dissolved oxygen analyzer user manual

Upload: abvsai

Post on 02-Jun-2018

284 views

Category:

Documents


4 download

TRANSCRIPT

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    1/92

    Operator Manual

    POLYMETRON 9182Oxygen analyzer (ppb)

    221=191=082 - Revision G - 20/11/2007

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    2/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    3/92

    Restriction of Hazardous Substances

    Note:The following only applies to exports of the product into the Peoples Republic of China.

    Marking

    Products contain toxic or hazardous substances or elements.

    Environment Protection Use Period Marking (years).

    Toxic or Hazardous Substances and Elements

    Part NameLead(Pb)

    Mercury(Hg)

    Cadmium(Cd)

    HexavalentChromium

    (Cr VI)

    PolybromBiphenyls

    (PBB)

    PolybromDiphenyls

    (PBDE)

    Transmitter box X

    CPU PCB O O

    Power PCB O O

    Relay PCB O O

    Module O

    Probes O

    Cable O

    O: Indicates that this toxic or hazardous substance contained in all homogeneous material for this part is below the limit requirement

    X: Indicates that this toxic or hazardous substance contained in at least one of thehomogeneous materials used for this part is above the limit requirement

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    4/92

    Oxyg en analyser (ppb) - 9182 -

    i

    This instrument conforms to the European Directives:

    - 89/336/CEE modified by the directive 93/68/CEE- 73/23/CEE modified by the directive 93/68/CEE

    Warning!

    There are no user-serviceable parts in either the transmitter or sensor. Only Hach Ultra Analytics personnel or their authorized representative should attempt repair of the system andonly components expressly approved by the manufacturer should be used. Any attempt torepair the instrument in contradiction of these guidelines may result in damage to theinstrument and injury to the person making the repair. It will also void the warranty and maycompromise the safe operation, electrical integrity or CE compliance of the instrument.

    Note:

    This equipment has been tested and found to comply with the limits for Class A digital device,pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonableprotection against harmful interference when the equipment is operated in a commercialenvironment. This equipment generates, uses, and can radiate radio frequency energy and, ifnot installed and used in accordance with the instruction manual, may cause harmfulinterference to radio communications. Operation of this equipment in a residential area is likelyto cause harmful interference in which case the user will be required to correct the interferenceat his own expense.

    Precautionary Labels:

    Read all labels and tags attached to the instrument. Personal injury or damage to thisinstrument could occur if not observed.

    This symbol, if noted on the instrument, references the instruction manual foroperation and / or safety information.

    Electrical equipment marked with this symbol may not be disposed of inEuropean public disposal systems after 12 August of 2005. In conformity with

    European local and national regulations (EU Directive 2002/96/EC), Europeanelectrical equipment users must now return old or end-of life equipment to theProducer for disposal at no charge to the user.

    Note: For return for recycling, please contact the equipment producer or supplierfor instructions on how to return end-of-life equipment for proper disposal.Important document. Retain with product records.

    !

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    5/92

    Oxyg en analyser (ppb) - 9182 -

    ii

    COPYRIGHT

    All rights reserved. No part of this publication may bereproduced, stored in a retrieval system, or transmitted, in anyform or by any means, electronic, mechanical, photocopying,recording or otherwise, without prior permission of Hach Ultra

    Analytics .

    Hach Ultra Analytics

    Hach Ultra Analytics can take no responsability for installationand/or use of its equipment if this is not done in accordancewith the appropriate issue and/or amendment of the relevantmanual.

    The user of this manual should ensure that it is appropriate inall details to the exact equipment to be installed and/oroperated. If in doubt, the user should contact Hach Ultra

    Analytics for advice.

    WARNING

    To maintain safety standards, regular maintenance,calibration and operation of this equipment by qualifiedpersonnel is essential. Read and understand Instructionmanual completely before operating or servicing. If anyfurther details are required which do not appear in thismanual contact Hach Ultra Analytics or their agent.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    6/92

    Oxyg en analyser (ppb) - 9182 -

    iii

    Warnings

    Changing the programming language

    The programming language is English when factory-programmed, when changingthe software version and when loading the default values. To change the languagefollow the procedure below (example for French):

    (Press 3 times)

    !

    AVERAGE

    CODESOFT ISSUEDEFAUT VAL.ADJUST mACONFIGURATIONPOLYMETRON

    SERVICE

    Select

    DISPLAY

    CALIBRATIONMAINTENANCEPROGRAMMING

    MENU

    Select

    SERVICE

    CONC : ppb/ppm

    TEMP. : C

    PRESSION : mmHg

    LANGUE :

    AFFICHAGE

    Choix

    F

    CONC : ppb/ppmTEMP. : C

    PRESSURE : mmHg

    LANGUAGE :

    DISPLAY

    Select

    GB

    CONC : ppb/ppm

    TEMP. : C

    PRESSURE : mmHg

    LANGUAGE :

    DISPLAY

    Select

    F

    6.59ppb

    23.2 C O2

    Disp2 Menu Enter

    Enter

    Enter

    Esc

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    7/92

    Oxyg en analyser (ppb) - 9182 -

    iv

    Programming the mains supply frequency:

    When your programming language is chosen, you may change the mains supplyfrequency if necessary. This change occurs at the first switching on andafterresetting the instrument. Follow the procedure below:

    AVERAGEDISPLAYCODESOFT ISSUEDEFAULT VAL.

    ADJUST mA

    POLYMETRON

    SERVICE

    Select

    CONFIGURATION

    CALIBRATIONMAINTENANCEPROGRAMMING

    MENU

    Select

    SERVICE

    FREQ :

    CONFIGURATION

    Select

    50Hz

    6.59ppb

    23.2C O2

    Disp2 Menu Enter

    Enter

    FREQ :

    CONFIGURATION

    60Hz

    Enter

    !

    Enter Resetthe instrument

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    8/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    9/92

    Oxyg en analyser (ppb) - 9182 -

    Table of contents

    Page

    Warnings ....................................................................................................................................iii

    Chapter 1: Introduction............................................................................................................. 1-1 1.1 Introduction........................................................................................................ 1-1 1.2 Principle of operation........ ......... ......... ........ ......... ........ ......... ........ ......... ........ ...... 1-1 1.3 Main characteristics............................................................................................ 1-3 1.4 Technical characteristics.... ........ ......... ......... ........ ......... ........ ......... ........ ......... .... 1-3 1.5 Dimensions ........................................................................................................ 1-6

    Chapter 2: Description of the analyser......... ......... ........ ......... ........ ......... ......... ........ ......... ........ 2-1

    2.1 Analytical part .................................................................................................... 2-1 2.2 Hydraulic mounting ............................................................................................. 2-2 2.3 Transmitter......................................................................................................... 2-3

    2.3.1 Presentation of the transmitter ........ ......... ........ ......... ........ ......... ........ ...... 2-3 2.3.2 Application fields .................................................................................... 2-5 2.3.3 Standard mounting possibilities (use of the red clamping bow) .... .... .... .... .... 2-5

    Chapter 3: Installation of the instrument......... ......... ........ ......... ........ ......... ........ ......... ........ ...... 3-1 3.1 Unpacking.......................................................................................................... 3-1 3.2 Inspection .......................................................................................................... 3-1 3.3 Mounting............................................................................................................ 3-1 3.4 Location............................................................................................................. 3-1 3.5 Electrical connections ......................................................................................... 3-2

    3.6 Description of the different terminals ........ ........ ......... ......... ........ ......... ........ ......... . 3-3 3.7 Mains connection ............................................................................................... 3-4 3.8 Starting the transmitter........ ......... ........ ......... ......... ........ ......... ........ ......... ........ ... 3-4 3.9 Adjusting the display contrast.......... ........ ......... ........ ......... ......... ........ ......... ........ 3-4

    Chapter 4: Using the instrument............. ......... ......... ........ ......... ........ ......... ........ ......... ........ ...... 4-1 4.1 Front panel keys................................................................................................. 4-1 4.2 Displays 1 to 4 (live displays)........ ........ ......... ......... ........ ......... ........ ......... ........ ... 4-2 4.3 Description of the function keys ........ ......... ........ ......... ........ ......... ........ ......... ....... 4-3 4.4 Icons ................................................................................................................. 4-3 4.5 Enter or modify a value..... ......... ......... ........ ......... ........ ......... ........ ......... ........ ...... 4-4 4.6 Warnings ........................................................................................................... 4-4

    Chapter 5: Programming the transmitter ........ ......... ........ ......... ........ ......... ........ ......... ........ ...... 5-1

    5.1 Main menu......................................................................................................... 5-1 5.1.1 Calibration Menu........... ......... ........ ......... ........ ......... ........ ......... ........ ...... 5-2 5.1.2 MAINTENANCE Menu......... ........ ......... ........ ......... ......... ........ ......... ........ 5-3 5.1.3 PROGRAMMING Menu.......... ........ ......... ........ ......... ........ ......... ........ ...... 5-3

    5.1.3.1 MEASURE Menu ....................................................................... 5-4 5.1.3.2 ALARMS Menu .......................................................................... 5-5 5.1.3.3 mA OUTPUTS menu............ ........ ......... ........ ......... ........ ......... .... 5-8 5.1.3.4 RS485 Menu.............................................................................5-11

