure water at different wavelengths and

    temperatures. The linearity of the data at constant temperature is consistent with the Cauchy

    expression, which predicts that the refractive index will be proportional to -2.

  • 8/10/2019 Dls Help Manual

    60/66

    As evident in the above figure, the solvent refractive index is also dependent on temperature.

    This dependence arises from the temperature dependence of the solvent density. As the

    temperature of the solvent is increased, the density decreases, leading to a subsequent increase in

    the velocity of the radiation and a decrease in the measured refractive index. The equation

    shown below, where T is the temperature in Celsius and is the wavelength in nm, is an

    empirical expression developed from the above pure water refractive index data. While notgrounded in theory, it does simplify calculations of H2O for DLS experiments under normal

    temperature and wavelength conditions.

    Because of the wide use of toluene as a Rayleigh ratio standard in static light scattering

    experiments, assorted temperature and wavelength dependent toluene refractive index values are

    also included here. As evident in the figure below, Tol is proportional to -2, consistent with

    the predictions of the Cauchy equation.

  • 8/10/2019 Dls Help Manual

    61/66

    Are o and d/dC temperature dependent?

    Both the refractive index () and specific refractive index increment (d/dC) are closely

    correlated with the density, and are hence sensitive to temperature. The correlation of therefractive index properties with density is evident in the figure below, which shows the

    temperature dependence (0-25 oC) of d/dC for selected globular proteins. The similarity in the

    (d/dC)/T tends is a direct consequence of the similarity in the partial specific volumes (1/)of the proteins.

    The solvent refractive index is also temperature dependent, as evident in the following figures,

    which show the reported (CRC Handbook of Chemistry and Physics) refractive index values for

    pure water and toluene respectively, plotted against the square of the temperature (T2) in Celsius

    at various wavelengths (). In the 1st figure, the slopes of the vs. T2 lines are independent ofwavelength, facilitating development of the empirical expression for H2O shown at the bottom

    of the figure. The dependence of the refractive index on -2 is discussed elsewhere. As shownin the legend of the 2nd figure, the slopes of the vs. T2 lines are wavelength dependent for

    toluene. Hence, development of an empirical expression relating the toluene refractive index to

    both the temperature and the wavelength is more problematic.

  • 8/10/2019 Dls Help Manual

    62/66

  • 8/10/2019 Dls Help Manual

    63/66

  • 8/10/2019 Dls Help Manual

    64/66

    Is it alright to estimate solvent viscosity values?

    The fundamental parameter extracted from the measured autocorrelation curve in dynamic light

    scattering studies is the diffusion coefficient (D). The solvent viscosity () enters the picture via

    the Stokes-Einstein equation for calculating the hydrodynamic radius (RH) of freely diffusing

    spheres, assuming non-slip boundary conditions.

    The standard error propagation expression for RH is shown below. Temperatures can be

    measured quite precisely. Hence, T/T is negligible. If the questionable assumption that the

    relative error in D (D/D) is negligible is also made, the best that can be expected is that the

    %error in the radius is equivalent to the %error in the estimated viscosity value.

    If the propagation of error ended here, small errors in the estimated viscosity would probably be

    acceptable. Typically however, our clients also desire the particle molecular weight, which

    cannot be directly calculated from the Stokes-Einstein equation and the measured diffusion

    coefficient. As such, the molecular weight can only be estimated, using MW vs. RH calibration

    curves developed from standards of know mass, size, and molecular conformation. Since these

    calibration curves are empirical, error in the solvent viscosity is propagated even further, if

    particle molecular weight is the desired parameter

    With regard to the question then of whether or not it is OK to estimate solvent viscosity, the

    answer depends upon how much error one is willing to accept in the target parameter and what

    other a prioriinformation is available. If one it interested only in the radius of a system of

    particles of singular size, then small errors in the viscosity might be acceptable. If however, the

    solution is more complex or the target parameter is the estimated molecular weight, error in the

    solvent viscosity becomes problematic. For example, the measured hydrodynamic radii for

    dimeric and tetrameric insulin are 1.4 and 1.7 nm respectively. With only a 21% difference in

    the size of the two inactive forms of insulin, a 10% error in the solvent viscosity can be fatal, if

    one is attempting to determine the predominant form under a given set of solution or preparatory

    conditions.

    The CRC Handbook of Chemistry and Physicscontains an enormous collection of temperature

    dependent viscosity values for solvents of various composition. The figure below shows a

    compilation of 20 vs. W% (in water) data for some typical additives, extracted from the 77th

    edition of the CRC. The empirical form of the 20 dependence on W% for all of the additives

    presented here is a 2nd order polynomial, the coefficients of which are given in the figure legend.

  • 8/10/2019 Dls Help Manual

    65/66

  • 8/10/2019 Dls Help Manual

    66/66

    What is the _____ value of ______?

    A number of solvent parameters, such as the viscosity and the refractive index, are incorporated

    into the calculations in dynamic and static light scattering studies. Depending upon theconditions, error in the values for the solvent parameters can lead to significant error in the

    results. Hence, questions of the form presented above are common. It would be impossible to

    address all the permutations; but we can certainly try. The figure below shows a screen shot of

    the new Archive utility developed at Protein Solutions, Inc. While it isnt as complete as the

    CRC Handbook of Chemistry and Physics, it is always available and you can edit and/or add

    your own information. This utility is currently available for download at the Protein Solutions,

    Inc. web page, and will be included in the Dynamics 6.0 package.

    ArchiveSetup Readme.txt