    5.1.4 SERVICE Menu.....................................................................................5-12 5.1.4.1 AVERAGE Menu.......................................................................5-12 5.1.4.2 DISPLAY Menu.........................................................................5-13 5.1.4.3 CODE Menu ........ ......... ........ ......... ......... ........ ......... ........ .........5-14

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    10/92

    Oxyg en analyser (ppb) - 9182 -

    5.1.4.4 SOFT VERSION Menu.............. ......... ........ ......... ........ ......... ......5-15 5.1.4.5 DEFAUT VAL. Menu..................................................................5-15 5.1.4.6 mA ADJUST Menu ........ ........ ......... ......... ........ ......... ........ .........5-16 5.1.4.7 CONFIGURATION Menu ........ ......... ......... ........ ......... ........ .........5-17 5.1.4.8 POLYMETRON Menu ........ ......... ........ ......... ........ ......... ........ .....5-17

    Chapter 6: Calibrating temperature compensation... ......... ........ ......... ........ ......... ........ ......... .... 6-1 6.1 Calibration of the temperature sensor............... ......... ......... ........ ......... ........ ......... . 6-1

    6.1.1 Automatic temperature compensation............... ......... ........ ......... ........ ...... 6-2 6.1.2 Manual temperature compensation...... ......... ......... ........ ......... ........ ......... . 6-3

    6.2 Calibration of the measurement. ........ ......... ........ ......... ........ ......... ........ ......... ....... 6-4 6.2.1 Slope calibration in the air + electrical zero ......... ........ ......... ........ ......... .... 6-4 6.2.2 Slope calibration in the air + chemical zero ......... ........ ......... ........ ......... .... 6-5 6.2.3 Slope process calibration + electric zero........... ......... ........ ......... ........ ...... 6-6 6.2.4 Slope process calibration + chemical zero ........ ......... ........ ......... ........ ...... 6-7

    Chapter 7: Start up, maintenance and troubleshooting ......... ........ ......... ......... ........ ......... ........ 7-1

    7.1 Start up ............................................................................................................. 7-1 7.1.1 Assembling the probe ............................................................................. 7-1 7.1.2 Probe connection................ ........ ......... ........ ......... ......... ........ ......... ........ 7-3 7.1.3 Main power supply connection ........ ......... ........ ......... ........ ......... ........ ...... 7-3 7.1.4 Starting up the analyser ......... ........ ......... ........ ......... ........ ......... ........ ...... 7-3

    7.2 Changing the membrane............. ......... ......... ........ ......... ........ ......... ........ ......... .... 7-4 7.2.1 Dismounting the membrane........ ........ ......... ......... ........ ......... ........ ......... . 7-5 7.2.2 Mounting the new membrane ......... ........ ......... ........ ......... ........ ......... ....... 7-6

    7.3 Oxygen electrode rejuvenation procedure ........ ......... ........ ......... ........ ......... ........ ... 7-7 7.4 Functional troubleshooting.. ........ ......... ......... ........ ......... ........ ......... ........ ......... .... 7-7 7.5 Electrical troubleshooting ......... ........ ......... ........ ......... ........ ......... ........ ......... ......7-11

    Chapter 8: Error messages......... ........ ......... ........ ......... ........ ......... ......... ........ ......... ........ ......... . 8-1

    Appendix 1: Pressure conversion table ......... ........ ......... ........ ......... ........ ......... ........ ......... ..... A1-1

    Appendix 2: Temperature conversion table........... ......... ........ ......... ........ ......... ........ ......... ..... A2-1

    Appendix 3: Default values........ ........ ......... ........ ......... ........ ......... ......... ........ ......... ........ ........ A3-1

    Appendix 4: Detailed cable connections ........ ......... ........ ......... ........ ......... ........ ......... ........ .... A4-1

    Appendix 5: Security data sheet........... ......... ........ ......... ........ ......... ........ ......... ........ ......... ..... A5-1

    Appendix 6: Spare parts list....... ........ ......... ........ ......... ........ ......... ......... ........ ......... ........ ........ A6-1

    Appendix 7: RS485 MODBUS-JBUS addressing............ ........ ......... ......... ........ ......... ........ ........ A7-1

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    11/92

    Oxyg en analyser (ppb) - 9182 -

    1-1

    Chapter 1: Introduction

    1.1 Introduction

    The 9182 oxygen analyser is a one-channel analyser for the measurement of dissolved oxygen inthe boiler feedwater, economizers, condensers and more generally all thermal equipment usingwater as a heat transfer liquid.

    1.2 Principle of operation

    The measurement of dissolved oxygen is based on the well-known Clark cell principle.

    An oxygen-permeable membrane isolates the electrodes from the sample water, thus obviating theneed for sample conditioning. Other reducible or oxidizable ions do not interfere, because theycannot pass through the gas-permeable membrane.

    A constant voltage supply powers two electrodes, maintaining each at a constant potential.

    A gold working electrode (cathode) reduces the dissolved oxygen to hydroxyl ions:

    O 2 + 2H 2O + 4e- 4OH -

    A large silver counter electrode (anode) provides the oxidation reaction which occurs on itssurface:

    4Ag + + 4Br - 4AgBr + 4e -

    The reduction of oxygen is the current limiting reaction, thus making the cell current linearlyproportional to the dissolved oxygen concentration.

    Electrochemical reactions and diffusion rates are temperature-sensitive. The measuring cell,therefore, is equipped with a temperature sensor which allows an automatic temperaturecompensation.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    12/92

    Oxyg en analyser (ppb) - 9182 -

    1-2

    Principle of oxygen cell

    1

    2

    3

    5

    2

    6

    7

    4

    27

    8

    9

    1 : Assembled electrode 6 : Membrane holder

    2 : Electrolyte 7 : Membrane

    3 : Probe body 8 : Cathode

    4 : Filling hole 9 : Sample

    5 : Anode

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    13/92

    Oxyg en analyser (ppb) - 9182 -

    1-3

    1.3 Main characteristics

    3 Range 0-2 ppm3 Calibration in the air

    3 Temperature compensation

    3 Programmable alarm levels, outputs on relays

    3 4-20 mA, 0-20 mA analogue outputs (standard) and RS485 (option)

    3 Wall- , panel- and tubing mounting

    1.4 Technical characteristics

    SAMPLE

    Number of channels 1

    Temperature 0-45 C (32-113 F)

    Working pressure Atmospheric pressure

    Flowrate 4...10 l/h

    ELECTRICAL CHARACTERISTICS Mains Standard version :

    100...240 VAC - 50/60 Hz

    Low voltage version :13...30 VAC 50/60 Hz18...42 VDC

    Power consumption 25 VA

    Connections 2,5 mm screw terminal

    Fuse 5 x 20 mm cartridge - T2AL - 250V

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    14/92

    Oxyg en analyser (ppb) - 9182 -

    1-4

    ANALYSIS

    Measuring range 0...2 ppmSensibility 0,5 ppb

    Reproducibility 0,5 ppb or 2% whichever the greater

    Detection limit 1 ppb

    Response time (90 %) 3 min

    Ambient temperature -20.. .60 C (-4...140 F)

    Relative humidity 10...90%

    TRANSMITTER

    Display Display in concentration unitsDisplay of the direct, concentrationdependent cell current in ADisplay of the sample temperature in C/FProgramming via menus

    Units nA, A, ppb-g/l, ppm-mg/l, C, F

    Calibration Electrical zero, chemical zero, slope calibrationin the air, slope process calibration bycomparison with a laboratory measurement

    Interferences Any gas liable to reduce to a potential> - 850 mV

    Analogue outputs 2 x 0/4... 20 mA isolated from input signal, 800Ohms load maximum- for the measure or for the temperature- mode: linear, bi-linear- accuracy: 0,1 mA

    Alarms Number: 4- Functions: alarm - system alarm - timer- Hysteresis: 0...10%- Delay: 0...999s

    - Breaking power: 250 VAC, 3A maximum30 VDC, 0,5A maximum

    Use a cable (rated 105 C and AWG22 to 14).The external cable insulation should be cut asclose as possible from the terminal block.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    15/92

    Oxyg en analyser (ppb) - 9182 -

    1-5

    Temperature compensation Automatic in the range of 0...45 C(32...113 F)

    RS485 (option) Speed: 300...9600 baudInsulation: galvanicStation number: 32 maximum

    Transmitter protection IP65 (NEMA 4X in option)

    Measurement category I (overvoltage less than 1500 V)

    Error reports Cell current > 999 ASample temperature > 45 C (113 F)or < 0 C (32 F)

    Ambient temperature > 45 C or < 0 C(32...113 F) during air calibrationSlope calibration errorZero calibration error (offset)

    ELECTROMAGNETIC COMPATIBILY

    European standards EN 61326-1997 and EN 61326 A1-1998(Industrial level for immunity)

    Low voltage standard IEC61010-1

    UL and CSA agreement File E226594

    MATERIALS

    Working electrode Cathode: gold

    Counter-electrode Anode: silver

    Membrane holder Noryl

    Membrane PFA

    Transmitter Aluminium + polyester painting

    Probe body Noryl

    Probe with optional immersion Stainless steel 316L

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    16/92

    Oxyg en analyser (ppb) - 9182 -

    1-6

    MAINTENANCE

    Approximative membrane lifetime 6 months depending on the sample

    Cleaning Clean the instrument with a soft tissue andwithout any aggressive agent

    OPERATING CONDITIONS

    Ambient temperature -20...+60 C

    Relative humidity 10...90 %

    Power supply voltage fluctuation 10 %

    Over voltage category 2

    Pollution degree 2 (as CEI 664)

    Altitude < 2 000 m

    Measurement category I (overvoltage less than 1500 V)

    1.5 Dimensions(Dimensions are in mm and [in inches])

    polymetron

    polymetron

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    17/92

    Oxyg en analyser (ppb) - 9182 -

    2-1

    Chapter 2: Description of the analyser

    2.1 Analytical part

    1

    2

    3

    4

    56

    Figure 2-1: Electrode cross-section Figure 2 -2: Complete probe

    1 : Connector 1 : Adjusting nut

    2 : Anode holder 2 : Electrode

    3 : Anode 3 : Probe body

    4 : Cathode holder 4 : Membrane

    5 : Cathode 5 : Filling screw

    6 : Temperature sensor 6 : Gasket

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    18/92

    Oxyg en analyser (ppb) - 9182 -

    2-2

    2.2 Hydraulic mounting

    (Dimensions are in mm [inches])

    1 : Cable 5 : Cell

    2 : Shrink ring 6 : 1/4" NPT connector (not provided)

    3 : Assembled probe 7 : M4 screws (not provided)

    4 : Probe gasket 8 : 1/4" NPT connector (not provided)

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    19/92

    Oxyg en analyser (ppb) - 9182 -

    2-3

    2.3 Transmitter

    2.3.1 Presentation of the transmitter

    Esc Enter

    18.6 Cppb

    2.10

    Polymetron

    O2Disp2 Menu

    S1

    S2 S4

    S3

    The electronic unit amplifies the signal of the amperometric measuring cell and converts it into adirect digital readout in ppm, mg/l, ppb, g/l, C and F. The transmitter comprises the following

    items:n Potentiostat which maintains the working electrode potential constantn Amperometric measuring modulen Analog multiplexern Microprocessor unit

    The analog multiplexer allows measurements to be acquired from the measuring cell, temperaturesensor and internal checkpoints. Further, the microprocessor operates the relays, the RS485interface (optional) and the analog outputs.

    The unit has a built-in concentration-autoranging feature and a microprocessor-operatedcalibration routine.

    The output of the potentiostat is monitored for possible overdriving of the potentiostat-output stage.This condition can occur with the connections to the measuring cell open, inoperable electrodes ora defective reference electrode.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    20/92

    Oxyg en analyser (ppb) - 9182 -

    2-4

    Transmitter synoptic below is:

    1 : Programmable potentiostat 6 : Auxiliary input

    2 : Polarization voltage amplifier 7 : Multiplexer

    3 : Reversing switch for a 2 or 3 electrodeoperation. In the 9181 the configuration is2 electrodes

    8 : A/D converter

    4 : Temperature measurement circuit 9 : Microprocessor

    5 : Measurement circuit current amplifier

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    21/92

    Oxyg en analyser (ppb) - 9182 -

    2-5

    2.3.2 Application fields

    Easy-to-use (installation, programming), this instrument is suitable for the following applications:

    - boiler feedwater,- ultrapure water.

    2.3.3 Standard mounting possibilities (use of the red clamping bow)

    The transmitter housing conforms to norm DIN 43700.

    Panel mounting

    Panel cutting : 138 138 mmFront panel dimensions : 144 144 mm

    2 screws 4 mm lg 16 flat head (provided) for panel thickness 0 to 4 mm 2 screws 4 mm lg 20 flat head (provided) for panel thickness 4 to 8 mm

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    22/92

    Oxyg en analyser (ppb) - 9182 -

    2-6

    Wall mounting

    2 screws 4 mm lg 60 flat head (not provided) / 80 mm center distance

    Vertical pipe mounting 2" maximum - 2 screws - 4 mm lg 60

    (provided)

    Horizontal pipe mounting 2" maximum - 2 screws 4 mm lg 60(provided)

    polymetron

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    23/92

    Oxyg en analyser (ppb) - 9182 -

    3-1

    Chapter 3: Installation of the instrument

    3.1 Unpacking

    The analyzer should be unpacked with great care. Make sure not to loose any accessory whenunpacking.

    3.2 Inspection

    The analyzer has been factory-checked and tested prior to shipment, it is however advisable toinspect all parts immediately upon receipt for any damage which may have occurred duringshipment. A damaged shipping container may indicate internal damage which may not beimmediately obvious. If there is any evidence of damage, keep the shipping container and refer toyour local agent or to:

    Hach Ultra Analytics6 route de Compois - CP212CH-1222 Vesenaz Geneva

    3.3 Mounting

    CAUTION!

    Mounting should be done by qualified service personnel only. No power should beapplied until the installation is complete.

    3.4 Location

    The analyser should be located in an accessible site.

    The site should permit the access for any checking or maintenance operation.

    The sample flowrate has to be maintained within the specifications (See 1-4).

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    24/92

    Oxyg en analyser (ppb) - 9182 -

    3-2

    3.5 Electrical connections

    3 MONEC

    Do not switch on the instrument until completion of the installation.

    An aluminium armour plate inside the MONEC gives a detailed description of the different terminalsand their connections:

    1. Microprocessor board2. Relay board3. Amperometric board4. Power supply board5. Program update connector

    The different terminals represented on the right side are accessible by removing the armour plate.

    1

    2

    3 45

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    25/92

    Oxyg en analyser (ppb) - 9182 -

    3-3

    3.6 Description of the different terminals

    Description Connection020 mA or 420 mA (n 1) [+] user

    020 mA or 420 mA (n 1) [-] user

    020 mA or 420 mA (n 2) [-] user

    0/4...20 mA outputsgalvanically insulated

    020 mA or 420 mA (n 2) [+] user

    userOption RS485user

    Description Colour ConnectionTemperature sensor [+] black temp +Temperature sensor [-] blue temp -

    Reference if use of3 electrodes

    not used on the 9182oxygen analyser

    Counter electrode (anode) red counterWorking electrode (cathode) white work

    Auxiliary input aux.

    Amperometric module

    External shield braid armour plate

    Internal shield brown GND

    Not used for the oxygen measurement n.c.

    Mains, 100240 VAC 50/60 Hz or 24 V AC/DC (special version)

    Description Connection Alarm 1, simple contact user Alarm 2, simple contact user Alarm 3 or alarm system, simple contact user

    Alarm 4 or timer, simple contact user

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    26/92

    Oxyg en analyser (ppb) - 9182 -

    3-4

    Electrical connections should remain dry to ensure a proper operation of theinstrument.

    Check the creeping of the cables when opening the transmitter. It is required to use shielded cables. This shielding should be connected to theearth central shielding.

    3.7 Mains connection

    Electrical connection should be performed only by qualified personnel. The power supply accepts100-240 VAC 10 %, (50/60 Hz) without changes in configuration. The terminal block for powerconnections can be lifted from its header for easier installation. For safety reasons, it is required to

    observe the precautions below:

    Use a three wire mains supply cable (2 core + PE) with a cross section between 0.35 and 2 mm 2 (AWG 22 to 14) rated at 105 C minimum. The external cable insulation should be cut as closeas possible from the terminal block.

    The instrument should be connected to the power supply by means of a breaker located closeto the instrument and be identified. The supply shall be fitted with an overcurrent protectiondevice rated at 20 Amp maximum.

    This breaker should switch off phase and neutral in case of electrical problems or when the userwish to service the instrument. However the power supply earth must always be connected.

    Before servicing the instrument, ensure that the power supply is switchedoff.

    3.8 Starting the transmitter

    Before switching on the transmitter, make sure the site voltage corresponds to the instrumentvoltage indicated on the identification plate.

    3.9 Adjusting the display contrast

    If your display contrast is not sufficient, you can adjust it with the potentiometer P1 (blue colour, seefigure page 3-2) which is located on the top left of the CPU board (after opening the enclosure).

    !

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    27/92

    Oxyg en analyser (ppb) - 9182 -

    4-1

    Chapter 4: Using the instrument

    4.1 Front panel keys

    The display may be programmed to indicate:

    - Sample concentration- Sample temperature- Diffusion current- Programming codes- Programming arguments

    Esc Enter

    18.6 C

    ppb 2.10

    Polymetron

    O2Disp2 Menu

    S1

    S2 S4

    S3

    Figure 4-1: Front panel

    Suppress an action orgo back to previous step

    Choice of the unit ornumeric value

    Oxygenapplication

    Validate and s witcho next step

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    28/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    29/92

    Oxyg en analyser (ppb) - 9182 -

    4-3

    4.3 Description of the function keys

    The function keys below have their signification highlighted at the bottom of the screen:

    Modify a parameter

    Select Scrolling in a list of menus

    Main Go back to the main display

    Menu Display the main menu

    Disp2 Display screen 2

    Disp3 Display screen 3

    Disp4 Display screen 4

    OK Validate the measure during the calibration

    Yes Confirm a command

    - Decrease a value

    + Increase a value

    4.4 Icons

    Symbol of waiting or instrument reset

    Alarm system for relay S3

    Timer symbol: countdown for relay S4

    P

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    30/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    31/92

    Oxyg en analyser (ppb) - 9182 -

    5-1

    Chapter 5: Programming the transmitter

    5.1 Main menu

    6.59ppb

    23.2 C O2Disp2 Menu

    S1

    S2

    S3

    S4

    CALIBRATIONMAINTENANCE

    PROGRAMMINGSERVICE

    MENU

    Select

    Choix

    F

    CONC. CALIB.TEMP. CALIB.PARAMETERS

    CALIBRATION

    Select

    54.90 ppb

    23.2 C8.05 A

    MAINTENANCE

    MEASUREALARMSmA OUTPUTS

    RS485

    PROGRAMMING

    Select

    AVERAGEDISPLAYCODE

    SOFT ISSUEDEFAULT VAL.ADJUST mACONFIGURATIONPOLYMETRON

    SERVICE

    Select

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    32/92

    Oxyg en analyser (ppb) - 9182 -

    5-2

    5.1.1 Calibration Menu

    + Any calibration should follow the procedure below:

    Configuration of the calibration characterics in the "PROGRAMMING" menu. Realization of the calibration via the EXECUTION menu.

    An access code may be required if it has been programmed (See 5.1.4.3CODE menu).See chapter 6 for a detailed programming of the calibrations.Some menus may appear or not according to how some parameters havebeen programmed.

    PARAMETERS

    DATE xx/xx/xx Date of the last calibration. The date programmed is notupdated automatically

    P x.xxx A/ppm Slope value

    T x.x C Drift between the theorical temperature (sensor response)Th and the measured temperature T m : T = T h - T m

    Choix

    F

    TEMP. CALIB.PARAMETERS

    CALIBRATION

    Select

    F

    DATE : 01/01/98

    P : 1.000 A/ppm

    T : 0.0 C

    PARAMETERS Choix

    FSLOPE

    CONC. CALIB.

    Select

    ZERO

    CONC. CALIB.

    Select

    TEMP. CALIB.

    EXECUTION

    Enter

    Enter

    Enter

    !

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    33/92

    Oxyg en analyser (ppb) - 9182 -

    5-3

    5.1.2 MAINTENANCE Menu

    Used for any maintenance operation in theinstrument.The transmitter continues to display thevariables measured.

    The relay status is not modified.The analogue output value depends on theconfiguration defined in the mA OUTPUTS/SPECIAL PROG./MAINTENANCE menu.

    5.1.3 PROGRAMMING Menu

    An access code may be required (See 5.1.4.3 CODE menu)!

    6.75 ppb

    21.6 C5.559 A

    MAINTENANCE

    ALARMSmA OUTPUTSRS485

    PROGRAMMING

    Select

    MEASURE

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    34/92

    Oxyg en analyser (ppb) - 9182 -

    5-4

    5.1.3.1 MEASURE Menu

    TEMPERATURE COMPENSATION

    TYPE - Auto- Manual

    Choice of a temperature measurement with automaticcompensation or with a manual compensation

    I f you have chosen a manu al temperature comp ensat ion, the TEMP.CA LI B . M enu i s no t acces s ib le anymor e

    TEMP. - xx.x C Possibility to enter the sample temperature in a manualcompensation mode

    SENSOR NTC/AD590 Type of temperature sensor used on the 9182 Oxygenanalyser. You have to program the NTC

    Choix

    MEASURE

    Select

    Choix

    F

    SENSOR : NTC

    TYPE : Manual

    TEMP. : 25.02 C

    TEMP. COMP.

    Select

    TEMP. COMP.

    Enter

    !

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    35/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    36/92

    Oxyg en analyser (ppb) - 9182 -

    5-6

    SYSTEM ALARM FUNCTION:

    The relay S3 may be used to indicate a faulty functioning of the analyser.

    It is required to connect the relay S3 to an external alarm system to control the faults traced by theanalyser.

    The relay S3 is activated as soon as a default appears.

    In case of a manual acknowledgment, the relay remains activated even if the default disappears. Itis required to press ENTER to desactivate the relay and the error message.

    In case of an automatic acknowledgment, the relay and the error message are desactivated whenthe default disappears.

    ALARM 3 (ALARM SYSTEM)

    MODE - No- Limit- Syst.

    The alarm S3 may be programmed as alimit function (see paragraph above) oras an alarm system function

    ACCEPT - Auto- Manu

    In the case of an alarm system, choicebetween a manual (key ENTER) or anautomatic acknowledgment

    RELAY - NO- NC

    Choice of S3 normally open or normallyclosed

    MODE :

    ACCEPT. : Auto

    RELAY : NC

    ALARM 3

    Select

    Syst.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    37/92

    Oxyg en analyser (ppb) - 9182 -

    5-7

    TIMER FUNCTION:

    Relay S4 may be affected to a timer function.

    ALARM 4 (TIMER)

    MODE - No- Limit- Timer

    Choice between a limit (seeparameters above) or a timer functionfor alarm 4

    INTERV XXXXmn Interval between 2 cycles (in minutes)

    IMPUL. X Number of pulses during an activecycle

    Ton XXXs Adjustment of the relay active time (inseconds) for each pulse

    Toff XXXs Adjustment of the relay inactive time(in seconds) for each pulse

    TmA XXmn Hold time for the analogue outputsafter each cycle T h e a n a l o g u e o u t p u t s t a t u s d e p e n d s o n t h e

    c o n f i g u r a t i o n o f t h e m e n u m A O U T P UT S /S P E CI A LPROG./TIMER

    Example of a timer operating cycle:

    Relay S4Pulse number

    Closed

    Open

    Ton Toff TmA

    Interval

    Time

    MODE :

    Interv : 1440 mn

    IMPUL. : 5

    Ton : 005 s

    Toff : 003 s

    TmA : 05 mn

    ALARM 4

    Select

    Timer

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    38/92

    Oxyg en analyser (ppb) - 9182 -

    5-8

    5.1.3.3 mA OUTPUTS menu

    The analogue output signals allow the transmission of the measurements from the analyser to anyexternal control system.

    It is highly recommended to use shielded cables for the output signals. This shielding should beconnected to the earth terminal on the armour plate.

    OUTPUT 2SPECIAL PROG.TEST

    mA OUTPUTS

    Select

    OUTPUT 1

    CALIBRATION

    SYST. ALARM.

    TIMER

    SPECIAL PROG.

    Select

    MAINTENANCE AFFECT. :TYPE : 4-20MODE : Lin

    LOW : 00001 nAUP : 100.0 A

    OUTPUTS 1/2

    Select

    Choix

    F

    MODE :

    VALUE : 00 mA

    SPECIAL PROG.

    Select

    Preset

    F

    VALUE : mA

    TEST

    00

    Enter

    Enter

    Enter

    Enter

    A

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    39/92

    Oxyg en analyser (ppb) - 9182 -

    5-9

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    40/92

    Oxyg en analyser (ppb) - 9182 -

    5-10

    OUTPUTS 1/2

    AFFECT - Conc- A- C/F

    Choice of the analogue output allocation to the cell current, tothe concentration or to the temperature measurement

    TYPE 0/204/20

    Choice of the analogue output type

    MODE - Lin- Dual

    Choice between a linear or dual range (see drawing below)

    LOWER XXXX Bottom of the scale value

    MIDD. XXXX Mid-scale value (only in dual mod e )

    UPPER XXXX Top of the scale value

    SPECIAL PROG.

    MODE - last- preset- live

    Characteristics of the analogue output during calibration, alarmsystem, maintenance or timer active cycles: frozen to the lateststored before any operation listed above, forced to a presetvalue, live measurement

    VALUE XX Preset value (0 to 21 mA)

    TEST Test the analogue outputs by steps of 1 mA (0...21 mA)

    Illustration of the linear/dual outputs scales:

    (%)

    Upper 20 20

    16 15

    Middle 12 10

    8 5

    Lower 4 0

    0 10 20 30 40 50 60 70 80 90 100

    (mA)

    Dual

    Linear

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    41/92

    Oxyg en analyser (ppb) - 9182 -

    5-11

    5.1.3.4 RS485 Menu

    If the RS485 optional board is installed in your transmitter, program the parameters of the menubelow.The RS485 optional board enables the connection between your analyser and a digitalcommunication system. The Communication protocol is JBUS/MODBUS. Refer to the instructionmanual JBUS/MODBUS communication (part number: 621=991=000) for further details and to the

    Appendix 7 for the address list.

    RS485

    N XX MONEC number (0.. .32)

    BAUD - 300- 600- 1200- 2400- 4800- 9600

    Transmission speed in baud

    PARITY - No- Odd- Even

    Without parity bitWith odd parity bitWith even parity bit

    BIT STOP - 1- 2

    1 bit stop2 bit stop

    N :

    BAUD : 9600

    PARITY : NoSTOP BIT : 1

    RS485

    Select

    00

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    42/92

    Oxyg en analyser (ppb) - 9182 -

    5-12

    5.1.4 SERVICE Menu

    An access code may be required if it has been programmed (see 5.1.4.3CODE Menu).

    5.1.4.1 AVERAGE Menu

    The measurement cycle lasts 4 seconds.

    AVERAGE Program a moving average on the concentration measurement

    AVERAGE X Define the number of measurements to calculate the average

    TEST Display the difference between a measurement obtained with andwithout average

    !

    AVERAGEDISPLAYCODESOFT ISSUE

    DEFAULT VAL.ADJUST mACONFIGURATIONPOLYMETRON

    SERVICE

    Select

    AVERAGE :TEST

    AVERAGE

    Select

    9 Average3.46 ppb

    Real3.49 ppb

    TEST

    Enter

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    43/92

    Oxyg en analyser (ppb) - 9182 -

    5-13

    5.1.4.2 DISPLAY Menu

    DISPLAY

    CONC - ppb/ppm- g-mg/l

    Choice of the concentration unit

    TEMP. - C- F

    Choice of the temperature unit

    PRESSURE - mmHg- mbar- InHg

    Choice of the atmospheric pressure unit

    LANGUAGE

    - F- GB- D- SP- I

    Choice of the language:

    - french- english- german- spanish- italian

    CONC :

    TEMP. : C

    PRESSURE : mmHg

    LANGUAGE : F

    DISPLAY

    Select

    ppb/ppm

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    44/92

    Oxyg en analyser (ppb) - 9182 -

    5-14

    5.1.4.3 CODE Menu

    Protection codes may be programmed to access to the PROGRAMMING, CALIBRATION, SERVICEmenus.

    This code may be desactivated by programming 0000.

    CODE

    CALIB. XXXX Access code to the calibrations

    PROG. XXXX Access code to the PROGRAMMING menu

    SERVICE XXXX Access code to the SERVICE menu

    If you forget your access code, press simultaneously ESC and ENTER to enter into themenu selected.

    CALIB. :

    PROG. : 0000

    SERVICE : 0000

    CODE

    Select

    0000

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    45/92

    Oxyg en analyser (ppb) - 9182 -

    5-15

    5.1.4.4 SOFT VERSION Menu

    This menu displays the software version installed in the instrument.

    5.1.4.5 DEFAULT VAL. Menu

    If you press YES, you load the defaultvalues and you loose the currentprogrammed values and calibrationparameters

    MONEC 9182

    OXY 1.00

    SOFT ISSUE

    !Loadingdefault

    values?

    DEFAULT VAL

    Yes

    !

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    46/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    47/92

    Oxyg en analyser (ppb) - 9182 -

    5-17

    5.1.4.7 CONFIGURATION Menu

    You have to program the operating frequency so it matches with the mains supply characteristics.When the frequency is changed, the instrument is automatically resetted.

    5.1.4.8 POLYMETRON Menu

    This menu is reserved to Hach Ultra Analytics qualifiedpersonnel.

    CODE :

    POLYMETRON

    0000

    FREQ :

    CONFIGURATION

    60 Hz

    Select

    FREQ :

    CONFIGURATION

    50 Hz

    Disp2

    Please

    wait...

    Disp2 Menu

    Start up ...

    Disp2 Menu

    Enter

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    48/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    49/92

    Oxygen analyser (ppb) - 9182 -

    6-1

    Chapter 6: Calibrating the instrument

    NOTESee chapter 5 for programming the commands.

    REMARK

    Any result (calibration or measurement) is always brought back to the reference temperature(25C, 77F). If the sample temperature is different from the reference temperature, it is requiredto execute a temperature compensation which can be either manual or automatic.

    6.1 Calibration of the temperature sensor

    The temperature sensor is located in the dissolved oxygen measurement probe. It is factory-preadjusted but needs to be calibrated in the sample on site. This calibration must be realisedbefore the oxygen measurement calibration (slope + zero).

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    50/92

    Oxygen analyser (ppb) - 9182 -

    6-2

    6.1.1 Automatic temperature compensation

    The sensor measures continuously the sample temperature. The concentration values areautomatically calculated according to the reference temperature (25 C) by a preprogrammedcompensation law in the transmitter.

    Follow the procedure below:

    PROGRAMMING

    Choice of an automatic temperaturecompensation.

    EXECUTION

    The execution is realised with theCALIBRATION menu.

    Enter the sample temperature value in C.Press OK to adjust the temperature displayedto the actual value of the sample temperaturemeasured with a precise thermometer.

    The gap between the calibration and thetheorical response curve of the sensor is givenfor information.

    MEASURE

    Select

    TEMP. COMP.

    TYPE :

    SENSOR : NTC

    TEMP : 25.0C

    TEMP. COMP.

    Select

    Enter

    Auto.

    TEMP. CALIB.

    Select

    EXECUTION

    21.2C

    6.87 ppmOK

    CAL

    21.2C

    1.2 CCAL

    2

    T : 0.0 C

    ENTER : ok

    ESC : cancel

    TEMP. CALIB.

    Enter

    Enter

    Press OK when themeasure is stable

    Possibility to changethe value

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    51/92

    Oxygen analyser (ppb) - 9182 -

    6-3

    6.1.2 Manual temperature compensation

    This type of temperature compensation should be used only if your sample temperature isconstant.

    PROGRAMMING

    Choice of a manual temperature compensation.

    Enter your sample temperature.

    EXECUTION

    Not applicable when a manual temperaturecompensation is chosen.

    MEASURE

    Select

    TEMP. COMP.

    TYPE :

    SENSOR : NTC

    TEMP. COMP.

    Select

    Enter

    TYPE :

    TEMP. : 25.0 C

    SENSOR : NTC

    TEMP. COMP.

    Select

    Enter

    Manual

    Manual

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    52/92

    Oxygen analyser (ppb) - 9182 -

    6-4

    6.2 Calibration of the measurement

    The slope calibration can be done in the air or in water, with electrical zero or chemical zero.

    6.2.1 Slope calibration in the air + electrical zero

    PROGRAMMING

    ZERO: Choice of an electrical zero. The electricalzero is carried out automatically by thetransmitter at a regular frequency. Choice ofthe slope in the air.

    SLOPE: Enter the atmospheric pressure in mmHg.Enter the ambient temperature in C.

    EXECUTION

    SLOPE: Humidify the wadding of the calibration cap andtake off the probe from the sample. Position thecap on the electrode and press ENTER.The probe should be positioned vertically withthe membrane downwards (see page 7.1).

    The flashing "cal" message indicates that theinstrument is in calibration mode. Wait for thecurrent stabilisation (approximately 10 min.)and press OK to validate the calibration. Theoxygen concentration value flashes during3 seconds.

    The analyser displays the date of the lastcalibration and the new slope calculated.Modify the date if you wish. Put back theelectrode in water.

    TYPE :

    P = 760.0 mmHg

    Tamb = 20 C

    SLOPE CALIB.

    Select

    AIR

    Enter

    SLOPE

    CONC. CALIB.

    Select

    ZERO

    ZERO CALIB.

    TYPE :

    ZERO CALIB.

    Select

    EXECUTION

    SLOPE CALIB.

    EnterEnter

    AutoElec.

    PROGRAMMING

    Select Select

    PROGRAMMING

    Enter

    PROGRAMMING

    SLOPE CALIB.

    Select

    EXECUTION

    DATE :P : 1 nA/ppm

    SLOPE CALIB.

    Enter

    01/01/98

    Possibility to enter thedate of the calibration

    21.2ppm

    25.0 C5.693 A OK

    CAL

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    53/92

    Oxygen analyser (ppb) - 9182 -

    6-5

    6.2.2 Slope calibration in the air + chemical zero

    PROGRAMMING

    ZERO: Choice of the chemical zero.Choice of the slope in the air.

    SLOPE: Enter the atmospheric pressure in mmHg.Enter the value of the ambient temperature.

    EXECUTION

    ZERO: Make sure the water used for the chemicalzero is really oxygen-free (< 1 ppb).Press ENTER, the "CAL" message flashesand indicates the instrument is in calibratingmode.Wait for the current stabilisation(approximately 10 min.) and press OK tovalidate the calibration. The oxygenconcentration value flashes during 3 seconds.The instrument displays zero.

    SLOPE: Humidify the wadding of the calibration cap andtake off the probe from the sample. Position thecap on the electrode and press ENTER.The probe should be positioned vertically withthe membrane downwards (see page 7.1).The flashing "cal" message indicates that theinstrument is in calibration mode. Wait for thecurrent stabilisation (approximately 10 min.)and press OK to validate the calibration. Theoxygen concentration value flashes during3 seconds.The analyser displays the date of the lastcalibration and the new slope calculated.Modify the date if you wish. Put back the

    electrode in water.

    PROGRAMMING

    ZERO CALIB.

    Select

    EXECUTION

    Enter

    21.2ppm

    25.0 C5.693 A

    OK

    CAL

    0ppm

    25.0 C5.693 A

    OK

    CAL

    TYPE :

    P = 760.0 mmHgTamb = 20 C

    SLOPE CALIB.

    Select

    AIR

    Enter

    SLOPE

    CONC. CALIB.

    Select

    ZERO

    ZERO CALIB.

    TYPE :

    ZERO CALIB.

    Select

    EXECUTION

    SLOPE CALIB.

    EnterEnter

    Chem.

    PROGRAMMING

    Select Select

    PROGRAMMING

    Enter

    PROGRAMMING

    SLOPE CALIB.

    Select

    EXECUTION

    DATE :P : 1 nA/ppm

    SLOPE CALIB.

    Enter

    01/01/98

    Possibility to enter thedate of the calibration

    21.2ppm

    25.0 C5.693 A OK

    CAL

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    54/92

    Oxygen analyser (ppb) - 9182 -

    6-6

    6.2.3 Slope process calibration + electric zero

    PROGRAMMING

    ZERO: Choice of the electrical zero. The electricalzero is carried out automatically by thetransmitter at a regular frequency.Choice of a process slope.

    EXECUTION

    SLOPE: The "CAL" message flashes and indicates theinstrument is in calibrating mode. Wait for thecurrent stabilisation and press OK.You have the possibility to adjust theconcentration value digit by digit with thefunction key to the value obtainedwith your reference analysis method.

    The analyser displays the date of the lastcalibration and the new slope calculated.Modify the date if you wish.

    PROGRAMMING

    SLOPE CALIB.

    Select

    EXECUTION

    DATE :P : 1 nA/ppm

    SLOPE CALIB.

    Enter

    01/01/98

    Possibility to enter thecalibration date

    21.2ppm

    25.0 C5.693 A OK

    CAL

    TYPE :

    SLOPE CALIB.

    Select

    PROCESS

    Enter

    SLOPE

    CONC. CALIB.

    Select

    ZERO

    ZERO CALIB.

    TYPE :

    ZERO CALIB.

    Select

    EXECUTION

    SLOPE CALIB.

    Enter

    Enter

    AutoElec.

    PROGRAMMING

    Select Select

    PROGRAMMING

    Enter

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    55/92

    Oxygen analyser (ppb) - 9182 -

    6-7

    6.2.4 Slope process calibration + chemical zero

    PROGRAMMING

    ZERO: Choice of the chemical zeroChoice of the slope process

    EXECUTION

    ZERO: Make sure the water used for the chemicalzero is really oxygen-free (< 1 ppb).Press ENTER, the "CAL" message flashesand indicates the instrument is in calibratingmode.Wait for the current stabilisation(approximately 10 min.) and press OK tovalidate the calibration. The oxygenconcentration value flashes during 3 seconds.The instrument displays zero.

    SLOPE: The flashing "cal" message indicates that the

    instrument is in calibration mode. Wait for thecurrent stabilisation (approximately 10 min.)and press OK to validate the calibration. Theoxygen concentration value flashes during 3seconds.The analyser displays the date of the lastcalibration and the new slope calculated. Modifythe date if you wish.

    PROGRAMMING

    ZERO CALIB.

    Select

    EXECUTION

    Enter

    21.2ppm25.0 C5.693 A

    OK

    CAL

    0ppm

    25.0 C5.693 A

    OK

    CAL

    PROGRAMMING

    SLOPE CALIB.

    Select

    EXECUTION

    DATE :P : 1 nA/ppm

    SLOPE CALIB.

    Enter

    01/01/98

    Possibility to enter thecalibration date

    21.2ppm

    25.0 C5.693 A OK

    CAL

    TYPE :

    SLOPE CALIB.

    Select

    PROCESS

    Enter

    SLOPE

    CONC. CALIB.

    Select

    ZERO

    ZERO CALIB.

    TYPE :

    ZERO CALIB.

    Select

    EXECUTION

    SLOPE CALIB.

    EnterEnter

    Chem.

    PROGRAMMING

    Select Select

    PROGRAMMING

    Enter

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    56/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    57/92

    Oxygen analyser (ppb) - 9182 -

    7-1

    Chapter 7: Start up, maintenance and troubleshooting

    7.1 Start up

    7.1.1 Assembling the probe

    The probe is composed of the following items:

    1 : Nut

    2 : Measuring electrode

    3 : Probe body

    4 : Membrane

    5 : Filling screw

    6 : Gasket

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    58/92

    Oxyg en analyser (ppb) - 9182 -

    7-2

    Proceed as follows for the assembling:

    A. Screw the membrane on the body till the dead stop .

    B. Fill in the probe body with 5 ml of electrolyte by means of the syringe.

    Check there is no impurity or bubble in the electrolyte .

    C. Insert the electrode in the probe body until you meet a resistance. Do not force.

    D. Complete the mounting by tightening the nut.

    E. Mount the filling screw and its gasket. Tight it sufficiently to avoid any leaking of electrolyte orexternal pollution from sample.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    59/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    60/92

    Oxyg en analyser (ppb) - 9182 -

    7-4

    7.2 Changing the membrane

    Never take off the electrode before removing the electrolyte filling plug.

    Never touch the membrane with the hands.

    Never use a worn membrane.

    Turn off the water supply. Remove theconnector from the probe.

    Unscrew the cell cap and remove theprobe.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    61/92

    Oxyg en analyser (ppb) - 9182 -

    7-5

    7.2.1 Dismounting the membrane

    A. Unscrew cap and gasket.

    B. Unscrew nut.

    C. Remove electrode without scratching the cathode.

    D. Empty the electrolyte left in the probe.

    E. Unscrew the worn membrane.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    62/92

    Oxyg en analyser (ppb) - 9182 -

    7-6

    7.2.2 Mounting the new membrane

    F. Screw the new membrane on the body till the dead stop.

    G. Replenish the probe body with 5 ml of electrolyte. Check there is no impurity or bubble in theelectrolyte.

    H. Preposition the electrode without forcing. It should take its place by simple gravity.

    I. Screw the probe nut till the electrode meets a resistance.

    J. Put back the cap and gasket, screw the cap to ensure watertightness.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    63/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    64/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    65/92

    Oxyg en analyser (ppb) - 9182 -

    7-9

    e) Cable or connections damaged when the probe has been replaced! Check the connection to the transmitter. If the connection is correct check the

    connections of the probe connector (See appendix 4 for the connectorconnections).

    f) The membrane is worn! Change the membrane.

    g) There is a AgBr (dark green) deposit on the silver! Polish the silver tube with soft abrasive (N 400 to 600)

    Problem: Imp ortan t ins tab i li ty in measuring mod e

    Causes and solutions:

    a) There is humidity or water in the probe connector! Dry the connector and check it is correctly screwed.

    b) Incorrect connexion! Check the connections of the transmitter to the probe.

    c) There are bubbles close to the cathod! Replenish the electroly te holder and check there is no bubble between the cathod

    and the membrane.

    d) The probe is violently shaked! Check the fixation and the stability.

    e) Important electromagnetic interferences close to the probe or transmitter cable! Find a better place for the cable and check the EMC level.

    f) Temporary interference with other gas! Mainly with H 2S

    g) Flowrate too weak (50 ml/h minimum)! Change the probe location

    h) Mud from an heterogen sample damages the membrane! Install a deflector or change the probe location

    Pro b l em: Lack o f accu racy

    Causes and solutions:

    a) The membrane permeability has changed (dirt deposit)! Calibrate the analyser and check if the concentration is back to normal.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    66/92

    Oxyg en analyser (ppb) - 9182 -

    7-10

    b) Pollution of the electrolyte! Check the parts to screw (membrane, filling screw) and change the electrolyte.

    c) Leak of electrolyte! Check the parts to screw (membrane, filling screw) and change the electrolyte.

    d) Interferences, mainly with H 2S! If the level of H 2S (or other pollutant) is stable, take into account its concentration

    to determine the O 2 concentation.

    e) Error during calibration or incorrect calibration! Calibrate again to check the parameters. If the error is confirmed, check the

    calibration current (too high, too low or unstable) and the concentration in the air.Refer to the problems described above.

    f) The temperature calibration has not been done! Check the temperature given by the transmitter and calibrate it (see page 6-2). If

    you have chosen the manual compensation check the value is correct.

    g) There are bubbles close to the cathod! Replenish the electrolyte holder and check there is no bubble at the bottom.

    h) Flowrate too low (4 l/h minimum)! Increase the flowrate.

    i) The sample temperature or pressure is out of the specifications! Change the probe location or modify the sample so that it meets the specifications.

    j) Mud or particles on the golden cathod! Clean the golden cathod with an absorbant tissue and rinse the membrane.

    k) The electrolyse potential is not equal to -0.85 V! Enter the correct value.

    Ot h e r p ro b l ems

    a) The probe current is null

    ! There is no electrolyte in the probe (leak).

    b) The probe current is negative! Connexion problem to the anode circuit (loose contact). ! AgBr deposit on the anode

    c) The sample temperature corresponds to the specifications but there is an error on thetemperature! There may be a short-circuit on the temperature connection.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    67/92

    Oxyg en analyser (ppb) - 9182 -

    7-11

    7.5 Electrical troubleshooting

    WARNING!

    Never attempt servicing before disconnecting the instrument from the main powerline

    MALFUNCTION POSSIBLE CAUSE REMEDIES

    No power; instrument is notconnected correctly

    Check for power, then check ifconnected properly

    Defective fuse Check fuse

    Instrument's power supply setfor wrong line voltage

    Check jumpers on power-supply board for correctvoltage settings

    Ribbon cable connectingpower with CPU board notproperly plugged in

    Check that the ribbon plugsare properly plugged in

    Connection between CPUboard and measurementmodule loose

    Check plug connections

    Short circuit in power-supplyboard

    Visually check power-supplyboard for shorts

    No indication

    Hardware is defective Call the Service Technician

    LCD displays undefinedcharacters

    Malfunctioning CPU board orprocessor

    Using the Instruction manual,program the instrument to loadthe default values

    CPU hardware RESET the instrument bytemporarily interrupting thepower (5-10 secs.)

    Call the Service Technician

    Keyboard does notoperate; all keys areinactive

    CPU malfunctioning, externalinterferences

    If there is no response, RESETthe instrument by temporarilyinterrupting the power (5-10seconds) Check each keyagain. If there is no change,call the Service Technician

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    68/92

    Oxyg en analyser (ppb) - 9182 -

    7-12

    MALFUNCTION POSSIBLE CAUSE REMEDIES

    Measurement is notcorrect

    Instrument was programmedincorrectly

    Recheck programmedparameters. Do they agree withthe probe's characteristics?

    System, including probe, notcalibrated correctly

    Calibrate the whole system(probe connected)

    Probe connected wrong Recheck all probe connections

    Probe malfunctioning,possibly incompatible with theapplication

    Visually check the condition ofthe probe. Is the applicationwithin the probe'sspecifications (data sheet)?

    CPU board is defective If error persists, call the

    Service TechnicianMeasurement is notstable

    Faulty probe Check the condition of theprobe. Is it contaminated?

    Probe connected wrong Is the probe connectedcorrectly?

    Interferences Are there any sources ofpotential interference,chemical, external,temperature, pressure, ...?

    Cable shield is not connected Check and connect

    Defective CPU board If problem persists, call theService Technician

    Probe connected wrong Is the probe properlyconnected? Check

    Temperaturemeasurement is notcorrect

    Temperature was notcalibrated

    Calibrate for temperature. Alsocheck the Pt100 for correctresistance (ohmmeter)

    CPU board is defective If problem persists, call theService Technician

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    69/92

    Oxyg en analyser (ppb) - 9182 -

    7-13

    MALFUNCTION POSSIBLE CAUSE REMEDIES

    Check to make sure that probeis connected correctlyInitiate a software RESET

    If steps 1-3 do not remedy theproblem, make a cold start(RESET): interrupt the powerfor 5-10 secs

    Reprogram the instrumentagain

    Display reading static;cannot be changed inany way

    Malfunctioning CPU boardand/or some other transmitterhardware is defective

    If problem persists, call theService Technician

    Relays not energised Instrument was programmedincorrectly

    Check whether the correctrelay parameters and setpointshave been programmed

    Hardware is defective Check that the programmedsetpoints are compatible withthe programmed measuringrange

    Check the relay characteristicsfor proper functioning using anohmmeter

    If problem persists, contact theService Technician

    Instrument was incorrectlyprogrammed

    Check the programmed output-current parameters

    Connection of the Monec withperipherals (recorder, etc.)are faulty, loose or defective

    Check the cables

    Compare the measured valuewith the output current range

    Wrong output current,output current remains at0 or 20 mA

    Hardware is defective

    If problem persists, call the

    Service TechnicianWrong configuration(3 electrode mode)

    Check the swiches under theamperometric board arecorrectly positioned ("on")

    Polarization voltageincorrect

    Wrong programming

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    70/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    71/92

    Oxyg en analyser (ppb) - 9182 -

    8-1

    Chapter 8: Error messages

    In case of errors, measurements are replaced by dashes " - - -" .

    Error messages Description

    Error messages during a measurement

    CONCENTRATIONTOO HIGH

    The concentration value is out of the limits. Check thecurrent value and the calibration parameters.

    STAND-BY1.99 ppb

    Electrode protection In case of measuring range exceeding "concentrationtoo hugh" over 2 minutes, the error message "STAND-BY" appears. From this moment on, the electrodepolarization voltage is switched off during 10 minutes.

    After these 10 minutes, the voltage is switched on.In STAND-BY the user m ay swi tch on the vo l tage atany t ime by p ressing ENTER.

    CONCENTRATIONTOO LOW

    The concentration value is out of the limits. Check thecurrent value and the calibration parameters.

    TEMPERATUREERROR

    The sample temperature is out of the limits. Check thecable polarity. Check if there is any short-circuit or opencircuit.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    72/92

    Oxyg en analyser (ppb) - 9182 -

    8-2

    CURRENTTOO LOW

    The current value is out of the limits. Check theelectrode (electrolyte and membrane).

    CURRENTTOO HIGH

    The current value is out of the limits. Check there is noshort-circuit on the measuring line.Check the polarization voltage.

    Error messages during a calibration

    TOUT OF LIMITS

    21.6C

    The temperature difference between the calibration andthe sensor theorical response is superior to theprogrammed limit.Limits: 20 C

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    73/92

    Oxyg en analyser (ppb) - 9182 -

    A2-1

    Appendix 1: Pressure conversion table

    hPa or mbars mm Hg mm H 2O psi inches Hg inches H 2O

    920.00 690 9503.60 13.34 27.16 374.16

    925.00 693.8 9555.25 13.42 27.31 376.19

    930.00 697.5 9606.90 13.49 27.46 378.22

    935.00 701.3 9658.55 13.56 27.61 380.26

    940.00 705 9710.20 13.63 27.76 382.29

    945.00 708.8 9761.85 13.71 27.90 384.32

    950.00 712.5 9813.50 13.78 28.05 386.36

    955.00 716.3 9865.15 13.85 28.20 388.39

    960.00 720 9916.80 13.92 28.35 390.43

    965.00 723.8 9968.45 14.00 28.50 392.46

    970.00 727.5 10020.10 14.07 28.64 394.49

    975.00 731.3 10071.75 11.14 28.79 396.53

    980.00 735 10123.40 14.21 28.93 398.56

    985.00 738.8 10175.05 14.29 29.08 400.59

    990.00 742.5 10226.70 14.36 29.23 402.63

    995.00 746.3 10278.35 14.43 29.38 404.66

    1000.00 750 10330.00 14.50 29.53 406.69

    1005.00 753.8 10381.65 14.58 29.68 408.73

    1010.00 757.5 10433.30 14.65 29.82 410.76

    1015.00 761.3 10484.95 14.72 29.97 412.79

    1020.00 765 10536.60 14.79 30.12 414.83

    1025.00 768.8 10588.25 14.87 30.27 416.86

    1030.00 772.5 10639.90 14.94 30.42 418.89

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    74/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    75/92

    Oxyg en analyser (ppb) - 9182 -

    A2-1

    Appendix 2: Temperature conversion table

    3 Conversion from C into F: F = 1,8 C + 323 Conversion from C into K: K = C + 273,15

    C F K C F K

    0 32 273,15 24 75,2 297,15

    1 33,8 274,15 25 77 298,15

    2 35,6 275,15 26 78,8 299,15

    3 37,4 276,15 27 80,6 300,15

    4 39,2 277,15 28 82,4 301,15

    5 41 278,15 29 84,2 302,15

    6 42,8 279,15 30 86 303,15

    7 44,6 280,15 31 87,8 304,15

    8 46,4 281,15 32 89,6 305,15

    9 48,2 282,15 33 91,4 306,15

    10 50 283,15 34 93,2 307,1511 51,8 284,15 35 95 308,15

    12 53,6 285,15 36 96,8 309,15

    13 55,4 286,15 37 98,6 310,15

    14 57,2 287,15 38 100,4 311,15

    15 59 288,15 39 102,2 312,15

    16 60,8 289,15 40 104 313,15

    17 62,6 290,15 41 105,8 314,15

    18 64,4 291,15 42 107,6 315,1519 66,2 292,15 43 109,4 316,15

    20 68 293,15 44 111,2 317,15

    21 69,8 294,15 45 113 318,15

    22 71,6 295,15

    23 73,4 296,15

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    76/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    77/92

    Oxyg en analyser (ppb) - 9182 -

    A3-1

    Appendix 3: Default values

    CALIBRATION

    CONC. CALIB.OFFSETType : AutoElecSLOPEType : AirP : 760 mmHgTamb. : 20.0 C

    PARAMETERSDATE : 01/01/98S : 1.000 A/ppmT : 0.0 C

    PROGRAMMING

    MEASURE

    TEMP. COMP.TYPE : AutoSENSOR : NTC

    ALARMS

    ALARMS S1/S2/S4 AFFECT. : Conc.LIM. : 0.00 ppbDIR. : Low

    DELAY : 000 sHYST. : 00%RELAY : NO

    ALARM S3 AFFECT. : System ACQUIT : AutoRELAY : NC

    m A OUTPUTS

    OUTPUT 1 AFFECT. : Conc.TYPE : 4-20MODE : Lin.LOW : 0.000 ppmUP : 1.000 ppm

    OUTPUT 2 AFFECT. : Conc.TYPE : 4-20MODE : Lin.LOW : 0.000 ppmUP : 1.000 ppm

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    78/92

    Oxyg en analyser (ppb) - 9182 -

    A3-2

    SPECIAL PROG.

    MAINTENANCEMODE : Last

    TIMERMODE : Last

    CALIBRATIONMODE : Last

    ALARM SYSTEMMODE : Last

    RS485

    N : 0BAUD : 9600PARITY : NoSTOP BIT : 1

    SERVICE

    AVERAGE

    AVERAGE : 1

    DISPLAY

    DISPLAYCONC. : ppb/ppmTEMP. : CPRESSURE : mmHgLANGUAGE : GB

    CODE

    CODECALIB. : 0000PROG. : 0000SERVICE : 0000

    CONFIGURATION

    CONFIGURATIONFREQ. : 50 Hz

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    79/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    80/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    81/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    82/92

    Oxyg en analyser (ppb) - 9182 -

    A5-2

    FIRST AID MEASURES

    In case of contact, immediately flush eyes or skin with copious amount s of water for at least 15 minuteswhile removing contaminated clothing and shoes. Assure adequate flushing of the eyes by separating theeyelids with fingers. If inhaled, remove to fresh air. If not breathing give artificial respiration. If breathing isdifficult give oxygen. If swallowed, wash out mouth with water provided person is conscious. Call aphysician immediately. Discard contaminated clothing and shoes.

    FIRE FIGHTING MEASURES

    Non combustible. Use extinguishing media appropriate to surrounding fire conditions. Do not use water.Wear self-contained breathing apparatus and protective clothing to prevent contact with skin and eyes.Reacts violently with water. Emits toxic fumes under fire conditions.

    ACCIDENTAL RELEASE MEASURES

    Evacuate area. Wear self-contained breathing apparatus, rubber boots and heavy rubber gloves. Sweepup, place in a bag and hold for waste disposal. Avoid raising dust. Ventilate area and wash spill site aftermaterial pickup is complete.

    HANDLING AND STORAGE

    Danger: under fire conditions material will melt and flow. Molten material reacts violently with water andcan react with aluminium, tin, zinc and their alloys to generate flammable and explosive hydrogen gas.Potassium hydroxide reacts with trichloroethylene to form s pontaneously flammable dichloroacetylene.

    EXPOSURE CONTROLS/PERSONAL PROTECTION

    Wear appropriate NIOSH/MSHA-approved respirator, chemical-resistant gloves, safety goggles, otherprotective clothing. Safety shower and eye bath. Use only in a chemical fume hood. Faceshield (8-inchminimum). Do not breathe dust. Do not get in eyes, on skin, on clothing. Avoid prolonged or repeatedexposure. Wash thoroughly after handling. Corrosive. Toxic. Keep tightly closed. Do not allow contactwith water. Store in a cool dry place.

    TOXICOLOGICAL INFORMATION

    Acute effects: harmful if swallowed, inhaled or absorbed through skin. Material is extremely destructive totissue of the mucous membranes and upper respiratory tract, eyes and skin. Inhalation may be fatal as aresult of spasm, inflammation and edema of the larynx and bronchi, chemical pneumonitis and pulmonary

    edema. Symptoms of exposure may include burning sensation, coughing, wheezing, laryngitis, shortnessof breath, headache, nausea and vomiting.

    DISPOSAL CONSIDERATIONS

    Contact a licensed professional waste disposal service to dispose of this material.

    The above information is believed to be correct but does not purport to be all inclusive and shall be used onlyas a guide.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    83/92

    Oxyg en analyser (ppb) - 9182 -

    A5-3

    Potassium bromide

    PHYSICAL AND CHEMICAL PROPERTIES

    White crystals

    Melting point 770 C

    Specific gravity 1.984

    LABEL PRECAUTIONARY STATEMENTS

    Harmful. Harmful by inhalation, in contact with skin and if swallowed. Irritating to eyes, respiratorysystem and skin. In case of contact with eyes, rinse immediately with plenty of water and seek medicaladvice. Wear suitable protective clothing.

    STABILITY AND REACTIVITY

    Incompatibilities Strong oxidizing agentsStrong acidsHazardous combustion or decompositionproducts: hydrogen chloride gas.

    FIRST AID MEASURES

    In case of contact, immediately flush eyes with copious amounts of water for at least 15 minutes. In caseof contact, immediately wash skin with soap and copious amounts of water. If inhaled, remove to freshair. If not breathing give artificial respiration. If breathing is difficult, give oxygen. If swallowed, wash outmouth with water provided person is conscious. Call a physician. Wash contaminated clothing beforereuse.

    FIRE FIGHTING MEASURES

    Extinguishing media Water spray.

    Special firefighting procedures Wear self-contained breathing apparatus andprotective clothing to unusual fire and explosionshazards.

    Unusual fire and explosions hazards Emits toxic fumes under fire conditions.

    ACCIDENTAL RELEASE MEASURES

    Wear self-contained breathing apparatus, rubber boots and heavy rubber gloves. Sweep up, place in abag and hold for waste disposal. Avoid raising dust. Ventilate area and wash spill site after materialpickup is complete.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    84/92

    Oxyg en analyser (ppb) - 9182 -

    A5-4

    PROCEDURES FOR CHECKING WORKER EXPOSURE

    Chemical safety goggles, rubber gloves, respirator mask approved by NIOSH/MSHA, safety showers andeye fountains, mechanical exhaust required. Avoid any contact and inhalation. Avoid any contact with theeyes, the skin and clothing.

    Wash well after handling. Noxious irritant solid. Keep the recipient tightly closed. Water absorbent. Storein a cool dry place.

    TOXICOLOGICAL INFORMATION

    Acute effects: harmful if swallowed, inhaled or absorbed through skin. Causes eye and skin irritation.Material is irritating to mucous membranes and upper respiratory tract. To the best of our knowledge, thechemical, physical, and toxicological properties have not been thoroughly investigated. Ingestion of large

    quantities can cause weakness, gastrointestinal irritation and circulatory disturbances.

    DISPOSAL CONSIDERATIONS

    For small quantities: cautiously add to a large stirred excess of water. Adjust the pH to neutral, separateany insoluble solids or liquids and package them for hazardous-waste disposal. Flush the aqueoussolution down the drain with plenty of water. The hydrolysis and neutralization reactions may generateheat and fumes which can be controlled by the rate of addition. Observe all federal, state and localenvironmental regulations.

    The above information is believed to be correct but does not purport to be all inclusive and shall be used onlyas a guide.

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    85/92

    Oxygen analyser (ppb) - 9182 -

    A6-1

    Appendix 6: Spare parts list

    Any other spare parts except those below in the table should be replaced in theinstrument.

    Description Part number

    RS485 board + manual 09125=A=2485

    Box of 4 pre-mounted membranes for 9182 and 9185 sensors 09185=A=3500

    Reference electrolyte 100 ml 363144,00100

    Oxygen electrode without probe body ppb 09182=A=1000

    Calibration cap 09182=A=1200

    Eprom for 9180 Oxygen 09180=A=6201

    Connection cable 10 m 09078=A=3010

    Probe body oxygen ppb 09078=C=1010

    Electrolyte filling screw 09078=C=1020

    Electrolyte filling washer 09078=C=1030

    Stainless steel flow cell 09078=A=2000

    Syringe 460150,21951

    CPU board 09125=A=1001

    Power supply 09125=A=2000

    Analog PC board for 9180 09180=A=1500

    Display board 09125=A=1010

    Monec D 9180 (standard) with software 09180=A=0300

    Monec D 9180 (RS485) with software 09180=A=0311

    9182 Instruction manual in English 621=191=082

    9182 Instruction manual in French 621=091=082

    9182 Instruction manual in German 621=291=082

    9182 Instruction manual in Italian 621=491=082

    9182 Instruction manual in Spanish 621=591=082

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    86/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    87/92

    Oxyg en analyser (ppb) - 9182 -

    A7-1

    Appendix 7: RS485 MODBUS-JBUS addressing

    CALIBRATION menu

    /CONC. CALIB. /ZERO

    /PROGRAMMING /Type

    /EXECUTION /SLOPE

    /PROGRAMMING /Type /P

    /Tamb /EXECUTION /TEMP. CALIB. /EXECUTION /PARAMETERS

    (0: ElecAuto, 1:Chemical)

    (0:Process, 1:Air)

    0121

    02210222

    0223

    MEASURE menu

    /MEASURE /TEMP.COMP.

    /Type /Temp.

    (0:Manual, 1:Auto) 12201230

    ALARMS menu

    /A L A RM 1 /Affect /Lim. /Dir. /Delay /Hyst. /Relay

    (0:conc, 1:C/F, 2:No)

    (0:Low, 1:Up)

    (0:N.O., 1:N.C.)

    212021302140215021602170

    /A L A RM 2 /Affect /Lim.

    /Dir. /Delay /Hyst. /Relay

    (0:conc, 1:C/F, 2:No)

    (0:Low, 1:Up)

    (0:N.O., 1:N.C.)

    22202230

    2240225022602270

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    88/92

    Oxyg en analyser (ppb) - 9182 -

    A7-2

    /A L A RM 3 /Mode

    /Affect /Lim. /Dir. /Delay /Hyst. /Relay /Accept

    (0:Limit, 1:Syst, 2:No)

    (0:conc, 1:C/F, 2:No)

    (0:Low, 1:Up)

    (0:N.O., 1:N.C.)

    2310

    2320233023402350236023702380

    /A L A RM 4 /Mode /Affect /Lim. /Dir. /Delay /Hyst. /Relay /Interv /Impul. /Ton /Toff /TmA

    (0:Manu, 1:Auto)

    (0:Limit, 1:Timer, 2:No)(0:conc, 1:C/F, 2:Non)

    (0:Low, 1:Up)

    (0;N.O., 1:N.C.)

    241024202430244024502460247024012402240324042405

    mA OUTPUTS menu

    /OUT PU T1

    /Affect /Type /Mode /Low /Mid. /Upp.

    (0:A, 1:C/F, 2:conc)(0:0/20mA, 1:4/20mA)(0:lin, 1:dual)

    411041204150413041604140

    /OUT PU T2 /Affect /Type /Mode /Low /Mid. /Upp.

    (0:A, 1:C/F, 2:conc)(0:0/20mA, 1:4/20mA)(0:lin, 1:dual)

    421042204250423042604240

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    89/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    90/92

    Oxyg en analyser (ppb) - 9182 -

    A7-4

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    91/92

  • 8/10/2019 Dissolved Oxygen Analyzer User Manual

    92/92